Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Similar documents
METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE)

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

Gas Mixture Two Components. Gas Mixtures Two Components

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

ISO INTERNATIONAL STANDARD

Multiple Gas#5 GC configuration Jan 2016

It s Just Calibration How Hard Can It Be?

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Agilent 490 Micro GC Natural Gas Analyzer

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

North Sea Flow Measurement Workshop th October 2017

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Vehicle and Engine Emissions Testing

Protocol Gas Verification Program (PGVP)

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

Experiment GC : Analysis of BTEX by GC-FID

Gas Metrology. SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

Rapid and Reliable Detection of Dissolved Gases in Water

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards

Nonrefillable Cylinders

GAS CHROMATOGRAPHY MAINTENANCE USING UNCERTAINTY BASED CBM

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

CALIBRATION. empanelled laboratories following standard procedures and using certified reference standards.

This test shall be carried out on all vehicles equipped with open type traction batteries.

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

DEPARTMENT OF ENVIRONMENTAL PROTECTION Air Quality. 35 P.S. Sec (Air Pollution Control Act) and Act 95 (75 PA C.S.

Improved Reliability in Natural Gas Energy Measurement with the Revised ISO Standard. Gerard Nieuwenkamp - VSL

Sr. No. General points CPCB Comments 1. Print option, acknowledgement of submission for CPCB OCEMS Protocol

Technical Manual. Sensepoint XCD Gas Detector

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

Retention Time Locking: Concepts and Applications. Application

Gases and welding distributors continue

IDEM Indiana Department of Environmental Management Office of Land Quality P.O. Box 6015 Indianapolis, IN OLQ PH: (317)

Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED. A Division of Norco, Inc.

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

PRECISION GAS MIXTURES

Comparison APMP.QM-S2.1 Oxygen in nitrogen at atmospheric level

Subparts AA and BB - The Other RCRA Air Rules. Beginners aabb

HiQ Specialty Gases and Specialty Equipment. Precision matters. In everything we do.

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

ISO 7941 First edition

Gas mixtures. Individual solutions specifically for your application

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

Instrument Performance Evaluation, comparing ISO and NORSOK I-106 Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk Analyse North Sea Flow

PENNSYLVANIA DEPARTMENT OF ENVIRONMENTAL PROTECTION

Proposed Uniform Standards for Equipment Leaks

The analysis of complex multicomponent

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

(NIST Special Publication )

Date of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Generating Calibration Gas Standards

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Automated Determination of Dissolved Gases in Water Anne Jurek

There is no protection without detection. HiQ specialty gases, equipment and services for environmental monitoring.

There is no protection without detection. HiQ Specialty Gases, Equipment and Services for Environmental Monitoring.

Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure

PQAO Training 2017 Karl Tupper, San Luis Obispo APCD

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions

GAS CHROMATOGRAPHY AND SAMPLE LET DOWN SYSTEM CONDITIONAL BASED MONITORING AND LIVE UNCERTAINTY CALCULATION

The Gold Standard in Calibration Mixtures

FUNDAMENTALS OF GAS CHROMATOGRAPHY, GAS QUALITY & TROUBLESHOOTING Larry Ewing Chandler Engineering Company LLC

EN ISO/IEC Technical Requirements

APPLICATION NOTE. GC Integrated Permeation Device

Research and Development. Money in the Pipeline

c) Loss of product or replacement when units are being serviced All of these are factors that need to be considered by the user.

Operation and Maintenance Manual

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司

AUK Environmental Sensor. Determining the refractive index of the air. AUK Environmental Sensor

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Thermo Scientific Model 146i Principle of Operation

Air Monitoring Directive Chapter 7: Calibration

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

VOC Emissions From the Drying of Red Alder Lumber

Applications Note: Use of "pentane equivalent" calibration gas mixtures

Scientific and Technical Gases Ltd

Leaders in Calibration Gas Solutions

Advanced Test Equipment Rentals ATEC (2832)

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Transcription:

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Today s Discussion QA/QC requirements under MACT CC for Gas Chromatographs (GCs) List of Compounds Surrogate Compounds Performance Specification (PS) 9 MACT CC modifications Areas of Concern/Interest How broadly applicable AMP for mass spectrometry compares (Alt 119) SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 2

Calibration Option 1 The owner or operator must use a calibration gas or multiple gases that include all of compounds listed in paragraphs (e)(2)(i)(a) through (K) of this section that may be reasonably expected to exist in the flare gas stream and optionally include any of the compounds listed in paragraphs (e)(2)(i)(l) through (O) of this section. All of the calibration gases may be combined in one cylinder. If multiple calibration gases are necessary to cover all compounds, the owner or operator must calibrate the instrument on all of the gases. (A) Hydrogen. (B) Methane. (C) Ethane. (D) Ethylene. (E) Propane. (F) Propylene. (G) n-butane. (H) iso-butane. (I) Butene (general). It is not necessary to separately speciate butene isomers, but the net heating value of trans-butene must be used for co-eluting butene isomers. (J) 1,3-Butadiene. It is not necessary to separately speciate butadiene isomers, but you must use the response factor and net heating value of 1,3-butadiene for co-eluting butadiene isomers. (K) n-pentane. Use the response factor for n-pentane to quantify all C5+ hydrocarbons. Optional - (L) Acetylene. (M) Carbon monoxide. (N) Propadiene. (O) Hydrogen sulfide. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 3

Calibration Option 2 The owner or operator must use a surrogate calibration gas consisting of hydrogen and C1 through C5 normal hydrocarbons. All of the calibration gases may be combined in one cylinder. If multiple calibration gases are necessary to cover all compounds, the owner or operator must calibrate the instrument on all of the gases. If the owner or operator chooses to use a surrogate calibration gas under paragraph (e)(2)(ii) of this section, the owner or operator must comply with paragraphs (e)(3)(i) and (ii) of this section. (i) Use the response factor for the nearest normal hydrocarbon (i.e., n-alkane) in the calibration mixture to quantify unknown components detected in the analysis. (ii) Use the response factor for n-pentane to quantify unknown components detected in the analysis that elute after n-pentane. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4

Performance Specification (PS) 9 40 CFR Part 60, Appendix B - SPECIFICATIONS AND TEST PROCEDURES FOR GAS CHROMATOGRAPHIC CONTINUOUS EMISSION MONITORING SYSTEMS IN STATIONARY SOURCES Presurvey Sample Analysis and GC Selection. Determine the pollutants to be monitored from the applicable regulation or permit and determine the approximate concentration of each pollutant (this information can be based on past compliance test results). Select an appropriate GC configuration to measure the organic compounds. The GC components should include a heated sample injection loop (or other sample introduction systems), separatory column, temperature-controlled oven, and detector. If the source chooses dual column and/or dual detector configurations, each column/detector is considered a separate instrument for the purpose of this performance specification and thus the procedures in this performance specification shall be carried out on each system Similar requirement for mass spec but can augment the minimum list. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 5

PS 9 - Sampling System. The sampling system shall be heat traced and maintained at a minimum of 120 C with no cold spots. All system components shall be heated, including the probe, calibration valve, sample lines, sampling loop (or sample introduction system), GC oven, and the detector block (when appropriate for the type of detector being utilized, e.g., flame ionization detector). SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 6

PS 9 - Calibration Gases. Obtain three concentrations of calibration gases certified by the manufacturer to be accurate to within 2 percent of the value on the label. A gas dilution system may be used to prepare the calibration gases from a high concentration certified standard if the gas dilution system meets the requirements specified in Test Method 205, 40 CFR Part 51, Appendix M...The calibration gas concentration of each target analyte shall be as follows (measured concentration is based on the presurvey concentration determined in Section 6.1). NOTE: If the low level calibration gas concentration falls at or below the limit of detection for the instrument for any target pollutant, a calibration gas with a concentration at 4 to 5 times the limit of detection for the instrument may be substituted for the low-level calibration gas listed in Section 7.1.1. Low-level. 40 60 percent of measured concentration. Mid-level. 90 110 percent of measured concentration. High-level. 140 160 percent of measured concentration, or select highest expected concentration. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 7

PS 9 - Performance Audit Gas. A certified EPA audit gas shall be used, when possible. A gas mixture containing all the target compounds within the calibration range and certified by EPA s Traceability Protocol for Assay and Certification of Gaseous Calibration Standards may be used when EPA performance audit materials are not available. The instrument relative error shall be 10 percent of the certified value of the audit gas. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 8

PS 9-7-Day Calibration Error (CE) Test Period. At the beginning of each 24-hour period, set the initial instrument setpoints by conducting a multi-point calibration for each compound. The multi-point calibration shall meet the requirements in Section 13.3. Throughout the 24-hour period, sample and analyze the stack gas at the sampling intervals prescribed in the regulation or permit. At the end of the 24 hour period, inject the three calibration gases for each compound in triplicate and determine the average instrument response. Determine the CE for each pollutant at each level using the equation in Section 9 2. Each CE shall be 10 percent. Repeat this procedure six more times for a total of 7 consecutive days. Calibration Precision and Linearity any one injection shall not deviate more than 5% from the average concentration. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 9

PS 9 - Performance Audit Test Periods. Conduct the performance audit once during the initial 7-day CE test and quarterly thereafter. Sample and analyze the EPA audit gas(es) (or the gas mixture prepared by EPA s traceability protocol if an EPA audit gas is not available) three times. Calculate the average instrument response. Report the audit results as part of the reporting requirements in the appropriate regulation or permit (if using a gas mixture, report the certified cylinder concentration of each pollutant). SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 10

PS 9 Calibration and Standardization Initial Multi-Point Calibration. After initial startup of the GC, after routine maintenance or repair, or at least once per month, conduct a multi-point calibration of the GC for each target analyte. The multi-point calibration for each analyte shall meet the requirements in Section 13.3. Daily Calibration. Once every 24 hours, analyze the mid-level calibration standard for each analyte in triplicate. Calculate the average instrument response for each analyte. The average instrument response shall not vary more than 10 percent from the certified concentration value of the cylinder for each analyte. If the difference between the analyzer response and the cylinder concentration for any target compound is greater than 10 percent, immediately inspect the instrument making any necessary adjustments, and conduct an initial multi-point calibration as described in Section 10.1. Calibration Precision and Linearity any one injection shall not deviate more than 5% from the average concentration. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 11

PS 9 MACT CC Modifications Follow the procedure in Performance Specification 9 of 40 CFR part 60, appendix B, except that a single daily mid-level calibration check can be used (rather than triplicate analysis), the multi-point calibration can be conducted quarterly (rather than monthly), and the sampling line temperature must be maintained at a minimum temperature of 60 C (rather than 120 C). SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 12

Comparison PS 9 vs Table 13 Specification PS 9 MACT CC Daily Calibration Check for Each Analyte Triplicate mid-level Single mid-level Multi-point Calibration Monthly Quarterly Sampling Line Temperature 120 C 60 C SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 13

Areas of Concern/Interest Sample line minimum temperature below 100 C could have issues of condensation (plugging) for sample system or analyzer. The beginning of the out-of-control period is the hour a performance check (e.g., calibration drift) that indicates an exceedance of the performance requirements is conducted. The end of the out-of-control period is the hour following the completion of corrective action and successful demonstration that the system is within the allowable limits. The owner or operator shall not use data recorded during periods the CPMS is out of control in data averages and calculations, used to report emissions or operating levels, as specified in paragraph (d)(3) of this section. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 14

Areas of Concern/Interest The Presurvey Sample Analysis in PS 9 allows for the use of past compliance test results. No mention of engineering evaluations No mention of how many samples are required or under what operating scenarios The CPMS must be capable of measuring the appropriate parameter over the range of values expected for that measurement location. Is the parameter NHV or analyte concentration? The more analytes with potentially high concentrations will increase the number of cylinders needed for daily CE and time spent in required monitoring system quality assurance or quality control activities. The data recording system associated with each CPMS must have a resolution that is equal to or better than the required system accuracy. Required accuracy is 10% of each analyte. SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 15

Questions? Herman Holm Director Specialty Environmental Services Spectrum Environmental Solutions, LLC herman@spectrumenvsoln.com 678-727-2147 SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 16