Loopamp Bovine Embryo Sexing Kit Instruction Manual

Similar documents
Bt-Cry1Ac & Bt-Cry2A Multi-trait ELISA Kit DAS ELISA for the detection of Bt-Cry1Ac & Bt-Cry2A proteins Catalog number: PSM 14900

Bt-Cry2Ab and Bt-Cry3Bb1 Multi-trait ELISA Kit DAS ELISA for the detection of Bt-Cry2Ab and Bt-Cry3Bb1 proteins Catalog number: PSM 26800

Gel Loading Practice MiniLab

CspB ELISA Kit DAS ELISA for the detection of CspB protein Catalog number: PSP 35500

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

MOLEBIO LAB #1: Microquantity Measurement

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

User s Manual. Autoclavable & UV resistant. Multi-channel digital micro pipette. In Vitro Medical Diagnostic Devices (98/79/EC)

A N F G J I C D D L A B C D E F

Lab 1. Instrumentation Familiarity: Using Micropipetters and Serological Pipettes

Pipetting and Determining Protein Concentration

Version BA D

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol

Pipettor. User Manual

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Transferpette. Lightweight precise rugged reliable!

Transferpette. Lightweight precise rugged reliable!

RayPette Plus Autoclavable Pipette. User Manual

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Chapter 2. Pipetting techniques. Guide to Pipetting

Biological Activity Assay for Gibco Recombinant Proteins Chemotaxis Assay

PIPETMAN F ENGLISH. Other languages available on our website:

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy

The pipettes cover a volume range from 0.1µl to 10ml.

Nichipet EX Plus II. User s Manual. Autoclavable & UV resistant. Digital micro pipette for liquid handling

User's Manual. Autoclavable Nichipet EX Plus Digital micro pipette for liquid handling CERTIFIED ISO9001

Pipetting Small Volumes

User s Manual. Autoclavable & UV resistant Nichipet EX II. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

Technical Procedure for General Laboratory Equipment

biosensis Human Lipocalin-2/NGAL ELISA Kit Protocol

Adjust the Volume Display

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

User s Manual. Autoclavable & UV r e s i s t a n t. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

A N B P F G J I C D D L A B C D E F

Using a micropipette to transfer exact liquid measurements in scientific experiments By: Nicole Hume

BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

Best Practices for Infinium Assays

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

biosensis Rat Fibronectin ELISA Kit Protocol

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol

Improving Pipetting Techniques

Transferpette. Single and Multichannel Pipettes

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol

Other languages available on our website:

RayBio Human vwf ELISA Kit

MULTICHANNEL. Labnet International, Inc.

Instruction Sheet. Calibration of Multiple 1720 and FT660 Turbidimeters

Biochemistry Laboratory I CHEM 4401 Pipetting & Scales

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

Real World Metric System

ab IgG1 Human ELISA Kit

Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit

Human ABCD1 ELISA KIT

SALX. Interpretation Guide. Salmonella Express System

Microprojectile Bombardment Protocol

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

APPENDIX A CORRECT USE OF DIGITAL PIPETTES

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

Eppendorf Research fix variable multi Instruction Manual Bedienungsanleitung Mode d emploi Instruzioni d impiego Manual de Instrucciones

A N B P G J I C D L D E

[USING THE VITROBOT April 9, Using the Vitrobot

Userts Manual. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC) Annex III self-declared ISO 8655 STANDARD

biosensis Human Soluble Tumor Necrosis Factor receptor type II (stnfrii) ELISA Kit Protocol

RayBio Human TNF-alpha ELISA Kit

WELLS JOHNSON HIGH VOLUME CANISTERS

biosensis Mouse Brain-derived neurotrophic factor (BDNF) ELISA Kit Protocol

TJF-Q180V Cleaning and Disinfection Checklist

MSD 96-Well MULTI-ARRAY CRP Assay

User Manual. Rotor Gene Q and Rotor Gene 3000

TapeStation Tips & Tricks

TJF-Q180V Cleaning and Disinfection Checklist

biosensis Mouse Interleukin-1 beta (IL-1β) ELISA Kit Protocol

Flow experiment with the ibidi pump system and µ-slide I 0.6 Luer

SoCal BioMed, LLC 4695 MacArthur Court, 11th Floor Newport Beach, CA Phone: +1 (424) Fax: +1 (323)

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

Exeter Clinical and Health Research

TITLE: DNA minitube - Red

Productivity. Technician. Maximized. Interpretation Guide. 3M Food Safety 3M Petrifilm Salmonella Express System

Varipette Operating manual

Ocular Irritation: Dilution Method

biosensis Rat Glial cell line-derived neurotrophic factor/gdnf total /ATF ELISA Kit Protocol

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

HUMAN IL6 KITS PROTOCOL

Department of Genetics

biosensis Rat Interleukin-1 beta, IL-1β ELISA Kit Protocol

qnano User Manual QN1-OQ-013D Page 1

Blue Thumb Test Procedures

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

How to Use: 1. Hold the pipette like you are going to shake hands with it.

Title of Manual: Specimen Collection Document Number: GPA To provide the proper collection protocol for Candida Vaginitis testing.

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE:

Fiber Optic SMA Z-Flow Cell Manual Version 2.1

QUIK-jet Micro-Injection System

Natural Reaction Boosters C. V. Krishnan

Application Note Temperature dependence of viscosity of liquid crystal

Transcription:

Loopamp Bovine Embryo Sexing Kit Instruction Manual 1

Instruments required Detection equipment Loopamp End Point turbidimeter LA-100(manufactured by TERAMECS) Pipettes Adjustable micro-pipettes(adjustable from a few μl~ several hundred μl for sampling manipulation) (Examples:0.5~10μL, 10~100μL, 100~1000μL) (If possible, try to use separate pipettes for reaction mix preparation and sample/ positive control handling. ) Tip with filter( RNase/DNase free) Tips Sterilized tubes for master mix preparation( RNase/DNase free) 1.5mL Sterilized tubes for sample pre-treatment( RNase/DNase free) 0.5mL Loopamp reaction tubes, 200μL 8 connected tube For remaining low temperature Equipments used on ice for reaction mix preparation (Example 1:Aluminum rack and crushed ice) (Example 2:Cold boxes) Ice box for sample and positive control handling Others Centrifuge for micro-tubes Centrifuge for 8-connected tubes(0.2ml) Tube rack(0.2ml, 0.5mL, 1.5mL) Vortex mixer (when needed) Disposable gloves Tips and vinyl bag for sealing the disposed tips(such as uni-pack etc.) Bovine Embryo Sexing Kit Kit contents (Bovine Embryo sexing kit) for 24 tests (Storage:-20 Expiration:in 12 months) Extraction Solution Reaction Mix.Ⅰ Reaction Mix.Ⅱ Bst DNA polymerase Control DNA EX RM Ⅰ RM Ⅱ Bst DNA polymerase Cont. DNA 1.5mL 1tube 0.25mL 2tubes 0.25mL 2tubes 60μL 1tube 0.1mL 1tube Reaction tubes 8 12sets 2

Sample Pre-treatment Thawing the Extraction Solution (EX) at room temperature, mix and spin down. Add 6μL of biopsy sample solution which containing bovine embryo cells into the sterilized tubes (for pre-treatment use). Add Extraction Solution 6μL to the tube and mix. Keep it at room temperature for above 5 minutes, mix and spin down. The solution is used as sample solution for the reaction. Biopsy sample solution Extraction Solution(EX) 6μL 6μL Leave at room temperature for at least 5 mins, mix and spin down. For biopsy, take 10% from the trophectoderm. The volume of biopsy sample should be 6μL. Avoid opening the cap of Extraction Solution as far as possible Operation procedures Master Mix for male specific reaction (20μL) Sample solution Positive Control Negative Control (5μL respectively) Master mix for male-female common reaction (20μL) Sample solution Positive Control Negative Control (5μL respectively) Mix and spin down Locate tubes to the End Point turbidimeter LA-100, Reaction start. About 40 mins Transfer the reaction tubes to the detection block Turbidity detection Result print out Sexing determination Equipment and reagents preparation Before handling the reagents, the End Point turbidimeter LA100 should be set ready for the detection. Take out reagents from the refrigerator. ( Thawing RMⅠ, RMⅡ at room temperature, Bst DNA polymerase on ice ) Mix RMⅠ, RMⅡby vortexing or by tapping, spin down the tube. For Bst DNA polymerase, only spin-down is necessary. Vortexing should be avoided. Keep on ice 3

Master Mix preparation Add the necessary amount of RMⅠ, Bst DNA polymerase into the sterilized tube (for preparing male-specific reaction master mix) Add the necessary amount of RMⅡ, Bst DNA polymerase into the sterilized tube (for preparing male-female common reaction master mix) Mix by tapping,or by repeatedly reversing mixing,or by vortexing(1second X 3 times) *; spin down the tube. * Too much fierce mixing by vortex mixer might inactivate the polymerase. Male specific reaction (amount for 1 sample) Male-female common reaction (amount for 1 sample) RMⅠ 20μL RMⅡ 20μL Bst DNA polymerase 1μL Bst DNA polymerase 1μL 1.5mL tube Total 21μL 1.5mLtube Total 21μL After mixing, spin down After mixing, spin down Example:in the case of handling 6 samples Prepare the amount for 8 tubes including PC(positive control), NC(negative control), and samples RM 20μL 8=160μL Bst DNA polymerase 1μL 8= 8μL Total 168μL Master mix dispensing (Reference: Cautions for Pipette manipulation (example 1)) Check to confirm there is no scratch or crack on the Loopamp reaction tubes. Dispense 20μL of the prepared master mix into each reaction tube and lightly close all caps. The reaction tubes containing the master mix should be kept on ice inside the clean bench. Male specific reaction master mix 20μL for each tube Male-female common reaction master mix 20μL for each tube Loopamp reaction tubes PC Samples NC PC Samples NC PC:Positive control NC:Negative control Cautions for handling tubes Please only use the reaction tubes exclusive for the kit. The usage of other reaction tubes might cause misjudgment. Reaction tubes are vulnerable to scratches, please handle with care. Damaged tubes can contaminate the tubidimeter, and the equipment might be disqualified for future use. The sterilized tubes (for master mix preparation) and the reaction tubes should not be exposed to UV light for sterilization. UV light can deteriorate the tube material and cause damages or color changes. These changes might cause non-specific amplification and lead to false results. 4

The preparation of controls and sample solution Positive control Thaw the Control DNA(Cont.DNA)at room temperature, mix and spin down, and then keep it on ice. Negative control Mix Extraction Solution and blastomere washing solution (or biopsy buffer solution) at the ratio 1:1. Mix well, spin down and keep on ice. Sample solution Keep the sample solution on ice. *Positive control can cause contamination, please take extra care. If the spin-down is not sufficient, the solution staying on the cap interior can disperse and cause contamination. Adding negative control ( Reference: Cautions for pipetting manipulation(example 2)) Open the cap of the tube containing negative control. Perform the pipetting manipulation once within the tube and then aspirate 5μL. Move the pipette to the reaction tube. While slowly rotating the pipette within the tube, perform pipetting manipulation for 5~10 times. *Please prevent the solution from being stuck to the inner tube wall. Finally, push the pipette s button to its 1st stop position to dispense the inside solution, and then remove the pipette. *Dispose the tip. *Perform the procedures to male specific reaction and male-female common reaction respectively. Adding sample solution Open the cap of the tube containing sample solution. Perform the pipetting manipulation once within the tube and then aspirate 5μL. Move the pipette to the reaction tube. While slowly rotating the pipette within the tube, perform pipetting manipulation for 5~10 times. * Please prevent the solution from being stuck to the inner tube wall. Finally, push the pipette s button to its 1st stop position to dispense the inside solution, and then remove the pipette. *Dispose the tip. *Perform the procedures to male specific reaction and male-female common reaction respectively Adding positive control Lastly, open the cap of the tube containing positive control. Use the pipette which is exclusive for the positive control, (the procedures are the same as adding negative control and sample solution), to add 5μL Control DNA into the reaction tube. *Dispose the tip. *Conduct the procedures to male specific reaction and male-female common reaction respectively. 5

Confirmations before amplification reaction Confirm the solution volume inside the reaction tubes. (Correct dispensing will assure that all dispensed solution is in the same volume, and the yellow color can be seen. ) Spin down and confirm no bubble inside the tubes. Confirm the reaction block in the turbidimeter has reached the temperature of 63.0. Amplification reaction and detection Locate the reaction tubes to the end-point turbidimeter LA-100, press START/ENTER button. Reaction proceeds at 63.0 for 35 minutes. And after the reaction, inactivate the enzyme at 80 for 2 minutes. While the message MEASURE is flashing, remove the reaction tubes from reaction block to the detection block. Press START/ENTER button. Detection Judging standard Confirm that positive control reaction shows a + and negative control shows a - sign. Male/ female determination: RMⅠ(male specific reaction) RMⅡ(male-female common reaction) Judgment + + male results - + - + - - female retest Cautions for locating reaction tubes Confirm that the tubes are not scratched, foggy, and no foreign matter or bubble inside. Locate the reaction tubes firmly to the reaction block and the detection block. Gently close the hot bonnet and the cover of the detection block. Cautions for handling amplification products After the detection, please wear gloves to remove the reaction tubes. The reaction tubes are vulnerable to damages, please handle with care. After the reaction, the cap of the tubes should not be open, especially while removing the reaction tubes from the turbidimeter. The disperse of amplification products will not only contaminate the turbidimeter and cause false judgment, but also disqualify the equipment for future use. After the reaction, keep the caps of the tubes closed and carry out incineration or dispose the tubes by sealing them in a vinyl bag. In order to avoid dispersing the amplification products, do not carry out high pressure sterilization on disposing the tubes. 6

(Reference) Cautions for pipetting manipulation Recommended procedures for pipetting manipulation on LAMP On carrying out LAMP reactions, pipetting manipulation is required for dispensing of master mix, adding samples/ controls, aspirating and discharging minute amount of solution. Performing the usual pipetting manipulation operation (pushing the button up to the 2 nd -stop position on discharging), it can cause bubbles, which are uneasy to remove. Therefore, it is recommended to push the button only up to the 1 st stop position on dispensing master mix to reaction tubes. Example 1:Dispensing 20μL master mix to each Loopamp reaction tube 1 Firstly, rinse the pipette tip once with the master mix: Push the pipette s button to the 1 st stop position, dip the tip 2~3mm into the master mix solution, gently perform aspiration, and then push the button to the 1 st stop position. 2 Keep tip end at the same position and gently aspirate 20μL from the master mix, and then hold the pipette for a few seconds. 3 Transfer the pipette carrying master mix to the reaction tube. Try to slightly place the tip end on the inner tube wall near the bottom, and gently push the button to the 1st stop position. Keep holding the pipette and slowly remove it from the tube. Although this allows a minute amount of master mix to be left in the tip, this should not be a problem for the detection. Keep holding the pipette and, for the next dispensing, dip the tip into the master mix to aspirate 20μL, remove the pipette and repeat step 3 to dispense it into the reaction tube. Repeat the procedures to dispense the master mix into each reaction tube. *The attempt to discharge all the master mix by pushing the pipette button to the end stop (push up to the 2 nd stop position) will result in the shortage of master mix or cause bubbles in the tubes. 1Pipette tip rinsing 2Aspiration Push to the 1 st stop Gently release the button Push to the 1st stop Slowly release the button Dip the tip of the tip 2~3mm into the master mix. Master mix aspiration Discharge the master mix Master mix aspiration 3Dispensing Gently push to the 1st stop Caution Not to push up to the 2nd stop Gently lay the tip to the tube wall near the bottom Discharge the master mix Keep holding the pipette and remove it from the tube It will cause bubbles in the tube 7

Example 2: Adding 5μL sample/controls into the reaction tube 1 Firstly, rinse the tip once with the sample solution: Push the pipette s button to the 1st stop position, dip the tip 2~3mm into the sample solution, gently perform aspirating and then push the button to the 1st stop position. 2 Let the tip stay at the same position and slowly aspirate 5μL from the sample solution. 3 Transfer the pipette carrying sample solution to the reaction tube. Dip the tip 2~3mm into the master mix in the reaction tube, gently push to the 1st stop position to discharge the sample. And then release the button, while lightly rotating the tip end in the solution, rinse the tip for 5~10 times according to the procedures in 1 (The same times of rinsing operation should be done to each tube). Finally, gently push to the 1st stop position, keep holding the pipette and then remove the tip from the tube. *Pushing up to the 2nd stop might cause bubbles in the tube. 1Pipette tip rinsing 2Aspiration Push to the 1st stop Gently release the button Push to the 1st stop Slowly release the button Dip the tip 2~3mm into the sample solution Sample solution aspiration Discharge the sample solution Sample solution aspiration Gently push to the 1st stop 3Mixing and discharging Release the button, and then again push to the 1st stop (Repeat for 5~10 times) Push to the 1st stop Caution Not to push up to the 2nd stop Dip the tip 2~3mm into the master mix in the reaction tube Discharge the sample to the tube Lightly rotate and rinse the tip. Keep holding the pipette and remove it from the tube It will cause bubbles in the tube. 8