CÜBE. User Manual. American Weigh Scales. CUBE-100 (100g x 0.01g) CUBE-650 (650g x 0.1g)

Similar documents
BCM-Series. User Manual. American Weigh Scales. BCM-100 (100g x 0.01g) BCM-650 (650g x 0.1g)

AXIS. User Manual. American Weigh Scales. AXIS-100 (100g x 0.01g) AXIS-650 (650g x 0.01g)

MS-500. User Manual. MS-500 (500g x 0.1g) Copyright 2010 Fast Weigh Scales All rights reserved. Rev. 2.0

EPSILON. User Manual. EP-5KG (5000g x 1g) Copyright 2011 American Weigh Scales, Inc. All rights reserved. Rev. 1.0

American Weigh Scales, Inc. BLADE Series D i g i t a l P o c k e t S c a l e BLADE-550. User Manual BLADE-400 BLADE-1000

American Weigh Scales, Inc. LB User Manual

STANDARD OPERATING PROCEDURE

User Manual. Portable Balances PL-E

KERN CM Version /2004

Operating Manual Body fat balance

StockWeigh 500 D3632 US-B

All Purpose AP-Series Operation Manual

ANALYTICAL BALANCE OPERATING MANUAL. U.S. Solid USS-DBS Models All Rights Reserved Page 1 USS-DBS MODELS

Sartorius Basic plus Service Manual. WBP5007-e95021

SPECIFICATIONS APCEPH1

Eclipse Precision & Analytical Balances

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

YAMATO CORPORATION 1775 S. Murray Blvd. Colorado Springs, CO USA Tel (719) Fax (719)

Ambient Weather WS-03 Thermo-Hygrometer

Sensoric 4-20 ma Transmitter Board Operation Manual

OEM Manual. MODEL ½ Digit DRUM SCALE

TABLE OF CONTENTS INTRODUCTION 3 SAFETY PRECAUTIONS 3 PACKAGE CONTENTS 4 DEVICE OVERVIEW 4 BUTTON OPERATION SUMMARY 5 BASIC OPERATION 6

DTG - LCD Digital Temperature Gauge USER MANUAL

Altimeter and Compass Watch Instruction Manual

SPS-5. Rate Computing Scale. Operation Manual

CHM Introductory Laboratory Experiment (r17sd) 1/13

Operation Instruction. BSM Series Electronic Precision Balances

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

SAUTER FK Version /2016 GB

Manual Weighingblock VB2 series and Uniscale

Reflow Oven HHL3000 INSTRUCTION MANUAL POHUA - jedyny autoryzowany przedstawiciel w Polsce

CHM 100 / Introductory Laboratory Experiment (r10) 1/11

Relative Humidity Calibration Kit

MicroTim XB. User Manual. Precision Digital Barometric Altimeter / Barometer / VSI. Document Revision 1.0 Firmware Version 3.0

DT 630 ALTIMETER, BAROMETER AND COMPASS WATCH OPERATING INSTRUSTIONS

How to Achieve Optimal Weighing Performance

Operating Instructions Part No

M104 Series: Digital Manometer: 0.1 % of Full Scale

KERN EG/EW Version /02

BASIC LABORATORY TECHNIQUES (Revised )

DIGITAL AUTOMATIC TYRE INFLATOR

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

INSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Fit Cuff. =Fit Cuff=!"#$% IA1B. Model

Infrared Ear and Head Thermometer Instruction Manual

Adam Equipment. DENSITY DETERMINATION KIT for AAA, ADA and ADP SERIES BALANCES OPERATOR S MANUAL

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Instruction Manual. Select Language. English French German Russian. All images in this manual are illustrations only.

Fingertip Pulse Oximeter

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

NOVALYNX CORPORATION MODEL RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL

RAM 4021 Operation Manual

User Manual GRI- 1500Li

ISO 9001 & ISO Rev. G. Authorized Representative in the E.U.: MT Promedt Consulting GmbH Altenhofstrasse 80 D St.

Pneumator. Testo Industrial Services More assurance, better service. Pressure calibrator and measuring instrument.

Adam Equipment PGL SERIES. (P.N. 9326, Revision C5, March 2009)

Adam Equipment. DENSITY DETERMINATION KIT for AFP, AAA and AAA/LE SERIES BALANCES OPERATOR S MANUAL

Harvard Trip Balance BA710. Impact Test Equipment Ltd & User Guide. User Guide

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

ACCUSTRIDE ACCUSTRIDEFM. Digital Clip-on Pedometer WC153. Digital Clip-on Pedometer WC154 ENGLISH

Service Manual. Sartorius Expert. LE-Models Electronic Analytical and Precision Balances. WLE5001-e04022

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

USER S GUIDE DepthTrax 1H HawkEye Portable Depth Finder (DT1H)

Heavy Duty Dissolved Oxygen Meter

Owner s Manual Humiport 10/20

Napa Technology Trouble Shooting. For Premier & Premier PLUS Models

General Accreditation Guidance. User checks and maintenance of laboratory balances

RayPette Plus Autoclavable Pipette. User Manual

PM105 PRESSURE METER USER S MANUAL

RAM Operation Manual

GE Measurement & Control

Infrared Thermometer Instruction manual

RAM 4021-DPX Operation Manual

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

Psyclone TM. Protimeter Thermo-Hygrometer. GE Sensing. Operating Manual. INS7800, Rev. A (September 2007)

I. GENERAL INFORMATION

INSTRUCTION MANUAL D Pedometer with Clip. VIDEO INSTRUCTIONS:

PTG100 Precision Test Gauge

The pipettes cover a volume range from 0.1µl to 10ml.

Sartorius Competence CP Series Sartorius Gem plus

Cover Page for Lab Report Group Portion. Pump Performance

ActiveYield. Operation and Adjustment

Operating Instructions Part No

Instruction Manual HI 9146 Dissolved Oxygen and Temperature Meter

BASIC LABORATORY TECHNIQUES (Revised )

Gilson Aqua-Check Moisture Tester MA-26X

FOOD INTAKE MONITOR. Hardware and Software User s Manual DOC-047 Rev 3.0. Copyright 2009 All Rights Reserved

ali-q TM Gravimetric Verification Procedure

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Application Notes. SLP85xD Load Cells

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Introduction to Scientific Notation & Significant Figures. Packet #6

Operating Instructions

Pipettor. User Manual

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Relative Humidity Calibration Kit

Transcription:

American Weigh Scales CÜBE User Manual CUBE-100 (100g x 0.01g) CUBE-650 (650g x 0.1g) Copyright 2014 American Weigh Scales, Inc. All rights reserved. Rev. 1.0

CUBE Manual Thank you for purchasing the American Weigh CUBE digital pocket scale. Please read all operating instructions carefully before use. Scales are precision measuring devices and should always be handled with proper care. To ensure years of reliable service, keep these simple tips in mind: Do not exceed the scale s maximum capacity. This scale has integrated overload protection, but any overload protection has its limits! Avoid exposure to extreme heat or cold. Scales perform best at normal room temperature. If temperatures have changed dramatically, recalibration may be necessary. Allow your scale to warm up for 30-60 seconds before performing calibration (if available). Store your scale in a clean, dry location. Dust, dirt, and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction. Avoid using your scale in close proximity to strong radio wave emitting devices such as cordless phones. Always weigh on a flat and level surface, free from vibrations and drafts. The corner of a room is usually the most stable. Gently apply all items to be weighed. Do not drop items onto the weighing platform. Avoid dropping your scale. The warranty does not cover damage due to rough treatment or overload. Check the batteries first if you are having any trouble with your scale. This simple step can remedy most scale issues. Calibration should be checked and adjusted occasionally as part of routine scale maintenance.

I. Part List 1. Scale 2. Hard Case / Tray 3. AAA Batteries (x2) II. Key Description 1. M I/O 2. 1. M - Tap to change weighing unit. Hold to enter calibration (see Section VII) 2. I/O - Turns the scale on or off when platform is clear. Zeros the weight of a container or tray if present. III. Battery Installation Proper Battery Orientation

CUBE Manual Low batteries can cause inaccuracies in digital scales. If you are experiencing issues with your scale, first try new batteries. Most times, this will solve the problem. 1. Before using the scale for the first time, check the battery compartment and remove any plastic insulation strips or packaging that may be in place to prevent battery drainage during shipping/storage. 2. If you are replacing the batteries, make sure the positive and negative contacts are properly aligned. IV. Operation 1. Place the scale on a flat, level surface and press the I/O key to turn the scale on. 2. Once the start-up process completes and the scale goes to zero you are ready to weigh. 3. Place the quantity to be weighed onto the weighing platform. 4. Wait a couple seconds for the display to stabilize before taking a weight reading. 5. To turn the scale OFF, press the I/O key again. V. Changing the Unit of Measure 1. Press the M key to change the scales displayed unit of measure. 2. If the scale is turned off then back on, it will default to the last unit used. CUBE-100 & CUBE-650 Units: g (gram) gn (grain) oz (ounce) ozt (troy ounce) ct (carat) tl (tael) dwt (pennyweight)

VI. Using the Tare Feature You can use this scale s tare feature to subtract the weight of an empty container or vessel for net weight determination. NOTE: The weight of the container will subtract from the available capacity (subtractive taring). VII. 1. Place the empty container or vessel on the weighing platform. 2. Tap the I/O key briefly. The display will then reset to zero after a couple seconds. If the container is removed, its weight will be displayed as a negative number. 3. You may now fill the container with the quantity to be weighed. The display will show the net weight of the contents. 4. To return to normal weight display (gross weight), remove any items from the weighing platform and press the I/O key again to re-zero the scale. Calibration Initial calibration is performed at the factory. However, the calibration can drift slightly due to stresses from shipping. Also, due to changes in local acceleration of gravity around the world, adjustment is recommended when moving the scale from one location to another. To check calibration: 1. Place the required calibration weight on the scale and note the displayed value. 2. If the displayed value is not within acceptable tolerance (see chart on next page), adjustment is required.

CUBE Manual Model Acceptable Calibration Recommended Tolerance (±) Weight Accuracy Class CUBE-100 0.02g 100g OIML M2 CUBE-650 0.2g 500g OIML M2 Adjustment WARNING: If you do not use the correct weight listed in the above chart for your scale, you WILL miscalibrate and lose the factory calibration. Calibration requires an accurate weight. VIII. 1. Press and hold the M button to enter calibration mode. The display will show the required weight. 2. Make sure no weight is on the platform then tap the M key again to set the zero point. The LCD will flash CAL followed by the required calibration weight again (see chart above). 3. Place the required weight on the center of the scale platform. The display will flash PASS then return to normal weighing. NOTE: PASS only indicates that calibration has completed and is not a confirmation that you have used the correct weight. 4. Test the accuracy by placing the same calibration weight back on the platform and checking the value. If the value is within the acceptable tolerance limit above, calibration is complete. If not, try repeating calibration under more stable conditions. Troubleshooting Problem: Scale will not turn on Solution: Check the batteries and battery contacts.

Problem: Displayed weight fluctuates randomly Solution: Make sure your workspace is stable and free of any air currents, vibration, and static electricity. Nearby radio wave emitting devices can cause fluctuations. Problem: Displayed weight is inaccurate Solution: Perform a calibration adjustment. If you are still having trouble, contact our warranty department. IX. Error Codes 1. OUTZ - Zero point has drifted out of the acceptable range. Try to perform a calibration adjustment. If the problem persists, the load cell may be permanently damaged due to overloading. 2. O-Ld - Overload. The maximum capacity has been exceeded. Remove the extra weight to avoid permanently damaging the load cell. 3. Lo - Batteries are low. Replace the batteries. X. Specifications CUBE-100 CUBE-650 Max. Capacity 100g 650g Readability 0.01g 0.1g Min. Weight 0.03g 0.2g Power 2 x 1.5V AAA Alkaline Dimensions 3.1 x 3.1 x 0.7 in. Platform Dimensions 3.1 x 3.1 in. Warranty 10 Year Limited Warranty

CUBE Manual XI. Capacity/Readability Charts CUBE-100 100g x 0.01g 3.527oz x 0.001oz 3.215ozt x 0.001ozt 64.30dwt x 0.01dwt 500ct x 0.05ct 1543.2gn x 0.2gn 2.666tl x 0.001tl CUBE-650 650g x 0.1g 22.92oz x 0.01oz 20.89ozt x 0.01ozt 417.95dwt x 0.1dwt 3250ct x 0.5ct 10031gn x 2gn 13.333tl x 0.001tl American Weigh Scales, Inc. Norcross, GA 30092 Made In China to our exact specifications and quality control.