A new method for noninvasive venous blood oxygen detection

Similar documents
Biomedical Sensor Systems. Laboratory

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

Instantaneous venous oxygenation estimation using the Photoplethysmograph (PPG) waveform

City, University of London Institutional Repository

KNOW YOUR EQUIPMENT: Pulse oximeter Verghese Cherian Associate Professor of Anesthesiology Hershey Medical Center

Oxygen Saturation Monitors & Pulse Oximetry

The Portable Gas Analyzer Based on the Spectrum

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

Measuring Sound Speed in Gas Mixtures Using a Photoacoustic Generator

RESPIRATORY PHYSIOLOGY, PHYSICS AND

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

sensors ISSN

A Coal Mine Multi-Point Fiber Ethylene Gas Concentration Sensor

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light

Section 2: Using the Invasive Blood Pressure Module - An Introduction

Linear Infl ation Technology Non-invasive Blood Pressure Measurement. inibp Case Report Experiences in OR

International Journal of Scientific & Engineering Research Volume 9, Issue 4, April ISSN

u = Open Access Reliability Analysis and Optimization of the Ship Ballast Water System Tang Ming 1, Zhu Fa-xin 2,* and Li Yu-le 2 " ) x # m," > 0

Bloodpressure measurement, improvement of oscillometric method.

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

Alphasense Application Note AAN NDIR: Gas Concentration Calculation Overview

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

Critical Gust Pressures on Tall Building Frames-Review of Codal Provisions

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math!

Smart-Walk: An Intelligent Physiological Monitoring System for Smart Families

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

Non-contact detection of oxygen saturation based on visible light imaging device using ambient light

Solutions for transcutaneous monitoring

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

The Physiologic Basis of DLCO testing. Brian Graham Division of Respirology, Critical Care and Sleep Medicine University of Saskatchewan

Proceedings of Meetings on Acoustics

Constant pressure air charging cascade control system based on fuzzy PID controller. Jun Zhao & Zheng Zhang

Instructions to User

Journal of Chemical and Pharmaceutical Research, 2014, 6(3): Research Article

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

Improved Algorithm for Measurement of Blood Pressure based on a Laser Doppler Flowmetry Signal

Measuring Venous Oxygen Saturation Using the Photoplethysmograph Waveform

Laser Spectrometers for Online Moisture Measurement in Natural Gas. Ken Soleyn GE M&C

PULSE oximetry is frequently used in clinical situations

OxyMon MkIII Optical Imaging made easy and affordable

67. Sectional normalization and recognization on the PV-Diagram of reciprocating compressor

Instructions to User

Acoustic Emission Testing of The Shell 0f Electromagnetic Valve

Research of Variable Volume and Gas Injection DC Inverter Air Conditioning Compressor

FLOW CONSIDERATIONS IN INDUSTRIAL SILENCER DESIGN

SPIRIT III Radiometer Saturation Effect

Blood Parameter Monitoring System 550

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

MATRIX-MG Series. Innovation with Integrity. Automated High-Performance Gas Analyzers FT-IR

Basketball field goal percentage prediction model research and application based on BP neural network

Implementation of Height Measurement System Based on Pressure Sensor BMP085

Identification of the dynamic component of the finger arterial pressure-volume relationship

SenTec OxiVenT Illuminate Ventilation and Oxygenation PCO2 PO2. Digital Transcutaneous Blood Gas Monitoring

as provided by other available near infrared oximeters. Moreover, OxiplexTS complements the pulse oximeter in

Wave phenomena in a ripple tank

Analog Front End Design of a Digital Blood Pressure Meter

Development of a Tissue Oxygen Saturation Detection System for Improving Surgical Training


TABLE OF CONTENTS 1 Overview Appearance Name and Model Intended Use Feature List Important Acroynms...

Level MEASUREMENT 1/2016

HCL MONITORING. ICAC MARAMA WORKSHOP May 18-19th, Presented by: Ty Smith. CEMTEK

Transactions on Ecology and the Environment vol 12, 1996 WIT Press, ISSN

MEASURING BLOOD PERFUSION IN BONE USING NIRS BONE OPTICAL SPECTROSCOPY. A Thesis Presented. Ali Ibrahim Alneami. The Department of

Lecture 8. Sound Waves Superposition and Standing Waves

RESPIRATORY MONITORING AND OXIMETRY

Development of a Non-Destructive Test Machine for Testing Tunnel Lining Concrete

Handlebar Bi-Directional Portable Pump on Vehicle Yingying Tang1, Yang Jiang1, Yugang Xin1 and Baida Yang1

CALIBRATION OF THE PLATOON DISPERSION MODEL BY CONSIDERING THE IMPACT OF THE PERCENTAGE OF BUSES AT SIGNALIZED INTERSECTIONS

Online DGA-monitoring of power transformers

City, University of London Institutional Repository

Folding Reticulated Shell Structure Wind Pressure Coefficient Prediction Research based on RBF Neural Network

Sensor. Directions For Use

Basic - Peripheral Vascular Assessment VASCULAR DOPPLER PHOTO-PLETHYSMOGRAPHY PULSE VOLUME RECORDING ANKLE AND TOE BRACHIAL INDEX

PROPAGATION OF LONG-PERIOD WAVES INTO AN ESTUARY THROUGH A NARROW INLET

Coupling and Analysis of 981 Deep Water Semi-submersible. Drilling Platform and the Mooring System

Cardiac Output Simulation for Specific Makes of Monitor. Each injection yields in a time-temperature curve whose area represents the cardiac output:

Resonance in Transverse Waves. Regents Physics

Tokyo: Simulating Hyperpath-Based Vehicle Navigations and its Impact on Travel Time Reliability

Vibration Analysis and Test of Backup Roll in Temper Mill

BLOOD PRESSURE SENSOR BT17i USER S GUIDE

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

Available online at ScienceDirect. Energy Procedia 53 (2014 )

Entry to the premium class!

DEPARTMENT OF THE NAVY DIVISION NEWPORT OFFICE OF COUNSEL PHONE: FAX: DSN:

PVDOMICS: Cardiopulmonary Exercise Testing (CPET) Training

RESPIRATORY REGULATION DURING EXERCISE

IMPC phenotyping SOPs in JMC

Heart Rate Estimation Using Wrist-acquired Photoplethysmography Under Different Types of Daily Life Motion Artifact

Novel empirical correlations for estimation of bubble point pressure, saturated viscosity and gas solubility of crude oils

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

Homeostasis and Negative Feedback Concepts and Breathing Experiments 1

Engineering World Health Youth EWH Design Projects That Matter

Analysis of Shear Lag in Steel Angle Connectors

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Determination of the Design Load for Structural Safety Assessment against Gas Explosion in Offshore Topside

Estimation and Analysis of Fish Catches by Category Based on Multidimensional Time Series Database on Sea Fishery in Greece

Available online at ScienceDirect. Procedia Engineering 84 (2014 )

Wave Motion. interference destructive interferecne constructive interference in phase. out of phase standing wave antinodes resonant frequencies

FLUORESCENCE DETERMINATION OF OXYGEN

Video Based Accurate Step Counting for Treadmills

Transcription:

DOI 10.1186/s12938-016-0208-8 BioMedical Engineering OnLine RESEARCH Open Access A new method for noninvasive venous blood oxygen detection Xu Zhang 1,2,3, Meimei Zhang 1,2,3, Shengkun Zheng 1,2,3, Liqi Wang 1,2,3 and Jilun Ye 1,2,3* *Correspondence: yejilun@126.com 1 School of Medicine, Shenzhen University, Shenzhen, China Full list of author information is available at the end of the article Abstract Background: Blood oxygen saturation of vein (SvO2) is an important clinical parameter for patient monitoring. However, the existing clinical methods are invasive, expensive, which are also painful for patients. Methods: Based on light-absorption, this study describes a new noninvasive SvO2 measurement method by using external stimulation signal to generate cyclical fluctuation signal in the vein, which overcomes the low signal-to-noise ratio problem in the measurement process. In this way, the value of SvO2 can be obtained continuously in real time. Results: The experimental results demonstrate that the method can successfully measure venous oxygen saturation by artificial addition of stimulation. Under hypoxic conditions, the system can reflect the overall decline of venous oxygen saturation better. When the results measured by the new method are compared with those measured by the invasive method, the root mean square error of the difference is 5.31 and the correlation coefficient of the difference is 0.72. The new method can be used to measure SvO2 and evaluate body oxygen consumption, and its accuracy needs improvement. Conclusions: Real-time and continuous monitoring can be achieved by replacing invasive method with noninvasive method, which provides more comprehensive clinical information in a timely manner and better meet the needs of clinical treatment. However, the accuracy of the new noninvasive SvO2 measurement based on lightabsorption has to be further improved. Keywords: Noninvasive method, SvO2, External stimulation signal Background SvO2 and SaO2 (blood oxygen saturation of artery) have been established as the basic parameters to evaluate oxygen delivery process and monitor patient s oxygen consumption. According to Fick Formula in Eq. (1)[1], the oxygen circulation in human body can be analyzed completely only by combining SvO2 and SaO2 together. Therefore, SvO2 has a significant meaning in evaluating the health of patients. SvO2 = SaO2 VO 2 (1.34 CO Hb) (1) On the other hand, there is no existing noninvasive or continuous device for measuring SvO2 in the market, even though light absorption method has already been widely 2016 The Author(s). This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Page 2 of 9 used to evaluate SaO2. Besides the high price and inconvenient, invasive method also has more risk in clinical SvO2 measurement. Due to the lack of fluctuation signal in vein, SvO2 can t be measured in the same way as SaO2 is measured, for which light absorption method is valid because of the obvious impulse in artery. To overcome this problem a new noninvasive light-absorption method for SvO2 measurement is proposed. In this method, we innovatively create vein impulse similar to artery impulse by artificially adding external stimulation signal during measurement, which enables the extraction of the stable venous signal from normal arterial pulse signal. Methods According to the Lambert Beer s law, the intensity of light through the medium has different degrees of attenuation which depending on the thickness and concentration of the solution. Because the transmission light intensity will decrease with the increasing of the concentration and the thickness of the solution, the absorbance of a single solution can be calculated by Eq. (2) ( ) Ii A = ln = a C L (2) It where A represents absorbance, the natural logarithm of the ratio of the incident light intensity to the transmitted light intensity; Ii represents the incident light intensity and It represents the intensity of the transmitted light, a represents the absorption coefficient, with different values for different solutions; C represents the solution concentration; L represents the optical path and is the distance that light travels through in the solution [2]. Be similar to mixed solution, the absorbance of blood can be calculated by optical absorption model for mixed solution as shown in Eq. (3), and there is no limit on the number of components in the solution. As we all know, the absorbance of the mixed solution satisfies the superposition theorem and is the superposition of all solution ingredients absorbance at different concentrations and different absorption coefficients. A = ln ( ) Ii = It a k C k L k=1 (3) Equation (4) is the mixed absorbency at 660 and 940 nm with oxygen free hemoglobin and oxygenated hemoglobin under static condition, where λ is the wavelength of the original light. ( ) Ii (λ) A(λ) = ln = a Hb (λ) ac Hb (λ) L + a HbO2 (λ) ac HbO2 (λ) L (4) I t (λ) In dynamic state, the intensity change of the transmission light is induced by the change of optical path due to congestion artery, as shown in Eq. (5) I(λ, t) = I 0 (λ)exp( (sβ 0 (λ) + (1 s)β r (λ))l(t)) (5)

Page 3 of 9 where I o and I represent the intensity of the incident light and the transmission intensity of the original light respectively. λ denotes the wavelength of the original light. t is the time. S is the oxygen saturation. β 0 and β γ are the absorption coefficients for the Hb and HbO2 in the solution. l(t) is the optical path from the light source to the photoelectric sensor [3]. Equation (5) can be easily reformulated as Eq. (6) d(ln I(λ, t)) = (sβ 0 (λ) + (1 s)β r (λ)) d(l(t)) (6) Then Eq. (7) can be derived from Eqs. (4), (5) and (6). S = dlni(λ R ) dlni(λ IR ) β r (λ R ) dlni(λ R) β r (λ IR ) (β 0 (λ IR ) β r (λ IR )) dlni(λ IR) (β 0 (λ R ) β r (λ R )) (7) Assuming that Eq. (8) holds. ( ) d(lni(λ, t)) I(t2, λ) = ln I(t 1, λ) (8) We can rewrite Eq. (7) to Eq. (9) to calculate oxygen saturation S = β r (λ R ) Rβ r (λ IR ) R(β 0 (λ IR ) β r (λ IR ))(β 0 (λ R ) β r (λ R )) (9) where R (molar extinction coefficient) [4] is defined below and all the other variables are constant. R = ln(i(t1, λ R)/I(t2, λ R )) ln(i(t1, λ IR )/I(t2, λ IR )) = dlni(λ R ) dlni(λ IR ) (10) According to Eqs. (7) and (9), the full parameters of the pulse oxygen saturation can be obtained by calculating the alternating current (AC) and direct current (DC) values of red light and infrared light of signals [5]. Usually, pulse fluctuation can lead venous oxygen signal overlap with the artery blood oxygen signal, which causes problem of low signal (SvO2)-to-noise (SaO2) ratio. Therefore, the extraction of the weak venous blood oxygen signal from artery blood oxygen signal is the key to non-invasive SvO2 measurement. Based on clinical studies, a new method has been proposed. It measures the periodic fluctuation vein signal, whereas a new system has been designed to verify the method. As Fig. 1 shows, the system consists of four main parts. Part.A is the monitor for data calculation and waveform display. Part.B is the system controller. This module can control the work of Part.A, Part.C and Part.D. Part.C is the photoplethysmography (PPG) signal sample module getting photoelectric volume pulse wave data. Part.D is the signal generation module. In this part, a ring-shaped inflatable air cuff (Part.D of Fig. 1) is used to produce stable signal as stimulate signal. By controlling Part.D, Part.C can gain the arterial PPG signal and the mixed PPG signal respectively, which is then displayed in Part.A.

Page 4 of 9 Fig. 1 Components of system Periodic pressure can be added to finger by inflating and deflating the air cuff periodically. NIBP (non-invasive blood pressure) is used as a controllable parameter to set the threshold pressure for different people, while the frequency of inflation and deflation are controlled by a high-precision timer. Figure 2 is the control chart for the venous signal enhancement system. The stimulation control system is the module that controls the venous signal enhancement. The feedback adjustment is controlled by MCU. The detailed process is as follows: firstly, PPG of arteries and veins can be detected by dual-wavelength pulse oxygen measurement system; secondly, the signal frequency and strength of the two kinds of PPG are used as feedback to increase the signal of vein; thirdly, with feedback of the relationship between the signal-to-noise ratio of vein and arterial, the stimulation control system to stay in a stable and acceptable condition can be maintained (the signal frequency of the two kinds of PPG is different). Figure 3 briefly Fig. 2 Stimulation feedback system

Page 5 of 9 Fig. 3 The difference between normal and stimulate signal shows the difference between normal signal and stimulated signal. In the signals of IR and RED, every PPG contains multiple excitation signals. Besides, under the stimulation condition, both arterial signal and venous signal are affected. The SvO2 signal extraction process generally includes three steps. Firstly, the stimulated signal is wiped out and the normal SpO2 signal is calculated. Secondly, the signal which is too far from our stimulation frequency is filtered and the peak of artificially stimulated signal is used as input data to calculate Mixed-SpO2 in the normal way which is same to the calculation of SpO2. Thirdly, SvO2 is calculated. In this paper, venous PPG generated by the stimulation signal is used to calculate the venous oxygen saturation, which is the same as arterial oxygen saturation PPG used to calculate the arterial oxygen saturation. The PPG is recorded at 660, 940 nm for red and infrared transmission intensity changes. In theory, if PPG of the arterial blood signal can be used to calculate the arterial oxygen saturation, PPG of the venous blood signal can also be used to calculate the venous oxygen saturation. Because the features of the two signals are the same, SvO2 can be calculated use the same method to calculate SpO2. In fact, the signal the system acquired is the mixed signal. In the case of ideal signal to noise ratio, venous oxygen saturation can be calculated by formula (11) and formula (12). Mixed_SpO2 = (SpO2 + SvO2)/2 ( R_SpO2 = R_SvO2 = ) ( I 660 t I 660 / t ) I 940 t I 940 t (11) (12) At last, the invasive blood gas analysis is utilized as the golden criterion to calibrate the data measured by our system, referring to the calibration process of SpO2. The blood gas analysis platform of the laboratory can control the oxygen concentrations by adjusting the intake of oxygen. The system records the values of SaO2 and SvO2 in noninvasive method, at the same time the blood samples of artery and vein are sampled in 100 70 % hypoxic state. 24 samples of venous blood were sampled near from our finger in consistent with the process showed in Fig. 4. The sampling was conducted at six oxygen

Page 6 of 9 Fig. 4 Sampling points saturation levels, with two samples collected at each level. This was repeated twice to acquire the 24 samples. Results and discussion In this manuscript, we introduced a system for the continuous and noninvasive measurement of SvO2. Our system utilizes a control system to modulate the adjustable stimulation signal that is essential to the measurement process, in which the stimulation signal rate shouldn t be the same as patients pulse rate. The power spectrum of PPG in Fig. 3 can be found in Fig. 5, which can be divided into low frequency spectrum for pulse and high frequency spectrum for stimulation signal. The peaks of the power spectrum of the two kinds of signal are separated in frequency domain. In this way the ordinary FIR filter can be used to extract the pulse spectrum for the calculation of arterial oxygen saturation easily, and the stimulation signal for the calculation of the venous oxygen saturation. The experimental results indicate that the method of addition artificial stimulation can successfully transform the vein interference signal of pulse oxygen saturation measurement into useful signal to venous oxygen saturation measurements without compromising the pulse oxygen measurement. In addition, under hypoxic conditions, the system can reflect the overall decline of venous oxygen saturation better. Fig. 5 Respiration peak and pulse rate peak

Page 7 of 9 To identify the stability and reliability of the system, we studied nine healthy adults. Five volunteers took part in the invasive experiment and others took part in the noninvasive experiment. Invasive blood gas analysis (co-oximetry, golden criterion) was used to benchmark the accuracy and stability of the system. Referring to the standard calibration process of SpO2, samples of venous blood near from subjects fingers were collected. The sampling point is approximately set in 6 levels of oxygen saturation, and recorded twice for each subject (Fig. 4). The difference between the values which measured by the system and co-oximetry is almost within 10 to 10, as shown in Fig. 6, which includes all the data points without excluding any outliers. Figure 7 displays changes in nsvo2 (noninvasive SvO2) and isvo2 (invasive SvO2) through different oxygen saturation levels from one subject. The root mean square error (the difference of SvO2-iSvO2) is 5.31 Fig. 6 The difference of SvO2 between co-oximetry and noninvasive measurements Fig. 7 SvO2 trend chart of the new method system and co-oximetry

Zhang et al. BioMed Eng OnLine (2016) 15:84 and the correlation coefficient (the difference of SvO2-iSvO2) is 0.72. In general, the system is stable but the accuracy should be further improved. Oxygen consumption is equal to SaO2 minus SvO2. Figure 8 shows is the oxygen consumption estimated by the noninvasive method, and Fig. 9 is the oxygen consumption estimated by the invasive method. Fig. 8 The oxygen consumption of body measured by noninvasive method Fig. 9 The oxygen consumption of body measured by invasive method Page 8 of 9

Page 9 of 9 The oxygen consumptions in the two figures are very close. Therefore, the new method can be used to assess the body oxygen consumption. Conclusions The experimental results indicate that the noninvasive method can be a feasible approach for SvO2 measurement. Adding the external stimulation signal to vein solved the problems of the lack of fluctuation signal and improved signal-to-noise ratio in the measuring process. This new method of SvO2 measurement can potentially meet the unmet clinical needs, reducing the time and cost involved. Additional file Additional file 1: Raw Data. Abbreviations SvO2: blood oxygen saturation of vein; SaO2: blood oxygen saturation of artery; PPG: photoplethysmography; nsvo2: noninvasive SvO2; isvo2: invasive SvO2. Authors contributions MMZ, SKZ and LQW carried out the measurement, analysis and drafted the manuscript. XZ and JLY concepted and designed the experiment and revised the manuscript critically. All authors read and approved the final manuscript. Author details 1 School of Medicine, Shenzhen University, Shenzhen, China. 2 National-Regional Key Technology Engineering Laboratory for Medical Ultrasound, Shenzhen, China. 3 Guangdong Key Laboratory for Biomedical Measurements and Ultrasound Imaging, Shenzhen, China. Acknowledgements Research was supported by science and technology innovations project of Shenzhen (JCYJ20140418182819121 and ZDSY20120612094855904). JCYJ20140418182819121 is the study of the key technology of the method of anesthesia Dept. ZDSY20120612094855904 is the study of ECG database and ECG analysis algorithm. JCYJ20140418182819121 supported the data collection of the study and ZDSY20120612094855904 supported the data analysis of the study. Competing interests The authors declare that they have no competing interests. Availability of data and materials The dataset supporting the conclusions of this article is included within Additional file 1. Received: 2 December 2015 Accepted: 13 July 2016 References 1. Zhang D. Venous blood gas analysis and its clinical significance[j]. J Med Postgrad. 2004;17(8):741 8. 2. Gao X, Liu X. Principle of pulse oxygen saturation testing and common probe connectors circnits[j]. J Press China Med Devices. 2010;25(6):57 9. 3. Nitzan M, Babchenkoz A, Khanokh B, Landau D. The variability of the photoplethysmography signal a potential method for the evaluation of the autonomic nervous system. Physiol Meas. 1998;19(2):93 102. 4. Baker CR Jr, Yorkey TJ. Method and apparatus for estimating physiological parameters using model-based adaptive filtering. US 6, 836, 679 B2[P]. 2004. 5. Cohn JN, Finkelstein S, McVeigh G. Noninvasive pulse wave analysis for the early detection of vascular disease[j]. Hypertension. 1995;26:503 8.