SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

Similar documents
SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA020ANM-S1 BREATHING AIR PANEL

SERVICE MANUAL MODEL BA200AA-S1 BREATHING AIR PANEL

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL

SERVICE MANUAL MODEL BA200AMST-S1 BREATHING AIR PANEL

MODEL RP050BMST RESPIRATORY PROTECTOR MANUAL

MODEL RP050BMST-S1 RESPIRATORY PROTECTOR MANUAL

MODEL RP200C(_)-S1 RESPIRATORY PROTECTOR MANUAL

INSTRUCTION MANUAL MST AFTERCOOLER SYSTEM MODEL

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

INSTRUCTION MANUAL & PARTS

ADVANCED AIRLINE FILTER INSTRUCTION MANUAL

3M TM Versaflo TM AP-600 Series Air Filtration Panels For use with 3M TM Versaflo TM Supplied Air Respirator Systems. Work. in a World of.

INTRODUCTION... 3 II II II II II II II II II II II II II I 4-5 ASSEMBLY... 6,7,8 II II II II II II II... 9

Important: Keep these instructions for reference.

PRODUCT OPERATING MANUAL

RAM 9025-XP Operation Manual

MODEL CALIBRATION GAS DELIVERY SYSTEM

BREATHING AIR PURIFIER

AIRLINE-SERIES HELMET AIR FILTER

PERSONAL AIR BREATHING UNIT

Operating Manual. Model BAC-30RT30/BAC-30DC

exceeds global breathing air quality standards B 1 breathing air systems flow capacity: scfm flow capacity: scfm ( Nm 3 /hr)

B-Series. breathing air systems flow capacity: scfm

A. Evaluation of the written program

Niltox Breathing Air Purifier Range

Technical Information

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

RAM 4021 Operation Manual

ADP3CA Free-Air Pump User Manual

Product Information News

Air Quality Test Kits Models: LP/HP-A4K and LP/HP-A445K

List of Contents vma. Compressed Air Conditioning 2. System description 2. Individual Units Pre-Filters 3. Combinations 6

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

accidents which arise due to non-observance of these instructions and the safety information herein.

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

Tsunami Breathable Air System

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

Model: 720-UL INSTALLATION OPERATION MAINTENANCE. Bermad Pressure Reducing Valve IOM. Model: FP -720-UL Sizes: 2"-12" BERMAD. Application Engineering

BS Series Basket Strainer

GREIG FILTERS, INC. OPERATION MANUAL

photo coming soon Product Family Respiratory Replacement Parts Canada (English) Page 1 of 17 Product Numbers & Ordering Information

TECHNICAL INFORMATION

Advantage 4000 Abrasi-Blast Respirator

Operating Manual. Appa. Models LP/HP-A4K & LP/HP-A445K Air Quality Test Kits Manual No. AIRTST01 (Rev 7 October 2008)

Oxygen Usage Best Practice Guide

RAM Operation Manual

RAM 4021-DPX Operation Manual

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

ADP20 Free-Air P u m p

Nitrogen Supply System

TECHNICAL INFORMATION

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

HPIC D Hydrostatic Test Pump PRODUCT INFORMATION AND OPERATING INSTRUCTIONS

600 o F High Temperature Aging Cell Instruction Manual

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Un-Pressurized Orefice Fittings FIO EZ. Parts List and Operation Instructions TECHNICAL MANUAL. Dn 2-6 Class Lbs

CleanAIR ARES CA-83. welding leather hood. Contents. 1 General information 2 Use 3 Technical specification 4 Maintenance 5 List of parts 6 Approvals

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

OPERATION MANUAL NTF-60 Plus

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Interface Devices, Inc. Hydraulic Mini Mule

MF2M Monnier Miniature Compressed Air Filter

Checklist for care of Respirators

Air Mask Low/High Pressure

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

photo coming soon Product Family Respiratory Replacement Parts Corporate - English Page 1 of 23 Product Numbers & Ordering Information

Breathing Air Purifiers. Comply with European Pharmacopoeia and International Breathing Air Standards

1 DRIVE INDUSTRIAL IMPACT WRENCH

Protocol Switchover Station

Carter Option X Valve connector assembly. Maintenance manual

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

PRESSURISED PAINT CONTAINER

Texas Department of Insurance. Respiratory Protection. Provided by. Workers Health & Safety HS01-027B (1-06)

Respiratory Protection for Producers

CS Controlled Dissolved Gas assembly with optional CS Air Aspirator assembly. Installation and Operation information

A Pocket Guide to Air-Line Systems

PremAire INSTRUCTIONS FOR DUAL-SUPPLY MODE OF OPERATION

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

Model PSI Compressor with 3-Gallon Air Tank 12VDC

RESPIRATORY PROTECTION PROGRAM OSHA Respiratory Protection

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions

Pure Water Power Purification System 4-Stage RODI Operations Manual

Operating Manual CBA Series Cascade Breathing Air Assemblies

ATD LBS. PRESSURE BLASTER INSTRUCTION MANUAL

Tri-Safem Model I1 Regulator. INSTRUCTIONS and PARTS

Right-Angle Tube Roller Model

100C Air Compressor Kit

Owner s Manual & Safety Instructions

Installation and commissioning instructions 255 series and 256 series

Transcription:

Product Manual MST Portable Air Filtration System SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 3/27/02 Copyright 2016 Burwell Technologies For more information: 1300 287 935 / mail@burwell.com.au / www.burwell.com.au

2 GENERAL OPERATION MANUAL WARNING: The Portable Air Panel Model BA100BMST: 1) WILL NOT remove Carbon Monoxide. 2) SHALL NOT be used when the air entering the filtering system is oxygen deficient. The Air Panel will not increase the oxygen content of the air. 3) SHALL NOT be used in an Immediately Dangerous to Life and Health Atmosphere (IDLH) unless it is used in conjunction with a Back-Up Escape system or a supplied air Self-Contained Breathing Apparatus (SCBA), where applicable. WARNING: The Carbon Monoxide Monitor used in this device will alarm if the incoming air's carbon monoxide content exceeds a pre-determined alarm set point (depending on the applicable standard). If the monitor's alarm should sound, remove respirator, or activate escape air supply or SCBA if applicable, and move immediately to a safe breathable atmosphere. Have the proper qualified personnel examine the equipment and make the appropriate corrections before using again. 4) SHALL NOT have air inlet pressure greater than 150 PSIG static (10.4 bar). Personal injury could result. 5) SHALL NOT have air outlet pressure that exceeds Manufacturers' Respirator/Hose Assembly pressure requirements. Personal injury could result. The Portable Air Panel is a Four Stage Filtration System designed to remove or reduce select contaminates found in compressed air lines while monitoring for carbon monoxide through an electrochemical Carbon Monoxide Monitor. The Portable Air Panel can be connected directly to shop air from a standard compressed air source to help provide breathing quality air to face masks, helmets, hoods and other supplied air-type breathing devices. For more information / 1300 287 935 2 of 13 mail@burwell.com.au / www.burwell.com.au

3 GENERAL FILTER SYSTEM DESCRIPTION (Refer to Figure No.1) Air entering the Portable Air Panel at the inlet (A) is usually contaminated with oil, water, dirt, rust, scale and gaseous Hydrocarbons. As the air passes through the First Stage (B) of the Prefilter, particulate matter is trapped and retained down to 0.3 microns. The air then enters the Second Stage (C) of the Prefilter which traps all particulate matter larger then 0.3 micron and coalesces liquid aerosols with an efficiency rating of 99.97% (D.O.P. 0.3-0.6 micron). MST s prefilter meets underwriters laboratory specification UL 586 for high efficiency, particulate, air filter units (HEPA rating). The liquid contaminates are trapped in the lower chamber of the prefilter and expelled out through the Automatic Float Drain (D). The Third and Fourth Stages (E & F) contain a deep bed of odor absorbing activated charcoal which collects various gaseous Hydrocarbons (such as oil fumes, benzene, etc.). The air is then passed through the Air Pressure Regulator (G), which is used to adjust the air pressure going to the respirator(s). A sample of the filtered air is taken at (H) and passed through the Carbon Monoxide Monitor (I). The Carbon Monoxide Monitor continuously checks the carbon monoxide levels per OSHA requirements and digitally displays the amount present in PPM (parts per million). An audio and visual alarm will alert operators if levels of carbon monoxide exceed OSHA (or other applicable standards) requirements. FIGURE NO. 1 For more information / 1300 287 935 3 of 13 mail@burwell.com.au / www.burwell.com.au

4 GENERAL INSTRUCTIONS The Model BA100BMST Portable Air Panel, a compressed air purifier, when used in accordance with the following instructions, will remove or reduce selected contaminants from compressed air lines. WARNING: The Portable Air Panel Model BA100BMST should not be used when the air entering the filter is oxygen deficient (air with a low oxygen content). The Portable Air Panel will not increase the oxygen content of oxygen deficient air. The Factory strongly recommends that a complete safety program be instated to ensure that the respiratory air is in compliance with all OSHA standards and other applicable laws regulating the use of supplied air respiratory systems, and the air quality be tested upon installation and periodically re-tested to ensure that the minimum requirements for breathing air are maintained. WARNING: The Portable Air Panel Model BA100BMST will not remove Carbon Monoxide. The Carbon Monoxide Monitor in the Portable Air Panel Model BA100BMST is a CSA Certified Intrinsically Safe monitor. Refer to the Monitor's Information Manual for charging, calibration, and operating instructions. The Factory will not assume any liability for accidents or personal injury resulting from the improper use of this equipment. Service on this equipment should only be performed by qualified personnel. This system is to be used only by trained qualified personnel in accordance with a respirator program as outlined in OSHA Regulation 29 CFR 1910.134(b). The air supply entering The Portable Air Panel must meet two (2) criteria to ensure maximum efficiency of the filter system: 1) Temperature range should be between 35-150 o F (2-65 o C) 2) Liquid size range should be between 2-8 microns. Conditioning equipment may be necessary, prior to the filter, to achieve this temperature range and bulk liquid removal. WARNING: Maximum pressure supplied to this unit IS NOT to exceed 150 PSIG. Personal injury could result. For more information / 1300 287 935 4 of 13 mail@burwell.com.au / www.burwell.com.au

5 The Regulator (80075) is used to adjust the air pressure supplied to the operator(s) to maintain the correct air exchange inside the mask, hood or helmet. Consult instructions on your particular breathing mask or hood for the proper pressure required. The Portable Air Panel Model BA100BMST is supplied with Industrial Interchange Quick-Disconnects. The quick-disconnects supplied on the mask and hose assembly from the manufacturer shall be used to validate the NIOSH approval on the mask and hose assembly. WARNING: DO NOT Exceed mask and hose assembly manufacturer's requirements for outlet pressures. Personal injury could result. The filter cartridges used in the Portable Air Panel Model BA100BMST have an average life expectancy of (1) one month, (depending on the quality of the air in your compressed air lines) under normal conditions. The Factory recommends changing all three (3) filter cartridges after one month of use unless conditions warrant more or less frequent replacement. WARNING: If the Monitor's alarm should sound, activate your emergency escape system, or self-contained breathing apparatus (SCBA) when applicable, (or if you ar NOT in an immediately dangerous to life and health atmosphere [IDLH] remove your mask or helmet) and move immediately to a non-contaminated atmosphere. Have the proper qualified personnel examine the equipment and make the appropriate corrections before using again. The Carbon Monoxide Alarm will sound to alert the operator that a problem has developed. The Monitor Alarm will sound due to one or more of the following conditions: A) Carbon Monoxide levels exceeding a pre-determined alarm set level is passing through the Portable Air Panel. The Portable Air Panel DOES NOT REMOVE or reduce the Carbon Monoxide content in the compressed air. Examine the air compressor, the air compressor intake air quality, and the complete system. Have the proper qualified personnel make the appropriate corrections before using again. B) Monitor out of Calibration. Calibrate Carbon Monoxide Monitor as outlined in the Monitor's Information Manual. C) The Monitor's carbon monoxide sensor may be bad. See the Monitor's Information Manual for all maintenance on the monitor. For more information / 1300 287 935 5 of 13 mail@burwell.com.au / www.burwell.com.au

6 SERVICE INSTRUCTIONS (Refer to Figure No. 2) WARNING: Always turn off air supply and bleed air pressure before disassembling unit or SERIOUS INJURY COULD RESULT. The Factory recommends replacing all three (3) filter cartridges after one (1) month of use unless conditions warrant more or less frequent replacement. To refill or replace the filter cartridges in the Portable Air Panel Model BA100BMST, follow these steps: 1. PREFILTER FIRST/SECOND DUAL STAGE ELEMENT REPLACEMENT A) First pull Drain Tube (1) up through case, then unscrew Prefilter bowl Assembly (2). Clean in mild soap and water and blow dry with low pressure air. B) Remove Dual Stage Element (3) by unscrewing End Cap Retaining Nut (4). C) Inspect the Prefilter Manifold (5) for dirt/contaminates and clean as required. Inspect "O"- Ring (6) for cuts, etc. and replace if required. D) Install new Dual Stage Element and tighten End Cap Retaining Nut. Be sure Element is seated squarely on Manifold boss and End Cap. E) Apply light film of petroleum jelly on Bowl's beveled edge to provide good seal between Bowl and "O"-Ring. HAND TIGHTEN ONLY. F) Guide Drain Tube back through hole in case bottom. G) Dispose of used Dual Stage Element according to local, state and federal regulations. 2. THIRD/FOURTH STAGE CARTRIDGE REPLACEMENT A) Loosen Bracket Screw (7) from Bracket (8). B) Loosen the five Manifold Bolts (9) and remove the front two Bolts. Now slide out the Third (10) and Fourth (11) Stage Aluminum Tube Assemblies. C) Remove the End Cap (12) from Third Stage Aluminum Tube Assembly and slide old Third Stage Filter Cartridge (13) out of aluminum tube. Clean aluminum tube in mild soap and water, dry and install new Third State filter cartridge. Be sure the Flow Direction Arrow on Third Stage Filter Cartridge is pointing down. Remove End Sealing Label (14) and install new End Cap. D) Follow same procedure for the Fourth Stage Filter Cartridge (15) replacement as in step (C). Be sure the Flow Direction Arrow on Fourth Stage Filter Cartridge is pointing up. Also be sure to remove End Sealing Label before installing new End Cap. E) Now slide the Third and Fourth Stage Aluminum Tube Assemblies back in place and install the front two Manifold Bolts. F) Tighten Manifold Bolts in sequence from center outward to 100 inch-pounds (1.15 Kg-M). Repeat sequence and re-torque bolts to 250 inch-pounds (2.88 Kg-M). G) Dispose of used cartridges according to local, state and federal regulations. For more information / 1300 287 935 6 of 13 mail@burwell.com.au / www.burwell.com.au

3. FINAL CHECK AND CALIBRATION A) Pressurize system and check for leaks. B) Flush system with compressed air for five (5) minutes. C) Calibrate Carbon Monoxide Monitor as outlined in Monitor's Information Manual. 7 FIGURE NO. 2 For more information / 1300 287 935 7 of 13 mail@burwell.com.au / www.burwell.com.au

8 AIR SAMPLE TO MONITOR ADJUSTMENT WARNING : SERIOUS INJURY could result if the AIR SAMPLE METERING VALVE is not properly adjusted. Proper sample air flow to CO monitor is required for monitor to give correct CO level readout. AIR SAMPLE METERING VALVE ADJUSTMENT A) Pressurize system and set regulator for proper air flow to respirator(s). B) Adjust Air Sample Metering Valve s adjustment knob so the black floating ball is within the GREEN BOXED area etched on valve body. Proper air sample is now being metered to the CO Monitor. Periodically check to be sure ball is floating in this area. 09/99 For more information / 1300 287 935 8 of 13 mail@burwell.com.au / www.burwell.com.au

9 RECORD KEEPING Record all periodical air quality checks, monitor calibration dates, filter cartridge change intervals and any other service performed on the Portable Air Panel. The Factory shall not be liable for any injury, loss or damage, direct or consequential, arising out of the use of or the inability to use this product, beyond the replacement of defective materials or workmanship. Users of supplied air respirators should evaluate their own particular application and perform their own tests for air quality to determine the suitability for use of this product. For further information, or questions about service or maintenance care of this unit, contact your local distributor or factory at (800) 542-6646. For more information / 1300 287 935 9 of 13 mail@burwell.com.au / www.burwell.com.au

10 SERVICE RECORD PORTABLE AIR PANEL MODEL BA100BMST DATE OF SERVICE SERVICE PERFORMED For more information / 1300 287 935 10 of 13 mail@burwell.com.au / www.burwell.com.au

11 SPECIFICATIONS PORTABLE AIR PANEL MODEL BA100BMST INLET PRESSURE (MAX.) RATED AIR FLOW (MAX.) OPERATING PRESSURE OUTLET PRESSURE RANGE 150 PSIG STATIC (10.4 BAR) 100 SCFM (47.2L/s) 100 PSIG DYNAMIC (6.9 BAR) 0-125 PSIG (0-8.6 BAR) OPERATING TEMP. RANGE 35-150 O F (2-65 O C) OUTSIDE DIMENSIONS 23.25"L X 20.75"W X 9"D (590mm X 527mm X 229mm) WEIGHT (INCLUDING MONITOR) REPLACEMENT FILTER SET 36 LBS. (16.3 Kg.) FB100 For more information / 1300 287 935 11 of 13 mail@burwell.com.au / www.burwell.com.au

For more information / 1300 287 935 12 of 13 mail@burwell.com.au / www.burwell.com.au

13 PORTABLE AIR PANEL MODEL BA100BMST PARTS 1 80025, (1), Black Case 14 80009, (1), Manifold Bracket 2 80086, (1), Blue Inlet Block 15 S006-148, (3), Bracket Bolts 3 80014, (1), Pressure Relief Valve 16 S011-040, (5), Manifold Bolts 4 S603-035, (3),Close Nipple - 3/4" 17 12021, (5), Manifold Washers 5 80104, (1), Prefilter-100 SCFM 18 80118, (2), Tubing St. Elbow 6 Drain Tube Clamp, (1) 19 80119-B, (1) Tubing 7 S711-020, (1) Drain Tube 20 80075, (1), Regulator 8 80001, (1), Blue Base 21 80074, (1), Blue Outlet Block 9 80005, (1), Al Tube Third Stage 22 S687-001, (2), Hex Plug - 1/8" 10 80005, (1), Al Tube Fourth Stage 23 S603-001, (1), Close Nipple - 1/8" 11 80114, (1), Base Bracket 24 80213, (1), Sample Metering Valve 12 80083, (1), Blue Manifold 25 80261,(1),Sampling Hose Connector 13 80127, (1), "CO" Monitor 26 80076, (1), Pressure Gauge, 0-160 For more information / 1300 287 935 13 of 13 mail@burwell.com.au / www.burwell.com.au