Oxygen Contamination

Similar documents
Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

Leak Checking Large Vacuum Chambers

Table of Contents Want full page pictures with most important parts labeled in the beginning.

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

JANIS OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS

Installation, Operation, and Maintenance Manual

Please refer F 90 of the present manual for the identification of the valves used in the detector.

Vacuum Systems and Cryogenics for Integrated Circuit Fabrication Technology 01

Freedom8 ShoeBox Compressor Manual

Glove Box Installation Manual

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

The Helium Leak Detector

BIMA cryogenics (Dick Plambeck, 18jan2007)

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Welker Constant Pressure Cylinders With Welker Solid Indicator, Syringe T-Handle, or Vortex Mixer

Installation, Operation, and Maintenance Manual

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

VACUUM & AIRLESS. Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs.

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

Vacuum. Dry Turbo Pumping Stations. Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators. Hermetically Sealed Pumps for He3 Recirculation

The Principles of Vacuum Technology

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER COMPOSITE GAS SAMPLER

Matching Vacuum Pump to Process

Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases.

Procedure of Xenon Transfer

Model VR6 System. Installation, Operation & Maintenance

Installation, Operation and Maintenance Manual for Back Pressure Regulator

Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.

TABLE OF CONTENTS SAFETY FIRST!...3 BASIC OPERATION...4 ADJUSTMENTS...5. Cocking Pressure...5. Three-Way Valve Adjustment...6

Installation and Service Manual

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Constant Pressure Crude Oil Container Model CPCCP

TactAir PCP Fill Kit/Adapter. Operation Manual

Bulletin TCR-104 & 109 Filling and adding to the Glycol pressure system

TECHNICAL DATA CAUTION

4400 TwinHybrid Gas Seal

VACUUM CHAMBER MANUAL

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

9.0 Equipment Cleaning and Repair

15ME-014 INSTRUCTION MANUAL OF STERN TUBE SEALING TYPE EVK2RV. URL

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

DURRIDGE STAINLESS STEEL SOIL GAS PROBE

MAINTENANCE PROCEDURE FOR X 650

Instructions for Installation, Operation, Care and Maintenance

AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide

RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats

Chapter 6 Leak detection

OPERATION MANUAL NTF-60 Plus

FE Petro vs. New Red Jacket Competitive Comparison

The latest automotive systems require innovative leak test methods and fixturing.

MODEL NUMBER: P-A AUTOMATIC PORTABLE COMPRESSOR

USER S INFORMATION MANUAL

Installation, Operation, and Maintenance Manual

GRAVITY FLOW DILUTE PHASE PNEUMATIC CONVEYING DENSE PHASE PNEUMATIC CONVEYING ORIFICE GATE. Handles powders, granules and pellets

450P AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

Table of Contents. Rules of Safe Shooting 3. Getting Started 4. Charging the Rifle 5. Loading 6. Safety 7. Looking After Your FX 8.

Aquor House Hydrant V2

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

INSTALLATION & MAINTENANCE INSTRUCTION

CSA Sample Draw Aspirator Adapter Operator s Manual

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER

BS ml Bottle Sampling Unit

School of Chemistry SOP For Operation Of Glove Boxes

Model Series 62 Constant Differential Relay

UltRo Dual Flow Reverse Osmosis Water System

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

OWNER S TECHNICAL MANUAL

OPERATION MANUAL NTF-15

Agilent 1290 Infinity Pump Head Maintenance

INSTRUCTION MANUAL GAS TRAK

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

INSTALLATION AND OPERATION INSTRUCTIONS

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Mass Spec will not Autotune

Instruction manual for FX Gladiator. Instruction manual for FX Gladiator

Major Design Features

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

G7S Hand Pump Owner s Manual

INSTRUCTION MANUAL Kobelco Eagle Marine Water Lubricated Stern Tube Seals

DUQUESNE UNIVERSITY. Liquid Nitrogen Safety

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

with O 2 Controller Instruction Manual

450P- RV AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

Microscopy Cryogenic WORKSTATION. Performance by Design

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER

CVI Valve Line. Exceeds the industry s highest standards for reliability and performance

Baffles and Traps Cryotrap

R-410A 50% R-32 50% R-125.

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

INSTRUCTIONS FOR MODELS UPH & UPHS HIGH FLOW, HIGH PURITY GAS REGULATORS

Transcription:

MPMS Application Note 1014-210 Oxygen Contamination This application note describes potential sources for oxygen contamination in the sample chamber and discusses its possible effects. Molecular oxygen, which undergoes an antiferromagnetic transition at about 43 K, is strongly paramagnetic above this temperature. The MPMS system can easily detect the presence of a small amount of condensed oxygen on the sample, which when in the sample chamber can interfere significantly with sensitive magnetic measurements. Oxygen contamination in the sample chamber is usually the result of leaks in the system due to faulty seals, improper operation of the airlock valve, outgassing from the sample, or cold samples being loaded. Oxygen Leaking into the Sample Chamber The sample chamber in the MPMS is leak-proof with respect to the ambient room atmosphere. A 5mm Hg pressure of helium gas is normally maintained in the sample chamber to provide for thermal exchange between the sample and the sample chamber walls. As the pressure in the sample chamber falls much lower than the ambient room pressure, any leaks that might develop will result in a steady flow of air into the sample chamber. Large and Small Air Leaks Large air leaks in the sample chamber, caused by a missing o-ring, for example, can cause the green ready light not to light up at the end of the airlock purging cycle. The green light doesn t light up because the pressure in the sample chamber isn t falling low enough to trigger the indicator lamp. Smaller leaks, however, can allow a continuous flow of air molecules into the sample chamber without significantly affecting the background pressure in the sample chamber. Leaks at Low and High Temperatures Accumulated oxygen is a consequence of an air leak in the sample chamber, particularly when making measurements below 100 K. At higher temperatures, oxygen that enters the sample chamber will diffuse out to the vacuum pump, resulting in a low concentration of oxygen in the sample chamber. Therefore, at higher temperatures the small amount of oxygen will have relatively little effect on the measurements. Contamination at Lower Temperature At lower temperatures, particularly below 100 K, the walls of the sample chamber, the sample rod, and even the sample itself act as cryopumps that cause the incoming oxygen to continuously condense on the cold surfaces. The easiest way to detect oxygen contamination Quantum Design MPMS Application Note 1014-210B 1

on your sample is to look for the antiferromagnetic transition in oxygen at 43 K as you are measuring the temperature dependence of your sample. The presence of this highly paramagnetic condensate can interfere with sensitive magnetic measurements by: giving incorrect values of the measured moment and introducing incorrect temperature and field dependencies into the data. making the antiferromagnetic transition of the oxygen contamination appear as an increased paramagnetism in the sample at about 43 K if the sample moment is measured as a function of temperature. making the standard deviation of the individual measurements higher than normal due to oxygen either condensing onto or evaporating off of the sample from one scan to the next if several scans are averaged together. As long as the temperature of the sample chamber walls and the sample itself are still settling, oxygen will continually move about the sample chamber, evaporating from one surface only to condense on another. allowing the standard deviations to decrease to a level typical of normal operation if additional time is allowed for the MPMS system temperature to stabilize. However, if the system is contaminated with oxygen, this means only that the oxygen is not moving appreciably during the time required to collect the data at one temperature. Figure 1 shows the temperature dependence of the magnetic susceptibility of solid oxygen. Figure 2 and Figure 3 show the effect of oxygen contamination on measurements of a palladium reference sample. Quantum Design MPMS Application Note 1014-210B 2

Figure 1. Temperature Dependence of the Magnetic Susceptibility of Solid Oxygen. The curve represents the susceptibility of 250.1 mg of bulk oxygen, redrawn from Stephen Gregory's paper [Physical Review Letters 40, 723 (1978)]. α, β, and γ are three different phases of solid oxygen. α and β are anti-ferromagnetic, whereas γ is paramagnetic. It is important to note that the β γ paraantiferromagnetic transition is a sharp step only for bulk solid oxygen. It is likely that any small leak into the MPMS sample chamber will result in a condensed film of oxygen rather than the bulk solid. This will tend to broaden the step transition, as in Figure 2 and Figure 3. The thinner the condensed film, the smaller and broader is the transition. Please see the reference listed above for further details. Quantum Design MPMS Application Note 1014-210B 3

Figure 2. The Effect of Oxygen Contamination on Measurements of a Palladium Reference Sample. The triangles represent the measured moments of a 0.2655g palladium sample in a 3 tesla field at various temperatures from 5 K to 100 K. The squares represent the measured moments of the same palladium sample in the same field. However, about 5.6x10-3 moles of oxygen were added to the sample chamber in addition to the few torr of helium already present. The sample chamber was then sealed off, trapping the oxygen. Figure 3. The Effect of Oxygen Contamination on Measurements of a Palladium Reference Sample that was introduced at 5K. The triangles represent the measured moments of the palladium sample (as in Figure 2) in a 3-tesla field. The squares represent the measured moments of the same palladium sample contaminated with about 5.6x10-3 moles of oxygen in a 3-tesla field. The difference from Figure 2 is that here the oxygen was added at 5K, and the system was warmed as measurements were made with the MPMS in normal mode (the sample tube was open to the pump). This results in the oxygen pumping away around 65K, resulting in a sharp step. It should be noted that the temperature at which the oxygen is pumped away is likely to be different for differing levels of contamination. Quantum Design MPMS Application Note 1014-210B 4

Removing Oxygen Contamination in the Sample Chamber 1. Raise the system temperature to about 250 K so that there won t be any cold spots above the sample chamber where the oxygen can recondense. 2. With the airlock valve open, press PURGE AIRLOCK on the front of the MPMS probe. This allows the system to carry out a purge cycle. This is similar to purging air from the airlock when loading a sample, except that the entire sample chamber will be flushed and purged instead of just the airlock. 3. If oxygen is building up in the sample chamber during a measurement sequence, use sequence control to purge the sample chamber to cleanse the sample chamber of oxygen. However, if the oxygen contamination is coming from a leak in your system, the oxygen will soon reappear. The level of contamination will increase steadily with time as the gas continuously condenses onto the cold surfaces in the sample chamber. If you are observing continual oxygen buildup in your instrument, perform a leak test for the sample chamber. Performing a Leak Test in the Sample Chamber If you observe oxygen contamination with your sample, check the sample chamber for leaks. Figure 4 shows the positions of the seals to be tested. 1. Set the system temperature to 310 K. Once the temperature has stabilized, load a sample rod without a sample in the instrument as if you are going to make a measurement. 2. Leave an atmosphere of helium gas in the sample chamber by activating Vent Sample Space. 3. Disconnect the smaller flexible pumping line at the rear of the MPMS control console and connect this line to a helium leak detector. If you need an adapting connector to connect the pumping line to your leak detector, please contact your Quantum Design service representative. 4. Vent and flush the sample chamber with air (not helium) several times to eliminate the background signal from the helium that is already in the sample chamber. You must flush the sample chamber through your leak detector. Do not use the MPMS controls. 5. Open and close the airlock several times to release helium trapped in the grease of the o- rings. 6. Once the helium leak detector is operating, you can find the leak in the system by spraying helium gas around each of the sealing areas indicated in Figure 4. To test the seals around the airlock, spray helium gas into the hole around the airlock valve shaft that protrudes from the front of the unit. If a leak is there, contact your Quantum Design service representative. 7. If you find a leak around the slide seal assembly, clean and re-lubricate the o-rings underneath the slide seal assembly (replace them if they are cracked or damaged) and perform the leak test again. 8. The test may be repeated using a different slide seal assembly each time you test each of your slide seal and sample rod assemblies. Quantum Design MPMS Application Note 1014-210B 5

9. After the tests are completed, reconnect the flexible pumping line to the rear of the MPMS console. Press PURGE AIRLOCK on the front of the unit to restore the system to its operating condition. Sources of Oxygen Contamination from Assemblies in the System The sample chamber on each unit is thoroughly tested before shipment to ensure there are no leaks between the sample chamber and the ambient atmosphere. However, under steady use of the instrument, oxygen can be introduced into the sample chamber through several mechanisms or by a leak at one of the seals between the sample chamber and the ambient atmosphere. The locations of the various seals protecting the sample chamber are shown in Figure 4. Figure 4. The Slide Seal Assembly Quantum Design MPMS Application Note 1014-210B 6

Air Leaks at the Slide Seal Assembly Lip seals experience the greatest amount of wear and are the most likely sources for air leaks. Lip seals are inside the blue plug of the slide seal assembly where the sample rod enters the sealed sample chamber. The slide seal assembly provides a double sliding lip seal for the sample rod so that helium gas from the dewar constantly flows through the region between the seals. This gas is then exhausted through the hose fitting on the rear of the sample transport mechanism. Air is dragged through the upper seal, diluted by the helium gas between the seals, and to a large extent, carried away by the flowing stream of helium. If the lip seals are damaged, air will continuously leak through both seals and directly into the sample chamber. In this case the slide seal assembly must be disassembled and the lip seals replaced. You can return slide seal assemblies to the factory to be rebuilt purchase required parts from Quantum Design if you wish to rebuild the units yourself. Contact your Quantum Design service representative for additional information. Quantum Design recommends replacing the lip seals in the slide seal assemblies at least once a year. Using a liberal amount of Apiezon M grease on the stainless steel sample rods keeps the seals well lubricated and makes them last longer. Also, when you remove a cold sample from the sample chamber, lift the rod slowly so that the sample rod can warm up before it is pulled through the lip seals. Drawing the sample rod through the lip seals while the grease and sample rod are cold causes them to freeze at the sealing point and tear. Air Leaks in Seals at the Top of the Sample Transport Unit Air leaks can also develop around the three o-rings directly under the slide seal assembly when the o-rings become dry or cracked or dirt and dust have accumulated around them. Leaks here can usually be repaired quickly and easily. Always keep the airlock valve in the closed position when performing this operation to allow easy retrieval of the small o-rings. It is easy to drop one down the sample port. 1. Remove the sample and close the airlock valve. If any of the o-rings are cracked or damaged, contact your service representative for replacements. 2. Remove and clean the o-rings and their grooves. 3. Apply a small amount of vacuum grease to the o-rings and reinstall them. Air Leaks in Seals under the Sample Transport A less likely place for an air leak is the double o-ring seal under the sample transport mechanism (see Figure 4). 1. Remove the sample transport mechanism to reach the o-rings. 2. Remove the sample and close the airlock valve. Remove the four screws that hold the transport mechanism in place (two at the front and the rear). 3. Lift the sample transport unit off the top of the probe. You can now remove the o-rings. If they are cracked or damaged, replace them with spares from the MPMS Utility Kit. Before reinstalling the o-rings and sample transport, clean the o-rings and the o-ring grooves and relubricate the o-rings with vacuum grease. Quantum Design MPMS Application Note 1014-210B 7

Other Sources of Air Leaks Other seals from which air can leak into the sample chamber are located at the top and bottom of the sight glass at the front of the sample transport and around the airlock valve assembly. Replacing any of these seals requires parts and more detailed assistance from your service representative. If you have checked the system for leaks and have definitely identified an air leak at one of these locations, contact your service representative for further assistance. Venting the Sample Chamber to Ambient Environment Even if the sample chamber is completely leak-tight, oxygen can enter the sample chamber if the airlock valve is opened and the top of the sample transport is opened to the ambient atmosphere. If the sample chamber temperature is sufficiently cold when exposed to the atmosphere, air will immediately begin to solidify onto the walls of the sample chamber. If the airlock is subsequently closed and a new sample installed in the machine in the normal manner, frozen atmospheric gasses will still be trapped in the sample chamber. In this case, purging the airlock and sample chamber can eliminate the contamination problem. However, to purge water vapor and carbon dioxide, the sample chamber temperature should be increased to about 350 K for 10-15 minutes before purging the sample chamber. Outgassing of Samples Oxygen can also accumulate in the sample chamber as a result of outgassing from samples or from air condensing on samples that are already cold when loaded into the instrument. If such samples are run continuously, and the sample chamber is never warmed up to evaporate the condensed oxygen, the oxygen contamination can become large enough to interfere with your measurements. Periodically warming the system to 250 K and purging the airlock and sample chamber can avoid this problem. Quantum Design MPMS Application Note 1014-210B 8