Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

Similar documents
6890 GC Site Preparation

7820A GC Site Preparation Checklist

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

490 Micro GC Site Preparation Checklist

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

490 Micro GC Site Preparation Checklist

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Customer Responsibilities

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

28 Site Preparation. Supplying valve actuator air

Agilent GC, GC/MS, and ALS

DD2 Site Preparation Checklist Hardware Site Preparation Specification

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Customer Responsibilities

3 Flow and Pressure Control

Customer Responsibilities

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Agilent 7697A Headspace Sampler

Customer Responsibilities

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1

Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions

Agilent G3180B Two-Way Splitter Kit

Micro GC Fusion Pre-Installation Checlist

Discovery TGA 55, TGA 550, TGA Site Preparation Guide

Agilent G Rebuilt Electron Capture Detector

Discovery HP-TGA 75/750. Site Preparation Guide

Pre-installation Checklist

UltraLo 1800 Alpha Particle Counter

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Advances in Capillary Column Backflushing

Agilent 7667A. Mini Thermal Desorber. Site Preparation

Best Practice for Identifying Leaks in GC and GC/MS Systems

no peaks missing peaks too small peaks increasing retention decreasing retention

Your local gas generation partner. Precision series Modular gas generation solution for GC.

TA Instruments Installation Requirements for the Thermal Discovery Series

Agilent 7890 Series Gas Chromatograph

Agilent 5975T LTM GC/MSD

Mass Spec will not Autotune

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Multiple Gas#5 GC configuration Jan 2016

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Super-Clean Gas Filters from Restek

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

CP-3800 GC Getting Started Manual

6890 GC Not Ready Messages (A16010)

Agilent 7697A Headspace Sampler

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Oxidation Stability of Gasoline and Aviation Fuels

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions

Agilent Inlet EPC Leak Test

Agilent Intuvo 9000 Gas Chromatograph

6850 Series Control Module User Information

5890II GC Standard Operating Procedure 9/2/2005

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

STANDARD OPERATING PROCEDURES

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Buck Scientific, Inc.

Tank Blanketing Pressure Regulators RHPS Series

HumiSys HF High Flow RH Generator

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Avio 500 ICP-OES. ICP-Optical Emission Spectroscopy PREPARING YOUR LAB. Preparation Checklist. Suitable Working Area

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

A New GC/MS System Designed for Helium Carrier Gas Conservation

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Description Advanced Features Typical Applications

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

System Overview. TCD Detector temperature setpoint Regulator

NGP-250/500 Nitrogen Generator Quick Start Guide

CP-3800 GC Operator s Manual

Zebron Gas Management

Instant Connect Helium Saver Injector Module

Retention Time Locking: Concepts and Applications. Application

Capital Controls Series NXT3000 Modular design gas feed system with self-contained automatic switchover capability.

FPG8601 Force Balanced Piston Gauge

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Mass Flow Controllers

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Preinstallation Checklist and Instructions for the TriStar These Preinstallation Checklist and Instructions were reviewed and approved by:

Carrier Gases in Capillary GC

Technical Information

CAPITAL CONTROLS SERIES NXT3000

Transcription:

Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key first step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide AND checklist prepared for you that outlines the supplies, consumables, space and utility requirements for your equipment for your site For additional information about our solutions, please visit our web site at http://www.chem.agilent.com/en-us/pages/homepage.aspx Customer Responsibilities Make sure your site meets the following prior to the installation date using the checklist below. For details, see specific sections within this document, including: the necessary laboratory or bench space is available. the environmental conditions for the lab as well as laboratory gases, tubing, the power requirements related to the product (e.g. number & location of electrical outlets) the required operating supplies necessary for the product and installation please consult Other/Special Requirements section below for other product-specific information If Agilent is delivering installation and familiarization services, users of the instrument should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information. Important Customer Information 1. If you have questions or problems in providing anything described as a Customer Responsibilities above, please contact your local Agilent or partner support/service organization for assistance prior to delivery. In addition, Agilent and/or its partners reserve the right to reschedule the installation dependent upon the readiness of your laboratory. 2. Should your site not be ready for whatever reasons, please contact Agilent as soon as possible to rearrange any services that have been purchased. 3. Other optional services such as additional training, operational qualification (OQ) and consultation for user-specific applications may also be provided at the time of installation when ordered with the system, but should be contracted separately. Issued: 12-May-2011 Revision: 1.0 Copyright 1 2011 Agilent Technologies Page 1 of 9

Dimensions and Weight Identify the laboratory bench space before your system arrives based on the table below. Pay special attention to the total height and total weight requirements for all system components you have ordered and avoid bench space with overhanging shelves. Also pay special attention to the total weight of the modules you have ordered to ensure your laboratory bench can support this weight. Instrument Description Weight Height Depth Width Kg lbs cm in cm in cm in G3440A Agilent 7890A 50 110 50 to 58 20 to 23 54 21 59 23 G2913A 7683 Auto-injector 3.1 6.8 42 above 17 above 12 5 12 5 G2614A 7683 Tray 3.0 6.6 20 8 34 13.4 30 Left of 12 left of G4513A 7693 Auto-injector G4514A 7693 Tray 50 above 20 above 2 in front of 1 in front of 45 left of 220 MS 23 51 49 19 64 25 25 10 18 left of Environmental Conditions Operating your instrument within the recommended temperature ranges insures optimum instrument performance and lifetime. Special Notes: 1. Performance can be affected by sources of heat & cold e.g. direct sunlight, heating/cooling from air conditioning outlets, drafts and/or vibrations. 2. The site s ambient temperature conditions must be stable for optimum performance. 3. The foreline pump exhausts most compounds introduced into the MS along with oil vapor from the pump. Check that the ventilation system is suitable for the foreline pump. Consult local regulations 4. An Oven Exhaust Deflector kit is available for attaching 10-cm (4 in) exhaust duct to exhaust the hot air. This adds about 13 cm to the back of the - Order option 306 or part number G1530-80650. For s with the exhaust deflector option installed, the exhaust is about 65 CFM (ft3/min /1.840 m3/min). Without the deflector, the exhaust rate is about 99 CFM (ft3/min /1.840 m3/min). Issued: 12-May-2011 Revision: 1.0 Copyright 2 2011 Agilent Technologies Page 2 of 9

Instrument Description Operating temp range C ( F) G3440A Agilent 7890A 15 to 35 C (59 to 95 F) 220 MS 16 to 30 C (61 to 86 F) Operating humidity range Heat Dissipation (%) (BTU/HR) 5 to 80 Std Oven Ramp 7,681 Fast Oven Ramp 10,071 20 to 80 Steady state 2,100 Startup 15,000 Power Consumption Special Notes: 1. If a computer system is supplied with your instrument, be sure to account for those electrical outlets. Instrument Description with Standard Oven with Standard Oven Line Voltage & Frequency (V, Hz) Americas: 120 single phase, 48-63 220/230/240 single/split phase, 48-63 Maximum Power Consumption (VA) 2250 2250 with Fast Oven Japan 200 split phase, 48-63 with Fast Oven 220/230/240 single/split phase, 48-63 220 MS 100/120/200/240 50-60 2950 2950 1500 Supplies Customer to Provide Special Notes: 1. For information on Agilent consumables, accessories and laboratory operating supplies, please visit http://www.chem.agilent.com/en-us/products/consumables/pages/default.aspx 2. Gas Supplies: Gases are supplied by tanks, internal distribution system, or gas generators. Tank supplies require two staged, pressure regulation. To connect tubing to the supply, it must have one 1/8-inch Swagelok female connector for each gas source. Make sure that your regulator has the appropriate sized adapter to end with a 1/8-inch Swagelok female connector. Agilent recommends a carrier and detector gas purity of 99.9995% or better. Issued: 12-May-2011 Revision: 1.0 Copyright 3 2011 Agilent Technologies Page 3 of 9

Tank Regulators All Agilent regulators are supplied with the 1/8-inch Swagelok female connector. Gas Type CGA Number Max Pressure Part Number Air 346 125 PSIG (8.6 Bar) 5183-4641 Hydrogen, Argon/Methane 350 125 PSIG (8.6 Bar) 5183-4642 Oxygen 540 125 PSIG (8.6 Bar) 5183-4643, Argon, 580 125 PSIG (8.6 Bar) 5183-4644 Air 590 125 PSIG (8.6 Bar) 5183-4645 Recommended Gas Supply Pressures The following tables list minimum and maximum pressures in psi for each electronic pneumatic control module (EPC). These requirements are for the input to the EPC module located at the back of the gas chromatograph. Conversions: 1 psi = 6.8947 kpa = 0.068947 Bar = 0.068 ATM. Detectors Gas FID NPD TCD ECD FPD Hydrogen 35-100 35-100 45-100 Air 55-100 55-100 100-120 Make up 55-100 55-100 55-100 55-100 55-100 Reference 55-100 Auxiliary EPC and Pneumatic Control channels The minimum supply pressure for AUX and PCM modules is 20 psi greater than pressure used in your method. For example, if you need a pressure of 20 psi for the method, the supply pressure must be at least 40 psi. AUX EPC PCM 1 PCM 2 or PCM AUX Maximum pressure 120 120 120 with Forward pressure control 50 with Back pressure control Inlets The minimum supply pressure for inlet modules is 20 psi greater than pressure used in your method. For example, if you need a pressure of 40 psi for the method, the supply pressure must be at least 60 psi. SSL 150 SSL 100 PCOC PPIP PTV MM Carrier max 170 120 120 120 120 120 Miscellaneous Gas Plumbing Information Cryogenic cooling with Liquid N2 requires 1/4-inch insulated copper tubing 25-30 PSI supply. Cryogenic cooling with Liquid CO2 requires 1/8-inch heavy-walled, stainless steel tubing 750-1000 PSI supply tank with dip tube. Internal Valco rotary Valve actuation requires a separate pressurized, dry air at 55 psi. If you have not requested option 305 (pre-plumbed ), you must supply pre-cleaned, 1/8-inch copper tubing and a variety of 1/8- gas supplies. Issued: 12-May-2011 Revision: 1.0 Copyright 4 2011 Agilent Technologies Page 4 of 9

WARNING: Never use liquid thread sealer to connect fittings. Never use chlorinated solvents to clean tubing or fittings. Gas purity and selection Agilent recommends a carrier and detector gas purity of 99.9995% or better. Air for flame detectors should be zero grade. Agilent also recommends using traps to remove hydrocarbons, water, and oxygen. When used with capillary columns, detectors require a separate makeup gas for optimum sensitivity. This table lists gas recommendations for capillary columns and the preferred makeup gas types. Detector Carrier gas Make up 1st choice Mass selective Hydrogen None Electron capture Hydrogen Argon/methane Argon/methane Argon/methane Argon/methane Flame ionization Flame photometric phosphorous Thermal conductivity Hydrogen Hydrogen Argon Hydrogen Must be same as carrier and reference Make up 2nd choice None Argon/methane none Must be same as carrier and reference Purge or reference Anode purge must be same as makeup Hydrogen and air for detector Hydrogen and air for detector Hydrogen and air for detector Reference must be same as carrier and makeup The inlet electronic pressure control (EPC) modules are calibrated for up to 4 carrier gases: Split/Splitless capillary (SS), Purged packed (PP), Programmable temperature vaporization (PTV), Multi-Mode (MM), and cool on-column (COC) are calibrated for, Hydrogen,, and Argon methane 5%. Volatiles inlet VI is calibrated for only and Hydrogen. Issued: 12-May-2011 Revision: 1.0 Copyright 5 2011 Agilent Technologies Page 5 of 9

220 MS CI Reagent purity and selection Liquid Methanol is required reagent for performance verification in CI mode. Reagent requirements Purity Comments Methanol 99.9% Reagent grade. Purge and trap grade recommended. Evaporation residue <.0001%. 220 MS CI Reagent purity and selection Gas (optional) Reagent requirements Typical pressure Typical flow(ml/min) Methane 21 to 34 kpa 1 to 2 (3 to 5 psi) Isobutane 21 to 34 kpa 1 to 2 (3 to 5 psi) Ammonia 21 to 34 kpa (3 to 5 psi) 1 to 2 Other Considerations Your Agilent 7890A comes with an analytical column: 19091J-413 (HP5, 30 meter, 0.32mm x 0.25µm). Our checkout standards are designed to work with this column. In many cases, you will need to select a different column for your application. Refer to http://www.chem.agilent.com/cag/cabu/gccolchoose.htm http://www.chem.agilent.com/cag/cabu/gccolchoose.htm for information on column selection. Refer to http://www.chem.agilent.com/cag/cabu/gcreflib.htm http://www.chem.agilent.com/cag/cabu/gcreflib.htm for topics including: guard columns, retention gaps, conditioning, and method development. Your comes with a few basic tools and consumables depending on the specific inlet and detector that you ordered. Issued: 12-May-2011 Revision: 1.0 Copyright 6 2011 Agilent Technologies Page 6 of 9

Tools and consumables that are provided with the instrument. Items Inlet wrench T10 and T20 Torx wrenches ¼-inch nut driver FID flow measuring insert Ceramic wafer column cutter 1/8-inch Tee, Swagelok, brass 1/8-inch nuts & ferrules, Swagelok, brass Inlet septa appropriate for type Inlet insert or liner Inlet wrench Std Checkout Column CP8944 Transferline alignment tool Nut driver Aluminum oxide, 600 grit PN 392027600 Applicator, cotton tipped, pkg. 100 PN 8899999000 O-rings for Transferline and Analyzer Assembly Purpose Replacing inlet septa and liners. Remove tray. Remove covers to access EPC modules, traps, and possible leaks. FID jet replacement. FID troubleshooting. Column installation. Connect gas supplies Connect gas supplies Injection port seal Injection port Replacing inlet septa and liners. Column 30 meter x.25mm x.25um FT Aligning Transferline with endcaps Trap oven removal Cleaning Trap Cleaning Trap Seals for vacuum components Useful tools and consumables that are not provided with the instrument. Item ECD/TCD Detector plug, 5060-9055 1/8-inch Ball Valve, 0100-2144 Digital flow meter Electronic gas leak detector Column cutters 1/8-inch tubing cutter (wire cutter type) Assorted wrenches: ¼, 3/8, 7/16, 9/16 Electronic vial crimper Purpose Inlet pressure decay test. Inlet pressure decay test. One per inlet. Verifying flows, checking for leaks and plugs. Pin pointing gas leaks. Safety checks when using Hydrogen. Cutting columns Cut gas supply tubing Gas supply and plumbing fittings. Ensures consistently air-tight vial closure no matter who does the crimping. This table lists consumables that you may wish to order. First time /MS users should consider adding the following supplies to maintain their system and prevent interruptions in the use of their system. Please refer to the Agilent Consumables and Supplies Catalog for part numbers and recommended maintenance periods. Useful tools and consumables that are not provided with the instrument. Consumable Part Number Gas Filter Replacement CP17973 Gate conductor 393055201 Gate 393055101 Wavy washer 1492000900 Assembly, multiplier 393031501 Assembly, filament disk with wires 393060191 DS-42 oil mist eliminator 393847701 Premium Foreline pump oil 9499390M001 O-ring, turbo pump to manifold 2820043800 Quad-ring, viton manifold 393010914 Quad-ring, viton transfer line 393010918 Issued: 12-May-2011 Revision: 1.0 Copyright 7 2011 Agilent Technologies Page 7 of 9

/MS Calibration compound, FC-43 392035300 Non-Agilent computer or software Agilent hardware, recommended computers, and software are thoroughly tested for compatibility and reliable operation. Your company may have selected other suppliers for parts of your system. This list summarizes some of the key requirements for computers and software. The following Agilent software products are thoroughly tested and compatible: MSWS Version 7.0 Mass Hunter Software o Qualification B.04.00 o Quanification B.04.00 SP3 G1035C Wiley 9th/NIST 08 Mass Spectral Database These software packages require the following: Windows 7 Professional (32-bit only) Windows XP Professional SP2 Windows XP Professional SP3 For MassHunter software requirements see the Mass Hunter Site Prep The following minimum computer requirements are recommended: Intel Core 2 Duo Processor (or higher) Video screen supporting 1280 x 1024 x 256 resolution or greater. 32-bit color is recommended. 2 GB RAM (or more) 100 GB Disk (or larger) CD/DVD-ROM drive Windows 7 or Windows XP Professional SP2/SP3 Issued: 12-May-2011 Revision: 1.0 Copyright 8 2011 Agilent Technologies Page 8 of 9

Document Control Logs Revision Log: Revision Date Reason For Update A.01.00 11-Apr-2011 First Release Approval Log: Revision Approver(s) Title of Approver A.01.01 Don Gage Product Support manager Issued: 12-May-2011 Revision: 1.0 Copyright 9 2011 Agilent Technologies Page 9 of 9