Best Practice for Identifying Leaks in GC and GC/MS Systems

Similar documents
Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Mass Spec will not Autotune

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Agilent G3180B Two-Way Splitter Kit

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Agilent Inlet EPC Leak Test

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Products Designed to Solve Nagging GC Problems

6890 GC Site Preparation

Mass Spec will not pump down

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Agilent 7697A Headspace Sampler

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

ms NoVent II Instruction Booklet

no peaks missing peaks too small peaks increasing retention decreasing retention

Super-Clean Gas Filters from Restek

Advances in Capillary Column Backflushing

Retention Time Locking: Concepts and Applications. Application

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Agilent G Rebuilt Electron Capture Detector

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

The Open-Split Capillary Interface. Part No

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

STANDARD OPERATING PROCEDURES

Carrier Gases in Capillary GC

3 Flow and Pressure Control

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

Agilent 7667A. Mini Thermal Desorber. Site Preparation

A New GC/MS System Designed for Helium Carrier Gas Conservation

INSTRUCTION MANUAL GAS TRAK

Agilent 490 Micro GC Natural Gas Analyzer

Zebron Gas Management

Tips and Tricks of Injector Maintenance. Abby Folk Agilent GC Technical Support August 2009

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Leak Checking Large Vacuum Chambers

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

Multiple Gas#5 GC configuration Jan 2016

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent 7697A Headspace Sampler

APPLICATION NOTE. GC Integrated Permeation Device

Your local gas generation partner. Precision series Modular gas generation solution for GC.

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

CSA Sample Draw Aspirator Adapter Operator s Manual

Addressing the World Shortage of Helium

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Connecting a SSL/PTV Backflush System

Welker Constant Pressure Cylinders With Welker Solid Indicator, Syringe T-Handle, or Vortex Mixer

7820A GC Site Preparation Checklist

Agilent 1290 Infinity Pump Head Maintenance

Agilent G3185B QuickSwap Accessory

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

GCMSD-Headspace Analysis SOP

Operation of the Perkin Elmer TGA-GC/MS

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

Instant Connect Helium Saver Injector Module

CalCheck. Instrument User Manual V1.8

Gases and welding distributors continue

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

PANACEA P200 STAINLESS STEEL USER MANUAL

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

HiQ laboratory gas generators.

100C Air Compressor Kit

PANACEA P125 USER MANUAL

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3

Pilot Check Valve: Metal Body Type

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

RD(H)20/25 Pressure-Reducing Regulator User Manual

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Transcription:

Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible, and accurate results. The symptoms typically observed from leaks are presented, along with a discussion of key areas of focus to identify leaks and how to correct them. Key facets discussed in obtaining optimal system performance include the use of high quality carrier gas, selection of the proper ferrule, proper just-tight-enough (JTE) ferrule installation, checking for leaks in GC and GC/MS systems, and an innovative self-tightening column nut for maintaining leak-free connections. Leak symptoms GC gas leaks fall into two distinct categories: large leaks that prevent the instrument from functioning and smaller leaks that allow the system to operate, but negatively impact chromatography. Large leaks typically prevent a system from reaching a ready state, leading to an electronic pressure control (EPC) safety shutdown. These types of leaks can result from a column not being installed in the expected inlet, a column not being connected to the expected detector, a broken column, broken or loose fittings, broken ferrules, cored septa, or tubing blockage, to call out a few possibilities. The cause of these symptoms typically can be rooted out quickly by visual inspection or review of the method settings. Identifying smaller leaks that allow the system to continue to operate can be more involved. Symptoms of smaller leaks can include constant cycling of actual pressure readings (oscillations greater than 0.02 psi), poor retention time reproducibility, higher than typical background, higher than typical bleed (particularly at temperatures greater than 230 C), baseline drift, higher than usual inlet activity, tailing peaks, the need for more frequent inlet maintenance, and poor area reproducibility.

Figure 1 shows the elution of US-EPA 8081 pesticides on an Agilent J&W DB-1701 phase before and after exposure to 1,000 µl/l oxygen in helium carrier gas. After just 10 injections, column bleed increased significantly and a shift to shorter peak retention times was apparent. pa 18 No O 2 exposure 16 14 12 10 8 6 4 6 8 10 12 14 16 18 20 min pa 18 16 14 12 10 8 6 10 injections O 2 exposure Increased bleed shift to shorter retention 4 6 8 10 12 14 16 18 20 min Figure 1. US-EPA 8081 pesticides before and after exposure to 1,000 µl/l oxygen in helium. Carrier Gas Considerations High quality carrier and detector gases of known purity are essential for obtaining optimal results in gas-phase analysis. Agilent specifies carrier and detector gas purity of at least 99.9995% (5.5 nines). Zero-grade air is recommended for flame detectors [1]. Inline indicating gas traps, such as Agilent Gas Clean Filters [2], are highly recommended to remove hydrocarbon, moisture, and oxygen. Gas certification testing and product descriptions vary by supplier and so obtaining a certificate of analysis (COA) for the gases in use is essential to understand gas quality. On the COA, key items to look for include tests conducted, specification for contaminants, and indications of whether testing was done on individual (preferred) or representative cylinders from a batch. Ferrule Selection Selecting an appropriate ferrule for the column tubing size and particular fitting being used are critical for minimizing potential leaks and keeping the flow path free of contamination. Agilent J&W columns require one size ferrule for 0.1 to 0.25 mm id columns, while 0.32 mm and 0.53 mm id columns each require ferrules with larger diameter holes to accommodate the wider outside diameter of these columns. Ferrule material choice is also important for achieving the desired results for specific applications. Graphite ferrules are a popular choice for general-purpose and high-temperature applications (above 350 C) but, typically, are not as contaminant-free as polyimide/graphite or metal ferrules. Further, graphite is a porous material and slightly permeable to gases, creating a very small continuous leak. Graphite also has a tendency to flake off, becoming a source of contamination. Pure polyimide ferrules are recommended for use outside of heated temperature zones only as they shrink dramatically with exposure to heat cycling. 2

Polyimide/graphite ferrules are a good choice for GC/MS and trace-level analysis, but they also have a tendency to shrink with repeated heat cycles, forcing the operator to snug the fitting repeatedly to avoid leaks. Ferrule shrinkage results in a tendency for analysts to over tighten fittings using polyimide/graphite ferrules. Flexible metal ferrules are recommended for use with Capillary Flow Technology (CFT) devices as they are specifically designed for the fittings in these devices [3]. Analysts are finding flexible metal ferrules an attractive alternative to other ferrules for standard column connection, such as the split/splitless inlet. Table 1 identifies some common benefits offered by various capillary column ferrules. Avoid Over Tightening Be aware that over tightening can break the column or permanently damage fittings and actually produce leaks. The Agilent UltiMetal Plus Flexible Metal ferrule was designed to reduce column breakage by compressing around the column. With a deactivated surface, these stainless-steel ferrules provide a robust and inert leak-free connection. Over tightening of Swagelok, SilTite or UltiMetal Plus Flexible Metal ferrules can damage fitting threads, making it impossible to obtain a seal, and resulting in costly instrument repairs. Carefully read and follow manufacturer s instructions on fitting installation and use to avoid chronic leaks from damaged fittings. Proper installation of graphite, polyimide/graphite, inlet seals, O-rings, and septa is also critical to maintaining leak free connections. Just tight enough (JTE) is the goal for proper installation of these somewhat pliable components. If tight is good, tighter is not better (TNB), as these pliable materials can easily be crushed beyond their design specifications, causing them to leak sooner and more often. In extreme cases with repeated over tightening of brass mass spec transfer-line nuts, the nuts themselves can crack and potentially cause permanent damage to the mass transfer line. The Agilent septum nut has a C-shaped clip at the top that should not be turned more than 3/4 of a turn past where it begins turning with the nut assembly when a septum is being installed. Over tightening of the septum nut will cause premature septum coring with repeated injections, which in turn causes the septum to leak during a run. The septum nut is another fitting that needs to be JTE. Table 1. Ferrule material selection attributes. Requirement Low torque needed; compatible with finger tight nuts Low cost Re-use of ferrule Graphite Polyimide/ graphite Will not fragment Inertness Leak free * High temp above 350 C Recommended for mass spectrometry interface Required for CFT device Pre-swage for precise height into fitting * When using self-tightening column nut, avoid over tightening. Checking GC Connections Flexible metal Checking all fittings for leaks immediately after installation, maintenance, and periodically while in use is an excellent practice. A handheld detector capable of detecting a helium leak of 0.0005 ml/min in air is available from Agilent [4]. Handheld leak detectors are particularly useful for finding leaks quickly either inside or outside the GC oven. It is good practice to always use a leak detector to check for leaks each time a column, fitting, or cylinder is changed. An excellent starting point for system troubleshooting is to first check for potential leaks. Avoid using water soap solutions, as these can be drawn back into the GC flow path, severely impacting chromatographic results even to the point of causing permanent column damage. Another very useful resource is the Agilent 7890 Series GC troubleshooting manual, which has an extensive checking for leaks section on page 126 [5]. A video describing an instrument software-based leak-checking tool on a 7890 GC is available on the Agilent website [6]. 3

Checking a GC/MS for Leaks A vacuum or ion gauge, if ordered with your instrument, is useful for monitoring vacuum pressures under typical operating conditions in GC/MS. An ion-gauge kit is also available for the Agilent 5977 Series GC/MS [8]. A vacuum gauge is very useful for isolating potential leaks to either the vacuum (MS) or pressurized (GC) side of the instrument. Vacuum readings in the 10 5 or 10 6 Torr range are typical for a system holding vacuum with a flow rate of 1.0 ml/min on a 30 m 0.25 mm, 0.25 µm GC column. When the MSD is capped and pumped down, vacuum readings typically drop to the 10 6 or 10 7 Torr range in the absence of a leak. If the vacuum pump does not reach these levels relatively quickly, a leak somewhere in the MS is indicated. Make sure the purge vent is closed, the transfer line fitting is installed correctly, and that the large O-ring on the vacuum side plate is positioned correctly. A software-based performance check of air and water is available under the tuning selection tab. This check looks at GC/MS ion traces of molecules typically found in air relative to ion 69 found in the calibrant. Ions 18 (water), 28 (N 2 ), 32 (O 2 ), 44 (CO 2 ), and 69 (typical base peak from PFTBA used during auto-tune) are all monitored. Nitrogen (28) levels above 10% relative to the 69 peak indicate that the system has not had sufficient time to pump down or that there is an air leak. An air leak will typically show nitrogen:oxygen in a 4:1 ratio. Water (18) is also typically present, particularly after a system had been vented and exposed to ambient air. An equilibrated leak-free system should show nitrogen (28) well below 10% with oxygen 32 at approximately ¼ of the signal seen for nitrogen, and ideally water (18) lower than the N 2 (28) peak. Troubleshooting leaks in GC/MS is a process of elimination, looking at each site where a leak can occur. A fluorocarbon (for example, 1,1,1,2-tetrafluoroethane, ions 69 and 83) or argon (ion 40) spray can with a plastic tube to direct the flow is very useful in isolating a leak. A short spray at a suspect point and monitoring the appropriate ions in manual tune is a powerful tool for isolation. Key points to check are the transfer line connection in the oven, septum nut, column nut, and the large O-ring on the vacuum plate of the MS. Once a leak has been isolated the leak can be remedied by replacing a septum, resetting a column connection, or cleaning the O-ring on the vacuum plate, and reinstalling it back into the groove on the plate [7]. Innovations to Minimize Leaks Figure 2 shows a total-ion chromatogram for an air and water check on a system that is operating normally. In this case, self-tightening column nuts were installed at the transfer line and inlet fittings. These column nuts provide a leak-free seal using a short polyimide/graphite ferrule at both column connections, without the need to retighten the fitting after more than 300 heat cycles [9]. Use of these column nuts eliminates the need to retighten the inlet or mass spec transfer-line connections after oven heat cycling. Furthermore, because very low torque is needed to make a leak-free seal when using the Agilent self-tightening column nuts, these nuts are installed using only fingers, not wrenches, which eliminates the risk of over tightening and damage to the fittings (Figure 3). 4

Figure 2. Example air and water check. Conclusions By using tools, supplies and best practices that provide a leak-free GC or GC/MS, analysts can improve performance and productivity of their system. Agilent UltiMetal Plus Flexible Metal ferrules provide robust leak-free column connections, along with an inert surface for fittings in the sample flow path. The Agilent innovative self-tightening column nuts using standard short polyimide/graphite ferrules eliminate the need to retighten GC column fittings, including the mass spec transfer line, after repeated heat cycling. These new fittings also have the advantage of using only short polyimide/graphite ferrules for inlet, detector, and mass-transfer-line connections. Figure 3. Agilent self-tightening column nuts installed at the transfer line and inlet fittings. Following the best practices described in this technical overview and accessing the references below will help GC and GC/MS users identify potential air leaks, where to find them, and how to fix and prevent them quickly. One rule of thumb is to adjust fittings, septa, and O-ring seals to be JTE for the best results. 5

References 1. Anon. Agilent 7890 Series GC Site Preparation Checklist, Revision 1.6. Agilent Technologies, Inc. Publication number G3430-90001 (14 November, 2013). 2. Anon. Agilent Gas Clean Filter User Manual. Agilent Technologies, Inc. Publication number 5973-1528 (2012). 3. Anon. Agilent ferrule selection. Agilent Technologies, Inc. 4. Anon. Agilent G3388B Leak Detector Operation Manual. Agilent Technologies, Inc. Publication number G3388-90005 (2012). 5. Anon. Agilent 7890 Series GC Troubleshooting Manual. Agilent Technologies, Inc. Publication number G3430-90053 (2013). 6. Anon. GC Leak Check Video. Agilent Technologies, Inc. 7. Anon. The right connections make all the difference. Brochure, Agilent Technologies, Inc. Publication Number 5991-3155EN (2014). 8. Anon. Agilent 5977 Series MSD Operation Manual (G3397B Ion Gauge Controller Kit). Agilent Technologies, Inc. Publication number G3870-90003 (2013). 9. K. Lynam Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application note, Agilent Technologies, Inc. Publication number 5991-3612EN (2013). www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2014 Printed in the USA January 23, 2014 5991-3899EN