CHAPTER 3 RESULTS AND DISCUSSION

Similar documents
Measurement of water vapour transport through a porous non-hygroscopic material in a temperature gradient

Generating Calibration Gas Standards

Air permeance of paper and paperboard (Sheffield method)

EASIDEW TRANSMITTER with Current Source Output

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

Application Note AN-107

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

CORESTA RECOMMENDED METHOD N 6

Measurement of water vapour transport through a porous non-hygroscopic material in a temperature gradient

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

Instruction Manual. Pipe Friction Training Panel

KATEDRA MATERIÁLOVÉHO INŽENÝRSTVÍ A CHEMIE. 123MEAN water vapor transport

The Discussion of this exercise covers the following points:

Accurate Measurement of Steam Flow Properties

Advanced Test Equipment Rentals ATEC (2832)

Gas Physics Pressure and Flow Topics Covered:

Monitoring Gas Pressure in Vacuum Insulation Panels

Pressure Measurement

Heat Engine. Reading: Appropriate sections for first, second law of thermodynamics, and PV diagrams.

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

LEAP CO 2 Laboratory CO 2 mixtures test facility

Automatic Permeability Testing: The Challenges and Solutions Author: Alyce Hartvigsen, PBI-Dansensor A/S, Ringsted, Denmark

Soil Water Characteristic Curve (SWCC) Using the WP4C

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

Chapter 13 Gases, Vapors, Liquids, and Solids

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

ASTM D86 - Procedural Outline

Gerald D. Anderson. Education Technical Specialist

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

M-06 Nitrogen Generator (Nitrogen Making Machine)

Model PDT Dewpoint Transmitter

An Automated High Precision Saturation/Dilution Calibration System for Trace Moisture Sensors

Hardware Triaxial and Consolidation Testing Systems Pressure Measurement and Control

FAS-W. Features. Applications. Automatic condensation hygrometer

3 1 PRESSURE. This is illustrated in Fig. 3 3.

Introduction. Objectives. Hazards. Procedure

Vibration-Free Joule-Thomson Cryocoolers for Distributed Microcooling

WP2 Fire test for toxicity of fire effluents

PMI Pulse Decay Permeameter for Shale Rock Characterization Yang Yu, Scientist Porous Materials Inc., 20 Dutch Mill Road, Ithaca NY 14850

LFE OEM TCD - Thermal Conductivity Detector

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Cover Page for Lab Report Group Portion. Drag on Spheres

What I have learned about SF 6 gas testing.a Practical explanation

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

Standard Test Methods for Water Vapor Content of Electrical Insulating Gases by Measurement of Dew Point 1

Columbus Instruments

Pneumatic dead-weight tester Model CPB3500

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

FTC130 Transmitter. Operating Manual

Effect of Coiled Capillary Tube Pitch on Vapour Compression Refrigeration System Performance

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

InstrumentationTools.com

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

1 SE/P-02. Experimental and Analytical Studies on Thermal-Hydraulic Performance of a Vacuum Vessel Pressure Suppression System in ITER

Truck-mounted Mass Flow Metering for LPG Delivery

CHARACTERISTICS OF LEAK DETECTION BASED ON DIFERENTIAL PRESSURE MEASUREMENT

AMT-Ex Dewpoint Transmitter

An experimental investigation of the influence of the expansion of the moist air on its relative humidity

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

LIQUID METER PROVING TECHNIQUES CT 4095

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

T EK-COR 1100A. Coriolis Mass Flowmeter. FLOW. Technology Solutions

Experiment Instructions. Circulating Pumps Training Panel

Chemistry 20 Unit 2 Gases FITB Notes. Topic A Characteristics of Gases

Experimental Characterization and Modeling of Helium Dispersion in a ¼-Scale Two-Car Residential Garage

STANDARD PROCEDURE OF A TIGHTNESS TEST OF A SOLAR COLLECTOR BOX. (EUROSUN 2000 SOLAR CONGRESS)

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

Technical Data Sheet MF010-O-LC

Traceable calibration of automatic weighing instruments in dynamic operation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

LAB 13: FLUIDS OBJECTIVES

MS.RAJA ELGADY/PRESSURE PAPER 3

METHOD 2E - DETERMINATION OF LANDFILL GAS PRODUCTION FLOW RATE. NOTE: This method does not include all of the

Mitos Fluika Pressure and Vacuum Pumps Datasheet

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Vapor Pressure of Liquids

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

Diffusivity in Gases Experiment ver 5.10b

Improve Process Reliability

Model DPC3500 Continuous Low Range Dew Point Analyzer. Operations Manual

Determination of Air Density with Buoyancy Artefacts

Instrumentation & Data Acquisition Systems

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

Relative Humidity Calibration Kit

METHOD 2D - MEASUREMENT OF GAS VOLUME FLOW RATES IN SMALL PIPES AND DUCTS. NOTE: This method does not include all of the

Topics For Discussion Acadiana Gas Measurement School

Vapor Sorption Analyzer

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

Flow in a shock tube

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Modeling Diffusion Rates of a Gas in an Enclosed Space

Transcription:

CHAPTER 3 RESULTS AND DISCUSSION The research included the development and verification of the earlier apparatus, and the collection and validation of moisture diffusion data under both isothermal and nonisothermal conditions. The main objective of the apparatus is to measure the moisture (water vapor) flux rate through a given hygroscopic material under various conditions. The test material for which data was obtained during the experimental procedure of this research is oriented strand board (OSB), which is a wood composite material. The test specimen consists of a disk of 0.4318 m (17 in.) in diameter and 6.4 mm (0.25 in.) in thickness. The performance of the controls systems of the apparatus as well as the data obtained are presented in this chapter. The first section covers the operation of both the instrumentation and the control systems. The results obtained both during isothermal and nonisothermal tests are presented in the second section. The data validation, results and discussion of additional experiments are presented in the last section of this chapter. 3.1 APPARATUS OPERATION The validation of the apparatus results relies on the accuracy of the temperature and the relative humidity maintained within the chambers. The validity of the data collected depends upon the accuracy in the control of the temperature and relative humidity. On the other hand, the control stability depends on the accuracy of the instruments and the reliability of the controllers. 27

The high end range of the relative humidity within the apparatus is limited by the lowest temperature encountered within the apparatus, since this temperature will dictate the dew point of the system. At temperatures of 0 o C to 5 o C, the maximum relative humidity allowed is approximately 68 per cent as a direct result of the low dew-point temperatures. At operating temperatures around room temperature the maximum allowable relative humidity is approximately 75 per cent. The temperature control system of the apparatus is capable of maintaining a temperature difference of 10 o C across the specimen with a maximum variation of ±0.05 o C. 3.1.1 Instrumentation As indicated in the previous chapter, it is extremely important to maintain close control of the temperature and relative humidity within the chambers. This section focuses on the accuracy of the instruments used to control the environmental conditions of the apparatus. The control system uses an instrument manufactured by Vaisala which measures temperature and relative humidity. The Vaisala transmitter was calibrated directly at the factory at the time of purchase. This calibration indicated an uncertainty of ±0.6 per cent R.H. and ±0.2 o C. The calibration and instrument verification as indicated by the manufacturer used dry nitrogen and a set of controlled aqueous salt solutions. The temperature calibration of the Vaisala instrument was performed at ambient temperature (25 C) against a Fluke 8520A/PRT. A separate calibration check was performed with the instrument assembled inside the chamber. A dew-point hygrometer (General Eastern) and a Fluke temperature transducer verified the relative humidity and temperature, respectively. The dew-point hygrometer used for the calibration of the instrument uses optical hygrometry principles. Optical hygrometry is a precise technique, which determines the water vapor content in gases by measuring dew- or frost-point temperatures. Optical condensation hygrometry is based on the chilled-mirror principle. The surface of the chilled-mirror is cooled until it reaches a temperature at which condensation begins to form on it, at this point an 28

optical sensor detects the condensation and its temperature is measured. The measured temperature corresponds to the dew point. The dew-point temperature within each chamber was calculated taking into account the actual pressure in the chamber, and compared to the read out corresponding to the hygrometers. The difference between the measured and calculated dew point temperatures (0.5 o C) were within the accuracy of both instruments, ±0.6 per cent R.H. for the Vaisala and ±0.2 o C for the hygrometer. The temperature indicated by the Vaisala instrument was compared to a Fluke temperature transducer using a type-t thermocouple. Once again the difference between the measured temperatures (±0.1 o C) were within the uncertainty of both instruments. Calibration was performed on each Vaisala instrument at different levels of relative humidity while running an isothermal test at 25 o C. 3.1.2 Temperature Control As discussed in the previous chapter, the environmental controls are a critical and essential part of the proper operation of the apparatus. The control stability allows the system to reach and sustain steady state conditions faster. The temperature control system was designed to maintain strong nonisothermal conditions across the sample material. In general, the nonisothermal tests were conducted under two different temperature differences, 5 and 10 degrees Celsius, corresponding to a temperature gradient of 7.87 and 15.74 C/cm, respectively. Figure 3.1 shows the temperature of each chamber over a period of 12 days. The accuracy and performance of the temperature control system can be observed in the enlarged plots depicted in Fig. 3.2. Even with 10 C temperature differences, the control maintained the temperature inside each chamber within ±0.1 o C of the set point. The plots seen in Fig. 3.1-2 were generated by data obtained by the computer every 10 minutes. 29

Temperature ( o C) 37 35 33 31 29 27 25 23 0 2 4 6 8 10 12 Time (days) Bottom Chamber Top Chamber Figure 3.1 Temperature control typical plot. 35.2 35.15 35.1 35.05 Temperature (oc) 35 34.95 34.9 34.85 34.8 0 2 4 6 8 10 12 Time (hours) Figure 3.2a Temperature variation, typical plot for bottom chamber. 30

25.2 25.15 Temperature (oc) 25.1 25.05 25 24.95 24.9 24.85 24.8 0 2 4 6 8 10 12 Time (hours) Figure 3.2b Temperature variation, typical plot for top chamber. 3.1.3 Relative Humidity Control Control of the relative humidity is essential to the adequate performance of the apparatus and the validity of results. As explained earlier, the relative humidity range of the apparatus depends on the temperature of the surroundings. For this reason, prior to every experimental run the dew-point temperature is calculated to ensure that no condensation occurs within the system. During the experimental procedure, the apparatus was run under conditions ranging from 22.8 to 52 per cent relative humidity. Figure 3.3 corresponds to the relative humidity conditions for Test 10. Figure 3.4 is a large-scale version of data shown in Fig. 3.3. It is clear from this plot that the maximum variation of the relative humidity control system is within ±0.1 of a percentage point from the set point. 31

Relative Humidity (%) 46 44 42 40 38 36 34 32 30 28 26 24 22 20 Top Chamber φ = 45% Bottom Chamber φ = 22.8% 0 0.5 1 1.5 2 2.5 3 3.5 4 Time (days) Figure 3.3 Relative humidity plot for Test No.10. 45.2 45.15 Relative Humidity (%) 45.1 45.05 45 44.95 44.9 44.85 44.8 0 2 4 6 8 10 12 Time (hours) Figure 3.4 Relative humidity control for Test No. 10. 32

3.1.4 Pressure Control As discussed in Chapter 2, the pressure control of the apparatus is essential to the moisture diffusion across the specimen. The pressure control system prevents the introduction of a pressure difference as an additional mechanism to the moisture diffusion. The pressure across the test material was controlled and maintained approximately equal by using rubber bladders as shown in Fig. 2.1. The pressure difference between the chambers was measured by a Datametrics micro-manometer, as described in Chapter 2. The pressure difference between the chambers and the atmosphere was measured with a U-tube mercury manometer. Although rubber bladders were used to equalize the pressure between the chambers, a very small pressure difference was observed. Several attempts were made to cancel the small difference in pressure, but it was later concluded that the difference is inherent for the FDC and cup apparatuses. The pressure difference is further analyzed in the next section. The control accuracy of the ASHRAE FDC allows the moisture transfer across the specimen to be measured, as a result only of the temperature and humidity gradients. The results obtained from the tests performed are discussed in the following section. 3.2 RESULTS Two different sets of experiments were conducted in order to validate the operation of the experimental apparatus. The first set of experiments (Moisture Permeance) measured the moisture transfer through the specimen under both isothermal and nonisothermal conditions. The second experimental set (Air Permeance) allows for the calculation of the air permeance of the specimen. This last experimental set was conducted as a result of findings encountered during the first experimental set. The purpose of moisture permeance measurements is to validate the diffusion model. Conditions were set according to the various existing models. As discussed 33

previously, a necessary condition for the validation of the model is that moisture transfer must cease when the diffusion potential is the same on both sides of the specimen. The models based on the gradient of chemical potential, moisture concentrations, activated moisture molecules, and vapor pressure were tested. A standard ASTM cup method test is included in this experimental set. As stated in Chapter 1, the cup test is an experimental method used to determine the moisture permeance for a given material under isothermal conditions. 3.2.1 Moisture Permeance The purpose of this experimental set includes investigating various moisture diffusion models, as well as validating the apparatus. This experimental set includes a series of isothermal and nonisothermal measurements conducted with the ASHRAE FDC, and a set of experiments based upon the ASTM standard cup test. Isothermal Tests Two isothermal tests were conducted in order to obtain information regarding the specimen and verify the operation of the control systems. The data obtained during these tests were used to calculate the permeance of the specimen. The data collected and permeance results as shown in Tables 3.1 and 3.4, respectively, are used in later sections for comparison and validation purposes. The calculation of the moisture transfer follows the procedure outlined in Sections 2.5 and 2.6. Table 3.1 Isothermal experimental results and conditions. Case n" x10 8 (kg/m 2 s) Uncertainty x10 8 (kg/m 2 s) φ t (%) T t (C) φ b (%) T b (C) Isothermal No. 1 3.505 0.221 58 24 43 24 Isothermal No. 2 3.136 0.196 23 25 33 25 34

Nonisothermal Test Current diffusion models were tested under nonisothermal conditions. A total of five diffusion models based upon the gradient of chemical potential, moisture concentration, water vapor concentration, vapor pressure and activated moisture molecules were tested. As explained previously, a necessary condition for the validation of the potential diffusion model is that the moisture diffusion must cease when the transfer potential or driving force is equal on both sides of the specimen. In order for the gradient of the potential to be zero, conditions were calculated such that, the potential given by each candidate model would be the same on both sides of the specimen. The calculation procedure for the values corresponding to the potential for each one of the models in each chamber is shown in Appendix A; the results are shown in Table 3.2. Table 3.2 summarizes the results obtained from the diffusion potential tests. This table shows the measured moisture flux rate across the specimen, the vapor pressure and the conditions corresponding to each chamber for the respective model. Figure 3.5 shows the results obtained from moisture diffusion tests. The models were tested while maintaining the conditions of the top chamber constant at 25 C and 45 per cent relative humidity. Among the models tested, the vapor pressure (P v ) and the water-vapor concentration (P v /T) models best satisfied the null flow condition. Although the watervapor concentration and the vapor pressure model yield similar results, the vapor pressure model best satisfied the null-flow condition needed for the validation of the diffusion model. Figure 3.6 shows the results obtained from moisture diffusion tests as a function of the vapor pressure for the bottom chamber. This figure depicts the moisture flux for each model while maintaining the corresponding potential equal to zero. A straight line was traced along the results obtained from each model. The constant slope of this line shows the permeance (g) to be constant under the range tested, where the permeance is given by, P γ= v n" & (3.1) 35

8.0E-08 Equal Chemical Potential 6.0E-08 Moisture Flux (kg/m 2 -s) 4.0E-08 2.0E-08 0.0E+00 Equal Water Vapor Concentration 1 Equal Activated Molecules Equal Water Vapor Concentration 2 Test No. 9 Test No. 12 Test No. 11 Equal Vapor Pressure -2.0E-08 Test No. 10-4.0E-08 Equal Moisture Concentration Figure 3.5 Evaluation of moisture diffusion theories. Although some uncertainties in the calculation of the conditions for the activated moisture model were considered, the results obtained for this case were significantly poorer for the null condition compared to the vapor pressure model. The sorption isotherm (Burch,1993) and the activated energy formulation for the OSB (Siau) are among these uncertainties. Additional ASHRAE FDC Tests Most of the experimental cases were conducted near room temperature with a temperature difference of either 5 o C or 10 o C across the specimen. However, in order to verify the results previously obtained from the vapor pressure model, a low temperature test (Test No. 12) was conducted. The results of this test are shown in Fig. 3.6. Each symbol (with estimated uncertainty bands) in Figures 3.5 and 3.6 represents the average of a series of runs (4-10) of each experimental case. 36

Table 3.2 Nonisothermal experimental results and conditions. Case Equal Chemical Potential Equal Moisture Concentration (Pv/T) t =(Pv/T) b Tb-Tt = 5C Equal Vapor Pressure Tb-Tt = 10C (Pv/T) t =(Pv/T) b Tb-Tt = 10C Equal Activated Moisture n" x10 8 (kg/m 2 -s) Uncertainty x10 8 (kg/m 2 -s) φ t (%) T t (C) φ b (%) T b (C) P vb (kpa) P vt (kpa) 6.741 0.714 45 25 52 30 2.226 1.425-3.036-0.303 45 25 45 20 1.052 1.425 0.623 0.195 45 25 34.1 30 1.443 1.425 0.554 0.190 45 25 25.3 35 1.425 1.425 0.790 0.315 45 25 26.1 35 1.467 1.425 2.968 0.391 45 25 31.9 30 1.793 1.425 Test No. 9-0.771 0.200 45 25 22.8 35 1.283 1.425 Test No. 10-1.951 0.445 22.8 35 45 25 1.283 1.282 Test No. 11-0.476 0.421 25.3 35 45 25 1.425 1.423 Test No. 12 0.330 0.0524 45 5 32 10 0.392 0.392 Note : Positive moisture flux denotes transfer from the bottom chamber to the top. Measured Moisture Flux Rate (kg/m 2 -sec) 8.0E-08 6.0E-08 4.0E-08 2.0E-08 0.0E+00-2.0E-08-4.0E-08 Chambers are designated B and T Note: positive flux denotes transfer from B to T Top chamber conditions are 25C and 45%R.H. unless otherwise noted. Ch B;45%R.H.,20C Ch B; 26.1%R.H.,35C Ch B; 34.1%R.H.,30C Ch B; 25.3%R.H.,35C Ch B;22.8%R.H.,35C Ch B;31.9%R.H.,30C Ch T; 25.3%R.H.,35C Ch B; 45%R.H.,25C Ch T; 22.8%R.H,35C Ch B; 45%R.H,25C Uncertainty bands Ch B; 52%R.H.,30C 1 1.2 1.4 1.6 1.8 2 2.2 2.4 Chamber B Vapor Pressure (kpa) Equal Chemical Potential Equal Moisture Concentration (Pv/T)t=(Pv/T)b Tb-Tt = 5 C Equal Vapor Pressure Tb-Tt = 10 C (Pv/T)t=(Pv/T)b Tb-Tt = 10 C Activated moisture content Test No. 9 Test No. 10 Test No.11 Figure 3.6 Moisture diffusion theories vs. vapor pressure difference. 37

8.0E-08 Measured Moisture Flux Rate (kg/m 2 -sec) 6.0E-08 4.0E-08 2.0E-08 0.0E+00-2.0E-08 Chambers are designated B and T Note: positive flux denotes transfer from B to T Ch T; 45%R.H.,5C Ch B; 32%R.H.,10C Test No. 12-4.0E-08 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 Chamber B Vapor Pressure (kpa) Figure 3.7 Evaluation of the vapor pressure diffusion model at low temperatures. A series of additional tests were conducted to determine the moisture diffusion behavior near the equal vapor-pressure point. Tests denoted as Test No. 9 and Test No. 10, were conducted to confirm the reversal of the moisture flow near the equal-vapor pressure condition. Test No. 11 was conducted to observe the effect of buoyancy (gravity) in moisture diffusion. The flow reversal test was performed by selecting the conditions of the lower chamber for which a change in the flow direction would be expected. As depicted in Fig. 3.5, the flow direction was changed when the vapor pressure gradient was reversed. Moiser (1996) reported that the effect of gravity in the moisture diffusion mechanism is negligible. The results of Test No. 11 suggest a small buoyancy effect. This test was performed by interchanging the conditions of the chambers and observing the direction of the moisture flow in contrasts to the original chamber state. For a given set of conditions for the top and bottom chamber, the moisture flow direction is considered to be positive if moisture is being transfer from the bottom chamber to the top. The results for the vapor pressure model test showed a very small positive flow of 38

moisture. When the chamber conditions were interchanged in Test No. 11, the flow was of the same order of magnitude but in the opposite direction. This last observation supports the statement by Moiser. The complete set of data attained for each of the experimental models is displayed in Appendix B. ASTM Standard Cup Test The cup experiments were based upon the ASTM test method 96-80 (ASTM 1989), and the research conducted by Burch and Thomas in 1992. This experimental method was used to determine the permeance of the OSB specimen by measuring the water-vapor transmission through the specimen. The complete experimental procedure of these tests has been discussed in Chapter 1. A total of three-cup measurements were made for the test material. Measurements made of the moisture transfer (WVT) in each cup allowed for the calculation of the permeability (µ, or permeance γ) as a function of the average relative humidity across the specimen. The measurements were conducted at a constant temperature of approximately 25 o C. The results obtained are shown in Table 3.3. Table 3.3 Cup test experimental results. Cup I Cup II Cup IV Mean R.H. 16.98 38 79.92 R.H. Difference 11.37 10.31 9.18 WVT x10 10 (kg/s) 3.839 3.142 5.486 γx10 11 (kg/m-sec-pa) 9.19 8.28 16.60 The relative humidity difference denotes the difference across the specimen generated by aqueous salt solutions, as shown in Table 1.1. The complete experimental data and results are shown in Appendix A. 39

3.2.2 Validation of Data The validation of the data is accomplished by comparing the results obtained from ASHRAE FDC and previously published results for the same material. The permeance of the test material was determined by using the moisture transfer data obtained during both isothermal and nonisothermal tests. As discussed in Chapter 1, the permeability (permeance) data obtained from isothermal tests are applicable to nonisothermal tests as well, (Galbraith and McLean, 1998). The reduced permeance data can be then validated by comparing it to published data of the same type. In this case the permeance information from the tests were compared to data obtained by the National Institute of Standards and Technology (NIST) (Burch and Thomas, 1992). Table 3.4 shows the data in its reduced form. Figure 3.7 shows the reduced data for the various tests (isothermal, nonisothermal and ASTM cup tests) as indicated in Table 3.4 and the data reported by NIST, for OSB at 24 C. Permeance data is not available either at lower or higher relative humidity as a result of the apparatus limitation by condensation. However, the ASHRAE FDC data agrees well with the data obtained from the conventional cup test and the published data. The dotted line in Fig. 3.7, indicates the linearity behavior of the permeance in the experimental range. Table 3-4 Permeance results from ASHRAE FDC and cup tests. Case Mean R.H. γ x10 11 (kg/m 2 -s-pa) Isothermal Test No. 1 50.0 7.84 Isothermal Test No. 2 28.0 9.47 Test No. 1 38.4 8.06 Test No. 2 45.0 8.14 Test No. 3 48.5 8.42 Cup 1 16.9 9.19 Cup 2 38.0 8.28 Cup 4 79.9 16.60 40

The permeance was determined by using a modified form of Fick s law given by Eq. 1.2. n γ= & P v (3.1) where n & is the flux of moisture and P v is the difference in vapor pressure. As the moisture flux approaches zero, the corresponding permeance is undefined. For this reason the permeance corresponding to the vapor pressure and the water-vapor concentration models test cannot be determined. Table 3.4 summarizes the results of the permeance data obtained from both the ASHRAE FDC tests and the ASTM standard cup tests. The temperature and relative humidity conditions corresponding to isothermal tests are shown in Table 3.1. The conditions of the tests labeled in Table 3.4, as Test No.1, 2, and 3, correspond to the conditions of those shown in Table 3.2 for the activated moisture, moisture concentration and chemical potential models, respectively. Permeance (kg/m 2 -s-pa) 3.5E-10 3.0E-10 2.5E-10 2.0E-10 1.5E-10 1.0E-10 Bursh/Thomas Data @24C New Data w / cups @ 25C Isothermal 1 Isothermal 2 Top: 25C,45%R.H. Bot: 30C,31.9%R.H. Top: 25C,45%R.H. Bot: 20C,45%R.H. Top: 25C,45%R.H. Bot: 30C,52%R.H. 5.0E-11 0.0E+00 0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0 Average Relative Humidity (%) Figure 3.8 Comparison of permeance data from the ASHRAE and cup apparatuses. 41

Unexpected Findings During the ASHRAE FDC experimental procedure, a small difference in the static pressure between the chambers was found. This pressure difference was first detected by a Magnehelic pressure differential gauge with a resolution of ±0.02 iwg. The pressure difference was attributed originally to leakage in the system. However, the difference was still observed after extensive leak-control measures were implemented. The static pressure was observed to be always higher in the chamber where moisture was being transferred. An experiment was designed to determine if the difference in static pressure across the specimen is attributed to the diffusion of moisture. The test utilized the experimental setup of the ASTM standard cup test and a digital micro-manometer transducer. Figure 3.8 shows the schematic of the experimental setup. A small adapter was designed to fit the ASTM cup. This adapter enables a digital Datametrics micromanometer as described in Chapter 2, to read the static pressure difference across the specimen between the cup and vessel. The specimen was perforated by a small stainless steel tube, which was connected to the instrument at the top of the vessel by a small diameter plastic tube. The instrument at the top of the vessel was connected to the digital micro-manometer, where the pressure difference was measured. The experimental results showed a static pressure difference of 0.0032 iwg across the specimen and opposing the moisture flow direction. The pressure difference and the moisture transfer were periodically measured for a period of several weeks. The pressure difference across the specimen is believed to be caused by bulk airflow generated by the process of air diffusion. The bulk flow of air governed by Darcy s equation balances the diffusion of air in the opposite direction as a result of the gradient in the partial pressure of (dry) air. 42

Additional attempts to use the digital micro-manometer to measure this difference failed in other cups and over extended times. The reason is that the micro manometer cannot tolerate the corrosive effects of the salts in the sealed environment. Cup's Pressure Vessel's Pressure Cup Digital Micro Manometer Figure 3.9 Static pressure difference experimental setup. 3.2.3 Air Permeance An experiment to measure the bulk flow as a result of the difference in static pressure found across the sample was designed. The flow of air is given by Darcy s Equation as n& a = µ a A s P s L (3.2) where µ a is the air permeability, A s is the cross-sectional area of the material, P s is the static pressure difference, and L is the thickness of the material. 43

The air permeance of the material was determined by measuring the pressure drops across the OSB specimen over a range of air flow rates. A schematic of the experimental setup is shown in Fig. 3.9. The airflow to the chamber was supplied by a dry-air generator and regulated by a needle valve in the flowmeter. The pressure was measured at various air flow rates, hence, permitting the calculation of the air permeability by manipulating Eq. 3.2 µ a n = & P s L A s = L ma s (3.3) where, m denotes the slope of the plot in Fig. 3.10. Figure 3.10 shows the data and the calculated air permeability (µ a ). A small discrepancy was found between the measurements made by the two flowmeters, as a result of their accuracy. However, the slopes corresponding to both flowmeters differed only one per cent. The air permeability of the material was determined by using the average of the slopes for each of the flowmeters. The air permeability for OSB was calculated to be 1.322x10-8 kg/pa-s-m. Flowmeter Air Flow Needle Valve 1 2 3 4 5 6 7 Digital Micro manometer Figure 3.10 Air permeance experimental setup. 44

40 35 30 y = 3268992.445x + 1.856 R 2 = 0.993 Pressure drop (Pa) 25 20 15 10 5 y = 3310784.631x - 0.109 R 2 = 0.986 Flow meter No. 1 Flow meter No. 2 0 0.0E+00 1.0E-06 2.0E-06 3.0E-06 4.0E-06 5.0E-06 6.0E-06 7.0E-06 8.0E-06 9.0E-06 1.0E-05 Air flow rate (kg/s) Figure 3.11 Air permeability data and results. (1 lb/h = 1.26E-04 kg/s), (1 iwg = 249 Pa) The results show that bulk airflow of approximately 1.890E-6 kg/s (0.015 lb/h) occurs through the specimen of OSB for the pressure difference of 0.7968 Pa (0.0032 iwg) measured in the ASTM cup test. Further investigation of this coupled effect on bulk flow by diffusion of water vapor and (dry) air is beyond the scope of the present research. 3.3 SUMMARY The objectives of the research included the validation of the existing apparatus operation and the study of the various moisture diffusion models. These two objectives have been accomplished. The permeance results obtained from the ASHRAE FDC and the ASTM cup tests agreed well with the data reported by the NIST. The satisfactory operation of the ASHRAE FDC is validated by the agreement between the measured and published results. The theory that moisture diffusion under nonisothermal conditions is governed by the gradient of the water-vapor pressure is strongly supported by the results obtained 45

from the ASHRAE FDC method. The results obtained from the tests of the vapor pressure model conducted near room temperature, and low temperature, were of good agreement with the null-flow condition. None of the other proposed models approached the null-flow necessary condition nearly as well. Further investigation of the significance of difference in static pressure across the OSB sample while experiencing moisture diffusion is recommended. 46