Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Similar documents
Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Instruction Manual. BZ7002 Calibration Software BE

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL

Columbus Instruments

Continuous gas analysis ULTRAMAT 23 Supplement 2 for Operating Instructions C79000-G52xx- C216-03

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Application Note AN-107

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

Overview. Front Panel: Keypad and Display

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

Manual of SF6 Comprehensive Tester

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

Advanced Test Equipment Rentals ATEC (2832)

Instruction Manual Dräger MSI P7 and MSI P7 plus

University of Minnesota Nano Center Standard Operating Procedure

HumiSys HF High Flow RH Generator

Generating Calibration Gas Standards

Standard Operating Procedure

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

RAM 4021-DPX Operation Manual

TR Electronic Pressure Regulator. User s Manual

Gas Diluter Portable Gas Dilution System User s Manual

Thermo Scientific Model 146i Principle of Operation

Instruction and Maintenance Manual

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

Standard Operating Procedure for H 2 S and SO 2 Converter

Using the UltraRAE. Firmware 2.35

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Pneumatic high-pressure controller Model CPC7000

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

AMT-Ex Dewpoint Transmitter

Pneumatic high-pressure controller Model CPC7000

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Technical Bulletin, Communicating with Gas Chromatographs

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

USER'S GUIDE. Vaisala CARBOCAP Carbon Dioxide Calibrator for the GMP220 series probes GM K220 U421EN-1.1

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Digital Vacuum Regulator

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Model Description

AUK Environmental Sensor. Determining the refractive index of the air. AUK Environmental Sensor

Transmitter CS 21 Operation Manual

KEM Scientific, Inc. Instruments for Science from Scientists

Model 130M Pneumatic Controller

Heat Pump Connections and Interior Piping

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

ACV-10 Automatic Control Valve

300 Held Drive Northampton Pa

Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer

Best Practice Guide, Servomex 2700

Continuous Gas Analysis ULTRAMAT 23 Supplement 1 for Operating Instructions C79000-G52xx-C Product Information

The Discussion of this exercise covers the following points:

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

STATE OF CALIFORNIA AIR RESOURCES BOARD AIR MONITORING QUALITY ASSURANCE VOLUME V AUDIT PROCEDURES MANUAL FOR AIR QUALITY MONITORING APPENDIX E

Bubble Tube Installations

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

8.1 Calibration of gas sensors

6200E Installation Requirements

FPG8601 Force Balanced Piston Gauge

NORDCALC Introduction... 2 Registration... 2 Flow Calculations tab Torque Calculation & Actuator Mounting Data tab... 21

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

RGC-IR Remote Gas Calibrator for IR400

AIR DRIVEN. Gas Boosters & SYSTEMS. Accepting VISA, MasterCard and American Express

RAM Operation Manual

CORESTA RECOMMENDED METHOD N 6

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Gas-Detection Instruments

Model PDT Dewpoint Transmitter

STANDARD OPERATING PROCEDURES

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

G200 Analyzer Range. Operating Manual

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Mass Spec will not Autotune

RAM 4021 Operation Manual

Improve Process Reliability

What I have learned about SF 6 gas testing.a Practical explanation

Transcription:

Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly accurate, reliable and stable quantitative gas monitoring system. It uses a measurement system based on the photoacoustic infra-red detection method, and is capable of measuring almost any gas that absorbs infra-red light. Gas selectivity is achieved through the use of optical filters. By installing up to five of these filters in the 1412, it can measure the concentration of up to 5 component gases and water vapour in any air sample. Although the detection limit is gas dependent, it is typically in the parts per billion (ppb) region. The accuracy of these measurements is ensured by the 1412 s ability to compensate for temperature and pressure fluctuations, water vapour interference and interference from other gases known to be present. Reliability of measurement results can be ensured by regular self-tests, which the 1412 performs. Figure D.1. Measurement system of the 1412. An interior pump draws air from the sampling point through the air-filter to flush out the old air in the measurement system and replace it with a new air sample. The new air sample is hermetically sealed in the analysis cell by closing the inlet and outlet valves (Figure D.1). Light from an infra-red light source is reflected off a mirror, passed through a mechanical chopper, which pulsates it, and then through one of the optical filters in the filter wheel. The gas being monitored selectively absorbs the light transmitted by the optical filter, causing the temperature of the gas to increase. Because the light is pulsating, the gas temperature increases and decreases causing an equivalent increase and decrease in the pressure of the gas (an acoustic signal) in the closed cell. Two microphones mounted in the cell wall measure this acoustic signal, which is directly proportional to the concentration of the monitored gas present in the cell. The filter 144

Page 2 of 19 wheel turns so that light is transmitted through the next optical filter, and the new signal is measured. The number of times this is repeated is dependent upon the number of gases being measured. The response time is down to approximately 13 seconds for one gas or water vapour, or approximately 40 seconds if five gases and water vapour are measured. The NH 3 s full scale is adjustable up to 2000 ppm. It has a lower detectable limit of 60 ppb. The INNOVA 1412 multi-gas analyzer is specified with a 1-second sampling integration time and fixed flushing time; 2 seconds for the chamber and 3 seconds for the tubing; and the response time for one single sampling cycle with NH 3, carbon dioxide and dew-point temperature measurements is approximately 22 seconds. The response time of the analyzer to step changes in gas concentrations was tested (Figure D.2). The analyzer was challenged with two NH 3 calibration span gases, 22.8 ppm and 60.8 NH 3 respectively in a N 2 balance (±2% accuracy) (Matheson Gas Products, Inc., Montgomeryville, PA) at flow rates of 8 L/min through two 80 ft long individual sampling lines. At the beginning of the test, the two sampling lines were full of N 2 gas. During the first two sampling cycles, the time taken for the readings to change from 0 ppm (N 2 gas) to 22.8 ppm was 10 cycles (220 seconds), and the time taken for the readings to change from 22.8 ppm (span gas) to 60.8 ppm was also 10 cycles (220 seconds). After the first two span gas changes, the time taken for the readings to change from 22.8 to 60.8 ppm or from 60.8 to 22.8 ppm within ± 2% difference was 4 cycles (88 seconds).the reason 10 cycles is required for the initial two changes is to purge the N 2 gas in the sampling lines. Once the sampling lines are full of the span gases, the NH 3 analyzer had a faster response time. It implied that separate sampling line should be used for each sampling location and air sample should be drawn continuously. Similarly, the time taken for the readings to change from 22.8 ppm to zero air was greater than 10 cycles when sampling lines were purged by zero gas. Thus, the results indicated that an 88-second sampling time (4 cycles) for the NH 3 analyzer would be sufficient to achieve 97% or better of the concentration value. 145

Page 3 of 19 INNOVA 1412 setup: 22 sec/sample (Chamber flush: 2 s; Tube flush: 3s; S.I.T.: 1 s) Percentage,% NH3,ppm Percentage,% ( Innova reading/gas concentration) 120 100 80 60 40 95 91 87 75 100 98 96 92 79 97 103 100 100 98 109 100100 99 100 96 90 98 100 96 81 109 110 101 102 102 100 101 98 101 96 100 96 111 102 100 70 60 50 40 30 20 NH 3 Concentration, ppm 20 10 0 11 0 16:27 16:31 16:35 16:39 16:43 16:47 16:51 16:55 16:59 17:03 17:07 17:11 Warming Up the Monitor Time,hour:min Figure D.2. Response of INNOVA 1412 intermittent exposure of 22.8, 60.8 ppm NH 3 (N 2 balance) and ambient air. Calibration The infra-red light source is very hot so the temperature in the analysis cell increases as calibration measurements proceed. Conditions within the cell tend to stabilize more quickly once the temperature inside the analysis cell is 15 C above the ambient room temperature. A period of 30-40 minutes is suggested to warm up the analysis cell before a calibration task is started. This will reduce the time required for calibration. The Basic Calibration Setup The general equipment required to perform the calibration is shown in Figure D.3. All the solenoid valves around the manifold (M1) will close using the manual control in the LabView 7 program. The tubing from the cal-gas cylinder will be connected to the flow meter which is mounted on the side of the instrument rack and connected to the sampling manifold (M2). 146

Page 4 of 19 INNOVA 1412 Flow meter API 101E M2 Exhaust Cal-gas Figure D.3. General equipment required for a calibration task Producing a Supply of Clean, Wet Air Bubble zero-gas through a thermostatically controlled water-bath to produce a supply of clean, wet air to the Monitor during humidity interference calibration of the filters (see Figure D.4). The temperature of the water bath should be at least 2 C below the ambient temperature of the room where calibration is to be performed. N2 gas To Flow meter Water Figure D.4. Schematic diagram of the equipment necessary to produce a supply of clean, wet air. Setting the Communication Parameters The communication parameters necessary for the monitor to communicate with the Gas Monitoring Software 7304 are shown below: Baud rate 9600 Stop bits 1 data bits 7 Parity Even Hardwire mode Leased line Handshake type Hardwire 147

Page 5 of 19 To prevent communication errors, the text line terminator, print data log and print error log must be set as shown below: Text line Terminator Print Data Log Print Error Log CR-LF NO NO Setup CONFIGURATION System General Test Self test Yes Print Data Log NO Print Error Log NO Calibration Gases Zero air, CO 2 in N 2 and NH 3 in N 2 are used to calibrate the instrument. The certified calibration gases will consist of zero air (99.999% accuracy, Acid Rain CEM zero), CH 4 in N 2 (NIST traceable, EPA Protocol, ±1% accuracy), Propane in N 2 (NIST traceable, EPA Protocol, ±1% accuracy), CO 2 in N 2 (NIST traceable, EPA Protocol, ±1% accuracy), NH 3 in air or nitrogen (manufactory standard, certified plus grade, ±2% accuracy). Zero and Span Calibration Procedure for Analyzer Only Change the setting of 1412 communication parameters. Turn off the 1412 and disconnect 1412 with the Compact Fieldpoint module. Reconnect the 1412 to the COM1 port of the PC. Turn on the 1412. Open the Calibration program and create a new calibration task. Detach the air inlet tube from the existing sampling manifold. Close the pipe adapter of the sampling manifold with a cap or plug. Attach the 1412 to a calibration manifold. Zero (N 2 ) Gas Calibration 1. Open the Calibration program and create a New task. 2. Type in the desired task name (TYSON ##_MMDDYYYY) and click OK. 3. Pull down the Sequence menu. Click Settings. 148

Page 6 of 19 4. Click on the Calibration index-card, if it is not already at the front. 5. Click to select Zero point radio-button. 6. Click on the Gas index-card and click in the Sample Integration Time field, and select the desired time: 5 s for Ammonia, 1 s for CO 2, 1 s for Nitrous oxide (N 2 O), 1 s for propane and 5 s for CH 4. 7. Click on the Sampling index-card and set the flushing time to desired time: Fixed time, 2 s for Chamber and 3 s for Tube. 149

Page 7 of 19 8. When all the settings are correct, click on OK. 9. Pull down the Sequence menu and click on Start. 10. Insert the 1/4 ID tubing (from N 2 gas regulator) to the flow meter mounted on the side of the instrument rack; then open the regulator valve to allow gas flow. Zero gas is now flowing from the cylinder to the 1412. 11. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the flow meter mentioned above). This provides a little extra zero air to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 12. Record time and analyzer display in lab notebook. 13. Continuously inject the gas for 5 min after display is stabilized. 14. Pull down the Sequence menu and click on Stop. 15. Close regulator and remove Zero gas tubing. 150

Page 8 of 19 Humidity Calibration 1. Pull down the Sequence menu. Click Settings. 2. Click on the Calibration index-card, if it is not already at the front. 3. Click to select Humidity Calibration. 4. Only Select Humidity Interference Calibration (All Filters). Do not select Perform Water Vapor Span Calibration. 5. Click on the Gas index-card and click in the Sample Integration Time field, and select the desired time as above. 6. When all the settings are correct, click on OK. 7. Pull down the Sequence menu and click on Start. 8. Connect a water bath device between the zero gas regulator and the flow meter mounted on the side of the instrument rack; then open the regulator valve to allow gas flow. Zero gas is now flowing from the cylinder to the 1412. 9. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the flow meter mentioned above). This provides a little extra zero air to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 10. Record time and analyzer display in lab notebook. 11. Continuously inject the gas for 5 min after display is stabilized. 12. Pull down the Sequence menu and click on Stop. 13. Close regulator and main valve and remove zero gas tubing. Span Gas Calibration (NH 3 ) 1. Pull down the Sequence menu. Click Settings. 2. Click on the Calibration index-card, if it is not already at the front. 151

Page 9 of 19 3. Click to select Gas Span Calibration radio-button. 4. Click in the Active filter field Ammonia and select the Ammonia, and check Perform Cross Interference Calibration and input the concentration of NH 3 gas. 5. Click on the Gas index-card and click in the Sample Integration Time field, and select the desired time as above. 6. When all the settings are correct, click on OK. 7. Pull down the Sequence menu and click on Start. 8. Insert the 1/4 ID tubing (from the NH 3 gas cylinder) to the flow meter mounted on the side of the instrument rack; then open the regulator valve to allow gas flow. NH 3 gas is now flowing from the cylinder to the 1412. 9. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the flow meter mentioned above). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 10. Record time and analyzer display in lab notebook. 11. Continuously inject the gas for 5 min after display is stabilized. 12. Pull down the Sequence menu and click on Stop. 13. Close regulator and main valve and remove tubing from the NH 3 gas cylinder. Span Gas Calibration (CO 2 ) 1. Pull down the Sequence menu. Click Settings. 2. Click on the Calibration index-card, if it is not already at the front. 3. Click to select Gas Span Calibration radio-button. 4. Click in the Active filter field Ammonia and select the Carbon Dioxide, and check Perform Cross Interference Calibration and input the concentration of CO 2 gas. 152

Page 10 of 19 5. Click on the Gas index-card and click in the Sample Integration Time field, and select the desired time. 6. When all the settings are correct, click on OK. 7. Pull down the Sequence menu and click on Start. 8. Connect the 1/4 ID tubing (from the CO 2 gas cylinder) with a Nafion tubing, and then connect to the manifold; then open the regulator valve to allow gas flow. CO 2 gas is now flowing from the cylinder to the 1412. 9. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the flow meter mentioned above). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 10. Record time and analyzer display in lab notebook. 11. Continuously inject the gas for 5 min after display is stabilized. 12. Pull down the Sequence menu and click on Stop. 13. Close regulator and main valve and remove tubing from the CO 2 gas cylinder. Span Gas Calibration (Propane) 1. Pull down the Sequence menu. Click Settings. 2. Click on the Calibration index-card, if it is not already at the front. 3. Click to select Gas Span Calibration radio-button. 4. Click in the Active filter field Ammonia and select the THC (propane), and check Perform Cross Interference Calibration and input the concentration of Propane gas. 153

Page 11 of 19 5. Click on the Gas index-card and click in the Sample Integration Time field, and select the desired time. 6. When all the settings are correct, click on OK. 7. Pull down the Sequence menu and click on Start. 8. Connect the 1/4 ID tubing (from the Propane gas cylinder) to the flow meter mounted on the side of the instrument rack; then open the regulator valve to allow gas flow. Propane gas is now flowing from the cylinder to the 1412. 9. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the flow meter mentioned above). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 10. Record time and analyzer display in lab notebook. 11. Continuously inject the gas for 5 min after display is stabilized. 12. Pull down the Sequence menu and click on Stop. 13. Close regulator and main valve and remove tubing from the Propane gas cylinder. Span Gas Calibration (CH 4 ) 1. Pull down the Sequence menu. Click Settings. 2. Click on the Calibration index-card, if it is not already at the front. 3. Click to select Gas Span Calibration radio-button. 4. Click in the Active filter field Ammonia and select the CH 4 (Methane), and check Perform Cross Interference Calibration and input the concentration of methane gas. 154

Page 12 of 19 5. Click on the Gas index-card and click in the Sample Integration Time field, and select the desired time. 6. When all the settings are correct, click on OK. 7. Pull down the Sequence menu and click on Start. 8. Connect the 1/4 ID tubing (from the Methane gas cylinder) to the flow meter mounted on the side of the instrument rack; then open the regulator valve to allow gas flow. Methane gas is now flowing from the cylinder to the 1412. 9. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the flow meter mentioned above). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 10. Record time and analyzer display in lab notebook. 11. Continuously inject the gas for 5 min after display is stabilized. 12. Pull down the Sequence menu and click on Stop. 13. Close regulator and main valve and remove tubing from the Methane gas cylinder. Span Gas Calibration (N 2 O) 1. Pull down the Sequence menu. Click Settings. 2. Click on the Calibration index-card, if it is not already at the front. 3. Click to select Gas Span Calibration radio-button. 4. Click in the Active filter field Ammonia and select the Nitrous Oxide, and check Perform Cross Interference Calibration and input the concentration of N 2 O gas. 5. Click on the Gas index-card and click in the Sample Integration Time field, and select the desired time. 6. When all the settings are correct, click on OK. 7. Pull down the Sequence menu and click on Start. 8. Connect the 1/4 ID tubing (from the N 2 O gas cylinder) with a Nafion tubing, and then connect to the manifold; then open the regulator valve to allow gas flow. N 2 O gas is now flowing from the cylinder to the 1412. 155

Page 13 of 19 9. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the flow meter mentioned above). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 10. Record time and analyzer display in lab notebook. 11. Continuously inject the gas for 5 min after display is stabilized. 12. Pull down the Sequence menu and click on Stop. 13. Close regulator and main valve and remove tubing from the N 2 O gas cylinder. Calculating Calibration Factors With the raw measurement data displayed on screen: 1. Open the Cursor Values dialogue. 2. Use two cursors and the statistical data are displayed in the Cursor Values dialogue to locate a suitable range of data. All values in the select interval should be very stable and the temperature should be above 40 C. 156

Page 14 of 19 3. When you have the desired region between the cursors, pull down the Sequence menu and click on Mark Interval. The two cursors are replaced by a pair of green lines. These lines have markings at the end of them, which show the type of calibration data lying between the lines. The markings are: ZP: shows a zero point calibration HI: shows a humidity interference calibration SHA: shows a span calibration (filter A) SHB: shows a span calibration (filter B) SHC: shows a span calibration (filter C) SHD: shows a span calibration (filter D) SHE: shows a span calibration (filter E) 157

Page 15 of 19 4. Select and mark all the intervals for every calibration. 5. Pull down the values menu and click on calculate. 6. When the calculation(s) is complete a Calculation Finished dialogue is displayed. To download the calculated values 1. Pull down the Values menu. Click on Download and the Download dialogue is displayed. 2. Select the Zero Point index card 3. Set ticks in the NH 3, CO 2, Propane and Methane and Water Vapor check boxes. 4. Repeat steps 2 and 3 in the Humidity and Gas index cards. 5. Click on OK. The calibration factors are now downloaded to the monitor. Post Calibration 1. Check and close main valves on all cylinders 2. Calibration of the 1412 monitor is complete 3. Turn off the 1412 and disconnect 1412 with the PC. 4. Reconnect the 1412 with compact Fieldpoint and turn on the 1412. 158

Page 16 of 19 5. Set the filter sampling integration time back to original setup: 1 s for each filter. Setup Configuration Filters Filter A YES 1s S.I.T Filter B YES 1s S.I.T Filter C YES 1s S.I.T Filter D YES 1s S.I.T Filter E YES Water Filter YES 1s 6. The communication parameters necessary for the monitor to communicate with the compact Fieldpoint are shown below: Baud rate 9600 Stop bits 2 data bits 7 Parity Even Hardwire mode Leased line Handshake type Hard wired Print data log and print error log must be set as shown below: Text line Terminator Print Data Log Print Error Log CR-LF Yes No Setup CONFIGURATION System General Test Self test Yes Print Data Log Yes Print Error Log NO Routine Checking Zero and Span Check Procedure for Analyzer Only Detach the air inlet tube from the existing sampling manifold. Close the pipe adapter of the sampling manifold with a cap or plug. Attach the 1412 to a calibration manifold. Zero Gas Check 1. Insert the 1/4 ID tubing (from the NH 3 gas cylinder) into the manifold connected to the valve; then open the regulator valve to allow gas flow. Zero gas is now flowing from the cylinder to the 1412. 2. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of the monitoring flow meter). This provides a little extra zero air to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 3. Close regulator and remove tubing from the zero gas cylinder after display is stabilized (typically 5 to 10 minutes). Span Gas Check (NH 3 ) 1. Insert the 1/4 ID tubing (from the NH 3 gas cylinder) into the manifold connected to the valve; then open the regulator valve to allow gas flow. NH 3 gas is now flowing from the cylinder to the 1412. 2. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of mounted on the side of the instrument rack). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 159

Page 17 of 19 3. Close regulator and main valve and remove tubing from the NH 3 gas cylinder after display is stabilized (typically 5 to 10 minutes). Span Gas Check (CO 2 ) 1. Connect the 1/4 ID tubing (from the CO 2 gas cylinder) with a Nafion tubing, and then connect to the manifold; then open the regulator valve to allow gas flow. CO 2 gas is now flowing from the cylinder to the 1412. 2. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of mounted on the side of the instrument rack). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 3. Close regulator and main valve and remove tubing from the CO 2 gas cylinder after display is stabilized (typically 5 to 10 minutes). 4. Check and close main valves on all cylinders. Span Gas Check (CH 4 ) 1. Insert the 1/4 ID tubing (from the CH 4 gas cylinder) into the manifold connected to the valve; then open the regulator valve to allow gas flow. CH 4 gas is now flowing from the cylinder to the 1412. 2. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of mounted on the side of the instrument rack). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 3. Close regulator and main valve and remove tubing from the CH 4 gas cylinder after display is stabilized (typically 5 to 10 minutes). 4. Check and close main valves on all cylinders. Span Gas Check (Propane) 1. Insert the 1/4 ID tubing (from the Propane gas cylinder) into the manifold connected to the valve; then open the regulator valve to allow gas flow. Propane gas is now flowing from the cylinder to the 1412. 2. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of mounted on the side of the instrument rack). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 3. Close regulator and main valve and remove tubing from the Propane gas cylinder after display is stabilized (typically 5 to 10 minutes). 4. Check and close main valves on all cylinders. Span Gas Check (N 2 O) 1. Connect the 1/4 ID tubing (from the N 2 O gas cylinder) with a Nafion tubing, and then connect to the manifold; then open the regulator valve to allow gas flow. N 2 O gas is now flowing from the cylinder to the 1412. 2. Adjust regulator valve until vent airflow is about 2.5 L/min (read from bottom of ball of mounted on the side of the instrument rack). This provides a little extra gas to the 1412 and keeps the pressure inside the manifold close to the atmospheric pressure. 160

Page 18 of 19 3. Close regulator and main valve and remove tubing from the N 2 O gas cylinder after display is stabilized (typically 5 to 10 minutes). 4. Check and close main valves on all cylinders. Manufacturer Contact Information Innova AirTech Instruments Energivej 30 DK-2750 Ballerup, Denmark Fax: (+45) 44 20 01 01 innova@innova.dk www.innova.dk Reference 1412 Photoacoustic Field Gas-Monitor Technical Document, INNOVA Air Tech Instruments. 161

Page 19 of 19 Maintenance/Calibration Record Sheet for INNOVA 1412 Multi-Gas Monitor Date of Calibration Calibrated by: Time Items Unit = Notes Ambient pressure, (mm Hg) : : Sample airflow (L/min) : : Zero air applied --- Cylinder P: psi : : NH 3 reading : : CO 2 reading N 2 O reading CH 4 reading Propane reading : : CO 2 ( ppm) applied (with Nafion tubing) --- Cylinder P: psi : : CO 2 reading : : NH 3 ( ppm) applied --- Cylinder P: psi : : NH 3 reading : : CH 4 ( ppm) applied --- Cylinder P: psi CH 4 reading : : Propane ( ppm) applied --- Cylinder P: psi Propane reading : : N 2 O ( ppm) applied --- Cylinder P: psi N 2 O reading : : New calibration Yes / No : : Zero air applied : : Zero humidity air applied : : NH 3 applied : : CO 2 applied (with Nafion tubing) : : Propane applied : : CH 4 applied : : N 2 O applied : : Download to bank Yes / No : : Time switch off --- : : Connect analyzer back to sampling system. Note: 162