Technical Information Conducal CLY421

Similar documents
CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

Technical Information Cleanfit CUA451

Operating Instructions Flowfit CUA252

Oxymax COS61/COS61D. Technical Information

Pneumatic high-pressure controller Model CPC7000

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

Indumax CLS54D. Operating Instructions

Precision pressure sensor Basic version Model CPT6020

Hand-held pressure calibrator with integrated pump Model CPH6600

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Pressure Measurement Single-range transmitters for general applications

High-performance submersible pressure transmitter For level measurement Model LH-10

High-performance submersible pressure transmitter For level measurement Model LH-10

Pneumatic high-pressure controller Model CPC7000

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

EPP4 Pressure Regulator Basic G 1/4"

Armatures for analytical probes

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

Glass tube flowmeters Series 2000

Operating Instructions Turbimax CUS52D


Mobile High Pressure Compressor Unit for Compressing Air and Breathing Air. PE 250-TE with compressor control (optional equipment)

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition

Digital pressure gauge Model CPG500

Improve Process Reliability

High-performance submersible pressure transmitter For level measurement Model LH-20

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

High-performance submersible pressure transmitter For level measurement Model LH-20

OUS31. Technical Information

Columbus Instruments

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Mass Flow Controller (MFC) for Gases

ph measuring system for hygienic process applications

Model PDT Dewpoint Transmitter

SAUTER FK Version /2016 GB

Mass Flow Meter (MFM) for gases

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

FPG8601 Force Balanced Piston Gauge

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

SITRANS P measuring instruments for pressure

Mass Flow Controller (MFC) for Gases

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

Pressure Measurement Single-range transmitters for general applications

Oxymax H COS21D. Technical Information

JUMO dtrans p30 Pressure Transmitter

SonoMeter 30 Energy Meters

AMT-Ex Dewpoint Transmitter

Digital pressure gauge Model CPG500

High-performance submersible pressure transmitter For level measurement Model LH-20

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

RAM 4021-DPX Operation Manual

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

Mass Flow Meter (MFM) for Gases

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

KTE6000 / KTU6000 Series OEM pressure transmitters for industrial media

Micro Motion 7812 Fiscal Gas Density Meter

Hydraulic hand spindle pump Models CPP1000-M, CPP1000-L

SonoSelect 10 and SonoSafe 10 energy meters

Pressure Transmitter P41

High-speed pressure sensor Model CPT6140

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

CCT-7320/ROC-2313 Reverse Osmosis Controller

Supplementary Operator s Manual 42/24-14 EN Rev. 3

BERMAD Waterworks. Level Control Valve with Altitude Pilot. 700 Series. Model X. Features and Benefits. Major Additional Features

SDX Submersible Depth Transmitter User Manual

MANUAL KPS Pressure Control Valve

2/2- and 3/2-way Rocker Solenoid Valve - bistable (impulse) version

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Glass tube flowmeters Series 2000 Variable area flowmeter for low flows of liquids and gases

ABB MEASUREMENT & ANALYTICS DATA SHEET. WaterMaster FEW530 Electromagnetic water meter

ELADOS EDP L Diaphragm Metering Pump Type

C-Flow Coriolis Mass Flow Meters

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Additel 761 Automated Pressure Calibrators

Additel 780 Series Pressure Controller

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

Digital deadweight tester Model CPD8500

Gas density monitor With integrated transmitter Model GDM-100-TI

Instruction and Maintenance Manual

water purification cylinders & accessories

Operating Instructions Part No

Additel 780 Series Pressure Controller

Tightness controls TC 1 3 and TC 4

Simple Set. Pressure Independent Control Valves 2 Way 1/2-2

Operating Instructions Part No

Instruction Manual. Digital Thermocouple Gauge Control Units Models DTC & DTC Models DTC-06M-115 & DTC-06M-230

Variable Area Flow Meters

SKP Series SKP25 U.. Pressure Regulating Gas Valve Actuator with Safety Shut-off Function

How to specify a product. Process Sensors and Mechanical Instruments

Digital dead-weight tester Model CPD8500

Industrial pressure controller Model CPC4000

Water Temperature Control - Groups of Fixtures

Sanitary pneumatic retractable fitting for use with Unical 9000 and Uniclean 900 automated control systems

Transcription:

TI00496C/07/EN/01.12 71142150 Products Solutions Services Technical Information Conducal CLY421 Conductivity calibration set for ultrapure water applications Factory calibrated measurement box Certified for comparison measurement in ultrapure water up to 20 μs/cm Application Ultrapure water is used in sensitive areas of the pharmaceutical and food industries and in process engineering. Calibration of process conductivity measurement is obligatory for these applications. Since calibration solutions with very low conductivity are unstable, there is no standardization for conductivities below 74 μs/cm. Conducal is a reference unit that permits the calibration of process measuring devices by means of a certified comparison measurement. Areas of application Calibration of quality-relevant process measurements in the ultrapure water range up to 20 μs/cm Checking of in-line measurements in the process, e.g. after inspections or interruptions in operation Checking of conductivity measuring equipment primarily in the pharmaceutical and food industries Checking and calibration for product quality assurance, e.g. in semiconductor production Your benefits Factory calibration traceable to SRM by NIST and DAkkS Factory calibration of Conducal system according to ASTM D-5391-93 Flow assembly with adjusting and monitoring functions according to ASTM D-5391 Flexible applicable in the field thanks to: - battery-driven system (independent of power network) - robust case IP67 with external power connection

Function and system design Measuring arrangements Comparison measurement with a bypass (preferred) Measuring system for bypass comparison measurement a0014022 1 Process transmitter 5 Bypass outlet 2 Process conductivity sensor 6 Flow assembly calibration set 3 Transmitter calibration set 7 Bypass inlet 4 Conductivity sensor calibration set 8 Sterile main pipe With this arrangement, it is important to make sure that the composition of the medium and the temperature at the process measurement point and comparison measurement point are the same. This is ensured by: Using short hose connections Waiting until the temperature in the flow assembly adjusts to match the process temperature. Advantage: Process stays hygienic Disadvantage: Waiting time for temperature harmonization 2 Endress+Hauser

Direct comparison measurement Measuring system for direct comparison measurement a0014023 1 Process transmitter process 5 Outlet 2 Process conductivity sensor 6 Flow assembly calibration set 3 Conductivity sensor calibration set 7 Inlet 4 Transmitter calibration set The flow assembly included with the calibration set has two mounting positions for conductivity sensors. This allows a direct comparison measurement. Advantage: No temperature difference, that means no waiting time Absolutely identical medium Disadvantage As it is necessary to remove the sensor from the process, the medium can be contaminated. Endress+Hauser 3

Flow assembly The Conducal calibration set contains a special flow assembly for the conductivity sensor CLS15D. 1 2 3 8 7 4 6 5 Flow assembly a0014030 1 Switch over from bypass (1) to direct (2) 5 Inlet for direct comparison measurement 2 Installation point for calibration set conductivity 6 Flow control valve sensor 3 Installation point for process conductivity sensor 7 Outlet 4 Inlet for bypass measurement 8 Flow monitoring Bubble trap To prevent corrupted conductivity measurement the medium has to be bubble free. Outgassing can occur when the medium relaxes tension. That means also in this bypass arrangement. Thanks to the integrated bubble trap the medium around the conductivity sensor stays bubble free. 1 2 3 4 Medium feeding at comparison measurement with bypass 1 Vent hole 2 Inlet 3 Second installation place 4 Lateral bore hole a0016297 When the flow assembly is used in the bypass mode, the second installation place is used as hydrocyclone. The tangential medium inlet induces a turbulent flow. The bubbles can exhaust upwards through the vent hole. The medium is fed to the measurement chamber via the lateral bore hole. 4 Endress+Hauser

Controlled conditions with flow monitoring For the calibration with Conducal according to ASTM D-5391, a minimum flow specified by the sensor manufacturer has to be kept. The flow assembly of the calibration set is designed according to the standard specifications of the conductivity calibration. Bubble trap (at the bypass measurement) Float-type flow meter Flow adjustment with control valve Temperature resistant up to 100 C (210 F) 2 3 4 5 1 6 Schematic of bypass arrangement a0014031 1 Inlet G½ 4 Flow adjustment control valve 2 Venting 5 Flow monitoring 3 Conductivity sensor calibration set 6 Outlet G½ Carrier for pipe mounting The flow assembly is equipped with a flexible carrier system for mounting at round and rectangular pipes. This makes comparison measurement in the field safe and easy. Carrier a0014036 Endress+Hauser 5

Calibration Flow chart a0015935-en 6 Endress+Hauser

Input Measured variables μs/cm or M cm; configurable Power supply Supply voltage Rechargeable battery External sensor cable connection Wide-area power pack 100 to 240 VAC, 47 to 63 Hz, Class II equipment with functional grounding Integrated lithium ion battery 14.4 V; 2.4 Ah The fully charged battery enables a calibration set operating time of over 80 hours. Buccaneer plug, 6-pin, IP 68 Performance characteristics Error calculation Reference system adjustment with standard NIST reference material Uncertainty of reference solution 0.2 % Uncertainty of temperature measurement << 0.1 % Uncertainty of reference system display 0.2 % Total uncertainty of reference system adjustment 0.3 % Conducal adjustment with 5 μs/cm (or 200 k cm) Uncertainty of reference system adjustment 0.3 % Uncertainty of reference system measurement at 5 μs/cm 0.6 % Uncertainty of Conducal display at 5 μs/cm 0.6 % Total uncertainty of Conducal adjustment with 5 μs/cm 0.9 % (Only corresponds to the uncertainty of Conducal. The adjustment of measuring points with Conducal requires an additional uncertainty analysis) The change of the CLS15D cell constant in the conductivity range between the standard reference material and 5 μs/cm is not considered. Depending on availability, different standard NIST reference materials are used to adjust the reference system. Normally SRM 3191 is used with a nominal value of 100 μs/cm. Reference devices Reference measuring device used Reference measuring cell used Liquiline CM42 Condumax CLS15D Endress+Hauser 7

Environment Ambient and storage temperature range 5 to 40 C (41 to 104 F) Humidity max. 80 % Altitude Degree of protection Up to 2000 m IP 30 with case open IP 67 at closed case without power cord Use indoors (pollution level II) Process Process temperature Process pressure Minimum flow 0 to 100 C (32 to 210 F) max. 6 bar (87 psi) 30 l/h (8 gal/hr) Mechanical construction Dimensions L x W x H 530 x 442 x 215 mm (20.9" x 17.4" x 8.5") Weight Approx. 12.7 kg (28 lbs) Materials Flow assembly Clamp seal Adapter PVDF EPDM PVDF Process connections Inlet G½ or Clamp ½" Outlet G½ or Clamp ½" Vent G½ 8 Endress+Hauser

Certificates and approvals Electromagnetic compatibility Factory calibration certificate Interference emission and interference immunity complies with EN 61326: 2005, Namur NE 21:2007 The Conducal CLY421 calibration set is calibrated and adjusted at the factory. With the aid of a reference system, the CLS15D reference sensor is calibrated precisely in accordance with the specifications of ASTM Standard D-5391-93. In this way, the cell constant and temperature offset of the sensor are determined and recorded in the factory calibration certificate. The certificate, which is created for each individual Conducal CLY421 kit, is enclosed in the calibration set. Removing covers or opening the housing results in termination of the factory calibration. Ordering information Product structure You can create a valid and complete order code with the configurator on the internet. To get to the product page enter the following address into your browser: www.products.endress.com/cly421 1. You can choose from the following options on the product page located on the right: 2. Click "Configure this product". 3. The configurator opens in a separate window. You can now configure your device and receive the complete order code that applies for the device. 4. Afterwards, export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of the page. Scope of delivery The delivery comprises: 1 Calibration set in the version ordered 1 Set of Operating Instructions for Conducal CLY421 1 Certificate of calibration Endress+Hauser 9

Accessories Calibration service Recalibration Depending on the operating conditions and the frequency of use, the conductivity calibration set must be calibrated regularly at the manufacturer's factory. Recommended recalibration interval: 1 year 10 Endress+Hauser

Endress+Hauser 11