Using the UltraRAE. Firmware 2.35

Similar documents
QRAE II Quick Reference Covers Diffusion & Pump Models with Firmware Version 3.12 and higher

Tiger Select. Instrument User Manual V2.7

RAM 4021-DPX Operation Manual

PERSONAL GAS MONITOR MODEL GM460

ACV-10 Automatic Control Valve

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Vers Gas Generator. Release III INSTRUCTION MANUAL. Advanced Calibration Designs, Inc.

Do Not Print This Page.

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021 Operation Manual

EVM Series: Frequently Asked Questions

Columbus Instruments

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Heavy Duty Dissolved Oxygen Meter

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Met One E-BAM Particulate Monitor

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

NIV EAU MATIC ST User Manual Made in Canada

User manual CF8-D/W-IN

Overview. Front Panel: Keypad and Display

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

QUICK START GUIDE TO DRINKING WATER MONITORING

SPECIFICATIONS APCEPH1

Atlas 6TM. Indoor Air Quality Monitor Instruction Manual. Measures CO 2, Temperature and Humidity DISTRIBUTED BY

CSA Sample Draw Aspirator Adapter Operator s Manual

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

G100 RANGE G100, G110, AND G150 OPERATING MANUAL

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

CONSOLE-320 ENGLISH. 230A: CONSOLE-320 with cable data output Item 230B: CONSOLE-320 with cable + wireless radio data output

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

MP15 Jockey Pump Controller

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Instruction Manual Dräger MSI P7 and MSI P7 plus

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

300 Held Drive Northampton Pa

XpressFill XF2500 Counter Pressure Filler. Operating Instructions

Operating Instructions Part No

RK-IR Sample Draw Aspirator Adapter Operator s Manual

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION

G100 Range. Operating Manual

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Operation and Maintenance of the EPV200 Portable Ventilator

Portable Gas Monitor GX Operating Manual

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

FROG-4000 TM Chemical Analysis System

PGA3510 Portable 3-Gas IR Analyzer. Operations Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual

FROG-5000TM Chemical Analysis System User s Manual

Sterisil Ac+ Installation & Operating Manual

Manual Weighingblock VB2 series and Uniscale

HumiSys HF High Flow RH Generator

COMPRESSED AIR SAVING PRODUCTS INDEX. Useful information AIR-SAVER G1 AIR-SAVER G2 LOCATOR. Condensate Management Specialist

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

User Manual. Pro CO LP Alarm. Carbon Monoxide Analyzer and Low Pressure Monitor with Alarm REV: 03.19

DURRIDGE STAINLESS STEEL SOIL GAS PROBE

FROG-4000 TM Chemical Analysis System

KANE455. Flue Gas Analyser with direct CO 2 measurement. Stock No: October Kane International Ltd

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev

G200 RANGE G200 AND G210 OPERATING MANUAL

RAM 9025-XP Operation Manual

Instruction Manual. Biogas Analyzer InCa Bio 04

Nuclear Associates

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

Gas-Detection Instruments

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Portable Toxic Gas Monitor SC Operating Manual (PT0-105)

G200 Analyzer Range. Operating Manual

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Operator Quick Guide ORBISPHERE 3654

Napa Technology Trouble Shooting. For Premier & Premier PLUS Models

Generating Calibration Gas Standards

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Altimeter and Compass Watch Instruction Manual

Operation & Maintenance Manual GDS-68XP

In Vivo Scientific, LLC INSTRUCTION MANUAL

SDM-2012 Docking Station Standalone Configuration Operator s Manual

Installation Instructions

P-TRAK Ultrafine Particle Counter Model 8525

Operating Instructions Part No

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

G200 Analyser Range. Operating Manual

DPC-30 DPC-100. Reference Manual

Digital Vacuum Regulator

Transcription:

Using the UltraRAE Firmware 2.35

Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes Datalogging Programming Mode Calibration Bar Code Reader

UltraRAE Features Combination photo ionization detector (PID) with vapor specific filter tubes 0 2000 ppm VOCs 0 200 ppm Benzene 0 200 ppm Halocarbons 0 200 ppm Butadiene

UltraRAE Features UltraRAE Measures trace benzene in a complex fuel mixture Benzene Amount of Compound Separation Time

UltraRAE Features Bar-coded RAE-SEP tubes absorb interfering vapors and let the desired vapor through Measure total VOCs continuously or Benzene intermittently (1 minute sampling time) 400 cc/min pump draw makes remote sampling easy Sample draw at least 50 feet Pump stall feature: when moisture is detected or when pump is blocked the pump will shut off, protecting the UltraRAE from potential damage Runs 10 hours with pump

UltraRAE Features Temperature range of 32 F to 113 F (0 C to 40 C) Must Adjust Sample time Rechargeable, snap-in, field-replaceable NiMH battery pack or alkaline RFI protection against radio interference Intrinsically safe: Class I, Division I, Groups A, B, C, D

Regulator & Calibration Equipment Setting up the UltraRAE Charger Calibration Gas Alkaline battery pack RAE Sep Tubes PID Cleaning Equipment & Charcoal Filters

Setting up the UltraRAE Unpack the charger

Setting up the UltraRAE Plug the charger into the wall and then into the UltraRAE Charge the UltraRAE overnight

Setting up the UltraRAE To install alkaline batteries, pull off the rubber boot and open the battery compartment

Setting up the UltraRAE Remove the battery

Setting up the UltraRAE Unpack the alkaline battery pack and install the batteries

Setting up the UltraRAE Line up the battery pack Drop the battery pack in This shim fits snuggly above the battery pack

UltraRAE: Physical Description Probe Tube Tip Breaker & Reservoir Tube Holder and Bar Code Reader Buzzer Data Port Gas Exit Port Charger Port

UltraRAE: Physical Description Tip-Breaking Hole Probe Tube Holder Base

User Interface Three buttons on a sealed membrane faceplate Light Sensor Alarm LEDs Charge LED Y/+ (alarm): clears and tests alarms N/- (light): turns on manual backlight for 60 seconds (on)

Turning On the UltraRAE Unplug UltraRAE from charger Hold key to turn on Alarm will beep once Watch display screen for configuration messages. Warm-up will take approximately 30 seconds The display will read Ready

Calibrate the Monitor You must calibrate the UltraRAE before using it

Calibrate Monitor Each compound/tube type must be zeroed and calibrated independently For example, for Benzene, a separate zero and Benzene span must be done Calibrating VOC with isobutylene DOES NOT calibrate the benzene or other channel MUST calibrate with the gas of interest No Correction Factors can be used Can use the same tube to zero and span

Calibrate Monitor To calibrate the monitor you will need: Fresh air Calibration gas corresponding to the tube type Tedlar bag, open cup or demand flow regulator UltraRAE must use either a Tedlar bag, open cup, or demand flow regulator to calibrate Never calibrate with the cylinder with a fixedflow regulator directly connected to the inlet of the instrument Never calibrate with a surrogate gas

Calibrate the Monitor Unpack the Tedlar bag, regulator and calibration gas

Calibrate the Monitor Unpack the charcoal filters

Calibrate the Monitor Unpack the charcoal filters and assemble parts as shown Charcoal filter with VOC zeroing tube (for zeroing VOC Mode) Effluent Scrubber with RAE Sep Tube (To capture Benzene exiting unit while calibrating)

Calibrate the Monitor Before hooking up the Effluent Scrubber cut a 1.5-inch piece of tubing from the scrubber tubing.

Span Calibration Put this section of tubing on the inlet of the Tedlar bag

Span Calibration Fill Tedlar bag with span gas. CAUTION: Gas flow is 3 LPM DO NOT connect directly to UltraRAE Open stopper (push in) Open gas valve 1/2 turn

Span Calibration Close the Tedlar bag by pulling the stopper out

Calibrate the Monitor To avoid breathing benzene while calibrating, attach the Effluent Scrubber before span calibration Screw the effluent hose barb to the outlet port on the side of the unit Attach the Charcoal Effluent Scrubber to the hose barb

Calibrate the Monitor Take out a RAE-Sep Tube Break both ends open using the tip breaker

Calibrate the Monitor Remove the probe Insert the tube with arrow pointing towards unit Give ¼ twist to seat the tube tightly Screw the probe back on

Programming Mode The UltraRAE must be in Programming Mode to calibrate Hold and keys for 5 seconds to get into Programming Mode If UltraRAE asks a question? Answer or To accept or move on to the next option use the key Repeatedly pushing the key will always eventually return the UltraRAE to the main display

Calibrate Monitor? The calibration menu is as follows: Fresh air calib? Span calib? Modify span value?

Fresh Air Calibration You must be in a location with clean ambient air Otherwise hook the charcoal filter to the tip of the probe (more important for VOC than benzene)

Fresh Air Calibration Press until Calibrate Monitor? appears Press Display reads Fresh air calib? Press Confirm or correct the tube type and press Wait until display reads 0.0 Done! The next sub menu will display and read Span calib?

Span Calibration Connect the Tedlar bag to the inlet port of the UltraRAE and open the stopper Open Closed

Span Calibration Press Confirm or correct the tube type and press Span gas = xx ppm? Confirm that your span gas is of the right concentration by comparing the displayed value to the value on the cylinder

Span Calibration Wait for the monitor to count down to zero Disconnect the Tedlar bag Press to calibrate to another tube type Press or to abort more calibrations and move on to the next menu item

Programming Mode You can also use the Programming Mode to: Change the alarm limits Change the site ID Change user ID Change lamp type Change real time clock and more

Programming Menus Calibrate monitor? Change alarm limits? View or change datalog? Change monitor setup? Change sensor configuration?

Turning on the Bar Code Reader Press until the display reads Monitor Setup? The monitor setup menu is as follows: Change Site ID? Change User ID? Change User Mode? Change Date? Change Time? Change Backlight? Change DAC/Alarm output? Change DAC range? Change Bar Code Reader?

Turning on the Bar Code Reader Press until the display reads Change Bar Code Reader? Press The display shows the current reader status. Press to toggle between On? and Off? Toggle to desired status and press Press to Save? When the bar code reader is on, the UltraRAE automatically identifies the tube type and applies the calibration and measurement time for the tube identified

Adjusting the Bar Code Reader Adjusting the Bar Code Reader may be necessary when: A new type of tube is used It gets dirty or misaligned with use A new batch of tubes is used

Adjusting the Bar Code Reader Start the unit by pressing the and keys simultaneously to enter Diagnostic Mode The display shows raw sensor readings. Press until CCD = [value] is displayed. With no tube CCD should read 64. If not, the bar code reader is damaged or misaligned. Insert a VOC tube; CCD should read 59-61. If not, use the key to adjust down or key to go up. Insert a benzene tube; CCD should read 47-52. If not, use the key to adjust down or key to go up. See TN-134 for more details

Idle Operation of the UltraRAE Now the sensor and pump are turned off (idle) and you can cycle through the idle operation menu: As you press The display will read: Last = [tube type], [reading in ppm] Change current tube type? [Battery voltage] [Date], [time], [temperature] PC Comm? Ready

Idle Operation of the UltraRAE Use the idle operation of the instrument to: Prep the instrument for taking a measurement Change current tube type Check Battery voltage Ready the instrument to communicate with the PC

UltraRAE Tubing Never use Tygon tubing! Absorbs chemicals like a sponge Reduces ppm readout when chemicals exist Causes false positives when chemicals don t exist Always use Teflon or similar non-reactive tubing Will not absorb most chemicals Clean with anhydrous methanol if it gets dirty

External Prefilter Use the white external prefilter in high moisture environments like rain and saturated headspace sampling. Replace filter when it looks very dirty or when it introduces PID drift. Replace filter when in pump alarm with the filter on and you can clear the pump alarm with the filter off. This filter can reduce the UltraRAE s response to low vapor-pressure/high boiling compounds. When sampling for these chemicals it is recommended to periodically remove this filter to note a significant change in readings.

Preparing for a Measurement To take an accurate measurement Read and understand the Data Sheet Calibrate the instrument Set the instrument to the correct type of gas (and tube) Use the correct type of RAE-Sep tube Set the Measure Wait Time to match the ambient temperature The UltraRAE comes from the factory with the tube barcode reader turned OFF (This may change)

Preparing for a Measurement Make sure the instrument is set for the correct gas (and tube) Press the key until the display reads: Last = [tube type], [reading in ppm] Displays the gas/tube type the UltraRAE is set-up for Displays the last measurement

Preparing for a Measurement If the instrument is not set up for the correct gas/tube, you must change the tube type Press the key until the display reads: Change current tube type? Select from Butadiene VOCs Benzene Halocarbons Select by pressing until the desired tube type appears. Press to select the tube type

UltraRAE Sep Tubes Four types of RAE Sep Tubes are available Butadiene VOCs Benzene Halocarbons Use a new tube for each measurement. Used tubes may cause low or high readings. Benzene and Butadiene require a 9.8 ev photoionization lamp Halocarbon requires an 11.7 ev lamp You MUST use the correct lamp to take an accurate, reliable measurement

Preparing for a Measurement Check the Data Sheet for the correct Measure Wait Time and adjust in Programming Mode For Benzene: Temp ( C) 2-10 10-15 15-30 30-40 Temp ( F) 36-50 50-60 60-86 86-104 Measure Time (sec) 150 90 60 40 Sample Vol. (ml) 900 540 360 240

Inserting a Tube Press the key until the display reads Ready Break the ends off of a new tube using the tube tip breaker on the head of the instrument Insert the new tube with the arrow pointing into the instrument The display will then read Tube in [tube type] Start? Tube tip breaker Arrow pointing inward

Taking a Measurement Press The display reads: Log on (if datalogging is turned on) Wait and counts down Measure Wait Time (60 sec for benzene; no count down for VOC) Display will alternate between Done! and the gas reading shown in ppm Press display reads Remove tube! Remove tube, display reads Ready repeat the process for a new measurement

After a Measurement Be sure to remove tube after measurement Moisture gradually absorbing onto the tube can form a corrosive liquid that will damage the instrument if sucked in!! Do not re-use tubes because readings are not guaranteed

Alarms: Pre-set Alarm Limits The UltraRAE has pre-set alarm limits: Cal Gas Cal Low High Span Alarm Alarm Benzene 5.0 ppm 5.0 ppm 10.0 ppm Butadiene 5.0 ppm 5.0 ppm 10.0 ppm Halocarbon 10.0 ppm 10.0 ppm 25.0 ppm VOC 100.0 ppm 50.0 ppm 100.0 ppm

Alarms: Lamp Alarm 0.0 Lamp Lamp display along with audible alarm indicates that PID lamp has failed to light If after a few minutes the Lamp message remains, turn off UltraRAE and restart If after restart Lamp message disappears, UltraRAE is ready for use If after restart Lamp message remains, the PID needs service

Alarms: Pump Check 0.0 Pump Every time the UltraRAE is used it is important to check pump flow With a tube in place in the tube holder, block the probe inlet, the UltraRAE will go into alarm and display Pump Reset pump alarm by pressing the key If pump does not go into alarm, check for leaks in the probe or service pump

UltraRAE Datalogging The UltraRAE is pre-set to datalogging ON if that option is included in the purchased features Sample data contains Sample number, time, tube name and gas concentration No TWA or STEL can be recorded because of intermittent sampling The UltraRAE stores sample data in groups called events Stores start time, user ID, site ID, serial number, last calibration date, alarm limits A new events is created when the instrument is turned on, or a configuration parameter is changed (e.g., change tube type)

UltraRAE: Maintenance Clean PID Lamp & Sensor When display creeps upwards after good zero When PID responds to moisture When movement of PID results in response on display How to Clean PID Sensor Always clean sample probe and replace or clean filters FIRST! If PID holds a stable zero after this step then further cleaning may not be necessary Use anhydrous methanol (Lamp cleaning solution) Clean lamp face with lens tissue Clean sensor by immersion in cleaning solution (an ultrasonic cleaner is required for cleaning)

UltraRAE: Maintenance Drying the PID Sensor Let air dry overnight Warm air (not hot) will speed drying