Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Similar documents
QUICK REFERENCE GUIDE

Life Pulse High Frequency Ventilator In-Service Manual

HIGH FREQUENCY JET VENTILATION (HFJV): EQUIPMENT PREPRATION

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

AUTOVENT 4000 VENTILATOR

3100A Competency Exam

Mechanical ven3la3on. Neonatal Mechanical Ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on 8/25/11. What we need to do"

SLE5000 Infant Ventilator with HFO

Pressure Controlled Modes of Mechanical Ventilation

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Technical Data and Specifications

Automatic Transport Ventilator

Ventilator Training Module

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Neonatal tidal volume targeted ventilation

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

Mechanical Ventilation

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

Mechanical Ventilation

2015 Hamilton Medical AG. All rights reserved. Printed in Switzerland.

Emergency Transport and Ventilation

Neonatal Assisted Ventilation. Haresh Modi, M.D. Aspirus Wausau Hospital, Wausau, WI.

Hospital and Transport for Controlled Breathing

Operating Instructions for Microprocessor Controlled Ventilators

The Basics of Ventilator Management. Overview. How we breath 3/23/2019

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

Hallowell EMC Anesthesia WorkStation Operating Manual

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

Unit 15 Manual Resuscitators

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

Operation and Maintenance of the EPV200 Portable Ventilator

Principles of Mechanical Ventilation: A Graphics-Based Approach

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Mechanical Ventilation


Introduction to Conventional Ventilation

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Medical Instruments in the Developing World

RESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02


How to use the SAVe II Ventilator

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

GE Healthcare. Centiva/5 Critical Care Ventilator. Meet a new level of expectations

Draeger Fabius GS Pre-Use Check

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

ASSISTANCE. Cardinal Health Pulmonetic Systems Medina Rd., Suite 100 Minneapolis, Minnesota

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

How Ventilators Work. Chapter 3

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

IMPACT. Uni-Vent. Eagle. Model 754 Portable Ventilator

SLE5000 Neonatal Ventilator with High Frequency Oscillation

Classification of Mechanical Ventilators

Astral in AirView: Improving patient care through connectivity. ResMed.com

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

IEC C LASSIFICATIONS:

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

S/5 Aespire. Quick Reference Guide

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

ONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg

VENTILATORS PURPOSE OBJECTIVES

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

Guide to Understand Mechanical Ventilation Waveforms

Flight Medical presents the F60

Astral in AirView: Improving patient care through connectivity. ResMed.com

Recommendations on Checking Anaesthesia Delivery Systems

carefusion.com CareFusion Yorba Linda, CA CareFusion

O-Two Self-Study Guide. Equinox Relieve

Complex Care Hub Manual: Pressure Ventilation on the LTV 1200

Volume Diffusion Respiration (VDR)

Selecting the Ventilator and the Mode. Chapter 6

e700 Automatic Transport Ventilator User Manual e700 User Manual (15PL Rev.14 28_8_2017)

Emergency Medicine High Velocity Nasal Insufflation (Hi-VNI) VAPOTHERM POCKET GUIDE

Flow meter. bellow vaporizer. APL valve. Scavenging system

Intensa GO Alternate Care Ventilator

Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator

fabian Therapy evolution Our 2-in-1 device

Using Common Ventilator Graphics to Provide Optimal Ventilation

Simplified Automated Ventilator Training Slides

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm


patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

excellence in care Procedure Management of patients with difficult oxygenation. For Review Aug 2015

fabian HFO Our 4-in-1 device

Welcome to the Specialized Medical Services Respiratory training webinar series!

Airox Supportair Ventilator

CCAT Mechanical Ventilation Clinical Practice Guideline

Other Oxygen Delivery Systems

Accumulation of EEV Barotrauma Affect hemodynamic Hypoxemia Hypercapnia Increase WOB Unable to trigger MV

Objectives. Respiratory Failure : Challenging Cases in Mechanical Ventilation. EM Knows Respiratory Failure!

Transcription:

Bunnell Incorporated n www.bunl.com n 800-800-4358 (HFJV) n info@bunl.com 436 Lawndale Drive n Salt Lake City, Utah 84115 n intl 801-467-0800 n f 801-467-0867 Bunnell LifePulse HFV Quick Reference Guide # 03202-00.1 2017 Bunnell Incorporated

Initial Ventilator Setup 1. Provide Electrical Power: Plug the LifePulse into a hospital grade electrical outlet in order to power it on. 2. Provide Gas Source: Connect the LifePulse to the air/ oxygen supply from a low flow blender (0-30 L/min.) or from the low flow output (2-100 L/min.) of a standard blender. 3. T urn on LifePulse: Toggle the Ventilator Power switch on the rear panel to the ON position. Press AUDIO PAUSED to cancel the high priority power on audio and visual alarm. OFF ALARM VOLUME 3 ON VENTILATOR POWER 20A QUARTZ HOURS 00000 0 1/10 (206.85-403.70 kpa) DO NOT BLOCK BATTERY ACCESS REFER TO USER MANUAL CIRCUIT BREAKER 2A MIXED GAS INPUT 30-60 PSI OXYGEN SAMPLE PORT ON ON 2 OFF DO NOT BLOCK 1 100-240/ 2A 50-60 HZ MONITOR 00.0 00.0 00.0 00.0 00.0 - SERVO PIP PEEP cm H2O CONTROLS 20. PIP CANNOT MEET PIP LOWER LIMIT HIGH PIP BATTERY DEPLETED cm H2O cm H2O MAP - cm H2O LifePulse - BPM HUMIDIFIER PAUSE 420 RATE SYSTEM TEST LOW GAS PRESS HIGH FREQUENCY VENTILATOR C o - STANDBY CHECK VENT VENT INOP UPPER LIMIT LOWER LIMIT.020 I-TIME SECONDS - ENTER I : 6.I I:E RATIO ALARMS TEMP 3 LOSS OF PIP UPPER LIMIT cm H2O ΔP AUDIO PAUSED ALARMS READY CHECK CIRCUIT GAS WATER HIGH TEMP LOW LEVEL bunnell INSPIRED INFANT CARE 1

Installing a Circuit WARNING: Latch the water inlet tube on the humidifier cartridge of the Patient Breathing Circuit into the water pump housing prior to connecting to the water supply. Failure to do so may result in cartridge overfill and delivery of water to the patient by gravity feed. WARNING: Clamp the water supply tube prior to opening the water pump door to prevent cartridge overfill and delivery of water to the patient by gravity feed. The water supply should be positioned at or below the level of the humidifier cartridge as an added precaution. A Patient Circuit must be installed in order to use the LifePulse. Follow these steps to install the Patient Circuit: 1. Open cartridge door and insert humidifier cartridge into receptacle 2. Open water pump door by lifting up. Secure the water inlet tube inside the water pump and snap the pump door closed. 3. Attach the green gas inlet tube to the green-coded Gas Out port on the LifePulse. 1 HUMIDIFIER 88.8 - PAUSE TEMPERATURE GAS WATER 4. Attach the purge tube to the yellow-coded Purge port on the LifePulse and to the yellow-coded port on the Patient Box. 3 Life Pulse HIGH FREQUENCY VENTILATOR GAS OUT PURGE bunnell INSPIRED INFANT CARE 5. Press PUSH TO LOAD on top of the Patient Box and gently stretch the pinch tube into the jaws of the pinch valve. PATIENT BOX 6. Connect the clear pressure monitoring tube from a LifePort adapter to its port on the Patient Box. BATTERY _ 2 7. Insert the green delivery tube into the Jet port on the LifePort adapter. 8. Attach the water transfer tube to the water inlet tube and spike the water supply. The water supply should be at or below the level of the humidifier cartridge. 9. Unclamp the water transfer tube. 4 8 Water Supply 9 5 7 4 6 2

Testing and Preparation 1. Perform System Test: Attach a LifePort adapter and ET tube to a test lung. Press SYSTEM TEST and verify the LifePulse runs through the test sequence and returns to Standby mode with an audible and visual alarm active. Press AUDIO PAUSED to cancel the high priority passed test alarm. WARNING: Patient connections must only be made in the Standby mode. Do not connect the LifePulse Patient Breathing Circuit to the LifePort adapter on the patient s ET tube while the LifePulse is running. Failure to comply risks high pressures and volumes being delivered to the patient, which may result in volutrauma. 2. Perform Operational Test: Perform an Operational Test. Press ENTER to activate the default settings (PIP: 20, Rate: 420, I-Time: 0.020) while attached to the test lung. Verify that the READY indicator illuminates, the monitored PIP reaches the set PIP, and PEEP is 0.0 ± 1.0 cm H 2 O when no PEEP is supplied by the conventional ventilator. 3. Place LifePulse into Standby: Place the LifePulse into Standby mode after the tests are completed and prior to connecting to the patient. 4. Before Connecting to Patient, Remove and Save Test Supplies: Remove the Test Supplies (LifePort, ET tube, and test lung) from the LifePulse circuit and Patient Box. Save for future use. 3

Beginning High Frequency Jet Ventilation WARNING: Do not leave the patient s bedside while the LifePulse water pump is running during initial start-up or following a circuit change. A properly trained person must observe the cartridge fill with sterile water for inhalation, USP, to the second water level sensing pin and the water pump stop pumping. All infants treated with the LifePulse should be connected to a conventional ventilator and appropriate physiologic monitors. They must be intubated with a standard ET tube with a Bunnell LifePort adapter attached. 1. Attach LifePort adapter to patient s ET tube: Replace standard ET tube adapter with appropriate size LifePort adapter. Connect pressure monitoring tube to port on Patient Box. 2. Connect LifePulse circuit to patient s LifePort adapter: Remove the cap on the jet port of the LifePort adapter and connect the patient end of the LifePulse circuit to the jet port. 2nd Pin 3. Set HFJV PIP: Adjust the set HFJV PIP to equal the monitored CV PIP (see Ventilation and Oxygenation flow charts on pages 6 and 7). 4. Set HFJV Rate: Adjust the set HFJV Rate to the frequency appropriate for the patient (420 bpm is recommended, or slower for larger patients or to address hyperinflation). 5. Set HFJV I-Time: Starting I-time is usually left at the default 0.020 seconds. WARNING: There will be no LOSS OF PIP alarm for the first 15 seconds after the ENTER Button is pushed. A properly trained person must observe the LifePulse reach the set PIP and have the Ready indicator illuminate before leaving the patient s bedside. 6. Begin High Frequency Jet Ventilation: Press ENTER to activate the Control settings. The PIP will quickly rise toward the set PIP. The LifePulse may pause momentarily when a conventional breath is delivered. 7. Eliminate LifePulse Interruptions: If the LifePulse is pausing, reduce the CV PIP until the pauses go away (CV PIP < Set HFJV PIP). 8. Lower CV Rate: Lower the CV rate to between 0 and 5 bpm. (Use 0 bpm if air leaks are the primary concern. Use 1-5 bpm and/or increase PEEP if atelectasis and oxygenation are the main concern.) 9. Assure Ready Indicator is On: Verify the Ready indicator illuminates, which indicates the pressures are stable and the LifePulse alarms are active. 10. Adjust PEEP: Adjust the CV PEEP setting to obtain the desired MAP and oxygenation. 11. Reassess Blood Gases: Observe blood gas monitored values and/or draw arterial blood gas samples after 30 minutes to determine if ventilator adjustments are necessary. 4

GENERAL RULES HFJV ΔP (PIP - PEEP) is the primary determinant of PaCO 2. HFJV I-time and Rate are secondary. Resting lung volume (FRC supported by set PEEP) and mean airway pressure (MAP) are crucial determinants of PaO 2. Avoid hypercarbia and hypoxemia by using optimal PEEP (see When to Raise PEEP below). Minimize IMV at all times, using very low rates (typically 0 5 bpm), unless IMV is being used to recruit lung volume or stabilize FRC. In general, keep CV PIP at a level necessary to achieve a moderate chest rise. To overcome atelectasis, IMV rates up to 5 bpm can be used for 10 30 minutes. Thereafter, IMV rate should be dropped back to as close to 0 as possible. If lowering CV rate worsens oxygenation, PEEP may be too low. Higher PEEPs and lower CV rates reduce risk of lung injury. Lower FiO 2 before PEEP when weaning until FiO 2 is less than 0.4. Patient Management During HFJV SETTING INITIAL WHEN TO RAISE WHEN TO LOWER HFJV PIP Whatever produces To increase PaCO To decrease PaCO 2 (Raise PEEP if necessary to keep SpO 2 desired PaCO 2 2 constant) HFJV Rate 420 bpm (neonates) 300 bpm (peds) To decrease PaCO 2 in smaller patients with low compliance To eliminate inadvertent PEEP or hyperinflation by lengthening exhalation time or to increase PaCO 2 when weaning HFJV I-TIme 0.020 seconds To increase delivered tidal volume and lower PaCO 2 0.020 is the minimum CV Rate 0-5 bpm To reverse atelectasis as a temporary recruitment maneuver (3 5 bpm) To minimize volutrauma, especially when air leaks are present, or to decrease hemodynamic compromise CV PIP PIP necessary to achieve moderate chest rise To reverse atelectasis or stabilize lung volume; PIP typically < HFJV PIP To minimize volutrauma, especially when air leaks are present, or to decrease hemodynamic compromise CV I-Time 0.4 seconds To reverse atelectasis or stabilize lung volume To minimize volutrauma, especially when air leaks are present, or to decrease hemodynamic compromise PEEP 7 12 cm H 2 O (Neonates) 10 15 cm H 2 O (Peds) To improve oxygenation and decrease hyper-ventilation To find optimal PEEP: Raise PEEP until SpO 2 stays constant when switching from IMV to CPAP Lower PEEP only when it appears that cardiac output is being compromised; or when oxygenation is adequate FiO 2 < 0.4, and when lowering PEEP doesn t decrease PaO 2 FiO 2 As needed Raise as needed after optimizing PEEP Lower FiO 2 in preference to PEEP when weaning until FiO 2 < 0.4 Special Air Leak Considerations: 1. Minimize IMV by using HFJV and adequate CPAP. 2. If oxygenation is compromised AND expiratory time has been optimized, raise PEEP, even if the lungs appear to be overexpanded on x-ray. 5

Ventilation Control Flow Chart Hypocarbia PaCO 2 too Low? PaCO 2 Hypercarbia PaCO 2 too High? To raise PaCO 2 try the following in order: 1. decrease HFJV PIP 2. decrease HFJV Rate 3. decrease HFJV I-time 4. increase CV PEEP To lower PaCO 2 try the following in order: 1. increase HFJV PIP 2. increase HFJV I-time 3. increase HFJV Rate* 4. decrease CV PEEP ** PaCO 2 too High? PaCO 2 too Low? PaO 2 too Low? Go to Oxygenation Flow Chart PaO 2 too Low? PaO 2 too High? Decrease FiO 2 until < 0.40 then decrease CV PEEP PaO 2 too High? Maintain Current Settings * Increasing HFJV Rate increases minute ventilation. However, if lungs are hyperinflated, decreasing HFJV Rate can lower PaCO 2 by increasing exhalation time. ** Decreasing CV PEEP increases Δ pressure (amplitude) and lowers PaCO 2, but it will also lower MAP, which may lower PaO 2. 6

Oxygenation Flow Chart Underinflation or Atelectasis? Hypoxemia PaO 2 is too Low Overinflation or P.I.E./Air leak? To raise MAP & PaO 2 try the following in order: 1. increase CV PEEP 2. increase CV Rate (3-5 bpm) 3. increase CV PIP 4. increase CV I-time 5. increase FiO 2 To decrease gas trapping & raise PaO 2 try the following in order: 1. decrease CV Rate 2. decrease HFJV Rate * (60 bpm at a time) 3. decrease HFJV PIP ** 4. decrease CV PEEP *** 5. increase FiO 2 PaC O 2 too High? PaCO 2 too High? Pa CO 2 too Low? Go to Ventilation Flow Chart Pa CO 2 too Low? PaO 2 too High? Decrease FiO 2 until < 0.40 then decrease CV PEEP PaO 2 too High? Maintain Current Settings * Decreasing HFJV Rate decreases minute ventilation. It may also lower PaCO 2 by increasing exhalation time. ** Decreasing HFJV PIP decreases pressure (amplitude) and minute ventilation ; PaCO 2 may increase.. *** Decreasing CV PEEP increases pressure (amplitude) and decreases MAP ; PaCO 2 and PaO 2 may decrease. 7

Optimizing MAP During HFJV MAP may be too low if FiO 2 > 0.5 or SpO 2 < 85% Underinflation or Atelectasis? In rare situations in which PEEP may be excessive: PIE/AIrleak or normal inflation? Make sure to provide 3-5 CMV breaths Uniform overinflation or compromised BP despite lower HFJV rates (240-300) and CV is set to CPAP? Increase MAP by raising PEEP 1-2 cm H 20 Increase MAP by raising PEEP 1-2 cm H 20 Turn CMV to CPAP Decrease MAP by lowering PEEP 1-2 cm H 20 SpO 2 improves in 1-15 minutes? NO YES SpO 2 or BP improves in 5-15 minutes? YES YES SpO 2 drops in 1-5 minutes? NO SaO 2 drops? YES Maintain MAP and work on lowering FiO 2 Titrating MAP up or down may improve oxygenation. Use your understanding of the interactions of the cardiopulmonary system to achieve optimal results. 8

The Importance of Servo Servo = driving pressure that automatically regulates flow. Servo changes with each change in HFJV or CV settings. Servo changes as lung volume or mechanics change. Servo Increases with: Improving compliance or resistance Loose tubing connections Moisture interference in LifePort Mechanical issues with LifePulse Servo Decreases with: Worsening compliance or resistance Patient needs suctioning Obstructed ET tube Tension pneumothorax Right mainstem intubation Servo changes are early indications of changes in the patient s condition or the LifePulse s performance (assuming HFJV and CV settings have not changed). Charting and trending Servo can help inform patient management decisions. 9

Weaning Patient from LifePulse 1. Lower HFJV PIP Slowly Avoid lowering PEEP/MAP until FiO 2 is < 0.40. Maintain HFJV Rate. Reduce the HFJV PIP (1-2 cm H 2 O) as needed per PCO 2. 2. Lower HFJV and CV PIP to Teens Continue to reduce the HFJV PIP while keeping the Rate constant. Reduce the CV PIP as needed to avoid interrupting the LifePulse PIP. Maintain appropriate MAP (may need to increase PEEP). If you are weaning to CV rather than NIV, increase the CV Rate in response to PCO 2 once the HFJV PIP is < 18 cm H 2 O. Interrupting the LifePulse breaths with the CV breaths may now be appropriate. 3. Evaluate Patient During CV or NIV Trial Observe the patient and monitored parameters to ensure patient is tolerating the weaning process. Increase the CV Rate or NIV support, if necessary, after the LifePulse is in Standby mode. If CV PIP has to be increased to > 18 cm H 2 O, the trial is a failure and the patient should be returned to the LifePulse. NOTE: During storage, plug the LifePulse into an electrical wall outlet in a properly ventilated room in order to maintain battery charge. See the LifePulse User Manual for more information on cleaning and storage. 4. Consider NIV Trial Consider NIV trial if HFJV PIP < 18 cm H 2 O, MAP 8 cm H 2 O, and FiO 2 is 0.40. Restart the LifePulse if the trial is unsuccessful. Try again in 12-24 hours. 10