Operation Manual. LB device KSV Revision 1.1

Similar documents
Complete Range of Force Tensiometers

RIPPLE TANK - with rippler & kit

INSTRUCTION MANUAL. January 23, 2003, Revision 0

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Chemistry - Scale-up F.32. Lara TM - Controlled - Lab Reactor. The Concept

HI 2314 HI 2315 HI 23151

Basic Flows in a Microfluidic Device

Tensiometer For the measurement of surface and interfacial tension

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Vibration isolation system 1VIS10W. User manual

Tensiometer For the measurement of surface and interfacial tension

Vapor Pressure of Liquids

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

SOP for use of the SX19 Stopped-Flow Instrument

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

AUTOMATIC VAPOR PRESSURE TESTER REID METHOD, DEMI SIZE MODEL AVP-30 INSTRUCTION MANUAL. Rev

EC214 EC215 - EC215R Bench Conductivity Meters

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

QUICK START GUIDE TO DRINKING WATER MONITORING

Heat Engine. Reading: Appropriate sections for first, second law of thermodynamics, and PV diagrams.

SERIES 20 CHAMBERS THE RC-25 AND RC-25F CHAMBERS ASSEMBLY

Vibration isolated tables. Individually tailored to your requirements.

The SPI Sputter Coater Handbook

VACUUM REGULATORS CONTENTS

Using the Akta Prime plus October 22, 2012

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

High Pressure Chem-SCAN Operating Manual

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

UsER manual for Watersens ph -REDOX

MPG-plus. Application example. Pneumatic 2-Finger Parallel Gripper Gripper for small components. Gripping force 38 N 175 N. Sizes

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

TJF-Q180V Cleaning and Disinfection Checklist

ACV-10 Automatic Control Valve

Vapor Pressure of Liquids

VISUAL PHYSICS School of Physics University of Sydney Australia. Why can a steel needle float but a larger piece of steel sink?

[USING THE VITROBOT April 9, Using the Vitrobot

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

STERILIZABLE DISSOLVED OXYGEN SENSORS

Cell Disruption System

215 Liquid Handler Up-Grade Manual

Heavy Duty Dissolved Oxygen Meter

Order No Projection Backboard

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Vapor Pressure of Liquids

#LZ400 LEAKALYZER. Water Loss Sensor INSTRUCTION MANUAL Country Dr. #190 St. Paul, MN

OPERATION MANUAL DIVE CALIPER NOT SCALE / IF IN DOUBT ASK* *ALL UNITS IN MM/DEGREES UNLESS SPECIF. Dive Caliper General DC-051

FAST EVAPORATION AT NORMAL PRESSURE

PRECISION DISSOLVED OXYGEN MEASUREMENT SYSTEMS

MoveRoll Conveyor Operating and Maintenance Manual

Vapor Pressure of Liquids

Operating Instructions Part No

Angstrom Dielectric Sputterer Operation Manual

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

Oxidation Stability of Gasoline and Aviation Fuels

Attension Precision made simple

SERIES 20 CHAMBERS THE RC-26, RC-26P AND RC-26G CHAMBERS ASSEMBLY

TEMPERATURE S RELATIONSHIP TO GAS & VAPOR PRESSURE

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Technical Data Sheet MF010-O-LC

Technical Note. Determining the surface tension of liquids by measurements on pendant drops

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Plasma Cleaner. Yamato Scientific America. Contents. Innovating Science for Over 125 Years. Gas Plasma Dry Cleaner PDC200/210/510 PDC610G.

Aerobic Respiration. Evaluation copy

YSC-8330 USER MANUAL XI AN TYPICAL INDUSTRIES CO.,LTD.

BUYER S GUIDE AQUAlogger 530WTD

Warnings: Notes: Revised: January 8,

10.2. Basketball Single Post System

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

TJF-Q180V Cleaning and Disinfection Checklist

Customer Responsibilities

Product information. Capacitive. Level detection in liquid VEGACAP 62 VEGACAP 63 VEGACAP 64 VEGACAP 66 VEGACAP 69. Document ID: 29983

Unisense In Situ UniAmp Connector System

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

7.9. Flash Column Chromatography Guide

DPZ-plus. Application example. Pneumatic 3-Finger Centric Grippers Sealed Grippers. Gripping force 520 N N. Sizes

Intermolecular Forces

1)! DO NOT PROCEED BEYOND THIS MARK

SCHEINWORKS Measuring and Analysis Systems by

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Rev. B Sept 06. VD510 INSTRUCTION MANUAL Part # DISPENSING VALVE MODEL VD510 INSTRUCTION MANUAL

MTS Restan. Automatic System for Residual Stress Measurement by Hole-Drilling

G1/8 G1/8. Model 0 bar -0,1 bar -0,2 bar -0,3 bar -0,4 bar -0,5 bar -0,6 bar -0,7 bar -0,8 bar

TR Electronic Pressure Regulator. User s Manual

MV5: Automated concentration system

Accurate, easy to use and reliable: Easy to use control panel. Clear, backlit LCD display: Quick connect spindles:

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Microfluidic Demonstration

PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS

Easy with a modular concept for future expansion

PFH. Application example. Pneumatic 2-Finger Parallel Grippers Long-stroke Grippers. Sizes Gripping force 510 N N

DETRMINATION OF A PLUNGER TYPE WAVE MAKER CHARACTERISTICE IN A TOWING TANK

HumiSys HF High Flow RH Generator

Cover Page for Lab Report Group Portion. Pump Performance

MODEL ANALOG ALTIMETER USER'S MANUAL 330KHZ, 1000M DEPTH RATED 0.2 TO 100FT OPERATING RANGE ANALOG OUTPUT

Microscopy Cryogenic WORKSTATION. Performance by Design

PSH (inch version) Application example. Pneumatic 2-Finger Parallel Grippers Long-stroke Grippers. Stroke per finger 14 mm.. 64 mm. Sizes 22..

Level MEASUREMENT 1/2016

Transcription:

Operation Manual LB device KSV 5000 Revision 1.1

Table Of Contents 1. INTRODUCTION------------------------------------------------------------------------- 1 2. PHYSICAL DESCRIPTION----------------------------------------------------------- 2 2.1. Overview ------------------------------------------------------------------------------2 2.2. System 1 -------------------------------------------------------------------------------3 2.3. System 2 -------------------------------------------------------------------------------4 2.4. System 3 -------------------------------------------------------------------------------5 2.5. Mechanics -----------------------------------------------------------------------------6 2.6. Additional Devices ---------------------------------------------------------------- 10 2.7. Special Troughs-------------------------------------------------------------------- 11 3. PRELIMINARIES -----------------------------------------------------------------------15 3.1. Connections------------------------------------------------------------------------- 15 3.2. Cleaning ----------------------------------------------------------------------------- 16 3.3. Dipper Substrate Attachment -------------------------------------------------- 19 3.4. Monolayer Material Spreading ------------------------------------------------ 19 3.5. Experimental Setup--------------------------------------------------------------- 20 4. TECHNICAL SPECIFICATIONS ---------------------------------------------------21 4.1. Deposition System----------------------------------------------------------------- 21 4.2. Film Pressure Measuring System---------------------------------------------- 21 4.3. Film Area Control System------------------------------------------------------- 21 4.4. Trough ------------------------------------------------------------------------------- 22 4.5. General ------------------------------------------------------------------------------ 22 5. CONTACT INFORMATION----------------------------------------------------------23

1. Introduction The Langmuir and Langmuir-Blodgett (LB) devices available from KSV Instruments are efficient and effective in investigating the properties of floating monolayers, precise deposition of multilayers onto solid substrates or simply as platforms for use in observing surface chemistry effects such as the breakdown of an enzyme or the crystalline structure of a surfactant. The wide range of available systems and the many modules available make customisation a charm, and if the troughs available do not suit your needs then contact us about designing a particular trough to your specifications. There are four series of LB systems available today, the KSV 5000, 2000, Minitrough and the Minimicro in descending order of size. The differences between the systems are mechanical, the same software and interface unit operates all of the LB devices. KSV 2000 is the basic multipurpose LB device. KSV 5000 adds a sturdy frame with automatic elevator arms for attaching additional measurement devices, becoming the most versatile LB device available today. The KSV Minitrough is a more compact and economical LB device. The KSV Minimicro is a small and elegant LB device brilliant for depositions on small substrates or to minimise sample volumes. This is the device-specific manual for the KSV 5000 LB device, featuring details such as physical descriptions and cleaning instructions. Please see the LB installation manual or the LB software manual for details on installation, calibration, measurement taking or data analysis. 1

2. Physical Description 2.1. Overview The general physical layout of the KSV 5000 Langmuir-Blodgett (LB) system consists of the trough with the subphase and the surfactant, barriers which can adjust the surface pressure by reducing the area available to the surfactant, a balance with a Wilhelmy plate which monitors the surface pressure and an interface unit to control the devices and communicate with the computer running the experiment. The available systems are: System Trough Devices 1 Flat Balance 2 Flat with well 3 Compartmentalised with half-circle well Balance and dipper Balance and twoarmed dipper Available experiments Langmuir film balance Multilayer LB-films on solid substrate and Langmuir film balance Alternating LB-films, one component LB-films and Langmuir film balance 2

2.2. System 1 The appearance of the instrument depends on the system in use. System 1 is the most basic setup, consisting of a balance and a trough with a barrier drive system. 1 4 2 3 Image 2.1: KSV 5000 System 1. 1) Balance 2) Trough 3) Barriers 4) LayerBuilder System 1 is designed for the study of floating monolayers. PTFE trough with a surface area of 870 cm 2, (L580 x W150 x D9 mm 3 ) and a subphase volume of 780 ml. 3

2.3. System 2 The System 2 additions over System 1 are the dipper, a second stand for the dipper and a trough with a well. 5 6 4 1 2 3 Image 2.2: KSV 5000 System 2. 1) Balance 2) Trough 3) Barriers 4) LayerBuilder 5) Dipper 6) Well System 2 is designed for the study of floating monolayers and for the deposition of multilayer films of one substance on a solid substrate. PTFE trough with a surface area of 870 cm 2, (L580 x W150 x D9 mm 3 ), a dipping well (L37 x W117 x D90 mm 3 ) and a subphase volume of 1172 ml. 4

2.4. System 3 In addition to system 1 the trough is fitted with a bigger, half circle well and a two-armed dipper is positioned above the well. Two pairs of barriers and two balances operate different sides of the trough. 5 1 6 4 2 3 Image 2.3: KSV 5000 System 3. 1) Balances 4) LayerBuilder 2) Trough 5) Dipper 3) Barriers 6) Well System 3 is in effect two different LB troughs where the subphases are connected through the dipping well. Two barrier drives operate independent pairs of barriers so the surface pressures of both sample compartments can be adjusted separately. A two-armed dipper with one arm operating above the surface and the other operating below the surface means that only the substrate breaks the surface. A central compartment with a clean water surface gives unlimited choice in which layer is next deposited onto the substrate and from which direction (immersing into or withdrawing from the subphase). PTFE trough with a surface area of 2 x 930 cm 2, (L775 x W120 x D10 mm 3 ), a half-circle (radius 133mm) dipping well in the centre divided into 3 compartments and a subphase volume of 6000 ml. 5

2.5. Mechanics A) Computer The computer controls all functions of the instrument. Because each device includes its own microprocessor, the computer sends only high level commands to them (for instance "Move with certain speed" or "Go to certain position" and so on). All devices are listening to every message, but they become active only when they recognise their own address. The computer should be a PC running Windows 2000/XP/Vista. A 1GHz processor and 512MB RAM memory are recommended. B) Barrier Driving System The barrier position is controlled by a micro step driven stepping motor. The motor moves the barrier holder using a tooth belt. The holder itself is attached to a linear motion system, which is equipped with ball bearings. The barrier driving system is equipped with adjustable safety switches which stop the barrier immediately when the barrier holder hits the switches. The position of the barrier is measured using an optical encoder. The position is relative and so a zero point at a known position must be set manually. The control electronics are located inside of the barrier driving unit. There is a 15-pin connector on the rear of the unit. The barrier driving unit should be connected to the Interface unit through this connector. 6

Image 2.4: Block diagram of the barrier driving system C) Surface Balance The surface pressure is measured by using the Wilhelmy plate method (round rod optional). The Wilhelmy plate is a carefully sandblasted platinum plate or a clean paper plate, which is put partly under the surface of the subphase. Normal practice is to position the plate so that one third of it is under the subphase. The force acting on the plate depends on surface pressure. This force is measured using the electro-balance. The total measurement range of the balance is +/- 2000 mg. The width of the platinum Wilhelmy plate has been chosen so that 1 mn/m corresponds to 4 mg. For paper Wilhelmy plates this value is commonly about half as great as with the platinum plates. 7

Image 2.5: Block diagram of the surface balance. D) Deposition system The deposition system is DC motor controlled to ensure vibration-free operation. The speed range of the dipping arm is 0.1...85 mm/min and maximum stroke is about 145 mm. The DC motor is servo controlled to keep the speed constant. The position is determined with an optical encoder. The deposition system is equipped with safety switches which stop the motor automatically in both ends. 8

Image 2.6: Block diagram of the deposition system. E) Trough and Barriers The troughs are made of a solid piece of PTFE. The trough is glued onto an aluminium base plate, in which there are water channels for the thermostation of the subphase. The temperature is adjusted with a water bath circulator, which is connected to the trough with special connectors. These connectors include valves in both parts, so the connection can be opened without any leakage. The maximum temperature allowed is 60C! The barriers are made out of hydrophilic material, which ensures that the film does not escape the barrier. The barrier material is polyacetal (Delrin) and it should be noted that Delrin is not resistant to acids or chloroform, so do not use these as cleaning agents. Chlorinated solvents such as these can still be used as spreading solvents. The barrier is equipped with a round rod, which fits into the barrier holder. 9

2.6. Additional Devices Various additional devices can be used to improve measurement conditions, enable additional measurement methods or to provide additional data. A) Thermostation The temperature of the subphase can be monitored with a temperature probe that connects directly to the Interface Unit. A temperature probe is shown in image 5.7. The temperature sensor is covered with a thin PTFE tube and it can be attached to the trough with a special holder. The display units are in degrees Celcius, though Fahrenheit is available on request. This keeps a full record of the temperature conditions of the experiment, but to adjust that temperature a water bath is required. KSV troughs have built-in piping in the aluminum block which holds the teflon trough. Connect the water bath to the ports along one end of the trough. KSV Instruments recommends the Julabo water bath also available from KSV, though others can be used as well. Connect the water bath to an available COM port on the computer. Use the software's Manual Control Unit to adjust the temperature settings for the water bath, see section 4.5 Control Panel for details. B) ph Probe The acidity of the subphase can be monitored with a ph probe that connects directly to the LayerBuilder interface unit. C) Stirrer A magnetic stirrer consists of a rotating magnetic field source and a small magnetic bar. The rate of rotations can be adjusted with the control knob on the power supply. A magnetic stirrer is used particularly with enzyme reaction studies. D) Cabinet A cabinet that can be placed around the LB device to isolate it more from the environment. Includes appropriate holes for water bath circulator tubes and cables. 10

E) Horizontal Dipping Clamp The standard dipper clamp can be replaced by a horizontal dipping clamp for the purpose of conducting Langmuir-Schaefer experiments. The two dipping clamps can be interchanged quickly and easily, and no software upgrade is necessary. Set the zero position of the dipper as the level at which the plate is wetted first. With longer experiments the dipping depth might need to be increased over time to reach the surface due to evaporation. F) Surface Potential Probe SPOT A surface potential probe (KSV SPOT) is a tool for measuring the surface potential of an interface or a solid. KSV SPOT uses the vibrating plate capacitor method. G) Brewster Angle Microscopy BAM Brewster angle microscopy (BAM) provides visual information of two dimensional structures in monolayer films. It utilises a laser and a CCD camera to obtain images at 25 fps. H) Interfacial Shear Rheometer ISR The ISR 400 provides information on the shear stress of interfaces, particularly useful for low viscosity applications. A teflon-coated magnetic needle rests at the interface, and as a magnetic field is induced the movement of the needle is observed with a microscope. 2.7. Special Troughs The choice of trough is one of the most important ones to make when acquiring and using a LB device. Several troughs are available designed, and many more can be manufactured according to specifications. The specialized trough types available at the moment are alternate dipping, conductivity, compartmentalised, enzyme reaction, oil/water interface and surface potential troughs. Additionally several features can be incorporated into most of the troughs, these are an injection port, a quartz window and low subphase volume. A) Alternate Dipping Trough The alternate dipping trough allows precisely constructed layers of two different surfactant substances for KSV 2000 System 3. Subphase volume reduced from 6000 ml to 1500 ml. Maximum size of substrate 20mm x 45mm. 11

B) Conductivity Trough The conductivity trough is used to measure the conductivity of the monolayer. Four electrodes are placed near the center of the trough, two on each side of a central rise. The central rise divides the trough into two compartments with a narrow band of glass connecting the subphases of the different compartments. Image 2.7: The conductivity trough The exterior electrodes introduce a current and the interior electrodes measure the potential between them. The subphase volume through which the current can flow is minimized with the use of the glass band. C) Enzyme Reaction Trough The enzyme zero order breakdown reaction can be observed with this special trough. The center compartment is connected to the outer compartments with glass bridges allowing the pressures to be adjusted without disturbing the center compartment. Image 2.8: The enzyme reaction observation trough. The enzyme is placed in the centre trough and the initial surface pressure is measured. A protein is placed through an injection port and as the protein 12

interacts with the monolayer the amount of monolayer material decreases and surface pressure decreases. The barriers approach the centre to keep the surface pressure (and thus mean molecular area) constant, and the area removed by them can be used to calculate the activity of the enzyme. A stirrer might be necessary to properly observe the activity of the enzyme. D) Oil/Water Interface The oil/water interface trough places the monolayer material between two liquids as opposed to between a liquid and a gas. The denser liquid, usually water or another polar liquid, is poured into the trough up to the first notch. The monolayer is placed onto the surface of the polar liquid. The organic liquid is then poured gently on top of the water so that the holes in the barriers are distinctly above the interface but covered by the oil. When the barriers are brought closer together the monolayer material is compressed. Image 2.9: The oil/water interface trough Alternatively the monolayer material can be mixed into the oil and as the oil is poured the monolayer material forms a monolayer at the interface due to the strong polar attractions between the polar subphase and the hydrophilic end of the monolayer material. E) Surface Potential Trough The trough especially designed for surface potential measurements is long and narrow to maximise compression area. Surface potential is much more sensitive than surface pressure, and changes in surface potential occur at higher mean molecular areas than changes in surface pressure. Additionally a long trough could be useful in investigating a material with curious properties at high mean molecular areas but with a collapse point at low mean molecular areas. 13

Image 2.10: The surface potential observation trough with a quartz window. F) Additional Trough Features A quartz window can be added to most troughs to enable live observation with a microscope or similar optical instrument. An injection port can be used to add a substance into the trough after the beginning of an experiment. As it is important not to disrupt the surface the injection port leads a needle tip from the side of the trough to the subphase. It is a standard feature with enzyme reaction troughs. A low subphase volume trough reduces the area that is available for the subphase, thus reducing the amounts of materials required. G) Custom Troughs Many situations arise where a custom designed trough is necessary to accomplish a particular task. KSV Instruments supplies many custom designed troughs, be they extensions of current designs or completely new concepts. Ask us for more information about a trough that suits your needs! Image 2.11: Some past custom-made troughs. 14

3. Preliminaries 3.1. Connections Check that all devices are attached to the LayerBuilder interface unit. Check that the power cable is connected to the interface unit and that it is turned on. Turn on the computer and start the sgserver for LB systems. It can be located from the Start menu under Programs > LB. Open the Manual Control Unit by clicking on the menu Control Panel. Open the Barrier1 tab. Image 3.1: The Manual Control Unit used to control attached devices. Check that the safety limit switches are properly set by first closing and then opening the barriers. Adjust the positions of the switches if necessary. 15

3.2. Cleaning Before an experiment the trough and barriers must be cleaned. Use rubber gloves to minimise oils from the skin contaminating the apparatus. The importance of cleanliness cannot be overstated! The troughs used by KSV can be removed from the frame and carried to a sink for ease of cleaning. A recommended cleaning procedure is to first brush with a soft brush covered in ethanol or another organic solvent (mechanical and chemical cleaning) and then rinse with ion exchanged water. Take care not to scratch the surface! Images 3.2 and 3.3: Brushing the trough with ethanol (left) and rinsing the barrier with pure water. If the trough has not been used for a longer period of time then first washing with hot water and a commercial detergent might be appropriate before performing the standard cleaning described above. Replace the trough in the frame, checking that the screws fit into the available notches. 16

Image 3.4: Replacing the trough so that the screws slot into the respective notches. Clean the barriers. The appropriate cleaning method is the same as with the trough. When one barrier is cleaned place it on top of the trough before cleaning the other barrier. Take special care in choosing the appropriate handhold on the barrier so that replacing it will be simple, avoid touching the delrin itself! With an Alternate Trough Experiment the second barrier drive must be attached on the opposite side from the first barrier drive. Fit the screw along the side of the trough to the respective notch on the barrier drive, just like with the first barrier drive. Rinse the aspirator tip with ethanol and ion exchanged water. On the Manual Control Unit click Open and wait until the barriers are in the desired position, most often fully open. Press Zero. Fill the trough with the subphase. In most cases pure water will be a good choice. Pour the water gently on to the trough so that the level of the water rises distinctly (at least 3mm) above the level of the trough. Turn on the aspirator and click on Close to start bringing the barriers together. Contaminants on the surface of the water will be picked up by the barrier. Run the tip of the aspirator along both barriers several times. The purpose of this is to remove contaminants and not to suck too much of the water so run the tip over the surface rather than plunging it into the water. When the barriers are fully closed run the aspirator tip over the surface between the barriers several times to pick up remaining contaminants. A systematic pattern such as three vertical and horizontal lines (#) is recommended. Remove enough water to lower the water surface until it is level with the trough. 17

Image 3.5: Sucking contaminants from the barriers. Clean the Wilhelmy plate. A thorough cleaning entails heating the platinum Wilhelmy plate over a flame (such as a Bunsen burner). Heat the plate over a blue flame hot enough to make the plate glow red in just a few seconds, a flame that is too cool will leave residues on the plate surface. Alternatively rinsing with ethanol and ion exchanged water should be enough for most cases of frequent use. Once the experiment is finished the Wilhelmy plate should be stored in a water-soluble organic solvent such as ethanol. Attention! The platinum Wilhelmy plate is an excellent catalyst for the combustion of methanol! Beware of operating the Wilhelmy plate around methanol, especially when hot. After the plate is completely glowing remove it from the flame and hold it for several seconds to allow it to cool. Dip the plate into the water in the trough and hang it from the balance hook. Set the plate so that about two thirds of it is covered and it is perpendicular to the barriers. For information on using a paper Wilhelmy plate, please see the LB software manual. Open the barriers. In the Manual Control Unit open the Balance tab and press Zero Balance1. The reading is now zeroed with pure water, and the measured difference in surface tension will be the surface pressure. Close the barriers. Pure water does not cause the surface pressure to change, so any change in surface pressure is caused by contaminants. If the surface pressure does not exceed 0.2mN/m the water can be considered sufficiently clean. 18

3.3. Dipper Substrate Attachment If about to perform a Dipping Experiment with a hydrophilic substrate then clean the sample as appropriate. In the Manual Control Unit open the Dipper1 tab and press Up and wait until the dipper arm moves well above the surface. Attach the substrate to the dipper arm and make sure that the substrate is still above the water surface. Take care not to touch the surface! Lower the substrate slowly from the Manual Control Unit and at when the water touches the substrate press Stop and then Zero. Then lower the substrate as far as it is meant to go for the experiment. Images 3.6 and 3.7: Attaching a substrate to the dipper (left) and zeroing the position of the substrate for a dipping experiment. 3.4. Monolayer Material Spreading Fill a precision syringe with a suitable volume of a prepared solution of monolayer material and a volatile organic solvent. The organic solvent should be non-polar so that it will not mix with the water and volatile so that it will evaporate from the surface, for example chloroform or hexane. A concentration of 1 mg/ml is usually a good choice to start investigations with. Placing the solvent on the water should be done with care, gently push on the syringe to get a drop out of the needle and then touch the surface with the tip of the needle. Do not let the drop fall from the needle as some of the sample might be lost to the subphase as micelles and the surfactant does not spread evenly! Furthermore the waves caused by drops can carry monolayer materials to the edges of the trough. 19

Image 3.8: Place the sample onto the water with care. Wait 10-15min for the solvent to evaporate. Use this time to fill in the Experimental Setup for this experiment. Press on the appropriate icon to start a new experiment, this will first open the Experimental Setup. Doublecheck that the dimensions of the trough are correct as this plays an important part in the analysis of the obtained data. 3.5. Experimental Setup The Experimental Setup is used to identify the appropriate procedure and as a record for later analysis. Access this screen by beginning a new isotherm, dipping or alternative trough dipping experiment. Fill in all of the relevant sections. To add items to the drop-down menus use the Edit Database function. For more details, and for instructions on how to run the experiments themselves, please see the LB Software Manual. 20

4. Technical Specifications 4.1. Deposition System Control and operation Deposition speed Speed adjust Optional speed range Software controlled unsupervised and automatic film deposition. User defined deposition parameters. 0.1 to 85 mm/min Increment 0.1 mm 0.2 to 170 mm/min Deposition cycles 1 to unlimited consecutive depositions Dwell times Deposition arm Max. size of substrate Type of motor 0 to unlimited seconds, individual adjustment for upper and lower end point of substrate movement. 145 mm max. stroke 100 x 100 mm Servo controlled DC motor 4.2. Film Pressure Measuring System Measuring principle Dynamic range Resolution Wilhelmy plate, platinum or paper sensor element or round rod connected to micro-electronic feedback system for surface pressure control. Software controlled operation, user defined measuring parameters. A floating barrier sensor element with the same features as above also available. 0 to 250 mn/m 4 µn/m 4.3. Film Area Control System Surface area regulation Surface barrier Two symmetrically moving surface barriers. Software controlled operation and user defined process parameters. Leak proof hydrophilic (Delrin) 21

Barrier speed Barrier motor barriers. Hydrophobic (Teflon) barriers by request or optional. 0.01 mm to 400 mm/min (max. relative speed 800 mm/min) High precision micro step driven stepper motor. Accuracy Greater than 99% 4.4. Trough Dimensions System 1: surface area 870 cm 2, (L580 x W150 x D9 mm 3 ), subphase volume 780 ml. Trough material Barrier material Temperature range 4.5. General Environmental protection Computer requirements System 2: surface area 870 cm 2, (L580 x W150 x D9 mm 3 ), dipping well (L37 x W117 x D90 mm 3 ), subphase volume of 1172 ml. System 3: surface area 2 x 930 cm 2, (L775 x W120 x D10 mm 3 ), half-circle dipping well (radius 133mm, depth 125mm), subphase volume of 6000 ml. Solid PTFE (Teflon), 1.5 mm bottom thickness, mounted on a thermo regulated base plate. Hydrophilic Delrin (polyacetal). Delrin is not suitable to be washed with acids or chloroform. Teflon barriers can be provided as option. 0 to +60 C Optional vibration isolated bases and laminar flow hoods available. PC with a 1 Ghz processor and 512 MB memory running Windows TM 2000/XP/Vista. Voltage Nominal 220 V, 50/60 Hz 110 V, 50/60 Hz Operational 176-264 V, 40/440 Hz 90-132 V, 40/440 Hz 22

5. Contact Information If any problems arise please feel free to contact a local distributor or KSV Instruments directly. KSV Instruments can be contacted from this address: KSV Instruments Ltd. Höyläämötie 7 FIN-00380 Helsinki Finland Tel. +358-(0)9-5497 3300 Fax +358-(0)9-5497 3333 E-mail info@ksvltd.fi for sales and support@ksvltd.fi for service or technical questions http://www.ksvltd.com Local distributors are listed at our website, www.ksvltd.com. 23