Operation and Maintenance Manual for GENTEC Medical Gas Outlets

Similar documents
Installation and Maintenance Manual. Compact Medical Gas Outlets

Installation and Maintenance Manual. Dual Medical Gas Outlets

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators

Product Description 3 Cleaning and Lubricating 4 Inspection and Testing 4

SUCTION REGULATORS, FLOWMETERS & ACCESSORIES

American Medical Gas Outlets

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

GM2 SERIES DOME BIAS AUTOMATIC MANIFOLD

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Oxygen and other Compressed Medical Gas Rules and Regulations Complying with the Law. Regulatory agencies:

Safety Policy Manual Policy No. 133

Introduction. Table of Contents. Andersen Medical Gas 12 Place Lafitte Madisonville, LA

Effective 2/1/17. SC1 Service Call PER Visit $76.00

Operating and Maintenance Manual. Flowmeter

Installation & Maintenance Manual. Economy Manifold

Installation & Maintenance Manual. C2 & C3 Manifolds

Hydraulic Punch Drivers

Medical Gas Control Panel

ALERT SERVICE LETTER

TECHNICAL DATA CAUTION

Installation Instructions for NP100 Series Nitrogen Control Panels

Booster Pump PB4-60 Replacement Kits

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Shut-Off Valves and Valve Boxes

Aluminum 13 cu tank Includes regulator Fill pressure: 3,000 psi (200 bar) Output: 1,100 psi +/- 10% (75bar)

35 TON HYDRAULIC PUNCH WARNING

Hydraulic Punch Drivers

Rev. B. Operating Instructions

REL-46 WARNING NOTICE 15 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD. Compatible with RELIABLE R15 and P Style dies. REL-46 Manual

CS150 CAP STAPLER OWNER S MANUAL

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

TABLE OF CONTENTS. Copyright 2010 Precision Medical, Inc. 10m 11/10 Printed in USA

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

User Operation Manual and Safety Instructions

AND OPERATION INSTRUCTIONS

- GAS REGULATOR - COMPACT STYLE

SERIES-O REGULATOR. Instructions for use

CRYOGENIC SERIES Regulators, Pressure Gauges, Shut-off Valves, Relief Valves Technology for a Better Future

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Entry to these spaces must be rigorously controlled to prevent serious injury or death.

HYDRAULIC PUNCH DRIVER 38456, 38520, 7306 / 7306SB, 7310 / 7310SB, 7506, 7606SB, 7610SB, 7625 / 7625Pg / 7625PgSB, 7646 / 7646Pg / 7646PgSB

Economy Manifold Alert -2. Installation and Maintenance Manual V1.6

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Technical Specification

COMPANY INTRODUCTION GENTEC Medical Products

REL-510H WARNING NOTICE 12 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD

Pressure Regulators. Instructions for Use


Oxygen Dialflow Meter. Instructions for Use

MEDICAL EQUIPMENT CATALOG

Johnston Invar Coupling Johnston Invar Coupling

PRODUCT OPERATING MANUAL

GATOR Battery-powered Pistol Grip Cable Cutters

Anti-flood device Model B-1

PDY TON HYDRAULIC CRIMPING TOOL WARNING

Model 7989T Steel Pipe Squeezer Sch. 40 & Sch. 80. Operations Manual

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Protocol Switchover Station

Flow meter. bellow vaporizer. APL valve. Scavenging system

Yoke Block Instruction Manual

AUTOMATIC GAS MANIFOLD

VACUUM CHAMBER MANUAL

TECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

Installation, Operation & Maintenance Manual for Flo-Max Coupler Bracket Model FM150

Yoke Block Instruction Manual

32374 R. Jazowski 4/4/ D. W. Mudge G. Sivik 1 of 8 ODN TITLE: Installation & Operation Manual for Nautilus G2 Connectors

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

AIR-OPERATED DOUBLE DIAPHRAGM PUMP USER S MANUAL

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

MODEL ANALOG ALTIMETER USER'S MANUAL 330KHZ, 1000M DEPTH RATED 0.2 TO 100FT OPERATING RANGE ANALOG OUTPUT

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

750 Cable Cutter INSTRUCTION MANUAL

709 and 710 Stud Punches

Micro Dial-Flowmeter. Instructions for Use

SCUBAPRO. Balanced Power Inflator

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

Pressure Regulator. Instructions for Use

SAFETY MEASURES. These important guidelines should always be followed to work safely with the PRIMATECH pneumatic stapler model Q180:


MUD 1. Vent check valves. Instruction and maintenance manual.

HPB25 Hydraulic Paving Breaker

623 Series Dome Load Regulators

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6

6A.004 SCBA MAINTENANCE. Policy Number. Page 1 of 5 SCOPE

TECHNICAL DATA. Q = C v P S

DC5A. Cyclone Separator Trap for Air. Copyright 2015 by TLV Co., Ltd. All rights reserved ISO 9001/ ISO MA-03 (DC5A) 19 June 2015

Instructions for Use. Plug connection for medical gas and vacuum acc. to DIN / 2004 DIN EN ISO

Poppetted Coaxial Drop Tube Assembly for use with Auto Limiter II. Installation Instructions

Auto-Rewind Hose Reels INSTRUCTION MANUAL FOR OXY-LPG MODEL

TECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY

TactAir PCP Fill Kit/Adapter. Operation Manual

Oxygen Dialflow Meter. Instructions for Use

MANUAL: Jumbo Siamese and Jumbo Wye

Transcription:

Operation and Maintenance Manual for GENTEC Medical Gas Outlets Genstar Technologies Co., Inc. 4525 Edison Avenue Chino, CA 91710 USA TEL 909-606-2726 FAX 909-606-6485 www.gentechealthcare.com

IMPORTANT SAFETY INSTRUCTIONS. READ AND UNDERSTAND THESE INSTRUCTIONS COMPLETELY BEFORE OPERATING THIS EQUIPMENT. If you do not understand any of these instructions, or if you have any questions regarding the use of this product, please contact your facility s training manager, your supervisor, the medical equipment dealer from whom the product was purchased, or the manufacturer before operating the equipment. Do not attempt to repair this device if you have not been properly trained. Doing so may create a hazardous situation that may result in death or serious injury. Carefully inspect and test this product before each use to ensure proper operation. Do not use the product if there are signs of damage or if it does not pass the initial inspection. It is imperative that your hands and all tools used be clean and free of any oil or other contaminants before working on these products. Wear appropriate protective clothing when working on medical gas and vacuum products and systems. Outlets may be under pressure, use care when disassembling. Do not stand in front of the outlet when removing the latch assembly. Genstar Technologies Co., Inc. (GENTEC ) manufactures four types of medical gas latch valve assemblies: Ohio (Ohmeda) compatible (also known as pin-indexed) Chemetron compatible (also known as positive-latch) Puritan-Bennett compatible (also known as geometrically keyed) DISS (Diameter Index Safety System). Each latch valve assembly will accept only the appropriate adapter style. GENTEC also manufactures six types of rough-in assemblies: Wall outlet with 90º pigtail Ceiling outlet with 180º pigtail Console outlet with 90º pigtail Console or column outlet with 180º pigtail Console outlet with 90º hose barb Console outlet with 90º DISS 1

The rough in assembly for any particular gas service will accept any of the four styles of latch valve assembly (i.e., any oxygen rough-in will accept Ohio, Chemetron, DISS or Puritan-Bennett oxygen latch valve assemblies). OUTLET MAINTENANCE Outlets should be checked for proper latching and unlatching at each use. Flow rates and pressures should be checked periodically, at least once per year. Testing should be carried out in accordance with NFPA 99 Health Care Facilities, CSA Z305.1-92 Nonflammable Medical Gas Piping System and state and local codes, by authorized personnel only. The outlets (rough-in and latch valve assemblies) are shipped from the factory cleaned for oxygen use, in individual sealed bags. The installed latch valve assembly and trim plate may be cleaned with a mild detergent or wiped with a disinfectant that is compatible with the powder coated wall trim plate, silk screened aluminum faceplate, and exposed ABS plastic surfaces. O-rings should be lubricated annually with a very small amount of oxygen-safe silicone lubricant. DO NOT USE OIL OR PETROLEUM-BASED PRODUCTS AT ANY TIME ON ANY MEDICAL GAS OR VACUUM PRODUCTS. When testing outlets, make sure there are no leaks, with and without the adapter inserted. Make sure that the outlet is properly latching on the adapter. If the latch mechanism will not properly latch, make sure that you are using the proper adapter by inspecting for the proper gas service and connection style. Old, worn, or damaged adapters may cause leaks, latch improperly, or create other unsafe conditions. Make sure your adapters are in good condition. Adapters for most patient care products (flowmeters, suction regulators, etc.), as well as the products themselves, are available from your GENTEC dealer. 2

Rough-in Exploded View (console shown, others similar) Ohio Compatible Latch Valve Assembly 3

DISS Latch Valve Assembly Chemetron Compatible Latch Valve Assembly 4

Puritan-Bennett Compatible Latch Valve Assembly MISCELLANEOUS PARTS Description Part # Trim Plate for Wall Outlet 304760640 Adapter, Ohio x 1/8 MNPT 3101-XXX-NT2 Adapter, DISS x 1/8 MNPT 3102-XXX-NT2 Adapter, Chemetron x 1/8 MNPT 3103-XXX-NT2 Adapter, P-B x 1/8 MNPT 3104-XXX-NT2 NOTE: Substitute Gas Service for XXX above as follows: Oxygen OXY Vacuum VAC Medical Air AIR Nitrous Oxide N2O WAGD WAG Nitrogen NIT Carbon Dioxide CO2 5

Latch Assembly & Adapter Identification Adapter Latch Assembly Ohio (Ohmeda) Compatible DISS (Diameter Index Safety System) Chemetron Compatible Puritan-Bennett Compatible 6

Genstar Technologies Co., Inc. 4525 Edison Avenue Chino, CA 91710 USA TEL 909-606-2726 or 800-333-0811 FAX 909-606-6485 or 800-999-1478 www.gentechealthcare.com Form No. 9-OM-MGO Copyright Genstar Technologies Nov 2007