Video Notes. LC Troubleshooting Series Retention Time Shifts

Similar documents
GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

Agilent Dimension Software for ELSD User Manual

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

More Speed and Resolution for Current Methods

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

no peaks missing peaks too small peaks increasing retention decreasing retention

Best Practice for Identifying Leaks in GC and GC/MS Systems

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Retention Time Locking: Concepts and Applications. Application

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

AKTA pure 25 New Owner s Intro

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Autosampler problems!

HPLC systems. Overview

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

SOP: Buck Scientific BLC-20P HPLC Operation

Using the Akta Prime plus October 22, 2012

Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design

Setting a New Standard in Flash Chromatography Performance

Dionex IonPac AS19-4µm

Carrier Gases in Capillary GC

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Prep LC Training Manual

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Prep LC Training Manual S.V.

Rapid and Reliable Detection of Dissolved Gases in Water

New improved robot for gas kinetics

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Experiment GC : Analysis of BTEX by GC-FID

Gradients in ion chromatography. Optimal separation with shorter analysis time

Agilent Auxiliary Gas Module AGM 2. User s Guide

Extended Application Note

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Running HPLC Methods on an ACQUITY UPLC System

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

HPLC troubleshooting guide

Standard Operating Procedure. Air Displacement Pipette Calibration

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

Leak Checking Large Vacuum Chambers

Supporting information for J. Med. Chem., 1994, 37(5), , DOI: /jm00031a018

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Packing Procedure Recommendation. For. YMC RP silica. ODS-A, ODS-AQ, Octyl, Butyl, YMCbasic

Carrier Gases in Capillary GC

Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors. User s Guide

The instrument will process the samples one by one till all the samples have been cleaned up.

Multiple Gas#5 GC configuration Jan 2016

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

GCMSD-Headspace Analysis SOP

TITLE: DNA minitube - Red

LC, RSLC, Bio RS and Dual Gradient Systems

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Operation of the Perkin Elmer TGA-GC/MS

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

Chromat Calibration Updated October 27th, 2017

Recovery collection complete sample recovery in preparative HPLC

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

EAS Basic Soil Gas Sampler

Modelling of Pressurised Pipes within InfoWorks ICM and CS

Mass Spec will not Autotune

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved

Gas Leak Measurement of TRT Barrel Module

Prof AH Basson, Pr Eng March Departement Meganiese en Megatroniese Ingenieurswese. Department of Mechanical and Mechatronic Engineering

SHIMADZU LC-10/20 PUMP

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

CORESTA RECOMMENDED METHOD Nº 67

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions

Lab 5- Cellular Respiration

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Shimming with VnmrJ 3.1 Software for Peak NMR Performance

WELCOME TO FORM LAB MAX

Quick run through of the four projects

Degassing ERC KK Saitama, Japan

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Joint industry project on foam

Buck Scientific, Inc.

Agilent 1260 Infinity Purification Solution

Introduction. Experimental

Technical Report. Comparison of Ventilation Standard Calibration Methods

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! There Are Many Reasons to Reduce Analysis Time

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

Transcription:

Video Notes LC Troubleshooting Series Retention Time Shifts Introduction John Henderson is an Applications Scientist for Agilent Chemistries and Supplies Division. Maureen Joseph is the HPLC Columns Product Manager. Some possible causes to consider for retention time changes are: Column aging Column contamination A poor column/mobile phase combination for your analytes Insufficient column re-equilibration Changes in flow rate or column temperature Figure 1 below shows both changing retention times and poor peak shape. In this case, the capillaries were connected to the autosampler valve improperly. Figure 1. Changing retention times and poor peak shape. Agilent Technologies, Inc. 2010 Page 1 of 7

Change in Column Dimensions or Flow Rate It s important to remember to adjust system and method parameters flow rate, injection volume, gradient time-- when you change your column dimensions. Try the Agilent LC Method Translator. Search for LC Method Translator at www.agilent.com. Figure 2 shows a gradient separation of pesticides and flame retardants in drinking water, first on a 100mm column and then on a 50mm column, but hardly any time savings are recognized with the shorter column. Figure 2. A gradient separation of pesticides and flame retardants in drinking water on a 100mm column and a 50mm column. Figure 3. Adjusting the gradient to accommodate a new column length. Because the column length was cut by 50%, the gradient time must also be cut in half to realize the time savings and keep the separation the same. When the gradient time is cut in half to match the column length and change in volume, Figure 3 shows that the analysis time is also cut in half. Agilent Technologies, Inc. 2010 Page 2 of 7

Figure 4 demonstrates the same principle as Figure 3, except the flow rate, instead of the column length has been changed to decrease analysis time. Figure 4. Altering the flow rate to decrease analysis time. When we increased the flow rate 60% to speed up our analysis, as we would in an isocratic method, we find that analysis time did not decrease 60%, and retention factors have changed, in fact they ve increased. We need to adjust the gradient to match the flow rate change. Figure 5 displays the new chromatogram: Figure 5. Adjusting the gradient to correspond with a new flow rate. As you can see, our retention factors have adjusted correctly, and we ve cut our analysis time. In Figure 5, many people might suspect there is a problem with the LC pump delivering the right flow rate in this case, it was just a method issue. This is the first thing you should check. Agilent Technologies, Inc. 2010 Page 3 of 7

The Gradient Equation t G1 F 1 /Vm 1 = t G2 F 2 /Vm 2 The retention factor in gradients, k*, depends on several parameters: Flow rate (F) Gradient time (t) Column void volume (proportional to column size) (Vm) Remember: Any change in flow rate, gradient time or void volume will change k* or retention factor. Vm is proportional to column length, so any change in column length will also change k*. To keep k* for method 1 the same when changing to method 2, these factors must balance. Figure 6. Maintaining retention factor (k*) when changing column length. In Figure 6, the column length was halved and the column length is directly proportional to the void volume, so since the void volume was halved, the gradient or flow rate must be doubled to keep the equation balanced; not doing so resulted in barely any time savings at all with the shorter column. Figure 7. Alkylphenone: Maintaining retention factor (k*) when changing flow rate. In Figure 7, flow rate (F2) in method 2 was increased, but the gradient time (tg2) was not adjusted, resulting in the surprising increase in retention factors. When we decreased the gradient time by 60% to balance the equation and keep k* equivalent, the retention factors matched. Agilent Technologies, Inc. 2010 Page 4 of 7

Buffers and Mobile Phase Buffers and mobile phases impact a column s retentiveness over time. Keep a chromatogram of your column s performance right out of the box, using a standard or your common method, so you can run periodic comparisons to evaluate its performance. Guard columns are a good way to extend the life of your analytical column. A change in mobile phase may change your retention. Sometimes mobile phases can differ because of the way they are mixed in the lab. Mobile phases can become contaminated over time. It s a good practice to degas your mobile phase, to prevent air bubbles from entering your flowpath and causing retention shifts. Figure 8 shows an analysis with a ZORBAX Rapid Resolution HD column on the 1290 Infinity. Note how the peak shapes and retention differ in this analysis from run to run, as the mobile phase is changed. Figure 8. The effect of equilibration time on peak shape with a ZORBAX RRHD Eclipse Plus C18 2.1 x 100mm column. Agilent Technologies, Inc. 2010 Page 5 of 7

Column Re-equilibration In gradients, column re-equilibration is necessary to keep your method reproducible. Measure your column void volume by: Measuring the time from the start of the chromatogram to the first disturbance on the baseline. Dividing this time in minutes by the flow rate in ml/min to find ml of void volume. A general rule is to use ten column volumes for method development analyses, but for established methods, the equilibration time can be evaluated and shortened. The optimal equilibration time needs to be determined by trial and error, running the gradient repeatedly and looking for changes in retention. In Figure 9, an example of a gradient separation for grapefruit juice analysis, we can see the column void volume and amount of time dedicated to column re-equilibration. In this case, equilibration time was 1.4 min., equivalent to 4 column void volumes. Figure 9. Gradient Separation in Grapefruit Juice. The rule of thumb is to use ten column volumes for method development analyses, but for established methods the equilibration time can be evaluated and shortened. Figure 10. Identical repeated injections. The minimum equilibration time has to be determined by trial and error, by running the gradient repeatedly and looking for changes in retention. If several repeated injections are identical, as in this Figure 10, equilibration time is sufficient. Agilent Technologies, Inc. 2010 Page 6 of 7

Summary Tips for keeping retention times consistent: Know your column, and keep a sample chromatogram of its performance in new condition, to compare over time. Make sure you adjust your method if you change your column size use the LC Method Translator Tool for help. Be sure you use a consistent method for mixing your mobile phase. If you re using gradients, test and confirm the number of column volumes you need to re-equilibrate during method development. If you require additional assistance, you can always contact Agilent technical support by logging on to www.agilent.com/chem/contactus. Agilent Technologies, Inc. 2010 Page 7 of 7