Mass Flow Controller (MFC) for Gases

Similar documents
Mass Flow Controller (MFC) for Gases

Mass Flow Meter (MFM) for Gases

Mass Flow Controller (MFC) for Gases

Mass Flow Controller (MFC) for Gases

Bypass Mass Flow Controller (MFC) for gases

Mass Flow Controller (MFC) for Gases

Mass Flow Controller (MFC) for Gases

Mass Flow Meter (MFM) for gases

Mass Flow Controller (MFC) for gases. Inline MFC for nominal flow rates from 25 I N. /min to I N

Mass Flow Controller (MFC)/ Mass Flow Meter (MFM) for gases

2/2-Way Solenoid Control Valve

2/2-Way Proportional Valve (motor-driven)

2/2-way proportional valve

2/2-way proportional valve

2/2-Way Solenoid Control Valve

2/2- and 3/2-way Rocker Solenoid Valve - bistable (impulse) version

4/2 way Pneumatic Solenoid Valve

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

Отсечные клапаны с электроприводом Burkert

Model 601CV. Mass Flow Controller. Product Features:

Design Features. General Description

C-Flow Coriolis Mass Flow Meters

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

Vortex Flow Sensor VA Di Ex-d for flow measurement in gases; with integrated, configurable transducer UVA in Ex-d housing

Armatures for analytical probes

Pilot HON 625. Entwurf. Product information. serving the gas industry worldwide

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

Pressure regulator for water

Series 44 Self-operated Pressure Regulators. Type 44-1 B Pressure Reducing Valve. Type 44-6 B Excess Pressure Valve. ANSI version

The affordable consumption counter for compressed air and gases R 3 /4 (DN 20)

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES

Precision pressure sensor Basic version Model CPT6020

Vortex flow sensor VA Di Ex-d with integrated, configurable transducer UVA in a flameproof enclosure for applications in explosive atmospheres

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Gas Pressure Regulator HON 200

T 2523 EN Type 2406 Excess Pressure Valve Self-operated Pressure Regulators ANSI version

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

Self-operated Pressure Regulators Type 2405 Pressure Reducing Valve

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

Series 44 Self-operated Pressure Regulators. Type 44-0 B Pressure Reducing Valve. ANSI version

KMT. Thermal Mass Flow Meter. for Compressed Air and Gases. Measuring range: m3n/h m3 N

Technical Data Sheet T 2025

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

HON 219 Pressure Reducer (D119a)

SonoMeter 30 Energy Meters

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司

Operating Instructions for Direct Flow Mass Flow Meter for gases. Model: DMW

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

Proportional pressure regulators VPPE

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

Rocker valve. Type Operating Instructions. Bedienungsanleitung Manuel d utilisation

LK-SX VOC. Application. Security Advice Caution. Notes on Disposal. Combined sensor mixed gas. Data sheet

3000 Series OPTIMASS Mass Flowmeter. Stainless Steel Hastelloy

GasMultiBloc Combined regulator and safety valve Infinitely variable air/ gas ratio control mode MBC-300-VEF MBC-700-VEF MBC-1200-VEF

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

MANUAL KPS Pressure Control Valve

EPP4 Pressure Regulator Basic G 1/4"

Ball Float Steam Trap UNA 45 MAX, UNA 46 MAX, UNA 46A MAX PN 40/Class 300 DN 40, 50, 65

Self-operated Pressure Regulators for special applications

Instruction Manual Contact Pressure Vacuum Gauge

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Gas Appliance Pressure Regulators with integrated gas filter. FRI/6 Series

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

The intelligent alternative for VA meters: mass flow meters for gases with flow display

Electropneumatic Positioner and Pneumatic Positioner Type 3760 JIS

Hydraulic hand spindle pump Models CPP1000-M, CPP1000-L

Level transmitter LT 100

Benefits. Rev.III_032018_Oxybaby 6.0_engl subject to technical alterations

Thermal flow sensor TA Di with integrated transducer for measuring mass flow, standard flow rate and air / gas consumption

SITRANS P measuring instruments for pressure

High-performance submersible pressure transmitter For level measurement Model LH-20

DMK 458. Pressure Transmitter for Marine and Offshore. Ceramic Sensor. accuracy according to IEC 60770: standard: 0.25 % FSO option: 0.

Technical Data Sheet. OPTIMASS 7000 Custody Transfer Mass Flowmeter

Precision Mass Flow Meters / Controllers for Liquids and Gases

Measurement of Low GAS & Liquid Flows

Brooks TMF-Zone 1 Models 5816-Ex/5864-Ex

SPCH/ Blowing block

Compact pressure switches for gas and air

Model PDT Dewpoint Transmitter

Old documentation - Only for your information! Product is not available anymore!

safety shut-off Valve HON 703 / HON 704

Mounting and Operating Instructions EB 3009 EN. Self-operated Regulators. Type RS Check Valve (backflow protection)

CALEFFI. Ball zone valves. series /03 GB

Thermal flow sensor TA Di for separate evaluation units for measuring mass flow, standard flow rate and air or gas consumption.

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

Products for Cryogenic Applications

Pressure Calibrators. Series PC..., PM...

Expert Hydrostatic Level Transmitters

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

High-performance submersible pressure transmitter For level measurement Model LH-20

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

GasBloc Multifunctional gas control Combined regulator and safety shut-off valves single-stage atmospheric operating mode GB-(LEP) 057 D01

Old documentation - Only for your information! Product is not available any more!

159 mm 154 mm. Subject to changes 10/2012

ATM/GR. Electronic pressure switches CUSTOMER BENEFITS

Pneumatic high-pressure controller Model CPC7000

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

Transcription:

Mass Flow Controller (MFC) for Gases Bypass MFC with capillary technology for nominal flow rates from 5 ml N /min to 15 l N /min Applicable for aggressive gases Compact design and digital communication Type 8715 can be combined with... Typ 33 3/2 or 2/2 way valve Typ 613 2/2 way valve Type 8715 controls the mass flow of gases through a sensor element which is not in direct contact with the gas itself. The measured value provided by the sensor (see the description on page 2) will be compared in the digital control electronics with the predefined set point according to the signal; if a control difference is present, the control value output to the proportional valve will be modified using a PI control algorithm. In this way, the mass flow can be maintained at a fixed value or a predefined profile can be followed, regardless of pressure variations or other changes in the system. Type 8715 can optionally be calibrated for two different gases, the user can switch between these two gases. As control element a direct-acting proportional valve guarantees a high sensitivity and a good control characteristics of the MFC. This instrument communicates digitally with master devices, no further A/D conversions needed. The MassFlowCommunicator software can be used for parameterisation and diagnosis. Technical data Full scale range 1) 5 to 15 ml N /min 2) (Q ) N 2 equivalent nom Control range 1:5 Operating gases Neutral, or aggressive gases Calibration gas Operating gas or air with conversion factor Max. operating pressure (Inlet pressure) 1 bar (145 psi), depending on the orifice of the valve Medium temperature - 1 to + 7 C (- 1 to + 6 C for oxygen) Ambient temperature - 1 to + 5 C 3), others on request Accuracy ± 1.5 % o.r. ±.3 % F.S. (after 3min. warm-up time) Repeatability ±.1 % F.S. Settling time (t 95 % ) < 3 sec. Materials Body Stainless steel Housing PC (Polycarbonate) or metal Seals FKM, EPDM or FFKM Port connections NPT ¼, G ¼, Screw-in fitting or sub-base, others on request Control valve (proportional valve) Valve orifice k VS -value Normally closed.5 to 2. mm.6 to.9 m 3 /h 1) The nominal flow value is the max. flow value calibrated which can be controlled. The nominal flow range defines the range of nominal flow rates (full scale values) possible. 2) Index N: Flow rates referred to 1.13 bar and C. Alternatively there is an Index S available which refers to 1.13 bar and 2 C 3) When an internal 5 W valve is used: Max. 4 C Electr. connection D-Sub plug 9 pin Power supply 24 V DC Voltage tolerance ± 1 % Residual ripple < 2 % Power consumption 3.5-11.5 W (depends on proportional vale) Communication Digital via RS485 (half duplex or full duplex), RS422 Protection class IP4 Dimensions [mm] See drawings on pages 5 and 6 Total weight ca. 85 g (stainless steel) Mounting position Horizontal or vertical Light emitting diode display (default, other allocations possible) Binary input (default, other functions possible) Binary output (default, other functions possible) Indication for Power, Limit Error Start Autotune One relay-output for 1. setpoint not reached, Max. load: 25 V, 1 A, 25 VA www.burkert.com p. 1/7

Measuring principle The measurement is based on the bypass principle. A laminar flow element in the main channel generates a small pressure drop. This drives a small flow, proportional to the main flow, through the bypass (sensor tube). Two heating resistors, which are connected in a measuring bridge, are wounded on this stainless steel tube. In the zero-flow state, the bridge is balanced, but with flow, heat is transported in the flow direction and the bridge becomes unbalanced. The dynamics of the measurement is limited by the tube walls, which act as a thermal barrier. Through use of suitable software in the controller, response times are obtained (in the range of a few seconds) that are adequate for a wide range of applications. With contaminated gases we recommend to install filter elements upstream. This avoids changes in the division ratio between main flow and sensor tube, as well as changes in the heat transmission caused by deposits on the walls of the sensor tube. With these sensors even aggressive gases can be controlled, because all essential parts in contact with the gas are fabricated in stainless steel. With this sensor principle it is also possible to convert between different gases. Q(Gas) = f x Q(N 2 ) gas factor f N 2 1. Luft 1. O 2.98 H 2 1.1 Ar 1.4 He 1.42 CO 2.77 By using the gas factors it is possible that the accuracy is not within the datasheet specification. For applications which need high accuracy it is recommended to calibrate under application conditions. The compatibility of the sealing materials of the MFCs should be checked before use with another gas. Notes regarding the selection of the unit For the proper choice of the actuator orifice within the MFC, not only the required maximum flow rate Q nom, but also the pressure values directly before and after the MFC (p 1, p 2 ) at this flow rate Q nom should be known. In general, these pressures are not the same as the overall inlet and outlet pressures of the whole plant, because usually there are additional flow resistors (tubing, additional shut-off valves, nozzles etc.) present both before and after the controller. Please use the request for quotation form on p. 7 to indicate the pressures directly before and after the MFC. If these should be unknown or not accessible to a measurement, estimates are to be made by taking into account the approximate pressure drops over the flow resistors before and after the MFC, respectively, at a flow rate of Q nom. In addition, please quote the maximum inlet pressure p 1max to be encountered. This data is needed to make sure the actuator is able to provide a close-tight function within all the specified modes of operation. The request for quotation form on page 7 contains the relevant fluid specification. Please use in this way the experience of Bürkert engineers already in the design phase and provide us with a copy of the request containing the data of your application together with your inquiry or order. p. 2/7

Ordering table for accessories Article Article no. 9 pin electrical connection D-Sub socket 9 pin solder connection with housing 917623 Adapters 4) USB adapter (version 1.1, USB-socket type B) USB connection cable 2 m Communication software MassFlowCommunicator 67693 7729 Info at www.burkert.com 4) The adapters serve mainly for initial operation or diagnosis. Those are not obligatory for continuous operation. Software MassFlowCommunicator for Communication with Bürkert MFC/MFM The communication software allows the user to program additionally various functions. For that purpose the MFC or MFM has to be connected to the computer by a RS232 adapter. p. 3/7

Pin Assignment 6 9 1 5 Pin Connection 1 Binary input (related to GND Pin 2) 2 GND 3 Power supply + 24 V DC 4 Relay, normally opened 5 Relay, normally closed 6 TX+ (RS485-Y) bridge with pin 9 at half duplex 7 TX- (RS485-Z) bridge with pin 8 at half duplex 8 RX- (RS485-B) 9 RX+ (RS485-A) Networking Pin 9, RS485 A Pin 8, RS485 B Pin 9, RS485 A Pin 8, RS485 B Pin 9, RS485 A Pin 8, RS485 B Master: i.e. PLC with RS485 Interface Slaves (max. 32) Mass Flow Controller p. 4/7

Dimensions [mm] Threaded version 2x M 4 6 1 97 17 28 2 approx. 3.5 1.7 8.5 61.5 12.5 G1/4 or NPT1/4 12 Sub-base version 4x 4.5 58.5 17.75 14 14 17.75 35.5 35.5 2x 8.8 2x 6 1 26 58.5 81 92 97 43 17 approx. 3.5 1.7 8.5 61.5 2 x M 4 5 p. 5/7

Dimensions [mm], continued Version with external valve 37 1 16 17 2x M 4.5 8.5 approx. 3.5 1.9 86.5 61.5 ca. 24 1.7 G1/4 or NPT1/4 12 12.5 p. 6/7

MFC/MFM-applications - Request for quotation Please complete and send to your nearest Bürkert sales centre Company Contact person Note You can fill out the fields directly in the PDF file before printing out the form. Customer No Address Postcode/Town Department Tel./Fax E-mail MFC-Application MFM-Application Quantity Required delivery date Medium data Type of gas (or gas proportion in mixtures) Density kg/m 3 5) Gas temperature [ºC or ºF] ºC ºF Moisture content g/m 3 Abrasive components/solid particles no yes, as follows: Fluidic data Flow range Q nom Min. l N /min 5) l S /min (slpm) 6) Max. m N3 /h 5) kg/h c m N3 /min 5) c m S3 /min (sccm) 6) l N /h 5) l S /h 6) 7) Inlet pressure at Q nom p 1 = bar(g) Outlet pressure at Q nom p 2 = bar(g) Max. inlet pressure P 1max bar(g) MFC/MFM port connection without screw-in fitting ¼" G-thread (DIN ISO 228/1) ¼" NPT-thread (ANSI B1.2) with screw-in fitting (acc. to specification for pipeline) mm pipeline (external Ø) inch pipeline (external Ø) Flange version Installation Ambient temperature horizontal vertical, flow upwards C vertical, flow downwards Material data Body Stainless steel Seal FKM EPDM FFKM Please quote all pressure values as overpressures with respect to atmospheric pressure bar(ü) 5) at: 1.13 bar(a) and C 6) at: 1.13 bar (a) and 2 C 7) matches with calibration pressure To find your nearest Bürkert facility, click on the orange box www.burkert.com In case of special application conditions, please consult for advice. Subject to alteration. Christian Bürkert GmbH & Co. KG 183/2_EU-en_895247 p. 7/7