no peaks missing peaks too small peaks increasing retention decreasing retention

Similar documents
6890 GC Site Preparation

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Zebron Gas Management

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Mass Spec will not Autotune

Experiment GC : Analysis of BTEX by GC-FID

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Retention Time Locking: Concepts and Applications. Application

Super-Clean Gas Filters from Restek

Multiple Gas#5 GC configuration Jan 2016

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Best Practice for Identifying Leaks in GC and GC/MS Systems

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Running HPLC Methods on an ACQUITY UPLC System

Carrier Gases in Capillary GC

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

GCMSD-Headspace Analysis SOP

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Agilent 7820A Gas Chromatograph

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

CORESTA RECOMMENDED METHOD Nº 67

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

7820A GC Site Preparation Checklist

Rapid and Reliable Detection of Dissolved Gases in Water

6890 GC Not Ready Messages (A16010)

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

5890II GC Standard Operating Procedure 9/2/2005

Video Notes. LC Troubleshooting Series Retention Time Shifts

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

Setting a New Standard in Flash Chromatography Performance

Carrier Gases in Capillary GC

3 Flow and Pressure Control

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Your local gas generation partner. Precision series Modular gas generation solution for GC.

The instrument will process the samples one by one till all the samples have been cleaned up.

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Simple and Worry Free Gas Filters

Tips and Tricks of Injector Maintenance. Abby Folk Agilent GC Technical Support August 2009

Agilent G3180B Two-Way Splitter Kit

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Advances in Capillary Column Backflushing

Generating Calibration Gas Standards

Instant Connect Helium Saver Injector Module

Agilent Dimension Software for ELSD User Manual

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Agilent G Rebuilt Electron Capture Detector

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Buck Scientific, Inc.

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Your local gas generation partner. Precision series Modular gas generation solution for GC.

CP-3800 GC Operator s Manual

GC9310 GAS CHROMATOGRAPH USER S MANUAL

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

Thermo Scientific Model 146i Principle of Operation

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Inert Air (N2) Systems Manual

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

FROG-4000 TM Chemical Analysis System

New improved robot for gas kinetics

Products Designed to Solve Nagging GC Problems

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

INSTRUMENT INSTRUMENT YL INSTRUMENT

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

FROG-5000TM Chemical Analysis System User s Manual

6850 Series Control Module User Information

Standard Test Method for Residual Acrylonitrile Monomer Styrene-Acrylonitrile Copolymers and Nitrile Rubber by Headspace Gas Chromatography 1

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

SOP: Buck Scientific BLC-20P HPLC Operation

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

Technical Bulletin AL-134

PerkinElmer Aanalyst100

ISO INTERNATIONAL STANDARD

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

System Overview. TCD Detector temperature setpoint Regulator

Transcription:

GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering peaks spikes, ghost peaks strong noise, waves peaks on hill fronting, tailing broad peaks cut tops negative peaks double peaks poor resolution baseline shift 1.Observation no peak 1. error in the detector power supply / electronics. 1. check detector/electronic power supply and cables. 2. no FID flame. 2. check FID; reignite it. 3. syringe defective / clogged. 3. use a different syringe or clean it. 4. temperature too low for the analytes, oven heating faulty. 5. detector / software / computer hardware failure. 4. check temperature program, oven temperature. 5. check integrator, cables; restart computer. 6. no gas flow. 6. check gas tubes, valves, seals; test gas flow; shorten front of GC column, change injection septum. 7. column connection leaks. 7. use new ferrules. 8. broken GC column. 8. if breakage is at the beginning or at the end, remove the short piece; breakage in the middle can be mended with a glass

connector; for multiple breakages: replace column. 2. Observation: Missing or overlapping peaks, poor separation efficiency 1. syringe defective / clogged. 1. use a different syringe or clean it. 2. sample too diluted. 2. increase injection volume; concentrate sample. 3. sample concentration too high. 3. decrease injection volume; dilute sample. 4. column connection leaks, column not properly installed 5. perforated injection septum. 5. replace septum. 4. check column installation; search for leaks; replace ferrules. 6. injector temperature too low. 6. check temperature program; increase injector temperature. 7. sample decomposes in the injector. 7. check temperature program; reduce injector temperature; replace liner; check capillary ends. 8. column oven too hot. 8. check temperature program, oven temperature (external thermometer); decrease temperature. 9. incorrect flow rate. 9. measure flow, control and adjust it if necessary. 10. column absorbs or decomposes analytes. 10. check capillary ends; check intact poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film c) use column with special selectivity 3. Observation: Peaks too small, poor quantification, concentrations not reproducible 1. dirty syringe. 1. use a different syringe or clean it. 2. concentration of sample too low. 2. increase injection volume; concentrate sample. 3. split too high. 3. reduce split. 4. sensitivity of detector too low. 4. inject standard in order to test detector

5. column connection leaks, column not properly installed. sensitivity. 5. check column installation; search for leaks; replace ferrules. 6. injector temperature too low. 6. check temperature program, increase injector temperature. 7. dirty ECD. 7. clean ECD. 8. FID, TCD gas flow too low. 8. correct flow according to manufacturers instructions. 9. sample decomposes. 9. check capillary ends; check intact poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film c) use column with special selectivity. 4. Observation: Increased or differing retention times / low gas flow 1. speed of gas too low. 1. increase flow. 2. column connection leaks, column not properly installed. 2. check column installation; search for leaks; replace ferrules. 3. oven temperature too low or unstable. 3. check temperature program, oven temperature (external thermometer); if the analytes are stable, increase temperature. 4. strong decrease of gas pressure. 4. replace septum; for an instrument with pressure / temperature control, flow pressure must be higher than 15 psi above the demand at max. temperature of the program. 5. tubes / capillaries / column constricted or blocked. 5. compare flow at column entrance and outlet with preset flow; check and / or clean gas tubes; in case of pressure buildup cut and remove one turn (20 cm) from the column or replace column. 5. Observation: Decreased or differing retention times

1. speed of gas too high. 2. oven temperature too high. 3. column length too short. 4. film thickness in column too low. 1. compare flow at column entrance and outlet with preset flow; check gas tubes and pressure gauge; control parameter settings; or replace column. 2. check temperature program, oven temperature (external thermometer); decrease temperature. 3. replace column. 4. replace column. 6. Observation: Constantly declining baseline 1. gas flow changes with temperature gradient. 2. contaminated gas/ poor gas quality (at constant inlet pressure). 1. check gas content in gas cylinder; pressure must be a few bar above the required pressure at max. temperature; otherwise exchange gas cylinder. 2. check gas supply. 3. column not properly installed. 3. check column installation (FID). 7. Observation: Constantly rising baseline 1. leak at column entrance or injection septum. 1. check column installation; search for leaks; replace ferrules. 2. injector contaminated. 2. make a run at lower injector temperature; if the baseline improves, replace liner, use low bleed or high temperature septa. 3. column contaminated. 3. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column. 4. detector contaminated. 4. clean detector. 5. increase of temperature too fast. 5. decrease temperature gradient and end temperature. 6. poor gas quality. 6. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC. 8. Observation: Increasing baseline, at high temperature bleeding or noise

1. decomposition of the stationary phase. 1. check for leaks; matrix check for compatibility with the column. 2. column contaminated. 2. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column. 3. increase of temperature too fast / end temperature too high. 3. decrease temperature gradient and end temperature. 4. column not properly conditioned. 4. condition column according to manufacturers instructions (while column is not connected to the detector). 5. detector contaminated. 5. clean detector according to manufacturers instructions. 6. poor gas quality. 6. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC. 9. Observation: Plateaus at certain temperatures 1. steps in temperature program too drastic. 1. avoid very short and strong heating periods. 10. Observation: Regular interfering peaks 1. poor gas quality. 1. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC. 2. FID: dust or contaminants in the detector. 3. electronic defect, damaged cable or detector. 2. clean detector; if particles are visible in the column or column ends are not cut precisely (frayed edges), cut two turns from the column entrance. 3. replace cable, contact your GC manufacturer. 4. bleeding of silicon septa. 4. replace injection septum, use low bleed or high temperature septa.

11. Observation: Irregular interfering peaks, spikes, ghost peaks 1. contamination from vials / septa or sample preparation. 1. control SPE and / or autosampler vials; use low bleed or high temperature septa. 2. derivatization not quantitative. 2. check derivatization protocol; use more reactive derivatization reagents. 3. dirty syringe. 3. use a different syringe or clean it. 4. sample decomposes. 4. check temperature program, oven temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner. 5. column absorbs or decomposes analytes. 6. sample volume too high, double injection. 5. check capillary ends; check intact poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film c) use column with special selectivity. 6. reduce sample volume or add a blank run after a high volume injection. 7. poor gas quality. 7. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC. 12. Observation: Strong noise, waves 1. leak at column entrance or injection septum. 2. bleeding of septum / injector contaminated. 1. check column installation; search for leaks; replace ferrules. 2. make a run with lower injector temperature; if the baseline improves, replace liner, use low bleed or high temperature septa. 3. septum particles in column entrance. 3. cut 1 turn from column entrance; replace injection septum. 4. column contaminated. 4. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column.

5. column not properly conditioned. 5. condition column according to manufacturers instructions (while column is not connected to the detector). 6. hardware defect. 6. check temperature program, oven temperature (external thermometer); contact your GC manufacturer. 7. detector contaminated (electronics). 7. clean detector according to manufacturers instructions; check electronics. 8. increase of temperature too fast. 8. decrease temperature gradient and end temperature. 9. poor gas quality. 9. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC. 13. Observation: Small peaks on fronting or tailing of bigger peaks 1. column not properly installed. 1. check capillary ends; check tight and correct fit in injector and detector. 2. temperature of injection too low. 2. check injector temperature; if analytes are stable, increase temperature. 3. solvent not compatible with GC phase. 3. change solvent. 4. splitter defect. 4. measure flow and adjust splitter. 5. poorly deactivated column, film thickness too low. 5. check capillary ends; check intact poor results shorten both column ends by about 10 cm; or replace column. 14. Observation: Fronting (strong broadening in the ascending part) Fronting 1. column overload. 1. decrease injection volume; dilute sample. 2. sample vaporizes too slowly, not evenly or condenses. 2. increase injector temperature (consider max. temperature limits of the column). 3. analytes coelute. 3. change temperature program or use column with different selectivity. 4. sample decomposes. 4. check temperature program, oven

5. column absorbs or decomposes analytes. temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner. 5. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film c) use column with special selectivity. 15. Observation: Tailing (strong broadening in the descending part) Tailing 1. sample vaporizes too slowly, not evenly or condenses. 1. increase injector temperature (consider max. temperature limits of the column). 2. high-boiling analytes. 2. derivatize polar, basic or highboiling compounds. 3. system leaks. 3. check column installation; search for leaks; replace ferrules. 4. analytes coelute. 4. change temperature program or use column with different selectivity. 5. sample decomposes. 5. check temperature program, oven temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner by a deactivated one. 6. column absorbs or decomposes analytes. 6. check capillary ends; check intact poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film c) use column with special selectivity. 7. split rate too low. 7. increase split rate. 8. analytes always tending to tail. 8. no chance for symmetric peaks. 9. column overload. 9. decrease injection volume; dilute sample.

16. Observation: Broad peaks 1. poor focussing. 1. decrease start temperature of the program. 2. flow too high or too low. 2. measure flow, control and adjust it if necessary. 3. split rate too low. 3. increase split rate. 4. column overloaded. 4. decrease injection volume, dilute sample or increase split flow. 17. Observation: Cut tops of peaks, broad peaks 1. detector overloaded. 1. decrease injection volume; dilute sample; increase the split flow. 2. column overloaded. 2. decrease injection volume; increase split flow. 3. zero point is outside the display. 3. change scale. 18. Observation: Negative peaks, negative signals 1. polarity of integrator is inverted. 1. invert polarity at the instrument. 2. column overload. 2. decrease injection volume; dilute sample / increase split rate. 3. pressure fluctuations. 3. check gas tubes, valves, seals; test gas flow; change injection septum; contact hardware manufacturer. 4. detector contaminated. 4. clean detector as specified by the manufacturer. 5. electronic artefacts. 5. check detector, A/D converter.

19. Observation: Double peaks, doubled peak tops 1. solvent and column not compatible. 1. change solvent or use guard column. 2. solvent mixtures with large differences in boiling point and polarity. 2. use just one solvent. 3. sample decomposes. 3. check temperature program, oven temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner by a deactivated one. 4. analytes coelute. 4. modify temperature program or use longer column; possibly change column polarity. 5. detector overload. 5. inject less; control make-up flow. 20. Observation: Short lifetime, poor resolution, lack of separation efficiency 1. impurities on the column. 1. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column. 2. contamination from vials / septa or sample preparation. 2. check SPE and / or autosampler vials; use low bleed or high temperature septa. 3. polymerization on the column. 3. use guard column (at least 10 m). 4. separation efficiency decreases for repeated injections, improves after reconditioning. 5. temperature too high / temperature increase too fast. 4. use guard column; reduce injection volume; modify temperature program; for repeated injections increase end temperature (if possible) and use longer temperature program. 5. decrease oven temperature and / or temperature gradient (should not be higher than 25 C/min). 6. cooling too fast. 6. do not open oven door at high temperatures. 7. temperature too low / condensation. 7. increase injector temperature and/or start temperature. 8. poor deactivation. 8. check capillary ends; check intact poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects:

a) use a column with thicker film c) use column with special selectivity. 9. oxygen / air in the system. 9. use oxygen absorber or gas grade with less oxygen. 10. water content too high. 10. reduce water content. 11. head-space analysis: permanent air injections. 11. displace oxygen from vials with an inert gas. 21. Observation: Baseline increases or decreases before or after a peak 1. injection volume too high. 1. decrease injection volume; dilute sample; clean injection system. 2. column bleeding due to poor conditioning. 2. condition column according to manufacturers instructions (while column is not connected to the detector). 3. pressure fluctuations. 3. check gas tubes, valves, seals; test gas flow; change injection septum; contact hardware manufacturer. 4. injector temperature too low. 4. check injector temperature; if the analytes are stable, increase temperature. 5. injection septum perforated. 5. replace septum. 6. wrong TCD gas flow. 6. adjust flow according to manufacturers instructions.