Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Similar documents
Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

TG-51: Experience from 150 institutions, common errors, and helpful hints

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Abstract: Introduction:

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

Relative Dosimetry. Photons

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Review of fundamental photon dosimetry quantities

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

Monitor Unit Verification for Small Fields

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

1. Question Answer cgy / MU cgy / MU 2. Question Answer

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Zoubir Ouhib Lynn Cancer Institute

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Proton Therapy QA Tools

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

THE development of more advanced techniques in radiotherapy,

Introduction. Materials and Methods

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Dosimetric Calculations. Lonny Trestrail

Constancy checks of well-type ionization chambers with external-beam radiation units

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC #

TEST REPORT BEAMSCAN WITH HALCYON

SnapShot IMRT with compensators and FFF beams

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Math Review. Overview

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

High Speed Direct SAD Radiosurgery Beam Scanner

Cover Page for Lab Report Group Portion. Pump Performance

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

Lab 1: Precision and Accuracy in Measurement (Using the right tool for the Job) and Density of Metals

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Thimble Ionization Chambers

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

E X R A D I N Spherical Ion Chambers A3. A4 REF DOC #

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

The dose distribution of medium energy electron boosts to the laryngectomy stoma

User Manual. RW3 Slab Phantom T29672 and T

Quality Management Determined from Risk Assessment

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections of your book.

ali-q TM Gravimetric Verification Procedure

IMRT in clinical practice at the UMC-Utrecht. Utrecht

Clinical Implementation of Volumetric Modulated Arc Therapy

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

General Accreditation Guidance. User checks and maintenance of laboratory balances

A beam-matching concept for medical linear accelerators

Perturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays*

Maintenance and general care of secondary standard and field instruments

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

Exposure System Selection

Sizing Pulsation Dampeners Is Critical to Effectiveness

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Performance of the Liquid Hydrogen Target

CALIBRATION PRODUCTS & PHANTOMS

OP CHECKLIST FOR 1D CONSOLIDATION LABORATORY TEST

EC214 EC215 - EC215R Bench Conductivity Meters

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING

Next Generation Quartz Pressure Gauges

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Measurements. Metric System

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at

Advanced Test Equipment Rentals ATEC (2832)

Technique Sheet 16. Using the Metric Ruler

USER MANUAL PROGRAM VERSION 1.0.1

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

. In an elevator accelerating upward (A) both the elevator accelerating upward (B) the first is equations are valid

Introductory Lab: Vacuum Methods

BekkTech s Procedures For Performing In Plane Membrane Conductivity Testing

Victoreen and

Variable Depth Bragg Peak Test Method

Standard Operating Procedure. Air Displacement Pipette Calibration

HASTAC High stability Altimeter SysTem for Air data Computers

CHM Introductory Laboratory Experiment (r17sd) 1/13

Density of Brass: Accuracy and Precision

Transcription:

Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas

Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) Only 8% to go 1200 1000 800 600 400 200 0 Nov-99 May-00 Nov-00 May-01 Nov-01 May-02 Nov-02 May-03 Nov-03 May-04 Nov-04 May-05 Nov-05 May-06 Nov-06 May-07 Nov-07 May-08 Nov-08 May-09

What s s the Holdup? New Air Kerma standard TG-51 TG-21 depending on the chamber Time and effort required (everyone is very busy) New equipment requirements (chambers and phantoms)

Equipment Needs Properly sized liquid water phantom (30x30x30 cm 3 ) Don t t use the scanning tank Adequate scatter conditions Easy reproducible setup

Chamber Holder and Positioner Holder Versatile to hold different chambers Rigid (sensitive volume perpendicular to water surface) No lateral displacement with depth Accurate sub-millimeter placement at any depth Verify accuracy prior to initial use Remote electronic control is nice

Ion Chambers TG-51 ion chambers vs NEW ion chambers Most are similar in design but now waterproof 1. Wall material 2. Radius of air cavity 3. Presence of Al electrode 4. Wall thickness AAPM working group to determine the k Q, k R50 ecal for new chambers k ecal 50,

Ion Chambers - Photons ADCL calibrated 0.6 cm 3 Smaller volume chambers (> 0.1cm 3 ) okay if traceable to another 0.6 cm 3 NO parallel plate chambers Waterproof (Go ahead and get one) Most common: Exradin A12, PTW 30013 Non waterproof needs a 1mm PMMA sleeve that does not leak!

Ion Chambers - Electrons Parallel-plate or cylindrical chambers okay Cylindrical for energies > 6 MeV per protocol (R 50 2.6 cm) Cylindrical = parallel plate if care in placement Always use a parallel plate chamber for 4 MeV beams Caution as to where the inside surface of the front window is located

Ion Chambers - Electrons All chambers must have an ADCL calibration coefficient coefficient EXCEPT PARALLEL PLATE CHAMBERS AAPM recommendation is to cross calibrate parallel plate chamber with cylindrical chamber in a high energy electron beam (worksheet C a la TG-39) ADCL N D,w good TG-51 k ecal bad Use of (N( D,w k ecal ) results in an error of 1-2% 1 ONE EXCEPTION Exradin P11 seems to be okay AAPM working group determining new k ecal values

Measurement Techniques Accurate placement of cylindrical ion chamber at depth Whether manual or electronic motor driven there must be a starting reference point Two techniques 1. Surface method Air Correct Position Water

Measurement Techniques 2. Cowboy method Cut ruler down to minimize surface area U-shape Water plastic surface attached flush with end of ruler Accuracy depends on cutting ruler Used for reference starting point Periodic check of depth Ion chamber Cut ruler by the chamber radius and wall thickness weights

Measurement Techniques Parallel plate ion chambers 1. Flat surface makes it easy to measure depth 2. Accurate ruler needed 3. Must know where the inside surface of the front window is located

Effective Point of Measurement and Beam Quality Photons 10 cm calibration depth Electrons calibration depth d ref point of measurement is the center electrode of a cylindrical chamber and the front window of a parallel plate chamber %dd(10) x beam quality R 50 Beam quality should always be measured using the effective point of measurement 0.6r cav 100 cm beam quality SSD 10 x 10 cm 2 field size 0.5r cav 100 cm 10 x 10 cm 2 shift to effective point 0.5r cav

Effective Point of Measurement Water surface Effective depth Physical depth X X Cylindrical Parallel plate

Get the lead out Photon beams ( ( 10 MV) Lead sheet 1 mm ± 0.2 mm 30 or 50 cm from phantom surface Determine %dd(10) Pb (percent values not fractional) %dd(10) x should be within 2.5% of %dd(10) Pb Interim alternative (No Lead Sheet) Measure %dd(10) without lead and use TG-51 eq 15 Introduces only 0.1-0.2% 0.2% error in k Q Saves time and minimizes chance of damage to chamber

Beam Quality Conversion Factors Photons k Q My favorite part Tabular values much easier to read Figures have a great deal of overlay For spreadsheets plot the tabular data and derive empirical fit for specific chamber Be sure to have an independent check of the empirical fit function

Beam Quality Conversion Factors Electrons k R 50 Only small figures, no tables Good figures at: http://www.physics.carleton.ca/~drogers/pubs/pa pers/tg51_figures.pdf

Beam Quality Conversion Factors Electrons 4 MeV beams (R 50 < 2.0 cm) Only use parallel plate chamber Need to extrapolate curve Equation good down to 1 cm 1

P TP Charge Measurements M = P ion P TP P elec P pol M raw TP correction factor Mercury thermometers and barometers most accurate (but they are no longer kosher) Hg barometers T&G corrections needed Quality aneroid or digital can be used Check annually against a standard Digital purchased with a calibration does not mean accurate but rather what it read at certain pressures or temperatures

P elec P pol Charge Measurements elec correction factor ADCL calibration for each scale needed pol correction factor Change polarity requires irradiation (600 to 800 cgy) ) to re-equilibrate equilibrate chamber Use of eq 9 in TG-51 requires that you preserve the sign of the reading or pol should be near unity for cylindrical chambers and slightly larger correction for parallel plate chambers P pol P pol = M + raw 2 + M M raw raw

Charge Measurements P ion 1.015 ion correction --- Monitor factor Response norm to 1st rdg (-300V) for a beam 1.010 Δ NEL'' fluctuation High dose rate capabilities result in higher P X PTW'' (All other chamber data are norm ion to monitor) Change 1.005 in bias requires irradiation (600-800 cgy) ) to re- equilibrate chamber. 1.000 0.995 0.990 0.985 0.980 0.975 A-12' 6 x @ d5 @ d10 PTW'' A-12 NEL'' 18 x 6 x @ d10 Cl 2100CD (210 5) data (75 mu) dated 19-Feb-00 Monitor's drift due to Ktp & machine 0 20 40 60 80 100 120 140 160 180 18 x @ d5 Sequence # 6 e @ dm 20 e @ dm

Charge Measurements P ion ion correction factor Use eqs.. 11 and 12 to calculate P ion As a check if using V H /V L = 2 (within 0.1%) P ion Pulsed beam : P ion = M H /M L if M H /M L < 1.02 Continuous beam : P ion = {(M H /M L 1)/3}+1 ion depends on chamber, beam energy, linac and beam modality Tends to increase with energy

Charge Measurements Electron beam gradient (P( gr ) correction factor No correction for photon beams since correction included in k Q Only for cylindrical ion chambers Ratio of readings at two depths ( d 0.5r ) ( d ) The reading at d ref +0.5r cav should have the same precision as the reading at d ref since: Dose = M(d ref ) (many factors) M(d ref +0.5r cav M(d P gr M ref + = M raw ref cav cav ) M(d ref )

Charge Measurements Electron beam gradient (P( gr ) correction factor E < 12 MeV; P gr >1.000 E 12 MeV; P gr 1.000 Why? Because for low electron energies d ref = d max and this places the eff.. pt. of measurement in the buildup region thus a ratio of readings greater than 1.000. At higher electron energies d ref is greater than d max and as such the eff.. Pt. of measurement is on the descending portion of the depth dose curve thus a ratio of readings less than 1.000.

Charge Measurements 100 80 Physical depth Percent Dose 60 40 100 x x 6 MeV Dose 20 0 80 0 2 4 6 60 Depth (cm) Effective depth ( d 0.5r ) M ref + M raw ( d ) ref cav

Charge Measurements 120 100 120 Physical depth 80 Percent Dose 60 100 40 20 Effective depth 80 0 0 5 10 15 Depth (cm) x d ref x d ref+0.5rcav ( d 0.5r ) M ref + M raw ( d ) ref cav

Clinical Depth Dose Always measured using the effective point of measurement Re-measurement not suggested for existing Linacs New Linacs or beams should incorporate shift Always use the clinical depth dose to make the correction from the calibration depth to the reference depth Measurement at depth will always equal calculation at the same depth (use same data to go to d max as is used to go back down to reference depth)

Clinical Depth Dose %dd(10) x = 67.0% Annual QA %dd 10 = 67.4% TPS %dd 10 = 66.6% (mu calc) 1.000 cgy/mu d max 0.666 cgy/mu 10 cm Joe patient

Clinical Depth Dose For photons do not use the beam quality value %dd(10) x to take dose from 10 cm to d max For electrons depth dose correction for 16 MeV is significant (~98.5% - 16 MeV and ~95.5% - 20 MeV) Caution!!! Super big problem if you use % depth ionization data (3-5% error for high energy electron beams)

Summary Implementation is straightforward Must read the protocol and follow the prescriptive steps Many suggestions to clarify confusion have been made RPC will assist you and answer questions Differences between TG-51 and other protocols such as TG-21 and TRS 398 are minimal.