CoriolisMaster mass flowmeter Diagnostics, verification and proof test

Similar documents
The benefits of the extended diagnostics feature. Compact, well-proven, and flexible

Diagnostics in the instrumentation More than just equipment monitoring

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

PositionMaster EDP300 Extended Diagnostics. Compact, well-proven, and flexible

ABB MEASUREMENT & ANALYTICS DATA SHEET. WaterMaster FEW530 Electromagnetic water meter

Measurement & Analytics Wet gas monitoring

9408 Series Dissolved Oxygen Systems. Dissolved oxygen systems designed for reliable operation with minimal operating costs

Service Calibration are you doing it properly?

Model NDA Differential pressure transmitter. Deltapi N Series A complete range of pneumatic pressure transmitters

ABB MEASUREMENT & ANALYTICS DATA SHEET series Dissolved oxygen systems

ABB MEASUREMENT & ANALYTICS DATA SHEET. AP100 ph/redox (ORP) cartridge sensors with water-wash option

T EK-COR 1100A. Coriolis Mass Flowmeter. FLOW. Technology Solutions

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

Automatic Isokinetic Sampler. ST5 Evo

2600T Series Pressure Transmitters Calibration Frequency Determination

Special Documentation Proline Promass 80, 83

ABB measurement & analytics data sheet. JAA, JAF and JAG Pneumatic indicating controllers

C1960. Multi-recipe profile recorder/controller. Measurement made easy

Supplementary Operator s Manual 42/24-14 EN Rev. 3

2600T Series Pressure Transmitters Calibration Frequency Determination. Pressure Measurement Engineered solutions for all applications

PHIL JANOSI MSC. International Product Manager, Coriolis Flowmeter KROHNE Ltd, UK.

Drilling Efficiency Utilizing Coriolis Flow Technology

S265 Remote seals with capillary tube. 2600T Series Pressure Transmitters. Engineered solutions for all applications

Ultima. X Series Gas Monitor

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid

deltaflowc deltaflowc Venturi or Probe

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

ABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas

Navigator 600 Silica analyzers

Pro-V Multivariable Flowmeter Model M22 In-line Vortex

Endura AZ40 Sensor assembly High temperature filter assembly kit

Pressure measurement Applying remote seal, isolation diaphragms between process fluids and transmitters

Automatic Isokinetic Sampler. ST5 Evo

FPD175 Compact orifice carrier assembly

Service ISO17025 explained

Endura AZ40 Sensor assembly Smart electronics PCB

Safety Manual. Process pressure transmitter IPT-1* 4 20 ma/hart. Process pressure transmitter IPT-1*

For the highest operational safety around the clock

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

S261 Remote seals with capillary tube

Pro-V Multivariable Flowmeter Model M22 In-line Vortex

ABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners

Vortex Meters for Liquids, Gas, and Steam

Safety Manual VEGAVIB series 60

Truck-mounted Mass Flow Metering for LPG Delivery

A review of best practices for Selection, Installation, Operation and Maintenance of Gas meters for Flare Applications used for Managing facility

Coriolis Mass Flow Meter

Diagnostics for Liquid Meters Do they only tell us what we already know? Terry Cousins - CEESI

InstrumentationTools.com

NEW VERSAFLOW CORIOLIS

Ultrasonic Flowmeter USZ 08

Safety Manual VEGAVIB series 60

Autocalibration Systems for In Situ Oxygen Analyzers

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Endura AZ40 Transmitter Main board spares kit

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

TEK-THERMAL 1700B. Thermal Mass Flowmeter. FLOW. Technology Solutions

TransPort PT878GC Panametrics Portable Gas Ultrasonic Flowmeter. GE Sensing. Applications. Features

Series 3730 and Series 3731 EXPERTplus Valve Diagnostics with Partial Stroke Test (PST)

EasyLine Continuous Gas Analyzers. So smart, they re simple

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Courses of Instruction: Controlling and Monitoring of Pipelines

THERMAL MASS FLOW MEASUREMENT FOR GASES

Proline Promass 200 Two-wire Coriolis measurement for the chemical industry

SIL Safety Manual. ULTRAMAT 6 Gas Analyzer for the Determination of IR-Absorbing Gases. Supplement to instruction manual ULTRAMAT 6 and OXYMAT 6

Safety manual for Fisher GX Control Valve and Actuator

TF- & DM-Series. Turbineflowmeter

New power in production logging

GAS FILLING MACHINES PROCESS MONITORING PRESSURE BALANCE

628 Differential Pressure Decay Leak Tester

Pro-V Multivariable Flowmeter

Vehicle- or rack-mounted liquefied gas meters, pump supplied

This manual provides necessary requirements for meeting the IEC or IEC functional safety standards.

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Flow and Energy Measurement F-5400 THERMAL MASS FLOW METER

Mass Flow Controller (MFC) for Gases

Vortex Flow Meter Wafer or Flange Connection. - Steam - Liquid - Gas

Optical Dissolved Oxygen Meter

Vehicle-mounted meters, pump supplied

DeZURIK. KGC Cast Knife Gate Valve. Safety Manual

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

T EK-COR 1100A Coriolis Mass Flowmeter

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

Reliable, low-maintenance pressure measurement with SITRANS P

Bubble Tube Installations

Differential Pressure Transmiter

TransPort PT878GC Panametrics Ultrasonic Portable Gas Flowmeter. GE Infrastructure Sensing. Features. Applications

SPE The paper gives a brief description and the experience gained with WRIPS applied to water injection wells. The main

FLOWSIC100 Flare Challenges in Ultrasonic Flare meter field installations. History of Evolution

RESILIENT SEATED BUTTERFLY VALVES FUNCTIONAL SAFETY MANUAL

CONTROL VALVE TESTING

C-Flow Coriolis Mass Flow Meters

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

Level, Pressure, and Density Instrumentation for the Petrochemical Industry

Oil and Gas Custody Transfer

ONH836 Series: Oxygen/Nitrogen/Hydrogen by Inert Gas Fusion

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

UNDERSTANDING, ACCELERATED

Transcription:

ABB MEASUREMENT & ANALYTICS WHITE PAPER CoriolisMaster mass flowmeter Diagnostics, verification and proof test Automatic in situ verification of accuracy for CoriolisMaster mass flowmeters. By Frank Frenzel and Reinhard Poft Measurement made easy Using ABB's VeriMass technology for selfcalibration of CoriolisMaster mass flowmeter Introduction Users often question whether their measuring devices are working properly and within specified measurement accuracy. The new CoriolisMaster flowmeter models by ABB include technology for integrated online diagnosis and accuracy verification. This technology constantly monitors the oscillating meter tube within the flowmeter for erosion and coating deposits. Self-monitoring technology avoids costly downtime needed for proof testing and provides other significant cost savings. It permits compliance with increasingly strict quality standards and legal requirements for verification of flowmeter accuracy.

2 CORIOLISMASTER MASS FLOWMETER DIAGNOSTICS, VERIFICATION AND PROOF TEST WP/CORIOLIS/VERIMASS/101-EN REV. B

CORIOLISMASTER MASS FLOWMETER DIAGNOSTICS, VERIFICATION AND PROOF TEST WP/CORIOLIS/VERIMASS/101-EN REV. B 3 CoriolisMaster mass flowmeter Diagnostics, verification and proof test 01 Oscillating meter tubes inside a Coriolis flowmeter Coriolis and VeriMass technology Aside from measuring highly accurate mass fluid flowrates, the Coriolis technique offers numerous other advantages. The Coriolis metering principle is independent of the fluid's density, temperature and conductivity, making it very flexible to use. It's also independent of the flow velocity profiles. So, it does not require upstream and downstream straight runs of piping. As a bonus, the flowmeter provides a measurement of fluid density within the tubes. It also includes a temperature sensor to compensate for dimensional and elasticity changes of the tubes with fluid temperature. Lastly, these flowmeters can measure nearly zero flow, where other measurement methods don't work or result in significant measurement errors. Each measuring device leaves the factory with extremely well-balanced meter tubes and minimized driver current consumption. Any mechanical change of the meter tubes caused by erosion or deposit formation affects the balancing of the meter tubes. This leads to higher driver currents to maintain the same level of oscillation. This fundamental principle, called VeriMass technology, can be used to detect tube changes and can serve as in-situ verification of accuracy. Imbalances of the meter tubes most frequently cause measured value errors. The detection of these factors constitutes a major step in Coriolis mass flow measurement technology. Self-monitoring capabilities greatly augment the above capabilities. The special metering tubes within the CoriolisMaster flowmeter are mechanically balanced. Balancing the meter tubes minimizes the energy for their oscillation. The amount of energy required is reflected in the electric current needed to drive the tube oscillations. 01

4 CORIOLISMASTER MASS FLOWMETER DIAGNOSTICS, VERIFICATION AND PROOF TEST WP/CORIOLIS/VERIMASS/101-EN REV. B 02 Control room Learning the baseline driver current VeriMass uses this basic principle and monitors the driver current to detect possible changes of the meter tubes. The major challenge for the use of this kind of verification is the elimination of any short-term process effects. These effects can be accommodated by a learning technique. After installation of a CoriolisMaster mass flowmeter, users can document the initial baseline driver current. The user defines an initial calibration period, during which the measuring device 'learns' the standard behavior of the flowmeter in the specific application. VeriMass technology monitors the driver current to determine its initial baseline, while eliminating any short-term effects. This selfcalibration period adapts VeriMass to demanding processes such as high viscosity liquids, temporary gas phases in liquids or liquid drops in gases. The initial calibration duration can be a few minutes up to several weeks. But for wellbehaved processes the self-calibration period rarely requires more than a few days following installation. The noisiness of the driver current signal during the self-calibration period sets the tightness of the alarm threshold. Smoother processes will be monitored with tighter alarm thresholds than noisier processes. This avoids possible false alarms. The self-calibration period also defines the verification monitoring time. At the end of each monitoring period VeriMass compares the newly found current level to the initial baseline. Verification monitoring repeats automatically with no further interaction by the user. If a new current baseline exceeds user-defined tolerances, VeriMass sets off an alarm. VeriMass technology works best for diagnosing errors caused by coating and / or erosion of the oscillating tubes. In the case of liquids, it suits processes with changing densities as well as those with only small viscosity changes and temporary gas bubbles. VeriMass can handle changes of different fluids with significantly different viscosities if the changes were present during the initial calibration period following installation. Liquids with erratic viscosities including changes of 50 mpas or more are NOT good application candidates. VeriMass technology also works well with gases having occasional liquid drops. 02 03

CORIOLISMASTER MASS FLOWMETER DIAGNOSTICS, VERIFICATION AND PROOF TEST WP/CORIOLIS/VERIMASS/101-EN REV. B 5 03 Example CoriolisMaster Verification Report. Measurement device status summary would indicate GOOD Operating VeriMass Users can set up VeriMass locally by navigating the menu on the LCD display of the CoriolisMaster flowmeter. More convenient, however, is setup using the CoriolisMaster Device Type Manager (DTM) or ABB s new Field Information Manager using FDI packages. The DTM and FDI packages include a rich and userfriendly graphical interface to make device configuration, maintenance, diagnostics and troubleshooting fast and easy. Users connect to the CoriolisMaster flowmeter via a host computer, HART commissioning device, or directly to a PC. VeriMass technology thus offers automatic monitoring / testing of the Coriolis meter tubes for possible erosion or deposit formation. ABB has developed and tested this innovative procedure to be a 'user-friendly' method to detect the most frequent sources of errors in Coriolis flowmeters. A report generation function supplements the benefits. The procedure includes monitoring of all key measuring device parts that affect the accuracy of the measuring device. The CoriolisMaster flowmeter DTM offers all the interaction capabilities provided by the local display. Additionally, the DTM can generate verification reports. The user can enter date and identification details of the operator who checks the verification, and generate two types of reports. A short report focuses on the erosion and coating monitor, including alarm status, event log, and alarm history. The longer, more complete, report version shows the flowmeters parametrization. The VeriMass option is available for all FCB100 and FCB400, as well as FCH100 and FCH400 CoriolisMaster models. Best practice suggests that a retest of a Coriolis flowmeters calibration should be conducted once a year. Factory calibration is the most complex and costly, but provides the greatest testing depth. In this case the user removes the flowmeter and sends it to the manufacturer for review and recalibration. Less costly on-site calibration with a reference device allows for limited testing depth. Using the VeriMass diagnostic procedure on the other hand, together with onsite inspections, greatly simplifies and speeds calibration. And it makes possible testing depths of > 90 % at very little expense without need for a socalled 'proof test' recalibration either on site or at a test lab. 03 G12268

6 CORIOLISMASTER MASS FLOWMETER DIAGNOSTICS, VERIFICATION AND PROOF TEST WP/CORIOLIS/VERIMASS/101-EN REV. B Other considerations The IEC 61508-2 and VDI / VDE2180 industry standards state that flexibility in determining calibration test intervals, including prolonging these intervals, is possible with existing diagnostic coverage and operating experience. NAMUR NE79 also mentions the desirability of prolonging intervals between test times through improved automatic diagnosis procedures with a wide degree of coverage. Conclusion: prolonging test intervals through internal self-monitoring and online diagnosis is permitted by the standards. Self-monitoring flowmeters have a variety of beneficial applications. Flowmeter usage for billing or allocation, such as for water, steam or gas lines, must guarantee accurate measurements. Applications subject to ISO 9001 Quality Management or ISO 50001 Energy Management also require this capability. Selfmonitoring can also fulfill legal requirements, such as for Emissions Trading (TEHG) and for retesting protective devices in accordance with IEC 61508. All devices from the ABB CoriolisMaster family provide comprehensive internal self-monitoring functionality. Aside from these applications, selfmonitoring offers numerous benefits. The most obvious is easy and quick testing on installed equipment no process disruption, no installation downtime, and no device removal. Yet users achieve a functional test from the sensed measured value through to the output signal with high testing depth. Users get regular verification of error-free operation with a clear pass / no-pass result. Calibration intervals may be prolonged without additional accessories. This in turn leads to increased productivity and installation efficiency.

Notes CORIOLISMASTER MASS FLOWMETER DIAGNOSTICS, VERIFICATION AND PROOF TEST WP/CORIOLIS/VERIMASS/101-EN REV. B 7

ABB Limited Measurement & Analytics Howard Road, St. Neots Cambridgeshire, PE19 8EU UK Tel: +44 (0)870 600 6122 Fax: +44 (0)1480 213 339 Mail: enquiries.mp.uk@gb.abb.com ABB Inc. Measurement & Analytics 125 E. County Line Road Warminster, PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183 ABB Automation Products GmbH Measurement & Analytics Dransfelder Str. 2 37079 Goettingen Germany Tel: +49 551 905-0 Fax: +49 551 905-777 Mail: vertrieb.messtechnik-produkte@de.abb.com abb.com/flow We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB. Copyright 2017 ABB. All rights reserved 3KXF411015R2801 WP/CORIOLIS/VERIMASS/101-EN Rev. B 09.2017