Influence of Ambient Temperature on Performance of a Joule-Thomson Refrigerator

Similar documents
A Numerical Study of the Performance of a Heat Exchanger for a Miniature Joule-Thomson Refrigerator

Cooling Characteristics of GM-type Pulse Tube Refrigerator with Neon as Working Gas

Helium-Liquefaction by Cryocooler for High-Field Magnet Cooling

Vibration-Free Joule-Thomson Cryocoolers for Distributed Microcooling

Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler

Earlier Lecture. In the earlier lecture, we have seen Kapitza & Heylandt systems which are the modifications of the Claude System.

Thermal Testing of an EM Two-Stage Coaxial Pulse Tube Cold Finger for Earth Observation Missions

Reliability Growth of Stirling-Cycle Coolers at L-3 CE

Open Cycle Joule-Thomson Cryocooling by Mixed Coolant

JT Micro Compressor Test Results

Dynamic Operation of a 4 K Pulse Tube Cryocooler with Inverter Compressors

CFD Simulation and Experimental Validation of a Diaphragm Pressure Wave Generator

Experimental Analysis on Vortex Tube Refrigerator Using Different Conical Valve Angles

CFD Flow Analysis of a Refrigerant inside Adiabatic Capillary Tube

Mid InfraRed Instrument (MIRI) Cooler Cold Head Assembly Acceptance Testing and Characterization

Recuperative Cooling Capability of Miniature Thermoacoustic Expanders (MTAEs) around 77 K

Development of a Two-Stage High -Temperature Pulse Tube Cooler for Space Applications

Gas Vapor Injection on Refrigerant Cycle Using Piston Technology

Single Phase Pressure Drop and Flow Distribution in Brazed Plate Heat Exchangers

An Experimental Performance Study of Vortex Tube Refrigeration System

An Investigative and Synoptic Review on Helium Liquefaction Using a Commercial 4 K Gifford- McMahon Cryocooler

AN EXPERIMENTAL INVESTIGATION ON HELIUM LIQUEFACTION USING TWO-STAGE GIFFORD McMAHON CRYOCOOLER

Development of High Efficiency 4K Two-Stage Pulse Tube Cryocooler

Effect of Coiled Capillary Tube Pitch on Vapour Compression Refrigeration System Performance

Experimental Verification of Integrated Pressure Suppression Systems in Fusion Reactors at In-Vessel Loss-of -Coolant Events

Development of a Passive Helium Heat Switch for Fast Cool Down by Two-Stage Cryocooler

Cryogenics The Basics. Lesson 2 D. Kashy

Oil-Lubricated Compressors for Regenerative Cryocoolers Using an Elastic Membrane

Figure 1 Schematic of opposing air bearing concept

Influencing Factors Study of the Variable Speed Scroll Compressor with EVI Technology

First Experimental Data of the Cryogenic Safety Test Facility PICARD

Effect of Hot End Obstruction and Nozzle on the Performance of Vortex Tube

HYDROGEN FAST FILLING TO A TYPE IV TANK DEVELOPED FOR MOTORCYCLES

Continuously Variable Inertance Tubes for Pulse Tube Refrigerators

Predicting the Suction Gas Superheating in Reciprocating Compressors

Two-stage pulse tube refrigerator with step displacer as phase shifter

Modeling and Optimization of a Mixed-Gas Joule-Thomson Cryoprobe System

Purdue e-pubs. Purdue University. Stephan Lehr Technische Universitaet Dresden. Follow this and additional works at:

CONSIDERATION OF DENSITY VARIATIONS IN THE DESIGN OF A VENTILATION SYSTEM FOR ROAD TUNNELS

Visual Observation of Nucleate Boiling and Sliding Phenomena of Boiling Bubbles on a Horizontal Tube Heater

A Numerical Simulation of Fluid-Structure Interaction for Refrigerator Compressors Suction and Exhaust System Performance Analysis

Experimental Investigation on Stirling type Thermally Coupled Three Stage Pulse tube Cryocoolers with U type Configuration

Transient Analyses In Relief Systems

Lab # 03: Visualization of Shock Waves by using Schlieren Technique

EXPERIMENTAL STUDY OF HOT INERT GAS JET IGNITION OF HYDROGEN-OXYGEN MIXTURE

Loss of Vacuum Experiments on a Superfluid Helium Vessel

An Investigation of Liquid Injection in Refrigeration Screw Compressors

TESTING AND ANALYZING ON P-V DIAGRAM OF CO 2 ROLLING PISTON EXPANDER

Experimental determination of deflagration explosion characteristics of methane air mixture and their verification by advanced numerical simulation

Cryogenic Engineering Prof. M. D. Atrey Department of Mechanical Engineering Indian Institute of Technology, Bombay

Schedule of Requirements THERMODYNAMICS LABORATORY- CHEMICAL ENGINEERING DEPARTMENT

Operation of Compressors Through Use of Load Stands

S.A. Klein and G.F. Nellis Cambridge University Press, 2011

Bradley Moore 1 University of Wisconsin - Madison

A. M. Dalavi, Mahesh Jadhav, Yasin Shaikh, Avinash Patil (Department of Mechanical Engineering, Symbiosis Institute of Technology, India)

Comparative Analysis of Two Types of Positive Displacement Compressors for Air Conditioning Applications

Flow and Heat Transfer Characteristics Over a NACA0018 Aerofoil and a Test Aerofoil A Comparative Study

Performance Testing of a High Capacity Compressor for a 20K 20W Cryocooler

Comparison of MARS-KS to SPACE for counter current flow limitation model

Start-up of the Three-shaft Closed Loop

ANALYSIS OF HEAT TRANSFER THROUGH EXTERNAL FINS USING CFD TOOL

3-Stage ADR for the SXS Instrument on the Astro-H Mission

NUMERICAL ANALYSIS OF SINGLE STAGE PULSE TUBE REFRIGERATOR

Flow in a shock tube

OPTIMIZATION OF RECUPERATER FIN GEOMETRY FOR MICRO GAS TURBINE

A NEW PROCESS FOR IMPROVED LIQUEFACTION EFFICIENCY

ME 200 Thermodynamics I Spring 2010 (Last) (First) Thermo Number: CIRCLE YOUR LECTURE BELOW

Heat Transfer Research on a Special Cryogenic Heat Exchanger-a Neutron Moderator Cell (NMC)

EXPERIMENTAL ANALYSIS OF FIN TUBE AND BARE TUBE TYPE TUBE-IN-TUBE HEAT EXCHANGER

IMECE IMECE2011-6

SIMULATION OF ENVIRONMENTAL FLIGHT CONDITIONS

Development of Prediction Models for Spent Nuclear Fuel Storage Cask Monitoring Hyong Chol Kim, Sam Hee Han, Dong Jae Park

United States Patent (19) Mikus et al.

Available online at ScienceDirect. Physics Procedia 67 (2015 )

CYCLE ANALYSIS OF LINEAR COMPRESSORS USING THREE- DIMENSIONAL CFD

Thermodynamics ERT 206 Properties of Pure Substance HANNA ILYANI ZULHAIMI

COMPUTATIONAL FLUID DYNAMICS ANALYSIS OF HELICAL NOZZLES EFFECTS ON THE ENERGY SEPARATION IN A VORTEX TUBE

SUPERSONIC GAS TECHNOLOGIES

New low temperature technologies of natural gas processing

FABRICATION AND OPTIMIZATION OF PERFORMANCE OF VORTEX TUBE PARAMETERS

Investigation to Improve Efficiency of Transcritical R744 Two-Stage Vapor Compression Systems

AIRFLOW GENERATION IN A TUNNEL USING A SACCARDO VENTILATION SYSTEM AGAINST THE BUOYANCY EFFECT PRODUCED BY A FIRE

EXPERIMENTAL PERFORMANCE ANALYSIS OF VORTEX TUBE FOR VARIOUS PARAMETERS

SALT RSS-NIR MID-TERM REVIEW MAY 20 & 21, 2009 THERMAL SYSTEM DESIGN

GAS MIXTURES. Department of Mechanical Engineering

Chapter 4, Problem 30.

STUDY OF CRYOGENIC SYSTEMS WITH ASPEN - HYSYS ANALYSYS

Investigation of Suction Process of Scroll Compressors

Improvement of the Volumetric and Isentropic Efficiency Due to Modifications of the Suction Chamber

CERTIFICATES OF COMPETENCY IN THE MERCHANT NAVY MARINE ENGINEER OFFICER

Thermodynamic Modeling and Mechanical Design of a Liquid Nitrogen Vaporization and Pressure Building Device

Effect of airflow direction on human perception of draught

Judson Technologies PB 219. JTC JOULE-THOMSON CRYOSTAT COOLING SYSTEM Operating Instructions. October, 2003

ENGG. THERMODYNAMICS

Influence of rounding corners on unsteady flow and heat transfer around a square cylinder

Basics of Low-temperature Refrigeration

University of Cincinnati

Development of Two Opposing Nozzle Vortex Tube Using Air as Working Fluid

E. MENDOZA, C. KEMPEN, Y. ESTERKIN, S. SUN, K. SUSKO and J. GOGLIA

Leak Checking Large Vacuum Chambers

Transcription:

Influence of Ambient Temperature on Performance of a Joule-Thomson Refrigerator Y. J. Hong, S.J. Park, J. Ko, H.B. Kim Korea Institute of Machinery & Materials Daejeon, 305-343, Korea ABSTRACT Miniature Joule-Thomson refrigerators consist of a counter flow heat exchanger with double helical tubes and fins, nozzles, a mandrel and a dewar. Typical performance factors for the Joule- Thomson refrigerator are cool-down time, cold end temperature, running time and gas consumption. The above performance factors depend on operating conditions such as the pressure of the gas, thermal environment and etc. In this study, effects of the ambient temperature on the thermal performance of the Joule- Thomson refrigerator were investigated by numerical analyses. Effectiveness-NTU approach was adopted to predict the thermodynamic behaviors of the heat exchanger of the Joule-Thomson refrigerator. A one dimensional isentropic model was used to calculate the mass flow rate through the nozzle. In the results, the Joule-Thomson refrigerator in the low ambient temperature has a large cooling capacity and mass flow rate for a regulated supply pressure of the gas. But, the low ambient temperature results in the decrease of the initial pressure of the gas in the vessel. The Joule-Thomson refrigerator, which is installed in the low ambient temperature condition, has a negative aspect in the pressure of the gas, but a positive aspect in the cooling effect. Hence, it has the cooling capacity for achieving fast cool-down. INTRODUCTION Miniature Joule-Thomson (J-T) refrigerators have been widely used for rapid cooling of infrared detectors, probes of cryosurgery, thermal cameras, missile homing head and guidance system, due to their special features of simple configuration, compact structure and rapid cooldown characteristics. Nitrogen (NBP 77.35 K) and argon gas (NBP 87.3 K) are typically used as cryogen due to their easy availability, low cost and ability to achieve a relatively low cryogenic temperature. The J-T refrigerator consists of a counter flow heat exchanger with a double helical tube and fins, a nozzle, a mandrel and dewar as shown in Figure 1. The cooling power of the J-T refrigerator is generated by the isenthalpic expansion of the high pressure gas through the throttling process. The cooling effect can be amplified by using an expanded gas to cool incoming gas within a recuperative heat exchanger. Cool-down time, temperature at the cold end, Cryocoolers 17, edited by S.D. Miller and R.G. Ross, Jr. International Cryocooler Conference, Inc., Boulder, CO, 2012 407

408 J-T AND SORPTION CRYOCOOLER DEVELOPMENTS Figure 1. Schematic diagram of the J-T refrigerator running time and gas consumption are the important indicators of the performance of the miniature J-T refrigerator. The above performance factors depend on operating conditions such as the pressure of gas, thermal environment and etc. To cool from an initial temperature to a cryogenic temperature, the net refrigeration produced by the J-T refrigerator must meet or exceed the heat to be removed from the cold end components over the temperature range. In order to insure fast cool-down, the design of the Dewar should minimize the thermal mass of the J-T refrigerator system 1. The thermal mass depends on the initial temperature of the system. In the operation of the J-T refrigerator in the high ambient temperature, the J-T refrigerator should overcome a large thermal mass and heat load. In general, the J-T refrigerator has a pressurized vessel in military applications. The change of the ambient temperature results in the initial pressure of the gas in the vessel, and effects of the gas pressure have significant effects on the performance of the J-T refrigerator. Hong, et al. 2 reported the experimental study of the J-T refrigerator with the gas pressure up to 12 MPa. Experimental results show that the J-T refrigerator has a rapid cool-down rate at the cold ambient condition. Chua, et. al. 3 developed the geometry model of the Hampson-type cryocooler with the argon gas and the steady state governing equations were solved numerically. Ng et al. 4 and Xue et al. 5 conducted the experimental and numerical study of the J-T refrigerator for steady-state characteristics with the argon gas. Hong, et. al. 6 adopted the effectiveness-ntu approach to predict the thermodynamic behaviors of the heat exchanger of the J-T refrigerator with different pressures and mass flow rate of the nitrogen and argon gas. The flow rate through the nozzle is relevant to the operation condition of the J-T refrigerator. Maytal 7 proposed an isentropic model to calculate the mass flux of the gas through a nozzle. In this study, effects of the ambient temperature on the thermal performance of the Joule- Thomson refrigerator were investigated by numerical analysis. ANALYSIS MODEL OF THE J-T REFRIGERATOR The J-T refrigerator has a convectional double helical tube and fin configuration. The heat exchanger has two passages for high pressure gas and a common passage for low pressure gas. The outer diameter and the length of the heat exchanger are 9.5 mm and 24 mm, respectively. Effectiveness-NTU approach 6 was adopted to predict the thermodynamic behaviors of the heat exchanger of the J-T refrigerator. A one dimensional isentropic model 7 was used to calculate the mass flux of the gas through the nozzle. Analyses were carried out for the argon gas with the gas pressure from 10 MPa to 60 MPa and the ambient temperature from -40 o C to +60 o C.

INFLUENCE OF AMBIENT TEMP ON J-T REFRIGERATOR 409 Figure 2. Comparison of the mass flow rate In order to validate the analysis model, the mass flow rate of the argon gas from analysis was compared with that of the experiment 2. The calculated mass flow rate follows the trend of experimental results as shown in Figure 2. RESULTS AND DISCUSSION In general military applications, the gas comes from a pressurized vessel. The change of the ambient temperature results in the initial pressure of the gas in the vessel. So the J-T refrigerator, which is installed in cold or hot ambient conditions, had variations in the cooling performance. The cooling capacity of the J-T refrigerator is given by the product of the mass flow rate of the gas and the cooling effects (enthalpy difference between supply gas and exhaust gas of the refrigerator), and therefore increases of mass flow rate of the gas give rise to increases of the cooling capacity. In an ideal J-T refrigerator with the argon gas, the cooling effect increases as the supply pressure of the gas increases up to the pressure of 55 MPa. But, the larger mass flow rate of the gas causes the degradation of the effectiveness of the heat exchanger. Figure 3 shows effects of the supply pressure of the gas on the performance of the J-T refrigerator. Analyses were carried out for the argon gas with a supply pressure of 10 MPa 60 Figure 3. Effects of the pressure of the gas: (a) mass flow rate and (b) cooling effects

410 J-T AND SORPTION CRYOCOOLER DEVELOPMENTS MPa and an ambient temperature of 20 o C. Up to the supply pressure of 40 MPa, the J-T refrigerator with the gas high pressure has a high mass flow rate and a high cooling effects. Above a pressure of 40 MPa, the gas mass flow rate increases as the gas pressure increases, but the cooling effect does not increases. At the elevated pressure of the gas, the large mass flow rate comes from the pressure of the gas, and the increase of the cooling capacity mainly comes from the increase of the mass flow rate. In the present study, to investigate effects of the ambient temperature on the performance of the J-T refrigerator, analysis was carried out previously for the constant supply pressure of the gas and the ambient temperature from -40 o C to +60 o C Performance of the Refrigerator with a Regulated Pressure of the Gas Figure 4 shows effects of the ambient temperature on the J-T refrigerator performance with a regulated supply pressure of argon gas. At a given gas supply pressure, as the ambient temperature increases, the gas mass flow rate and the cooling effect decrease, so the cooling capacity decreases rapidly. At a high ambient temperature, the thermal mass of the Dewar and the J-T refrigerator would be increased. These phenomena will result in a slow cool-down characteristics of the J-T refrigerator at high ambient temperature. Figure 4. Effects of the ambient temperature for the regulated pressure of the gas: (a) mass flow rate, (b) cooling effects, and (c) cooling capacity

INFLUENCE OF AMBIENT TEMP ON J-T REFRIGERATOR 411 In the experiment 2, the cool-down period of the J-T refrigerator at the ambient temperature of +40 o C is 5-6 times longer than that at the ambient temperature of -40 o C. By estimation, the thermal mass of the Dewar and the J-T refrigerator at the ambient temperature of +40 o C is raised by a factor of 1.6 times. So, during cool-down period, the cooling capacity of the J-T refrigerator at the ambient temperature of +40 o C might be estimated 3-4 times lower than that at the ambient temperature of -40 o C. In present analysis, at the ambient temperature of -40 o C, the J-T refrigerator with the gas pressure of 10 MPa has the cooling capacity of 11.7 W as shown in Figure 4. Otherwise the J-T refrigerator has the cooling capacity of 2.9 W at the ambient temperature of +40 o C. These results show agreement with the experimental results. Figure 5 shows the pressure-enthalpy diagram of the J-T refrigerator for different ambient temperatures. It is obvious that the J-T refrigerator in the cold ambient condition has a large cooling effect due to the outcome of the isenthalpic expansion from the low gas temperature at the inlet of the J-T nozzle. The increase of the mass flow rate mainly comes from the low temperature of the gas at the inlet of the J-T nozzle. Performance of the Refrigerator with a Pressurized Vessel In the cool-down period, the change of the ambient temperature results in a change in the gas supply pressure in the vessel. Figure 6 illustrates the calculated variation of the argon gas pressure with a change in the temperature. The gas pressure in a pressurized vessel increases as the ambient temperature increases. There is a two-fold increase in the pressure over the charging pressure range of 10-50 MPa. Figure 7 shows the effect of the ambient temperature on the performance of the J-T refrigerator in a pressurized vessel. In the figure, pressure represents the gas charging pressure. By contrast to the case of the regulated pressure of the gas, the mass flow rate for the given pressure increases as the ambient temperature increases. These increases in the mass flow rate result from the pressure rise as shown in Figure 6. In the cold ambient condition, the J-T refrigerator has no great change in the aspect of the cooling effect by comparison with the case of the regulated pressure of the gas. But, at the pressure of 60 MPa and hot ambient condition, there is the rapid degradation in the cooling effect due to the decline in the performance of the heat exchanger. Figure 5. P-H diagram of the J-T refrigerator.

412 J-T AND SORPTION CRYOCOOLER DEVELOPMENTS Figure 6. Variation of the pressure with the change of the ambient temperature Figure 7. Performance of the J-T refrigerator with a pressurized vessel: (a) mass flow rate, (b) cooling effects, and (c) cooling capacity. The J-T refrigerator in the low ambient temperature has the large cooling capacity. But, the low ambient temperature results in the decrease of the initial pressure of the gas in the vessel. It

INFLUENCE OF AMBIENT TEMP ON J-T REFRIGERATOR 413 could be concluded that the J-T refrigerator, which is installed in the low ambient temperature condition, has a negative aspect in the pressure of the gas, but a positive aspect in cooling effect. So the J-T refrigerator has the big cooling capacity for achieving fast cool-down. SUMMARY In the present study, the effectiveness-ntu approach and the isentropic model were adopted to predict the thermodynamic behaviors of the J-T refrigerator, and performances of the J-T refrigerator in different ambient temperatures were investigated. For a regulated supply pressure of the gas, the J-T refrigerator for a given supply pressure in the cold ambient condition has the large mass flow rate and the big cooling effect. These phenomena result in the high cooling capacity. On the other hand, for a gas supplying from a pressurized vessel, the J-T refrigerator in the cold ambient condition has the small mass flow rate due to the change of the gas pressure in the vessel. But the J-T refrigerator has the high cooling capacity on account of the big cooling effect. So the J-T refrigerator has a rapid cool-down characteristics in the cold ambient condition. ACKNOWLEDGMENT This work was supported by Korea Institute of Machinery & Materials and Ministry of Knowledge Economy. REFERENCES 1. Bonney, G. E., Stubbs, D. M., Design fundamentals of rapid Cool-down Joule-Thomson (JT) cryostats and sensors, SPIE Cryogenic Optical Systems and Instruments, Vol. 2227 (June 1994), pp. 98-108. 2. Hong, Y. J., Kim, H. B., Park, S. J., Effects of environmental temperature on performance of the Joule-Thomson refrigerator, Adv. in Cryogenic Engineering, Vol. 57, Amer. Institute of Physics, Melville, NY (2012), pp. 1805-1812 3. Chua, H. T., Wang, X., Teo, H. Y., A Numerical study of the Hampson-type miniature Joule- Thomson cryocooler, Int. J. of Heat and Mass Transfer, Vol. 49 (2006), pp. 582-593. 4. Ng, K. C., Xue, H., Wang, J. B., Experimental and numerical study on a miniature Joule-Thomson cooler for steady-state characteristics,, Int. J. of Heat and Mass Transfer, Vol. 45 (2002), pp. 609-618. 5. Xue, H., Ng, K. C., Wang, J. B., Performance evaluation of the recuperative heat exchanger in a miniature Joule-Thomson cooler,, Applied Thermal Engineering, Vol. 21 (2001), pp. 1829-1844. 6. Hong, Y. J., Park, S. J., Choi, Y.D., A Numerical Study on the performance of the Miniature Joule Thomson Refrigerator, Cryocoolers 15, ICC Press, Boulder, CO (2009), pp. 379-386. 7. Maytal, B. Z., Real gas choked flow conditions at low reduced-temperatures, Cryogenics, Vol. 46, Issue: 1 (January 2006), pp. 21-29.