Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem

Similar documents
Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Mass Spec will not Autotune

PE NexION/Elan Nebulizers

UsER manual for Watersens ph -REDOX

Agilent 1290 Infinity Pump Head Maintenance

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3

OPERATION. Estimated kerf width compensation. HPR260 Manual Gas Instruction Manual 4-9

Agilent Auxiliary Gas Module AGM 2. User s Guide

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Latvin Luxury Shower Panel. Telephone Product Specification. ~ Minimum Working Pressure 1.0 bar ~ Maximum Working Pressure 3.

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Chamber Test Testing Theory

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Installation, Operation, and Maintenance Manual

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

Mass Spec will not pump down

Halsey Taylor Owners Manual

Using the Akta Prime plus October 22, 2012

Aridus3 Desolvating Nebulizer System

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

1)! DO NOT PROCEED BEYOND THIS MARK

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

Halsey Taylor Owners Manual

Navigator 600 Silica analyzers

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

AIR DRIVE ROLL FEED CARTON STAPLER MODEL: H-1031 MANUAL

Armfield Distillation Column Operation Guidelines

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

5 Channel Calibrator

FINAL REVISED TECHNICAL SPECIFICATION OF INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER (ICP-MS) WILL BE AS FOLLOWS:- GENERAL SPECIFICATIONS

Thumb Shifter Plus Thumb Shifter

Met One E-BAM Particulate Monitor

Microfluidic Demonstration

Best Practice for Identifying Leaks in GC and GC/MS Systems

Standard Operating Manual

Flexfence Swaging Unit. Product Manual. Release 02/16.

STS ICP-RIE. Scott Munro (2-4826,

Halsey Taylor Owners Manual STOP!

Description Advanced Features Typical Applications

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

Basic ICP Operating Procedures

Technical Operating Instructions for Magnum Cutting Systems. for online instructions

Glove Box Installation Manual

User Manual. Quantos Automated Dosing Liquid Module

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Service and Repair Manual

Nordiko Metal Sputtering System Standard Operating Procedure

Assembling & Installing your Pool Cleaner

Table of contents. Table of contents 1. Installation 2. Introduction ( Display, monitor ) 3. Introduction ( Operating elements ) 4

Product Information News September 26, 2003

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

QUORUM TECH 150RES THE FIRST AND THIRD WEEK WILL BE SET UP FOR CARBON COATING THE SECOND AND LAST WEEK WILL BE SET UP FOR GOLD COATING

ASSEMBLY & OPERATIONS MANUAL. Straight & Clean Torch Cuts. Every Time.

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

Vapor Sorption Analyzer

Halsey Taylor Owners Manual

Usage Policies Notebook for STS DRIE System

Urgent Next visit Not urgent. All customers Selection of customers No customers

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

Flexi-pro Aerosol Delivery System (FADS) Manual

PANACEA P100 LOW PRESSURE USER MANUAL

Section 3- Repair and Calibration Procedures

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Customer Responsibilities

FX134, FX1234, FX3030 Problem and Solutions

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

BPS-2, Rev TABLE OF CONTENTS

Halsey Taylor Owners Manual 4410 Freeze Resistant Tubular Fountain STOP!

Float Operated Level Controllers

SS32L Dissolved O 2 Probe

New product release. Universal Rebreather Monitor (URBM) Single O2 cell and Dual HP package

Indian Institute of Technology Kanpur Samtel Centre for Display Technologies

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Addendum to the Waters Micromass Quattro Premier XE Mass Spectrometer Operator s Guide

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

Linkam Scientific Instruments USER GUIDE. LTS420 and LTS120. Temperature Controlled Stage

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

User s Guide. Vacuum Controller for liquid delivery systems

Thermo K-Alpha XPS Standard Operating Procedure

Helios, a CyTOF System

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

RM-80 respiration monitor

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Transcription:

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem Rev. B 2017, January, 20 Next >

Preconditions 7900, 7800, 7700, 8800, or 8900 ICP-MS is used. Sample introduction type is PeriPump. If you are using a special configuration such as LC/GC integration or ISIS, setup the instrument with standard configuration. You have the Hardware Maintenance Manual for your instrument. Some details of the operations are not in this tool. You should refer to the manual. <Previous Next >

Legends te Reference Good Bad tes or tips that help troubleshoot. You need to refer to another document for detail of the steps. Good example of situation. Bad example of situation. Did the ion lens test pass? Questions are displayed in blue you need to click buttons such as Yes or Based on your answer the tool will navigate you to next step. <Previous Next >

1.1.1 Check Water flow and Gas flow errors (1) Select MassHunter Instrument Control Window Main menu View Instrument Status to display Instrument Status Pane te If Main menu is not displayed, press F10 key. (2) Click the dropdown below and check the error number for errors that are generated at the date/time of the attempt to ignite plasma. (3) Are error numbers 1440, 1445, or 1446 included in the latest errors? <Previous Yes

1.2.1 Check water and argon gas (1) If error 1440 has been generated, check if cooling water valve is open and chiller or heat exchanger is working correctly. (2) If error 1445 or 1446 have been generated, check if argon gas valve is open and remaining amount is enough. (3) Did this solve the problem? Yes

1.3.1 Check timing when Plasma goes out (1) When did Plasma go out? t ignite at all Several seconds after ignition One to several minutes after ignition When "Plasma Correction is going on 20 minutes or later after ignition

2.1.1 Check condition of argon gas line condition (1) Are there any situations below? - Ar cylinder has been replaced recently - Was there any work on Ar line recently? - Is the instrument used after a long time ( > 1 month). Yes

2.2.1 Check condition of argon gas line (1) Purge Ar line. (2) Do leak check on Ar line. (3) Check specification of Ar gas to have 99.99% purity (In case cylinder has been replaced). (4) Try alternative source of Ar gas (In case cylinder has been replaced and any alternatives are available) (5) Does this fix the problem? Yes

2.3.1 Isolate the torch and try to ignite (1) Block end of the torch with plastic film and try to ignite. (2) Can it ignite? Yes

2.4.1 Check torch box (1) Open the torch box and observe to check if there are situations below, following instruction from the next page. - Damaged torch - Damaged bonnet - Damaged shield plate - Damaged shield contact - n Agilent torch - Dirty torch stand - Wrongly assembled torch <Previous Next >

2.4.2 Check torch box (2) Check torch Bad Bad Damaged torch or bonnet ( image) n Agilent torch ( image) <Previous Next >

2.4.3 Check torch box (3)Check if shield plate is damaged. Good Good shield plate. Bad Damaged shield plate. <Previous Next >

2.4.4 Check torch box (4) Check if shield/torch is correctly assembled. Good Set the projection on torch in the hole of the shield plate. Good Set the projection on torch in the notch of the holder. Good Good contact between shield plate and shield contact. <Previous Next >

2.4.5 Check torch box (5)Check shield contact. Good Shield contact in good condition. Bad Damaged shield contact. <Previous Next >

2.4.6 Check torch box (6) Check torch stand Good Torch stand with clean contact <Previous Next >

2.4.7 Check torch box (7) Is there any failure of torch, shield plate, shield contact, and torch stand? <Previous Yes

2.5.1 Torch box issues cannot be fixed by customers (1) Are there any situations below? - Damaged torch - Damaged shield contact Yes

2.6.1 Torch box issues can be fixed by customers (1) Are there any situations below? - Damaged bonnet - Damaged shield plate - n Agilent torch - Dirty torch stand - Wrongly assembled torch Yes

2.7.1 Fix torch box issues (1) Replace or maintain the parts. (2) Did this fix the problem? Yes

2.8.1 Check work coil shape (1) Check work coil shape. Reference Hardware Maintenance Manual Maintenance Torch area RF Coil (2) Any problem found? Yes

2.9.1 Fix work coil shape (1) Fix work coil shape. Reference Hardware Maintenance Manual Maintenance Torch area RF Coil (2) Did this fix the problem? Yes

2.10.1 Check cooling water (1) Does the cooling water have any additive substance, such as antifreeze or antiseptic? Yes

2.11.1 Check cooling water (1) Replace water with pure water. (2) Did this fix the problem? Yes

3.1.1 Check photo sensor cable (1) Open torch box and remove torch. Reference Hardware Maintenance Manual Maintenance Torch area Torch Removal (2) Are there any situation below? - Cable for photo sensor is not correctly installed (not inserted into the hole on torch stand). - Cable for photo sensor is damaged. Bad Photo sensor cable installed not correctly. Yes

3.2.1 Check if photo sensor cable is damaged (1) Is the Cable for photo sensor damaged? Yes

3.3.1 Check if photo sensor cable is damaged (1) Insert the cable to the hole. (2) Did this fix the problem? Good Photo sensor cable installed correctly. Yes

4.1.1 Check IF/BK vacuum pressure errors (1) Select MassHunter Instrument Control Window Main menu View Instrument Status to display Instrument Status Pane te If Main menu is not displayed, press F10 key. (2) Click the dropdown below and check the error number for errors that are generated at the date/time of the attempt to ignite plasma. (3) Are error numbers 1118, or 1119 included in the latest errors? Yes

4.2.1 Check sampling cone (1) Check if sampling cone has a distorted or expanded orifice. (2) Any trouble found? Good Sampling cone in good condition. te Distortions of cones are subtle. If it is not clear, regard it as Yes. Yes or not clear

4.3.1 Change sampling cone (1) Exchange sampling cone. Reference Hardware Maintenance Manual Maintenance Interface (2) Does this fix the problem? Yes

4.4.1 Check Analyzer vacuum pressure errors (1) Select MassHunter Instrument Control Window Main menu View Instrument Status to display Instrument Status Pane te If Main menu is not displayed, press F10 key. (2) Click the dropdown below and check the error number for errors that are generated at the date/time of the attempt to ignite plasma. (3) Are error numbers 1124, or 1129 included in the latest errors? Yes

4.5.1 Check skimmer cone (1) Check if skimmer cone has a distorted or expanded orifice. (2) Any trouble found? Good Skimmer cone in good condition. te Distortions of cones are subtle. If it is not clear, regard it as Yes. Yes or not clear

4.6.1 Change skimmer cone (1) Exchange skimmer cone. Reference Hardware Maintenance Manual Maintenance Interface (2) Does this fix the problem? Yes

4.7.1 Isolate the torch and try to ignite (1) Block end of the torch with plastic film and try to ignite. (2) Can it ignite? Yes

4.8.1 Check usage of User Tune (1) Is User Tune function in use? Yes

4.9.1 Check User Tune parameters (1) Modify tune parameters in user tune like below. Sampling Depth 8mm Carrier Gas flow MicroMist: 1.0L/min MicroFlow: 0.6L/min Make Up Gas flow MicroMist: 0L/min MicroFlow: 0.3L/min (2) Does this fix the problem? Yes

4.10.1 Re-adjust User Tune parameters (1) Check if the new tune parameters give good sensitivity. If not, adjust tune parameters so that it gives enough sensitivity. OK

5.1.1 Check dilution gas connection (1) Check dilution gas connection Bad loose nut. Push dilution gas line into connector. Confirm line is inserted into the bottom of connector. Confirm nut has been tightened. Next >

5.2.1 Check dilution gas connection (2) Did this fix the problem? With the other end of the connector, confirm connection in the same way <Previous Yes

6.1.1 Check spray chamber (1) Remove top cover of spray chamber chiller. (2) Is the spray chamber filled with liquid? Bad Spray chamber filled with liquid. Yes

6.2.1 Drain spray chamber (1) Pull out drain tube and drain liquid from the spray chamber Next >

6.2.2 Drain spray chamber (2) Remove torch. Then clean it with pure water and dry it. Reference Hardware Maintenance Manual Maintenance Torch area Torch Removal (3)Attach the torch again. Reference Hardware Maintenance Manual Maintenance Torch area Torch Installation (4)Check and fix drain tube Reference Hardware Maintenance Manual Maintenance Peristaltic Pump Area Peristaltic Pump Tubing Installation <Previous OK

6.3.1 Check Water flow and Gas flow errors (1) Select MassHunter Instrument Control Window Main menu View Instrument Status to display Instrument Status Pane te If Main menu is not displayed, press F10 key. (2) Click the dropdown below and the check error number for errors that are generated at the date/time of the attempt to ignite plasma. (3) Are error numbers 1440, 1445, or 1446 included in the latest errors? Yes

6.4.1 Check water and argon gas (1) If error 1440 has been generated, check if cooling water valve is open and chiller or heat exchanger is working correctly. (2) If error 1445 or 1446 has been generated, check if argon gas valve is open and remaining amount is enough. (3) Did this solve the problem? Yes

7.1.1 Air leak check (1) Check carrier gas connection. Push carrier gas line into connector. Push carrier gas line into connector (In case of glass nebulizer). Push connector toward nebulizer (In case of glass nebulizer). Next >

7.1.2 Air leak check Check if ferrule is correctly installed (In case of PFA nebulizer). Bad loose nut. Tighten the nut, if it is loose. (In case of PFA nebulizer). <Previous Next >

7.1.3 Air leak check (2) Check makeup gas connection. Bad loose nut. Push makeup gas line into connector. Confirm line is inserted into the bottom of connector. Confirm nut has been tightened. <Previous Next >

7.1.4 Air leak check (3) Check loose nebulizer. Good Correctly installed nebulizer Bad loose nut. Bad t inserted to the end. <Previous Next >

7.1.5 Air leak check (4) Check loose spray chamber cap. Check that the cap is not loose. (5) Did this fix the problem? <Previous Yes

8.1.1 Check rotary pump (1) Check and Maintain rotary pump. Reference Hardware Maintenance manual Maintenance Maintaining Other Parts Yes

9.1 End of procedure Close this window. Return to top

9.2 End of procedure To solve the problem an engineer from Agilent is needed. Please contact Agilent using the link below. http://www.agilent.com/en-us/contact-us/page Close this window. Return to top