SAFETY PRINCIPLES FOR PRESSURE REGULATORS FOR MEDICAL OXYGEN CYLINDERS

Similar documents
CATALOGUE OF EIGA MEDICAL GASES PUBLICATIONS

USE OF RESIDUAL PRESSURE VALVES

USE OF NON-METALLIC MATERIALS IN HIGH PRESSURE OXYGEN BREATHING GAS APPLICATIONS

SAFETY TRAINING LEAFLET 06 CARBON DIOXIDE

REFERENCE GUIDE ON MEDICINAL GAS CYLINDER VALVES

300 BAR HIGH STRENGTH SEAMLESS STEEL GAS CYLINDERS

SAFETY TRAINING LEAFLET 02 BASIC RULES FOR SAFETY AND GOOD HOUSEKEEPING

SAFETY FEATURES OF PORTABLE CRYOGENIC LIQUID CONTAINERS FOR INDUSTRIAL AND MEDICAL GASES

METHODS TO PREVENT THE PREMATURE ACTIVATION OF RELIEF DEVICES ON TRANSPORT TANKS

DESCRIPTION OF THE PRESSURE TEST METHODS USED DURING CYLINDER MANUFACTURE

LABELLING OF MEDICAL GAS CONTAINERS

CYLINDER RETEST STATIONS

DEVELOPMENT OF CALCULATION FORMULAE FOR CYLINDER WALL THICKNESS

FLEXIBLE CONNECTIONS IN HIGH PRESSURE GAS SYSTEMS

A REFERENCE GUIDE FOR REQUALIFICATION OF GAS CYLINDERS

DESIGN CONSIDERATIONS TO MITIGATE THE POTENTIAL RISKS OF TOXICITY WHEN USING NON- METALLIC MATERIALS IN HIGH PRESSURE OXYGEN BREATHING GAS SYSTEMS

MEDICAL GAS CYLINDERS COLOUR CODING TD 09/15/E

300 BAR HIGH STRENGTH SEAMLESS STEEL GAS CYLINDERS

Pressure Regulator. Instructions for Use

CYLINDER VALVE GUIDELINES TD 12/15/E

HAZARDS ASSOCIATED WITH THE USE OF ACTIVATED CARBON CRYOGENIC GAS PURIFIERS

ISO 9539 INTERNATIONAL STANDARD. Gas welding equipment Materials for equipment used in gas welding, cutting and allied processes

Oxygen Dialflow Meter. Instructions for Use

ISO INTERNATIONAL STANDARD. Flow-metering devices for connection to terminal units of medical gas pipeline systems

DESIGN CONSIDERATIONS AND GUIDANCE FOR THE SAFE USE OF MEDICAL GAS VIPR

Pressure Regulators. Operating Instructions. Instrumentation

MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE CYLINDERS

GUIDELINES FOR HANDLING EXTERNALLY CONTAMINATED MEDICAL CYLINDERS IN A PANDEMIC SITUATION

Pressure regulators for use with medical gases. Part 2: Manifold and line pressure regulators

for the of and the EM55-1

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

Dialflow Regulator. Instructions for Use

GOOD MANUFACTURING PRACTICE GUIDE PART I FOR MEDICAL GASES

SAFETY BULLETIN 01/05

Oxygen Usage Best Practice Guide

Oxygen Dialflow Meter. Instructions for Use

RECERTIFICATION GUIDELINES FOR HIGH PRESSURE CYLINDERS TD 30/18/E

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

300 BAR HIGH STRENGTH SEAMLESS STEEL GAS CYLINDERS

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

This document is a preview generated by EVS

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS

REDLINE pressure regulators for GENIE cylinders.

ISO INTERNATIONAL STANDARD. Gas cylinders Cylinder valves with integrated pressure regulators Specification and type testing

Micro Dial-Flowmeter. Instructions for Use

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

SERIES-O REGULATOR. Instructions for use

DUPLEX PRESSURE REDUCING SET

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

User manual. ZX trolley.

Tri-Safem Model I1 Regulator. INSTRUCTIONS and PARTS

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

ISO INTERNATIONAL STANDARD. Gas cylinders Safe handling. Bouteilles à gaz Sécurité de manutention. Second edition

ISO 2503 INTERNATIONAL STANDARD

GUIDANCE FOR MANAGING THE RISKS ASSOCIATED WITH END OF LIFE LPG VEHICLES

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

Operating Manual for the Cylinder and Line Pressure Regulators of the M51/M52/M53 series and E51/E52/E53 series

SAFE DESIGN AND OPERATION OF ON SITE GENERATION OF OXYGEN 93% FOR MEDICAL USE

Gases Under Pressure: Why we need to manage the risks Philip Brickell, General Secretary European Industrial Gases Association

MAINTENANCE & ASSEMBLY INSTRUCTIONS FUSIBLE LINK. Inline Thermal Check Valve

SAFETY POLICY AND PROCEDURE MANUAL All Euramax Subsidiaries. Number K-3.0 Welding Safety Procedures

IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions

Safe Handling of Pressurized Gases and Cylinders. Dr. Michael Hanisch

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

Cal/OSHA T8 CCR 1536 Cal/OSHA T8 CCR 4799 Cal/OSHA T8 CCR 4845 Cal/OSHA T8 CCR 4848

SOUTH AFRICAN NATIONAL STANDARD

Safety, Set-Up, and Operating Instructions

COSASCO 6000 PSI SINGLE ISOLATION SERVICE VALVE (FR) MAINTENANCE

for Series FE51/52/53-SP

623 Series Dome Load Regulators

ISO Anaesthetic and respiratory equipment Conical connectors Part 2: Screw-threaded weight-bearing connectors

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD

A guide to Part 10 of the

92831 TEL: (714) FAX:

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

BCGA GUIDANCE NOTE 17

MAINTENANCE AND OPERATING MANUAL

MANUAL GAS MANIFOLDS

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

1 DRIVE INDUSTRIAL IMPACT WRENCH

Fisher and Whisper Disk Diffusers

Pilot HON 625. Entwurf. Product information. serving the gas industry worldwide

MPC1M, MPC2M and MPC2AM Monnier High Efficiency Compressed Air Filter / Regulators

Operating and maintenance manual Filter and reducing station Series / 1.0

BOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare

Operating Manual. R280 Pressure regulator made of brass. 1. Intended Use

Pressure Regulators. Instructions for Use

RK-IR Sample Draw Aspirator Adapter Operator s Manual

This document is a preview generated by EVS

Quebec Quebec Tel : Fax : MSDS: Methane, Oxygen in Nitrogen Mixture PRODUCT INFORMATION

COSASCO 6000 PSI DOUBLE ISOLATION SERVICE VALVE MAINTENANCE

GAS CYLINDERS AND VALVES WITH RESTRICTED USE IN THE EU

Decant Filling of Gas Cylinders

Peninsula Dental Social Enterprise (PDSE)

ISO INTERNATIONAL STANDARD. Transportable gas cylinders Inspection and maintenance of cylinder valves

ISO INTERNATIONAL STANDARD

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

Transcription:

SAFETY PRINCIPLES FOR PRESSURE REGULATORS FOR MEDICAL OXYGEN CYLINDERS Doc 104/16 Revision of Doc 104/03 EUROPEAN INDUSTRIAL GASES ASSOCIATION AISBL AVENUE DES ARTS 3-5 B 1210 BRUSSELS Tel: +32 2 217 70 98 Fax: +32 2 219 85 14 E-mail: info@eiga.eu Internet: www.eiga.eu

Doc 104/16 SAFETY PRINCIPLES FOR PRESSURE REGULATORS FOR MEDICAL OXYGEN CYLINDERS Prepared by EIGA WG-15 Medical Equipment Disclaimer All technical publications of EIGA or under EIGA's name, including Codes of practice, Safety procedures and any other technical information contained in such publications were obtained from sources believed to be reliable and are based on technical information and experience currently available from members of EIGA and others at the date of their issuance. While EIGA recommends reference to or use of its publications by its members, such reference to or use of EIGA's publications by its members or third parties are purely voluntary and not binding. Therefore, EIGA or its members make no guarantee of the results and assume no liability or responsibility in connection with the reference to or use of information or suggestions contained in EIGA's publications. EIGA has no control whatsoever as regards, performance or non performance, misinterpretation, proper or improper use of any information or suggestions contained in EIGA's publications by any person or entity (including EIGA members) and EIGA expressly disclaims any liability in connection thereto. EIGA's publications are subject to periodic review and users are cautioned to obtain the latest edition. EIGA 2016 - EIGA grants permission to reproduce this publication provided the Association is acknowledged as the source EUROPEAN INDUSTRIAL GASES ASSOCIATION AISBL Avenue des Arts 3-5 B 1210 Brussels Tel +32 2 217 70 98 Fax +32 2 219 85 14 E-mail: info@eiga.eu Internet: www.eiga.eu

Table of Contents 1. Introduction... 1 2. Scope and purpose... 1 2.1 Scope... 1 2.2 Purpose... 1 3. Definitions... 1 3.1 Publication terminology... 1 4. Recommendations for the selection and use... 2 4.1 Risks... 2 4.2 Selection... 2 5. Implementation and use... 3 6. Routine maintenance... 4 7. Maintenance... 4 8. References... 4 Amendments to 104/03 Section Change Editorial to align style with EIGA template Rewrite to reflect current standards and practices

1. Introduction The pressure regulators that are mounted on transportable gas cylinders for medical use, and particularly those for oxygen, constitute a critical interface in the area of safety in the use of gases. Their use is very diverse including for example hospital services, ambulances and home care patients. The design, manufacture, operation and maintenance of these pressure regulators shall only be carried out by suitably qualified persons with expertise and knowledge in the areas concerned. These aspects shall comply with the applicable regulations and standards. Regulators placed on the market and into service within the European Union shall comply with the Medical Devices Regulations 1.[1] Regulators shall take account of factors such as: The pressure in the cylinder, possibly up to 300 bar; Many common materials that are considered not to be combustible will burn in oxygen; and The conditions of use, that can sometimes be harsh such as impacts and environmental conditions. The pressure regulators shall be operated in accordance with the manufacturers operating instructions and the users shall be trained. 2. Scope and purpose 2.1 Scope Pressure regulators intended for use with medical oxygen or medical breathing gas mixtures with an oxygen partial pressure above 30 bar. 2.2 Purpose This publication provides guidance and safety information for the selection, use, and maintenance of pressure regulators for oxygen cylinders in medical service. It is intended for those who specify, operate, manage and use medical pressure regulators 3. Definitions For the purpose of this publication, the following definitions apply. 3.1 Publication terminology 3.1.1 Shall Indicates that the procedure is mandatory. It is used wherever the criterion for conformance to specific recommendations allows no deviation. 3.1.2 Should Indicates that a procedure is recommended. 1 References are shown by bracketed numbers and are listed in order of appearance in the reference section. 1

3.1.3 May Indicates that the procedure is optional. 3.1.4 Will Is used only to indicate the future, not a degree of requirement. 3.1.5 Can Indicates a possibility or ability. 4. Recommendations for the selection and use The principles for the selection and use of pressure regulators are covered in the following sections. 4.1 Risks The main risks with pressure regulators are: Failure to provide the desired flow and/or pressure of medical gases; Failure to indicate the correct remaining contents of the cylinder; Ignition, that can be caused by adiabatic compression, materials incompatible with oxygen and contaminants such as grease; Release of toxic products in the case of internal combustion; and Uncontrolled release of high pressure and/or ejection of parts. 4.2 Selection The following points need to be taken into account when selecting a medical oxygen regulator. Given the possibility of violent combustion of aluminium alloys in oxygen, their use shall be limited and preferably should be avoided. Suitable alternative materials shall be used such as copper alloys. All parts in contact with oxygen shall be compatible with oxygen. In order to ensure correct operation of the pressure regulator, it should be equipped with an inlet filter made from material compatible for use in oxygen, such as bronze. The use of non-metallic materials in the manufacture of the pressure regulator shall be minimised and chosen taking into consideration the recommendations provided in EIGA Doc 73, Non-metallic materials in high pressure breathing gas applications. [2] The pressure regulator shall be fitted with a pressure relief device, The pressure gauge, if supplied, shall include a flow restriction device, The level of cleanliness shall be in accordance to ISO 15001 Anaesthetic and respiratory equipment -- Compatibility with oxygen, [3]. NOTE Recommendations for cleaning for oxygen service are given EIGA Doc 33, Cleaning of equipment for oxygen service [4]. 2

In the case where lubricants are used, the quantities of these shall be minimised, and they shall be compatible for use with oxygen. The regulator shall pass the adiabatic compression test as specified in ISO 10524-1, Pressure regulators with or without flow-meter for medical gas systems [5]. A regulator complying with ISO 10524-1 [5] and CE marked meets the above criteria. 5. Implementation and use The users, whether they are hospital professionals or home care patients, shall be familiar with the correct use and potential risks before first use. This information shall be provided by the operating instructions supplied with the pressure regulator or by training provided by a competent person. The precautions for use shall include the following: Never use oxygen in the vicinity of a potential point of ignition; Do not use flammable organic materials such as oils, greases, creams; Do not operate the pressure regulator with dirty or greasy hands; Keep the cylinder valve/pressure regulator interfaces clean and free of greasy materials; Prior to connecting the pressure regulator to the cylinder: Check that the pressure regulator has a rated pressure, (see marking on the pressure regulator), above or equal to the working pressure of the oxygen cylinder; Check that the pressure regulator is appropriate for the gas service and application. In particular, never use a medical gas regulator with an industrial gas cylinder; Check the general condition of the pressure regulator. In the event of signs of damage, impact or fall, remove the equipment from service and have it checked and repaired by an authorised specialist; Check the integrity of the cylinder valve/pressure regulator connection, for example; o o Cleanliness and good mechanical condition of the mating surfaces; and Presence, good condition and appropriate positioning of any gaskets. Prior to opening the valve: Ensure that the person operating the valve is behind the valve, opposite the pressure regulator; Check that the flow selector is on zero and nothing is connected downstream of the pressure outlet. Opening the valve: Open the cylinder valve gradually and in case of adjustable regulators, ensure that the regulator screw is in the zero position before opening the cylinder valve. Both will avoid dangerous adiabatic pressure impact; In the event of a leak in the high-pressure side, close the cylinder valve. Never try to re-tighten a defective connection under pressure; 3

In the event of a persistent leak, remove the equipment from service and have it repaired by an authorised specialist. Disconnecting the pressure regulator from the cylinder valve: Make sure that the cylinder valve is closed and that the upstream pressure of the pressure regulator has dropped to atmospheric pressure; and Store the pressure regulator in a clean and safe place. 6. Routine maintenance Routine maintenance is usually performed by the user. Routine maintenance instructions shall be available to the user and this shall include the list of materials required for the routine maintenance, such as gaskets. If it is necessary to replace the gasket make sure that the new gasket is the one recommended by the manufacturer (material / dimension). Any spare gaskets shall be stored in a closed plastic bag to avoid contamination with dust, dirt, grease, etc. It is imperative to use the correct gasket as failure to do this is a common cause of ignition. 7. Maintenance Only technicians who have been suitably qualified and trained shall undertake maintenance of the pressure regulators. Maintenance by users shall be limited to the routine maintenance only and in no case dismantling of the pressure regulator. It is essential to: Ensure that maintenance is carried out at the periods specified by the manufacturer; NOTE Adverse service conditions can require a reduced maintenance period. Comply with the servicing recommendations of the manufacturer; Systematically check the performances of the pressure regulator: Accuracy of the flow delivered at the flow outlet over the full range of settings; and Range of the pressure at the pressure outlet. Use only the spare parts recommended by the manufacturer; Use only cleaning and disinfection products recommended by the manufacturer of the pressure regulator and according to the instructions for use of these products. Prevent ingress of cleaning or disinfection products into the pressure regulator, and always check that there no remaining products in the internal passages of the regulator prior to placing back into service. 8. References Unless otherwise stated the latest edition shall apply. [1] Medical Device Regulations, (under revision) www.ec.europa.eu [2] EIGA Doc 73, Non-metallic materials in high pressure breathing gas applications www.eiga.eu 4

[3] ISO 15001: Anaesthesia and respiratory equipment - compatibility with oxygen www.iso.org [4] EIGA Doc. 33 Cleaning of equipment for oxygen service www.eiga.eu [5] ISO 10524-1 Pressure regulators with or without flow-meter for medical gas systems www.iso.org 5