Adiabatic Compensation for Digiquartz Intelligent Transmitters

Similar documents
ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

REASONS FOR THE DEVELOPMENT

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

HART Communications Board

Description of Device Parameters Proline Prowirl 200 HART. Vortex flowmeter. Products Solutions Services. Main menu Language.

Series 3730 and Series 3731 EXPERTplus Valve Diagnostics with Partial Stroke Test (PST)

Using Modbus Protocol with the ALTUS Net Oil Computer. Instruction Manual

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

Hydro-Control V User Guide

OFI MODEL 20 CONSTANT SPEED BLENDER INSTRUCTIONAL MANUAL

ACV-10 Automatic Control Valve

NATIONAL UNIVERSITY OF SINGAPORE. EE3302/EE3302E Industrial Control Systems E1: ADVANCED CONTROL SYSTEMS

INSTRUCTION MANUAL. Smart PT. February Campbell Scientific (Canada) Corp Avenue NW Edmonton AB T5L 4X4

KEM Scientific, Inc. Instruments for Science from Scientists

EZ Boom 2010 System for the EZ Guide 500 Lightbar Triangle Ag Services Users Guide

Digital Level Control One and Two Loops Proportional and Integral Control Single-Loop and Cascade Control

Autodesk Moldflow Communicator Process settings

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

Completion of Hydrometric Survey Notes (HSN)

Fail Operational Controls for an Independent Metering Valve

HOBO MX Water Level Logger (MX2001-0x) Manual

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

BUBBLER CONTROL SYSTEM

DeeterFlow Automatic Liquid Dispenser

Surface Tracking Feature

Cochran COMMANDER EMC-20H With Helium Owner's Manual

COMPRESSOR PACK SUCTION CONTROLLER TYPE: LP41x

Autopilot setup. VRF (Virtual Rudder Feedback) calibration. Software setup NSS evo2 Installation Manual

Figure SM1: Front panel of the multipatcher software graphic user interface (GUI) at the beginning of multipatcher operation.

Micro Motion Multiphase Applications

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Fail operational controls for an independent metering valve

Level Process Control. Penn State Chemical Engineering

HKA TECHNOLOGIES INC. TEMPERATURE COMPENSATION

GNX 20/21. Owner s Manual

USER MANUAL. Intelligent Diagnostic Controller IDC24-A IDC24-AF IDC24-AFL IDC24-F IDP24-A * IDP24-AF * IDP24-AFL * IDP24-F * 1/73

Data Sheet T 8389 EN. Series 3730 and 3731 Types , , , and. EXPERTplus Valve Diagnostic

PRODUCT MANUAL. Diver-MOD

Gerald D. Anderson. Education Technical Specialist

PMI Pulse Decay Permeameter for Shale Rock Characterization Yang Yu, Scientist Porous Materials Inc., 20 Dutch Mill Road, Ithaca NY 14850

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

HEATEC TEC-NOTE. Setting Siemens Pressure Transmitter. Used on Heatec Vertical Asphalt Tanks. Publication No , Revised

Waves & Interference

Digital Pressure and Temperature Sensor

Cochran EMC-20H With Helium Owner's Manual

Engineering: Measurement Technology Pressure/Level (SCQF level 6)

Quick Start Setup Instructions for Raven LRC & AgXcel Harness for NH3 Profile plus GX40 Synergist

BERNOULLI EFFECTS ON PRESSURE.ACTIVATED W ATER LEVEL GAUGES

Rotel RX-1052 RS232 HEX Protocol

Oxymax COS61/COS61D. Technical Information

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

DIVECOMPUTER.EU USER MANUAL V3.0

PRODUCT MANUAL. Diver-DCX

Technical Bulletin, Communicating with Gas Chromatographs

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

Preparation for Salinity Control ME 121

Calibration Summary of Test Report No.: Sample

Next Generation Quartz Pressure Gauges

Module 9. Advanced Data Processing

The Univentor 1250 Anaesthesia Unit

Flow Vision I MX Gas Blending Station

UNIVERSITY OF WATERLOO

Doc. no.dit om005 PRODUCT NAME. IO-Link/ELECTRO-PNEUMATIC REGULATOR. MODEL / Series / Product Number ITV*0*0-IO****-X395

Average Accuracy within 0.1 mph, Calls Out Speeds, Tracks Results

1999 Microcomponents Technology SLYA017A

TR Electronic Pressure Regulator. User s Manual

Technical Information Conducal CLY421

Ranger Walking Initiation Stephanie Schneider 5/15/2012 Final Report for Cornell Ranger Research

Micro Motion Advanced Phase Measurement

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

CCT-7320/ROC-2313 Reverse Osmosis Controller

S-CURVE DOCUMENTATION AND TUTORIAL FOR THE S-CURVE ON THE COOL MUSCLE

Overview. 2 Module 13: Advanced Data Processing

Injection Controller Program User Manual (for FloBoss 107 and ROC800-Series)

Application Notes. SLP85xD Load Cells

PTG100 Precision Test Gauge

Touch Screen Guide. OG-1500 and OG Part # T011

Application Note AN-107

STICTION: THE HIDDEN MENACE

XSeries G4 Plunger Application rev. AA

Burner Management System DEMO Operating instructions

GNX 20/21. Owner s Manual

Safety Systems Based on ND Pressure Monitoring

Cochran EMC-20H FO2/PO2 Owner's Manual

SDI mA Pressure Transducer (Submersible + Dry) Model WL2100 Operating Manual

CHAPTER 7: THE FEEDBACK LOOP

Tutorial for the. Total Vertical Uncertainty Analysis Tool in NaviModel3

High Ping Rate Profile Water Mode 12

Metatron 21. The proven specialist for professional milking management. GEA Farm Technologies The right choice. GEA Milking & Cooling WestfaliaSurge

Special Documentation Proline Promass 500 Coriolis Flowmeter Modbus RS485 Petroleum Application Package

ISO INTERNATIONAL STANDARD

APPLICATION NOTE. OJ-DV Modbus protocol. OJ Drives A DRIVES PROGRAMME DEDICATED TO VENTILATION SOLUTIONS E 03/18 (PDJ) 2018 OJ Electronics A/S

Deuterium more than just for locking: Part II. H Topshim. Donna Baldisseri Boston Massachusetts Pre-ENC Workshops 2014, March 22

Characterizers for control loops

Air Compressor Control System for Energy Saving in Manufacturing Plant

Safety Manual VEGAVIB series 60

DBI version 002 User Manual US version Free Balloon Flight Instrument

Code Basic module and level control complete with optionals code

Transcription:

Adiabatic Compensation for Digiquartz Intelligent Transmitters Paroscientific, Inc. 4500 148 th Ave. N.E. Redmond, WA 98052, USA Tel: (425) 883-8700 Fax: (425) 867-5407 www.paroscientific.com support@paroscientific.com The standard by which other standards are measured 2012 Paroscientific, Inc. 1

Adiabatic Compensation for Digiquartz Intelligent Transmitters 1. Introduction Paroscientific recently introduced the ability to compensate for adiabatic effects in its line of Intelligent Transmitter products. Products that support the feature include the 1000 and 9000 Series Intelligent Pressure Transmitters and 8000 Series Intelligent Depth Sensors built with 4000 Series pressure transducers. The feature is available in firmware version R5.12 and later. The adiabatic compensation feature was designed to address a specific application involving a consistent, repeatable positive-going pressure event. It is not possible to compensate for negative-going, inconsistent, or non-repeatable pressure events at the present time. 2. Concept Adiabatic compensation involves characterizing the adiabatic effect, setting a group of serial configuration parameters accordingly, and verifying that the adiabatic compensation has been properly configured. The serial configuration parameters are as follows: Command AA AC AH AM AR Description Read/write the adiabatic excursion amplitude in current pressure units. Read/write the adiabatic compensation time constant in seconds. Read/write the adiabatic compensation trip threshold in current pressure units. Read/write the adiabatic compensation mode enable. Read/write the adiabatic compensation re-arm threshold in current pressure units. 3. Characterizing the Adiabatic Effect The adiabatic effect is characterized by acquiring pressure data during the pressure event, graphing it vs. elapsed time, and analyzing the pressure data per Figure 1. When values A through H have been determined, set the AA, AC, AH, and AR parameters accordingly. Set the AM parameter to 1 to enable adiabatic compensation. Once again, acquire pressure data during the pressure event, graph it vs. elapsed time, and verify adiabatic compensation. It may be necessary to tune one or more parameters for optimum performance. Complete documentation for the serial commands used to configure adiabatic compensation can be found below. 2012 Paroscientific, Inc. 2

Figure 1: Pressure Pulse Showing Adiabatic Effects (not to scale) A Adiabatic pressure overshoot amplitude from the steady state level [E]. This value is entered into the AA serial command (see below) in the current pressure units. B Adiabatic pressure pulse time constant that is entered into the AC serial command (see below) in seconds. C Pressure pulse trip threshold turns ON the active adiabatic compensation and is entered into the AH serial command (see below) in the current pressure units and establishes the start time. D Pressure pulse re-arm threshold turns OFF the active adiabatic compensation and resets the algorithm. This value is entered into the AR serial command (see below) in the current pressure units. E Pressure pulse steady state level that would be generally found greater than five time constants [B] after the start of the pulse. F Pressure pulse duration before returning below the re-arm threshold [D]. G Adiabatic pressure overshoot level after one time constant [B], which is down approximately 37% from the peak overshoot amplitude (A * e -1 ). H Quiescent pressure level prior to and after the pressure pulse. 2012 Paroscientific, Inc. 3

4. Serial Command Descriptions Definitions: * = Start character. HH = Host address ASCII decimal. UU = Device address ASCII decimal. rl = Carriage return/line feed (0DH 0AH) AA : Read/write the adiabatic excursion amplitude in current pressure units. ACTION: Reads or writes (EW command required) of the adiabatic excursion amplitude is a floating point number specified in the currently used pressure units. This value is used in the calculation of the adiabatic pressure error as the variable: A (see above calculation, also [A] of figure 1). The value would generally be determined by noting the steady state pressure of the pulse after the adiabatic effects diminish (after at least five time constants) and subtracting this value from the experienced peak pressure value of the adiabatic pressure overshoot. *UUHHAArl Read adiabatic excursion amplitude value. *UUHHEW{rl}*UUHHAA=cccccccrl Write new adiabatic excursion amplitude. *HHUUAA=cccccccrl Report adiabatic excursion amplitude value. ccccccc = Adiabatic excursion amplitude (floating point ASCII decimal value.) AC : Read/write the adiabatic compensation time constant in seconds. ACTION: Reads or writes (EW command required) of the adiabatic compensation time constant is a floating point number in seconds. This value is used in the calculation of the adiabatic pressure error as the variable: T TC (see above calculation, also [B] of figure 1). *UUHHACrl Read current time constant value. *UUHHEW{rl}*UUHHAC=cccccccrl Write new time constant value in seconds. *HHUUAC=cccccccrl Report current time constant value. ccccccc = Time constant (floating point ASCII decimal value.) 2012 Paroscientific, Inc. 4

AH : Read/write the adiabatic compensation trip threshold in current pressure units. ACTION: Reads or writes (EW command required) of the adiabatic compensation trip threshold is a floating point number in current pressure units. This value is used in the calculation of the adiabatic pressure error by determining the start time (T START, see above calculation) of the pressure pulse and adiabatic over-shoot (see [C] of figure 1). Also, this value is always greater than or equal to the re-arm threshold of the AR serial command. If the value entered is less than the current AR value, AR will be made equal to the new AH value otherwise, it is left unaltered. *UUHHAHrl Read current trip threshold value. *UUHHEW{rl}*UUHHAH=cccccccrl Write new trip threshold value. *HHUUAH=cccccccrl Report current trip threshold value. ccccccc = Trip threshold (floating point ASCII decimal value.) AM : Read/write the enable state of the adiabatic compensation mode enable. ACTION: Enabling/Disabling the adiabatic compensation mode is done by writing (EW required) an ASCII 0 or 1 in accordance to the following chart: Argument Description 0 Adiabatic compensation is disabled 1 Adiabatic compensation is enabled *UUHHAMrl *UUHHEW{rl}*UUHHAM=drl *HHUUAM=drl d = Boolean value (ASCII 0 or 1) Read adiabatic compensation enable state. Write argument (see above). Returns the current adiabatic compensation enable state. 2012 Paroscientific, Inc. 5

AR : Read/write the adiabatic compensation re-arm threshold in current pressure units. ACTION: Reads or writes (EW command required) of the adiabatic compensation re-arm threshold is a floating point number in current pressure units. This value is used to determine when the active adiabatic compensation is disabled (sampled pressure is less than the AR value) and also to reset the algorithm for the next pressure pulse event that exceeds the trip threshold of the AH command (see also [D] of figure 1). This value cannot be set greater than the trip threshold of the AH command. Thus, as a value of AR is set larger than the current AH value, AH will be made equal to the new AR value. An AR value set less than the current AH value will leave the AH value unchanged. *UUHHARrl Read current re-arm threshold value. *UUHHEW{rl}*UUHHAR=cccccccrl Write new re-arm threshold value. *HHUUAR=cccccccrl Report current re-arm threshold value. ccccccc = Re-arm threshold (floating point ASCII decimal value.) 5. Formulas Adiabatic Pressure Error Calculation: P e ( TCAPTURE TSTART ) TTC = A* e Adiabatic Pressure Compensation: P CORRECTED = P RAW P e where: P e = adiabatic pressure error in current pressure units. P CORRECTED = pressure corrected for adiabatic effects. P RAW = temperature compensated pressure value containing adiabatic pressure effects. A = adiabatic excursion amplitude in current pressure units (see AA, [A] from Figure 1). T CAPTURE = time captured at pressure acquisition(s) after trip threshold event in seconds. T START = start time captured at pressure acquisition of the trip threshold event in seconds. (see AH) T TC = adiabatic time constant in seconds (see AC, also [B] of figure 1). 2012 Paroscientific, Inc. 6