Shimming with VnmrJ 3.1 Software for Peak NMR Performance

Similar documents
Pulsed-Field Gradient Shimming With VNMR

Deuterium more than just for locking: Part II. H Topshim. Donna Baldisseri Boston Massachusetts Pre-ENC Workshops 2014, March 22

Typically NMR Sample Configuration

An Advice about Shimming in High-Resolution Nuclear Magnetic Resonance

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer

Things to know before operating an NMR spectrometer André Boltjes, University of Groningen DD NMR Lab v

Variable Temperature Operation

User guide for the NMR spectrometer AVA400Stud

AKTA pure 25 New Owner s Intro

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Standard Operational Procedure for Capillary Electrophoresis

RM-80 respiration monitor

High Pressure Chem-SCAN Operating Manual

1)! DO NOT PROCEED BEYOND THIS MARK

User guide for the NMR spectrometer AVA300

Fast 3D gradient shimming by only 2 2 pixels in XY plane for NMR-solution samples

Quadruple mass spectrometers (transducer type)

NMR Service. How to Prepare Samples for NMR

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

DCIF Training Guide Variable Temperature Bruker B400 (last update )

Setting a New Standard in Flash Chromatography Performance

Updated: Variable Temperature

Organic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results

KEM Scientific, Inc. Instruments for Science from Scientists

Allocation of referees, hours and pistes User manual of Engarde - August, 2013

Boyle s Law: Pressure-Volume Relationship in Gases

Console Acceptance Tests & Specifications. MERCURYplus NMR Spectrometer Systems Pub. No , Rev. B0902

Residual Gas Analysis Systems for Industry

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

FIG: 27.1 Tool String

An intuitive and innovative answer to your anaesthesia challenges

Oxygen Meter User Manual

Complete Range of Force Tensiometers

PINMRF Checkout Quiz - Bruker/TopSpin Version

ATL. Take Control of your Helium Supply. Advanced Technology Liquefier The next generation of helium recycling

MATRIX-MG Series. Innovation with Integrity. Automated High-Performance Gas Analyzers FT-IR

ZIN Technologies PHi Engineering Support. PHi-RPT CFD Analysis of Large Bubble Mixing. June 26, 2006

Courseware Sample F0

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

Specifications for the CMI 2409 Magnetocardiograph

Aerobic Respiration. Evaluation copy

To Logon On to your tee sheet, start by opening your browser. (NOTE: Internet Explorer V. 6.0 or greater is required.)

We release Mascot Server 2.6 at the end of last year. There have been a number of changes and improvements in the search engine and reports.

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Road Force Elite. The world's fastest diagnostic balancer NEW

The WIRTGEN GROUP User Magazine // N o 05. WIRTGEN GROUP solutions for automation and process optimization: On the leading edge

APA software instruction manual

Micro Motion 3098 Gas Specific Gravity Meter

CONTROL SYSTEM FOR POOL AND SPA

Cover Page for Lab Report Group Portion. Pump Performance

SPECIFICATION FOR AN MRBR 9.4 TESLA/310MM/AS CRYO-COOLED MAGNET SYSTEM

Specifications for Synchronized Sensor Pipe Condition Assessment (AS PROVIDED BY REDZONE ROBOTICS)

Using the Akta Prime plus October 22, 2012

New Cortex Software User Guide

Standard Operating Manual

Instruction Manual Dräger MSI P7 and MSI P7 plus

Optical Dissolved Oxygen Meter

Distributed Control Systems

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

These Terms and conditions apply to audit trails incorporated in weighing and measuring devices and systems 1.

Section 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

Lab 4: Transpiration

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

STARLOG. Capacitive Water Depth Probe

ACV-10 Automatic Control Valve

Leak Checking Large Vacuum Chambers

SOLUBILITY OF A SOLID IN WATER

PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

Vapor Pressure of Liquids

FLUORESCENCE DETERMINATION OF OXYGEN

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

UNIVERSITY OF WATERLOO

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

ON736 Series: Oxygen/Nitrogen by Inert Gas Fusion

Pressure calibration. 20

[MYLAPS INTEGRATION]

ONH836 Series: Oxygen/Nitrogen/Hydrogen by Inert Gas Fusion

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

ZIPWAKE DYNAMIC TRIM CONTROL SYSTEM OUTLINE OF OPERATING PRINCIPLES BEHIND THE AUTOMATIC MOTION CONTROL FEATURES

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

EasyLine Continuous Gas Analyzers. So smart, they re simple

Landsat 8's Emergency RMM

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

RSKtools for Matlab processing RBR data

Drift-Chamber Gas System Controls Development for the CEBAF Large Acceptance Spectrometer

ISO INTERNATIONAL STANDARD

The Tekran Series 2600-IVS Difference!

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Transcription:

Shimming with VnmrJ 3.1 Software for Peak NMR Performance Application Note Author Ron Crouch Agilent Technologies 13000 Weston Parkway Cary, NC 27513 USA Shimming is the term used to describe the process of adjusting the shape of the effective magnetic field in the region of the sample in an NMR magnet, in order to yield good resolution in the NMR experiment. The term shimming is historical, dating to the very early days of NMR spectroscopy when scientists actually inserted mechanical shims into the magnet system to affect the shape of the applied field. Modern NMR consoles now provide adjustment through multiple coils controlled by computers and digital-analog converters (DACs) designed to correct virtually all field shape. Shimming is done in both a maintenance mode and sample-to-sample, in order to assure the highest resolution results. The Agilent NMR console offers a variety of technologies to provide facile control of the adjustment process. This technical overview describes the use of four easy-to-use tools in the VnmrJ 3.1 Software to perform both sample-to-sample and maintenance shimming and assure superior NMR performance.

Sample-to-Sample Shimming For best performance, shimming is done with each sample, using the appropriate tools in the VnmrJ 3.1 software. The default shimming technique is the use of Z-PFG to adjust the most important Z-shims that are axial to the long axis of the sample. Z-PFG Changes in the position of the NMR tube, the height of the sample, or the type of solvent can have significant effects on spectral line shape, due to changes in the magnetic field shape. These can be offset with Z-Axis Pulse Field Gradient Shimming (Z-PFG). There is a convenient menu entry point for managing all PFG shimming tools in the VnmrJ 3.1 software. The implementation of Z-Axis PFG shimming is fully automatic before any sample, as a routine step in data acquisition, and can be selected to be performed before any NMR experiment as well. The NMR operator has the choice to accept and use the default PFG shim map or a different one. For example, the spectrometer may be running in automation, with the default Z-PFG map set for 2 H observe, and the sample in pure H 2 O. In such a case, the operator might select a 1 H PFG map for use with this sample. Dramatic improvement with Z_PFG Shimming Figure 1. Automatic Z-PFG shimming of a spectrum improves resolution. The left spectrum was acquired before shimming, and the right spectrum was acquired following Z-PFG shimming. Both of these spectra are scaled identically. Given a high lock sensitivity (which is the case with the OneNMR probe), 2 H Z-PFG shimming can be done with a single scan for any solvent. This means that the important step of doing Z-PFG shimming on a sample-to-sample basis typically takes only a total of 10 to 20 seconds. Also, when a series of multiple experiments is chosen to be run, each of which is quite long, each experiment can be quickly Z-PFG shimmed before beginning acquisition of data, at the choice of the operator. Figure 1 Illustrates a realistic loss of resolution and/or line shape due to differences in homogeneity between samples caused by different solvents and significantly different tube liquid column heights. The improvements in resolution and sensitivity resulting from an automatic Z-PFG shimming that was about 20 seconds in duration are remarkable. 2

Proshim Proshim (introduced in VnmrJ 3.1) provides easy touch-up shimming along the x and y axes of the sample for best line shape. Proshim optimizes the shim values one at a time in a very fast manner. Available shim methods are in the menu list (Figure 2). After selecting a desired method, shimming can be initiated manually or on-demand. Proshim can also be used for maintenance shimming (see the Maintenance Shimming section). Proshim methods are simple text files presented as a list of the shims to optimize in a given order, and can be run in automated mode. When an extensive Proshim method such as allshims or 4th-order_xyz is used, it can be helpful to follow it with a final Z-PFG shim. The user can define custom methods that will also appear on the menu (see the VnmrJ 3 Automation User Guide, Pub. No. 91001987 for details on creating methods). Proshim is provided in the VnmrJ 3.1 software to both complement and enhance the traditional Simplex lock autoshim. Simply choose the desired method from the menu and click the Proshim Now button Figure 2. Push button on-demand shim methods are defined by a simple menu list. 3

Three-Dimensional Shimming VnmrJ 3.1 provides a flexible and easy-to-use interface for adjusting all shims (along x, y and z axis) using gradient methods. 3D shimming can be done using any probe with or without gradient coils. 3D Shimming requires a very strong signal because of the large number of fids which must be acquired. For example, an 8x8 3-D matrix requires 128 fids for each shimming step. During the mapping process these 128 fids must be acquired for each of the 27 shims in a typical 27 shim set. Because of the lack of a sufficiently strong signal in an organic chemistry environment, 3D shimming is generally only useful for maintenance shimming. 3D Shimming is potentially usable for sample-to-sample shimming with Bio-NMR samples because they are typically in 90% H 2 O solution and a strong signal is always available. Water suppression is an essential component for most Bio-NMR experiments, and optimal calibration of all of the shims using the 3D shim interface could be very valuable. Figure 3 demonstrates the value of 3D shimming using a sample of ubiquitin in 90% H 2 O. The quality of water suppression with presaturation methods is a direct reflection of the quality of shimming. Note the nearly complete lack of residual H 2 O near 4.6ppm and the sharp appearance of the protein methyl resonances near -0.2ppm. Figure 4 shows a series of two-dimensional NOESY experiments acquired with or without 3-D shimming. The quality of information is best in the experiment obtained after 3-D shimming with this ubiquitin protein sample. Simplex Lock Autoshim The traditional Simplex lock autoshim has been a feature with Agilent NMR systems for several years. It can still be employed to do shimming, using a large selection of provided shimming methods. The VnmrJ 3 Automation User Guide contains detailed information on how to create a custom method. Figure 3. 1D spectrum of ubiquitin in 90% water in a Shigemi tube using a normal presaturation, directly after 3D shimming. No processing methods for suppressing water were used. The total procedure was less than 10 minutes, after the sample temperature equilibrated. Figure 4. Shimming of three NOESY experiments using only presaturation for water suppression, and the 3D shim interface. The shims were first significantly altered to give bad shims before shimming. For the left spectrum, 3D shimming was performed before the NOESY experiment. Normal Z-PFG gradient shimming was performed on the middle spectrum, along with some quick manual shimming of the x and y shims. For the right spectrum, only manual shimming was done. There is an obvious improvement in water suppression from right to left, and spectra features are improved with better shims. 4

Maintenance Shimming Maintenance shimming should be performed on a regular basis to assure proper operation over the long term. The easy-to-use maintenance tools accessed through the intuitive interface of the VnmrJ 3.1 software provide for efficient maintenance shimming. The need for maintenance shimming often dramatically decreases as a newly installed magnet settles in. Some touchup shimming might be desired on a daily basis in the first weeks after a magnet is commissioned, and this interval typically reduces with time. Three-Dimensional Shimming VnmrJ 3.1 provides a flexible and easy to use interface for adjusting all shims using gradient methods. 3D shimming can be done using any probe with or without gradient coils. It is important to note that use of 3D shimming as a maintenance tool is tied to a specific sample and that use on a sample-to sample-basis (other than for Bio-NMR samples in 90% H 2 O) is not appropriate. However, for maintenance shimming performed with a specifically selected sample 3D shimming is very powerful. The 3D shim interface is fully automatic and provides menu selection to tailor to a list of suggested samples. Perhaps the most convenient sample to use is a sample in 90% H 2 O because of the speed with which the mapping process can take place. Once the sample type is selected from a menu, all calibrations, mapping steps, and shimming are carried out without further input from the operator. A variety of samples is supported and use is routine, even with probes that have no PFG gradients at all. 3D shimming can also be used for probe installation and affords a very rapid and thorough probe maintenance procedure. In essence, 3-D Shimming is the ultimate maintenance shimming tool. Figure 5 illustrates the power of 3D Shimming to adjust a magnet that is very far from optimal values. For this figure a map created with a 90% H 2 0 shim map was used. The 3D shim on the water sample was followed by submission of the line shape sample with only the automatic Z-PFG adjustment. There is The power of 3D Shimming An excellent method for fast maintenance shimming a dramatic improvement in line shape. After magnet installation and the settlingin period are completed, typical monthly shim maintenance with 3D Shimming on a 90% H 2 O sample provides good line shape results and assures ongoing magnet homogeneity for optimal performance (Figure 6). Figure 5. 3D Shimming can correct a magnet that is very far from optimal values, and give very significant improvement in line shape. The figure on the left shows line shape with a very poorly shimmed magnet. The figure on the right shows the same line shape sample after a 2 minute 3-D shim on a 90% water sample followed by just Z-PFG shimming with the line shape sample. Figure 6. Monthly shim maintenance with 3-D Shimming on a 90% H 2 O sample provides good line shape results and assures magnet homogeneity for optimal performance. The spectrum shown is from a 5% CHCl 3 sample in acetone-d6. 5

Proshim Proshim provides both on-demand and scheduled maintenance shimming. It utilizes a straightforward scheduling interface to enable the administrative user to define a shimming method for a line shape sample reserved in the sample tray. Proshim tells the console to conduct unattended maintenance shimming at any time/day desired, either as a single event or on a recurring schedule (Figure 7). Once the job is defined (revealed by the Delete button being active and the Create button being grayed out), the day, hour and minute for desired maintenance shimming are entered and set by clicking the Shim Scheduler button (Figure 7). Using a line shape sample with the Fast Lineshape automated method, which includes a final Z-PFG line shape adjustment, can provide scheduled magnet maintenance in an automated fashion. On-demand maintenance is also available with the push of a button. Figure 7. To schedule Proshim maintenance shimming, simply define the desired name for the shimming study and select the method from the choices in the ShimMethod menu in the Shim Maintenance Scheduler tool (top panel). In addition, the position of the line shape sample in the sample rack and the sample solvent are specified. Click Create and the shimming job is in place.once the sample name, location, and shim method are set, the bottom portion of the scheduler can be used to define the time and frequency for the maintenance to be done. The Help facility guides the user through these steps. 6

Conclusions The VnmrJ 3.1 software provides a variety of very powerful automatic shimming tools for both sample-tosample and maintenance shimming. These range from simple Z-PFG for rapid routine shim adjustments for all samples, to a complete 3D PFG shimming interface for facile installation of new probes of any type. Proshim provides rapid, easy and automated fine adjustment of radial shims for both sample-to-sample and maintenance shimming, while 3D shimming is used primarily in the maintenance mode. Proshim is certainly preferred over the traditional Simplex shimming method, as it is faster and its methods are easier to create. Maintenance shimming is made easy with the shim scheduling tool. It can be used to create a defined method to submit a study to automatically adjust homogeneity using the line shape sample, to run on any given day and at any given time. The VnmrJ 3.1 software provides powerful and easy-to-use tools to meet any and all sample-to-sample and maintenance shimming needs, in order to assure superior NMR performance. 7

For more information Learn more: www.agilent.com Buy online: www.agilent.com/chem/store Find an Agilent customer center in your country: www.agilent.com/chem/contactus U.S. and Canada 1-800-227-9770 agilent_inquiries@agilent.com Europe info_agilent@agilent.com Asia Pacific adinquiry_aplsca@agilent.com www.agilent.com Product specifications and descriptions in this document are subject to change without notice. Agilent Technologies, Inc., 2011 Published in USA, April 8, 2011 Publication Number 5990-7860EN