EVALUATION OF INTRAVENOUS THERAPY DEVICES IN THE HYPERBARIC CHAMBER

Similar documents
NN- NAVY EXPERIMENTAL DIVING UNIT SECNIALRPOR NO. 6-9 NAV EPEIMETAL DI VING UIT~ TIC-99 DTIC QUAJLIXDISPECTED X

200602q PULMONARY FUNCTION AFTER OXYGEN-ACCELERATED DECOMPRESSIONS FROM REPETITIVE SUB-SATURATION AIR DIVES

POTENTIAL BENEFITS OF NAVY DIVE COMPUTER USE IN SHIPS HUSBANDRY DIVING: ANALYSIS OF DIVES CONDUCTED ON THE USS RONALD REAGAN (CVN-76)

D TIC 0 F U AD-A IllllllIII ELECTE f1 JUN2 71M4 DEPARTMENT OF THE NAVY

EVALUATION OF THE KMS 48 FULL FACE MASK WITH THE VIPER VERY SHALLOW WATER UNDERWATER BREATHING APPARATUS

ANALYSIS OF GASES PRODUCED BY THREE UNDERWATER CUTTING DEVICES

DTIC 1111II IGI~ ~I HI II17 04 AD-A NAVY EXPERIMENTAL DIVING UNIT REPORT NO OMMERCIALLY AVAILABLE IDI SCUBA REGULATORE

UNMANNED TEST AND EVALUATION OF THE TELEDYNE ANALYTICAL INSTRUMENTS R-10DN OXYGEN SENSOR FOR USE IN THE MK 16 MOD 1 UNDERWATER BREATHING APPARATUS

AD-A DIVING UNIT NAVY EXPERI:MENTAL ...DTIC ILKICT[ 000 CTL14 Ift,

NAVY EXPERIMENTAL DIVING UNITC OTIC_

S... q. Psychometric Evaluation of the Mindstreams Neuropsychological Screening Tool. Navy Experimental Diving Unit

ELEDT S APR24 19 DECOMPRESSION SICKNESS AFFECTING THE TEMPOROMANDIBULAR JOINT. Author:

AD-A Ott, pp, ELECTE NAVY EXPERIMENTAL DIVING UNIT

Subj: HYPERBARIC OXYGEN TREATMENT IN NAVY RECOMPRESSION CHAMBERS

Alaris Pump module FAQs

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

AVExPERIMENTAL DVING UNIT

XVAL-HE-4B: A MAXIMUM PERMISSIBLE TISSUE TENSION TABLE FOR REAL-TIME THALMANN ALGORITHM SUPPORT OF CONSTANT 1.3 ATM PO 2 -IN-HELIUM DIVING TO 200 FSW


BODYGUARD 2CH TWIN CHANNEL INFUSION PUMP

DIVE DIVING MEDICAL (MED) Checklist

EVALUATION OF THE KMS 48 FULL FACE MASK CONFIGURED FOR OPEN CIRCUIT UBA WITH CROSSOVER TO EITHER MK 16 MOD 0 UBA OR MK 25 UBA

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

Challenges in Undersea Medicine

Recommendations on Checking Anaesthesia Delivery Systems

REPORT TO THE 2009 LEGISLATURE REPORT ON THE UNIVERSITY OF HAWAI I S HYPERBARIC TREATMENT CENTER. ACT 158, SLH 2008 (Part 3, Section 90.

Defense Technical Information Center Compilation Part Notice

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Multi Channel Solutions

:Q WfiEi~RW DIVING UNIT~

University System of Maryland (USM) Dive Plan Approval Form To be submitted to the USM Scientific Diving Safety Officer

Introduction. Medley TM Medication Safety System. Alaris Medical Systems

Defibtech s Latest Advancement The Lifeline ARM Automated Chest Compression (ACC) Device for Professionals

LEADER SEARCH RADAR LIFE LOCATOR

Training and Evaluation Outline Report

MILITARY STANDARD INTERFACE STANDARD FOR SHIPBOARD SYSTEMS SECTION 106 COMPRESSED AIR SERVICE FOR SURFACE SHIPS

Hospital and Transport for Controlled Breathing

Temporary Flying Restrictions Due to Exogenous Factors

DEPARTMENT OF THE NAVY NAVY EXPERIMENTAL DIVING UNIT Panama City, Florida NAVY EXPERIMENTAL DIVING UNIT REPORT NO. 3-79

Anaesthetic infrastructure and supplies

First Class. Monitoring. solutions for. Range of monitors CO 2 O 2 SpO 2 ECG NIBP. More Choices to meet your needs

TECHNICAL REPORT No MANNED EVALUATION OF THE CARLETON 1.3 ATA P0 2 PRIMARY ELECTRONICS ASSEMBLY WITH THE MK 16 UNDERWATER BREATHING APPARATUS

US NAVY DIVING PROGRAM: DIVING TO 300 FT DEPTHS USING SURFACE-SUPPLIED AND SATURATION FLY-AWAY DIVING SYSTEMS. Introduction

Low-Cost Dry Submersible Technology & Dry Combat Submersible Light

Making Life Easier Easypump II

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

HAZARDOUS INCIDENT SITE SAFETY PLAN

DEVELOPMENT AND EVALUATION OF AN ONLINE AIR QUALITY MONITOR (DIVEAIR2) FOR DIVING COMPRESSORS

MILPERSMAN (703) (901) (901) NAVPERSCOM CUSTOMER SERVICE CENTER

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

Advanced Test Equipment Rentals ATEC (2832)

Commandant United States Coast Guard

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Air-Sea Interaction Spar Buoy Systems

DIVE PLAN SUBMITTAL FORM

Commandant United States Coast Guard

EXECUTIVE SUMMARY USN DIVE MANUAL REVISION 7 Change A

Introduction. U.S. Navy Diving Manual (Mar-1970)

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR PRESSURE GAUGE. This publication replaces T.O. 33K dated 30 January 2012.

Compatible Baxter IV Sets

Emergency Transport and Ventilation

Basic life support updates

Remote Monitoring of Dolphins and Whales in the High Naval Activity Areas in Hawaiian Waters

NONIII F/S 13/10 EVALUATION OF ME 12 SURFACE-SUPPORTED DIVING SYSTEM (,SSDS) MODI-ETCIUI JUN 81 E H PAHL UNCLASSIFIED NEDU-7-81

PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN

USF/FIO Dive Plan Approval Form To be submitted to both the: Keys Marine Lab & the USF/FIO Scientific Diving Office

DEPARTMENT OF THE NAVY HEADQUARTERS UNITED STATES MARINE CORPS WASHINGTON, DC MCO C21 21 Dec 92

NAVSEA STANDARD ITEM CFR Part 1915, Occupational Safety and Health Standards for Shipyard Employment

Certified Hyperbaric Specialist Application

NOAA DIVING PROGRAM MINIMUM STANDARD FOR DECOMPRESSION DIVING USING OPEN-CIRCUIT SCUBA EQUIPMENT

No additional equipment must be stored in/on the cardiac arrest trolley/grab bag without prior agreement with the Resuscitation Committee

Chest Drainage Systems for Cardiac & Thoracic Surgery. Including portable suction with Digital Air Leak Monitoring

Common Discrepancies Uncovered on ADCI Diving Contractor & Training Program Audits

M2.50-cal. Multishot Capabilities at the US Army Research Laboratory

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control

DIVE ADMINISTRATION (5AD) Checklist

PRC CO ² -LASER PRESENTATION

Requirements for Reduced Supervision of Power Plants, Thermal Liquid Heating Systems, and Heating Plants

Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator

DG100 Digital Depth Gauge and Timer. Training Module, Rev B

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

High-performance submersible pressure transmitter For level measurement Model LH-10

Transmitter CS 21 Operation Manual

Eliminating Sources of Error in IV Pump Testing

DTICD I JTMMUTON STATEMEN DIVING UNIT NAVY EXPERIMENTAL. Ditrbuio. Uli ite j N REPORT NO NAVY EXPERIMENTAL DIVING UNIT

Operation and Maintenance of the EPV200 Portable Ventilator

COMPETENCIES FOR SPECIFIC TASK AREAS UNDER THE SCIENTIFIC AND ARCHAEOLOGICAL DIVING ACoP

Mobile Breathing Air Systems

'40. j 1.8. IIJI25 11= L.6l MICROCOPY RESOLUTION TEST CHART NATIONAL BUREAU OF STANDARDS 1963 A

AS/NZS :2013. Training and certification of occupational divers AS/NZS :2013. Part 2: Surface supplied diving to 30 m

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Pressure Regulator. Instructions for Use

DEPARTMENT OF THE NAVY NAVAL AIR SYSTEMS COMMAND RADM WILLIAM A. MOFFEIT BUILDING BUSE ROAD, BLDG 2272 PATUXENT RIVER, MARYLAND,

Tourniquet System. Ideal for single or bi-lateral orthopaedic surgery and pain management

Negative Pressure Wound Therapy

Volumat Agilia _Volumat_Presentation_customer.ppt

ali-q TM Gravimetric Verification Procedure

Oxygen Usage Best Practice Guide

Ultrasound Dose Calculations

DEPARTMENT OF THE NAVY HEADQUARTERS UNITED STATES MARINE CORPS WASHINGTON, DC MCO C2I 13 Mar 92

Transcription:

Navy Experimental Diving Unit NEDU TR 03-21 321 Bullfinch Rd. December 2003 Panama City, FL 32407-7015 EVALUATION OF INTRAVENOUS THERAPY DEVICES IN THE HYPERBARIC CHAMBER Navy Experimental Diving Unit 2006021 3 101 Authors: D. F. Stanga DISTRIBUTION STATEMENT A: HMC / SCW / DSW Approved for public release; Principal Investigator distribution is unlimited.

la. REPORT SECURITY CLASSIFICATION Unclassified REPORT DOCUMENTATION PAGE lb. RESTRICTIVE MARKINGS 2a. SECURITY CLASSIFICATION AUTHORITY 3. DISTRIBUTION/AVAILABILITY OF REPORT 2b. DECLASSIFICATION/DOWNGRADING AUTHORITY DISTRIBUTION STATEMENT A: Approved for public release; distribution is unlimited. 4. PERFORMING ORGANIZATION REPORT NUMBER (S) 5. MONITORING ORGANIZATION REPORT NUMBER(S) NEDU TECHNICAL REPORT No. 03-21 6a. NAME OF PERFORMING 6b. OFFICE SYMBOL 7a. NAME OF MONITORING ORGANIZATION ORGANIZATION (If Applicable) Navy Experimental Diving Unit 02 6c. ADDRESS (City, State, and ZIP Code) 7b. ADDRESS (City, State, and Zip Code) 321 Bullfinch Road, Panama City, FL 32407-7015 8a. NAME OF FUNDING 8b. OFFICE SYMBOL 9. PROCUREMENT INSTRUMENT IDENTIFICATION SPONSORING ORGANIZATION (If Applicable) NUMBER BUMED 6.4 8c. ADDRESS (City, State, and ZIP Code) 10. SOURCE OF FUNDING NUMBERS 2300 E. Street N.W., Washington D.C. 20372-5300 PROGRAM PROJECT TASK WORK UNIT ACCESSION ELEMENT NO. NO. NO. NO. FTOO- 10 11. TITLE (Include Security Classification) EVALUATION OF INTRAVENOUS THERAPY DEVICES IN THE HYPERBARIC CHAMBER 12. PERSONAL AUTHOR(S) DARYL F. STANGA, HMC /DSW /SCW, USN 13a. TYPE OF REPORT 13b. TIME COVERED FROM 14. DATE OF REPORT (MMM/YYYY) 15. PAGE COUNT Technical Report 2001 to 2003 December 2003 12 16. SUPPLEMENTARY NOTATION 17. COSATI CODES 18. SUBJECT TERMS (Continue on reverse if necessary and identify by block number) FIELD GROUP SUB-GROUP Hyperbaric chamber, safe intravenous pump (IV), medical equipment 19. ABSTRACT: Ventilators, cardiac monitors, intravenous (IV) pumps, and other equipment are being evaluated to meet a U.S. Navy-identified need to improve patient care levels in the hyperbaric environment. Though the MTP MILITARY is currently the only IV pump that has been found suitable for use in the hyperbaric chamber, a search of the market has identified three pumps that afford increased capability for patient care. This report presents results from a series of tests conducted to increase the IV pump options available to commands operating hyperbaric chambers. The pumps evaluated each have capabilities that may make them appealing in different operational scenarios. 20. DISTRIBUTION/AVAILABILITY OF ABSTRACT 21. ABSTRACT SECURITY CLASSIFICATION UNCLASSIFIED/UNLIMITED 1XI SAME AS RPT. I IDTIC USERS Unclassified 22a. NAME OF RESPONSIBLE INDIVIDUAL 22b. TELEPHONE (w/area Code) 22c. OFFICE SYMBOL NEDU Librarian (850) 230-3100

CONTENTS Report Documentation Page... Contents... Page i ii SECTION: Introduction... 1 Methods... 1 General... Experimental Design and Analysis... 1 1 Equipment and Instrum entation... 3 Results... 3 Discussion... 5 Conclusions/Recom mendations... 6 References... 9

INTRODUCTION Since the U.S. Navy has identified a need to improve its level of patient care in hyperbaric environments, ventilators, cardiac monitors, intravenous (IV) pumps, and other equipment are being evaluated to meet this need. The only IV pump currently found suitable for use in the hyperbaric chamber is the MTP Military Transport Infusion Pump (Medical Technology Products, Inc.; Huntington Station, NY). After researching the market and contacting several clinical hyperbaric facilities, the Navy Experimental Diving Unit (NEDU) has identified three intravenous pumps with features that may provide fleet commands with options that meet their varied needs: the IMED Gemini PC-1, the ALARIS Medsystem Ill, and the Infusion Dynamics Power Infuser. This report presents test results on and documents the safety and performance of these three pumps, and, on those bases, evaluates what and how each might contribute to fulfilling Navy needs for such units in hyperbaric settings. GENERAL METHODS Testing was conducted both on the surface and at various depths in the NEDU treatment chamber. The units were reviewed in several phases to evaluate their 1. technical information, 2. physical characteristics, 3. surface functioning at atmospheric pressure, and 4. test depth functioning. EXPERIMENTAL DESIGN AND ANALYSIS Following a search of military and civilian organizations to identify standards for operating ventilators in a hyperbaric chamber, the U.S. Navy Diving and Manned Hyperbaric System Safety Certification Manual was found to define such standards for U.S. Navy hyperbaric chambers. 1 The National Fire Protection Association (NFPA), in its NFPA 99 Health Care Facilities, also provides standards for electronic devices used in oxygen-enriched environments, 2 a standard specifying current flows for operating medical devices in 0 2 -enriched environments. Test Parameters 1. To determine whether the pumps could be safely exposed in the hyperbaric environment, NEDU engineers inspected the units per NFPA 99 and the U.S. Navy Diving and Manned Hyperbaric System Safety Certification Manual. 1,2 1

2. Unmanned pressure testing to a depth of 247 feet of seawater (fsw) was conducted to ensure that increased pressures would not damage electrical components of the units. Results were documented and reviewed. 3. If a unit had an internal battery, that battery was tested per NEDU Test Plan 01-23.3 4. Function testing was conducted on the surface to establish unit operating characteristics and baseline operating parameters. The following measurements of pump function and settings were taken: For the ALARIS Medsystem Ill and the IMED Gemini PC-1 a. Flow rate at 20 ml/h (KVO), with total volume 20 ml b. Flow rate at 100 ml/h, with total volume 100 ml c. Flow rate at 400 ml/h, with total volume 400 ml For the Infusion Dynamics Power Infuser* a. Flow rate at 200 ml/h, with total volume 200 ml b. Flow rate at 1000 ml/h, with total volume 1000 ml c. Flow rate at 2000 ml/h, with total volume 2000 ml * The Power Infuser is designed as a high-volume rehydration pump and cannot be set to the levels of the other two pumps. 5. The IV pumps and calibrated containers were set up in the NEDU treatment chamber, and results from the parameter settings were recorded at 30, 60, and 165 fsw. Test Parameters at Treatment Depths in the Chamber On the surface, the IV pumps were set at one setting within their operating ranges. The chamber was then compressed to 30, 60, and 165 fsw. Upon reaching each depth, the inside tender monitored the units for one hour and measured each pump's output in the calibrated cylinders. Each test was repeated four times at each setting at each depth, and all results were recorded and reviewed. 2

EQUIPMENT AND INSTRUMENTATION - Infusion Dynamics Power Infuser - ALARIS Medsystem Ill - IMED Gemini PC-1 - Three (3) 1 000-mL calibrated cylinders - NEDU treatment chamber RESULTS The batteries, tested per NEDU Test Plan 01-23, were found to be safe. Their discharge times are listed in Table 1. Table 1. Operation after 24-hour charge Pump Predive Duration Postdive Duration IMED Gemini PC-1w 7 hours, 30 minutes 8 hours, 10 minutes ALARIS Medsystem III' 8 hours, 10 minutes 8 hours, 0 minutes Power Infuser 10 hours, 15 minutes N/A The pumps were set up in the NEDU treatment chamber and operated at the specified settings (see EXPERIMENTAL DESIGN AND ANALYSIS, Test Parameters, 4), and pump outputs were measured and documented as shown in Table 2. Table 2. Pump output volume as measured by calibrated cylinder IMED Gemini DEPTH SETTING cycle1 cycle2 cycle31 cycle4 Standard PC-ill (fsw) Deviation 0 20 ml/h 20 ml 20 ml 21 ml 20 ml 0.5 30 20 mlh 20 ml 20 ml 20 ml 20 ml 0 60 20 mlh 22 ml 22 ml 20 ml 20 ml 1.15 165 20 ml/h 20 ml 20 ml 20 ml 24 ml 2.0 0 100mL/h 100mL 102mL 100mL 100mL 1.0 30 100mlh 100mL 102mL 102mL 102mL 1.0 60 100mL/h 102mL 102mL 102mL 102mL 0 165 100mL/h 103mL 104mL 104mL 100mL 1.90 0 400 ml/h 410mL 405mL 400mL 415mL 6.45 30 400 ml/h 430mL 425mL 415mL 425mL 6.29 60 400 ml/h 425mL 415mL 425mL 425mL 5.0 165 400 ml/h 400mL 410mL 412mL 446mL 20.03 3

ALARIS DEPTH SETTING cycle 1 cycle 2 cycle 3 cycle 4 Standard Medsystem Il (fsw) I I I IDeviation 0 20 ml/h 20 ml 19 ml 20 ml 20 ml 0.5 30 20 ml/h 19 ml 20 ml 20 ml Alarm 0.58 60 20 mlh 22 ml 20 ml 20 ml 20 ml 1.0 165 20 mlh 20 ml 20 ml 20 ml 20 ml 0 0 100mL/h 102mL 102mL 100mL 102mL 1.0 30 100mUh 100mL 100mL 100mL 100mL 0 60 100lmUh 102mL 100mL 102mL 102mL 1.0 165 100mL/h 102mL 104mL 100 ml 104mL 1.91 0 400 mlh 405mL 410mL 405mL 405mL 2.5 30 400 mlh 400 ml 400 ml 410 ml Alarm 5.77 60 400mUh 410mL 415mL 425mL 425mL 7.5 165 400mL/h 420mL 410mL 400mL 412mL 8.23 Power Infuser DEPTH SETTING cycle I cycle 2 cycle 3 cycle 4 Standard (fsw) I I [ Deviation 0 200 ml/h 206mL 208mL 210mL 198mL 5.26 30 200mL/h 168mL 189mL 184mL 170mL 10.34 60 200mL/h 205mL 198mL 21omL 200mL 5.38 165 200 ml/h 200mL 200mL 200mL 200mL 0 0 1000 ml/h 890 ml 890 ml 950 ml 950 ml 34.64 30 1000 ml/h 1015 ml 1000 ml 1020 ml 1000 ml 10.31 60 1000 ml/h 910 ml 910 ml 920 ml Alarm 5.77 165 1000 ml/h 950 ml 960 ml 940 ml 960 ml 9.57 0 2000 ml/h 2000 ml 2050 ml 1990 ml 2000 ml 27.08 30 2000 ml/h 2050 ml 1980 ml 1850 ml 1900 ml 88.13 60 2000 ml/h 1925 ml 2000 ml 2050 ml 1870 ml 79.62 165 2000 mlh 1950 ml 1800 ml 1700 ml 1820 ml 102.75 4

DISCUSSION Technical review of supporting documentation for each of the three IV pumps found no components that caused significant concern about their use in the hyperbaric chamber. When batteries were removed and evaluated, all were found to be safe and functional. The IMED Gemini PC-1 and the ALARIS Medsystem Ill pumps were placed on their respective chargers for 24 hours to ensure that their batteries were fully charged. The Power Infuser operates on AAA batteries, which already are authorized for chamber use and do not require charging. Each pump was operated until its battery discharge was low enough to cause a low power alarm. The pumps were then pressurized 11 times, after which they were reexamined. The IMED Gemini PC-1 and ALARIS Medsystem Ill batteries were recharged and operated until a low power alarm was given (see Table 1). During testing at depth, low power alarms from the IMED Gemini occurred on two occasions. Once, 30 minutes into the fourth cycle of 100 ml/h at 30 fsw, the alarm sounded, but the pump continued to operate until the end of the test, with no apparent change in the flow rate. The unit had been in operation for five hours at the time of this alarm. The second alarm occurred 45 minutes into the fourth cycle of 11 ml/h at 60 fsw; again, the pump completed the cycle with no change in its flow rate. To assess the functional characteristics and establish the performance accuracy of the units, each IV pump was operated at the surface. As the results of Table 2 show, the Power Infuser is designed as a rehydration pump and is unable to deliver the small volumes that the other pumps can. The ranges set during the testing represent low-, mid-, and high-output ranges for each of the pumps. Once the accuracy of the pumps was established, the chamber was compressed to 30 fsw, and the pumps were operated through four one-hour runs at each of the given drip rates. The amounts collected in their calibrated cylinders were measured, and these cylinders were emptied at one-hour intervals. After the cycles at 30 fsw were completed, the chamber was compressed to 60 fsw, and all tests were repeated. Because of concerns about the decompression status for the inside tenders, the first cycles were for one hour, and the following three cycles at 165 fsw were shortened to 30 minutes, with the measured volume then doubled for charting purposes. 5

CONCLUSIONS/RECOMMENDATIONS All three pumps completed all testing with satisfactory results. Each of the three units has particular features that address specific needs: Infusion Dynamics Power Infuser. This small (9.8 cm x 7.0 cm x 5.0 cm), light (330-g) infusion pump powered by six AAA batteries rated to last approximately eight hours under typical operation can deliver fluids at rates from 0.2 to 6.0 L/h. It uses a scuttle mechanism pump and can deliver fluids to the patient even when the bag and unit are level with or below the patient. Tested by the U.S. Army, the Power Infuser is certified for air medivac use 4 ' 5 and is well adapted to facilities and units with limited available space, particularly to commands with transportable recompression chamber systems (TRCS). Its small size allows the unit to meet space limitations of deploying units. Medications, however, cannot be delivered by this unit: the filter system in its infusion set will easily clog if any substance other than IV solution is passed through it. Infusion Dynamics Power Infuser 6

ALARIS Medsystem III. This medium-sized unit (20.0 cm x 15.2 cm x 5.3 cm) weighs 1828 g and is powered by a nickel cadmium battery with an expected operating time of six hours or more when all three channels are operating. As a hospital intensive care-level IV pump, the Medsystem III can deliver three different medications or solutions simultaneously and can do so with great precision, at rates from 0.01 to 9.9 L/h. Its dose rate calculation function is also capable of delivering complex doses of medication. The pump is sensitive to air in the lines, however, and is often difficult to clear and restart if air or occlusions cause alarms to resound. This unit is a good choice for a hospital-based chamber expected to treat patients with multiple and complex medical problems. ALARIS Medsystem I1l 7

IMED Gemini PC-1. As a large unit (8.4 inches x 10.8 inches x 6.5 inches) weighing 11.2 Ib, this pump is rugged and versatile. Under normal rates of operation, an internal lead acid battery can power the Gemini for five hours. This unit, which is simple to operate and the least expensive of the pumps tested, can deliver fluids and medications at rates from 0.01 to 9.9 L/h. As a hospital-level pump with enough resilience to be transported to remote locations and into adverse conditions, it is well suited for multifunction hyperbaric treatment units. The pump worked well in all attitudes, both above and below the patient, and since it has been used extensively in military medical facilities throughout the world, most military healthcare providers as well as established biomedical repair and maintenance personnel are familiar with it. IMED Gemini PC-1 8

REFERENCES 1. Naval Sea Systems Command, U.S. Navy Diving and Manned Hyperbaric System Safety Certification Manual, Publication SS800-AG-MAN-010/P-9290 (Arlington, DC: NAVSEA, 1998). 2. National Fire Protection Association, NFPA 99 Health Care Facilities (Quincy, MA: NFPA, 2002. 3. R. R. Lowe, D. F. Stanga, and G. Conley, Unmanned Evaluation of Batteries for Use in Hyperbaric Chambers, NEDU TP 01-23, Navy Experimental Diving Unit, November 2001. 4. Commander, U.S. Army Medical Materiel Agency, memorandum, Aeromedical Certification of the Infusion Dynamics Power Infuser Model 10OB-3A for the MEDEVAC UH-60A Helicopters, MCMR-UAD (70-45a), dtd 10 Jul 2002. 5. D. R. Mandel, Testing and Evaluation for the Infusion Dynamics Power Infuser, Final Report, U.S. Air Force 31 1 th Human Systems Program Office/YAML, April 2003. 9