Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Similar documents
Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

ATIS OPERATION MANUAL

STANDARD OPERATING PROCEDURES

METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Gas Diluter Portable Gas Dilution System User s Manual

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

3 Flow and Pressure Control

METHOD TO-2 Revision 1.0 April, 1984

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Retention Time Locking: Concepts and Applications. Application

Automated Determination of Dissolved Gases in Water Anne Jurek

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

Carrier Gases in Capillary GC

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

GCMSD-Headspace Analysis SOP

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

FROG-4000 TM Chemical Analysis System

Rapid and Reliable Detection of Dissolved Gases in Water

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa

no peaks missing peaks too small peaks increasing retention decreasing retention

Generating Calibration Gas Standards

Name: Class: Date: SHORT ANSWER Answer the following questions in the space provided.

FROG-4000 TM Chemical Analysis System

Experiment 13 Molar Mass of a Gas. Purpose. Background. PV = nrt

6890 GC Site Preparation

Quantitative Analysis of Hydrocarbons by Gas Chromatography

CORESTA RECOMMENDED METHOD Nº 67

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

User manual. TC-20 Sample conditioning and dry-purge rig. QUI-0003 Version 6.3 September

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Standard Test Method for Residual Acrylonitrile Monomer Styrene-Acrylonitrile Copolymers and Nitrile Rubber by Headspace Gas Chromatography 1

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

STANDARD OPERATING PROCEDURES

Experiment GC : Analysis of BTEX by GC-FID

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Zebron Gas Management

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

STANDARD OPERATING PROCEDURES

NCASI METHOD IM/CAN/WP IMPINGER/CANISTER SOURCE SAMPLING METHOD FOR SELECTED HAPS AND OTHER COMPOUNDS AT WOOD PRODUCTS FACILITIES

Agilent G3180B Two-Way Splitter Kit

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Carrier Gases in Capillary GC

EXPERIMENT 8 Ideal Gas Law: Molecular Weight of a Vapor

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

This standard is due for Five Year Review. This process is required by the SEMI Regulations to ensure that the standard is still valid.

The Polyethylene Casing as Diffusion Barrier for Polyurethane Insulated District Heating Pipes

1. Compute the TWA for the following respirable dust exposures:

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

A New GC/MS System Designed for Helium Carrier Gas Conservation

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

5890II GC Standard Operating Procedure 9/2/2005

Honors Chemistry Unit 7 Gas Laws Notes

APPLICATION NOTE. GC Integrated Permeation Device

8. Now plot on the following grid the values of T (K) and V from the table above, and connect the points.

You should be able to: Describe Equipment Barometer Manometer. 5.1 Pressure Read and outline 5.1 Define Barometer

Best Practice for Identifying Leaks in GC and GC/MS Systems

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

Unit 8: Gases and States of Matter

3. Coupled Techniques in Thermal Analysis

STANDARD OPERATING PROCEDURES

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4

MV5: Automated concentration system

MMS Mercury Monitoring System for Natural Gas

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Kinetic Molecular Theory imaginary Assumptions of Kinetic Molecular Theory: Problems with KMT:

A SCENT OF FUTURE. Master DHS. Dynamic Head Space Sampler. Operating Manual

MicroJet Cryo-Trap MODEL MJT-1030Ex OPERATION MANUAL

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Quick Reference Guide. OPTIC 3 - PTV on Shimadzu GC-2010 Gas Chromatograph

CHM 111 Unit 5 Sample Questions

FROG-5000TM Chemical Analysis System User s Manual

World of Chemistry Notes for Students [Chapter 13, page 1] Chapter 13 Gases

Space product assurance

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Super-Clean Gas Filters from Restek

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

NOTE: Performance of this method should not be. attempted by persons unfamiliar with the operation of a gas

Buck Scientific, Inc.

Example: 25 C = ( ) K = 298 K. Pressure Symbol: p Units: force per area 1Pa (Pascal) = 1 N/m 2

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

UNIT 10 - GASES. Notes & Worksheets - Honors

System Overview. TCD Detector temperature setpoint Regulator

Gilbert Kirss Foster. Chapter 10. Properties of Gases The Air We Breathe

Funsheet [WHAT IS PRESSURE AND TEMPERATURE] Gu 2015

Determination of the Gas-Law Constant (R) using CO2

Transcription:

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National Environmental Monitoring Conference August 4, 2014 sigma- aldrich.com/analy1cal

What will be Covered What is Thermal Desorption Air Concentration Units Overview of the tube spiking technique we ve been using with compressed gas standards. Liquid Calibration Standards µg/ml of solvent e.g. Methanol Compressed Gas Calibration Standards PPBv by (volume or by mole)

What is Thermal Desorption? A sample preparation technique for gas chromatography. The sample is collected onto an adsorbent packed glass or stainless steel tube. The sample is concentrated on the adsorbents. The packed tube is heated (Thermal) and the compounds are released into the carrier gas (Desorption) where they are swept onto the GC column and analyzed by the gas chromatograph. GC Detector Carrier Gas ~300 C ~300 C Thermal Desorber Transfer Line Gas Chromatograph

Multi-Bed Thermal Desorption Tube 3-Bed Tube (Carbotrap 300)

Wrong Assumptions of Air Concentration Units µg/l is equivalent to ppbv - this is false µg/m 3 is equivalent to ppbv - this is false The M.W. of the compound, the temperature, and barometric pressure are needed to convert between the units Using a Molar Volume 24.46 L/mole gets me close enough - this is false (This value only applies when your at the beach)

Calibration Process Using Liquid Standards Typically the data handling system is setup to calculate the mass of contaminate(s) found on the tube (e.g. ng/tube, or µg/ tube). (Online or Offline) The sample volume that passed through the tube is divided by the mass found on tube to calculate the air concentration in mass per volume units. (e.g. ng/l, or µg/m 3 ) If spike tubes using liquid standards the process is straight forward, since the mass of all the compounds are at the same mass concentration (e.g. 2000 µg per ml of solvent). Simple dilutions of the liquid standard allow us to make the multi-point calibration curves.

Calibration Process Using Gas Standards Most compressed gas standards are produced so the compounds are at the same ppmv, or ppbv conc. But, the mass of each compound will be different for each compound due to it's molecular weight. Creating a curve based on units of Mass (e.g. ng/tube): Requires the user to manually enter the mass concentration for each compound in the data handling system while creating the curve. OR, Creating a curve based on units of Concentration (e.g. ppbv): Only a single value is required for each level of the curve. But for a ppbv curve to calculate the correct concentration, the sample volume must be accounted for by using scaling factors in the data handling system prior to analysis.

Creating Calibrations Curves the Hard Way To create calibration curves using gas standards often times users develop elaborate processes to dilute the gas mix using mass flow controllers etc., or create complex dilutions using a series of air sampling bags to spike their tubes. There is an an easier way By simply removing a series of representative syringe volumes direct from the gas cylinder without any dilution.

Advantage of Compressed Gas Standards Methods like EPA TO-17 require the calibration curve to bracket a range of 0.5 to 25ppbv. Using gas standards makes this easy, since the compounds are in the same units. This task is not so easy when using liquid standards, since the ppbv concentration for each compound will be different related to air.

Spiking Tubes by EPA Methods EPA TO-17, and the new proposed Method 325B discusses the option of using either gas or liquid phase standards. Section 6.5 of Method 325B talks about using a GC injection unit, or suitable device that has the means of controlling a carrier gas flow through the tube when spiking calibration standards.

How to use Gas Standards to meet the Methods

How Much of the Gas Mix Do I Need? V syringe = C C expected gas mix x V sample V syringe = Syringe Volume needed in (Liters) C expected = Concentration Expected in the Field (ppbv) C gas mix = Concentration of the Gas Mix (ppbv) V Sample = Sample Volume to be collected on the TD Tubes The amount in the syringe will contain the appropriate mass of each compound, as if you sampled this specific sample volume. C 1 V 1 = C 2 V 2

Example of Creating a Curve Desired Calibration Curve Range: 0.5 to 25 parts per billion (ppbv) Sample Volume to be collected is 4-Liters Concentration of the gas mix: 1000 ppbv Curve Syringe Volume Required 0.5 ppb 2 ml 1 ppb 4 ml 2.5 ppb 10mL 5 ppb 20 ml 10 ppb 40 ml 25 ppb 100 ml Gas-tight syringes with on/off valve

Converting the Gas Mix from ppbv to ng/l Step 1 Step 2 ppbv x M.W. = ng/l 24.46 L/mole Use the Molar Volume based on the Temp & Pressure of your lab ng/l x Syringe Volume (L) = ng in syringe

How to Calculate Molar Volume for my Location? Equipment Required: Thermometer Barometer MolarVolume = o Lab Temp ( C) + 273.15 Lab Pressure(mmHg) x 62.36 Universal Gas Constant Values for Different Pressure Units: atm: 0.08205 kpa: 8.315 mm Hg or (torr): 62.36 Inch Hg: 2.246

Mass per Tube Based on a 4-Liter Sample Temperature ( C) Pressure (mmhg) Molar Volume 0.5ppb 1ppb 2.5ppb 5ppb 10ppb 25ppb 23 735 25.11 Syringe Volume (Liters) Actual Values from Cylinder Certificate Mass Conc. from Cylinder 0.002 0.004 0.010 0.020 0.040 0.100 Compound M.W. ACTUAL (ppbv) ng/liter ng in syringe Vinyl chloride 62.5 1050 2613 5 10 26 52 105 261 1,3-Butadiene 54.1 1050 2262 5 9 23 45 90 226 Acrylonitrile 53.1 1070 2261 5 9 23 45 90 226 Methylene chloride 84.9 1040 3517 7 14 35 70 141 352 Chloroform 119.4 1050 4992 10 20 50 100 200 499 1,2-Dichloroethane 99.0 1050 4138 8 17 41 83 166 414 Benzene 78.1 1050 3266 7 13 33 65 131 327 Trichloroethene 131.4 1050 5494 11 22 55 110 220 549 Tetrachloroethene 165.8 1050 6934 14 28 69 139 277 693

How Much Benzene is contained in 10mL Syringe @ different Laboratory Temperatures & Pressures Lab in Denver, CO (5000 C) My Lab Bellefonte, PA (1000 C) STP Sea Level (0 C) Temp 23 23 C 25 C Sta5on Pressure 626 mm of Hg 735 mm of Hg 760 mm of Hg Molar Volume 29.5 L/Mole 25.1 L/Mole 24.5 L/Mole Mass Contained in the 10mL syringe Concentration of the Gas Mix = 1000ppbv 26.5ng 31.1ng 31.9ng % diff from STP 21% 3% 0%

The Tube Spiking Process Thermometer Multi-Bed TD Tube Injection Glassware Carrier Gas Inlet Heating Block ---- Small M.W. Chloromethane ---- Mid M.W. Benzene ---- High M.W. Dichlorobenzene

Getting the Gas out of the Cylinder Using a syringe adapter makes filling the syringe simple. Set the pressure regulator to ~10 psig when working with glass syringes. Syringe Adapter Septum Inside

Details of Spiking Tubes with a Gas Standard 1. Set the regulator pressure to ~10 psig 2. Insert the needle into syringe adapter 3. Open valve on the syringe 4. Fill syringe to the desired volume 5. Close valve on the syringe, and remove it from adapter 6. Insert needle into a small vial of de-ionized water 7. Open syringe valve to equalize the pressure in the syringe. 8. Close the valve, and remove the syringe from the vial of water. 9. Insert needle of the syringe into a Tube Injector. (carrier gas ~50mL/min) 10. Slowly inject the contents of the syringe. (2-3 second per ml) 11. Remove the syringe & wait 2 to 5 minutes before removing the tube.

How to Equalize to Atmospheric Pressure Once the gas is at atmospheric pressure - it s easy to calculate the mass of the each compound contained in the syringe. 1. The water provides a visual - to when the pressure is equalized. 2. It prevents any additional gas from escaping until we close the syringe valve.

Using a Tube Injector (Commercially Available, or Homemade) Tube Connects Here Carrier Gas Homemade Device Glassware used in the device to the left Commercially Available

Tube Injector- Dry Carrier Gas

Examples of How Things Can Go Wrong Details about the gas mix used for these examples: 43-Compound Gas Mix at 1000ppbv (Many of them are suitable for TO-17 and Method 325B) Range: (Chloromethane 52 M.W. to Hexachlorobutadiene 261 M.W. Gas Mix is contained in small cylinder (12" high ) pressurized to 1800 psig. The cylinder contains 110 Liters of gas. (<1000 injections can be made if using 100mL syringe).

Pressure Not Released from the Syringe pa 60 50 FID1 A, (NEMC\TIME0008.D) More Mass Contained was contained in the same volume 40 30 20 10 2.5 5 7.5 10 12.5 15 17.5 20 22.5 FID1 A, (NEMC\TIME0001.D) pa 60 Proper Technique min 50 40 30 20 10 2.5 5 7.5 10 12.5 15 17.5 20 22.5 min

Not Enough Carrier Gas pa 60 50 FID1 A, (NEMC\TIME0002.D) 20 seconds = 17 ml Transfer Volume Not enough carrier gas to flush the compounds from the injector 40 30 20 10 pa 2.5 5 7.5 10 12.5 15 17.5 20 22.5 FID1 A, (NEMC\TIME0001.D) min 60 50 5 minutes = 250 ml Transfer Volume Prober Technique 40 30 20 10 2.5 5 7.5 10 12.5 15 17.5 20 22.5 min

Recent Calibrations Done in Our Lab

Tube Injector Parameters Carrier Gas Flow Rate Block Temp Gas-Tight Syringes* Range of the Curve Gas Mix and Concentration Nitrogen 50 ml/min 65 C 10 ml & 50 ml 0.5, 1, 2.5, 5, 10, 25ppbv @ 4-Liters 9-compound mix at 1000ppbv * 10 ml used for the 2,4, and 10mL injections * 50 ml used for the 20, 40 & 100mL injection (2qty 50mL injections)

Analytical Parameters Thermal Desorber: PerkinElmer TurboMatrix Tube Desorb: 330 C for 5 min. (Primary Tube: Carbotrap-300) Trap Desorb: 330 C for 5 min. (Trap: Carbopack-B & Carboxen-1000) Valve & Transfer Line: 175 C Desorb Flow: 30 ml/min Inlet Split: 5 ml/min Outlet Split: 5 ml/min Column: Equity-1 60 m x 0.25 mm ID x 1.0 µm film (28050-U) Flow Rate: 2.5mL/min 40 C for 5min,10 C/min hold for 0min 15 C/min/ to 225 C hold for 5min. GC/FID: H 2 40 ml/min, Air 400 ml/min, Makeup (N 2 ) 40 ml/min

6-point Calibration Curve using Dry Nitrogen as the Carrier Gas Linear Regression - Origin Ignored Average Response Factor Compounds Slope Intercept Corr Vinyl chloride 21.720 8.375 0.99918 1,3-Butadiene 50.390-3.293 0.99999 Acrylonitrile 17.078 0.043 0.99942 Methylene chloride 10.805-1.182 0.99999 Chloroform 8.949-0.787 0.99999 1,2-Dichloroethane 23.489-4.048 0.99997 Benzene 69.435-7.103 0.99998 Trichloroethene 24.880-4.323 0.99993 Tetrachloroethene 25.843-6.419 0.99985 Compounds Average Stdev %RSD Vinyl chloride 24.171 1.296 5% 1,3-Butadiene 49.287 0.894 2% Acrylonitrile 16.147 1.593 10% Methylene chloride 10.224 0.524 5% Chloroform 8.513 0.430 5% 1,2-Dichloroethane 21.510 1.809 8% Benzene 66.424 2.696 4% Trichloroethene 23.040 1.530 7% Tetrachloroethene 23.000 2.442 11% Range of the Curve 0.5, 1, 2.5, 5, 10, 25ppbv @ 4-Liters

Humidity

Humidity Can Cause Problems Problems During Sampling: Water vapor can be retained by the adsorbent Can mask the available sites of the adsorbent Can in (some cases) displace the compounds Problems During Analysis: Water vapor can alter split flow ratios during desorption Can create chromatographic separation issues Can blow out the FID flame Reduces the vacuum of an MS detector

Water Vapor Retained by Adsorbents Silica gel is hydrophilic and will retain ~40% of its own weight in water. Reason why silica gel is not typically used for thermal desorption applications. Some adsorbents are more hydrophobic than others. Porous Polymers & Graphitized Carbons are very hydrophobic and virtually no water will be retained on them while sampling in humid conditions. Carbon Molecular Sieves are still classified as hydrophobic. At high humidity some water is retained, but the bond to the water is weak. It can be driven off at ambient temperatures using a dry purge prior to analysis.

Should We Be Spiking Tubes Using Humidified Carrier Gas? It depends on how you interpret the specific method. Both EPA TO-17 and the proposed 325B describe using humidified gas when testing tube performance, but during calibration it s vague. With all of the problems humidity can cause, maybe we should spike our tubes using a humidified gas stream? If so - It's easy to do.

Device For Spiking Using Humidified Carrier Gas DI Water Two: Flow Controls

6-point Calibration Curve using Humidified Carrier Gas 50% RH (A dry purge volume of 0.5 Liters was needed) Linear Regression - Origin Ignored Average Response Factor Compounds Slope Intercept Corr Vinyl chloride 22.338 8.401 0.99940 1,3-Butadiene 46.004-6.094 0.99984 Acrylonitrile 16.437-1.847 0.99969 Methylene chloride 10.820-1.249 0.99992 Chloroform 9.121-0.655 0.99990 1,2-Dichloroethane 24.040-3.548 0.99989 Benzene 71.178-4.402 0.99989 Trichloroethene 25.632-3.309 0.99988 Tetrachloroethene 26.341-4.912 0.99988 Compounds Average Stdev %RSD Vinyl chloride 25.657 2.692 10% 1,3-Butadiene 44.428 0.847 2% Acrylonitrile 15.568 1.764 11% Methylene chloride 10.510 0.257 2% Chloroform 8.953 0.227 3% 1,2-Dichloroethane 22.856 0.841 4% Benzene 71.342 3.746 5% Trichloroethene 24.802 0.779 3% Tetrachloroethene 24.788 0.962 4% Range of the Curve 0.5, 1, 2.5, 5, 10, 25ppbv @ 4-Liters

Conclusions Compressed Gas Standards offer an alternative to using Liquid Standards. A carrier gas is required to distribute the compounds to the adsorbent(s) in the Thermal Desorption Tube. Equalizing the pressure of the syringe to atmospheric pressure is required for representative curves. Spiking tubes with a humidified gas stream can make the calibration more representative of actual sampling conditions.

References EPA TO-17. Determination of Volatile Organic Compounds in Ambient Air Using Active Sampling Onto Sorbent Tubes. Proposed EPA Method 325B. Volatile Organic Compounds From Fugitive and Area Sources, available from the Federal Register Vol. 79 No.125 June 30, 2014 Dinardi, Salvatore. Calculation Methods for Industrial Hygiene: Van Nostrand, 1995. Print

Thank You for Attending! jamie.brown@sial.com 814-359- 5917