STERILIZABLE DISSOLVED OXYGEN SENSORS

Similar documents
STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

OxyProbe II Instruction Manual

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

LogIT Dissolved Oxygen Probe

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

PRODUCT SHEET. Order probe only as RXPROBE02

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

DISSOLVED OXYGEN SENSOR BT34i

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

SS32L Dissolved O 2 Probe

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Oxymax H COS21D. Technical Information

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Best Practice for Calibrating LTH Conductivity Instruments

Dissolved Oxygen/Temp Waterproof Tester

Vinmetrica Dissolved Oxygen System User Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Oxygen Sensor with Electrode

Unisense In Situ UniAmp Connector System

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

& Dissolved Oxygen Sensor. Instruction Manual

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

General Specifications

Heavy Duty Dissolved Oxygen Meter

12mm Dissolved Oxygen Sensor Care and Use Instructions

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Oxygen Measurement Systems High Reliability and Wide Application Coverage

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

QUICK START GUIDE TO DRINKING WATER MONITORING

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

Dissolved Oxygen Sensor

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

Instruction Manual DO 6. Economy Hand-held Dissolved Oxygen Meter. 68X Rev 2 Apr 08. Technology Made Easy... Part of Thermo Fisher Scientific

Free Chlorine Sensor For additional information, please visit our website at

NEULOG OXYGEN LOGGER SENSOR GUIDE

UsER manual for Watersens ph -REDOX

ph measuring system for hygienic process applications

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

OWNER S MANUAL NMB SERIES

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

SensoDirect Oxi200 Dissolved Oxygen Meter Operating instructions

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

MODEL GT820 OXYGEN SENSOR

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

MULTICHANNEL. Labnet International, Inc.

Swell UK. The UK's leading online aquatic retailer.

A LAZAR LABORATORIES HANDBOOK DISSOLVED OXYGEN PRIMER

Quick Start Guide LIQ-QSG-499ACL-03, Rev I July Rosemount 499ACL-03. Monochloramine Sensor

Radnoti Liver Perfusion System

Transmitter CS 21 Operation Manual

Non-Membrane Disinfection Sensors Instruction Manual

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Dissolved Oxygen Probe (Order Code DO-BTA)

Dissolved Oxygen Guide

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

PERFORM Operating Document

PRO20 USER MANUAL. English. Français. Español. Deutsch

RayPette Plus Autoclavable Pipette. User Manual

NEW: CO2 incubator with sterilizable sensor

Optical Dissolved Oxygen Meter

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

T02.0.3X.6C-06 Page 1 of 6. Operating Manual for Dissolved Oxygen-Measuring transducer OXY 3610 MP

Instruction Manual MicroChem 2 Chlorine Probes Series CL4000

The pipettes cover a volume range from 0.1µl to 10ml.

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Oxygen Dipping Probe PSt3 / PSt6 / PSt9

Dissolved Oxygen Probe (Order Code DO-BTA)

HI 2314 HI 2315 HI 23151

Aerobic Respiration. Evaluation copy

HI 9143, HI 9143M HI 9145

EC214 EC215 - EC215R Bench Conductivity Meters

Quick Start Guide LIQ-QSG-499AOZ, Rev M July Rosemount 499AOZ. Dissolved Ozone Sensor

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES

PERFORM Operating Document

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Pipettor. User Manual

Optical Dissolved Oxygen Meter

Walchem Disinfection Sensors WCDB, WFCB, WOZB and WPAB Instruction Manual

NEW: CO2 incubator with sterilizable sensor

Instruction Manual DO 6+ Dissolved Oxygen/Temp

InFlow Modular perfusion bioreactor

Bioengineering RALF Advanced

Technical Information Oxymax COS22D/COS22

Transmitter CS 21 Operation Manual

Improve Process Reliability

Instruction Manual HI 9146 Dissolved Oxygen and Temperature Meter

User Manual DOCUMENT # Pro20 Pro20i USER MANUAL. English

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Membrane covered optical CO 2 sensor

Transcription:

STERILIZABLE DISSOLVED OXYGEN SENSORS USER MANUAL Table of Contents 1. Introduction 2. Operation 3. Calibration 4. Electrode Verification 5. Mounting 6. Maintenance 7. Storage 8. Instructions for Exchange of Electrolyte and Membrane 9. Ordering Information 10. Trouble Shooting

1. INTRODUCTION: The ADI do 2 sensors are used for measuring the partial pressure of dissolved oxygen in media for biotechnological processes. The sensors are available in two versions: a standard version with a stainless steel membrane tip and a low-drift version with a titanium membrane tip. Both types of do 2 sensors are available in different sizes for application in the head plate of a 2-20 liter autoclavable bioreactor and in the bottom DN25 insert of the 20-200 liter Bio Bench / Pilot In Situ reactors. Control of dissolved oxygen tension can be obtained in combination with the ADI 1030 (autoclavable systems) or the ADI 1060 (in situ systems) Bio Controller. The polarographic ADI do 2 sensor is based on the principle of a Clark-cell. It consists of a platinum working electrode (cathode) and a silver counter (reference) electrode (anode); the gas permeable membrane separates the electrode from the medium. The polarographic amplifier supplies a constant polarization voltage of - 675 mv. The oxygen molecules that diffuse through the semipermeable membrane (and the thin layer of electrolyte) are reduced at the cathode while the silver anode is oxidized simultaneously. Polarization causes a current to flow which is measured by the amplifier; this current is proportional to the partial pressure of oxygen in the medium. The slope of the O 2 sensor is temperature, pressure and medium dependent. Therefore calibration must be performed under operating conditions (same temperature, pressure and medium!). To enable automatic temperature correction the standard version sensor with the AppliSens logo on the connector is fitted with an integrated NTC resistor (k = -4%, R = 22 kohm @ 25 C, B25/50 = 3560) 2. OPERATION: The sensor comes with a protective cap to avoid mechanical damage of the sensor tip and prevent evaporation of the electrolyte; this cap must be removed before using the sensor. AppliSens sterilizable do 2 sensors are supplied factory-tested and certified with membrane module in place. When the system is operated for the first time or when the sensor has been disconnected from the controller for longer than 5 minutes, the sensor must be polarized (connected to the do 2 amplifier) for at least 6 hours prior to calibration. If the sensor is only disconnected for a couple of minutes, less than 6 hours will be sufficient. Insufficient polarization will cause an increased residual signal and therefore lead to measurement errors (see section 4: Electrode Verification). In case of autoclavable systems, the electrode cable must be removed before sterilization; cover the electrode connector with a cap or cotton with sterilizable (aluminum) foil! 3. CALIBRATION: If the measurement is performed under sterile conditions, the system should be calibrated after sterilization since sterilization may alter the sensor slope. After cooling, the fermentor must be aerated long enough to reach a constant oxygen pressure; the signal is allowed to stabilize and the reading of the amplifier adjusted to the desired value (e.g. 100 % air saturation, 20.9% oxygen saturation or e.g. 9.17 mg O 2 /l pure water at 20 o C and 760 mmhg). - In case of autoclavable systems, the sensor must be connected to the amplifier immediately after sterilization in order to polarize it; it is recommended to polarize the electrode for at least several (6) hours before calibration. Preferably calibration should take place just before inoculation (the longer the polarization period, the higher the measuring accuracy!). - Homogenize the medium by stirring during calibration in order to prevent a diffusion gradient. A "one point" calibration is normally enough to enable accurate measurements over the full range between 0 and 100% do 2 (at 0% do 2 the polarized sensor system does not generate any current). However, it is good practice to verify the measured signal at 0% do 2. Strip all oxygen from the medium by pure N 2 gas (99.98%) and follow the reading of the sensor until it is stabilized. In case the final reading is > 0.5% of air saturation, the sensor needs maintenance (refer to section 10). The sensor should be pre-sterilized once before calibrating in order to prevent measurement errors. After the first sterilization, new membrane modules may show a slope-alteration of some percent of the measuring signal; after a second sterilization only a fraction of the first alteration will be observed.

4. ELECTRODE VERIFICATION: It is good practice to verify the sensor before it is put to use. An elevated residual signal (signal at 0% oxygen saturation) indicates malfunctioning of the O 2 sensor. If the reading at 0% do 2 is too high (> 0.5% of air saturation) the sensor should be inspected; see section 10, Trouble Shooting. Verify the sensor by placing it in an oxygen-free environment, either in pure N 2 gas or in a solution from which all oxygen was stripped with N 2 (99.98%). - Do not use CO 2 gas for stripping oxygen from the calibration/verification solutions since this will introduce an error. After 5 to 10 minutes the reading should be stable. In case the bioreactor is sparged with N 2, take enough time to strip all oxygen from the solution. 5. MOUNTING: When mounting the sensor in the reactor (head plate or bottom insert), be careful not to damage it; both the membrane and inner body are fragile. The low drift type has a PG 13.5 thread that can be mounted into the DN25 port sensor holder for ph and DO electrodes for In Situ applications. 6. MAINTENANCE: Visually inspect the membrane for damage prior to each calibration. If the membrane is contaminated, clean it with a moist, soft cloth. Depending on the nature of the medium, the electrolyte may have to be changed periodically; it must be refreshed at least every six months. If the medium contains components that diffuse through the membrane (like acetic acid, aldehyde or neutral mercaptanes), the electrolyte must be refreshed more frequently! The membrane module must be replaced when it starts to show: - elevated residual signal, - sluggish response, - drifting or noisy readings, - mechanical damage, etc. (see section 10, Trouble Shooting). 7. STORAGE: Filled with electrolyte for the ADI do 2 sensor, the electrode can be stored for several months provided that the protective cap is covering the membrane module (to prevent evaporation of the electrolyte). During this period of time, the electrode can be kept polarized so it is always ready for use. For storage over longer periods, the sensor should be stored dry e.g. in the membrane module without electrolyte. 8. INSTRUCTIONS FOR EXCHANGE OF ELECTROLYTE AND MEMBRANE: WARNING: The O 2 electrolyte is strongly alkaline (ph = 13). Contact of electrolyte with skin, especially mucous membrane or eyes, must be avoided (use safety glasses and gloves). If such contact occurs, the affected area must be thoroughly rinsed with water. Get medical attention if adverse symptoms appear. When the membrane starts to shows signs of failure (refer to section 6, Maintenance), it must be replaced. When replacing the membrane module and electrolyte, strictly follow the instructions below: - Hold the sensor in a vertical position (membrane downwards) and unscrew the old membrane module, - Rinse the interior electrode body and the annular space between this and the lower part of the steel shaft with d.i. water and dry it carefully with a lint free tissue, - Visually inspect the O-ring for mechanical defects and replace if necessary, - Hold the membrane module with a tissue and fill it with 1 ml (40 drops) of electrolyte.

Hold the electrode vertically and screw the membrane module very slowly onto the shaft. - After refreshing electrolyte for the ADI do 2 sensor, the sensor has to be repolarized (see paragraph 2), - After replacing the membrane module, the sensor has to be sterilized, calibrated (see paragraph 3) and verified (see paragraph 4). When this instruction is not followed carefully, the membrane in the module may be damaged! 9. ORDERING INFORMATION: Normal autoclavable do 2 -sensor in different lengths, detachable cable (head plate application): Z010023510: Sensor Dissolved Oxygen L=235MM 1-5L = 12mm Z010038510: Sensor Dissolved Oxygen L=385MM 7-15L = 12mm Z010059010: Sensor Dissolved Oxygen L=590MM 20L = 12mm Z110200010: Cable D. Oxygen sensor L=2.0m BNC Z110200020: Cable DO sensor + NTC L=2.0m leads Normal In Situ sterilizable do 2 -sensor with detachable cable (DN25 port application): Z010007010 Z110380010 Z110500010 Z110500020 Sensor Dissolved Oxygen L=70mm DN-25 Cable D. Oxygen sensor L=3.8m BNC Cable D. Oxygen sensor L=5.0m BNC Cable DO sensor + NTC L=5.0m leads Low drift autoclavable do 2 -sensor in different lengths, detachable cable and titanium module (head plate application): Z010006521 Z010023520 Z010038520 Z010059020 Z100200010 Sensor Dissolved Oxygen L= 65 = 12mm 2P (fixed PG13.5 for 250L BioSep) Sensor Dissolved Oxygen L=235 1-5L = 12mm 2P Sensor Dissolved Oxygen L=385 7-15L = 12mm 2P Sensor Dissolved Oxygen L=590 20L = 12mm 2P Cable ph/do-2p sensor L=2.0m BNC Low drift In Situ sterilizable do 2 -sensor with detachable cable and titanium module (DN25 port application): Z010011020 Z101002510 Z100002520 Z010015420 Z100002530 Z100380010 Z100500010 Sensor Dissolved Oxygen L=110 = 12mm 2P for DN25 ph/do Sensor Holder or for Sensor holder DN25 A Tracfix Sensor Dissolved Oxygen L=154 = 12mm 2P for Sensor holder DN25 B Tracfix Cable ph/do-2p sensor L=3.8m BNC Cable ph/do-2p sensor L=5.0m BNC Accessories and spare parts: Z110000410 Z110000420 Z110005010 Membrane kit Dissolved Oxygen sensor (4 modules, electrolyte 50ml, 4 O-rings) Membrane kit Dissolved Oxygen sensor titanium (4 modules, electrolyte 50ml,4 O-rings) Electrolyte Dissolved Oxygen sensor 50ml

10. TROUBLE SHOOTING: PHENOMENON REASON VERIFICATION WHAT TO DO 1. Slope out of range (Slope >0.6 at 37 o C) a Short circuit in cable Check reading with connected cable but without sensor b Short circuit in sensor Check reading with detached module and carefully dried inner body c Ground loop Earthing potential difference between fermentor head plate and measuring device / controller Replace cable if reading more than 0.5 % of air saturation If reading more than 0.5% of air saturation, the sensor is defective Earth fermentor and measuring device at the same earth point d Damaged membrane Visually inspect the membrane Replace membrane module 2. No sensitivity a No polarization voltage Amplifier Verify polarisation voltage b Cable rupture Verify resistance of cable Replace cable c Not enough electrolyte in membrane module Visual inspection. Electrolyte must contact anode Refresh electrolyte solution d Membrane completely fouled Impermeable layer on membrane Clean membrane module or replace it e Malfunction in sensor Sensor Replace sensor 3. Sensitivity too low (Slope <0.3 at 37 o C) a Depleted electrolyte After long term measurements of more than 6 months b Contaminated anode Black silver ring (brown is o.k.) (e.g. caused by S 2- ) c Contaminated cathode Not visible (can only be identified by Replace electrolyte solution d Membrane fouled Layer on membrane Clean with soft tissue or replace module 4. Elevated residual signal (Verify with O 2 -free medium) a Short circuit in cable Check reading with connected cable but without sensor b Short circuit in sensor Check reading with detached module and carefully dried inner body c Ground loop Earthing potential difference between fermentor head plate and measuring device / controller Replace cable if reading more than 0.5 % of air saturation If reading more than 0.5% of air saturation, the sensor is defective Earth fermentor and measuring device at the same point d Damaged membrane Visually inspect the membrane Replace membrane module e Test with CO 2 gas instead of N 2 The do 2 sensor shows a CO 2 interference in absence of O 2 f Contaminated cathode Not visible (can only be identified by Use N 2 or Bisulphite solution (catalysed by Co(II)salt) g Sensor not polarized long enough Polarization < 6 hours Polarize more than 6 hours

PHENOMENON REASON VERIFICATION WHAT TO DO 5. Unexpected high reading during measurement a O 2 -free and CO 2 saturated medium CO 2 interference of do 2 sensor b Interference of air/o 2 bubbles May be observed with air lift reactors or in cultures of high cell density Place sensor in a region with less air/o 2 bubbles c Elevated pressure in the bioreactor E.g. caused by an increased flow resistance in sterile filter Avoid increased pressure 6. Sluggish response a Membrane fouled Layer on membrane Clean with soft tissue or replace module b Damaged membrane Visually inspect the membrane Replace membrane module c Contaminated cathode Not visible (can only be identified by d Very viscous medium Sluggish response is normal in case of a viscous medium 7. Unstable signal a Air/O 2 bubbles sticking on membrane May be observed with air lift reactors or in cultures of high cell density Change stirring and/or aeration conditions or change probe location b Damaged membrane Visually inspect the membrane Replace membrane module c Crack in glass body at the cathode Visual inspection. Unstable signal especially at low po 2 If crack in glass body, sensor is defective d Electrostatic interference Apply equipment according to certified conditions 8. Drift too large a After sterilization (In-Situ systems only) Time between cooling down and calibration too short Increase cooling interval b Membrane fouled Layer on membrane Clean with soft tissue or replace membrane module c Damaged membrane Visually inspect the membrane Replace membrane module d Contaminated cathode Not visible (can only be identified by e Contaminated anode Black silver ring (brown is o.k.) (e.g. caused by S 2- ) f Ground loop Earthing potential difference between fermentor head plate and measuring device / controller Earth fermentor and measuring devices at the same point g Change of pressure in bioreactor E.g. caused by an increase of the flow resistance in sterile filter Avoid changes in pressure h Change of medium temperature Verify medium temperature Avoid changes in temperature 9. Membrane torn Too much electrolyte in the module Add only 1 ml (40 drops of electrolyte and screw the module very slowly onto the shaft.