Agilent 490 Micro GC Natural Gas Analyzer

Similar documents
Agilent Dimension Software for ELSD User Manual

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent Flowmeter ADM1000

ISO INTERNATIONAL STANDARD

Agilent 1290 Infinity Pump Head Maintenance

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Multiple Gas#5 GC configuration Jan 2016

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Technical Manual. Sensepoint XCD Gas Detector

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Flow Vision I MX Gas Blending Station

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

EnCal 3000 Gas Chromatograph Commissioning and Maintenance Manual Biogas and THT-Devices

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Continuous gas analysis ULTRAMAT 23 Supplement 2 for Operating Instructions C79000-G52xx- C216-03

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

Best Practice for Identifying Leaks in GC and GC/MS Systems

Describe Flammable Gas Measurement

Applied Technology and Best Practices in CEE. Conference

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

TABLE OF CONTENTS LIST OF FIGURES

This GC has two columns (I think new columns can be ordered from Restek)

ISO INTERNATIONAL STANDARD. Welding consumables Gases and gas mixtures for fusion welding and allied processes

Solubility of Gases in Water

GASES NATURAL GAS INDUSTRY SPECIALTY. Gases & Equipment

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Discovery HP-TGA 75/750. Site Preparation Guide

Gas Metrology. SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons

Retention Time Locking: Concepts and Applications. Application

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

SRI Multi Gas Analyzer 2016

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

800.Keen.Gas

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

APPLICATION NOTE. GC Integrated Permeation Device

Agilent G3180B Two-Way Splitter Kit

Nonrefillable Cylinders

Chapter 22 Testing Protocols

New improved robot for gas kinetics

System Overview. TCD Detector temperature setpoint Regulator

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

1 PIPESYS Application

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

FUNDAMENTALS OF GAS CHROMATOGRAPHY, GAS QUALITY & TROUBLESHOOTING Larry Ewing Chandler Engineering Company LLC

Agilent Inlet EPC Leak Test

Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors. User s Guide

Atmospheric Hazards. Hazardous Atmosphere 3/29/2017. Oxygen deficiency and enrichment Flammable gases and vapors Toxic contaminants

GasLab SAN-102 Micro Oxygen Monitor User Manual

Our Air- Why We Should Care!

Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED. A Division of Norco, Inc.

General Specifications

General Specifications

PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

GasLab SAN-102 Micro Oxygen Monitor User Manual

Operation of the Perkin Elmer TGA-GC/MS

Misuse of Combustible Gas Meters

Mass Spec will not Autotune

Instrument Performance Evaluation, comparing ISO and NORSOK I-106 Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk Analyse North Sea Flow

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

AKTA pure 25 New Owner s Intro

GAS CHROMATOGRAPHY MAINTENANCE USING UNCERTAINTY BASED CBM

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Experiment GC : Analysis of BTEX by GC-FID

Dr. Rogers Chapter 5 Homework Chem 111 Fall 2003

VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control

490 Micro GC Site Preparation Checklist

General Specifications

Rapid and Reliable Detection of Dissolved Gases in Water

GASES PROCESS CHEMICAL SPECIALTY. Gases. Pure Gases & Gas Mixtures. VOC Mixtures EPA Protocols. Propane Methane Nitrogen Air Acetylene Propylene Argon

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

stat0il Tone Orke Reservoir Fluid Study for Statoil, Well by Arne M.Martinsen LAB Den norske stab ol.ieselskap a.

Instrumental Technique GAS CYLINDER

USER'S MANUAL HOT-WIRE TYPE GAS SENSOR NAP-100A SERIES ( NET10A(R) ) (For All Combustible Gases)

High-purity gases. Customer-oriented offering highest quality

16. Studio ScaleChem Calculations

Lecture Bottles Purchase & Disposal

GAS SUPPLY APPLICATION GUIDE

ISO 9539 INTERNATIONAL STANDARD. Gas welding equipment Materials for equipment used in gas welding, cutting and allied processes

Agilent Precision Gas Flow Meter

Gases and welding distributors continue

Instruction Manual 742 5/1/2009. Eclipse Ratio Regulators ES Series Version 1

General Specifications

PRECISION GAS MIXTURES

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

5890II GC Standard Operating Procedure 9/2/2005

A New GC/MS System Designed for Helium Carrier Gas Conservation

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

ISO/TR TECHNICAL REPORT. Natural gas Calibration of chilled mirror type instruments for hydrocarbon dewpoint (liquid formation)

Aneer Engineers Pvt. Ltd.

Agilent 7697A Headspace Sampler

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Transcription:

Agilent 490 Micro GC Natural Gas Analyzer User Manual Agilent Technologies

Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G3582-90002 Edition First edition, May 2012 Printed in USA Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial computer software as defined in DFAR 252.227-7014 (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non-dod Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Contents 1 Agilent 490 Micro GC - Natural Gas Analyzers 2 Checkout Information Introduction 5 Types of Agilent 490 Micro GC - Natural Gas Analyzers 5 Agilent 490 Micro GC - Natural Gas Analyzer A and the Extended Version 6 Natural Gas Analyzer A 6 Natural Gas Analyzer A Extended 6 Agilent 490 Micro GC - Natural Gas Analyzer B and the Extended Version 7 Natural Gas Analyzer B 7 Natural Gas Analyzer B Extended 7 Introduction 8 3 Tuning the Backflush Time Natural Gas Analyzer A and the A Extended Version 10 HayeSep A straight 10 6 meter CP-Sil 5 CB 11 HayeSep A backflush 13 4 meter CP-Sil 5 CB 14 8 meter CP-Sil 5 CB 16 Natural Gas Analyzer B and B Extended Version 18 PoraPLOT U 18 6 meter CP Sil 5 CB 20 CP-Molsieve 5A 22 Tuning the Backflush Time on CP-Molsieve 5A, PoraPLOT U, or CP-Sil 5 CB 24 Tuning procedure for the backflush time 24 Tuning the Backflush Time on a HayeSep A 26 Tuning procedure for HayeSep A channel 26 Natural Gas Analyzer 3

A B C D E Appendix A: Certificate of the NGA Gas Calibration Standard Appendix B: Certificate of the Universal Gas Calibration Standard Appendix C: Typical Method Settings for Natural Gas Analyzers Appendix D: Carrier Gas Type Configuration Procedure to Change the Carrier Gas Type 32 Appendix E: Method for CP-Molsieve 5A Channel with Carrier Gas Argon Natural Gas Analyzer 4

Agilent 490 Micro GC Natural Gas Analyzer User Manual 1 Agilent 490 Micro GC - Natural Gas Analyzers Introduction Natural gas is a fossil fuel. Millions of years ago, the remains of plants, animals, and micro-organisms were buried underneath layers of soil and sediment. Due to the weight of the soil and sediment, the pressure and temperature increased. This resulted in the slow transformation of the organisms into natural gas and oil. The composition of natural gas varies widely. Natural gas is a combustible mixture of hydrocarbon gases that mainly consists of methane, smaller amounts of light hydrocarbons such as ethane, propane, n-butane, n-pentane, hexanes plus, and other gases such as nitrogen, carbon dioxide, oxygen, hydrogen, and hydrogen sulfide. Natural gas can also contain traces of argon, helium, neon, and xenon. Natural gas in its pure form is colorless and odorless. Natural Gas Analyzers have been designed to analyze the different compositions of natural gas. Types of Agilent 490 Micro GC - Natural Gas Analyzers The following four types of Natural Gas Analyzers can be used to analyze natural gas: The Agilent 490 Micro GC - Natural Gas Analyzer A The Agilent 490 Micro GC - Natural Gas Analyzer A Extended The Agilent 490 Micro GC - Natural Gas Analyzer B The Agilent 490 Micro GC - Natural Gas Analyzer B Extended Agilent Technologies 5

Agilent 490 Micro GC - Natural Gas Analyzers 1 Agilent 490 Micro GC - Natural Gas Analyzer A and the Extended Version Natural Gas Analyzer A The Agilent 490 Micro GC - Natural Gas Analyzer A is used to analyze natural gas samples containing methane, carbon dioxide, and hydrocarbons up to n-nonane. The Agilent 490 Micro GC - Natural Gas Analyzer A is a dual cabinet micro GC equipped with two channels: HayeSep A channel (straight) 6 meter CP-Sil 5 CB channel Natural Gas Analyzer A Extended The Agilent 490 Micro GC - Natural Gas Analyzer A Extended is used to analyze methane, carbon dioxide, and hydrocarbons up to n-dodecane. The Agilent 490 Micro GC - Natural Gas A Extended is a quad cabinet micro GC equipped with three channels: HayeSep A channel with a backflush option 4 meter CP-Sil 5 CB channel with a backflush option 8 meter CP-Sil 5 CB channel Natural Gas Analyzers 6

Agilent 490 Micro GC - Natural Gas Analyzers 1 Agilent 490 Micro GC - Natural Gas Analyzer B and the Extended Version Natural Gas Analyzer B The Agilent 490 Micro GC - Natural Gas Analyzer B is used to analyze methane, carbon dioxide, hydrogen sulfide, and light hydrocarbons up to n-nonane in natural gas. The Agilent 490 Micro GC - Natural Gas Analyzer B is a dual cabinet micro GC equipped with two channels: PoraPLOT U channel with a backflush option 6 meter CP-Sil 5 CB channel Natural Gas Analyzer B Extended The Agilent 490 Micro GC - Natural Gas Analyzer B Extended is used to analyze hydrogen, nitrogen, oxygen, methane, carbon dioxide, carbon monoxide, hydrogen sulfide, and hydrocarbons up to n-nonane. The Agilent 490 Micro GC - Natural Gas Analyzer B Extended is a quad cabinet micro GC equipped with three channels: CP-Molsieve 5A channel with a backflush option PoraPLOT U channel with a backflush option 6 meter CP-Sil 5 CB channel The Natural Gas Analyzer B Extended version is equipped with a dual carrier gas option for the CP-Molsieve 5A channel. It is made flexible for analysis of hydrogen and helium on the CP-Molsieve 5A channel if required. For the analysis of helium on the CP-Molsieve 5A channel, the carrier gas must be argon. All analyzers are factory tuned on carrier gas helium. Before starting up the Agilent 490 Micro GC - Natural Gas Analyzer, ensure that correct carrier gases are connected for all channels. For all types of analyzers, the required carrier gas is helium. However, to analyze helium on the CP-Molsieve 5A, the carrier gas must be argon. The required pressure for carrier gas is 550 kpa (80 psi). For more information on the Agilent 490 Micro GC, see the Agilent 490 Micro GC User Manual. Natural Gas Analyzers 7

Agilent 490 Micro GC Natural Gas Analyzer User Manual 2 Checkout Information Introduction The Agilent 490 Micro GC - Natural Gas Analyzers are factory tuned on carrier gas helium including appropriate settings for back flush times for the following channels: HayeSep A 4 meter CP-Sil 5 CB CP-Molsieve 5A PoraPLOT U The final checkout of the Agilent 490 Micro GC - Natural Gas Analyzer is performed with a NGA Gas Calibration standard and a Universal Gas Calibration standard. The NGA Gas Calibration standard contains nitrogen, methane, ethane, carbon dioxide, propane, iso-butane, n-butane, iso-pentane, n-pentane, and n-hexane. This natural gas standard does not contain all components specified for each type of Natural Gas Analyzer. Therefore, a second checkout sample, the Universal Gas Calibration standard is analyzed on the Natural Gas Analyzer. The Universal Gas Calibration standard contains helium, neon, hydrogen, oxygen, nitrogen, methane, ethane, ethylene, carbon dioxide, carbon monoxide, acetylene, propane, methyl acetylene, n-butane, n-hexane, and n-heptane. Not all components in the Universal Gas Calibration standard are specified for the Natural Gas Analyzer. Agilent Technologies 8

Checkout Information 2 More detailed information about the NGA Gas Calibration standard and the Universal Gas Calibration standard can be found in Appendix A on page 27 and Appendix B on page 28. The components specified for each channel are described in Agilent 490 Micro GC - Natural Gas Analyzer A and the Extended Version on page 6 and Agilent 490 Micro GC - Natural Gas Analyzer B and the Extended Version on page 7. The NGA Gas Calibration standard and the Universal Gas Calibration standard are shipped with the analyzer and used by the Agilent Customer Engineer at installation. The factory tuned method, final chromatogram (Test Report), and the Natural Gas Analyzer User Manual are supplied on the Analyzer CD. To perform reference checks on the analyzer, use the NGA Gas Calibration standard or the Universal Gas Calibration standard. If OpenLAB CDS EZChrom is used, load the method from the analyzer CD or create a method using the settings from the method pdf-file available on the analyzer CD. For a quick start, see the method settings for Natural Gas Analyzers listed in Appendix C on page 29. Connect and inject the NGA Gas Calibration standard. To analyze helium on the CP-Molsieve 5A channel, use the Universal Gas Calibration standard for the checkout. Ensure that the carrier gas argon is supplied and configured before starting the analysis. For more information on how to configure the carrier gas for the analyzer see Appendix D on page 31. Method settings for this analysis can be found in Appendix E on page 35. The NGA Gas Calibration standard and the Universal Gas Calibration standard do not contain hydrogen sulfide. A reference chromatogram of hydrogen sulfide is included in this manual. Natural Gas Analyzers 9

Checkout Information 2 Natural Gas Analyzer A and the A Extended Version HayeSep A straight The HayeSep A channel of the Natural Gas Analyzer A is specified for the analysis of methane, carbon dioxide, ethane, and propane. Analysis of the NGA Gas Calibration standard results in the chromatogram shown in Figure 1. Analysis of the Universal Gas Calibration standard results in a chromatogram similar to Figure 2 300 250 Nitrogen 200 150 100 50 0 Figure 1 Methane Carbon Dioxide Ethane 0 10 20 30 40 50 60 70 80 90 100 NGA Gas Calibration standard analyzed with the HayeSep A channel on Analyzer A Propane. 12 10 Nitrogen 8 6 4 2 0 Methane Carbon Dioxide Ethylene Ethane Acetylene Propane 0 10 20 30 40 50 60 70 80 90 100 Figure 2 Universal Gas Calibration standard analyzed with the HayeSep A channel on Analyzer A Natural Gas Analyzers 10

Checkout Information 2 6 meter CP-Sil 5 CB The 6 meter CP-Sil 5 CB channel of the Natural Gas Analyzer A is specified for the analysis of light hydrocarbons from propane up to n-nonane. Analysis of the NGA Gas Calibration standard results in a chromatogram similar to Figure 3. Analysis of a sample containing neo-pentane results in a chromatogram similar to Figure 4. 500 400 300 Ethane n-propane 200 100 0 i-butane n-butane i-pentane n-pentane 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 55.0 n-hexane Figure 3 NGA Gas Calibration standard analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer A 500 Ethane 400 Propane 300 200 100 i-butane n-butane neo-pentane i-pentane n-pentane n-hexane 0 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0 32.5 35.0 37.5 40.0 42.5 45.0 47.5 50.0 52.5 Figure 4 Sample containing neo-pentane analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer A Natural Gas Analyzers 11

Checkout Information 2 Analysis of the Universal Gas Calibration standard results in a chromatogram similar to Figure 5. Analysis of a sample containing heavier hydrocarbons up to n-nonane results in a chromatogram similar to Figure 6. If a sample contains hydrocarbons up to n-nonane ensure that the total run time is increased sufficiently to detect all hydrocarbons on this channel. 10.0 Propane 7.5 5.0 2.5 i-butane n-butane n-hexane n-heptane 0.0 10 20 30 40 50 60 70 80 90 Figure 5 Universal Gas Calibration standard analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer A 30 Propane 20 10 0 n-butane n-pentane n-hexane n-heptane n-octane 0 50 100 150 200 250 300 350 400 n-nonane Figure 6 Sample containing hydrocarbons up to n-nonane analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer A Natural Gas Analyzers 12

Checkout Information 2 HayeSep A backflush The HayeSep A channel, with backflush option, of the Natural Gas Analyzer A Extended is specified for the analysis of methane, carbon dioxide, ethane, and propane. Analysis of the NGA Gas Calibration standard results in the chromatogram shown in Figure 7. Analysis of the Universal Gas Calibration standard results in a chromatogram similar to Figure 8. olt 150 100 50 0 Figure 7 Nitrogen Methane Carbon Dioxide Ethane 0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 NGA Gas Calibration standard analyzed with the HayeSep A channel on Analyzer A Extended Propane 11 Methane olt 10 9 Nitrogen Carbon Dioxide Ethylene Ethane Acethylene Propane 8 0 10 20 30 40 50 60 70 80 90 100 110 Figure 8 Universal Gas Calibration standard analyzed with the HayeSep A channel on Analyzer A Extended Natural Gas Analyzers 13

Checkout Information 2 4 meter CP-Sil 5 CB The 4 meter CP-Sil 5 CB channel of the Natural Gas Analyzer A Extended is specified for the analysis of light hydrocarbons propane, n-butane, iso-butane, iso-pentane, and n-pentane. Analysis of the NGA Gas Calibration standard results in the chromatogram shown in Figure 9. Analysis of a sample containing neo-pentane results in a chromatogram similar to Figure 10. 150 Ethane olt 100 Propane 50 0 Figure 9 i-butane n-butane 8 10 12 14 16 18 20 22 24 26 28 30 32 34 NGA Gas Calibration standard analyzed with the 4 meter CP-Sil 5 CB channel on Analyzer A Extended i-pentane n-pentane 200 Ethane 150 Propane olt 100 50 i-butane n-butane neo-pentane i-pentane n-pentane 0 Figure 10 10 12 14 16 18 20 22 24 26 28 30 32 Sample containing neo-pentane analyzed with the 4 meter CP-Sil 5 CB channel on Analyzer A Extended Natural Gas Analyzers 14

Checkout Information 2 Analysis of the Universal Gas Calibration standard results in the chromatogram shown in Figure 11. olt 15.0 12.5 10.0 7.5 5.0 2.5 0.0 Ethane Propane i-butane 10 11 12 13 14 15 16 17 18 19 20 21 22 n-butane Figure 11 Universal Gas Calibration standard analyzed with the 4 meter CP-Sil 5 CB channel on Analyzer A Extended Natural Gas Analyzers 15

Checkout Information 2 8 meter CP-Sil 5 CB The 8 meter CP-Sil 5 CB channel of the Natural Gas Analyzer A Extended is used for the analysis of hydrocarbons from n-hexane up to n-dodecane. The NGA Gas Calibration standard and the Universal Gas Calibration standard do not contain heavy hydrocarbons up to n-dodecane. Reference chromatograms of the NGA Gas Calibration standard and the Universal Gas Calibration standard are included in this manual. For heavy hydrocarbons up to n-dodecane an additional sample is analyzed for identification. Analysis of the NGA Gas Calibration standard results in the chromatogram shown in Figure 12. Reference check of the Universal Gas Calibration standard results in a chromatogram similar to Figure 13. 15 10 5 n-hexane 0 15 20 25 30 35 40 45 50 55 Figure 12 NGA Gas Calibration standard analyzed with the 8 meter CP-Sil 5 CB channel on Analyzer A Extended 7.5 5.0 2.5 n-hexane n-heptane 0.0 15 20 25 30 35 40 45 50 55 Figure 13 Universal Gas Calibration standard analyzed with the 8 meter CP-Sil 5 CB channel on Analyzer A Extended Natural Gas Analyzers 16

Checkout Information 2 Analysis of a sample containing heavier hydrocarbons up to n-decane results in the chromatogram shown in Figure 14. Analysis of a sample that contains heavier hydrocarbons up to n-dodecane results in a chromatogram similar to Figure 15. 1.2 1.0 0.8 0.6 0.4 0.2 n-heptane n-octane n-nonane n-decane 0.0 10 20 30 40 50 60 70 80 90 100 110 Figure 14 Hydrocarbon gas mix containing n-heptane to n-decane analyzed with the 8 meter CP-Sil 5 CB channel on Analyzer A Extended 2.0 1.5 n-hexane n-heptane n-octane 1.0 n-nonane 0.5 n-decane n-undecane n-dodecane 20 40 60 80 100 120 140 160 180 200 220 Figure 15 Hydrocarbon gas mix containing n-hexane to n-dodecane analyzed with the 8 meter CP-Sil 5 CB channel on Analyzer A Extended Natural Gas Analyzers 17

Checkout Information 2 Natural Gas Analyzer B and B Extended Version PoraPLOT U The PoraPLOT U channel of the Agilent 490 Micro GC - Natural Gas Analyzer B and Extended version is specified for the analysis of methane, carbon dioxide, ethane, hydrogen sulfide, and propane in natural gas samples. The total sample path of this channel is deactivated (Ultimetal), which results in better performance and peak shape for hydrogen sulfide. Analysis of the NGA Gas Calibration standard results in a chromatogram similar to Figure 16. 400 300 Nitrogen Methane 200 100 Carbon Dioxide Ethane Propane 0 20 25 30 35 40 45 50 55 60 65 Figure 16 NGA Gas Calibration standard analyzed with the PoraPLOT U channel on Analyzer B Natural Gas Analyzers 18

Checkout Information 2 Analysis of the Universal Gas Calibration standard results in a chromatogram similar to Figure 17. Analysis of a sample containing hydrogen sulfide results in a chromatogram similar to Figure 18. 20 Nitrogen 10 0 Methane Carbon Dioxide Ethylene Ethane Acetylene Propane 20 25 30 35 40 45 50 55 60 65 Figure 17 Universal Gas Calibration standard analyzed with the PoraPLOT U channel on Analyzer B 75 Methane 50 25 0 Nitrogen Carbon Dioxide Ethane Hydrogen Sulfide Propane 20 25 30 35 40 45 50 55 60 65 Figure 18 Sample containing hydrogen sulfide analyzed with the PoraPLOT U channel on Analyzer B Natural Gas Analyzers 19

Checkout Information 2 6 meter CP Sil 5 CB The 6 meter CP-Sil 5 CB channel of the Natural Gas Analyzer B and the B Extended is specified for the analysis of hydrocarbons from propane to n-nonane. Analysis of the NGA Gas Calibration standard results in the chromatogram shown in Figure 19. Analysis of the Universal Gas Calibration standard results in a chromatogram similar to Figure 20. 500 400 300 Ethane n-propane 200 100 0 i-butane n-butane i-pentane n-pentane 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 55.0 n-hexane Figure 19 NGA Gas Calibration standard analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer B. 10.0 Propane 7.5 5.0 2.5 i-butane n-butane n-hexane n-heptane 0.0 Figure 20 10 20 30 40 50 60 70 80 90 Universal Gas Calibration standard analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer B Natural Gas Analyzers 20

Checkout Information 2 Analysis of a sample containing neo-pentane results in a chromatogram similar to Figure 21. Analysis of a sample containing hydrocarbons up to n-nonane results in a chromatogram similar to Figure 22. For samples that contain hydrocarbons up to n-nonane, increase the total run time to detect all hydrocarbons on this channel. 500 Ethane 400 300 200 100 Propane i-butane n-butane neo-pentane i-pentane n-pentane 0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 n-hexane Figure 21 Sample containing neo-pentane analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer B 30 Propane 20 10 0 n-butane n-pentane n-hexane n-heptane n-octane 0 50 100 150 200 250 300 350 400 n-nonane Figure 22 Sample containing hydrocarbons up to n-nonane analyzed with the 6 meter CP-Sil 5 CB channel on Analyzer B Natural Gas Analyzers 21

Checkout Information 2 CP-Molsieve 5A The CP-Molsieve 5A channel of the Natural Gas Analyzer B Extended is used for the analysis of permanent gases such as hydrogen, oxygen, nitrogen, methane, and carbon monoxide. Analysis of the NGA Gas Calibration standard results in a chromatogram similar to Figure 23. Analysis of the Universal Gas Calibration results in the chromatogram shown in Figure 24. 15 Oxygen 10 5 0 Neon Nitrogen Methane 10 15 20 25 30 35 40 45 50 55 60 Figure 23 NGA Gas Calibration standard analyzed with the CP-Molsieve 5A channel on Analyzer B Extended 12.5 Nitrogen 10.0 7.5 5.0 2.5 0.0 Figure 24 Neon Hydrogen Oxygen Methane 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 Universal Gas Calibration standard analyzed with the CP-Molsieve 5A channel on Analyzer B Extended Carbon Monoxide Natural Gas Analyzers 22

Checkout Information 2 CAUTION The CP-Molsieve 5A channel is factory tuned on carrier gas helium. If analysis of helium or hydrogen is required on the CP-Molsieve 5A channel, the carrier gas must be changed and configured for argon. The procedure for configuring the carrier gas in OpenLAB CDS EZChrom is described in Appendix D on page 31. Analysis of the Universal Gas Calibration standard results in a chromatogram similar to Figure 25. 8 Hydrogen 6 4 Helium 2 0 Neon Oxygen Nitrogen Methane Carbon Monoxide 20 30 40 50 60 70 80 90 100 110 Figure 25 Universal Gas Calibration standard analyzed with the CP-Molsieve 5A channel on Analyzer B Extended Natural Gas Analyzers 23

Agilent 490 Micro GC Natural Gas Analyzer User Manual 3 Tuning the Backflush Time Tuning the Backflush Time on CP-Molsieve 5A, PoraPLOT U, or CP-Sil 5 CB Tuning the backflush time is necessary for each new CP-Molsieve 5A channel, PoraPLOT U channel, and 4 meter CP-Sil 5 CB channel. This chapter describes how to tune the backflush time for these channels. The goal for tuning the CP-Molsieve 5A is to get all methane on the column while later-eluting components such as moisture, carbon dioxide, ethane, and higher hydrocarbons are backflushed. The PoraPLOT U is tuned on the propane peak while all other components that elute after propane are backflushed. The 4 meter CP-Sil 5 CB is tuned on the n-pentane while all other hydrocarbons that elutes after n-pentane are backflushed. Tuning procedure for the backflush time 1 Set the backflush time to 0 seconds and analyze the checkout sample or a proper sample for the specific channel. The goal of this is to identify the components in the calibration standard. 2 Change the backflush time to 10 seconds and perform a run. The following can be observed: When the backflush time is set too early, the peaks of interest are partially or totally backflushed. If the backflush time is set too late, the unwanted components are not backflushed and show up in the chromatogram. 3 Perform runs with different backflush times until there is no huge difference in the peak of interest. To fine tune the backflush time, set smaller steps (for example 0.10 seconds) until you find the optimal backflush time. Agilent Technologies 24

Tuning the Backflush Time 3 Figure 26 shows a simple example of tuning the backflush time for the CP-Molsieve 5A channel. Tuning backflush for CP-Molsieve channel Area_peak of interest 9 10 11 12 13 Backflush me (s) Figure 26 Effect of the backflush time on the peak of interest Natural Gas Analyzers 25

Tuning the Backflush Time 3 Tuning the Backflush Time on a HayeSep A 100 For each new HayeSep A channel, with a backflush option, it is necessary to tune the backflush time properly. The tuning procedure of the HayeSep A channel is different than the tuning procedure of a CP-Molsieve 5A, PoraPLOT U, or 4 meter CP-Sil 5 CB channel. This section describes how to tune the backflush time for the HayeSep A channel. The goal for tuning the HayeSep A channel is get all peaks of interest, components up to propane, on the HayeSep A column while all unwanted peaks that elute after propane are backflushed. Tuning procedure for HayeSep A channel 1 Set the backflush time of the HayeSep A channel to 0 seconds. 2 Set an appropriate run time for the first analysis (for example 300 seconds or longer). 3 Analyze the NGA Gas Calibration standard and identify all components in the calibration standard. 4 When all peaks of interest are identified, select a proper backflush time after propane peak. Figure 27 shows an example of the tuning procedure of HayeSep A channel. In this example the propane peak elutes around 90 seconds, proper backflush time for the HayeSep A here is around 120 seconds. Consider that the total run time must be sufficient to backflush all unwanted components from the column. The ideal total run time is approximately twice the backflush time or higher. So in this example, a total run time of 240 seconds is sufficient to backflush all unwanted components from the HayeSep A channel. 80 olt 60 40 20 Nitrogen Methane Carbon Dioxide Ethane Peaks of Interest Propane Appropriate Backflush Time Unwanted Peak i--butane 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 Figure 27 Selecting backflush time for a HayeSep A channel Natural Gas Analyzers 26

Agilent 490 Micro GC Natural Gas Analyzer User Manual Appendix A: Certificate of the NGA Gas Calibration Standard Part number: 5184-3536 Agilent Technologies 27

Agilent 490 Micro GC Natural Gas Analyzer User Manual Appendix B: Certificate of the Universal Gas Calibration Standard Part number 5184-3541 Agilent Technologies 28

Agilent 490 Micro GC Natural Gas Analyzer User Manual Appendix C: Typical Method Settings for Natural Gas Analyzers Typical method settings for the Natural Gas Analyzers are given in Tables 1-4. Table 1 Method for Agilent 490 Micro GC - Natural Gas Analyzer A Method settings * HayeSep A Carrier gas Helium Helium Injector temperature ( C) 110 110 Injection time (ms) 40 40 Column temperature ( C) 60 70 Pressure (kpa) 260 150 Sample line temperature ( C) 110 110 CP-Sil 5 CB_6m * For more details of the method, see the pdf-method file available on the Natural Gas Analyzer CD. Table 2 Method for Agilent 490 Micro GC - Natural Gas Analyzer A Extended Method settings * HayeSep A CP-Sil 5 CB_4m CP-Sil 5 CB_8m Carrier gas Helium Helium Helium Injector temperature ( C) 110 110 110 Injection time (ms) 20 40 40 Backflush time (s) 120 12 - Column temperature ( C) 90 60 150 Pressure (kpa) 340 150 200 Sample line temperature ( C) 110 110 110 * For more details of the method, see the pdf-method file available on the Natural Gas Analyzer CD. The backflush time must be tuned for each new CP-Molsieve 5A, PoraPLOT U, 4 meter CP-Sil 5 CB, and HayeSep A channel. Agilent Technologies 29

Appendix C: Typical Method Settings for Natural Gas Analyzers Table 3 Method for Agilent 490 Micro GC - Natural Gas Analyzer B Method settings * PoraPLOT U Carrier gas Helium Helium Injector temperature ( C) 110 110 Injection time (ms) 40 40 Backflush time (s) 17 - Column temperature ( C) 60 70 Pressure (kpa) 175 150 Sample line temperature ( C) 110 110 CP-Sil 5 CB_6m * For more details of the method, see the pdf-method file available on the Natural Gas Analyzer CD. The backflush time must be tuned for each new CP-Molsieve 5A, PoraPLOT U, 4 meter CP-Sil 5 CB, and HayeSep A channel. Table 4 Method for Agilent 490 Micro GC - Natural Gas Analyzer B Extended Method settings * CP-Molsieve 5A PoraPLOT U CP-Sil 5 CB_6m Carrier gas Helium Helium Helium Injector temperature ( C) 110 110 110 Injection time (ms) 40 40 40 Backflush time (s) 11 17 - Column temperature ( C) 80 60 70 Pressure (kpa) 200 175 150 Sample line temperature ( C) 110 110 110 * For more details of the method, see the pdf-method file available on the Natural Gas Analyzer CD. The backflush time must be tuned for each new CP-Molsieve 5A, PoraPLOT U, 4 meter CP-Sil 5 CB, and HayeSep A channel. Natural Gas Analyzers 30

Agilent 490 Micro GC Natural Gas Analyzer User Manual Appendix D: Carrier Gas Type Configuration The Agilent 490 Micro GC - Natural Gas Analyzers are factory tuned using the carrier gas helium. When analysis of helium is required, the carrier gas must be changed to argon. Instruments with the dual carrier option are typically configured as shown in Figure 28. Natural Gas Analyzer B Extended, quad channel cabinet Channel 1 Channel 2 Channel 3 Channel 4 MolSieve 5A PoraPLOT U CP-Sil 5 CB Empty slot Carrier gas 2 Carrier gas 1 Figure 28 Typical configuration of a dual carrier instrument Agilent Technologies 31

Appendix D: Carrier Gas Type Configuration Procedure to Change the Carrier Gas Type If you are using Agilent OpenLAB CDS EZChrom edition, use the following procedure to change the carrier gas type. 1 In the Agilent OpenLAB Control Panel Navigation pane, select the instrument. 2 In the Actions toolbar, select Configure Instrument. See Figure 29. Figure 29 OpenLAB Control Panel Natural Gas Analyzers 32

Appendix D: Carrier Gas Type Configuration 3 Double-click Agilent 490 Micro GC. See Figure 30. Figure 30 Configuration window Agilent 490 Micro GC 4 In the EZChrom Configuration dialog box, select the Instrument configuration tab. 5 In the Carrier gas list, select Argon. See Figure 31. Figure 31 The Instrument configuration tab Natural Gas Analyzers 33

Appendix D: Carrier Gas Type Configuration 6 Select Apply Configuration. The new configuration is sent to the instrument as shown in Figure 32 Figure 32 Sending configuration to instrument 7 When the configuration is complete, in the Reboot Instrument dialog box select OK. See Figure 33 Figure 33 Request after configuration change 8 Turn off the instrument. 9 Change the carrier gas on the instrument. 10 Restart the Agilent 490 Micro GC - Natural Gas Analyzer. The Agilent 490 Micro GC is now configured for carrier gas Argon. Natural Gas Analyzers 34

Agilent 490 Micro GC Natural Gas Analyzer User Manual Appendix E: Method for CP-Molsieve 5A Channel with Carrier Gas Argon Typical method setting for analysis of helium and hydrogen on the CP-Molsieve 5A channel. Table 5 Method setting for analysis of helium with Natural Gas Analyzer B Extended Method settings CP-Molsieve 5A Carrier gas Argon Injector temperature ( C) 110 Injection time (ms) 40 Backflush time (s) 11 Column temperature ( C) 80 Pressure (kpa) 200 Sample line temperature ( C) 110 Invert signal Yes Agilent Technologies 35

Agilent Technologies Agilent Technologies, Inc. Printed in USA, May 2012 *G3582-90002* G3582-90002