Recommendations for the minimal monitoring of the patient during anaesthesia.

Similar documents
Guidelines on Monitoring in Anaesthesia

Recommendations on Checking Anaesthesia Delivery Systems

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Monitoring, Ventilation & Capnography

Mechanical Ventilation

ekom Compressor for Medical Air Sonata with Ekom

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Anesthesia monitoring

Recommendations for Standards of Monitoring and Alarms During Cardiopulmonary Bypass

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

Anaesthetic infrastructure and supplies

Comparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia

AQAI SIS MODULES AND MODELS PAGE 1. AQAI Simulator Interface Software AQAI SIS

Operating Instructions for Microprocessor Controlled Ventilators

Recommended Minimum Facilities for Safe Anaesthetic Practice in Operating Suites

RESPIRATORY PHYSIOLOGY, PHYSICS AND

Nitrous Oxide Sedation

Selecting and Connecting Breathing Systems

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

MEDUMAT Standard a MEDUMAT Standard. The Ventilation Standard for Professionals

Let s talk about Capnography

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

Keeping anaesthetic pollution to minimum in theatre environment

Oxylog 3000 plus Emergency & Transport Ventilation

Breathing Circuits. Product training

Apollo Safety Features

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

VENTILATORS PURPOSE OBJECTIVES

fabian HFO: Modular, Powerful, Versatile.

A V A L A N C H E R E S C U E

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

Section 14: Equipment failure

ROUTINE PREOXYGENATION

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Automatic Transport Ventilator

First Class. Monitoring. solutions for. Range of monitors CO 2 O 2 SpO 2 ECG NIBP. More Choices to meet your needs

A simple method of partial inflation of the LMA cuff before insertion in children to allow cuff pressure without adjustment after insertion

Multicentre Study of Preoxygenation Practices by Anaesthesia Providers

Page: 1 of 6 Responsible faculty: (Signature/Date)

Mechanical Ventilation

Draeger Fabius GS Pre-Use Check

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

4/2/2017. Sophisticated Modes of Mechanical Ventilation - When and How to Use Them. Case Study 1. Case Study 1. ph 7.17 PCO 2 55 PO 2 62 HCO 3

Hospital and Transport for Controlled Breathing

Medical Instruments in the Developing World

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

COPYRIGHT Ramsey Medical, Inc.

Historically, rebreathing anesthesia systems

Difficult Oxygenation Distribution: Sydney X Illawarra X Orange X

Emergency Transport and Ventilation

Day-to-day management of Tracheostomies & Laryngectomies

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

Respiratory Signs: Tachypnea (RR>30/min), Desaturation, Shallow breathing, Use of accessory muscles Breathing sound: Wheezing, Rhonchi, Crepitation.

Electric Blower Based Portable Emergency Ventilator

Recommendations. For Standards of. Monitoring and Safety during. Cardiopulmonary Bypass

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS

excellence in care Procedure Management of patients with difficult oxygenation. For Review Aug 2015

Capnography. The Other Vital Sign. 3 rd Edition J. D Urbano, RCP, CRT Capnography The Other Vital Sign 3 rd Edition

Endotracheal Suctioning: In Line ETT

Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future

Practical considerations in anaesthetising exotic species. By Keith Simpson BVSc MRCVS AMIIE(Electronics).

o An experienced and trained assistant is available to help you with induction.

Effect of Fresh Gas Flow on Isoflurane Concentrations during Low-flow Anaesthesia

Minimum size for maximum performance

Health Professional Info

Made to meet your needs DRÄGER SAVINA 300

Section 2.9 Decannulation

Notes on BIPAP/CPAP. M.Berry Emergency physician St Vincent s Hospital, Sydney

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

D Beyond essentials DRÄGER SAVINA 300 SELECT

D Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS

Monnal T60. Touch and Breathe.

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Patient Information for the: Humanitarian Device for use in the Control of Air Leaks

Technical Data and Specifications

C O R P O R AT E B R O C H U R E

Vital Signs Monitoring

AUTOVENT 4000 VENTILATOR

Update on Thoracic Anesthesia: Perioperative Lung Protective Strategies in One- and Two-Lung Ventilation

Astral in AirView: Improving patient care through connectivity. ResMed.com

QED-100 Clinical Brief

Selecting the Ventilator and the Mode. Chapter 6

Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS

HOBIT Trial Checklist

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

Cardiac Arrest General

Flight Medical presents the F60

II. Set up the monitor

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

Astral in AirView: Improving patient care through connectivity. ResMed.com

Transcription:

Guidelines for Safety in Veterinary Anaesthesia: Enclosure 2 Recommendations for the minimal monitoring of the patient during anaesthesia. Members of the ISVRA Task Force on Guidelines for Safety in Veterinary Anaesthesia: Alessandro Bonioli, Federico Corletto, Paolo Franci, Cristiano Gualtierotti, Lorenzo Novello, Alberto Perini, Nicola Ronchetti, Eriberta Ros, Cristina Spinnato, Enrico Stefanelli, Flavio Valerio Wishing to conform to European directives issued at various levels, ISVRA has decided to publish the following set of rules, as other Scientific Societies have already done. ISVRA also hopes that this first publication of its suggestions may be of some use to Colleagues, with the aim of attracting the attention of bodies and people in charge of the technological and organizational updating of both public and private veterinary practices. The monitoring of vital functions is an important part of anaesthesia, and its improvement will allow enhancement of safety in anaesthesia, bettering the quality of assistance we give our patients. An appropriate monitoring of the patient may permit, under certain clinical conditions, to promptly detect alterations in vital functions which may be caused by side-effects of administered drugs, by faulty functioning of medical instruments, or by physiopathological alterations related to surgery. In human medicine, minimal monitoring standards during anaesthesia, which were proposed by the Anaesthesia Department of Harvard Medical School (1) in 1986 have been widely accepted and applied in the USA and have been adopted by the American Society of Anaesthesiologists (2). Similar standards have been proposed in various European Countries and in South Africa (3-7). This document, created by ISVRA Workgroup on Safety in Veterinary Anaesthesia after a similar SIAARTI document (SIAARTI stands for Italian Society of Anaesthesia, Analgesia, Resuscitation and Intensive Care), is meant to be a recommendation. It may be subject to further revision to update it to the clinical and technological progress in our field, while we hope it will be inserted in future suggestions from other societies or bodies. Using a minimal instrumental monitoring system does not exempt the physician who carries out anaesthesia (since now on we will call him/her anaesthetist ) from a constant and scrupulous clinical observation of the 1

patient. According to the anaesthetist in charge of the procedure, the monitoring levels recommended here will have to be adapted to each case or increased under certain peculiar pathological conditions, as is the case in cardiac and neuro surgery. The definition of monitoring systems suggested in the present document has taken due account of simplicity of use, reliability, precision and noninvasiveness. All the suggested equipment are not to be seen as simple alarm systems, but as instruments which can provide an essential monitoring system for patient s physiological and functional parameters and their modifications over the time. In the following text, monitoring will be divided into constant and periodical. By the word constant, we mean prolonged and uninterrupted. This feature is typical of some monitoring techniques such as ECG, body temperature, ventilation, end tidal CO 2, oxygen saturation. The term periodical means regularly repeated at short time intervals. This term refers to observation of the patient, recording of instrumental parameters or to special monitoring techniques which perform automatic surveys at fixed times (i.e. non-invasive arterial blood pressure monitoring. As for available equipment, material or device, we mean an instrument not physically present in the theatre, but quickly available if needed. Care of the patient during general anaesthesia Constant assistance during anaesthetic and sedation techniques General anaesthesia should be carried out by veterinary surgeon who has skills and knowledge in anaesthesia: his constant presence is necessary to provide proper anaesthesia and to monitor patient s vital functions. The same rule is applicable to regional anaesthesia and to sedation techniques which may imply the risk of losing consciousness or respiratory/cardiovascular complications. Needs for distant control and/or monitoring When a temporary physical risk for staff is present (x-rays, radiotherapy, CT, MRI), the direct control of the patient may be replaced, for a short time span, by an appropriate remote control, maintaining observation of the patient and of monitored parameters. It will be the anaesthetist s task to decide if this alternative is safe for the patient. Taking over from an anaesthetist The same anaesthetist who starts an activity on a patient should also conclude it. If this is not possible, the in-coming anaesthetist should receive adequate information on how anaesthesia is proceeding and on the settings of all instruments in use. 2

The time of substitution and all functional parameters of the patient at that moment should be recorded in the anaesthetic record. Pre-anaesthetic check of drugs and equipment We would like to stress that a correct functioning of the instruments to be used and the safety of connections with gas sources should be thoroughly checked before anaesthesia starts. The availability of both ordinary and emergency drugs and technical devices should also be controlled. The use of an anaesthetic gas analyser is recommended if the anaesthetist is using a lowflow circular breathing system, and necessary if using a closed circular breathing system. For the anaesthesia machine check, you may refer to the ISVRA paper on the subject. Clinical and instrumental monitoring of respiratory function Oxygenation To make sure oxygen in the inspired gas mixture (FiO 2 ), blood and tissues of the patient is adequate the following approach is recommended: Evaluation of blood and peripheral tissue oxygenation Besides a periodic clinical evaluation of the patient, the use of a pulseoxymeter as a quantitative method to check oxygenation is recommended. Monitoring of oxygenation will have to be prolonged until dismissal from the operating theatre. Inspired oxygen concentration assessment It is highly desirable that each anaesthetic machine should have its own oxygen analyser to constantly measure the oxygen concentration in the inspired gas mixture. This device should have a visible and acoustic alarm, triggered by minimum and maximum concentrations. Ventilation Monitoring of ventilation is necessary at all times during anaesthesia. This monitoring should be done by the following methods: - during a short-term anaesthesia, under spontaneous or manual ventilation using a mask, the observation of the rib cage expansion/reduction range and volume changes in the respiratory bag, chest auscultation, along with the monitoring of colour of mucosal membranes, may be considered an adequate survey. If available, the use of a MicroStream capnograph is advisable. - During spontaneous or mechanical ventilation with tracheal intubation, ventilation should be monitored with the following systems, in addition to those reported above: a) a capnograph, to determine the level of end-of-expiration CO 2, is recommended if available; b) a spirometer, with sensor placed on the expiratory limb of the breathing system to assess gas volumes expired by the patient, is a 3

very effective monitoring to use in conjunction with the capnograph; the use of a spirometer is also recommended if available. These two instruments should be equipped with visible and acoustic alarms, triggered by minimum and maximum levels. Monitoring of patient s connection to the anaesthetic machine The mechanical ventilator, if available, should be provided with an acoustic alarm system based on insufflation pressure, capable of signalling a patient s disconnection when the pressure in the circuit goes under a pre-fixed value for more than 20 seconds. The spirometer, but above all the capnograph, also represent useful alarm systems in case of patient s disconnection. Clinical and instrumental monitoring of the cardiovascular system We recommend the periodic clinical monitoring of the cardiovascular activity during general anaesthesia is not suspended, and we suggest that continuous ECG and heart frequency monitoring should become a common practice in all patients. This monitoring will have to be prolonged until dismissal from the operating theatre. If a specific device is available, we suggest arterial pressure under general anaesthesia might be determined at intervals depending on the patient s clinical condition and, at any rate, not exceeding 10 minutes. Useful information on the haemodynamic situation may also be obtained from capnography (ETCO 2 ). Monitoring of body temperature The maintenance of an adequate body temperature has to be ensured throughout every anaesthetic, bearing in mind that prevention of hypothermia should start at premedication and should end at recovery of standing position or later. Even though the anaesthetist may make use of clinical signs to detect raising/falling temperature, a continuous measuring method for body temperature is recommended in every operating theatre. If the patient is at risk of malignant hyperthermia, or when the use of temperature modifying techniques has been decided (i.e. hypothermia), body temperature should be constantly monitored. Here too, capnography may provide useful information. Monitoring of neuromuscular functions When neuromuscular blocking agents (NMBA) are used, a neuromuscular function monitoring device should be available to monitor recovery of the neuromuscular junction. If a neuromuscular function monitoring device is not available, an indirect monitoring of recovery (i.e. a capnograph, or a 4

spirometer displaying airways pressure trace and tidal volume) can be used instead. Recording monitored parameters All instruments should be provided with a recording system of the monitored parameters which also allows a review of what has happened immediately before and after the event that triggered the alarm. All parameters subject to clinical and instrumental observation should be inserted in the anaesthetic record. Special procedures may be needed, which will require an interruption of instrumental monitoring: the anaesthetist s task will be to specify the time and reasons of this interruption in the record. When all this occurs, clinical monitoring should be intensified. ISVRA Task Force for Guidelines on Safety in Veterinary Anaesthesia References 1. Eichhorn JH et al. Standards for patient monitoring during anesthesia at Harvard Medical School. JAMA, 1986;256(8):1017-20. 2. American Society of Anesthesiologists. Standards for basic intra-operative monitoring. ASA Newsletter 1986;50(12):12. 3. European Academy of Anaesthesia. International Task Force on Safety in Anaesthesia. Eur J Anaesth, 1993;10:1-44. 4. Belgian Standards for safety in anaesthesia. Acta Anaesth Belg 1989;40:231-8. 5. Canadian Anesthetist Society. Guidelines to the practice of anesthesia. 1993. 6. Guidelines for anaesthesia care in Nordic countries. Acta Anaesthesiol Scand, 1992;36:741-4. 7. South Africa Society of Anaethetists. Guidelines for practice. 1990. 8. ACVA American College of Veterinary Anaesthesiologists. Anaesthesiology guidelines developed. JAVMA 1995; 206(7):936-937. Last updating: August 2002 5