NIBP OEM Module. (Model : M_NIBP) Contents. 1. Concept 2. Composition 3. PCB Dimension 4. Specification 5. Protocol 6. Cuff & Electrical Connection

Similar documents
hlon Invasive Blood Pressure OEM board NIBP2O2O UP

Section 2: Using the Invasive Blood Pressure Module - An Introduction

Non-Invasive Blood Pressure (NIBP)

Instruction Manual. Auto Inflate Blood Pressure Monitor

BLOOD PRESSURE SENSOR 0377i

BLOOD PRESSURE SENSOR BT17i USER S GUIDE

Digital Blood Pressure Monitor. Model UA-787

Connections The C20B can be connected either using the Molex locking socket, or by connection to four plated through holes. The plated through

PRELIMINARY DATASHEET

Model UB-328. Wrist Digital Blood Pressure Monitor. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Digital Blood Pressure Monitor. Model UA-705. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni 1WMPD B

Datasheet: K-30 ASCII Sensor

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

Fresenius Medical Care North America Service Bulletin

Linear Infl ation Technology Non-invasive Blood Pressure Measurement. inibp Case Report Experiences in OR

Noninvasive Blood Pressure Measurement. D. J. McMahon rev cewood

Overview. Front Panel: Keypad and Display

Mouse and Rat Tail Cuff Blood Pressure Systems from 1-24 Animals. Mouse rat tail cuff blood pressure system components

Non-Invasive Blood Pressure

Model AUTOMATIC UPPER ARM Blood Pressure Monitor

Declaration of Equivalence Form

Analog Front End Design of a Digital Blood Pressure Meter

Semi-Automatic Blood Pressure Monitor with Memory

Fixed Ozone Monitor User Manual December, Automatikprodukter

APPLICATION NOTE. OJ-DV Modbus protocol. OJ Drives A DRIVES PROGRAMME DEDICATED TO VENTILATION SOLUTIONS E 03/18 (PDJ) 2018 OJ Electronics A/S

INSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Fit Cuff. =Fit Cuff=!"#$% IA1B. Model

MODEL CH-432B INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR. - Eng 1 - English Español Português Deutch. Italiano Français CH-432B

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Operating Instructions Part No

CF8-W-Disp-CO. User manual. CO 2 / CO sensor with built-in general purpose controller

User Manual. Hopkins IMPACT Digital Wrist BP RELATED ITEMS. Item #526014

Device Equivalence Evaluation Form

ICS Supersedes EN :1997. English Version

GasSense NDIR User Manual

Sports Wrist Digital Blood Pressure Monitor Item #

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

AMS 6915 Board mount pressure sensor with digital output (I²C)

Contents. Safety Notice. Wrist Type

6833_INSTRUCTIONS MANUAL DIGITAL WIRST BLOOD PRESSURE

REF CH-302B IDENTIFICATION OF PARTS

Oregon Scientific Wrist Blood Pressure Monitor (BPW211)

Additel 761 Automated Pressure Calibrators

Accutorr. Quick Reference Guide

REF CH-602B IDENTIFICATION OF PARTS

Operating Instructions Part No

INTRODUCING CARDIOSAVE Hybrid MORE THAN A PUMP. A REVOLUTION. SPECIFICATIONS

User Manual. asense miii. / CO sensor with built-in general purpose controller

DIY - PC - Interface for Suunto Cobra/Vyper/Mosquito

Talking Wrist TypeBlood Pressure Monitor. Talking Wrist Type Blood Pressure Monitor Model: BPW810. User Manual

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

Blood Parameter Monitoring System 550

User Manual. asense VAV. CO 2 / temperature sensor with built-in general purpose controller

Talking Wrist Style Blood Pressure Monitor Model INS-LAB-RevB08

REASONS FOR THE DEVELOPMENT

Doc. no.dit om005 PRODUCT NAME. IO-Link/ELECTRO-PNEUMATIC REGULATOR. MODEL / Series / Product Number ITV*0*0-IO****-X395

Introduction. Table of Contents. Automatic Wrist Blood Pressure Monitor With Voice-Guided Operation. Model No.: BP5K

Bloodpressure measurement, improvement of oscillometric method.

User Manual DMD1001 1

PRODUCT MANUAL. Diver-MOD

Application Notes. SLP85xD Load Cells

Opti 24-hour ABPM. User Guide

BUBBLER CONTROL SYSTEM

Additel 761 Automated Pressure Calibrators

Additel 761 Automated Pressure Calibrators Selection Guide

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Additel 761 Automated Pressure Calibrators

M-BUS communication protocol

CATALOG NO. FIG DESCRIPTION NIB A Replacement Hose for Critikon-type devices-models 8100, 8100T, 9300XL, & 1846SX. 12

HOME BLOOD PRESSURE KIT INSTRUCTIONS TRILINGUAL INSTRUCTION GUIDE MODEL UA-100

Levitronix LEVIBOOST 75 or 140. up to 75 or 140 l/min up to 20 or 37 gpm MANUAL

SDI mA Pressure Transducer (Submersible + Dry) Model WL2100 Operating Manual

Stand-Alone Bubble Detection System

Additel 761 Automated Pressure Calibrators Selection Guide

DPC-30 DPC-100. Reference Manual

TD-3140 Blood Pressure Monitor. Owner s Manual

Fully Fuzzy Auto Digital Wrist Blood Pressure Monitor. tensio control. INSTRUCTION MANUAL Model GP-6220

User Manual DMD1001. Arm Type. Digital Blood Pressure Monitor. version:1.0

USE, CARE & MAINTENANCE DIAGNOSTIX TM 972 MERCURIAL BLOOD PRESSURE INSTRUMENT

Digital Blood Pressure Monitor Model UA-621 Instruction Manual

Certification Exhibit FCC ID: R7PIWRS4. FCC Rule Part: ACS Report Number: W03.11.A

Digital Blood Pressure Monitor for the Upper Arm INSTRUCTION MANUAL

Upload-pressurized Automatic Sphygmomanometer.

400H HARDMOUNT AIR COMPRESSOR KIT PART NO H HARDMOUNT AIR COMPRESSOR KIT PART NO

English Español Português Italiano Deutch Français INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR MODEL CH-656C. - Eng 1 -

User Manual DMD1001. Arm Type. Digital Blood Pressure Monitor. version:1.0

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

ProBP 2400 Digital Blood Pressure Device

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

Blood Pressure Monitor

Simulation for emergency care

WiiMaze Design Document March 13, 2008

Precision digital pressure gauge Model CPG1000

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

TD-3140 Blood Pressure Monitor. Owner s Manual

Domino DUEMMEGI. Domino. Communication Interface DFTS User s Manual. Release September 2007

Burner Management System DEMO Operating instructions

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

GoWISE USA R Blood Pressure Monitor Wrist Type. Instruction Manual

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

Transcription:

1 NIBP OEM Module (Model : M_NIBP) Contents 1. Concept 2. Composition 3. PCB Dimension 4. Specification 5. Protocol 6. Cuff & Electrical Connection

2 1. Concept Rev V1.00 Oscillometry : The oscillometric method of blood pressure measurement is a non-invasive method that monitors the amplitude of cuff pressure changes during cuff deflation to determine arterial blood pressure. The cuff pressure is first elevated above the patient systolic blood pressure level and the cuff begins to deflate at a certain rate. The initial rise in amplitude of these pressure fluctuations during cuff deflation corresponds closely to the systolic blood pressure. As the cuff is further deflated, these pressure fluctuations increase in amplitude until a peak is reached which is usually referred to as the mean arterial pressure(map). As cuff deflation continues, the diastolic pressure can be determined based upon the rapidly diminishing amplitude of the pressure fluctuations. Thus systolic, MAP and diastolic blood pressures can be accurately obtained by supervising the pressure fluctuations while controlling the cuff deflation rate. 2. Composition 2-1. Module Selection Model : M_NIBP 2-2. Components Discription Accessory Fig. No. Order No. NIBP Board Fig 1 M_NIBP_PC B RS232C & Power Harness Harness to Connect RS-232C Communication & Supply Power Fig 2 M_NIBP_C1 UART TTL & Power Harness Harness to Connect UART TTL Communication & Supply Power Fig 3 M_NIBP_C2 M_NIBP Components Cuff Adult Set(A:Large, B:Medium, C:Small) Fig 4 Cuff_Adt_* Neonate Set(A:Large, B:Medium, C:Small) Fig 5 Cuff_Neo_* Hose Silicon Tube 1meter Fig 6 D3_1000 PC Program CD WindowsXP(only for evaluation test module) Fig 7 M_NIBP_SW

3 2-3. Evaluation Kit Discription Accessory Fig. No. Order No. NIBP Board Fig 1 RS232C & Power Harness Harness to Connect RS-232C Communication & Supply Power Fig 2 M_NIBP Evaluation Kit Cuff Adult Cuff_Adt_B Fig 4 Neonate Cuff_Neo_B Fig 5 M_NIBP_EV Hose Silicon Tube 1meter Fig 6 PC Program CD WindowsXP(only for evaluation test module) Fig 7 2-4. Figures Fig 1 NIBP Board(M_NIBP_PCB) J4 is an Amp part number 103635-9(90 connector).

4 Fig 1 NIBP Board(M_NIBP_PCB) J4 is an Molex part number 5268-10(90 connector). Fig 2 RS232C & Power Harness(M_NIBP_C1) Fig 3 UART TTL & Power Harness(M_NIBP_C2)

5 Fig 2 RS232C & Power Harness(M_NIBP_C1) Fig 4 Adult Cuff(Cuff_Adt_*) Fig 5 Neonate Cuff(Cuff_neo_*) Fig 6 Silicon Hose

6 3. PCB Dimension

7 Dimension Weight Power Source Power consumption Measuring Method 105.4 69.4 38mm(Width Length Height) Appro. 150g DC from 6.0 to 15Volt ± 5% Maximum12Watt, Typical 6Watt Oscillomeric Adult 40 270mmHg SYSTOLIC of Blood Pressure Pediatric 40 150mmHg Neonate 40 130mmHg Adult 30 260mmHg Measuring range MAP of Blood Pressure Pediatric Neonate 30 140mmHg 30 120mmHg Adult 20 250mmHg DIASTOLIC of Blood Pressure Pediatric 20 130mmHg Neonate 20 110mmHg Pulse Rate Measuring range 30-300bpm Measuring Accuracy Initial Inflation Pressure Blood Pressure Pulse Rate Adult Pediatric Neonate ±3mmHg between 0 mmhg and 300 mmhg for operating conditions between 0 C and 50 C. ± 2% or ± 3 BPM, whichever is greater 180mmHg(Defualt), Variable 120-280mmHg 130mmHg(Defualt), Variable 100-160mmHg 120mmHg(Defualt), Variable 80-140mmHg Startup Initialization Period 5 seconds 4. Specification

8 Maximum cuff inflation time is limited to 35 seconds Patient Safety Operation Condition Storage Condition Altitude cuff pressure exceeds 300 mmhg (Adult & Pediatric modes) or 150mmHg (Neonate mode) at any time the cuff has been inflated for 180 seconds (Adult & Pediatric modes) or 90 seconds (Neonate mode) 0 C to 50 C, 15% to 95% non-condensing humidity 20 C to 65 C, 15% to 95% non-condensing humidity Measurement accuracy is not affected by altitude Interfacing UART TTL UART RS-232C Asynchronous Operation Baud rate 9600bps, 8 Data bit, 1 Stop bit, no parity bit The Module meets all relevant parts of the following Safety Standards: IEC60601-1:1997 IEC/EN60601-2-30:1999/2000 AAMI SP10:1992/2002 EN1060-1:1996 EN1060-3:1997 5. Protocol 5.1 Protocol UART [TTL Level (0~5[V])] / [RS-232C Level] Asynchronous Operation Baud rate : 9600bps 8 Data bit 1 Stop bit no parity bit

9 5.2 Host Command Packet Structure() HOST START BYTE COMMAND BYTE DATA BYTE(S) CHECKSUM BYTE : character (0x3A) 1Byte 1Byte Supplemental command data. The number of data bytes varies according to the command. LSB 1Byte [0x100 - BYTE(Startbyte + Command bytes + Data bytes)] COMMAND DATA DESCRIPTION COMMAND DATA DESCRIPTION 0x17 Pressure Data(2Bytes) Sets initial pressure for next BP Data 0x20 None Starts Adult Mode BP 0x87 None Starts Pediatric Mode BP 0x28 None Starts Neonate Mode BP 0x79 0x01, 0x00 Aborts current BP reading 0x79 0x05, 0x00 Returns current cuff pressure 0x79 0x03, 0x00 Returns data from last BP measurement 0x0C B0, B1, B2 External control of pump & valves This command should never be used when the cuff is attached to a patient. B0 = Pump (0x00 = Off, 0x01 = On) B1 = Control Valve (0x00 = Open, 0x01 = Closed) B2 = Dump Valve (0x00 = Open, 0x01 = Closed) Sending the Starts BP command will automatically place the module in the Adult monitoring mode, causing excessive cuff pressures on a neonate. It is the responsibility of the Host system to always determine Adult, Pediatric or Neonate mode when initiating a BP measurement. 5.3 Module Packet Structure(Module Host) MODULE START BYTE PACKET BYTE DATA BYTE(S) CHECKSUM BYTE > character (0x3E) total number of bytes in the packet. the Module data response to the COMMAND issued by the Host. LSB 1Byte [0x100 - BYTE(Startbyte + Command bytes + Data bytes)] 5.4 Example to set initial inflate to 180mmHg pressure(decimal 180 Hex 0xB4) 0x3A 0x17 0xB4, 0x00 0xFB

10 Module Host( OK packets) 0x3E 0x04 0x4F 0x6F 0x3E 0x04 0x4B 0x73 5.5 Example to start adult mode BP 0x3A 0x20 (None) 0xA6 Module Host (acknowledges O about the start command) 0x3E 0x04 0x4F 0x6F Module Host ( K after BP measurement is completed) 0x3E 0x04 0x4B 0x73 When the Module is taking a BP measurement, the only commands that are "valid" are the abort BP and the returns current cuff pressure commands. The Module will respond to all other commands with a "Busy" data packet (data byte = "B") as shown below. Module Host ( B module is busy) 0x3E 0x04 0x42 0x7C 5.6 Example to start Pediatric mode BP 0x3A 0x87 (None) 0x3F Module Host (acknowledges O about the start command) 0x3E 0x04 0x4F 0x6F

11 Module Host ( K after BP measurement is completed) 0x3E 0x04 0x4B 0x73 When the Module is taking a BP measurement, the only commands that are "valid" are the abort BP and the returns current cuff pressure commands. The Module will respond to all other commands with a "Busy" data packet (data byte = "B") as shown below. Module Host ( B module is busy) 0x3E 0x04 0x42 0x7C 5.7 Example to start Neonate mode BP 0x3A 0x28 (None) 0x9E Module Host (acknowledges O about the start command) 0x3E 0x04 0x4F 0x6F Module Host ( K after BP measurement is completed) 0x3E 0x04 0x4B 0x73 When the Module is taking a BP measurement, the only commands that are "valid" are the abort BP and the returns current cuff pressure commands. The Module will respond to all other commands with a "Busy" data packet (data byte = "B") as shown below. Module Host ( B module is busy) 0x3E 0x04 0x42 0x7C

12 5.8 Example to abort BP 0x3A 0x79 0x01, 0x00 0x4C Module Host( AK packets : Module Response when BP measurement in progress) 0x3E 0x04 0x41 0x7D 0x3E 0x04 0x4B 0x73 Module Host( A packets : Module Response when BP measurement not in progress) 0x3E 0x04 0x41 0x7D 5.9 Example to return current cuff pressure 0x3A 0x79 0x05, 0x00 0x48 Module Host(current pressure of 258mmHg) 0x3E 0x05 0x02, 0x01 0xBA Module Host(current pressure of 142mmHg) 0x3E 0x05 0x8E, 0x00 0x2F 5.10 Example to return data from last BP measurement 0x3A 0x79 0x03, 0x00 0x4A Module Host 0x3E 0x18 SYS, DIA, N1, N2, N3, HR, MAP, EC, N4, N5 0x100 Modulo 256 sum of all bytes

13 DATA PACKET Data Description Number of Bytes SYS Systolic value in mmhg (unsigned integer, LSB first) 2 DIA DDD Diastolic value in mmhg (unsigned integer, LSB first) 2 N1 Not used 1 N2 Not used 1 N3 Not used 8 HR Heart Rate in beats per minute (unsigned integer, LSB first) 2 MAP Mean Arterial Pressure (MAP) in mmhg (unsigned integer, LSB first) EC Error Code (unsigned byte) 1 N4 Not used 1 N5 Not used 1 2 Error Code List & Definitions Error Code Description Corrective Action: 0 = 0x00 Good BP Reading None 1 = 0x01 2 = 0x02 4 = 0x04 Weak or no oscillometric signal Artifact / erratic oscillometric signal Exceeded measurement time limit 85 = 0x55 Pneumatic Blockage 86=0x56 87 = 0x57 BP reading terminated by user Inflate Timeout, Air Leak or Loose Cuff Check that the cuff is in the correct position. Check the patient. Check that the cuff is properly tightened. Check that there is no excessive clothing between the arm and the cuff. Check that the correct size cuff is being applied. The patient may have been moving too much. Check that the cuff is in the correct position. Check that the correct size cuff is being applied. The patient may have been moving too much. Check that the cuff is properly tightened. Check that the cuff is in the correct position. Check that the correct size cuff is being applied. Check that there is no excessive clothing between the arm and the cuff. Check that the hose has no sharp bends or is pinched. Check that the patient is not lying on the cuff. Check that the cuff is in the correct position. Check the patient. Take another BP reading. Check that the hose is connected to the system and the cuff. Check that the cuff is properly tightened. Check that the cuff is in the correct position. Check that the correct size cuff is being applied. Check that the cuff is not leaking air. Check that the hose connections are not damaged or loose.

14 89 = 0x59 Cuff Overpressure 90 = 0x5A Power supply out of range or other hardware problem 97 = 0x61 Transducer out of range 98 = 0x62 ADC out of range 99 = 0x63 EEPROM calibration data failure Check that the correct size cuff is being applied. Check that the hose has no sharp bends or is pinched. Check that the cuff is in the correct position. Check that the patient is not lying on the cuff. Check that V PUMP and V LOGIC are within the voltage specifications. Check the data/power input connection. Service may be required. Call a manufacturer representative. It is not correct calibration, need to recalibrate. Service may be required. Call a manufacturer representative. Service may be required. Call a manufacturer representative. It is not correct calibration, need to recalibrate. Service may be required. Call a manufacturer representative. 5.11 CONTROL_PNEUMATICS Caution : This command should never be used when the cuff is attached to a patient. Data definitions: 0x3A 0x0C B0, B1, B2 B0 = Pump B1 = Control Valve B2 = Dump Valve (0x00 = Off, 0x01 = On) (0x00 = Open, 0x01 = Closed) (0x00 = Open, 0x01 = Closed) 0x100 Modulo 256 sum of all bytes Example to close the Control Valve and the Dump Valve: 0x3A 0x0C 0x00, 0x01, 0x01 0xB8 Module Host( OK packets) 0x3E 0x04 0x4F 0x6F 0x3E 0x04 0x4B 0x73

15 6. Cuff & Electrical Connection 6.1 Power and communication connector (PCB J4) Pin NO.(PCB J4) Description Name 1 No Connect or Optional RS-232 TxD (data out) TXD-RS 2 No Connect or Optional RS-232 RxD (data in) RXD-RS 3 Ground of Pump GND PUMP 4 Ground of Pump GND PUMP 5 +VCC of Pump V PUMP 6 +VCC of Pump V PUMP 7 Logic Level RxD (data in) or No Connect RXD-TTL 8 Logic Level TxD (data out) or No Connect TXD-TTL 9 Ground of Logic GND Logic 10 +VCC of Logic V Logic

16 6.2 J4 is an Amp part number 103670-9(straight connector) or 103635-9(90 connector). The mating connector is Amp part number 104257-9. Pins for the mating connector are Amp part number 104480-7. OR J4 is an Molex part number 5267-10(straight connector) or 5268-10(90 connector). The mating connector is Molex part number 5264-10. Pins for the mating connector are Molex part number 5263PBT9. 6.3 Caution to communication connection Do not attempt to connect to both Logic Level and RS232 simultaneously. If use to connect to Logic Level, have to remove R137 on PCB. Rear View of PCB