GasLab SAN-102 Micro Oxygen Monitor User Manual

Similar documents
GasLab SAN-102 Micro Oxygen Monitor User Manual

RAM 4021-DPX Operation Manual

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev

RAM 4021 Operation Manual

Calibration Requirements for Direct Reading Confined Space Gas Detectors

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

RAM Operation Manual

Phosgene Sensor. Sensoric COCl2 3E 1

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

EM151/EM201. Electronic Manometer - + INSTRUCTION MANUAL EM Washington Street Melrose, MA Phone Toll Free

Operation Manual. Pro Trimix Alarm TM. Helium & Oxygen Gas Analyzer. Rev

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

EM151/EM201. Electronic Manometer - + INSTRUCTION MANUAL. EM201 PSI

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

User Manual. Pro CO LP Alarm. Carbon Monoxide Analyzer and Low Pressure Monitor with Alarm REV: 03.19

In Vivo Scientific, LLC INSTRUCTION MANUAL

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

O3 3E 1 F Gas Sensor Module

DEEPSEA. LED Micro SeaCam POWER & LIGHT. Operator s Manual. Original Instruction Ruffin Road San Diego, CA USA

COMBAT SWIM BOARD. RJE International, Inc. RJE International Feb 1, TAC-300 Revision 6

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

Operation Manual. O2 Quickstick. Oxygen Analyzer

Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y

Advanced Test Equipment Rentals ATEC (2832)

Atmospheric Hazards. Hazardous Atmosphere 3/29/2017. Oxygen deficiency and enrichment Flammable gases and vapors Toxic contaminants

COMBAT SWIM BOARD. RJE International, Inc. RJE International January 16, TAC-300 Revision 4

Verification Of Calibration for Direct-Reading Portable Gas Monitors

RAM 9025-XP Operation Manual

URC Voltage Sensor SEN-VOLT for use with MRX units containing sensor ports

CALIBRATION KIT. Instruction Manual

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

DIVER NAVIGATION BOARD. RJE International, Inc. RJE International Feb 1, TAC-100D Revision 10

MODEL RP050BMST RESPIRATORY PROTECTOR MANUAL

Columbus Instruments

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

453 Series Steam Heated Vaporizing Regulator

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

Ambient Weather WS-03 Thermo-Hygrometer

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL RLD10

PERSONAL AIR BREATHING UNIT

CSA Sample Draw Aspirator Adapter Operator s Manual

WARRANTY, QUICK START GUIDE

Overview. Front Panel: Keypad and Display

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors

MODEL CALIBRATION GAS DELIVERY SYSTEM

Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements

Operation Manual. Pro CO Analyzer TM. Pro CO with High Temp Alarm Carbon Monoxide Analyzer

User Instruction Manual

4-in-1 Professional Inflator Kit

Psyclone TM. Protimeter Thermo-Hygrometer. GE Sensing. Operating Manual. INS7800, Rev. A (September 2007)

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Ozone. Sensoric O3 3E 1

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. C

Portable oxygen concentrator

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Model ASSEMBLY and OPERATING INSTRUCTIONS

HI 2314 HI 2315 HI 23151

Return Policy / Procedures

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

MODEL RP050BMST-S1 RESPIRATORY PROTECTOR MANUAL

Nitrogen Supply System

SDM-2012 Docking Station Standalone Configuration Operator s Manual

EC214 EC215 - EC215R Bench Conductivity Meters

GE Measurement & Control

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

210 Series Transmitter with External Electrochemical Sensor

700 SERIES BMS II BUBBLER MONITORING SYSTEM

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

PS-425 Pitot-Static/Air Data Tester

MODEL 100 NITROGEN INFLATION CART

User Manual. PolyGard AT-1130 V3. Electrochemical Nitrogen Dioxide Transmitter. July, 2011

GasSense NDIR User Manual

GAS DETECTION: CALIBRATION GAS

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. K

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL

COMBAT SWIM BOARD. RJE International, Inc. TAC-300 Revision C. RJE International August 3, TAC-300 Revision C

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

RK LEL Sample Draw Aspirator Adapter Operator s Manual

SPECIFICATIONS APCEPH1

RGC-IR Remote Gas Calibrator for IR400

ANALOX O2 PORTABLE OPERATORS MANUAL

Dual-gas Oxygen and Carbon Monoxide SMART Analyzer User Manual

SMART Oxygen Analyzer. User Manual

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

Describe Flammable Gas Measurement

STORM WATER RETENTION TANKS MAINTENANCE INFORMATION FOR HEALTH AND SAFETY FILE

Transcription:

GasLab SAN-102 Micro Oxygen Monitor User Manual The GasLab SAN-102 is a wearable, personal safety meter designed to monitor ambient oxygen levels in real time. It is designed to protect workers in confined spaces. Features Audible and visual low / high oxygen level alarm Silicone boot and metal clip are light and comfortable to wear Electrochemical sensor, no consumable parts Includes ambient temperature and % relative humidity sensors Lithium-ion rechargeable battery. USB Charging port Runs 4+ days per charge Applications Stored Gases in Confined Spaces. Industrial use of stored or piped gases creates the possibility that oxygen in breathable air could be displaced by gases such as nitrogen, helium, argon or carbon dioxide resulting in a hazardous condition. Underground Tunnels, Caves or Mines. Over time oxygen can be displaced by other naturally occurring gases like carbon dioxide, carbon monoxide, ammonia, methane and the oxidation of sulfides resulting in lower oxygen levels than is required for personal safety. Combustion in Confined Spaces. Combustion first uses up the available oxygen, then produces partial oxidation of carbon (carbon monoxide), hydrogen cyanide, and nitrogen oxides. Hydrogen gas can also be produced and displace oxygen. Please take a moment to read these instructions before use. They will provide you with all the necessary information for the correct use of your oxygen monitor. Copyright 2017, CO2Meter, Inc. All Rights Reserved. Last modified March 16, 2018 1

Theory of Operation The SAN-102 Oxygen Monitor uses an electrochemical sensor to determine the concentration of oxygen in air samples. Air reaches the sensor by diffusing through the openings on the back of the unit. Normal air movement is enough to carry the sample to the sensor. Device 1. OLED display 2. Power button 3. Fahrenheit and Celsius switching button 4. Alarm LED 5. Micro USB charge port 6. Vent 7. Clip 8. Silicone boot OLED Display 1. Oxygen concentration 2. Air Temperature ( F or C) 3. % Relative Humidity 4. Oxygen Concentration Units (%) 5. Battery Charge Indicator OPERATION Plug the USB cable into the meter and a 120VAC wall adapter. It will take approximately 2 hours to receive a full charge. Once the meter is charged, it can run 4+ days before recharging. Press the power button to turn on the meter. When the meter is first turned on, it performs a 10 second countdown for device warm up, then enters normal display with current oxygen level readings displayed. The device starts taking measurements when powered on and updates readings every 3 seconds. The OLED display will turn off automatically after 120 seconds of inactivity to conserve battery power. However, the sensor will continue to monitor the O2 level. Briefly press the Power button to view the display. Hold down the Power button for 3 seconds to turn the meter off. This will conserve the battery s charge until you need to use the meter again. Copyright CO2Meter, Inc. All Rights Reserved PAGE 2 of 6

Low and High Oxygen Level Alarm The SAN-102 is equipped with audible and visual alarms to alert you when the ambient oxygen concentration exceeds either of the two factory preset alarm levels: Danger Low Oxygen Alarm: 19.5% - LED will flash and audible alarm will sound 3x / sec. Warning High Oxygen Alarm: 23.5% - LED will flash and audible alarm will sound 2x / sec. Normal atmosphere contains between 20.8% and 21% oxygen. OSHA defines any atmosphere that contains less than 19.5% oxygen as oxygen deficient and any atmosphere that contains more than 23.5 percent oxygen as oxygen enriched. These levels are factory set and cannot be changed. Fahrenheit and Celsius Switching Press the F/ C button to switch between the two temperature units. Press the Power button briefly to save the temperature setting and return to normal display. Low Battery Alarm The meter continuously measures the rechargeable battery s voltage. If it drops to a critical level, the low battery alarm will sound until the battery is recharged or the battery power is depleted. Note that the meter can continue to monitor oxygen levels while charging. ATTENTION! - If a Warning or Danger alarm is triggered while using the instrument as a safety monitor, leave the area and seek fresh air immediately. Remaining on site under such circumstances can cause serious impairment or even lead to death. CLEANING AND STORAGE The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow drying completely before using. Do not use aromatic hydrocarbons or chlorinated solvents for cleaning. Copyright CO2Meter, Inc. All Rights Reserved PAGE 3 of 6

SPECIFICATIONS Oxygen Sensor Specifications Measurement Range 0 to 30.0% vol. Resolution 0.1% vol. Temperature Range -20 to 50 C (-4 F ~ 122 F) Humidity Range 0 to 99%RH Baseline Drift (-10-50 C) <1% change in output @ 3 months Expected Lifespan 24 months Temperature Sensor Specifications Temperature Range -20.0 to 60.0 C (-4 to 140 F) display Display Resolution 0.1 C (0.1 F) Display Options C/ F(switchable) Accuracy ±0.5 C( ±0.9 F) Response Time 5~30 seconds(device must equilibrate with environment) Humidity Sensor Specifications Measurement Range 0.0 to 99.9%RH Display Resolution 1%RH Accuracy ±4.5%RH Response Time <8 seconds for 63% of step change General Operating Storage Power Supply Dimensions Weight -20 C to 50 C (-4 F to 122 F),0~99% RH non-condensing -20 C to 60 C(-4 F to 140 F),<99% RH non-condensing Li-ion battery(4.2v, 400mAh) 60x35x25mm (2.36x1.38x0.98 ) 45 grams (1.58oz.) TROUBLESHOOTING GUIDE Symptom / Issue Possible Cause / Resolution Device does not power ON Reading does not change Make sure the power supply is connected properly and that there is power at the outlet. Use the supplied adapter. Verify that your meter has had enough time to charge before turning it on. In fresh air, you will rarely see oxygen levels change. This is normal. Copyright CO2Meter, Inc. All Rights Reserved PAGE 4 of 6

OXYGEN RECOMMENDED LEVELS > 23.5% Oxygen enriched, presents significant fire and explosion risk 20.8-21% Normal oxygen level in fresh air 19.5% Lowest safe oxygen level limit (OSHA) < 16% Headache, nausea, difficulty performing manual tasks < 12% Difficulty breathing, impaired coordination and perception < 10% Poor judgement and bluish lips < 8% Fatal within minutes if not treated < 6% Fatal in 40 seconds SUPPORT The quickest way to obtain technical support is via email. Please send all support inquires to support@co2meter.com. Please include a clear, concise definition of the problem and any relevant troubleshooting information or steps taken so far, so we can duplicate the problem and quickly respond to your inquiry. WARRANTY This meter comes with a 1 YEAR (warranty period) limited manufacturer s warranty, starting from the date the meter was shipped to the buyer. During this period of time, CO2Meter.com warrants our products to be free from defects in materials and workmanship when used for their intended purpose and agrees to fix or replace (at our discretion) any part or product that fails under normal use. To take advantage of this warranty, the product must be returned to CO2Meter.com at your expense. If, after examination, we determine the product is defective, we will repair or replace it at no additional cost to you. This warranty does not cover any products that have been subjected to misuse, neglect, accident, modifications or repairs by you or by a third party. No employee or reseller of CO2Meter.com s products may alter this warranty verbally or in writing. LIABILITY All liabilities under this agreement shall be limited to the actual cost of the product paid to CO2Meter.com. In no event shall CO2Meter.com be liable for any incidental or consequential damages, lost profits, loss of time, lost sales or loss or damage to data, injury to person or personal property or any other indirect damages as the result of use of our products. Copyright CO2Meter, Inc. All Rights Reserved PAGE 5 of 6

RETURNS If the product fails under normal use during the warranty period, a RMA (Return Material Authorization) number must be obtained from CO2Meter.com. After the item is received CO2Meter.com will repair or replace the item at our discretion. To obtain a RMA number, call us at (386) 256-4910 or email us at support@co2meter.com. When requesting a RMA please provide reason for return and the original order number. If we determine that the product failed because of improper use (water damage, dropping, tampering, electrical damage etc.), or if it is beyond the warranty date, we will inform you of the cost to fix or replace the product. For additional warranty information visit our website: www.co2meter.com/pages/faq CONTACT US We are here to help! For information or technical support, please contact us. support@co2meter.com (386) 256-4910 (Technical Support) (386) 872-7665 (Sales) www.co2meter.com CO2Meter, Inc. 131 Business Center Drive Ormond Beach, FL 32174 USA Copyright CO2Meter, Inc. All Rights Reserved PAGE 6 of 6