Contacts. Quick Start Guide

Similar documents
MANITOBA RENAL PROGRAM

Indications for Use: Caution: Note:

Quiz for Module 2: Aquarius Training (Rev. 1) Lesson 1: Introduction to Aquarius

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

AQUARIUS Continuous Renal Replacement Therapy Competency

2008T Hemodialysis Machine with CLiC Device Training Guide

2008T Hemodialysis Machine Operator s Manual

Module 2: Aquarius Training

MANITOBA RENAL PROGRAM

Liberty Handi-Guide. A Patient s Reference to the Most Commonly Asked Questions About the Liberty Cycler

Handi-Guide A Patient s Reference to the Most Commonly Asked Questions About the Liberty Select Cycler

Liberty Cycler Handi-Guide

IV IV Training Guide

Troubleshooting Guide Aquarius

Liberty Handi-Guide A Patient s Reference to the Most Commonly Asked Questions About the Liberty Cycler

Liberty PDx Cycler Handi-Guide

Haemodialysis. Bloodlines, Tubing Systems, Accessories Product Range. Product name / Headline Subline

Bloodlines, Tubing Systems, Accessories

Shared Haemodialysis Care Handbook. Name Hospital number: Shared HD Care Nurse: Date: Machine type: Dialysis Unit:

Liberty Cycler User s Guide

Liberty PDx Cycler User s Guide

Micro-X Dialyzer Reprocessing Concentrate

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

Liberty Select Cycler User s Guide

INTERNATIONAL MARKETING. 61, avenue Tony Garnier Lyon France Tel.: 33 (0) Fax: 33 (0)

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

HAEMOFILTRATION GUIDELINES

Alfred ECMO Circuit Priming and Storage Guide

100 Series Dialyzer Reprocessing System MODEL RS8310 MODEL RS SERIES S/N S 1,000 - UP

CRRT with Prismaflex LEADS TO More Flexibility, Ease of Use and Safety

WARNING: All medications should be closely monitored to detect any alterations in the effective concentration due to the ultrafiltration process.

Blood Pressure Monitoring: Arterial Line

1020 Industrial Drive, Orlinda, TN fax

Introduction. Medley TM Medication Safety System. Alaris Medical Systems

MyoSure Setup Simplified

1020 Industrial Drive, Orlinda, TN fax

FAULT CODE TROUBLESHOOTING INDEX

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Lucid System In-Service Training

Summary of Cardiac Anesthesia Technician Responsibilities

CircuFlow Quick Setup Guide

OPERATOR S MANUAL HOTLINE 3 Blood and Fluid Warmer

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS. Hereditary Angioedema

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

1 Introduction on the appearance, configuration and function

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

OPERATOR S MANUAL. P/N GB Rev. 008

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

BD Cytopeia Fluidic Kit User s Guide

My CoughAssist. A patient guide to CoughAssist T70. Please visit

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES

Using the Akta Prime plus October 22, 2012

TJF-Q180V Cleaning and Disinfection Checklist

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Alaris Pump module FAQs

Hemoglobin Reference Failure Troubleshooting

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

TJF-Q180V Cleaning and Disinfection Checklist

HemoCue Hb Procedure Template

New. The high security connection

OPERATION MANUAL BFX-2

CADD-Micro. Model 5900 Operator s Manual. Note: Colors are Black and Cyan (as second color) Deltec. Ambulatory Infusion Pump

Fresenius Medical Care North America Service Bulletin

Operation Manual. Carbon Filtration Industrial Drive, Orlinda, TN fax

Xcela Hybrid PICC. Clinician Reference Tool

Smart JVA Module WARNING. For use with AquaLink RS systems with rev R firmware or newer.

Varithena Delivery System Instructions for Use

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

A complete family of Pediatric Perfusion Systems

OPERATING MANUAL LO-CROSS-FLOW CF M

Installation of Your SprayMaster System

Blood Parameter Monitoring System 550

SS32L Dissolved O 2 Probe

Comfortmax Thermax Hot/Cold Water Circulation System. Instructions for Use

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

Using the UltraRAE. Firmware 2.35

with O 2 Controller Instruction Manual

Ensuring Patient Safety and Quality- Water Standards. Eileen Lischer MA, BSN, RN,CNN University of California San Diego

C O L O G Y. New. g u O N. l o t a C a. The high security connection

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

IVIG Educational Presentation

Varithena Delivery System Instructions for Use INSTRUCTIONS FOR USE Please read the Product Monograph before using the product.

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

InFlow Modular perfusion bioreactor

RH800 & 2000 BASIC OPERATIONS

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

BiPAP A30. Set-up guide. Connection to oximetry. Getting started. Guidelines for ventilation set-up. Main screen. Menus.

HS653 Negative Pressure Ventilation Controller

USER GUIDE, VOL.2.0 MYOGRAPH SYSTEM - 112PP/114P

Oxygen Dialflow Meter. Instructions for Use

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker.

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

Skills & Scenarios 81

Transcription:

Contacts Clinical Support Specialist: Phone: Cell Phone: Email: Fresenius Renal Technologies A division of Fresenius Medical Care North America 920 Winter Street Waltham, MA 02451 Technical Service Customer Service Phone: 800-227-2572 Phone: 800-323-5188 Quick Start Guide Caution: Federal (US) law restricts this device to sale by or on the order of a physician. Note: Read the Instructions for Use for safe and proper use of this device. For a complete description of hazards, contraindications, side effects and precautions, see full package labeling available at www.fmcna.com. 2010, 2014, 2015 Fresenius Medical Care, All Rights Reserved. Fresenius Medical Care, Fresenius Renal Technologies, the triangle logo, 2008 and DIASAFE are trademarks of Fresenius Medical Care Holdings, Inc and/or its affiliated companies. All other trademarks are the property of their respective owners. P/N 490161 Rev C 02/2015

Notes NOTE Indications for Use: The 2008T hemodialysis machine is indicated for acute and chronic dialysis therapy. This 2008T Hemodialysis Machine Quick Start Guide is intended to be a supplement to the 2008T Operator's Manual (P/N 490122) using single-use dialyzers and the Prime Amount method of priming. Operators must read and understand the Operator s Manual supplied with the equipment. Attention must be given to all Warning and Caution statements in the Operator s Manual. WARNING Failure to operate and maintain the machine according to the manufacturer s instructions may cause operator or patient injury or death. WARNING To maintain sterility of the blood path, do not allow the ends of the bloodlines to come into contact with non-sterile solutions or surfaces which may contaminate the blood path. Setup Concentrate 1. Press the POWER key on the control panel. The Select Program screen will appear on the monitor. 2. Flip down the Touchpad. 3. Plug the acid (red) concentrate connector into acid concentrate source. 4. Select the Dialysis button, the Dialysate screen will appear on the monitor. 5. Confirm concentrate. If the concentrate must be changed to match prescription, select the CONC button, enter the concentrate type using or on keyboard, then confirm. 6. Select the Base Na+ and Bicarbonate buttons and enter prescribed value with the data entry keys and confirm. 7. Select the Home button. The Home screen will appear on the monitor (Air Detector alarm will be displayed in the Status Box). 8. Plug the bicarb (blue) concentrate connector into bicarbonate concentrate source. NOTE To confirm a changed parameter, do any of the following: Tap the Touchpad Press the CONFIRM key Press either of the Enter keys on the keyboard 1 2

Setup Concentrate (continued) WARNING! Acid concentrate products are used as one component in mixing dialysate bath. These acid products contain chemical compounds that, after mixing, yield acetate (and citrate in certain products) in the dialysate. (Please refer to the acid concentrate product labeling for specific acetate/citrate amounts.) After diffusion across the dialyzer membrane, acetate (and citrate when present) is metabolized by the liver to serum bicarbonate and adds to the serum bicarbonate that separately results from the diffusion of dialysate bicarbonate across the dialyzer membrane. During dialysis, the dynamic of diffusion and concentration gradients prevent serum bicarbonate concentration from exceeding the dialysate bicarbonate concentration. The bicarbonate concentration of the dialysate is the bicarbonate setting on the dialysis machine, and is the bicarbonate dose prescribed by the physician. On the 2008 series hemodialysis machines, the bicarbonate dose may be set in a range between 20 and 40 milliequivalents per liter, but may be set in different ranges in other machines. When the dialysis session terminates, acetate (and citrate when present) that has not yet metabolized may remain in the blood and will be converted to serum bicarbonate after diffusion ceases, without possibility of diffusion out of the blood. The post dialysis metabolism of acetate (and citrate when present) could thus briefly increase serum bicarbonate concentration above the prescribed bicarbonate concentration of the dialysate. Physicians should consider this possibility in prescribing bicarbonate dose. Prescription of insufficient bicarbonate may contribute to metabolic acidosis; excessive bicarbonate may contribute to metabolic alkalosis. Both conditions are associated with poor patient outcomes, including increased mortality risk. Setup Bloodline Arterial Line: 1. Close medication port clamp. 2. Snap arterial drip chamber into holder. 3. Connect monitor line to arterial pressure port with a transducer protector and verify the line is unclamped. 4. Insert blood pump segment into blood pump and close door. 5. Snap bloodline into red guides on modules. 6. Aseptically place the patient end of the arterial line into priming bucket clip. Snap the dialyzer end of the arterial line into the dialyzer holder clip. Venous Line: 1. Close medication port clamp. 2. Roll the venous drip chamber into holder with filter below sensor heads; close and latch door. 3. Connect monitor line to pressure port with a transducer protector and verify the line is unclamped. 4. Snap bloodline into blue guides on modules (do not insert into venous clamp yet). 5. Snap dialyzer end of bloodline into dialyzer holder clip. 6. Aseptically place patient end of venous line into priming bucket clip. Dialyzer: 1. Place dialyzer into holder, arterial end up. 3 4

Prime Extracorporeal Circuit 1. Connect dialyzer end of arterial line to the arterial port of the dialyzer. Rotate the dialyzer to the arterial end down position. 2. Connect the dialyzer end of the venous line to the venous port of the dialyzer. 3. Hang a saline bag and attach an administration line, if not already attached, to the saline port on the arterial bloodline. Aseptically spike the saline bag. 4. Gravity prime the patient end of the arterial bloodline below the saline T with saline. When primed, clamp off the patient end of the arterial bloodline. 5. If the heparin pump is to be used: connect heparin syringe, prime heparin line and load heparin syringe into heparin pump. If heparin pump is not used, clamp the heparin line. 6. Press the Prime key. 7. Press the blood pump Start/Stop key. Set blood speed to 150 ml/min. 8. Fill arterial drip chamber to an acceptable level using Level Adjust key. Clamp arterial monitor line and disconnect from machine so pressure port is open to atmosphere. 9. The blood pump will continue to run until the pre-set amount of saline has been flushed through the circuit. When blood pump stops, clamp the patient end of the venous bloodline. 10. Insert the venous line in venous line clamp and optical detector and close the optical detector door. 11. Fill venous drip chamber with Level Adjust key, clamp venous monitor line and disconnect from machine so pressure port is open to atmosphere. 12. Aseptically connect patient ends of arterial and venous bloodlines together and open arterial and venous bloodline clamps. 13. Set the blood pump rate to 350-400 ml/min. Press the blood pump Start/Stop key to start the pump and begin recirculation. If necessary, press RESET to clear any alarms. 14. Ensure that the extracorporeal circuit is free of air bubbles. 5 6

Test 1. To run the tests, verify: Dialysate lines are on the shunt interlock with the interlock door closed. The machine is alarm-free. Arterial and venous pressure monitor lines are clamped and disconnected so the pressure ports are open to atmosphere. UF and SVS are off. 2. Select the Test & Options screen button and press CONFIRM. 3. Select and confirm the Both Tests button to begin running the pressure and alarm tests. 4. When the message Test Complete is displayed in the Status Box, press RESET to clear message. Recirculation 1. Rotate the dialyzer so the arterial end is up. 2. Check the conductivity and ph of the dialysate and test for residual disinfectant. WARNING Always verify the conductivity and approximate ph of the dialysate solution through independent means (e.g., using a conductivity meter or ph paper or meter, as applicable) before initiating each dialysis treatment. Verify that the conductivity is reasonably close to the theoretical value (TCD) and the ph is between 6.9 and 7.6. If they are not, do not initiate dialysis. 3. Attach dialysate connectors to dialyzer (using counter-current flow, red to red and blue to blue) and close shunt door. 4. Connect the arterial and venous pressure monitor lines to their respective pressure ports. Unclamp the lines. 5. When the dialysate compartment is filled, rotate the dialyzer so the arterial inlet is down. 7 8

Recirculation (continued) 6. Press RESET to clear all alarms. Set the blood pump rate to 350-400 ml/min and start the blood pump to begin recirculating saline through the circuit. 7. Test the level detector by pressing on the level adjust down arrow key on the level detector to lower the fluid level in the venous drip chamber. Verify that the blood pump stops and the venous clamp closes. Use the up arrow key to raise the level back to an acceptable level. 8. Check blood tubing to ensure there are no kinks. 9. Replace saline bag with a fresh bag if necessary. 10.Check for a normal dialysate flow by watching the float in the dialysate line. Open the shunt door to verify the machine goes into bypass by observing the float in the dialysate line. Close the shunt door again. 11. Set treatment parameters as prescribed. Starting Dialysis 1. Complete patient assessment per unit policy. 2. Wrap blood pressure cuff around the patient s non-access arm. 3. Verify ultrafiltration (UF) is off (UF light on control panel is off) and that the UF Removed button is reset to zero. 4. Verify the venous line is in the venous clamp and the optical detector. Verify that the optical detector door is closed. WARNING Discard the recirculated saline and fill the bloodlines with fresh, sterile saline. 5. Lower the blood pump rate to 150 ml/min and press the Start/Stop key to stop blood pump. 6. Connect the patient and initiate treatment per unit protocol. 7. Press the Start/Stop key to start blood pump and set QB and QD to prescribed rates. 8. Rotate the dialyzer to arterial end up. 9. Select the Tx Clock button and press CONFIRM to start the treatment, the Status Box will now show Dialysis. 10. Check that UF/SVS/Heparin are on, if prescribed. If applicable, a blood pressure measurement is initiated. 9 10

Completion of Dialysis 1. Press RESET to clear any alarms. 2. Select thetx Clock and press CONFIRM to stop the treatment. The Status Box will now show Dialysis Paused. 3. Press the Start/Stop key on the blood pump to stop the pump. 4. Replace saline bag with a fresh bag, if necessary. 5. Rinse the blood in the patient end of the arterial bloodline back to the patient: Using a hemostat, clamp the arterial bloodline directly above the saline T. Open the saline line clamps and rinse the blood in the tubing below the saline T back to the patient. When the blood in the line has been rinsed back to the patient, close the saline line clamps. 6. Rinse the remaining blood in the bloodline back to the patient: Clamp the arterial bloodline directly under the saline T. Remove the clamp on the bloodline above the saline T and open the saline line clamps. Start the blood pump and set a rate of 150-200 ml/min. When the blood has been returned to the patient, turn the blood pump off and close the saline line clamps. 7. Clamp the arterial and venous bloodlines and the patient s arterial and venous access lines and aseptically disconnect them. 8. Empty the dialyzer: Open the shunt interlock door and return the blue dialysate line to the shunt interlock. Place a dialyzer cap on the open dialyzer port. Rotate the dialyzer so the red connector is down then close the shunt door. The Status Box will display Emptying. When Emptying stopped is displayed, open the shunt door, return the red dialysate line to the shunt interlock and close the shunt door. Place a dialyzer cap on the open dialyzer port. 9. Discard the bloodlines, transducer protectors, and dialyzer according to unit policy. 10. Clean or disinfect the machine exterior according to unit policy. 11. If there is another patient: Press the New Tx key and then CONFIRM. The treatment parameters will be reset. If there is not another patient: Return acid and bicarb connectors to their respective ports securely and proceed with cleaning and disinfecting. 11 12

13 Acid Clean & Heat Disinfect 1. To run the cleansing/disinfecting programs, verify: Dialysate lines are securely on the shunt. Acid and bicarb connectors are securely inserted into machine ports. 2. After the last treatment of the day, select the Acid Clean button on the Select Program screen and press CONFIRM. The lines will be rinsed. 3. When prompted, connect acid (red) and bicarb (blue) connectors to jug containing acid cleaner. 4. Press CONFIRM to start the Acid Clean program. 5. When completed, the message Put Connectors in ports will appear. Put acid and bicarb connectors securely into respective machine ports and press CONFIRM to exit. 6. If the Auto Heat Disinfect service mode option is set: Leave the machine on the Select Program screen. The heat disinfect program will run at the scheduled time. If the Auto Heat Disinfect service mode option is not set: Select the Heat Disinfect button on the Select Program screen and press CONFIRM to start the Heat Disinfect program. The lines will be rinsed with hot water. After the heat disinfect is complete, press CONFIRM to exit. Press and hold the POWER key to turn off the machine. Acid Clean & Chemical Rinse 1. To run the cleansing/disinfecting programs, verify: Dialysate lines are securely on the shunt. Acid and bicarb connectors are securely inserted into machine ports. 2. After the last treatment of the day, select the Acid Clean button on the "Select Program"screen and press CONFIRM. The lines will be rinsed. 3. When prompted, connect acid (red) and bicarb (blue) connectors to a jug containing acid cleaner. 4. Press CONFIRM to start. 5. When the Acid Clean program is completed, the message Put Connectors in ports will appear. Put acid and bicarb connectors securely into respective machine ports and press CONFIRM to exit. 6. Select the Chemical/Rinse button on the Select Program screen and press CONFIRM. The lines will be rinsed. 7. When prompted, connect the red acid connector to a jug containing chemical disinfectant and press CONFIRM. 8. When completed, the message Put Connectors in Ports will appear. Return red acid connector securely to machine port and machine will initiate a mandatory post-rinse. 9. If applicable, the machine will then run an automatic DIASAFE filter test. When the message Diasafe Test Passed is displayed, press CONFIRM to exit. 10. Test for residual disinfectant prior to starting treatment following a chemical disinfection. When the residual disinfectant test results are negative, press and hold the POWER key to turn off the machine. 14