Toxalert International, Inc.

Similar documents
O3 3E 1 F Gas Sensor Module

MODEL GT820 OXYGEN SENSOR

MODEL CALIBRATION GAS DELIVERY SYSTEM

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Model 4501-XX 2-Wire Series Toxic Gas Sensor Module

Model B Wire Wet Oxidant Gas Transmitter. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

Operation and Maintenance Manual

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

210 Series Transmitter with External Electrochemical Sensor

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

RAM 4021 Operation Manual

Operating Instructions

BI-680 Online Dissolved Oxygen Controller Instruction Manual

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

HLM-2000-TX Gas Transmitter User Manual

Installation and Operation Manual

Air Check Lite Toxic Gas Monitor

RAM Operation Manual

HLM-2000-TX Gas Transmitter User Manual

Detcon MicroSafe Model DM-534C Oxygen Deficiency Sensor (0-25% O2)

Model TX-FM/FN Sensor Head

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User Manual. µ-gard MD Electrochemical Carbon Monoxide Transmitter Serial No. MD April, 2011

GasSense NDIR User Manual

ACV-10 Automatic Control Valve

Electro-Pneumatic Converter YT-940 SERIES

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

RAM 4021-DPX Operation Manual

Columbus Instruments

Operation and Maintenance Manual

SC505 Sof Care Inflator. Operating Instructions and Service Manual

CSA Sample Draw Aspirator Adapter Operator s Manual

! Warning, refer to accompanying documents.

Best Practice Guide, Servomex 2700

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

Advanced Test Equipment Rentals ATEC (2832)

Sensoric 4-20 ma Transmitter Board Operation Manual

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

User Manual. PolyGard DT V4. Electrochemical Carbon Monoxide Transmitter Serial No. DT December, 2008

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Datasheet: K-30 ASCII Sensor

RGC-IR Remote Gas Calibrator for IR400

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

T i m i n g S y s t e m s. RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA (408)

CALIBRATION KIT. Instruction Manual

Using the UltraRAE. Firmware 2.35

Technical Data Sheet MF010-O-LC

BGA244 Binary Gas Analyzer

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Overview. Front Panel: Keypad and Display

5 Channel Calibrator

Model 106 DPI "Micro-switch" Installation and Operating Instructions

Model DPC3500 Continuous Low Range Dew Point Analyzer. Operations Manual

GAS FUEL VALVE FORM AGV5 OM 8-03

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

DPC-30 DPC-100. Reference Manual

SDX Submersible Depth Transmitter User Manual

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Operation Manual. Rev. C LPT Low Power Transmitter (Electrochemical Sensor)

with O 2 Controller Instruction Manual

Manual of SF6 Comprehensive Tester

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

In Vivo Scientific, LLC INSTRUCTION MANUAL

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

SDX Submersible Depth Transmitter User Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual

User Manual. PolyGard AT-1130 V3. Electrochemical Nitrogen Dioxide Transmitter. July, 2011

LVU2800 Series. Ultrasonic Level Transmitter

PURA. Pure Gas Dewpoint Transmitter. Users Guide

DTG - LCD Digital Temperature Gauge USER MANUAL

A4s Operation Manual

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

INSTRUCTION MANUAL Detcon Model DM-200

Simrad GD10P Extended On-site Maintenance

Transmitter CS 21 Operation Manual

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual

Media-Isolated Mag Plus Probe Low Pressure-ISO Kit

INSTRUCTION MANUAL. January 23, 2003, Revision 0

NOVALYNX CORPORATION MODEL BAROMETRIC PRESSURE SENSOR PCB MODEL BAROMETRIC PRESSURE SENSOR IN NEMA ENCLOSURE INSTRUCTION MANUAL

MODELS (2 foot Version) (4 foot Version)

A4 Operation Manual. Fig.1-1 Controller Socket Diagram

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

PERSONAL AIR BREATHING UNIT

EX-5150-MOS Sensor/Transmitter For PPM or %LEL Manual

SPECIFICATIONS APCEPH1

STERILIZABLE DISSOLVED OXYGEN SENSORS

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

Ultima. X Sampling Module, DC Aspirated Model

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions

Transmitter CS 21 Operation Manual

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

MODEL 154XST INSTALLATION AND SERVICE INSTRUCTIONS FOR MODEL 154XST STROBE LIGHT FOR USE IN HARSH ENVIRONMENTS/ HAZARDOUS LOCATIONS

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Transcription:

MODEL TOX45-XX 2-Wire Series TOXIC GAS SENSOR MODULE APPLICABILITY & EFFECTIVITY Effective for all Model TOX45-XX Modules manufactured after September 1, 2010. Instruction Manual Part Number T13022 Rev. B

PAGE LEFT BLANK

TABLE OF CONTENTS 1. PRODUCT DESCRIPTION... 3 1.1 GENERAL... 3 1.2 MOUNTING CONFIGURATION... 3 1.3 THEORY OF OPERATION... 3 1.4 MECHANICAL... 4 1.5 INTERCONNECT WIRING... 4 1.6 POWER REQUIREMENTS... 4 2. CAUTIONS & WARNINGS... 6 2.1 INTRODUCTION... 6 2.2 GAS SENSOR MODULES - GENERAL... 6 2.3 WIRING... 6 2.4 CALIBRATION FREQUENCY... 6 3. QUICK START... 7 3.1 OVERVIEW... 7 3.2 WIRING... 7 3.3 MODULE INSTALLATION... 7 3.4 TRANSMITTER INSTALLATION... 7 3.5 START-UP & OPERATION... 7 3.6 ZERO STABILIZATION... 7 4. INSTALLATION... 8 4.1 SENSOR MODULE LOCATIONS... 8 4.2 WIRING... 8 4.3 ENCLOSURE INSTALLATION... 9 4.4 TRANSMITTER AND SENSOR INSTALLATION... 10 5. OPERATION... 11 5.1 INTRODUCTION... 11 5.2 DATA ENTRY KEYPAD... 11 5.3 MAIN MENU... 12 5.4 SETUP SUB-MENU... 13 5.5 MAINTENANCE SUB-MENU... 15 6. CALIBRATION... 16 6.1 CALIBRATION FREQUENCY... 16 6.2 CALIBRATION PREPARATION... 16 6.3 CALIBRATION GAS DELIVERY METHODS... 16 6.4 SENSOR EXPOSURE TO GAS... 16 6.5 CALIBRATION SUB-MENU... 17 7. SERVICE... 18 7.1 SENSOR MODULE CONFIGURATION... 18 7.2 ENCLOSURE REPLACEMENT... 18 7.3 TRANSMITTER REPLACEMENT... 19 7.4 SENSOR REPLACEMENT... 19 7.5 INSTALLATION INSPECTION... 19 7.6 INSPECTION AND TROUBLESHOOTING GUIDE... 20 Page: 1

8. APPENDICES... 21 8.1 APPENDIX A: SPECIFICATIONS... 21 8.2 APPENDIX B: MODEL NUMBERS & PARTS LIST... 24 8.3 APPENDIX C: LIMITED WARRANTY... 25 8.4 APPENDIX D: TOX45-25 GAS SENSOR MODULE... 26 LIST OF FIGURES 1-1: MODEL TOX45-XX-AL HYDROGEN SULFIDE GAS SENSOR MODULE.. 5 1-2: MODEL TOX45-XX HYDROGEN SULFIDE GAS SENSOR MODULE ORIENTATION OPTIONS..... 5 4-1: TYPICAL MOUNTING OPTIONS.... 8 4-2: 4-20 MA CONNECTION.... 9 4-3: CONNECTOR LOCATIONS....10 5-1: FACE PLATE WITH OPERATOR KEYPAD..... 11 6-1: MODEL 5358-01 CALIBRATOR ADAPTOR.... 16 7-1: MODULE COMPONENTS.. 18 8-1: MODULE COMPONENTS... 26 8-2: SENSOR ASSEMBLY EXPLODED VIEW.... 28 8-3: MODEL TOX45-25-AL..... 29 8-4: MODEL TOX45-25-SS..... 29 LIST OF TABLES 5-1: MASTER MENU.... 12 5-2: KEY OPERATIONAL DISPLAY...12 5-3: SET-UP CONFIGURATION.....14 5-4: MAINTENANCE SUB MENU..... 15 6-1: CALIBRATION... 17 Page: 2

1. PRODUCT DESCRIPTION 1.1 GENERAL The is a 2-wire transmitter with 4-20 ma output and it offers a broad array of features including: 2-wire loop powered 180 day calibration interval Selectable ranges (if appropriate for gas type) Non-intrusive one-person calibration Integral LCD 2-line, 20-character display Stainless steel enclosure and sensor housing (Optional) 4-20 ma output Long sensor life Rotatable faceplate enables multi-directional conduit installation. Members of the TOX45-XX Two-Wire Toxic Gas family include: TOX45-03 Oxygen Gas Sensor Module TOX45-04 Carbon Monoxide Gas Sensor Module TOX45-05 Hydrogen Sulfide Gas Sensor Module TOX45-06 Chlorine Gas Sensor Module TOX45-08 Chlorine Dioxide Gas Sensor Module TOX45-10 Sulfur Dioxide Gas Sensor Module TOX45-12 Nitrogen Dioxide Gas Sensor Module TOX45-25 Ammonia Gas Sensor Module 1.2 MOUNTING CONFIGURATION Various module mounting configurations can be implemented without special fixtures. Where applicable, these options are factory configured prior to shipment. Mounting configuration can be selected by the installer or field technician and are fully described in this manual. Sensor must always be oriented downward. 1.3 THEORY OF OPERATION Electrochemical sensors are fuel cell-like devices consisting of an anode, cathode, and electrolyte. The components of the cell are selected so a subject gas, allowed to diffuse into the cell, will cause a chemical reaction and generate a current. The cells are diffusion limited so the rate the gas enters the cell is solely dependent on the gas concentration. The current generated is proportional to the rate of consumption of the subject gas in the cell. Toxalert electrochemical sensors provide improved reliability by allowing the gas to diffuse into the sensor through a capillary port, rather than diffusing through membranes. The result is an extremely stable sensor with very low temperature and pressure coefficients and the capability to monitor gas as ppm. Page: 3

1.4 MECHANICAL The sensor module is comprised of the following three primary components: 1.4.1 ENCLOSURE Explosion proof, rain-tight aluminum or stainless steel electrical housing with three ¾ FNPT conduit hubs. The enclosure cover has a viewing window. The 3 conduit hubs combined with the rotatable faceplate enable the user to mount the module in a variety of configurations. See Figure 1-2 for details. 1.4.2 TRANSMITTER ELECTRONICS Electronic Assembly consisting of one printed circuit board assembly mounted under a cover plate. Wiring connections for power and signal interface are located on the back side of the display. 1.4.3 SENSOR ASSEMBLY The sensor assembly includes a stainless steel, explosion proof, housing containing the gas sensor and a wiring harness for connection to the transmitter. The sensor assembly threads into one hub of the enclosure. The exposed end of the sensor assembly is threaded to allow connection of a rain-shield or calibration gas delivery fitting. 1.5 INTERCONNECT WIRING Not supplied with the sensor module, but necessary to the installation and operation is the two conductor wiring shielded which connects the module to its power source and controller. Before this wiring is installed it is important to read and understand the control system installation instructions to determine wiring requirements and alternatives. 1.6 POWER REQUIREMENTS The TOX45-XX module operates on DC power between 14 VDC and 30 VDC. The DC power must be supplied by the loop controller. Page: 4

Figure 1-1 Model TOX45-05-AL or SS Hydrogen Sulfide Gas Sensor Module Figure 1-2 Model TOX45-XX Hydrogen Sulfide Gas Sensor Module Orientation Options Page: 5

2. CAUTIONS & WARNINGS 2.1 INTRODUCTION Although the TOX45-XX Gas Sensor Module is designed and constructed for installation and operation in industrial applications including hostile environments, caution should be taken to insure that the installation is made in compliance with this instruction manual and that certain procedures and conditions are avoided. This chapter discusses the necessary cautions. Read the entire chapter prior to installation of the equipment. 2.2 GAS SENSOR MODULES - GENERAL Avoid installing sensor modules where they will be unnecessarily exposed to wind, dust, water (i.e. direct hose down), shock, or vibration. Observe temperature range limitations. Sensors may be adversely affected by prolonged exposure to certain materials. Loss of sensitivity, or corrosion, may be gradual if such materials are present in low concentrations. These materials include: Halides (compounds containing chlorine, fluorine, bromine, iodine), acid vapors, caustic liquids or mists. Care has been taken by the manufacturer to ship modules in protective packaging to avoid contamination prior to installation. It is recommended that the modules remain protected during installation and that the covering be removed immediately prior to system start-up. During normal use the O 2, CO, H 2 S, SO 2 and NO 2 sensors are protected from dirt and oil contamination by a sintered metal cover. If this cover becomes clogged, the response of the sensor will be reduced. Protect the sensor from contamination by careful placement, or by use of rain shields and dust shields. Sensor modules must not be painted. Paint may contain compounds which will contaminate the sensor. Paint will cause clogging of the sintered metal cover and will cause difficulties during attachment of the calibration head or other maintenance activity. It is recommended that the module be tagged "DO NOT PAINT". 2.3 WIRING The manufacturer recommends that extra caution be taken where the installation is near any sources of electromagnetic or radio frequency interference. Precautions include: Avoid running sensor module cable close to high power cables, radio transmission lines, or cables subject to pulses of high current. Avoid running cables near large electric motors or generators. Analog devices require shielded cable. In conduit installations and cable applications the shield should be connected to instrument ground. All splices must be via either a termination hardware system or soldered. Improperly spliced cable can result in corrosion, resistance changes and system errors. NOTE Installation and wiring must be in accordance with the National Electrical Code. Voltage AC conductors are not to be run in the same conduit as voltage DC conductors. 2.4 CALIBRATION FREQUENCY The TOX45-XX calibration frequency is six months (180 days). However, prudent gas detector maintenance practices normally suggest a simple recalibration following incidences of exposure to high levels of gas that would lead to alarm activation within any system utilizing the output of the gas sensor module. Page: 6

3. QUICK START 3.1 OVERVIEW The gas sensor module has been supplied factory calibrated and ready for immediate installation and operation. An installer familiar with installation and operation of gas detection products can use this section to begin immediate use of the module. 3.2 WIRING Provide two conductor twisted shielded wiring from the power supply/control device to the sensor module location. Use wire that is 18 AWG or larger. 3.3 MODULE INSTALLATION The TOX45-XX can be mounted in a variety of configurations supported by the conduit. See figure 1-2 to determine which configuration is best for your specific application. The default configuration enables the modules to be put in line with other modules with the sensor element below the transmitter. To change the configuration simply remove the transmitter and rotate to the appropriate configuration and remount the standoffs and transmitter. The module is designed to be installed on a ¾ conduit. Two important warnings: The installation must meet any hazardous environment codes for electrical equipment. The sensor module enclosure mounting must be spaced far enough from any vertical surface to allow removal and replacement of the sensor assembly which is threaded into one ¾ conduit entry. Sensor housing must be oriented vertically pointing downward. 3.4 TRANSMITTER INSTALLATION To install the transmitter printed circuit assembly into the housing, carefully turn the faceplate so that the printing is in the correct horizontal position for the mounting configuration and slide the assembly over the two stand-offs in the enclosure. Hand tighten the two captive panel thumb screws into the stand-offs. Replace the enclosure cover prior to providing power to the transmitter. If the transmitter is installed in a classified hazardous area, replace the threaded cover prior to providing power. 3.5 START-UP & OPERATION To begin operation of the sensor module activate the instrument loop with 14-30 VDC. Each time the sensor module is powered up it will perform a warm-up for approximately 1.0 minutes (60 seconds). During this time the display will read Starting. The loop output will be held at 4 ma. NOTE: If the sensor is uncalibrated, the startup display will state UNCALIB instead of STARTING NOTE: The TOX45-06, TOX45-08, and TOX45-25 require 300 seconds warm-up period After the warm-up period has expired, the display will indicate the gas concentration. Also, the instrument loop will be released to output current in the range of 4 to 20 ma. The actual current is linear with the gas concentration and depends on the selected range. For instance, when the range of 0 to 50 PPM has been selected, a current of 4 ma corresponds to 0 PPM and a current of 20 ma corresponds to 50 PPM. 3.6 ZERO STABILIZATION All electrochemical sensors require at least 30 minutes on power prior to calibration. This allows the electrode potentials to equilibrate, resulting in a stable zero signal level. Page: 7

4. INSTALLATION 4.1 SENSOR MODULE LOCATIONS The gas sensor module utilizes a diffusion type sensor which should be located close to either the expected source or destination of the gas hazard. If the gas is heavier than air, the sensor module should be installed within 24 inches of the ground or floor. If it is lighter than air, move it above 6. After optimum locations are determined based on the above recommendations, consideration should be given to placing the sensor modules in locations which are accessible for calibration service. Slight adjustments to the location of the sensor module may have little impact on effectiveness but major effect on accessibility. Modules should be placed in areas accessible for calibration. Model Gas Specific Gravity N/A Air 1.00 TOX45-04 CO 0.97 TOX45-05 H 2 S 1.19 TOX45-06 Cl 2 2.49 TOX45-08 ClO 2 3.09 TOX45-10 SO 2 2.26 TOX45-12 NO 2 1.59 TOX45-25 NH 3 0.60 Table 4-1 The cover should face out from the wall for easy access. Specific Gravities Sensors should be pointed down and the conduit should include an inverse trap to reduce moisture (condensation) from accumulating in the electronics enclosure. 4.2 WIRING Figure 4-1 Typical Mounting Options Wire should be 18 AWG minimum at a maximum distance is 5,000 feet. Install conduit as required by local code or construction specifications. See figure 4-2 for typical 2 terminal and 4 terminal connections. Page: 8

TOX45-XX Figure 4-2 4-20 ma Connection 4.3 ENCLOSURE INSTALLATION To protect the transmitter and sensor assembly they should be removed from the enclosure and preserved until final installation and wiring termination. Prior to installation and wiring: 1. Remove the transmitter from the module housing by: Unscrew the two captive panel screws on the faceplate. Lift the transmitter out of the enclosure. Unplug the sensor cable from transmitter connector J1. Remove the sensor assembly from the enclosure hub. 2. Install the module enclosure onto the end of the supply conduit and/or bolt into position as required. NOTE When enclosure earth grounding is required for the installation a grounding lug is located in the base of the enclosure. Install the earth ground under the green ground screw. Page: 9

4.4 TRANSMITTER AND SENSOR INSTALLATION When all pre-wiring is complete: 1. Install sensor assembly in the open hub on the module enclosure. The sensor assembly thread must be fully seated into the hub and tightened to maintain explosion proof assembly. 2. Connect the sensor assembly cable to transmitter connector J1. (Figure 4-3) 3. Connect the loop wires (Loop + and Loop -) to transmitter connector P1 4-20 ma + and (Figure 4-3). NOTE Conductor P1 is removable 4. Carefully return the transmitter to the enclosure installing it over the two standoffs. Finger tighten the retaining screws into the standoffs. Loop Power Sensor Connector Figure 4-3 Connector Locations Page: 10

5. OPERATION 5.1 INTRODUCTION The TOX45-XX Toxic Gas Sensor Module utilizes a visual menu system operated by means of a magnet with the cover in place. A magnet stick is supplied for this purpose. Access the menu system to calibrate the sensor module, and for maintenance and diagnostic procedures. The following adjustments can be made to the sensor module: Range adjustments (see section 5.4) 4mA and 20mA calibration adjustments (see section 5.4) Span adjustment (see section 6.3) 5.2 DATA ENTRY KEYPAD The module menu system is operated by means of directing the magnet stick toward each of four independent hall-effect magnetic switches. Each switch functions as if it is a manually activated panel key. The keys are located under the faceplate above and below the LCD and are labeled M, E, and as shown in Figure 5-1. Key : MODE Key ENTER Key : UP (+) Key : DOWN (-) Figure 5-1 Face Plate with Operator Keypad Page: 11

5.3 MAIN MENU Key Function Display Description Reference Mode Up Down Mode Down Down Down TOX45-XX CAL 180 TOX45-XX Version TOX45-XX Starting XXX.PPM CALIB: SETUP: MAINT: Exit-? Switch [M] Switch [E] Switch [ ] Previous Menu Switch [ ] Next Menu First screen at power-up, Model and days to calibration Second screen at start-up, Version (e.g. V300aA) Third screen at start-up, will remain for 60 seconds (TOX45-05) Default Display (example TOX45-05) Mode Function: Calibration (example TOX45-05) Table 6-1 Mode Function: Setup (example TOX45-05) Table 5-3 Mode Function: Maintenance (example TOX45-05) Table 5-4 Exit Mode (example TOX45-05) Table 5-1 Master Menu Table 5-2 defines the key operational displays on the operator interface. Display - - STARTING SENSOR FAIL NO SENSOR C.## PPM F.## PPM XXX DONE Table 5-2 Key Operational Display Description Warm-up at start-up (example TOX45-05) Sensor failure Sensor not found or plugged in Calibration Mode Failure detected, sensor is attempting recovery Finished with step Page: 12

5.4 SETUP SUB-MENU The sensor module set-points menu is used to initially set-up the range and 4-20 ma calibration. Range: Once the Set-up menu is selected, press key [ ] twice and press [E] to activate the Range Set-up menu. Use the [ ] or [ ] keys to select Range menu and press [E]. When Range is selected menu provides any choices of ranges available for the gas type selected. Use the [ ] or [ ] keys to select the desired range. If the User range is selected, use the [ ] or [ ] keys to adjust the high end of the range desired. See specifications below for optional ranges. Factory Range Set-up Model Range Other Available TOX45-03 5-25% 0-25% TOX45-04 0-500 PPM Variable TOX45-05 0-100 PPM Variable TOX45-06 0-10 PPM None TOX45-08 0-3 PPM None TOX45-10 0-100 PPM Variable TOX45-12 0-20 PPM None TOX45-25 0-100 PPM Variable 4-20mA: Once the Set-up menu is selected, press key [ ] three times and press [E] to activate the 4-20mA Set-up menu. Use the [ ] or [ ] keys to select Calib menu and press [E]. The Calib section of the menu allows the user to calibrate the 4 ma and 20 ma outputs. To calibrate the 4 ma and 20 ma outputs it is necessary to have an amp meter connected to the TOX45-XX and upon selecting the 4 ma output calibration then the [ ] or [ ] keys can be used to adjust the 4 ma reading on the amp meter until it reads 4 ma. Similar steps can then be performed for the 20 ma output. Page: 13

Key Function Display Description Reference XX PPM Default Display (example TOX45-05) Mode CALIB: Down SETUP: Gas Range Adjustment Example Range S.P. Function - Range Adjust Down Down Down *100 PPM 10 PPM USER 100 PPM 4-20mA Calib Out: 4mA 4mA ACK Select [E] to select or or to select another and press [E] Press [E] if selecting 0-10 PPM range Press [E] if selecting User adjustable range Select [E] to select or or to select another and press [E] S.P Function 4-20 ma Calibrator Use or keys to select 4 ma or 20mA Selects 4 ma 4-20 ma Adjustment Example S.P. Function - Calib Adjust Momentary Acknowledge of new Set Point Table 5-3 Set-Up Configuration Page: 14

5.5 MAINTENANCE SUB-MENU The maintenance menu enables the operator to view sensor and software versions. Toxalert Intl technical support has access to other values as needed. The maintenance menu operation is described in Table 5-4. Key Function Display Description Reference Mode XXX PPM Default Display(example TOX45-05) CALIB Go to previous maintenance menu item Down SETUP: Down MAINT: VX.XX Version number of firmware : CCC001 Configuration Version Table 5-4 Maintenance Sub Menu Page: 15

6. CALIBRATION 6.1 CALIBRATION FREQUENCY The sensor module must be calibrated every 180 days at a minimum. Periodic functional tests are advisable for critical applications and hostile environments. The sensor module microprocessor software includes high-level self checking algorithms which provide continuous sensor diagnostic and self adjustment. 6.2 CALIBRATION PREPARATION Calibration of the sensor is accomplished by simple menu based steps and application of span gas. NOTE If an error is made during any stage of the calibration process, hold the magnet stick at the [M] for 10 seconds. A scrolling display will indicate calibration aborted and the sensor module will exit the calibration activity and return to normal operating mode. The calibration procedure can then be restarted. Calibration must be performed only when the area is known to be clear of the gas of interest. When in doubt, use a portable instrument to confirm that there is no background. 6.3 CALIBRATION GAS DELIVERY METHODS Calibration gas can be delivered to the sensors via the Model 5358-01: Calibration Adapter (Figure 6-1) - used with portable calibrators. CALIBRATION PROCEDURE The Calibration Menu is described on Table 6-1: The procedure requires that the menu keys be activated using the magnet stick. Each key press steps through the process of setting the zero value for clean air and then setting the span value. At each of these steps, apply calibration gas of the value corresponding to the setting accepted on the sensor module display. 6.4 SENSOR EXPOSURE TO GAS Calibration gas must be delivered to the sensor using the flow rate and duration listed in below: Model Gas Flow Period TOX45-03 Oxygen 300 cc of Zero Air or exposure to Ambient Air (3 minutes) TOX45-04 Carbon Monoxide 300 cc/min Until Stable (minimum 3 minutes) TOX45-05 Hydrogen Sulfide 300 cc/min Until Stable (minimum 3 minutes) TOX45-06 Chlorine 300 cc/min Until Stable (minimum 3 minutes) TOX45-08* Chlorine Dioxide 300 cc/min Until Stable (minimum 3 minutes) TOX45-10 Sulfur Dioxide 300 cc/min Until Stable (minimum 3 minutes) TOX45-12 Nitrogen Dioxide 300 cc/min Until Stable (minimum 3 minutes) TOX45-25 Ammonia See Appendix D * Note: Use 10PPM Cl 2 to calibrate Sample Gas In Figure 6-1 Model 5358-01 Calibration Adapter Page: 16

6.5 CALIBRATION SUB-MENU Key * TOX45-03 Function Display Description Reference Mode Down Down Mode XXXPPM CALIB:-- Cal Zero ACK 10PPM-SP CXXX PPM CXXX PPM CXXXPPM CAL-OK WAIT-300 10PPM-SP 25PPM-SP Adj-SPAN 25PPM-SP C 25PPM C 25PPM CAL-FAIL WAIT-300 (Any) XXXPPM Default Display Mode Function - Calibrate Banner: Apply zero gas, enter <E> when done Operation: Confirm area clear of gas, or apply zero air to sensor. Zero gas setting acknowledged Banner: Select span, enter <E> when done* Banner: Apply span gas, then enter <E> to calibrate gas sensor Operation: Apply calbration gas. Operation: As gas is applied the reading will increase - wait 3 minutes Calibration Passed - now remove gas Operation: Five minute time out before sensor is returned to service. Sub-Routines Banner: Select span, enter <E> when done Operation: Change Span Gas Value to 25PPM Operation: User adjustable value Operation: Ready for user adjustment Banner: Apply 25PPM gas, then enter <E> to calibrate gas sensor Operation: Apply calbration gas. Operation: No calibration gas applied, or sensor did not respond correctly. Operation: Five minute time out before sensor is returned to service. Operation: Hold magnet over Mode Switch for ten senconds to abort calibration Banner: Calibration Aborted Default Display Table 6-1 Calibration Oxygen Calibration: If area is known to be safe with adequate air flow, you may use the surrounding air as the calibration gas. If Oxygen levels are unknown, it is recommended to use a Zero Air gas standard. When calibrating an Oxygen module, there is no zero gas process in the calibration procedure. Only span adjustments can be made. Normal ambient Air = 20.9% Oxygen. Begin calibration procedure using the magnet on the Mode key until CALIB is displayed. to begin procedure. Select cal gas value (20.9), to confirm. Apply span gas for 3 minutes or use surrounding air if acceptable. Select when done, and procedure is complete. Sub A Sub B Sub A Sub B Sub C NOTE The Analog output values to be transmitted during the calibration procedure are selected in the maintenance configuration. Normal 20.9% value provides 17.3 ma. Page: 17

7. SERVICE 7.1 SENSOR MODULE CONFIGURATION The gas sensor module is comprised of the following sub-assemblies (Figure 7-1): TOX45-XX Gas Sensor Module SPL27085 Aluminum Enclosure (SPL 27077 for Stainless Steel Enclosure) TOX27071 Transmitter Assembly XXXXXXX Sensor Assembly (see Appendix B) XXXXXXX Sensor (see Appendix B) There are no field serviceable components below the sub assembly level. NOTE Area must be determined to be non-hazardous before opening enclosure. 7.2 ENCLOSURE REPLACEMENT The enclosure should be replaced if the cover threads or conduit threads have been damaged, or if the enclosure has been damaged sufficiently that it no longer meets the required NEMA classification. To replace the enclosure follow the transmitter and sensor assembly removal instructions, remove the damaged enclosure from its conduit, install a new enclosure and continue the transmitter and sensor assembly replacement instructions. Figure 7-1 Module Components Page: 18

7.3 TRANSMITTER REPLACEMENT The transmitter assembly should be replaced when it is determined that it is unreliable, noisy or cannot be adjusted for calibration. This may occur due to age, corrosion or failed components. To replace the transmitter assembly: a. Remove the cover of the main enclosure. b. Unscrew the two thumb screws in the top of the cover plate, lift the assembly and rotate 90 o to relieve the wiring service loop. c. Remove the two wires from P1 terminals (see figure 4-3). d. Unplug the sensor connector from the transmitter. e. Reverse the preceding steps to install the new transmitter. f. Restore power and allow a minimum of 30 minutes for stabilization before re-calibration. 7.4 SENSOR REPLACEMENT NOTE: Refer to Appendix D for information on sensor electrolyte replacement for TOX45-25. The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of the housing. The gas sensor needs replacement when: It is no longer possible to obtain correct Zero and Span values. The LOW SENS message appears after calibration. The sensor output signal is noisy, causing erroneous gas level readings. To replace the sensor: a. Confirm that system power has been removed. b. Remove the gas sensor module enclosure cover. c. Unscrew the two thumb screws in the top of the faceplate, lift the transmitter assembly and rotate 90 o to relieve the wiring service loop. d. Unplug the sensor connector from the transmitter. e. Unscrew the old sensor assembly from the enclosure conduit hub. Remove the sensor assembly with its harness. f. Unscrew sensor housing cover from the sensor. g. Carefully pull the old sensor straight up from the socket. h. Press the new sensor into the socket. i. Reverse the preceding steps to install the sensor assembly. j. Allow the new sensor to stabilize for a minimum of 30 minutes and then calibrate using the procedure in Section 6. 7.5 INSTALLATION INSPECTION Prior to system start-up or trouble shooting, the entire system should be visually inspected. The following are guidelines for that inspection: 7.5.1 CONTROLLER INSTALLATION Controller installed in conformance to manufacturer s instruction manual recommendations. 7.5.2 CABLING INSTALLATION All splices are soldered or via terminal block. Cabling is away from sources of electrical noise or RFI where possible. 7.5.3 SENSOR MODULE INSTALLATION Module installation in conformance with this manual. Modules accessible for calibration. Wiring terminations clean and correct. Page: 19

7.5.4 MOISTURE TRAPS AND RAINSHIELDS Conduit seals and drains installed to avoid moisture build up in electronics enclosure. Water accumulation in sensor module enclosures is a major cause of damage and system failures - take precautions to seal electrical conduits and provide moisture traps and drains to avoid water damage. Rain-shields installed where applicable. 7.5.5 STANDARD VOLTAGES Loop voltage to be applied to the sensor module must be between 14 VDC and 30 VDC. 7.6 INSPECTION AND TROUBLESHOOTING GUIDE The inspection and troubleshooting guide can be used to step through the system start-up and to determine the corrective action if a fault occurs. 7.6.1 IF MODULE DOES NOT RESPOND TO GAS 1. Repeat calibration procedure. 2. Remove the gas and wait for the timer to completely count down. 3. Apply calibration gas and verify that the sensor sees calibration gas after calibration. 4. If the sensor still does not see gas, power cycle the unit and repeat calibration. 7.6.2 IF THE MODULE DISPLAYS STARTING FOR MORE THAN 1 HOUR 1. Make sure the sensor is place in an ambient room temperature environment. 2. Power cycle the sensor. 3. Ensure that the sensor is not exposed to the gas of interest during warm-up. 7.6.3 IF THE MODULE DOES NOT DISPLAY THE CORRECT PPM 1. Power cycle the unit. 2. Recalibrate the sensor. 7.6.4 IF THE DISPLAY SHOWS F OR NO SENSOR 1. Power down the unit. 2. Open the enclosure and unplug the sensor assembly from the transmitter board. 3. Plug the sensor back into the transmitter board carefully and ensure a secure fit. 4. Check all wire connections. 5. Power up the unit. 6. If the system still shows F, apply magnet to M key, then E key. The module will start scrolling an error message. Please write that message down and contact Toxalert Intl for assistance. 7.6.5 IF THE DISPLAY SHOWS C 1. Complete calibration and exit to operating mode. Page: 20

8. APPENDICES 8.1 APPENDIX A: SPECIFICATIONS Sensor: Type: Electrochemical (1) Optional ranges available (2) Response time to 90% full signal value for applied concentration (3) Sensor life typical for use at standard temperature and pressure with occasional exposure to gas of interest (4) Sensor electrolyte needs to be recharged every six months Output: Max Range (1) Zero Drift Repeatability Linearity Resolution Response Time (2) Sensor Life (3) TOX45-03 O 2 0-25% Vol. +/- 0.5 Vol. +/- 0.1 % +/- 0.1 % 0.2% Vol. <10 seconds 20 months TOX45-04 CO 0-1200 PPM +/- 1 PPM +/- 1 PPM +/- 1 PPM 0.5 PPM <25 seconds 2 years TOX45-05 H 2S 0-100 PPM +/- 0.5 PPM +/- 1 PPM +/- 1 PPM 0.1 PPM <30 seconds 2 years TOX45-06 Cl 2 0-10 PPM +/- 0.5 PPM +/- 0.5 PPM +/- 0.5 PPM 0.1 PPM <60 seconds 2 years TOX45-08 ClO 2 0-3 PPM +/- 0.5 PPM +/- 0.5 PPM +/- 0.5 PPM 0.1 PPM <60 seconds 2 years TOX45-10 SO 2 0-100 PPM +/- 0.5 PPM +/- 1 PPM +/- 1 PPM 0.5 PPM <20 seconds 2 years TOX45-12 NO 2 0-20 PPM +/- 0.5 PPM +/- 0.5 PPM +/- 0.5 PPM 0.2 PPM <35 seconds 2 years TOX45-25 NH 3 0-100 PPM - - - +/- 1 PPM <30 seconds (4) Electrical Data: Display: Signal Output: Loop Resistance Power consumption: Connection type: Input voltage: RFI/EMI Protection: LCD, 2-line, 16-characters 2-wire sensor powered analog 4-20 ma (Trouble: 3.6 ma) 800 ohm 0.2 watts 2 wire loop-powered 24 VDC nominal: 14-30VDC EN50081-2, EN50082-2 Page: 21

Operating Range: Ambient Temp ( o F) Ambient Temp ( o C) Relative Humidity TOX45-03 O 2-4 to 122 o F -20 to 50 o C 15 90% TOX45-04 CO -4 to 122 o F -20 to 50 o C 15 90% TOX45-05 H 2S -4 to 122 o F -20 to 50 o C 15 90% TOX45-06 Cl 2-4 to 122 o F -20 to 50 o C 15 90% TOX45-08 ClO 2-4 to 122 o F -20 to 50 o C 15 90% TOX45-10 SO 2-4 to 122 o F -20 to 50 o C 15 90% TOX45-12 NO 2-4 to 122 o F -20 to 50 o C 15 90% TOX45-25 NH 3-14 to 113 o F -10 to 45 o C 20 90% Extended Range Sensor and 4-20 ma will continue to operate to -40 o F (-40 o C) but the Operation: LCD display will decay over time if exposed to temperatures below -4 o (-20 o C) for significant time (not applicable to TOX45-25) Storage Temperature: -4 o to 122 o F(-20 o to 50 o C) Enclosure: Dimensions: 7.0 x 3.6 x 4.0 inches (H x W x D) (17.8 x 9.1 x 11.2 cm) Weight: 2.2 lb. (1.0Kg) Al 4.6 lb. (2.1 Kg) SS Material of Construction: 316 Stainless Steel, Aluminum Housing: NEMA 4X, NEMA 7 Warranty: Limited Warranty: 1 year Specifications subject to change without notice Page: 22

Cross Sensitivity Data: TOX45-04 Carbon Monoxide Gas Conc. Response H 2S 15 PPM ~38 PPM SO 2 5 PPM ~3 PPM NO 35 PPM ~10 PPM NO 2 5 PPM ~-3 PPM H 2 100 PPM <60 PPM C 2H 4 100 PPM <100 PPM HCN 10 PPM ~5 PPM HCl 5 PPM 0 PPM TOX45-05 Hydrogen Sulfide Gas Conc. Response CO 400 PPM <.1 PPM SO 2 20 PPM <10 PPM NO 50 PPM <3 PPM NO 2 10 PPM <-30 PPM Cl 2 10 PPM <-25 PPM H 2 400 PPM 1 PPM C 2H 4 400 PPM <0.1 PPM NH 3 20 PPM <0.1 PPM TOX45-06 Chlorine Gas Conc. Response CO 300 PPM 0 PPM H 2S 15 PPM 0 to -3.38 PPM SO 2 5 PPM -0.05 PPM NO 35 PPM 0 PPM NO 2 5 PPM ~5 PPM H 2 100 PPM 0 PPM C 2H 4 100 PPM 0 PPM HCN 10 PPM 0 PPM HCl 5 PPM 0 PPM TOX45-10 Sulfur Dioxide Gas Conc. Response H 2S 15 PPM 0 PPM NO 35 PPM -7 to 0 PPM NO 2 5 PPM -5 PPM Cl 2 5 PPM -1.5 to 0 PPM NO 2 5 PPM ~5 PPM H 2 100 PPM 0 PPM HCN 10 PPM <5 PPM HCl 5 PPM 0 PPM C 2H 4 100 PPM 0 PPM TOX45-12 Nitrogen Dioxide Gas Conc. Response CO 300 PPM 0 PPM H 2S 15 PPM -1.5 to 0 PPM SO 2 5 PPM -0.05 to 0 PPM NO 35 PPM 0 PPM Cl 2 1 PPM ~1 PPM H 2 100 PPM 0 PPM HCN 10 PPM 0 PPM HCl 5 PPM 0 PPM C 2H 4 100 PPM 0 PPM TOX45-25 Ammonia Gas Conc. Response CO 1000 PPM 0 PPM SO 2 15 PPM >40 PPM NO 100 PPM 0 PPM NO 2 60 PPM 0 PPM H 2 4% 0 PPM CH 4 4% 0 PPM HCN 10 PPM 75 PPM H 2S 3 PPM 75 PPM TOX45-08 Chlorine Dioxide Gas Conc. Response Cl 2 3.1 PPM 10 PPM Page: 23

8.2 APPENDIX B: MODEL NUMBERS & PARTS LIST Sensor Module TOX45-03 Sensor Module 2-wire, O 2 TOX45-04 Sensor Module 2-wire, CO TOX45-05 Sensor Module 2-wire, TOX45-06 Sensor Module 2-wire, CI 2 TOX45-08 Sensor Module 2-wire, ClO 2 TOX45-10 Sensor Module 2-wire, SO 2 TOX45-12 Sensor Module 2-wire, NO 2 TOX45-25 Sensor Module 2-wire, NH 3 Options 5311-00 Rainshield 5311-02 Rainshield, with calibration port Calibration Items 5358-01 Calibration Adapter - Direct, Standard 5358-51 Calibration/Configuration Magnetic Tool, TOX45-XX TOX27057 Spare Parts TOX21834 TOX21832 TOX21830 TOX21833 TOX21844 TOX21845 TOX21840 TOX27071 TOX27082 TOX27083 TOX27074 TOX27084 TOX27087 TOX27088 SPM33044 SPM33030 SPM33041 SPM33050 SPM33052 SPM33054 SPL27085 SPL27077 Kit, Sensor Recharge, for TOX45-25 Sensor Assy, Aluminum, TOX45-03 Sensor Assy, Aluminum, TOX45-04 Sensor Assy, Aluminum, TOX45-05 Sensor Assy, Aluminum, TOX45-06/TOX45-08 Sensor Assy, Aluminum, TOX45-10 Sensor Assy, Aluminum, TOX45-12 Sensor Assy, Aluminum, TOX45-25 Transmitter for TOX45-XX Sensor Assy, 316SS, TOX45-03 Sensor Assy, 316SS, TOX45-04 Sensor Assy, 316SS, TOX45-05 Sensor Assy, 316SS, TOX45-06/TOX45-08 Sensor Assy, 316SS, TOX45-10 Sensor Assy, 316SS, TOX45-12 Sensor for TOX45-03 Sensor for TOX45-04 Sensor for TOX45-05 Sensor for TOX45-06 & TOX45-08 Sensor for TOX45-10 Sensor for TOX45-12 Module Enclosure AL Module Enclosure SS Page: 24

8.3 APPENDIX C: LIMITED WARRANTY TOXALERT INTERNATIONAL INC warrants its products to be free from defects in workmanship or material under normal use and service for one year after date of shipment. TII will repair or replace without charge any equipment found to be defective during the warranty period. Final determination of the nature and responsibility for defective or damaged equipment will be made by TII personnel. All warranties hereunder are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without TII approval or which have been subjected to accident, improper maintenance, installation or application, or on which original identification marks have been removed or altered. This Limited Warranty also will not apply to interconnecting cables or wires, consumables (i.e. calibration gases, batteries, sensors), nor to any damage resulting from battery leakage. In all cases TII s responsibility and liability under this warranty shall be limited to the cost of the equipment. The purchaser must obtain shipping instructions for the prepaid return of any item under this warranty provision and compliance with such instruction shall be a condition of this warranty. Except for the express warranty stated above, TII disclaims all warranties with regard to the products sold hereunder including all implied warranties of merchantability and fitness and the express warranties stated herein are in lieu of all obligations or liabilities on the part of TII for damages including, but not limited to, consequential damages arising out of/or in connection with the use or performance of the product. Page: 25

8.4 APPENDIX D: TOX45-25 GAS SENSOR MODULE 1. Calibration of TOX45-25 2. Equipment Required The following tools and equipment will be required for calibration: - Calibration Kit (TOXNH 3 CALKIT) - Calibration Adapter (Model 5358-01). Figure 8-1 Module Components Page: 26

Electrolyte/Membrane Replacement (See Fig. 8-2) The electrolyte needs to be replaced at least every six months, or if there is evidence of low sensor output. The membrane and O-Ring should be replaced at the same time that the electrolyte is replaced. 1. Equipment Required - Electrolyte Recharge Kit (TOX27057) consisting of the electrolyte, package of 5 membranes, one (1) 0-Ring and alcohol wipes to clean the electrode. - Tweezers to aid in removing and replacing the membrane. 2. Preparation a. Transfer the necessary electrolyte into a plastic wash bottle. b. Confirm that system power has been removed. c. Remove the transmitter electronics board from the main housing and unplug the sensor harness from the transmitter electronics. d. Unscrew the sensor assembly from the bottom of the enclosure. e. Remove the assembly cover from the sensor assembly. CAUTION: Only remove the sensor cover from the assembly, DO NOT attempt to unscrew the sensor from the mounting assembly or you could damage the wiring. 3. Electrolyte Replacement (See Fig. 8-1) a. Invert the sensor assembly so that the membrane is upward. b. Insert the sensor assembly in a clamp, if available, and unscrew the box nut counterclockwise, and take out the hold-down ring, membrane, O-Ring, and washer. Then pour out the old electrolyte into a beaker for disposal. CAUTION: Avoid contact of the electrolyte with skin, eyes and clothing. In case of contact wash immediately - rubber gloves and safety glasses are recommended. c. Clean the electrode with the alcohol wipes included in the electrolyte replacement kit. All oxidation residue must be removed before re-assembly. When dirt is found inside the sensor assembly, wash the dirt off with electrolyte. d. Hold the sensor assembly with the sensing electrode up. Close the pressure compensation port with a tissue. Pour approximately 10cc of fresh electrolyte from the wash bottle into the sensor, and rinse the inside of the sensor. Empty the electrolyte into the beaker. e. Place the old washer back over the sensor and place the O-ring on top of the washer. Then transfer one drop of electrolyte from the plastic wash bottle to the exposed electrode surface. f. Place a new membrane into the hold-down ring and carefully place the plastic hold-down ring on top of the sensor body with the ring ears in the slots of the sensor body. Then fasten the sensor cover over the assembly. CAUTION: Never touch the membrane with your hands. Use rubber gloves or tweezers. g. Invert the sensor so that the membrane is down. If this is the first time that electrolyte has been replaced, there should be a round Mili Seal covering the pressure compensation screw. This Mili-Seal prevented loss of electrolyte during shipment. Remove and discard the round Mili Seal and remove the screw to expose the electrolyte supply port. h. Transfer electrolyte from the plastic wash bottle into the electrolyte supply port (see Fig. 8-1) until the level is at the maximum level line on the side of the sensor body. i. Replace the pressure compensation screw into the electrolyte supply port. Page: 27

j. Replace the sensor housing cover. k. Install the sensor assembly into the enclosure hub and tighten firmly. l. Reconnect the sensor harness to the transmitter, carefully install the transmitter into the housing and restore system power. m. Recalibrate the transmitter following the instructions in Section 6. Figure 8-2 Sensor Assembly Exploded View Page: 28

Toxalert International, Inc. Figure 8-3 Model TOX45-25-AL Figure 8-4 Model TOX45-25-SS Page: 29

INDEX 4-20MA... 13 AMMONIA... 26 CALIBRATION... 6, 16 DATA ENTRY KEYPAD... 11 ENCLOSURE... 4, 9, 18 INSTALLATION... 7, 19 MAINTENANCE... 15 MOUNTING... 3 POWER... 4 RANGE...13 SENSOR...4, 6, 8, 16, 18, 19 SET-UP...13 SPECIFICATIONS...21 THEORY OF OPERATION...3 TRANSMITTER...4, 7, 10, 19 TROUBLESHOOTING...20 WIRING...6, 7 Page: 30