Use of Instrument at Different Altitudes

Similar documents
O2 Portable. User Manual

ANALOX O2 PORTABLE OPERATORS MANUAL

ANALOX ATA - Trimix Analyser. Technical Manual. Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK

ANALOX O2 EII Oxygen Analyser User Manual for your Personal Protection

O2 EII Pro. Technical Manual

Engineering Note. Algorithms. Overview. Detailed Algorithm Description. NeoFox Calibration and Measurement. Products Affected: NeoFox

MEC OEM. User Manual. Commercial in Confidence

! All activations or settings on the revodream always start when the unit is OFF (except switching OFF)

MEASURING WATER CONTENT IN COMPRESSED BREATHING AIR IN COMPLIANCE WITH EN 12021:2014 ULTRA-PRECISE AND ULTRA-RELIABLE WITH POLYMER TECHNOLOGY

Holistic Atmosphere Management System (HAMS)

CP Chapter 13/14 Notes The Property of Gases Kinetic Molecular Theory

Diving at Altitude. Part One: Starting with Residual Nitrogen. By John Adsit January 30, 2018

Operation manual Oxygen sensor

Sub MkIIP Oxygen and Carbon Dioxide Monitor. User Manual

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

Ambient Weather GL150-B Nautical Barometer User Manual

Ascent to Altitude After Diving

Exercises The Atmosphere (page 383) 20.2 Atmospheric Pressure (pages )

DIVING PHYSICS EXAMPLE QUESTIONS

Kinetic-Molecular Theory

Air Monitor Calibration, Bump Testing, and Sensor Challenge

PHYSICS. Light: Sound:

To convert to millimeters of mercury, we derive a unit factor related to the equivalent relationship 29.9 in. Hg = 760 mm Hg.

Chem 110 General Principles of Chemistry

How does atmospheric pressure vary? Measuring atmospheric pressure at different altitudes above sea level

Planning Decompression Diving Gas Requirements in Open Water

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

The Discussion of this exercise covers the following points: Range with an elevated or suppressed zero Suppressed-zero range Elevated-zero range

Chapter 10 Gases. Characteristics of Gases. Pressure. The Gas Laws. The Ideal-Gas Equation. Applications of the Ideal-Gas Equation

EYE DOMINANCE. You WILL be one of the three. If you re not sure, use the provided test sheet to check.

Service ISO17025 explained

NMISA New Gas Flow Calibration Facility

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

Problem Solving. Gas Laws

Service Calibration are you doing it properly?

Exploring the relationship between Heart Rate (HR) and Ventilation Rate (R) in humans.

Engineering: Measurement Technology Pressure/Level (SCQF level 6)

Example: 25 C = ( ) K = 298 K. Pressure Symbol: p Units: force per area 1Pa (Pascal) = 1 N/m 2

How does atmospheric pressure vary? Measuring atmospheric pressure at different altitudes above sea level

c) Loss of product or replacement when units are being serviced All of these are factors that need to be considered by the user.

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Boyle s Law Practice Problems Name:

Setting up and carrying out a check source measurement

PURE SUBSTANCE. Nitrogen and gaseous air are pure substances.

Let s talk about Capnography

PRODUCT CONFORMITY CERTIFICATE

CHAPTER 31 IDEAL GAS LAWS

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CP Chapter 13/14 Notes The Property of Gases Kinetic Molecular Theory

Thank you for downloading. Help Guide For The Lay Back Horse Racing Backing Spreadsheet.

DFT Valves. Selecting Control Valves

THE WAY THE VENTURI AND ORIFICES WORK

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

FLUID POWER FLUID POWER EQUIPMENT TUTORIAL OTHER FLUID POWER VALVES. This work covers part of outcome 2 of the Edexcel standard module:

Chemistry Honors - Gases

8. Now plot on the following grid the values of T (K) and V from the table above, and connect the points.

HYDROGEN - METHANE MIXTURES : DISPERSION AND STRATIFICATION STUDIES

Behavior of Gases. Gases are mostly The molecules in a gas are separate, very small and very

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

Moisture levels in compressed breathing air. Prepared by QinetiQ Limited for the Health and Safety Executive 2006 RESEARCH REPORT 427

Phase Changes * OpenStax

This test shall be carried out on all vehicles equipped with open type traction batteries.

13.1!"#$#%"&'%()$*+%,+-.$+/*$#

Trade of Metal Fabrication. Module 1: Basic Fabrication Unit 3: Tools and Equipment Phase 2

The University of Hong Kong Department of Physics Experimental Physics Laboratory

(AS AT 31 st MARCH, 2002)

INSTRUCTIONS FOR USING HMS 2016 Scoring (v1)

Variation of Pressure with Depth in a Fluid *

Digital Melting Point Apparatus

Hyper-Gas MkII TM Bell Unit

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

Auto-Zero Calibration Technique for Pressure Sensors

Unit 8: Gases and States of Matter

Lung Volumes and Capacities

BREATH-BY-BREATH METHOD

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

12m. Instructions For Use. 14m. 16m. 18m. Recreational Dive Planner. 20m DISTRIBUTED BY PADI AMERICAS, INC.

OIL AND GAS INDUSTRY

Flow Visualisation Studies - I

Pressure Measurement

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

Welcome to Flogas. Metered Estates. You would never know you were off the grid. Metered Estates Welcome Pack.indd 1 18/01/ :48

Instruments for Measurement of Dissolved CO 2 and O 2 in Beverages

AUK Environmental Sensor. Determining the refractive index of the air. AUK Environmental Sensor

REAL LIFE GRAPHS M.K. HOME TUITION. Mathematics Revision Guides Level: GCSE Higher Tier

Experiment 12: MOLAR VOLUME OF AN IDEAL GAS

Battery Powered Thermal Mass Flow Meters for Gases

Best Practice for Calibrating LTH Conductivity Instruments

User Manual for the Mars Calibration Bench

MicroTim XB. User Manual. Precision Digital Barometric Altimeter / Barometer / VSI. Document Revision 1.0 Firmware Version 3.0

T-TOUCH II User s Manual

Chapter 1 The Atmosphere

Chapter 12. Properties of Gases

Life Support Team Mission Day Instructions

1. [Chang7 5.P.013.] Convert 295 mmhg to kpa. kpa Convert 2.0 kpa to mmhg. mmhg

Transcription:

ANALOX O2EII - Oxygen Analyser Use of Instrument at Different Altitudes Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, Cleveland, TS9 5PT, UK T: +44 (0)1642 711400 F: +44 (0)1642 713900 W: www.analox.net E: info@analox.net

Introduction The Analox O2EII analyser displays oxygen in the range 0.0-100.0 %. It is intended for use at atmospheric pressure. If the analyser is used as recommended in the User Manual, it will work for you in almost all situations. However, at extreme altitudes it may not be possible to achieve the correct calibration. This paper aims to explain to you exactly how the analyser works, and how to use the analyser at extreme altitude if the normal calibration method fails. Any users still unsure should contact Analox for further advice. 2

Effect of atmospheric pressure variation Let us assume that the analyser is calibrated in air to read 20.9% when it is at an atmospheric pressure of 1013.25mBar (Standard atmospheric pressure). If the analyser is then moved to a location with a different atmospheric pressure (but still 20.9% oxygen), then the reading will either rise with increasing pressure, or fall with decreasing pressure. This is because the sensor is sensitive to the partial pressure of oxygen. At 1013.25mBar pressure, the partial pressure of 20.9% oxygen is 211.77 mbar ppo2 (20.9% of 1013.25). So by adjusting the calibration knob to make the display read 20.9%, what we are actually doing is normalising the display to read 20.9% for a ppo2 of 211.77. If the atmospheric pressure increased to say 1030mBar, then the partial pressure of oxygen in the atmosphere would now be 215.27mBar ppo2. The display reading would rise to 215.27/211.77 * 20.9 = 21.2% Similarly if atmospheric pressure decreased to say 980mBar, then the partial pressure of oxygen in the atmosphere would now be 204.82mBar ppo2. The display reading would fall to 204.82/211.77 * 20.9 = 20.2%. Increases or decreases in atmospheric pressure are experienced as different weather fronts move across a region. Typically the figures quoted above would only be experienced over a matter of several hours or even days and hence do not interfere with the sequence of an air calibration followed by a check on a bottled sample. Exercise caution if ever you are attempting to calibrate and check a gas bottle in the immediate vicinity of a storm when atmospheric pressure may vary much more dramatically. Much more extreme cases of low pressure occur simply by taking the analyser to a higher altitude. Remember though that the constant of importance is that fresh air still contains 20.9% oxygen. Earlier versions of the O2EII manual contained a table which stated atmospheric pressure at various altitudes. Table 1 shows a more widely recognised table which we are now using. So as an example of altitude effects, if the O2EII mentioned above was calibrated at sea level at standard atmospheric pressure, and then taken to say an altitude of 10000 feet (3048metres), the atmospheric pressure would be 697 mbar, and the partial pressure of oxygen would be 145.67mBar ppo2. The display would read 145.67/211.77 * 20.9= 14.4% oxygen. 3

Table 1: Atmospheric pressure variation with altitude Figure 1 Altitude (feet) Altitude (metres) Pressure (Bar) Pressure (kpa) 0 0.0 1.013 101.3 1000 304.8 0.977 97.7 2000 609.6 0.942 94.2 3000 914.4 0.908 90.8 4000 1219.2 0.875 87.5 5000 1524.0 0.843 84.3 6000 1828.8 0.812 81.2 7000 2133.6 0.782 78.2 8000 2438.4 0.753 75.3 9000 2743.2 0.724 72.4 10000 3048.0 0.697 69.7 11000 3352.8 0.670 67.0 12000 3657.6 0.644 64.4 13000 3962.4 0.619 61.9 14000 4267.2 0.595 59.5 15000 4572.0 0.572 57.2 16000 4876.8 0.549 54.9 17000 5181.6 0.527 52.7 18000 5486.4 0.506 50.6 19000 5791.2 0.485 48.5 20000 6096.0 0.466 46.6 21000 6400.8 0.446 44.6 22000 6705.6 0.428 42.8 23000 7010.4 0.410 41.0 24000 7315.2 0.393 39.3 25000 7620.0 0.376 37.6 26000 7924.8 0.360 36.0 27000 8229.6 0.344 34.4 28000 8534.4 0.329 32.9 Atmospheric Pressure variation with Altitude Altitude (Feet) 40000 12000 35000 11000 10000 30000 9000 25000 8000 7000 20000 6000 15000 5000 4000 10000 3000 5000 2000 1000 0 0 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 Pressure (Bars) Altitude (Metres) 4

Does all of this matter? No not really! We say in the user manual that air calibration is essential before every use. Why do we say this? It s not because the oxygen cell specifically needs to be calibrated so often, but it is to ensure that the variability of atmospheric pressure is removed from the equation. So let s take our example above at an altitude of 10000 feet. Air calibrate the O2EII at 10000 feet to read 20.9%. This means that we have set up the analyser to display 20.9% for a partial pressure of oxygen of 145.67mBar ppo2. Let s assume that we then want to check the contents of a bottle that is believed to contain 40% oxygen and 60% nitrogen. 40% oxygen released to the ambient atmospheric pressure will equate to a partial pressure of 278.8 mbar ppo2. This will cause the analyser (with its just calibrated settings) to read 278.8 / 145.67 * 20.9 = 40.0%. Surprise, surprise. The analyser reads 40% and we didn t really need to understand all the calibration factors we only had to air calibrate to 20.9%. However, an understanding of what the analyser is doing will hopefully reassure you that it is working correctly. Prior to calibration at 10000 feet altitude, you may have been concerned to note that it was reading 14% oxygen, but as you can see from the above, this is entirely expected. The converse would happen if you now took the analyser back to sea level from 10000 feet altitude. If you took an air reading before calibrating, you would now expect the analyser to read 211.77/156.67 * 20.9 = 28.3%. No, the atmosphere hasn t suddenly experienced a dramatic increase in oxygen levels you simply need to recalibrate to the new atmospheric pressure. 5

So when would I need to understand all of this? What we ve been saying above is that you really shouldn t have to understand the internal operation of the sensor to measure the percentage content of your oxygen mix. However, as with all types of technical information, there may come a time when you do need to understand it. In previous versions of the O2EII manual, we included a section for use at altitude which may have confused some users. What we were trying to say was that at extreme altitudes, you may run out of adjustment of the calibration knob and be unable to obtain a reading of 20.9%. It isn t really possible to state a specific altitude at which this will occur, because it will also depend on the age of your oxygen sensor, and the ambient temperature. If you can t obtain the required 20.9% figure, then the alternative is to use the display to show the partial pressure of oxygen. Some users assumed they had to do this at altitude irrespective of whether they managed to achieve the 20.9% air calibration. 6

Using as a partial pressure analyser You should only have to do this if you are unable to achieve an air calibration as described on page 5. However, there is no reason why you can t use the analyser this way if you want to work in partial pressures. The downside is that you ll probably need a calculator to assist you, whereas if you use the analyser as recommended you will not need a calculator! So let s go back to some of the locations described above At sea level and standard atmospheric pressure of 1013.25 mbar, we know that fresh air contains 20.9% oxygen, so we should calibrate the analyser to 20.9% of 1013.25 = 212 mbar (which you can obtain by adjusting the calibration knob to make the display read 21.2. Ignore the fact that the decimal point is on simply read the display as 212mBar. Now if you measure a 40% oxygen mix, the reading you will obtain will be 40/20.9* 212 = 405 mbar. Remember we re at 1013.25mBar, so 40% oxygen has a partial pressure of 405.3, so that s exactly what we expect. Now take the analyser to 10000 feet altitude (697mBar atmospheric pressure). Without re-calibrating the sensor, we would expect it to read 20.9% of 697mBar = 146mBar. You ll actually find that the analyser needs very little adjustment (if any) to achieve this setting. So let s assume the analyser now reads 146mBar. If you now test your 40% oxygen mix, we would expect the reading to be 40% of 697mBar = 279mBar. So for the purpose of measuring the oxygen mix if we don t know what it contains, we have to take the partial pressure reading (279mBar) and divide this by the atmospheric pressure to work out the fraction of oxygen in the mix. 279mBar/697mBar gives a fraction of 0.40 (which equates to 40% oxygen). 7

Conclusions/Recommendations This paper aims to inform you of exactly how your O 2 EII works. In 99.9% of applications, if you follow our advice to calibrate the analyser every time you use it, then you will not need to understand the complexity of the pressure calculations. However, in rare circumstances, you may be unable to achieve the required calibration. On such occasions, you could choose not to dive because you are unable to measure your oxygen mix, OR you could continue your diving activities if you use the analyser as a partial pressure analyser. You would have to be prepared to do some calculations, and you may also need a barometer, but if you understand the notes, there is no reason why you can t measure your oxygen mix safely and accurately. 8