Validation of SEC/GPC Software

Similar documents
User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Service Calibration are you doing it properly?

Application Note AN-107

CORESTA RECOMMENDED METHOD N 6

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

INTERNATIONAL STANDARD

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

The University of Hong Kong Department of Physics Experimental Physics Laboratory

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

ANALYSIS OF AERODYNAMIC CHARACTERISTICS OF A SUPERCRITICAL AIRFOIL FOR LOW SPEED AIRCRAFT

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007

The Estimation Of Compressor Performance Using A Theoretical Analysis Of The Gas Flow Through the Muffler Combined With Valve Motion

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Retention Time Locking: Concepts and Applications. Application

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

TPM TIP. Oil Viscosity

HASTAC High stability Altimeter SysTem for Air data Computers

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Physical Analysis Model Report

PIQCS HACCP Minimum Certification Standards

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

POWER Quantifying Correction Curve Uncertainty Through Empirical Methods

Legendre et al Appendices and Supplements, p. 1

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

VITLAB micropipette. Standard Operating Procedure (SOP)

Traceable calibration of automatic weighing instruments in dynamic operation

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

Completion of Hydrometric Survey Notes (HSN)

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Alpha Series. Main Features > Data displayed Selected speed r.p.m Selected spindle SP

Safety Manual VEGAVIB series 60

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004

Transferpette -8/-12

EVALUATION OF ENVISAT ASAR WAVE MODE RETRIEVAL ALGORITHMS FOR SEA-STATE FORECASTING AND WAVE CLIMATE ASSESSMENT

Safety Manual VEGAVIB series 60

Agilent Dimension Software for ELSD User Manual

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

GCMSD-Headspace Analysis SOP

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Technical Datasheet ZHA Series Gear Flow Meters Aluminum Version with ball bearing for lubricating media

Gas mixtures. Individual solutions specifically for your application

FIG: 27.1 Tool String

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

Fungilab s new branch in New York (US)

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

In addition to reading this assignment, also read Appendices A and B.

ISO/TR TECHNICAL REPORT. Natural gas Calibration of chilled mirror type instruments for hydrocarbon dewpoint (liquid formation)

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

North Sea Flow Measurement Workshop th October 2017

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

GAS CHROMATOGRAPHY MAINTENANCE USING UNCERTAINTY BASED CBM

WIND CONDITIONS MODELING FOR SMALL WIND TURBINES

Introduction to Roundabout Analysis Using ARCADY

Pressure Measurement Unit HEMP

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

EasyOne Pro LAB Measurement Technology Background

Presented to the International Technical Rescue Symposium, November Abstract

Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers

ZHM ST Series. Technical Datasheet. Gear Flow Meters for lubricating, non lubricating, abrasive and corrosive fluids

Chapter 11. Recall: States of Matter. Properties of Gases. Gases

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Name: Class: Date: SHORT ANSWER Answer the following questions in the space provided.

Real-time PCB Gas Monitoring System (MOHMS-21GP)

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates

Novel empirical correlations for estimation of bubble point pressure, saturated viscosity and gas solubility of crude oils

Transferpette. Lightweight precise rugged reliable!

PEAPOD. Pneumatically Energized Auto-throttled Pump Operated for a Developmental Upperstage. Test Readiness Review

Study on the Influencing Factors of Gas Mixing Length in Nitrogen Displacement of Gas Pipeline Kun Huang 1,a Yan Xian 2,b Kunrong Shen 3,c

Evaluation of the Mass Technology Precision Mass Measurement System on Bulk Field-Constructed Tanks (120,000 Gallon Vertical Tank Evaluation)

GMC-I Service GmbH Calibration Center

7 th International Conference on Wind Turbine Noise Rotterdam 2 nd to 5 th May 2017

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

Surface Texture Gage study in the qs-stat Measurement System Analysis Module

Ozobot Bit Classroom Application: Boyle s Law Simulation

Fluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

A Flexible, Multi-Purpose, Multi-Detector HTS Automated Platform

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Lab 1. Adiabatic and reversible compression of a gas

A SEMI-PRESSURE-DRIVEN APPROACH TO RELIABILITY ASSESSMENT OF WATER DISTRIBUTION NETWORKS

Setting a New Standard in Flash Chromatography Performance

Transcription:

Validation of SEC/GPC Software M. Gray, P. KiIz*, G. Reinhold* PSS-USA, 13531Cedar Creek Lane, Silver Spring, MD 20904 *PSS Polymer Standards Service GmbH, POB 3368, D-55023 Mainz Poster presented at the International GPC Symposium, San Diego, 1996

Introduction In many laboratories certification (ISO 9000) and accreditation (EN 45000) plays a dominant role. GLP and GMP regulations also require validated analytical procedures. The ISO/EN/DIN SEC/GPC standard confirmity is important, but cannot replace complete and systematic software validation. Due to the complextity of todays software systems, it is very difficult for laboratory managers to proof to the authorities that their data acquisition systems handle analytical data precisely and adaquately.

Method The following software validation methods are used at PSS to proof and document the proper functionality of the PSS WINGPC software! traditional software validation procedures! world-new PSS AUTOVAL validation process

Method Pro Con traditional each step analyzed time consuming directed debugging data flow dependent not tranferable to user no installation validation independend validation for each test step PSS verifies on-site installation degugging difficult AUTOVAL repeated on-site user tests complete integral analysis tests data flow paths tests software interactions indicates computer problems incorporation into system suitibility user training option easy testing of modified routines

The PSS AUTOVAL Validation Method 1. Step Creation of raw data (c(v), P(V), I (V) and M(V)) h from theoretical molar mass distributions (with well-known molecular parameters) 2. Step Import of validation raw data into the PSS WINGPC Software. Test of all software features on-the-fly 3. Step Comparison of the theoretical molar mass distribution with the software-calculated results (numerical values and distribution curves). The results must be identical (taking numerical errors into account) for the test to be successful.

Results Validation of WINGPC Calculations with Direct Calibration Run Conditions and Data Processing Parameters Polymer Type Eluent Temperature Calibration K [ml/g] a Polystyrene THF 25 C Polystyrene 0.01363 0.714 W(log M) 1.2 1.1 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 10 3 10 4 10 5 10 6 10 7 g/mol Mol mass Fig. 1: Molar mass distribution using conventional calibration as calculated by PSS WINGPC (the theoretical dsitribution is not shown, since no differences are visible) PSS WinGPC scientific V 3.0 b-36 M [D] n theoret. calculated error [%] 150000 150300 0.200 M w [D] 300000 300070 0.023 M z [D] 450000 450010 0.002 D 2 1.997 0.150 M v [D] 280869 280230 0.228 [η] [ml/g] 105.68 105.69 0.009 c [g/l] 4.343 4.343 0.000 Very good agreement of all parameters between theory and software testing Minor deviations caused by limited number of data-points (52).

Testing GPC Viscometry Setups (Example: Diff. Viscometer) Run Conditions and Data Processing Parameters Polymer Type Eluent Temperatur Univ. Calibration K [ml/g] a e Polystyrene THF 25 C PMMA 0.0129 0.688 η sp [η] R g g 0.0225 ml/g nm Viscosity 0.0200 0.0175 0.0150 0.0125 0.0100 0.0075 0.0050 0.0025 0.0000 10 2 10 1 10 2 10 1 1.06 10 0 1.14 1.12 1.10 1.08 1.04 1.02 10 4 10 5 10 6 g/mol molecular mass UE Fig. 2: Molar mass dependance of spezific and intrinsic viscosity, radius-of-gyration (R g) and branching (g ) in GPC-Viscometry calculations in PSS WINGPC PSS WinGPC scientific V 3 0 b-36 theoret. calculated error [%] M n [D] 150000 150110 0.073 M w [D] 300000 298230 0.590 M z [D] 450000 447220 0.618 D 2 1.987 0.650 M v [D] 278997 278510 0.175 [η] [ml/g] 105.68 105.69 0.001 K [ml/g] 0.01298 0.0137 0.514 a 0.688 0.714 0.000 c [g/l] 4.343 4.343 0.000

Testing GPC Light-Scattering Setups (example: LALLS) Run Conditions and Data Processing Parameters Polymer Type Eluent Temperature λ [nm] h [grd] n dn/dc [ml/g] 0 Polystyrene THF 25 C 633 6 1.403 0.187 0.16 0.14 0.12 0.10 0.08 0.06 0.04 0.02 10 6 10 5 10 4 10 5 10 6 g/mol molecular mass BE Fig. 3: Light scattering intensity and molar mass by LALLS detection, and comparison of M w(ls) vs. fitted (interpolated) molar mass in the LS-module of PSS WINGPC; straight line indicates perfect fit of measured M w(ls) PSS WinGPC scientific V 3.0 b-36 theoret. calculated error [%] M n [D] 150000 150700 0.467 M w [D] 300000 300240 0.080 M z [D] 450000 450220 0.049 D 2 1.992 0.400 M v [D] 278997 278560 0.157 c [g/l] 4.343 4.343 0.000 Very good agreement of all parameters between theory and software testing Minor deviations caused by limited number of data-points (52).

Testing Combined GPC-Viscosity-LS Configurations (example shown: diff. viscometer with RALLS) Run Conditions and Data Processing Parameters Polymer Type Eluent Temp. Calibration λ [nm] h [grd] n dn/dc K [ml/g] a 0 Polystyrene THF 25 C PMMA 633 6 1.403 0.187 0.01298 0.688 0.16 0.14 0.12 0.10 0.08 0.06 0.04 0.02 10 6 10 5 10 4 10 5 10 6 g/mo molecular mass UE Fig. 4: Molar mass distribution using GPC with viscometry and RALLS detection; measured molar masses with dyssymetry correction (linear) and without ( ) using the triple detector calculations in PSS WINGPC Software PSS WinGPC scientific V 3.0 b-36 theoret. Visco-LS error [%] M n [D] 150000 150680 0.453 M w [D] 300000 299760 0.080 M z [D] 450000 447410 0.576 D 2 1.989 0.550 M v [D] PMMA 278997 278510 0.175 [η] [ml/g] 105.68 105.76 0.076 K [ml/g] 0.01363 0.0137 0.514 a 0.714 0.714 0.000 c [g/l] 4.343 4.343 0.000 Very good agreement of all parameters Minor deviations by limited number of data (52) Expected M LS deviation at about 100000 D Accurate M with viscometer correction

Influence of Detector Noise on GPC Calculations 5% statistical noise on all raw data in worst case scenario results obtained without smoothing and despiking in software LALLS RALLS IV RI 0.250 225 0.0200 0.009 0.225 0.008 200 0.0175 0.200 0.0150 0.007 175 0.175 0.006 0.0125 150 0.005 0.150 0.0100 125 0.004 0.125 0.0075 100 0.003 0.100 0.0050 75 0.002 0.0025 0.075 50 0.001 0.0000 0.050 25 0.000-0.0025 0.025-0.001 0-0.0050 Voltage [V] 10 12 14 16 18 20 Elution volume [ml] LALLS RALLS IV RI PSS WinGPC scientific V 3.0 b-36 Fig. 5: LALLS, RALLS, Viscometry and RI raw data shown with baselines. All signals have a 5% random noise (S/N: 20:1) to test the robustness of PSS WINGPC algorithms; baselines are set as shown at 9.4 ml and 20.3 ml; integration limits can be set separately (see further explanation in text)

theoretical conventional universal LALLS LS-Visko M n [D] 150000 131630 150200 169120 151040 M w [D] 300000 303310 306260 299670 294420 M z [D] 450000 477670 478210 427110 415290 D 2 2.304 2.039 1.772 1.949 M v [D] (PS) 280869 282240 285670 284030 276170 [η] [ml/g] 105.68 106.79 106.86 105.90 104.59 K [ml/g] 0.01363 n/a 0.01457 n/a 0.01295 a 0.714 n/a 0.708 n/a 0.719 c [g/l] 4.343 4.327 4.276 4.284 4.296 conventional calibration: underestimated M n, overestimated M z (about 10%) universal calibration: larger molar mass deviations good bulk-iv agreement Mark-Houwink coefficient surprisingly close to reality light scattering: big discrepancies: M n clearly overestimated polydispersity clearly underestimated improvement of the results: integration-limits set to significant data (S/N>3). This corresponds to a reduction of the calculated injected mass of about 1%, but improves precision dramatically

Summary All PSS WINGPC software packages and modules are validated and conform to the GPC/SEC standards ISO/CD 13885 and DIN 55672. With the efficient PSS AUTOVAL test procedure, the GPC-user has! more security in the selection of GPC software packages! better control of GPC calculation methods! time-savings for otherwise costly software validation! the security of correct data reduction.! the option of testing future or in-house calculation procedures! ideal for training users for optimzed baseline settings, smoothing etc The PSS AUTOVAL validation:! is independent on any preconditions! applicable for all software-algorithms! tests the complete flow of data! can be used the in-house as required