FMECA Analyses of radiological over-exposure accident to patients in brachytherapy

Similar documents
Risk Assessment Guidance for QM in Brachytherapy. Conflicts. Learning Objectives. Example Procedure

Safety Analysis Methodology in Marine Salvage System Design

Quality Management Determined from Risk Assessment

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

Analysis of hazard to operator during design process of safe ship power plant

Safety-Critical Systems

Determining Occurrence in FMEA Using Hazard Function

Marine Risk Assessment

Physicist's Responsibility On End-Of-Life for Brachytherapy Devices and Software

Hazard Identification

(C) Anton Setzer 2003 (except for pictures) A2. Hazard Analysis

Zoubir Ouhib Lynn Cancer Institute

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

CHAPTER 4 FMECA METHODOLOGY

Safety assessments for Aerodromes (Chapter 3 of the PANS-Aerodromes, 1 st ed)

Lecture 04 ( ) Hazard Analysis. Systeme hoher Qualität und Sicherheit Universität Bremen WS 2015/2016

THE CANDU 9 DISTRffiUTED CONTROL SYSTEM DESIGN PROCESS

Understanding safety life cycles

Risk Management Qualitatively on Railway Signal System

Physics and Quality Assurance

Review and Assessment of Engineering Factors

Application of fuzzy logic to explosion risk assessment

Federal Aviation Administration Safety & Human Factors Analysis of a Wake Vortex Mitigation Display System

Hazard Operability Analysis

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

How to Define Your Systems and Assets to Support Reliability. How to Define Your Failure Reporting Codes to Support Reliability

A study on the relation between safety analysis process and system engineering process of train control system

Using what we have. Sherman Eagles SoftwareCPR.

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

A quantitative software testing method for hardware and software integrated systems in safety critical applications

To comply with the OHS Act, the responsible manager must carry out and document the following:

C. Mokkapati 1 A PRACTICAL RISK AND SAFETY ASSESSMENT METHODOLOGY FOR SAFETY- CRITICAL SYSTEMS

Safety Critical Systems

The Best Use of Lockout/Tagout and Control Reliable Circuits

Product Overview. Product Description CHAPTER

Real-Time & Embedded Systems

University of Cincinnati. Radiation Safety Committee Operations Guidelines Statement of Policy (RSC Guidelines) RSC Guidelines (revision 5)

Aeronautical studies and Safety Assessment

Failure Modes And Effects Analysis Fmea Tool

Safety of railway control systems: A new Preliminary Risk Analysis approach

Safety Manual VEGAVIB series 60

FLIGHT TEST RISK ASSESSMENT THREE FLAGS METHOD

New product release. Universal Rebreather Monitor (URBM) Three cell / Independent backup

D-Case Modeling Guide for Target System

Managing for Liability Avoidance. (c) Lewis Bass

Reliability engineering is the study of the causes, distribution and prediction of failure.

Risk Management Series Article 8: Risk Control

DeZURIK. KGC Cast Knife Gate Valve. Safety Manual

BUBBLER CONTROL SYSTEM

University Safety Policy Supplement

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

FMEA- FA I L U R E M O D E & E F F E C T A N A LY S I S. PRESENTED BY: AJITH FRANCIS

Safety Analysis: Event Classification

Three Approaches to Safety Engineering. Civil Aviation Nuclear Power Defense

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

A Novel Gear-shifting Strategy Used on Smart Bicycles

Apollo Safety Features

SIL Safety Manual. ULTRAMAT 6 Gas Analyzer for the Determination of IR-Absorbing Gases. Supplement to instruction manual ULTRAMAT 6 and OXYMAT 6

4. Hazard Analysis. Limitations of Formal Methods. Need for Hazard Analysis. Limitations of Formal Methods

2017 Edition GUIDANCE NOTE GN 8 INVERTER DRIVES FOR FANS AND PUMPS GN 8 - INVERTER DRIVES FOR FANS AND PUMPS

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

UNIVERSITY OF EDINBURGH GENERIC RISK ASSESSMENT - RADIATION RP/RA8 - WORK WITH X-RAY FLUORESCENCE ANALYSERS

PI MODERN RELIABILITY TECHNIQUES OBJECTIVES. 5.1 Describe each of the following reliability assessment techniques by:

Justification and use of hand-held X-ray equipment

Purpose. Scope. Process flow OPERATING PROCEDURE 07: HAZARD LOG MANAGEMENT

INSTALLATION MANUAL. M-WRG-S/Z-T(-F, -FC) and InControl pushbutton sensor

DeZURIK. KSV Knife Gate Valve. Safety Manual

Failure modes and models

Transformer fault diagnosis using Dissolved Gas Analysis technology and Bayesian networks

ACV-10 Automatic Control Valve

AUSTRALIA ARGENTINA CANADA EGYPT NORTH SEA U.S. CENTRAL U.S. GULF. SEMS HAZARD ANALYSIS TRAINING September 29, 2011

USING HAZOP TO IDENTIFY AND MINIMISE HUMAN ERRORS IN OPERATING PROCESS PLANT

Gas Network Craftsperson

A systematic hazard analysis and management process for the concept design phase of an autonomous vessel.

Qi Litre Ultrasonic Cleaner

Basic STPA Tutorial. John Thomas

Systems Theoretic Process Analysis (STPA)

Surge suppressor To perform its intended functions, an AEI site must have the components listed above and shown in Fig. 4.1.

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

Introducing STAMP in Road Tunnel Safety

Failure Modes and Effect Analysis of Electro-Pneumatics System

HEATEC TEC-NOTE Publication No Revised

Training Fees 3,400 US$ per participant for Public Training includes Materials/Handouts, tea/coffee breaks, refreshments & Buffet Lunch.

A4s Operation Manual

Reliability of Safety-Critical Systems Chapter 3. Failures and Failure Analysis

SWS Option: Well Pressures. SWS Option: Stretch Correction (Corrected depth) SWS Option: MMD (Magnetic Marks Detection) ASEP Products

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

Module No. # 04 Lecture No. # 3.1 Case studies (continued) (Refer Slide Time: 00:10)

The Criticality of Cooling

STICTION: THE HIDDEN MENACE

in84061.txt at Page 1 of 6 2/25/00

10 Best Practices for Electrosurgical Unit Testing

Maintenance and general care of secondary standard and field instruments

Technical Data Sheet MF010-O-LC

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

RAM 4021 Operation Manual

LECTURE 3 MAINTENANCE DECISION MAKING STRATEGIES (RELIABILITY CENTERED MAINTENANCE)

SEMEM3-78 Testing and calibrating instrumentation and control equipment and circuits

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

Transcription:

FMECA Analyses of radiological over-exposure accident to patients in brachytherapy M. Casamirra, F. Castiglia, M. Giardina, E. Tomarchio Department of Energy, Nuclear Section, University of Palermo, Viale delle Scienze, 90128 Palermo, Italy E-mail: mgiardina@din.unipa.it Abstract. This paper presents safety analyses of accidental events which can involve patient during High Dose Rate brachytherapy treatment in over-exposures. The study has been performed by using the well-known techniques FMECA modified by Fuzzy logic theory. Moreover, fuzzy HEART methodology was employed in order to evaluate human error probabilities for each treatment stage. The obtained results, aimed to obtain a list of the deviations with a reasonable probability to produce significant adverse outcomes, provided some recommendations for procedures and safety equipments to reduce the occurrence of radiological over-exposure accidents. Keyword: High Dose Rate, radiation therapy, radiological risk, FMECA, fuzzy RPN. 1. Introduction In 2001, Committee 3 (protection in medicine) of the International Commission on Radiological Protection (ICRP) formed a Working Party to study problems associated with the widespread introduction of High Dose Rate (HDR) brachytherapy, which concluded that a full report addressing such problems was desirable. Subsequently, a Task Group on Prevention of HDRate Brachytherapy Accidents performed studies to identify reported accidents and incidents by category (transport, handling, procedures, dosimetry, etc..); to identify potential events; and to analyze the geographic distribution of machines and trends. It is estimated that about 500,000 procedures (administrations of treatment) are performed by HDR units annually. This technique delivers a very high dose, of the order of 1.6 5.0 Gy/min, so mistakes can lead to under- or over-dosage with the potential for clinical adverse effects. More than 500 HDR accidents (including some deaths) have been reported along the entire chain of procedures from source packing to delivery of dose, and the revision of their root causes confirms that the accidents rarely occur due to a single failure or human error [1]. They are instead caused by the combination of different events that denote weaknesses in the radiological practice. So the prevention of radiological exposure accidents requires design of proper radioprotection systems with more quality assurance and highly qualified maintenance, together with accurate risk analyses which should allow to highlight the causes of the possible shortages and, if necessary, to predispose the safety devices against those events with high occurrence frequency, especially in case of radiotherapy procedure where the over-dosage accidents have devastating and sometimes fatal consequences. In order to achieve this aim, the paper presents the results obtained in the safety analyses of some representative accidental events involving patient during HDR brachytherapy treatment in over-exposures, carried out by using modified FMECA (Failure Mode, Effects, and Criticality Analysis) methodology, for obtaining an exhaustive list of the deviations with a reasonable probability to occur. FMEA (Failure Mode and Effect Analysis)is a powerful qualitative tool for systematically identifying the initiating event and the root causes of unacceptable outcomes, though this tool has difficulties to identify accident sequences and dependences between equipment or human actions. In FMECA analysis, to classify each critical failure the calculation of Risk Probability Number (RPN) is used. In the present paper RPN fuzzy rule-based assessment model based on fuzzy logic theory [2] is proposed to systematically review of system or component failures as well as human errors in Brachytherapy treatment. Moreover to evaluate the error probability of human actions the fuzzy HEART methodology has been employed, as proposed in [3, 4]. The obtained results have allowed to characterize the critical points of the radioprotection systems, highlighting the human errors which principally contribute in occurrence frequency of high dose rate patient exposures. Moreover, it has been possible to suggest the introduction of suitable operating procedures that reduce the over-dose risk.

2. Safety analyses of the patient exposure in HDR brachytherapy Brachytherapy treats cancer by placing radioactive sources directly into or next to the area requiring treatment, enabling clinicians to deliver a high dose with minimal impact on surrounding healthy tissues. Among the devices for the medical applications, the use of remote afterloading of radioactive sources is becoming increasingly popular in much countries because these units offer both the potential for superior dose distributions and the practical advantages of better radiation protection. In this field, the safety assessment in high dose-rate HDR treatment delivery practices at the Oncological Unit of Paolo Giaccone Policlinic of Palermo (Italy) has been performed. The examined system consists of a motor-driven source transport device for automatically transferring radioactive material between a shielded safe and the treatment applicator (MicroSelectron HDR manual)[5]. The device contains a small, sealed, 450 GBq 192 Ir stepping source, mounted at the end of a stainless steel drive wire. The afterloader is connected to the implanted applicator, catheter or needle using flexible transfer tubes. Typically, the dose is calculated at representative points by the Treatment Planning System (TPS) and then the therapy is performed by Treatment Control System (TCS) which enables an operator to apply, by remote control, a radionuclide source into the body. The TCS supports amongst others the following functions: Make a new plan; Add a plan (from the Library); Import a plan (from a treatment planning system); Load a patient study (from the database); Edit, print and save Plan information; Execute treatment; Start and monitor treatment; Print a treatment report. For any computer system that produces isodose curves for remote afterloaders, it is imperative that the user understand the algorithm and exactly how the doses are computed. The user may choose to enter key parameters specific to a radionuclide and source model, or select parameters from an existing menu. The after-loader unit contains (Fig. 1): Geiger-Muller counter to check the source in safe; Stepping Motor (SM) to drive the source Direct Current (DC) motor available for source retraction in case of SM failure, supplied by backup battery in case of electrical blackout; check cable, identical in appearance to the source cable (but not radioactive) to check the treatment channel, transferring tube and the distance of the pre-programmed source positions; opto-pair to verify source position; several channels for source transport and transfer tube connector; primary and secondary timers for measurement the pre-programmed exposure times. A last resort mechanical system for manually returning the source to the safe, in the event of other electrical source return mechanisms fail, is included. The check cable is connected with a second stepper motor. When the appropriate stepper motor rotates the source or the check cable is advanced in the treatment channel and positioned with an accuracy of + 1 mm. The afterloader system drives the source mechanically to pre-programmed locations in the applicator (dwell positions) and holds it in place for certain pre-programmed periods of time (dwell time). A primary timer measures the exposure time for each source dwell position, whereas a secondary timer measures the time between the source channel transit time and the dwell time for each pre-programmed locations. The software system in TCS checks the consistence between the measurements performed by the timers, and the data are also compared with those reported in the treatment plan. If error occurs, the source is withdrawn into the safe position. In case of failure of both timer, the operator in the (Control Room) CR is

alerted by light alarm and warning in the consol display. Prior to the initial use of a new (or replacement) applicator, it is necessary to verify that the source dwell positions correspond to the radiographic marker positions used in simulation and treatment planning. The dose distribution can be optimized by adjusting the dwell time at each source position. Moreover, for each type of HDR treatment, the procedures and duties of each team member should be carefully defined. A radiation therapist or physicist assisting the physician during the applicator placement should document: the applicators type used, body anatomy, surface markers. Fig. 1 Components of the after-loader unit. The indexer guides the source into one of the 18 source channels. Before the active source is inserted into any of the channels, a dummy source is inserted to check for obstructions. The control console, located inside the CR operates the after-loader, shows the source position on the display, the treatment progress, and prints out a report of the treatment. Tree emergency stop push-buttons, located on the HDR unit, inside the Treatment Room (TR) and on the control console located in the CR, respectively, allow interruption of the procedure at any time. When the treatment is in progress, an electrical switch detects if the TR door is closed. If the operator erroneously opens the door during the treatment, the irradiation process is interrupted by the DC safety motor, which withdraws the source in safe. 3.1 FMECA Analysis FMEA may be performed to identify failure scenarios in the examined facility, i.e. potential accident initiators by systematically reviewing the failure of each system or component in terms of its potential consequences. The FMECA analysis is a procedure that is performed after a FMEA analysis to classify each potential failure effect according to its severity and probability of occurrence (Pillay and Wang, 2003). In particular,

three numerical values can be used to describe each failure mode: Occurrence (O) which describes to the probability that a particular accidental event occurs ; severity (S) which is a measure of the severity of the consequences resulting from the failure mode if it is not detected and corrected; Detectability (D) which describes the probability that the failure will be detected before the treatment commences or the failure is effective. Multiplying these three numbers together yields a Risk Probability Number (RPN) which can be used for prioritizing quality control tests and activities. In general, these three parameters are estimated by experts in accordance with a scale from 1 to 10 based on commonly agreed evaluation criteria. Higher value points to critical situation. Tables 1 through 3 summarize the evaluation criteria for occurrence, severity, and detect ratings, respectively, which is used practically in high-risk medical applications. The calculated RPN rank will be between 1 and 1000, and some users define priorities in their FMECA procedure as: - acceptable if RPN < 200 - undesirable 200 < RPN <500 - unacceptable if RPN > 500 In this paper to evaluation of RPN number, a fuzzy rule-based assessment models, similarly to one suggested by Guimarees et al. [6, 7], is used to identify the critical events relevant both human errors and potential failures. As well known, a fuzzy rule can be created by combining various fuzzy sets, to form the premise (input fuzzy sets) and the consequence (output fuzzy sets) of the rule. The fuzzy inference sequence are performed by the following steps: fuzzy input composition, rule evaluation, fuzzy output composition, defuzzification. The fuzzy input and output composition is the process of decomposing input and outputs variables into one or more fuzzy sets. For example, the first step to perform a risk study based on fuzzy logic is to define the linguistic terms connotations and the relation with the factors used on risk estimation. After the inputs and outputs have been decomposed into fuzzy sets, a set of fuzzy if-then-else rules is used to process the inputs and produce a fuzzy output. Each rule consists of a condition and an action where the condition is interpreted from the input fuzzy set and the output is determined on the output fuzzy set. In other words fuzzy inference is a method that interprets the values in the input vector and, based on some set of rules, assigns values to the output vector. In this research, the parameters O, S, and D (used as input fuzzy sets) are combined as linguistic data. The fuzzyfication process is based on five fuzzy triangular and trapezoidal membership functions that show the degree of potential attribute as follows: Very High, High, Moderate, Low, and Very Low, denoted as VH, H, M, L, and VL. The graphical representation of fuzzy membership function to Occurrence, Severity and Detection are identical and only one, Occurrence is shown in Fig. 2. The fuzzy output RPN was scaled in the range 0 through 1000 in order to be compatible with the classic results (Fig. 3) and the corresponding five membership functions are: Acceptable (A), Almost Acceptable (AA), Undesirable (U), Almost Unacceptable (AU), and Unacceptable (U).These RPN linguistic representations taken into account the classification above described. Concerning the fuzzy IF-THEN rules, it assumes the form: If x is A then y is B (1) where A and B are the linguist variable defined by fuzzy sets on the ranges (universe of discourse) X and Y, respectively. In the present study, the fuzzy rule base has 125 (5 O 5 S 5 D). A simple example follows: If Occurrence is L and Detection is VL and Severity is VH then RPN is VH (2) Table 4 presents a sample of the inference rules adopted for fuzzy RPN evaluation. The finally step is the defuzzification to obtain a crisp value.

1,2 O, S, D fuzzy sets definition degree of membership 1 0,8 0,6 0,4 0,2 Very Low (VL) Low (L) Moderate (M) High (H) Very High (VH) 0 0 2 4 6 8 10 Fig. 2- Occurrence fuzzy sets definition. 1,2 RPN fuzzy sets definition 1 degree of membership 0,8 0,6 0,4 0,2 Acceptable (A) Almost Acceptable (AA) Undesirable (U) Almost Unacceptable (AU) Unacceptable (U) 0 0 200 400 600 800 1000 Fig. 3- RPN fuzzy sets definition. RPN Table 1 - Occurrence rating. Probability of Human error occurrence Component failure Rank occurrence probability occurrence probability Linguistic value Failure unlikely One time per 10 year < 5 10-6 1 VL Few Failures Some time per 5 year 5 10-6 10-4 2, 3 L Occasional failures Some time per 2 year 10-4 5 10-3 4, 5, 6 M Repeated failures One time per year 510-3 5 10-2 7, 8 H High Inevitabile failure More time per year < 0.5 9, 10 VH

Table 2 Severity rating. Severity ranking Rank Linguistic value Very minor No effect 1 VL Very Low Minor effect 2, 3 L Moderate Potential ineffectiveness 4, 5, 6 M High Regulatory non-compliance 7, 8 H Very High Injury or death 9, 10 VH Table 3 - Detection rating. Likelihood of detection ranking Rank Linguistic value Almost Certain 1 VH High 2, 3 H Moderate 4, 5, 6 M Remote 7, 8 L Absolute uncertainty 9, 10 VL Table 4 - Inference rules adopted for fuzzy RPN evaluation in case of Occurrence Very Low. Occurrence VL Severity VL L M H VH Detection VH VL VL VL L M H VL VL L M M M VL L M M H L L M M H H VL M M H H VH 3.1. FMECA analyses in HDR Brachytherapy By applying fuzzy FMECA technique, a systematically review of system or component failures as well as human errors in Brachytherapy treatment have been studied. To evaluate human errors probability the fuzzy HEART methodology has been employed. The obtained results are reported in Table 5 in terms of triangular fuzzy error probability sets and crisp value of the probability expressed as error per years, obtained by using the center area defuzzification method. The number of patient per years is considered equal to 100. The probabilities have been evaluated by taking into account that the operating medical team is trained in operating practices, whereas the conditions that can favour the error are assessed to be Shortage of time for error detection or correction, Noisy signal, A means of suppressing or overriding information, and Mismatch between perceived and actual risk. The obtained results in classic FMECA approach are reported in Tables 6 and 7 for components failures and human errors, respectively. As shown in Figure 3, the more critical events are rated in the following order: data insertion errors in TPS (ID 11), error in data entry of dwell time or dwell position programming (ID 14), backup battery failure (ID 7), dose calculation errors in TPS (ID 10), incorrect identification of the patient (ID 12). In Figure 4 the results obtained in terms of fuzzy RPN numbers are shown. In this case the rating is: backup battery failure (ID 7), data insertion errors in TPS (ID 11), error in data entry of dwell time or dwell position programming (ID 14), dose calculation errors in TPS (ID 10), during treatment, the stop button in the console did not retract the wire source (ID 9), failure of the computerized security program with incorrect calculation after wrong data entry or incorrect use of source strength, or step size, tip length (ID 8), incorrect identification of the patient (ID 12).

500 450 400 350 300 250 200 150 100 50 0 1000 900 800 700 600 500 400 300 200 100 0 RPN in FMECA analyses RPN ID1 ID2 ID3 ID4 ID5 ID6 ID7 ID8 ID9 ID10 ID11 ID12 ID13 ID14 ID15 Figure 3. RPN indexes classic FMECA analyses. Fuzzy RPN in FMECA analyses Fuzzy RPN ID1 ID2 ID3 ID4 ID5 ID6 ID7 ID8 ID9 ID10 ID11 ID12 ID13 ID14 ID15 Figure 4. Fuzzy RPN indexes. Table 5. Triangular membership function of human error probabilities and relevant crisp number Human errors Probabilità fuzzy probability expressed in error per year x 1 x 2 x 3 Dose calculation error in TPS 1.98E-3 2.84E-3 3.84E-3 0.2 Data entry error in TPS 4.22E-3 6.24E-3 8.64E-3 0.4 Incorrect identification of the patient 1.08E-2 2.50E-2 4.98E-2 1.0 Error in load a patient study (from database) 4.22E-3 6.24E-3 8.64E-3 0.4 Error in data entry of dwell time or dwell position programming 4.22E-3 6.24E-3 8.64E-3 0.4 Incorrect medical application of catheter, or applicator 1.54E-3 2.16E-3 2.88E-3 0.6

On the basis of the results described above, some recommendations on safety equipment and procedures can be adopted to reduce the risk of radiological over-exposure. For example, periodic maintenance of the backup battery can prevent component faults, whereas an acoustic alarm can be provided to signal the condition of uncharged battery. When the treatment is in progress, an electrical switch detects if the TR door is closed, if the operator erroneously opens the door during the treatment, the irradiation process is interrupted by the DC safety motor, which returns the source to the safe. This safety device allows also to dispose a redundant system in case of stop button in the console failure to withdraw the source in safe, if necessary. For the medical staff tasks, TPS and TCS treatment data can be checked and compared both by Therapist and Physicist. Moreover passport type photo of the patient can be enclosed in treatment plan. It is worth to highlight that the fuzzy approach procedure allows a more accurate ranking classification of the critical events. 4. Conclusion The paper presents the results obtained by safety analyses of accidental events which can involve patient during HDR brachytherapy treatment in over-exposures, performed by using modified FMECA techniques. In particular, the calculation of the Risk Probability Number (RPN) is performed by using fuzzy rule-based assessment model based on fuzzy logic theory. Moreover, the fuzzy HEART methodology is also employed in order to evaluate human errors for each treatment stage. As expected, the obtained results suggest that the events related to human error are very significant and important in the accidental scenarios. On the basis of the obtained results in terms of fuzzy RPN, it is worth to highlight that the fuzzy approach to RPN evaluation produces a more accurate ranking about the critical events importance, so it is more immediate to provide some recommendations for procedures and safety equipment to reduce the occurrence of radiological over-exposure accidents. Acknowledgments The authors tanks the Nuclital for furnishing manual of the MicroSelectron afterloading unit. References 1. ICRP, 2005, Prevention of high-dose-rate brachytherapy accidents, (ICRP Publication 97), Ann. ICRP 35. 2. Zadeh, L.A., 1975, The concept of a linguistic variable and its application to approximate reasoning, Parts 1 and 2, Inf. Sci. 8, 199 249, 301 357. 3. Castiglia F, Giardina M, Tomarchio EA (2010). Risk analysis using fuzzy set theory of accidental exposure of medical staff during brachytherapy procedures. JOURNAL OF RADIOLOGICAL PROTECTION, vol. 30 (2010), p. 49-62. 4. Casamirra M, Castiglia F, Giardina M, Tomarchio E (2009). A fuzzy modelling of HEART methodology: application in safety analyses of accidental exposures in irradiation plants. RADIATION EFFECTS AND DEFECTS IN SOLIDS, vol. 164. 5. MicroSelectron-HDR, User Manual, Nucletron, Holland, 2005. 6. Guimaraes, A. C. F., Lapa C. M., F. Fuzzy FMEA applied to PWR chemical and volume control system, Progress in Nuclear Energy, Vol. 44, No. 3, pp. 191-213, 2004. 7. Guimaraes, A. C. F., Lapa C. M., Hazard and operability study using approximate reasoning in lightwater reactors passive systems, Nuclear Engineering and Design, 236, pp1256-1263, 2006.

Table 6 - FMECA of components faults ID Component Failure mode Failure effect (system) Failure Detection method Failure mode frequency (1/h) 1 SM stepper motor Loss of power HDR unit stopped its operation DC motor withdraws the source in safe 2 DC safety motor Loss of power HDR unit stopped its operation Operator goes in TR to manually return the source in safe 3 Opto-pair sensors 4 Dwell Position Distance control device light sensor fault Stepper motor failure 5 Primary Timer Electronic fault 6 Secondary Timer Electronic fault Light alarm Warning in user control panel Light alarm Warning in user control panel Patient Failure O S D RPN effect 4.57E-7 No 1 1 3 3 1.67E-7 Patient overexposure Source position not verified Warning in user control panel 2.0E-9 Treatment not completed correctly Source position not correct Source dwell time error Not correct check of primary timer 7 Backup battery Power-off DC motor fault in case of electrical blackout 8 Software Power-off Failure of the computerized security program with incorrect calculation after wrong data entry OR incorrect use of source strength, or step size, tip length 9 Stop button in Contact fault During treatment, the stop the console button in the console did not retract the wire source Radiographic marker position not corresponding to dwell position Inconsistence between the two timers measurements source is withdrawn into the safe position Source is withdrawn into the safe position No No 2.0E-7 1.0E-5 1.0E-5 2.41E-5 1.0E-9 Erroneous treatment Treatment not completed correctly Treatment not completed correctly Patient overexposure Patient overexposure No 2.28E-7 Patient overexposure 1 9 3 27 1 5 4 20 1 7 3 21 2 5 2 20 2 5 3 30 3 9 9 243 1 7 9 63 3 7 9 189

ID Component Failure mode Failure effect (system) 10 physicist dose Incorrect HDR calculation treatment errors in TPS 11 therapist data insertion Incorrect HDR 12 medical operator errors in TPS Incorrect identification of the patient 13 therapist Error in load a patient study (from database) 14 therapist Error in data entry of dwell time or dwell position programming 15 medical operator Incorrect medical application of catheter, or applicator treatment incorrect data are used in TCS incorrect data are used in TCS incorrect data are used in TCS Incorrect HDR treatment Tale 7 FMECA of human errors Failure Detection method Failure mode frequency Patient Failure (error per year) effect no 0.2 Patient overexposure no 0.4 Patient overexposure Therapist asks the patient name 1.0 Patient overexposure first of beginning treatment Data are not corresponding to TPS 0.4 Patient overexposure no 0.4 Patient overexposure A radiation therapist, or physicist assisting the physician during the applicator placement Radiographic marker position not corresponding to treatment plan 0.6 Patient overexposure O S D RPN 2 8 9 144 4 8 9 288 8 9 2 144 4 8 2 64 4 8 9 288 5 9 1 45