Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y

Similar documents
CAL. Handi+ OPERATING MANUAL & INSTRUCTIONS FOR USE R218P12. R218M12 Rev. J

Handi+N2 OPERATING MANUAL & INSTRUCTIONS FOR USE. R218M06 Industrial. R218M06 Rev. E

Handi+ OPERATING MANUAL & INSTRUCTIONS FOR USE. R218P15 Industrial. R218M15 Rev. D

BAT CAL. MAXO 2 + AE Industrial. MaxO 2 + OPERATING MANUAL & INSTRUCTIONS FOR USE. R217M72 Rev. E

Nitrogen Analyzer INSTRUCTIONS FOR USE R217P65, R217P66, R217P67. R217M65 Rev. G

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. C

AND OPERATION INSTRUCTIONS

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. K

PORTER INSTRUMENT COMPANY, INC.

USER MANUAL. SAVE THESE INSTRUCTIONS (For latest revision, go to

OxiQuant S. Operating Instructions

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

TED 191 PORTABLE OXYGEN MONITOR

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

CSA Sample Draw Aspirator Adapter Operator s Manual

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

with O 2 Controller Instruction Manual

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Flow meter. bellow vaporizer. APL valve. Scavenging system

Push-To-Set Intermittent Suction Unit (PTS-ISU) Instructions for Use

Operation Manual. O2 Quickstick. Oxygen Analyzer

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

RAM 4021 Operation Manual

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Yoke Block Instruction Manual

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

GasLab SAN-102 Micro Oxygen Monitor User Manual

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

453 Series Steam Heated Vaporizing Regulator

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Advanced Test Equipment Rentals ATEC (2832)

SMART Oxygen Analyzer. User Manual

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

User s Guide Temperature Sensor Converter TSC-599

GasLab SAN-102 Micro Oxygen Monitor User Manual

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

TOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev

Ambient Weather WS-03 Thermo-Hygrometer

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

RAM 4021-DPX Operation Manual

VT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features

U S E R M A N U A L CAUTION. Models: 19MFA Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

Blendox AII5000F Inhalation Antalgic System. User s Manual :SP-Y Series No. Version :A1. Date of Review:

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

Psyclone TM. Protimeter Thermo-Hygrometer. GE Sensing. Operating Manual. INS7800, Rev. A (September 2007)

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

NITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS

O3 3E 1 F Gas Sensor Module

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

E X R A D I N Spherical Ion Chambers A3. A4 REF DOC #

Golf Performance Monitors. PureContact Operating Guide. Version of 9

RAM Operation Manual

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

MODEL CALIBRATION GAS DELIVERY SYSTEM

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Certified ISO 9000 INSTALLATION AND OPERATION INSTRUCTIONS. Before Installing or Operating, Read and Comply with These Instructions

RK LEL Sample Draw Aspirator Adapter Operator s Manual

MACS Mask CPAP System

Operation Manual. Pro O 2 Pro O 2 Remote. Oxygen Analyzers. Rev 08.17

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 (SHOWN WITH OPTIONAL EQUIPMENT)

PM105 PRESSURE METER USER S MANUAL

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC #

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

GE Measurement & Control

Yoke Block Instruction Manual

Digital Vacuum Regulator

USER MANUAL. AIR-OXYGEN BLENDER (DISS and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS

TOP VALVE. Pat. #5,857,486 & 5,944,050. High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves. Instruction Manual

623 Series Dome Load Regulators

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

AUTOVENT 4000 VENTILATOR

Model ASSEMBLY and OPERATING INSTRUCTIONS

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Introduction. Table of Contents. Automatic Wrist Blood Pressure Monitor With Voice-Guided Operation. Model No.: BP5K

VAPOMATIC ANESTHETIC VAPORIZER

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

SINGLE STAGE ULTRA-HIGH PRESSURE REDUCING REGULATOR INSTALLATION AND OPERATION INSTRUCTIONS. Certified ISO 9001

Fingertip Pulse Oximeter

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

User s Guide. CO2 Controller for miniature incubators

Overview. Front Panel: Keypad and Display

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Transcription:

Check and Go Disposable Operating Manual 2005 Medical Depot, Inc Port Washington N.Y. 11050 visit us at www. drivemedical.com Check and Go Disposable Part# 18580 2005 Medical Depot, Inc Port Washington N.Y. 11050 www. drivemedical.com

WARNING Before use, all individuals who will be using this product must become thoroughly familiar with the information contained in this Operation Manual. Strict adherence to the operating instructions is necessary for safe and effective product performance. This product will perform only as designed and only if installed and operated in accordance with the manufacturer's operating instructions. WARNING Although the sensor of this device has been tested with various anesthesia gases including nitrous oxide, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane and found to have acceptably low interference, the device in entirety (including electronics) is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. Only the threaded sensor face, flow diverter, and T adapter may be allowed to contact such a gas mixture. CAUTIONS The Drive Check & Go Disposable oxygen sensor is a sealed device containing a weak acid electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and should be disposed of properly, or returned to Maxtec for proper disposal or recovery. The Check and Go Disposable is not intended for steam, ethylene oxide or radiation sterilization. Do not autoclave or expose the sensor to high temperatures. Do not immerse the Check and Go Disposable oxygen analyzer in any cleaning solution. The flow diverter provided with the Check and Go Disposable is for use with flowing gases only. Do not use the diverter when performing static sampling (e.g., in incubators, oxygen tents, oxygen hoods). Do not attempt any repairs or procedures, which are not described in this Operation Manual. Drive cannot warrant this product from damage resulting from misuse, unauthorized repair or improper maintenance of this product. Federal (USA) law restricts this device to sale by or on the order of a physician. Effect of Interferent Gases and Vapors Volume % Dry Interference in O 2 Nitrous Oxide 75% < 2% Halothane 5% < 2% Isoflurance 5% < 2 % Enflurane 5% < 2 % Sevoflurane 6% < 2 % Desflurane 15% < 2 % Carbon Dioxide 10% < 2 % Helium 70% < 2 % Warranty Drive warrants the Check and Go Disposable to be free from defects of workmanship or materials for a period of two-(2) years from the date of shipment from Drive, under normal operating conditions and provided that the Check and Go Disposable is properly operated and maintained in accordance with Drive s operating instructions. Should Check and Go Disposable fail prematurely, the replacement Check and Go Disposable is warranted for the remainder of the original analyzer warranty period. Based on Drive s product evaluation, Drive s sole obligation under the foregoing warranty is limited to making replacements, repairs or issuing credit for equipment found to be defective. This warranty extends only to the buyer purchasing the equipment directly from Drive or through Drive s designated distributors and/or agents as new equipment. Routine maintenance items are excluded from this warranty. Drive shall not be liable to the purchaser or other persons for incidental or consequential damages or equipment that has been subject to abuse, misuse, misapplication, alteration, negligence or accident. THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. 1 14

General Specifications Sensor Type: Measurement Range: Resolution/ Display: Response Time: Drive galvanic cell w/temperature Compensation (Non-Replaceable) 0.0-99.9% oxygen (gas). 0.1% - The three digit LCD indicates values between 0.0-99.9% oxygen. Over range indicated by one decimal point on display located after the first digit. < 15 seconds for 90% step change. (at 25 C) Table of Contents Page Declaration of Conformity. 1 Warning and Cautions.. 2 Introduction. 4 Features and Functions.. 5 Calibrating the Check and Go Disposable 7 Linearity/ Accuracy: @ 15 to 40 C Power: Battery Life: Sample Port Diverter Fitting: Operating Temperature: Storage Temperature: Warm-up Time: Expected Storage Life: Operating Pressure: + 1 % of full scale at constant temperature, R.H. and pressure when calibrated at full scale. + 3% actual oxygen level over full operating temperature. Powered by one internal, non-replaceable Lithium battery. Power on push button automatically shuts off after 80 seconds time-out. Low Battery Indication: Display will turn on and immediately off after button is pushed. Electronics rated general purpose; not for use in hazardous areas or for use with flammable gases. Approx. 1850 hours (74,000 cycles) or 24 months under normal usage. M-16 x 1 thread with diverter fitting and barbed tubing adapter. Fits medical standard 15mm T adapter. 15 to 40 C -15 to 50 C None Required (Instant On) Two months. Special freshness seal on sensor. Atmospheric pressure to 3psig. Environmental: General purpose housing equivalent to NEMA 1. The CHECK AND GO DISPOSABLE TM is not waterproof. 0-95% RH, non-condensing. Warranty: Weight: Twenty-four months in normal operating conditions. Approx. 60 grams Operating the Check and Go Disposable 9 Problem-Solving.. 10 Cleaning and Maintenance 11 Factors Affecting Calibration Effects of Temperature.. 12 Effects of Pressure 12 Effects of High Humidity.. 13 Accessories. 13 General Specifications.. 14 Effect of Interferent Gases and Vapors... 15 Warranty... 15 ON I AUTO OFF O 2.25 inch 57.15 mm Handi CAL OVER RANGE %O 2 1.00 inch 25.40 mm TM 2.85 inch 72.39 mm.875 in. 22.22 mm 13 2

Introduction The Check and Go Disposable is designed to monitor oxygen concentration in the patient-breathing environment. It is one of a full line of oxygen analyzers. The Check and Go Disposable utilizes the Drive oxygen sensor and is engineered for fast response, maximum reliability and stable performance. The Check and Go Disposable is designed primarily for spotchecking of oxygen levels delivered by medical oxygen delivery equipment and respiratory care systems. Its lightweight, compact size, extended battery life, and "auto off" feature makes this oxygen analyzer ideal for portable oxygen analysis by qualified heath care professionals. Humidity Effect Humidity has no effect on the performance of the Check and Go Disposable other than diluting the gas, as long as there is no condensation. Depending on the humidity, the gas may be diluted by as much as 4%, which proportionally reduces the oxygen concentration. The device responds to the actual oxygen concentration rather than the dry concentration. Environments where condensation may occur are to be avoided since condensate may obstruct passage of gas to the sensing surface, resulting in erroneous readings and slower response time. For this reason, the following is recommended: Avoid usage in environments greater than 95% relative humidity. When used in a breathing circuit, place the sensor upstream of the humidifier. 3 12

TM Factors Influencing Calibration The primary factors influencing the Check and Go Disposable are temperature, pressure and humidity. Effects of Temperature The Check and Go Disposable will hold calibration and read correctly within ±3% when in thermal equilibrium within the operating temperature range. The device must be thermally stable when calibrated and allowed to thermally stabilize after experiencing temperature changes before readings are accurate. For these reasons, the following is recommended: Allow adequate time for the sensor to equilibrate to a new ambient temperature. When used in a breathing circuit, place the sensor upstream of the heater. For best results, perform the calibration procedure at a temperature close to the temperature where analysis will occur. Pressure Effect Readings from the Check and Go Disposable are proportional to the partial pressure of oxygen. The partial pressure is equal to the concentration times the absolute pressure. Thus the readings are proportional to the concentration if the pressure is held constant. Flow rate of sample gas can affect pressure at the sensor in that back pressure at the sensing point may change. For these reasons, the following is recommended: Calibrate the Check and Go Disposable at the same pressure as the sample gas. If sample gases flow through tubing, use the same apparatus and flow rates when calibrating as when measuring. The Check and Go Disposable oxygen sensor has been validated at pressures up to 2 atmospheres absolute. Calibration or operation above this pressure is beyond the intended use. LCD DISPLAY "ON" BUTTON ON I CONFIGURATION A FLOW DIVERTER "T" ADAPTER TOP VIEW Handi AUTO OFF OVER RANGE O CALIBRATION KNOB %O 2 TM CAL Features and Functions SIDE VIEW HANDI OXYGEN ANALYZER PRODUCT LABEL CONFIGURATION B BARBED ADAPTER CLEAR TUBING (TO GAS) 11 4

LCD Display: A 3-digit display provides a direct readout of oxygen concentration in the range of 0-99.9%. The display is blank when the Check and Go Disposable enters its Sleep (power off) mode. The Check and Go Disposable will automatically enter the Sleep mode after approximately 1.3 minutes from the last time the ON button was pressed. ON Button/Auto OFF: Use this button to turn the Check and Go Disposable on. When the Check and Go Disposable is in the Sleep (power off) mode, the LCD display is blank. When the ON button is pressed once, the analyzer will display the oxygen concentration for 1.3 minutes. Pressing the ON button during this 1.3 minute "window" will prolong the ON period to 1.3 minutes from the most recent time that the button was pushed. Calibration Knob: Rotate the calibration (CAL) knob to adjust the calibration value to reflect the known oxygen concentration. Flow Diverter: The flow diverter is designed to fit industry standard 15 mm I.D. "T" adapters. Over Range Indicator: The appearance of a decimal point after the first digit means that the Check and Go Disposable is reading in excess of 99.9%. Example: 0.0.0 = 100% 0.0.1 = 101% 0.0.2 = 102% etc. 3. I have calibrated to 99.9% but when I check my oxygen delivery equipment, the Check and Go Disposable reads ".0.4" or greater (Over Range Indicator) It is recommended that you conduct the calibration procedure again to get another reading. The most likely cause is that the Check and Go Disposable has received a "false " calibration value. Make sure that the calibration gas is connected to the Check and Go Disposable at 2 liters per minute for a minimum of 2 minutes prior to proceeding with calibration. This 2-minute equilibration time is necessary to insure that the sensor is completely saturated with the calibration gas. 4. I have found the reading to drift greater than ± 3% from a know source value. What is the possible cause? The sensor may be at or near its useful life. Replace your Check and Go Disposable. Sensor life is dependent on the oxygen concentration exposure. For example, a sensor which is used to check flow meters once a week for 8 hours will outlast one which is used to analyze oxygen blender performance 24 hours per day, 5 days a week Cleaning and Maintenance When cleaning or disinfecting the Check and Go Disposable, take appropriate care to prevent any solution from entering the analyzer. The Check and Go Disposable surface may be cleaned using a mild detergent and moist clothe. The Check and Go Disposable may be disinfected using standard topical disinfectants. The Check and Go Disposable is not intended for steam, ethylene oxide or radiation sterilization. Store the Check and Go Disposable in a temperature similar to its ambient environment of daily use. 5 10

Problem-Solving Problem-Solving Table Condition Likely Cause Recommended Action LCD goes on momentarily and then returns to blank display (Sleep mode). While analyzer on, LCD shows a decimal point (".") after first digit. ON button was not actuated. Battery is depleted. Analyzer is in Over Range mode, meaning that an oxygen concentration greater than 99.9% was identified. Frequently Asked Questions If condition continues after making sure ON button is actuated, then analyzer is Low Battery disabled. Replace analyzer with a new Check and Go Disposable. Contact Drive Customer Service to reorder. 0.0.0 = 100% 0.0.1. = 101% 0.0.2. = 102% If display reads > 0.3. the Check and Go Disposable. should be re-calibrated. 1. I cannot adjust the calibration when I rotate the CAL knob in a clockwise (counterclockwise) direction. Remove the Check and Go Disposable from the calibration sample gas stream and re-calibrate in room air with a gas with known oxygen concentration. If any calibration procedure is unsuccessful, replace the Check and Go Disposable with a new one. To reorder, contact Drive Customer Service. 2. After I calibrate to 20.9%, the display changes as much as ± 1%. 9 This is within the normal Check and Go Disposable operating tolerance of ± 1% when temperature and pressure are constant. Calibrating the Check and Go Disposable Before You Begin A protective film covering the threaded sensor face must be removed and the sensor allowed to breath for at least 30 minutes, next, the Check and Go Disposable Oxygen Analyzer should be calibrated. Thereafter, Drive recommends calibration on a weekly basis. However, more frequent calibration will not adversely affect product performance. Calibration of the instrument should be performed when the temperature of the gas stream changes by more than 3 degrees Celsius. Changes in elevation result in calibration error of approximately 1% of reading per 250 feet. In general, calibration of the instrument should be performed when the geographic elevation at which the product is being used changes by more than 500 feet. In addition, calibration is recommended if the user is unclear when the last calibration procedure was performed or if the measurement value displayed is in question. It is best to calibrate the Check and Go Disposable to a known oxygen value, which has been documented at a pressure and flow similar to your clinical application. Calibrating the Check and Go Disposable at lower concentrations with a known oxygen value is also acceptable and may provide additional accuracy if the calibration gas is closer to the environment in which the Check and Go Disposable will be used. A "known" value of oxygen is defined as an oxygen source, which has a traceable certificate and/or USP certification. Note: Before beginning calibration the Check and Go Disposable must be in thermal equilibrium. You may also need to be aware of other factors, which affect device calibration values. For more information, refer to "Factors Influencing Calibration and Performance" on page 12 of this manual. In Line Calibration (Configuration A) 1. Put the Check and Go Disposable in an upright position such that you can read the product label. 2. Connect a sample supply hose to a standard "T" piece. The Drive "T" is precision-tapered to insure a tight connection with O-rings of the MAX-250 oxygen sensor diverter. 3. Insert the Check and Go Disposable in the center position of the "T" piece. 6

4. Attach an open-ended reservoir to the end of the "T" piece. Then start the calibration flow of oxygen at 2 liters per minute. Six to 10 inches of corrugated tubing works well as a reservoir. A calibration oxygen flow to the Check and Go Disposable of 2 liters per minute is recommended to minimize the possibility of obtaining a "false" calibration value. 5. Allow the oxygen to saturate the sensor. Although a stable value is usually observed within 30 seconds, allow at least 2 minutes to ensure that the sensor is completely saturated with the calibration gas. 6. If the Check and Go Disposable is not already turned on, do so now by pressing the analyzer ON button. 7. Rotate the Calibration knob on the Check and Go Disposable until you read the calibration gas value on the analyzer display. Turning the Calibration knob in a clockwise direction will increase the displayed value. Turning the Calibration knob counterclockwise will decrease the displayed value. Direct Flow Calibration (Configuration B) 1. Attach the Barbed Adapter to the Check and Go Disposable. 2. Connect the clear sampling tube to the Barbed Adapter. 3. Attach the other end of the clear sampling tube to a source of oxygen with a known oxygen concentration value and initiate flow of the calibration gas to the unit (2 liters per minute is recommended). 4. Allow the oxygen to saturate the sensor. Although a stable value is usually observed within 30 seconds, allow at least 2 minutes to ensure that the sensor is completely saturated with the calibration gas. 5. If the Check and Go Disposable is not already turned on, do so now by pressing the analyzer ON button. 6. Rotate the CAL knob on the Check and Go Disposable until you read the calibration gas value on the analyzer display. Turning the CAL knob in a clockwise direction will increase the displayed value. Turning the CAL knob in a counterclockwise direction will decrease the displayed value. Operating the Check and Go Disposable To Check the Oxygen Concentration of a Sample Gas: 1. Maintain the Check and Go Disposable in an upright position such that you can read the product label. 2. Place the Check and Go Disposable in the sample gas stream. IMPORTANT:. When using a standard "T" adapter, make sure that the sensor is mounted in the adapter with the flow diverter pointing down.. Make sure that there is a tight fit between the flow diverter and the "T" adapter. 3. Start the flow of the sample gas to the sensor. 4. Allow the oxygen sensor to remain in the flow of the sample gas until stable. 5. If the Check and Go Disposable is not already turned on, do so now by pressing the analyzer ON button. 6. Read the value displayed on the LCD. Note: If the Check and Go Disposable is used to measure the oxygen concentration with equipment using a heated or humidified gas stream, it is recommended that the Check and Go Disposable be placed upstream of the heater and/or humidifier. For more information, refer to "Factors Influencing Calibration and Performance" on page 12 of this manual. 7 8