Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Equinox II NITROUS OXIDE / OXYGEN ADMINISTRATION EQUIPMENT 01EQ7000E

2 PRE USE FUNCTION CHECK 3 OPERATING PROCEDURE 4 SERVICING. 5 CAREvent ALS ACCESSORIES 11

O-Two Equinox Relieve Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System

USER MANUAL (For G2005 Model)

45A, Armthorpe Road, Brampton, Ontario, Canada, L6T 5M4. Made in Canada by O-Two Medical Technologies Inc.

USER MANUAL AUTOMATIC AND MANUALLY TRIGGERED RESUSCITATOR. O-TWO MEDICAL TECHNOLOGIES INC. Innovation in Resuscitation

O-Two Self-Study Guide. Equinox Relieve

Equinox Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System USER MANUAL

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

Pressure Regulators. Instructions for Use

Yoke Block Instruction Manual

AUTOVENT 4000 VENTILATOR

Blendox AII5000F Inhalation Antalgic System. User s Manual :SP-Y Series No. Version :A1. Date of Review:

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

CPR MEDICAL DEVICES INC. OPERATING MANUAL

NITRONOX USER S MANUAL

Rev. B. Operating Instructions

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

Dialflow Regulator. Instructions for Use

SERIES-O REGULATOR. Instructions for use

Yoke Block Instruction Manual

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Pressure Regulator. Instructions for Use

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

RG2200 BC Integrated Octopus

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

Oxygen Demand Valve. Instructions for Use

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

RG3100 and RG3100Ice Regulator System

User Operation Manual and Safety Instructions

PORTER INSTRUMENT COMPANY, INC.

NITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS

Paradise Pump Series Operator s Manual

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

CSA Sample Draw Aspirator Adapter Operator s Manual

Oxygen Dialflow Meter. Instructions for Use

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

NeoNatalie Resuscitator Oxygen kit

OXY-life TM. Oxygen Resuscitator OPERATING MANUAL

U S E R M A N U A L. (CGA 870 Connection)

MEDICAL BREATHING CIRCUIT SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

Model ASSEMBLY and OPERATING INSTRUCTIONS

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

NeoNatalie Resuscitator Oxygen kit

Data Sheet... RUDOLPH, inc. HANS RUDOLPH,

AND OPERATION INSTRUCTIONS

Mirage Vista Nasal Mask

RK-IR Sample Draw Aspirator Adapter Operator s Manual

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

Analgesic Demand Valve. Instructions for Use

M5400 Operators Manual

Service and Repair Manual

MACS Mask CPAP System

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

English. ivolve Full Face Mask. User Manual

Invacare Twilight II Nasal Mask

User Instruction Manual

oxyacetylene welding/cutting torch set

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

User manual. ZX trolley.

Peninsula Dental Social Enterprise (PDSE)

TactAir PCP Fill Kit/Adapter. Operation Manual

C-1 C-7. C-6 Short air tubing Upper Velcro straps. C-a C-5 C-2 C-4 C-3 B-1 B-9 B-10 B-2 B-7 B-8 B-3 B-6 D-1 D-2 D-13 D-3 D-12 D-10 D-11 D-4 D-5 D-8

Frigitronics CRYO-PLUS

OXI-dive 3 OXYGEN RESUSCITATION UNIT OPERATING MANUAL

SCUBATECH. Instructions for use. Demand regulators. Series models R1, R2, R3,R5, V2 (ICE, PRO, TEC, SPECIAL, OCTO, MONO)

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution:

RK LEL Sample Draw Aspirator Adapter Operator s Manual

MOBILE E TANK STAND INSTALLATION AND INSTRUCTIONS

RG1200 Regulator System

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

4-in-1 Professional Inflator Kit

Flow meter. bellow vaporizer. APL valve. Scavenging system

- GAS REGULATOR - COMPACT STYLE

AUTOMATIC TRANSPORT VENTILATORS

Operation and Maintenance of the EPV200 Portable Ventilator

Binks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP

Freedom8 ShoeBox Compressor Manual

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

SIX R (P ) AND SEVEN R (P ) POSITION CYLINDERS Service Information

Micro Dial-Flowmeter. Instructions for Use

North RU8500 Series Half Masks

PRODUCT INFORMATION AND INSTRUCTIONS OM-810

RAM Operation Manual

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Operating Manual. Model MP-4R and MP-4RSS MULTI-PAK TM RESCUE AIR CARTS Manual No. PAK013 (Rev 1 January 2001)

Instruction Manual Updated 7/26/2011 Ver. 2.2

WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

RAM 4021 Operation Manual

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

Transcription:

TM controlled ventilation Equinox II NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E USER TABLE OF CONTENTS CHAPTER TITLE PAGE 1 INTRODUCTION 1.1 Equinox II Administration System 2 1.2 Warranty Information 2 1.3 Safety Precautions 3 1.4 Performance Specification 4 2 OPERATING PROCEDURE 2.1 Demand Breathing 5 2.2 Operating Instructions 5 2.3 Scavenging of Waste Nitrous Oxide/ 6 Oxygen Using the Scavenger Adapter USER MANUAL 3 SERVICING 3.1 Routine Maintenance 7 3.2 Cleaning and Disinfection 8 4 REPLACEMENT PARTS/ ACCESSORIES 10 Made in Canada by O-Two Medical Technologies Inc. Part Number: 17MP1342 Rev 9 Apr. 2015 1

1. INTRODUCTION 1.1 Equinox II ADMINISTRATION SYSTEM The Equinox II Administration System is intended for administration of 50%/50% N 2 O/O 2 analgesic gas mixture and is specifically designed for use in emergency and obstetric analgesia. The Equinox II Administration System consists of a demand valve, a single stage pressure regulator, a supply hose and a universal patient face mask. It can be operated anywhere 50% Nitrous Oxide/50% Oxygen mixture cylinders are available. The Equinox II Demand Valve is lightweight, portable, and extremely durable and is intended to administer 50%/50% N 2 O/O 2 analgesic gas on patient s demand. The pressure regulator is intended to lower the high, variable cylinder pressure to a lower constant working pressure for the Demand Valve. Note: Only those individuals trained in the administration of Nitrous Oxide/Oxygen mixtures should use this equipment. Thoroughly review the instruction manual before use. 1.2 WARRANTY Equinox II Administration System is manufactured from the finest quality materials. Each individual part is subject to strict quality control tests to ensure exceptionally high standards. The manufacturer warrants to the purchaser of Equinox II Administration System that its component parts are free from defects in material and workmanship for a period of two years from the date of purchase. The manufacturer will replace and /or repair all parts of the device at its option for two years from the date of purchase at no cost to the purchaser, upon the notification of the defects, in writing by the purchaser. All shipping costs shall be borne by the purchaser. The manufacturer shall be liable under this warranty only if the device and its parts have been used and serviced in the normal manner described in the instruction manual. There are no other expressed or implied warranties. This warranty gives no specific legal rights. You may also have other rights which may vary according to local regulations. 1.3 SAFETY PRECAUTIONS The Equinox II Administration System is intended for use by suitably trained and qualified personnel. The following precautions should always be observed: 1. CAUTION: FEDERAL LAW IN THE U.S.A. RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A LICENSED PRACTITIONER. 2. WHEN THE UNIT IS IN USE, DO NOT SMOKE OR USE NEAR OPEN FLAME EITHER DURING USE OR WHEN CHANGING THE CYLINDER. 3. WHEN NOT IN USE, ALWAYS TURN OFF THE CYLINDER. 4. NEVER ALLOW OIL OR GREASE TO COME INTO CONTACT WITH ANY PART OF THE CYLINDER, REGULATOR OR DEMAND VALVE. 5. DO NOT DISASSEMBLE ANY PART OF THE DEMAND VALVE EXCEPT WHERE DESCRIBED IN THIS MANUAL AS ANY UNAUTHORIZED DISASSEMBLY WILL INVALIDATE THE WARRANTY. 6. AFTER USE, ALWAYS ENSURE THAT ALL COMPONENTS ARE CLEANED IN ACCORDANCE WITH THE INSTRUCTIONS PROVIDED IN THIS MANUAL. 7. ALWAYS USE THE CHECK LIST TO ENSURE THAT ALL COMPONENTS ARE REASSEMBLED CORRECTLY AND READY FOR USE. 8. AFTER USE, ALWAYS ENSURE THAT A FULL GAS CYLINDER IS ATTACHED BEFORE RETURNING THE UNIT TO ITS NORMAL STORAGE POSITION. 2 3

1.4 PERFORMANCE SPECIFICATION Demand Valve (Minimum) Peak Flow rate...... 140 LPM Exhalation Resistance.... 0 to 6 cmh 2 O @ 60 LPM Inhalation Resistance.... 0 to -6 cmh 2 O @ 60 LPM Input Pressure..... 50 PSI Initiation Pressure..... -2 to -2.5 cmh 2 O Operating Temp.....-18 0 C to 50 0 C (0 0 F to 122 0 F) Storage Temp..-40 0 C to 60 0 C (-40 0 F to 140 0 F) Input Connection......... 5/8-18 TPI Patient Connector...... 15/22 mm Weight...210 grams (7.4 oz) Dimensions (inches)... Dia. 51mm x 81mm (2.0 x 3.2 ) Pressure Regulator Inlet Connection...CGA 965 N 2 O/O 2 Pin Indexed Yoke Outlet Connector...... 5/8 Thread Operating Pressure...500-3000 PSI Output Pressure.. 50-60 PSI Weight...450 grams (1.0 lb) 4 2. OPERATING PROCEDURE 2.1 DEMAND BREATHING The O-Two Medical Equinox II Demand Valve is designed to operate on compressed Nitrous Oxide/ Oxygen gas. It is intended to provide 50/50% Nitrous Oxide/ Oxygen mixture to the patient at a safe delivery pressure and adequate flow rate upon demand. An inspiratory effort by the patient will activate the demand valve and Nitrous Oxide/ Oxygen will flow to the patient at a rate and volume in line with their inspiratory effort. 2.2 OPERATING INSTRUCTIONS 1. Ensure that the Equinox II Demand Valve/Regulator connection is secured and no leaks present. 2. Turn the cylinder hand wheel counter clockwise to the ON position (at least one full turn). 3. Ensure that the regulator contents gauge reads at least 500 psi. If not, replace with a full cylinder. 4. Attach the Universal Facemask to the Equinox II Demand Valve outlet. 5. Instruct the patient in its operation. Most protocols require the gas mixture be self-administered by the patient to prevent inadvertent excessive administration by an attendant. 6. To trigger the demand valve and obtain an adequate flow of gas, it is necessary to make a good mask to face seal. Simply place the mask on the patient s face, covering both the nose and mouth. Instruct the patient to hold the mask on his/her face with a light pressure and breath normally. The sound of gas flowing should be heard with each inspiration. If not, adjust the patient s hold on the mask for a better seal and ask the patient to breath more deeply. 5

7. Continue administration as per local management protocol. 8. Ensure that the cylinder is turned OFF after each use and remaining gas released from the system by placing the bleed pin attached to the regulator into the center hole on top of the Equinox II Demand Valve and depress until gas flow ceases. 2.3 SCAVENGING OF WASTE NITROUS OXIDE/ OXYGEN USING THE SCAVENGER ADAPTER The optional Scavenger Adapter is designed to replace the Nonrebreathing Patient Valve assembly to scavenge exhaled air from the patient during inhalation analgesia. The 19 mm exhalation port meets the requirements of the International Standards for Conical Connectors for Low Pressure Medical Gas Systems and will allow the attachment of corrugated tubing so that the patient s exhaled breath can be scavenged from the exhalation valve to the exhaust method chosen by the facility using the device. 3. SERVICING 3.1 ROUTINE MAINTENANCE WARNING: The Equinox II Administration System is designed to provide years of reliable service in all emergency situations. Failure to follow the maintenance and inspection routines properly could result in incorrect operation of the demand valve. To ensure proper operation of the Equinox II Administration System, regular inspection and checking of the demand valve and accessories for correct function should be undertaken by a responsible member of staff on a regular basis. This check is to ensure that all of the accessories and demand valve components are present, the gas mixture cylinder is full and that the demand valve is in working order. MAINTENANCE Function parameters should be checked at least every six months, and more frequently in high use applications. Units with test pressures outside of the ranges listed in the product specifications should not be used. O-Two Medical products are not designed for field disassembly or service outside that indicated in this manual. Any malfunctioning units should be returned to the manufacturer or an Authorized O-Two Medical Technologies Dealer. Unauthorized repairs will nullify the product warranty. 6 7

3.2 CLEANING AND DISINFECTION Routine cleaning and/or disinfection of the O-Two Medical Equinox II Administration System should be undertaken after each patient use to maintain the equipment in a clean condition. D Demand Valve Body B Diaphragm A Patient Valve C Anti-entrainment Ring CLEANING/ DISINFECTION PROCEDURE 1. Ensure that the cylinder is turned off (completely clockwise) and bleed any remaining gas out of the gas supply hose. 2. Remove the facemask from the Equinox II Demand Valve, and then disconnect the demand valve from the regulator. The regulator can be cleaned with a clean soft cloth. 3. Unscrew patient valve (A) from the unit body being careful to ensure that the diaphragm (B) is retained (Fig1). 4. Remove the diaphragm and anti-entrainment ring (C) from the patient valve swivel housing and wash these components thoroughly in a mild soap solution; 5. Disinfect the face mask, patient valve and diaphragm using commonly available cold chemical disinfecting solutions. Rinse thoroughly with water after disinfection. 6. Sterilization, if required, may be achieved by other methods compatible with the materials of manufacture of these components including Ethylene Oxide Gas and Cidex. 7. The demand valve body (D) and gas supply hose can be wiped over with a soft cloth and mild soap solution and may be disinfected using a hard surface disinfectant that is compatible with aluminium. 8. Dry all components thoroughly. 9. Re-assemble and check performance. Fig.1 WARNING: Do not immerse the demand valve unit or the pressure regulator into cleaning solution or disinfectant. Discontinue to use and discard the face mask, the patient valve and the diaphragm should discoloration or deterioration be observed. 8 9

4. REPLACEMENT PARTS/ ACCESSORIES MANUFACTURERS AND DISTRIBUTORS OF: 01EQ7001E 01RE05S10-B 02FM4999-Cs 17MP1329 01EQ7006 17MP1528 17MP1320-Equinox 17MP9039 Equinox II Demand Valve Equinox Pressure Regulator Universal Face Mask (case/12)* Equinox II 7 Supply Hose Equinox II Demand Valve Bleed Pin with Chain Silicone Diaphragm Supply Fitting assembly Non-rebreathing Valve Assembly CPR Bio-Barrier Face Shields Universal Pocket Ventilators Rescue Breather CPR Protective Barriers Demand Valve Resuscitators CAREvent Automatic Transport Ventilators O-Two eseries Electronic Automatic Transport Ventilators O-Two Pressure Regulators for Use with Medical Gases SMART Bag Bag-Valve-Mask Resuscitators EasyGrip Bag-Valve-Mask Resuscitators Burn Relief Burn Kits and Dressings Equinox II Nitrous Oxide/ Oxygen Administration Systems Equinox Relieve and Equinox Advantage Nitrous Oxide/ Oxygen Gas Mixing and Delivery System Your Representative is: * Other sizes and seal types are available. Contact your nearest O-Two Medical authorized distributor for a complete catalogue on all O-Two Medical manufactured products. O-TWO MEDICAL TECHNOLOGIES INC. Innovation in Resuscitation EU Representative: Marcel Houben Rue Vinâve, 32 4030 LIEGE Belgique 7575 Kimbel Street Mississauga, Ontario Canada L5S 1C8 Tel: (905) 677-9410 Fax: (905) 677-2035 Website: www.otwo.com E-Mail: resuscitation@otwo.com 10 For your nearest Authorized O-Two Medical Tech. Distributor In North America Toll Free 1-800-387-3405