VETROSON. OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure

Similar documents
SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

Welcome to the Specialized Medical Services Respiratory training webinar series!

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

2010 Technical Education. Topics to be covered: Set Up. Training Manual Overview Set Up / Specifications & Preventive Maintenance

NGP-250/500 Nitrogen Generator Quick Start Guide

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

Workhorse-15 Oxygen Generator

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

Operating Manual. Model BAC-30RT30/BAC-30DC

RIX COMPRESSORS. MICROBOOST Owner s Manual CONFIGURATION A MB-115 MB-230. November 21, 2002 APPLICABILITY [ A]

Customer Responsibilities

OPERATION AND MAINTENANCE INSTRUCTION MANUAL

Workhorse 8 Oxygen System Model No 5665, 5673, 5679

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATORS OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY GLOSSARY OF SYMBOLS. : ON (power switched on)

Universal Valve Company Inc

Oxygen and other Compressed Medical Gas Rules and Regulations Complying with the Law. Regulatory agencies:

OX SCFH Oxygen Concentrator. Installation and Operation Manual

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY. by Supera Anesthesia Innovations GLOSSARY OF SYMBOLS. : ON (power switched on)

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

Best Practices Pneumatics Machine & Design. Written by Richard F. Bullers, CFPPS as published in Fluid Power Journal, July/August 2016

OPERATING INSTRUCTIONS

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

INFINITE 02. Storage System with Compressor. Installation and operation Manual

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

NGN Series nitrogen generator

INSTALLATION GUIDE Gas Rangetops

Universal Valve Company Inc

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

Emergency Mobile Unit Medicinal gases everywhere, always on time

Instruction Manual Recovery Station

Code AWC20HP Air Compressor

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

OPERATION MANUAL NTF-60 Plus

High Performance Oilless Commercial Refrigerant Recovery Unit. User Manual Rev. 4 November 2010

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

Budget Range Operators Handbook

AA-75CF Portable Air Compressor NSN:

PumpAgents.com - Click here for Pricing/Ordering Nominal psi (bar) Cut-In Cut-Out

NTF-230 OPERATION MANUAL

Oxygen Generating Systems Ozone Medical Welding Glassblowing Aquaculture Pulp & Paper Wastewater

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

AirSep Corporation Centrox PSA Oxygen Concentrator

High Performance Oilless Commercial Refrigerant Recovery Unit. User Manual

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

OPERATION MANUAL NTF-15

Accu-Tab Systems 1000 Series by Axiall Corporation

Reverse Osmosis System Installation Guide and Operation Manual.

Nitrogen Supply System

BEFORE HANDLING REFRIGERANTS, READ MATERIAL SAFETY DATA SHEET FROM REFRIGERANT MANUFACTURER.

NTF-515. Operation Manual

VERTICAL AIR COMPRESSORS

Instruction Manual Updated 7/26/2011 Ver. 2.2

Fiber Cable Puller with Tuf-Lugger lite

MEDICAL EQUIPMENT CATALOG

Accu-Tab Systems 2000 P Series by Axiall Corporation

ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION

RPS900W Redundant Power Supply. Installation Guide.

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

I. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:


Ever. EverFlo. Home oxygen system. Oxygen simplified

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: WARRIOR 55 PART NO: (110V) , (230V) , LS0512

R E D I C O N T R O L S

ARCHITECTURAL GUIDELINES FOUR-PIECE RESIDENTIAL

400H HARDMOUNT AIR COMPRESSOR KIT PART NO H HARDMOUNT AIR COMPRESSOR KIT PART NO

Instructions for: COMPRESSORS 270ltr belt drive 5.5 / 7.5 hp

Oxygen Concentrator Instruction

Anesthesia Machine Intro

COMPRESSOR MANUAL. (888)

Nidek Medical Products, Inc Mark 5 Nuvo Lite Oxygen Concentrator Service Manual

Model PSI Compressor with 3-Gallon Air Tank 12VDC

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Standard Operating and Maintenance Instructions for Pumping System Model PS-90

NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18

APPS-160 Gen 2 User Manual

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

200 PSI COMPRESSORS - MODEL NUMBERS

Light Commercial Reverse Osmosis System. EE-1000 Manual

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

Model 3600-R410A. Part No Instruction Operation & Maintenance

High Performance Oilless Commercial Refrigerant Recovery System. User Manual REV E 06/30/03

Oxygen concentrators. A user guide.

CONGRATULATIONS ON YOUR PURCHASE OF YOUR THUNDER COMPRESSOR For your personal safety please read, understand and follow the information provided in

Manual. FRESH AERO OneOP Model PC (Portable Compressors) Tire Inflator. Where Imagination Meets Innovation

100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

Portable Oil Lube Air Compressors

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Refrigerant Recovery System Model 3600 Part No

Model OG-15 & OG-15 Compressorless Oxygen Generator Installation, Operation & Maintenance Manual

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

/ /

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

Avio 500 ICP-OES. ICP-Optical Emission Spectroscopy PREPARING YOUR LAB. Preparation Checklist. Suitable Working Area

Mark 5 Plus Oxygen Concentrator Domestic Models 115V; 60Hz

Anes Machine Use and Pre-use Check - CA1 focus

Transcription:

VETROSON OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure

Q & A VETROSON OXY-GEN SYSTEMS 1) Q. How does the VETROSON OXY-GEN System work? A: A simplified explanation is that air (21% oxygen, 78% nitrogen, 0.97% argon and 0.03% other gases) is drawn into a powerful compressor in the generator. The oxygen is separated from the compressed air by passing through the molecular sieve which contains zeolite. Zeolite may be compared to the soda lime in your anesthesia machine which absorbs CO2. Zeolite adsorbs the nitrogen and sends it back into the room. The oxygen is delivered to the receiver tank where it is ready to be delivered throughout the hospital. A diagram of this is below. Molecular Sieves 2) Q. What is the difference between your oxygen generator system and an oxygen concentrator? A. The terms have mistakenly been used interchangeably. Generally, the term oxygen concentrator describes a smaller home medical system used in a 1-to-1 situation producing oxygen at only 5 PSI. Concentrators are placed near a patient s bed. They cannot deliver oxygen any appreciable distance from the unit. Also, the purity of oxygen (%) falls drastically at a LPM flow necessary to supply oxygen for the requirement of a veterinary hospital. If an attempt is made to increase the PSI with an external booster, LPM will be decreased in addition to a decrease in oxygen purity. Oxygen generators are large capacity units. Our OXY-GEN SYSTEMS have built-in compressors and are modified to boost the delivery of oxygen to the practice s central oxygen location (i.e. manifold) at 93% +/- 3% oxygen purity. We have a 20 PSI model and a 50 PSI model.

3) Q. I don t think I have room for one. Is it big like a refrigerator? A: No. The specifications are: 20 PSI, 15 LPM 115V 50 PSI, 15 LPM 230V Height 38 38 Width 25" 25" Depth 17 17 Weight*** 145 lbs 220 *** Weight shown is for the generator only The oxygen receiver tank is 7 ½" in diameter, weighing 34 lbs. and holds approximately 30 liters of oxygen. Its stand measures 27.5" high x 13" wide x 15" deep. 4) Q: What is the function of the Receiver Tank? A: The Receiver Tank has two functions: 1) First having an extra supply of oxygen allows you to flush an anesthesia machine without depressing the flow on the others. 2) In the event of a power failure and loss of electricity, the receiver tank has a sufficient oxygen supply (approximately 20 minutes on the average) to allow you to complete surgery and at the same time turn on your back up H tanks. It should be noted that our special valve on the receiver tank is constantly releasing a small amount of oxygen when you are using a low flow demand. Your VETROSON Oxy-Gen System is a constant flow system. 5) Q: How much noise does it make? A: The noise level averages about 65-75 decibels depending upon the model. It may be compared with a new window air conditioner. When the system is On the special valve on the receiver tank continually releases approximately 1 LPM depending on the model. This results in a hissing sound, which is normal. 6) Q: How can I calculate the LPM of oxygen required for my clinic? A: To calculate your usage you must know your maximum oxygen use. Total the LPM you use on the flow meter for each anesthesia machine, ICU and any other oxygen consuming items in the clinic (1 for Anesthesia machines, 10 to fill Snyder or Cage Door ICU- 5 for maintenance, and 4 as an average for ventilators). Add the amount of the projected use of any new equipment that you plan to add to the clinic. This will give you the estimated maximum LPM requirement. This figure plus 20% more LPM will show you the model which will be best for most practices. We would suggest that veterinarians with very large facilities that require more than 15 LPM contact Summit Hill Laboratories. We recommend the piggy-backing of two units or "zoning" if our system is to be installed in a new practice. 7) Q: You have machines at 20 PSI & 50 PSI that deliver oxygen at 93% +/- 3% purity. Which machine should I buy? 2

A: For a practice with only anesthesia machines and/or ICU units, a 20 PSI unit would be perfect. For practices with ventilators, the 50 PSI unit is recommended, as ventilators require 50 PSI to handle the mechanical side. 8) Q: How quickly would the VETROSON OXY-GEN SYSTEM pay for itself? A: For example, if you lease the 20-15 VETROSON OXY-GEN SYSTEM, the leasing price would be about $156.00*/month for five years. Compare this figure to the cost to your monthly oxygen bill. The difference between a monthly leasing price of $156.00 and your monthly oxygen cost is your monthly savings. On the average, your monthly oxygen bill will be 2-3 times the leasing amount. In addition, you will have an endless supply of oxygen at pennies per day. Remember, if you lease, you will own the system after five years for $1.00. Also, you can review your oxygen purchases for the past year and divide the figure into the cost of the system to determine the number of months it would take to pay for the system. Either way, you will note the dollar savings and convenience using the VETROSON OXY-GEN SYSTEM. *The leasing figure used here was current at the time of printing. It is only one example of a leasing cost. We use it only to illustrate potential monthly savings. Lease rates may vary by lessor. 9) Q: Is 50 PSI required to operate a veterinary anesthesia machine or an intensive care unit? A: No. Flow rate (i.e. LPM) is the most important factor in an oxygen delivery system. Our 20 PSI VETROSON Oxy-Gen System will supply the needs of veterinary practices that do not have ventilators. When our 20 PSI unit is installed in an existing practice, a drop of about 10% will be noted on the flowmeters. Adjust the flow to where it was previously and oxygen will be supplied at the rate that it had been receiving at 50 PSI from the tank prior to our machine's installation. Our present unit supplies 15 LPM. 10) Q: You make reference to ventilators in your oxygen usage chart. I understand ventilators will only run at 50 PSI no lower. What s the answer here? A: Most anesthesia machines with ventilators have dual inlets. One is directly to the line that supplies oxygen to the patient, the other is to power the bellows. A 20 PSI Oxy-Gen System can easily supply the needs of the animal side; however, the bellows requires 50 PSI. Therefore, a 50 PSI unit should be purchased if a ventilator is being employed. The LPM is determined by the weight of the animal. (Note: There are at least two companies that market anesthesia machines with built in ventilators. These require our 50 psi unit. They use considerably more oxygen per # of animal than conventional ventilators.) If a practice has our 20 PSI generator and adds a ventilator, an air compressor set at 50 PSI may be used to run the mechanical side only. * PLEASE NOTE: The use of a ventilator (requiring 50 PSI) with a 20 PSI System will over-tax the molecular sieve beds by asking for a higher pressure than the system is capable of producing. This will result in contamination of the molecular sieve beds and a possible machine breakdown. 11) Q: Your literature says your units produce 93% +/- 3% oxygen. What does the remaining % consist of? A: Nitrogen and Argon, both gases harmless to the patients. 3

12) Q: Can your VETROSON OXY-GEN SYSTEM handle the O2 supply for an ICU? A: Yes. Our 20-15 system has been tested supplying O2 to the Snyder Mfg. Co. ICU 2000 H double cage. This is a 6 ft. model.* To attain 35% O2 concentration, the flow meter on the Snyder unit was set at 8-10 LPM (no higher). It reached the 35% level in less than 25 minutes. The flow meter was then set at 4-5 LPM for maintenance. Our 2015 VETROSON OXY-GEN SYSTEM can easily handle the ICU requirements plus 3-4 anesthesia machines simultaneously at the cost of the electricity consumed by one light bulb. * Please note: We have found that there is no "stop" on these ICU flow meters. By turning the flow meter knob all the way up, as much as 70 LPM may be delivered. Therefore, setting the flow meter on an ICU higher than 10 LPM may result in exceeding the capacity of the model. This, of course, depends on the LPM of your model and how high above 10 LPM the knob on the flowmeter is turned. If the capacity is exceeded, the molecular sieves will break down. This type of breakdown is NOT covered by the SHL warranty. The same situation exists on cage door ICUs. 13) Q: How difficult is it to connect the VETROSON OXY-GEN SYSTEM to the VETROSON Receiver Tank? A: It's rather simple. The receiver tank and generator will be packed in separate boxes. The large box contains the generator. It will be shipped on a pallet, strapped in an upright position. Slide the box off of the pallet. Open the box and remove all of the packing. Roll the generator out of the box. Place it where it will be located. Open the smaller box with the receiver tank. It will also have the additional components (O2 equalizer valve and two oxygen hoses). One end of the 3 ft. oxygen hose is already connected to the receiver tank at the factory. Connect the male fitting on the other end to the female fitting on the oxygen outlet on the generator. A 9 ft. oxygen hose is attached to the elbow on the receiver tank. It should be connected to the low pressure side of the practice s central oxygen system manifold. 14) Q: Once the system is connected, what will it look like? A: The OXY-GEN SYSTEM consists of the oxygen generator and a receiver tank. The generator has an Off/On switch, a flow meter, an hour meter, an oxygen pressure gauge to show the line pressure produced. Please consult the "How To Set Up Your VETROSON OXY-GEN SYSTEM" sheet for the details. At the lower right side of the cabinet is a circuit breaker to protect the generator in case of power overload. The receiver tank has two oxygen hoses: the 3 ft. one connects to the generator, and the 9 ft. one connects to the central oxygen system manifold. The oxygen hoses have a series of check-valves to prevent back flow from the manifold to the OXY-GEN SYSTEM. 4

Low Pressure Tank Regulator 15) Q: My present oxygen manifold system has a low pressure alarm. It sounds if the pressure drops. Will your 20 PSI unit work with this system? A: The low pressure alarm must adjusted to 15 PSI or be turned off when a 20 PSI VETROSON System is being used. The 50 PSI System should not affect the alarm. 16) Q: I have a high pressure manifold for my present oxygen system. How can I hook up the VETROSON OXY-GEN SYSTEM? A: A high pressure manifold has a built in pressure regulator. Our systems cannot be connected into the high pressure manifold in this situation. It should be connected into the supply line exiting the manifold or be plugged into any of the quick-release outlets within the central oxygen line using a special connector. 17) Q: How do I connect the VETROSON OXY-GEN SYSTEM to my anesthesia machine if my practice does not have a central oxygen system with a manifold? A: Just connect the 9 ft. oxygen line from the receiver tank directly to the machine requiring oxygen. A "Y" can be used to supply a second unit. 18) Q: I have E tanks on all of my anesthesia machines and no central system. What do I have to do to connect my anesthesia machines to your VETROSON OXY-GEN SYSTEM? A: You must plumb in oxygen lines to all of your anesthesia machines using an oxygen manifold system. The plumbing, connectors etc. for each oxygen location would cost approximately $600.00 plus $600.00 to connect to the oxygen manifold. Therefore, if you have four oxygen locations, the cost would be about $3000.00. The "E" tanks should remain in place in the event of a power failure or one "H" tank could be 5

purchased and tied into the central oxygen manifold for backup. The savings in using the VETROSON OXY-GEN SYSTEM in this case would be significant and pay for the plumbing. MATRX OX YGEN MANIFOLD SY STEM (Low Pressure) SUPPLY LINE GOING T O ANESTHESIA MACHINES ETC. LINE FROM "H" TANK ATTACHE S HERE THE 9' LINE FROM VETROSON O X Y-GEN SYSTEM RECEIVER TANK ATTACHES HERE 19) Q: I have four anesthesia machines and they all have ventilators connected to them. What is your recommendation? A: Purchase two 50-15 VETROSON OXY-GEN Systems which will produce approximately 26 LPM at 50 PSI. These would be piggy-backed with each unit s 9 ft. hose being connected into a Y. One line would exit the Y and connect to the manifold. You will have more than enough LPM to run the practice. 20) Q. Where should I place my VETROSON OXY-GEN System? A: We would prefer that our unit be placed in the room where your oxygen tanks and manifold are located. However, the room must be climate controlled, ventilated with a 100 cfm exhaust fan, clean and have the minimum dimensions of 1000 cubic feet. If the practice has Chemtron or Ohio quick-disconnects, place the unit where any quick-disconnect is positioned. The following explains how this may be accomplished: Connecting the VETROSON Oxy-Gen System to the Chemtron & Ohio oxygen quick-disconnects Photo #1 - Fron t view 6 Photo #2 - Bottom Side View

Photo #1 - You are looking at Ohio and Chemtron oxygen Quick Disconnects as they come from a wall attachment. Our 9 ft. oxygen hose should be attached to the 3/8" DISS Oxygen fitting on the "T" between the wall plate and face plate. This makes it possible to connect our VETROSON Oxy-Gen System at any location in the hospital that has either oxygen quick disconnect. Photo #2 - Shows a side view of both oxygen quick disconnects with the face plate to the left and the wall plate to the right. Our 3/8" DISS oxygen fitting on the "T" is positioned between the two. The 9 ft. oxygen hose attaches to the "T". Please read the Set Up & Installation Instructions in our Installation, Operation & Maintenance manual before installing our VETROSON Oxy-Gen System. It is required that a 20 amp circuit be dedicated to both our 115V and 230V systems. The 115V units are supplied with a 3-pronged ground fault protected plug. The 230V 60Hz (US) units are supplied with a 20 amp-230v slotted plug, Type #6-20P. This requires a 6-20R receptacle. Note: Our 230V units require 230V single phase electrical service. Triple phase service provides only 208V. If a practice has a 208V system, it cannot be used to supply electric to our 230V system. A 208 voltage supply will cause premature burn out of our major components. In such cases a step-up transformer may be wired into the 208 line at a relatively low cost. We recommend a Hammond QC50ERCB step up transformer. Contact your local electrician to purchase and install this. The unit should never be placed in or around where animal hair is being clipped. Our machine has a strong pull that will bring an abundance of hair into the unit, blocking the filters. 21) Q: What are the electrical requirements for each model? A: Model Delivers Volts Starting Amps Running Amps Power Watts VGS2015 20 PSI, 15 LPM 115 20 8 920 VGS5015 50 PSI, 15 LPM 230 15 7 1610 The 230V units are supplied with a 20 amp-230v slotted plug- type# 6-20P. This requires a 6-20R receptacle. Please remember that our units require a dedicated line. 22) Q: What maintenance is required for the VETROSON OXY-GEN SYSTEMS and when? A: Depending upon the model, one or more pleated HEPA Air Intake Filters (VGS1500) are used.vgs1500 consists of 2 parts: the HEPA Filter component and a foam filter which caps it. They fit into the round holes at the rear of the unit. Large particle filters (VGS1600) are located in the cabinet door. The large particle filters and the foam filter that is part of VGS1500 should be checked weekly and brushed off or vacuumed and washed as indicated. Both filter systems should be replaced every 6 months or sooner if indicated. The Switch Valve (VGS2278) should be replaced after each 1000 hours of use. 7

Note: The generators provide the best performance at temperatures between 40 & 80 F. and when the feed air has a dew point of 40 F or less. The hospital should be between 40 and 80 F for peak performance of VETROSON Oxy-Gen System. Therefore, if temperatures in the hospital or a mobile hospital unit are outside of the peak performance zone the temperature of the hospital should be brought into the peak performance zone before operating the generating system. When operating your VETROSON Oxy-Gen System, please make sure the cabinet door is closed and locked. This is not only for safety reasons. Leaving the door open when running the machine may cause overheating of the compressors. 23) Q: What is the warranty? A: The VETROSON OXY-GEN SYSTEMS are warranted for one (1) year or 1000 hours of use, whichever comes first. 24) Q: If I have a technical question or a breakdown, how should I handle these? A: For technical questions, contact Summit Hill Laboratories at 1-800-922-0722. For repairs/breakdowns or set up questions contact: Summit Hill Laboratories. Customer Service: 1-800-922-0722 E-mail: info@summithilllaboratories.com COPYRIGHT SUMMIT HILL LABORATORIES - ALL RIGHTS RESERVED Jan 1, 2016 8