Version BA D

Similar documents
TapeStation Tips & Tricks

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Pipettor. User Manual

Gel Loading Practice MiniLab

RayPette Plus Autoclavable Pipette. User Manual

UsER manual for Watersens ph -REDOX

Technical Procedure for General Laboratory Equipment

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

The pipettes cover a volume range from 0.1µl to 10ml.

EcoCell Vertical Electrophoresis Unit

[USING THE VITROBOT April 9, Using the Vitrobot

User Manual MFCS TM -EZ

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Loopamp Bovine Embryo Sexing Kit Instruction Manual

SOP for use of the SX19 Stopped-Flow Instrument

Pipetting Small Volumes

Compact Ball Valves ASAHI AV VALVES. Contents. User s Manual. (Page) (1) Be sure to read the following warranty clauses of our product 1

NovelBright LED Illuminator

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Underwater Housing for Panasonic Lumix DMC-ZS60, DMC-TZ80

Department of Genetics

BD Cytopeia Fluidic Kit User s Guide

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

TITLE: DNA minitube - Red

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Vacuum Bagging Wings Instruction Manual Purdue University

UNIVERSITY OF ROCHESTER MEDICAL CENTER

Lab 1. Instrumentation Familiarity: Using Micropipetters and Serological Pipettes

Velocity USER GUIDE CPU WATER BLOCK. 2nd Revision, October 18th 2018

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Using the Akta Prime plus October 22, 2012

PIPETMAN F ENGLISH. Other languages available on our website:

Underwater Housing for Canon G7X Mark II

Paradise Pump Series Operator s Manual

Oxygen Sensor with Electrode

Varipette Operating manual

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

Lipodrop 2.0 device coating

SEMEMI SQA Unit Code H2B7 04 Assisting in the installation of compressed air equipment

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

How to Measure R7.1. Reference. I. Linear dimensions

VAC-U DRIVE Swinging float arm for Acrobat like agility 36 Fin disk with DEEP VEE Scoop

Using a micropipette to transfer exact liquid measurements in scientific experiments By: Nicole Hume

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Elcometer 138. Bresle Kit and Patches. Operating Instructions

14500 Coy Drive, Grass Lake, Michigan Isolation Glove Box Instruction Manual

Cryolys is a registered trademark of Bertin Technologies.

W I L D W E L L C O N T R O L FLUIDS

Standard Operational Procedure for Capillary Electrophoresis

SEMIC05 Installing mechanical equipment

Elcometer 138. Bresle Kit and Patches. Operating Instructions

General remarks regarding ALS robot system:

Sea Rebel Kayak Owner s Manual

Velocity USER GUIDE CPU WATER BLOCK. 1st Revision, September 6th 2018

Wet Lamination of Dry Film Photoresist for the Hobbyist

MOLEBIO LAB #1: Microquantity Measurement

AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide

Thank you for purchasing your new Empire Reloader B Sound-Activated 3-Speed Paintball Hopper!

Pressure Dump Valve Service Kit for Series 2300 Units

Digital Melting Point Apparatus

User s Manual. USB Golf Simulator. Contents

Chapter 2. Pipetting techniques. Guide to Pipetting

Best Practices for Infinium Assays

User Manual. Rotor Gene Q and Rotor Gene 3000

Manual. Bypass Feeder. Installation Maintenance Repair Manual. Chemical Addition and Filtering

HemoCue Hb Procedure Template

SPUD OWNER S MANUAL ALWAYS WEAR A COAST GUARD APPROVED FLOTATION DEVICE WHEN OPERATING THE SPUD KAYAK.

Standard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442

Standard. Abdominal Aortic Aneurysm (AAA) Repair Trainer. Part No: 60610

INSTRUCTION MANUAL. January 23, 2003, Revision 0

Congratulations, once again, on buying an E RIDER Model 18!!!!

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

Quick Reference Guide

PURDUE UNIVERSITY TEFLON CORE PCB THRU-HOLE PREPARATION FOR BLACK HOLE APPLICATION RF LAB. Professor in charge: William Chappell MSEE 289

REL-46 WARNING NOTICE 15 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD. Compatible with RELIABLE R15 and P Style dies. REL-46 Manual

Calibration Procedure

Abdominal Aortic Aneurysm (AAA) Repair Trainer User Guide

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy


Installation Instructions

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

REL-510H WARNING NOTICE 12 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD

User's Manual. Autoclavable Nichipet EX Plus Digital micro pipette for liquid handling CERTIFIED ISO9001

ANTARES. User Manual ANTARES OVAL DRY GLOVE SYSTEM BY SI TECH. Photo: Jonas Andersson, Lysekil

QUICK START GUIDE TO DRINKING WATER MONITORING

Flow experiment with the ibidi pump system and µ-slide I 0.6 Luer

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

USER MANUAL. MACMAG magnetic separator MHK MHF

Microfluidic Demonstration

Decloaking Chamber NxGen Manual Digital programmable pressure system for heat-induced epitope retrieval

AR STYLE FIREARMS OWNER'S MANUAL: OPERATION, HANDLING, DISASSEMBLY / REASSEMBLY & SAFETY INSTRUCTIONS

SEMPEO SQA Unit Code FP2M 04 Maintaining mechanical devices and equipment

Written By: Scott Havard

Handle the film very carefully. As with aluminum foil, once creased, the film will remain creased.

Fiber Optic SMA Z-Flow Cell Manual Version 2.1

Transcription:

Aurora Dispo osable Cartridge Handling Manual For Aurora cartridge 210 0001 CA D Version 1.00 106 0010 BA D

2011 Boreal Genomics, Inc. All rights reserved. All trademarks are property of their owners. http://www.borealgenomics.com support@borealgenomics.com +1 (604) 822 4111 2

Table of Contents Introduction... 4 Cartridge compatibility... 4 Safety guidelines... 4 Procedure overview... 4 Preparing the sample... 4 Preparing the cartridge... 5 Load the sample... 10 Prepare the Aurora... 10 Begin the run... 10 Sample extraction... 10 Disposal... 11 Troubleshooting... 11 Ordering and support... 12 3

Introduction This manual describes the proper method of preparing a disposable cartridge for use in the Aurora, loading the sample in the cartridge and extracting the purified DNA from the cartridge once the Aurora purification is complete. Details on preparing a variety of sample types prior to loading them into the Aurora cartridge can be found in separate Aurora protocol documents on the Boreal Genomics website. Cartridge compatibility The procedures described in this manual apply to Aurora cartridge 210 0001 CA D. Safety guidelines Cartridges are made from non hazardous plastics, metal, TBE buffer and agarose. Appropriate precautions should be taken if hazardous samples are used with the cartridges. Wear gloves during all stages of the protocol. Avoid skin contact with all reagents. Refer to safety guidelines in Boreal Genomics Protocol documents for further information. Procedure overview Prepare the sample Prepare the cartridge Load the sample Prepare the Aurora Begin the run Sample extraction Preparing the sample Please see your Aurora protocol documentation for details on the method of preparing your sample prior to loading it into the Aurora cartridge. 4

Preparing the cartridge Buffer chamber C Extraction well Buffer chamber D Buffer chamber F Sample chamber Buffer chamber E Sample gel Buffer chamber B Concentration gel Buffer chamber A Figure 1. Top view of the cartridge. Opening the cartridge To remove a cartridge, open the packaging by cutting the endd of the bag open using scissors. Inspecting the cartridge Upon removal, each cartridge should be inspected for damage or defects. Examine: The graphite electrodes for chips or cracks, The cartridge plastic for cracks, The concentration and sample gels for presence of bubbles, The cartridge for signs of leakage (wetness of the cartridge or its packaging), and The buffer chambers and the sample chamber for presence of buffer; each should be nearly full of liquid along with some air bubbles. If you observe any damage or defects, please contact support@borealgenomics.com. 5

6

Removing the sealing tape In order to gain access to the sample and buffer chambers, the clear adhesive sealing tape must be removed from the top of the cartridge. To remove the sealingg tape from the cartridge: Place the cartridge on a sturdy, flat surface. Hold the cartridge down firmly. Peel off the sealing tape, starting from the exposed flap at one corner of the cartridge. Figure 2. Removing the sealing film from the cartridge. Removing packing buffer Before using the cartridge, excess packing buffer must be removed in order to prevent buffer overflow during cartridge operation. The buffer in each chamber is 0.25x TBE. To remove excesss buffer: Remove and discard 2 ml of buffer from each of the 6 buffer chambers (A F), using a 1 ml pipette or a serological pipette to accesss the buffer chambers through the access holes. Remove and discard all of the buffer (approximatelyy 9 ml) from the sample chamber taking care not to disturb the sample gel or extraction gel. Preparing the extraction well A fresh 60 µl sample of buffer must be added to the extraction well prior to the run in order to ensure consistent run results. The extraction well must then be sealed in order to prevent output volume fluctuations. To prepare the extraction well: Remove the small piece of adhesive film covering thee extraction well. Extract and discard the buffer from the extraction well (approximately 60 µl). Transfer 60 µl of buffer from any of the buffer chambers into the extraction well. Ensure the surface of the cartridge above the concentration gel is dry, so that the adhesive film will provide a good seal. Wipe off excesss buffer with a Kimwipe if necessary. Remove the white backing from the supplied piece of adhesive film and place it on top of the extraction well as shown in Figure 3 below. One edgee of the film should stick up above the cartridge slightly to allow for easier removal of the adhesive film after the run. With a gloved 7

finger, firmly press down on the film, ensuring a goodd seal on all edges and taking care to avoid wrinkles and bubbles. Align the extraction well sealing tool to the extractionn well, as shown in Figure 4. Firmly press down on the tool until it is fully compressed, as shown in Figure 4. Release and then press down a second time to ensure a good seal. Adhesive film Figure 3. Cartridge with adhesive film over extraction well. 8

Extraction well sealing tool Figure 4. Extraction well sealing tool aligned and uncompressed (left) and fully compressed (right). Placing the cartridge on the cold plate To load the cartridge into the Aurora instrument: Pull gently outward at both left and right sides of thee cartridge drawer until the drawer slides out and latches in the open position Examine the cold plate and the metal bottom of the cartridge for debris that may prevent the cartridge from sitting flat on the cold plate. Remove any debris that is found. Pipette 1 ml of water onto the center of the cold plate and place the cartridge onto the cold plate oriented so the concentration gel is on the left hand side, ass pictured Figure 5 below. Ensure that the cartridge is sitting flat on the surface of the cold plate and is not resting on top of the cartridge registration features. These steps ensure good thermal contact between the cartridge and cold plate. Final cartridge inspection Inspect the empty sample chamber prior to loading your sample. If any buffer has leaked into the sample from the other reservoirs, use a pipette to remove the buffer and immediately proceed to the next section, Load the sample. For best results, the sample should be loaded and the run started as quickly as possible once you have removed the buffer from the sample chamber. 9

Cartridge Cartridge registration features Cartridge drawer Figure 5. Cartridge oriented correctly on the cold plate. The focusing gel is on the left hand side. Load the sample Bring the volume of the sample you prepared earlier up to 5 ml with deionized water. Transfer the entire 5 ml sample into the empty sample chamber (see Figure 1). Close the cartridge drawer. Prepare the Aurora Prepare the Aurora and chiller as described in the user manual. Briefly, turn on the chiller and the Aurora. Start the chiller and ensure that a plus or minus symbol appears on the display to indicate that thermal control is operating. If a star appears, press Start again. Begin the run From the home screen of the Aurora software, load and run the protocol appropriate for your application (ex. 106 0001 BA D AURORA DNA CLEAN UP PROTOCOL) as described in the Operating the Aurora section of the Aurora User Manual. Create an experiment folder for the run to save the logs. The run will typically complete in 2 4 hours depending on the protocol that you are running. Sample extraction After a run is complete, to extract the concentrated sample: 10

Pull gently outward at both left and right sides of thee cartridge drawer until the drawer slides out and latches in the open position. Lift the cartridge from the platform, taking care not to spill any liquid, and place the cartridge on a flat work surface. Examine the concentration gel and ensure that theree are no large bubbles present. Presence of large bubbles indicates thatt the gel overheated during the run, which will adversely affect the quality of the output. Using forceps or gloved fingers, remove the clear film covering the extraction well in the center of the gel. Transfer concentrated DNA from this well using a pipette, taking care not to disturb the concentration gel. Typically, two pipetting steps are required in order too extract most of the liquid from the extraction well. Be sure to extract any sample that is suspended on the walls of the extraction well. The expected output volume is 50 60 µl, but may vary from 40 700 µl depending on sample and run conditions. Disposal Dispose of the cartridge and contents following all applicablee policies, laws, and regulations. Aurora cartridges, buffers, and gels are non hazardous as shipped, however cartridges that have contained hazardous samples (including many nucleic acid stains) may be considered hazardous in your jurisdiction. Important The user is responsible for the safe use, transport, storage, and disposal of any materials that may be considered a chemical or biological hazard. Refer to accompanying material safety data sheets (MSDS) for safety information and handling instructions, and comply with all federal, state/provincial, municipal/local, and institutional requirements and guidelines for disposal. Troubleshooting Please seee the Aurora User Manual for more information about troubleshooting instrument faults. Troubleshooting information related to processing your sample can be found in your protocol documentation. 1. The Aurora or the camera does not connect to the computer. Unplug the instrument s USB cable, wait five seconds, and reconnect the cable. 11

2. A cartridge fails a contact test. Before each run, the Aurora checks to make sure the instrument has good electrical contact with the cartridge. Poor contact can be caused by misalignment of the cartridge. Re seat the cartridge firmly within the registration features on the cold plate. Check that the correct contact plate is installed. If your contact plate has spring pin electrodes, check to ensure that none of the spring pins are jammed in the up position. If you notice stuck pins, see the Aurora manual for cleaning instructions. If your contact plate has coiled spring electrodes, check to make sure there is nothing blocking contact with any of the springs and the graphite electrodes on the cartridge, such as PCR tape or salt. 3. A sample current is below minimum current limit warning appears. The Aurora may not be making electrical contact with the cartridge. Pause the run, open the cartridge drawer, check that the cartridge is firmly seated within the registration features on the cold plate, close the cartridge drawer, and press the play button to resume. If the cartridge fails the contact check, see above. 4. A temperature controller error appears. Make sure the chiller is on and operating. If you see a * on the display, the chiller is pumping coolant but temperature control is not active. Press Start/Stop on the chiller to enable temperature control. 5. Error: The Aurora control software warns that the sample is too conductive. Running high conductivity samples will result in lower yields and may cause gel damage and other issues during the run. The Aurora instrument will give the warning Injection Conductivity test failed. Sample conductivity is too high. Injection might fail for highly conductive samples. Conductivity for a 5 ml sample should be 100 μs/cm. Please refer to the Troubleshooting section in your Protocol document for further instructions on reducing sample conductivity. Ordering and support For support for Aurora protocols or cartridges, or to order additional cartridges, please contact support@borealgenomics.com. This protocol uses cartridge part number 210 0001 CA D. 12