BUOYANCY EFFECT OF PURGE GAS IN THE DETERMINATION OF CARBON BLACK CONTENT IN POLYETHYLENE

Similar documents
ISO INTERNATIONAL STANDARD

Operation of the Perkin Elmer TGA-GC/MS

A SYSTEM DYNAMIC MODEL FOR DECISION SUPPORT SYSTEM IN LEAN CONSTRUCTION ALI CHEGENI

DSC 204 HP Phoenix. High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications

The Polyethylene Casing as Diffusion Barrier for Polyurethane Insulated District Heating Pipes

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

JABATAN KIMIA Kategori Arahan Kerja

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

The Effect of Air Column in Transport Canisters on Measured Gas Contents

ANALYSIS OF STROKE DISTRIBUTION BETWEEN PROFESSIONAL, INTERMEDIATE AND NOVICE SQUASH PLAYERS DIYANA ZULAIKA BINTI ABDUL GHANI

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007

Procedures for operation of the TA Instruments DSC

CFD SIMULATIONS OF GAS DISPERSION IN VENTILATED ROOMS

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

INFLUENCE OF INITIAL PRESSURE ON THE IGNITION AND COMBUSTION OF A PROPELLING CHARGE. Patrice Franco

Long Term Strength And It s Consequences for PEX

Gun Vacuum Performance Measurements February 2003

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Traceable calibration of automatic weighing instruments in dynamic operation

TECH TIP 14. Minimum Surface Area Measurements with Micromeritics Physisorption Analyzers

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter

ValidatingWindProfileEquationsduringTropicalStormDebbyin2012

Exercises The Atmosphere (page 383) 20.2 Atmospheric Pressure (pages )

The Effects of Inflation Gas on Tire Laboratory Test Performance

Retention Time Locking: Concepts and Applications. Application

VALIDATING AND DEVELOPING A NEW AGILITY TEST FOR KARATE

Effect of Fluid Density and Temperature on Discharge Coefficient of Ogee Spillways Using Physical Models

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Sample Solution for Problem 1.a

TI 176B Flow Rate Audit Calculations

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

STANDARD SCORES AND THE NORMAL DISTRIBUTION

Elimination of Oil Residual inside the Copper Pipe Using Ladder Technique

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Recommendations on Checking Anaesthesia Delivery Systems

Experimental Characterization and Modeling of Helium Dispersion in a ¼-Scale Two-Car Residential Garage

3. Coupled Techniques in Thermal Analysis

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

PURE SUBSTANCE. Nitrogen and gaseous air are pure substances.

PEDESTRIAN UTILIZATION; ENHANCING FROM EXISTING : A STUDY CASE OF PANTAI CHENANG, LANGKAWI AND MELAKA HISTORICAL CITY, MELAKA

ISO INTERNATIONAL STANDARD. Welding consumables Gases and gas mixtures for fusion welding and allied processes

EXPERIMENTAL STUDY ON THE DISCHARGE CHARACTERISTICS OF SLUICE FOR TIDAL POWER PLANT

Test Method of Trap Performance for Induced Siphonage

Effect of airflow direction on human perception of draught

New Technological Guidelines Required for Pipette Management (Proposal of a Pipette Management Method)

Simulation analysis of the influence of breathing on the performance in breaststroke

INTEGRATED FARE PAYMENT SYSTEM IN MULTI OPERATORS SINGLE MARKET PUBLIC BUS NETWORK SAFIZAHANIN BINTI MOKHTAR UNIVERSITI TEKNOLOGI MALAYSIA

Characterizing a Cu/Mn Alloy for Extracting Oxygen from Inert Gas. Streams

Experimental Research on Oxygen-enriched Air Supply System for Gasoline Engine

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

DURING the course of certain investigations it became

Analysis of Pressure Rise During Internal Arc Faults in Switchgear

EN1279 part 3 assessment of insulating glass: IGU with Duralite sealant

PRESSURE REGULATOR LBM SERIES

Laboratory Activity Measurement and Density. Average deviation = Sum of absolute values of all deviations Number of trials

DELTI. Vac. DELTI-Vac MIL. Application. Principle. Characteristic

SIMULATION OF ENTRAPMENTS IN LCM PROCESSES

Technical Note. Determining the surface tension of liquids by measurements on pendant drops

PAB Deployment Simulation with Curved Retainer

The Ideal Gas Constant

Organic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results

THE TIRE LABORATORY MAC DOT - NHTSA AUTHOR EVANS JOHN HARRIS TRANSPORTATION RESEARCH INSTITUTE DEVELOPMENT AKRON RUBBER ITEC 2008 PAPER 8C-1

Study on Fire Plume in Large Spaces Using Ground Heating

EFFICIENCY OF TRIPLE LEFT-TURN LANES AT SIGNALIZED INTERSECTIONS

Experimental Design Unit. Grade 10 Science Ms. Hayduk

Chapter 22 Testing Protocols

Available online at ScienceDirect. Procedia Engineering 112 (2015 )

BASIS Lesson Plan. *Note to teachers: Detailed standards connections can be found at the end of this lesson plan.

INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK

Laboratory studies of water column separation

Gas mixtures. Individual solutions specifically for your application

Vacuum Insulation Panels for Buildings and Technical Applications

Gas Laws. Introduction

REGISTRATION EXAMINATION, NOVEMBER 2017 TRADESMAN GASFITTER QUESTION AND ANSWER BOOKLET. Time allowed THREE hours

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

Experimental Analysis on Vortex Tube Refrigerator Using Different Conical Valve Angles

ISO 6570 INTERNATIONAL STANDARD. Natural gas Determination of potential hydrocarbon liquid content Gravimetric methods

Comparison of Methods and Devices for High Pressure Vessel Passivation B. Zitouni (1), L. Denies (1),M. Peukert (1)

An Engineering Approach to Precision Ammunition Development. Justin Pierce Design Engineer Government and International Contracts ATK Sporting Group

Overview. 2 Module 13: Advanced Data Processing

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

THE WAY THE VENTURI AND ORIFICES WORK

Influence of rounding corners on unsteady flow and heat transfer around a square cylinder

FOURIER TRANSFORM INFRARED SPECTROSCOPY

AIR PRESSURE CONTROLLED MASS MEASUREMENT SYSTEM

1 PIPESYS Application

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii

DESIGN AND PERFORMANCE OF THE NEW SARTORIUS 1 KG VACUUM MASS COMPARATOR AT PTB

Naval Postgraduate School, Operational Oceanography and Meteorology. Since inputs from UDAS are continuously used in projects at the Naval

Flow transients in multiphase pipelines

Determination of the wind pressure distribution on the facade of the triangularly shaped high-rise building structure

End of Chapter Exercises

Correction of Pressure Drop in Steam and Water System in Performance Test of Boiler

Experimental Analysis of Inclined Narrow Plate-Fins Heat Sink under Natural Convection

Putting Report Details: Key and Diagrams: This section provides a visual diagram of the. information is saved in the client s database

Transcription:

Journal of Industrial Technology 8 (1), 1999, 35-41 BUOYANCY EFFECT OF PURGE GAS IN THE DETERMINATION OF CARBON BLACK CONTENT IN POLYETHYLENE M.Y. Ahmad Fuad*, Z. Ismail, S. Rahmad and H. Hanim Plastics Technology Centre SIRIM Berhad, P. 0. Box 7035, 40911 Shah Alam, Malaysia. RINGKASAN: Oksigen dan juga udara adalah gas yang lazim digunakan sebagai gas mencahar (purge gas) dalam kaedah penentuan kandungan karbon hitam dalam polietilina. Kajian menunjukkan bahawa ketepatan kandungan sebenar karbon hitam dipengaruhi oleh penggunaan oksigen sebagai gas mencahar. lni dipercayai disebabkan oleh kesan apungan yang timbul akibat perbezaan densiti antara oksigen dan nitrogen. Udara yang mempunyai densiti yang hampir sama dengan nitrogen adalah pilihan terbaik sebagai gas mencahar untuk memberikan keputusan yang lebih jitu dan kebolehulangan yang tinggi. ABSTRACT: Oxygen and air have been interchangeably used as the purge gas in the determination of carbon black content in polyethylene. This study shows that accuracy of the carbon black filler content is significantly affected by the choice of oxygen as the purge gas used to oxidise the carbon filler. It is believed that the difference in the density of the oxygen give rise to the buoyancy effect. Air, having closer density to the initial nitrogen purge gas in the analysis, is the better choice for purge gas for accuracy of analysis and reproducibility of results. KEYWORDS: Carbon black, purge gas, buoyancy, density, oxygen, air, polyethylene To whom correspondence should be addressed. 35

M. Y. Ahmad Fuad, Z. Ismail, S. Rahmad and H. Hanim INTRODUCTION Carbon black is one of the most common fillers used in rubber and thermoplastics materials. In rubber, it serves as an efficient reinforcement agent that control the key properties of the compound whilst in polyolefin materials, it acts as an effective thermal and ultra-violet stabiliser (Hepburn, 1984; Cotten, 1987; Coran, 1993; Bair, 1981 ). Thermogravimetric analysis technique is a common method to measure the carbon black content in these materials (Fuad et al., 1994; Cassel, 1976). Maurer has reported on the factors that influence precision and accuracy of the analysis by this technique giving due considerations to the experimental conditions (Maurer, 1981 ). The basic experimental procedure is quite simple; the sample is heated up at a constant rate in an atmosphere of nitrogen. After the polymer has been completely burnt off (e.g. at 550 C), oxygen or air is introduced. Subsequent weight loss will be due exclusively to the oxidation of the carbon black in the system. Oxygen and air have been interchangeably used as the oxidising gas for carbon black determination and the applications have been reported in numerous established literature and technical reports (Cassel, 1978; Widmann et al., 1987; Mettler, 1995; Perkin Elmer, 1989). It appears the type of the purging gas is immaterial to the result of the analysis. In this study, we wish to show that the choice of gas, either air or oxygen, do have significant effect on the accuracy of the carbon black quantification. It is suspected that at elevated temperature, the buoyancy effect due to the difference in density between the initial nitrogen purge gas and the subsequent oxygen introduced to the system plays a significant role. This effect is particularly enhanced when the quantity of carbon black to be determined is very small. In the case of carbon black incorporated in high density polyethylene (HOPE) pipe, the stipulated content requirement in the standard is only 2.25 % ± 0.25 (ISO/DIS 4427). Any small change affecting the weighing of the sample will be reflected in the subsequent analysis of the level of carbon black. It is also believed this buoyancy effect is further aggravated by the design of the thermogravimetric analyser, i.e., one with a vertical weighing/furnace system that is susceptible to the influence of the purge gas. Since the analysis forms the basis for decision on acceptance or rejection of the HOPE resins for pipe water application, this variation has to be resolved technically during the thermogravimetric analysis for the filler content. MATERIALS AND METHOD Three HOPE resin samples, namely samples A, B and C were analysed for their carbon black content. The analysis were performed using the Mettler TG50 thermogravimetry 36

Buoyancy Effect of Purge Gas in the Determination of Carbon Black Content in Polyethylene analyser utilising the M3 microbalance with a vertical weighing/furnace system. The sample was placed in an alumina crucible (with lid) and was scanned from 30 C to 550 C at a heating rate of 20 C/min, in an atmosphere of nitrogen. At 550 C, oxygen was introduced and the heating continues till 800 C. The analysis was repeated using air instead of oxygen as the oxidising gas. The carbon black level was calculated using STEP analysis available in the Mettler Graphware data evaluation software. Blank run was performed prior to the analysis using identical test parameters. The flow rate for all purge gases was fixed at a constant rate of 200 ml/min. throughout the analysis. RESULTS AND DISCUSSION A typical thermogravimetric scan for carbon black content analysis is shown in Figure 1. The hump at 550 C is due to the introduction of the second gas and is presumed caused by the difference in the density of the gases. It was noted that the scan in oxygen results in bigger hump than the scan in air. Several computing techniques were applied in the analysis of the carbon black to show the effect of oxygen as compared to air as the oxidising gas. Sample 83 - Carbon BlBGk - N2/02 File: 00013.001 Rate: 20.0 "C/min 23.622 mg Step Analysis Height -0.66 mg -2.79 % ResiC. -0.25 mg -1. 04 % Dpeak 573.3 "C Na '' ' \... Na. TG METTLER 17-Dec-96 Plastics Centre. SIAIM BHD!dent: o. o I--- Air -- --------,J r+---- ------~--... ~ - -. -,... _.. Step Analysis Height -0.43.mg -2.04 % 450 500 550. 600 650 700 750 Figure 1. Typical thermogravimetric curves for carbon black content analysis in oxygen (top) and in air (bottom). Note the bigger hump for the analysis in oxygen. 37

M. Y. Ahmad Fuad*, Z. Ismail, $. Rahmad and H. Hanim Method I The step analysis was performed from 530 C to say 700 C and the weight loss during this stage was taken to be the carbon black level. Table 1 shows the result of the analysis in oxygen (column 2) and in air (column 3). It can be observed clearly that the analysis in the oxygen gives significantly higher values than those carried out in air. Table 1. Carbon Black Content in HOPE analysed under various conditions OXY Carbon Black Content (%) N2 only BCORR AIR 2.94 3.01 2.55 2.83 0.25 2.14 2.12 2.31 2.14 0.13 2.10 1.97 0.07 1.97 0.06 2.01 0.01 2.96 2.68 2.79 2.81 0.14 2.12 2.04 0.06 2.13 2.24 1.75 2.04 0.26 2.12 2.24 1.74 2.03 0.26 2.04 2.02 1.99 2.02 0.03 2.51 3.29 2.63 2.81 0.42 2.13 2.02 0.07 2.10 2.25 1.89 0.18 2.22 1.89 2.07 0.17 2.07 2.08 0.01 SAMPLE A1 A2 A3 MEAN SD 81 82 83 MEAN SD C1 C2 C3 MEAN SD TGA850 Note: OXY run in oxygen; AIR - run in air; N2 - run in air but step evaluation in nitrogen region only; BCORR - run in oxygen, after buoyancy correction; TGA850 - run using Mettler TGA850 c/w horizontal furnace. Method II To negate the effect of oxygen for the samples analysed in the oxygen purge gas, the step analysis was performed from 50 C to 530 C and the residue at 530 C was taken to be carbon black. This is only acceptable if no other mineral filler or ash is known to be present in the samples. The result is shown in column 4 of Table 1. It may be seen that if analysis is performed before the introduction of oxygen, then the carbon black content results tend to be closer to the previous values as analysed in air. Method Ill When the step analysis was performed for the samples in oxygen, it was observed that the final mass of the microbalance (ResiC in Figure 1) was not zero but gave substantial 38

Buoyancy Effect of Purge Gas in the Determination of Carbon Black Content in Polyethylene negative results. This indicates a buoyancy (lifting up) effect due to the oxygen purge gas. If the carbon black content is recalculated by considering the buoyancy effect, i.e., by subtracting the ResiC values, then the carbon black content results (column 5, Table 1) tend to agree with those performed in air (column 3) or those calculated in the nitrogen region (column 4). This clearly shows that the buoyancy effect of the oxygen is the cause of the significant increase in the result of the carbon black content analysis. To validate the results of the analysis, some 'calibrated' samples of carbon black in HOPE were prepared. A known amount of carbon black filler was introduced into HOPE resin granules and the samples were subjected to the same analyses as above, i.e. in oxygen and air. The results are shown in Table 2. It was observed that, as expected, the analysis in oxygen gives significantly higher results. For the analysis performed in air or corrected versions of analysis in oxygen, the analysed carbon black content values are marginally lower than the actual input. This is perfectly acceptable since trace losses of the carbon powder is inevitable. Basically the analysed results are in agreement with the actual amount of carbon black put into the samples. Thus the accuracy of the thermogravimetric analyser and the computing technique can be validated and may support the above argument on the buoyancy effect of oxygen. Good repeatability of the TGA measurements are indicated by the small value of the standard deviations for all the 3 samples at various conditions. For additional cross-checking, the samples A, Band C were also analysed using the latest Mettler TGA850 thermogravimetric module with a horizontal furnace; a better design to minimise the disturbing influence of the purge gas. The results of the analysis as shown in column 6 of Table 1 (run in air), agree with the previous run (in air) using the TG50 system. These results further show that the inflated values of the carbon black initially acquired using the TG50 system are due to the buoyancy effect of the oxygen purge gas. It is believed that this effect is prominent for analysis performed by the TG50 analyser due to the vertical design of the balance/furnace system. Table 2. Carbon Black Content Analysis using Calibrated Samples SAMPLE OXY C10XY C20XY C30XY C1AIR C2AIR C3AIR Carbon Black Content (%) AIR N2 only BCORR 2.97 3.51 1.75 2.36 2.73 2.21 1.82 2.25 1.72 2.35 2.72 INPUT 1.87 2.45 2.86 2.36 1.96 2.36 Note: CnOXY - calibrated samples analysed in oxygen ; CnAIR - calibrated samples analysed in air; INPUT - actual percentage carbon black content input into the samples. 39

M. Y. Ahmad Fuad*, Z. Ismail, S. Rahmad and H. Hanim CONCLUSION For greater accuracy of carbon black determination using the Mettler TG50,thermogravimetric analyser, air (instead of oxygen) is the better choice for purge gas. It is also noted that air gives better reproducibility of results when compared to oxygen. ACKNOWLEDGMENTS Ms S. L. Moh is acknowledged for running some of the TG scans. REFERENCES Bair, H. E. (1981). In: Thermal Characterization of Polymeric Materials. ed. Turi, E. A., Academic Press, Inc, San Diego, pp 845. Cassel, B. (1976). Recent Advances in Thermogravimetric Analysis of Polymer and Elastomer formulations. In: Proceedings of the Sixth North American Thermal Analysis Society Conference, Princeton, New Jersey. Cassel, B. (1978). Evaluating polymer additives by thermal analysis. Plastics Engineering; pp 40. Coran, AV. (1993). The Dispersion of Carbon Black in Natural Rubber. In : Proc. of the Sec. Inter. Cont. on Carbon Black, Mulhouse, France. Cotten, G. (1987). Plas. Rubber Process. Appl., 7 : pp 173. Draft ISO standard, ISO/DIS 4427 - Polyethylene (PE) pipes for water supply - specifications. Fuad, M.Y. A., Zaini, M. J., Mustafah, J., Ishak, Z. A. M. and Omar, A. K. M. (1994). Determination of Filler Content in Rice Husk Ash and Wood Based Composites by Thermogravimetric Analysis. J. Appl. Polym. Sci.. 51 : pp 1875. Hepburn, C. (1984). Plas. Rubber Process. Appl., 9: pp 11. Maurer, J. J. {1981 ). In : Thermal Characterization of Polymeric Materials. ed. Turi, E. A., Academic Press, Inc, San Diego, pp 572. 40

Buoyancy Effect of Purge Gas in the Determination of Carbon Black Content in Polyethylene Mettler Toledo Company Brochure. (1995). Thermal Analysis: Thermogravimetry for exacting requirements. TGA850 Module, 6. Perkin Elmer Company Brochure. (1989). TGA7 Thermogravimetric Analyzer, 5. Widmann, G. and Reisen, R. (1987). In: Thermal Analysis - Terms, Methods, Applications. ed. Huthig, Germany, pp 113. 41