User s Guide. CO2 Controller for miniature incubators

Similar documents
User s Guide. CO2 & O2 Controller for miniature incubators

User s Guide. Vacuum Controller for liquid delivery systems

User s Guide. Pressure Controller for liquid delivery systems

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

MODEL CALIBRATION GAS DELIVERY SYSTEM

User Manual MFCS TM -EZ

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

PH01 Perfusion Cannula Manual

RK LEL Sample Draw Aspirator Adapter Operator s Manual

THE UNIVENTOR 400 ANAESTHESIA UNIT

Overview. Front Panel: Keypad and Display

RK-IR Sample Draw Aspirator Adapter Operator s Manual

User Instruction Manual

Portable oxygen concentrator

INFUSION SYSTEM (110V) (220V) USER S MANUAL

GETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised )

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

The Univentor 1250 Anaesthesia Unit

CSA Sample Draw Aspirator Adapter Operator s Manual

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

Radnoti Liver Perfusion System

INSTALLATION AND OPERATING INSTRUCTIONS BIG TEX VENTED GAS LOGS

Issue OWNER S MANUAL. GAMMA RACQUET SPORTS GAMMA Sports All Rights Reserved

PERSONAL AIR BREATHING UNIT

InFlow Modular perfusion bioreactor

MODEL 100 NITROGEN INFLATION CART

SPECIFICATIONS APCEPH1

Techcon Systems TS1254 Pressure Pot

LEAK TROUBLESHOOTING

with O 2 Controller Instruction Manual

Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

MANUAL BE SERIES Test Benches

AUTO PUMP STATUS CENTER

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

TL-12 FLUSH TANK. Manual Ref. # MN-??? Polyurethane Machinery Corp. REVISION 1.0

STATEMENT OF LIABILITY

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

UsER manual for Watersens ph -REDOX

ROPV R40 E Series User Manual

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

DPC-30 DPC-100. Reference Manual

PRO SINK -R 1 and 2. Dilution system CONTENTS

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required)

Marschalk Model # 94000

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

ATD /8 x 50 Retractable Air Hose Reel Owner s Manual

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

D Series Air and Water Kit Part# 02550

LSA-BAWMBL. Wall-Mounted Boom Arm INSTRUCTIONS

HI 2314 HI 2315 HI 23151

AA GWR Water Retention Meter for Accurate Measurement of Water Retention

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

ACCU-MIX 1200 / / 26270

Model # HC-MIS INSTRUCTION MANUAL

Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

VAPOMATIC ANESTHETIC VAPORIZER

GALISO SOAP OPTION FOR PCT-ADW-D

Instruction Manual ibidi Gas Incubation System

J Air and Water Kit Instructions Part# 02584

PBT1100 POWER BANDING TOOL

Techcon Systems TS1258 Pressure Pot

G7S Hand Pump Owner s Manual

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Oxygen Delivery Systems:

EM151/EM201. Electronic Manometer - + INSTRUCTION MANUAL EM Washington Street Melrose, MA Phone Toll Free

QE-1, Rev Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Digital Vacuum Regulator

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Research Grade Blood Pressure Transducer User's Manual

EM151/EM201. Electronic Manometer - + INSTRUCTION MANUAL. EM201 PSI

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

SINORIX Valve Reconditioning

EC214 EC215 - EC215R Bench Conductivity Meters

UNIVERSAL MASK TEST SYSTEM

NOVALYNX CORPORATION MODEL WEATHER STATION INSTRUCTION MANUAL

User Manual November 2014

Natural Gas Conversion Kit

ELECTRONIC PLC SERIES Test Benches

Standard Operating and Maintenance Instructions for Pumping System Model PS-150

Oxygen Therapy Devices

Interface Devices, Inc. Hydraulic Mini Mule

Rejuvenation Instructions

EMT EnvironmentalMonitoringTECHNOLOGIES

DGT-CO2BX, Digital CO 2 Controller. Manual

Portable Compressed CO 2 Regulator

Work Smarter...Not Harder!

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

92831 TEL: (714) FAX:

TOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS HAZARDOUS AREA CLASSIFICATION

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Tentec. Instruction Document. Mini Air Driven Pump Unit Model: HTT.627X Series. Part Identifier

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

Paddle Bar Replacement

LRS(H)4 Pressure-Reducing Regulator User Manual

Transcription:

CO2 Control User s Guide CO2 Controller for miniature incubators Precise CO2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for any microscope Compatible with Imaging systems Ph: 877-853-9755 www.biosciencetools.com

Contents Specifications............................................... 3 Introduction................................................ 3 Installation Guide............................................ 4 Front Panel Controls.......................................... 7 Inputs, Outputs and Back Panel controls........................ 8 Using Heated Humidifier CO2-500ML............................ 9 Warranty................................................... 9

Specifications Range 0 to 20% CO2; ships calibrated to provide 5% level of CO2 inside miniature incubators TC-MIS, TC-MI, or any custom enclosure Input 150PSI max Sensors uses built-in CO2 sensor Feedback from CO2 sensor adjustable GAINs Output 750SCCM max Size (Controller) : 8Wx4Hx9D in. Power Supply 94 to 234 V AC, 50/60 Hz 35W Input Port 4mm O.D. tubing (10-32 threaded) includes adapters for different size tubing Output Port 1/8in O.D. tubing (10-32 threaded) Introduction The complete CO2 control system comes with a controller, tubing to connect to miniature incubators, and fitting to connect to a source of CO2 (a cylinder, for example, or a wall outlet). A source of CO2 is required to operate the system. The CO2 source needs to be regulated, since the maximum pressure of input CO2 should not exceed 150 PSI. The controller ships adjusted for input pressure 40 PSI. During operation, the controller is continuously monitoring CO2 inside the incubator.

Installation Guide 1 Using provided fitting and clamps, connect CO2 source (cylinder or wall outlet) to the controller. Some tubing and additional fitting might be required to connect to your CO2 source as designs vary. Usually some luer-lock fitting or other easy-connect adapters are used to splice different diameter tubing connecting your source to 4mm O.D. translucent tubing, which fits inside INPUT port on the back of the controller. After splicing provided 4mm tubing to CO2 source, simply push the tubing inside IN port all way, and slightly pull back to clamp. In order to disconnect the source, push YELLOW rim inside the connector, and pull the tubing out. Make sure the regulator on CO2 source does not show more than 150 PSI of output pressure. Pressures around 40 PSI should be sufficient to operate the system. The controller ships tuned to work with 40 PSI input CO2 pressure Similarly, insert a piece of 1/8in. O.D. BLACK tubing inside OUTPUT port on the back of the controller, and connect the other end of tubing to the incubator, or heated humidifier CO2-500ML. If a humidifier is used, connect the output of humidifier to the incubator. Make sure the provided check-valve is connected between the water bath or the incubator lid and the controller. Otherwise, you might have back flow of liquid inside the controller, which can result to the controller damage. Connect power cable. Plug the power cable into wall outlet. Connect grey cable to DB-9 connector on one end, and to the incubator lid on another end. 4

SENSOR I O INPUT PRESSURE OUTPUT 2 Turn the controller ON. The controller will self-calibrate itself for 5min, so make sure there are no residual CO2 under the lid of the incubator. Simply open the lid and let it ventilate for at least 30min. After 5min, The display will show INPUT CLOSED. Rotate the front knob to open INPUT - the controller is ready to use out of the box. Although the controller ships adjusted for use with your incubator, it can be re-adjusted using the procedure below. IMPORTANT: since the controller calibrates in clean air, the incubator should be open for at least 30min before turning the controller ON. 1. By pressing the front button, go through displays to adjust CO2 % level (5% for example) and FLOW rate (in SCCM, 200 SCCM should be enough for small incubators). The lid can be placed on the incubator at this point. 2. Open the source of CO2 gas. Note: the input pressure should not exceed 150 PSI (usually 40 PSI is enough). Rotate the front button to open the input port - the display will show: output % 0% 2. Press the front button until the display will show: input PSI 1.0 Rotate the pressure regulator on the back clock-wise to adjust this reading to around 3 PSI (higher pressure might be required for larger incubators). 3. Go to DC% screen by pressing the front knob. Although the controller adjusts DC level by itself, you can decrease the level, if there is indication that too much gas flowing into the incubator. 4. AC level is used to increase sensitivity of internal sensors while regulating FLOW rate. Very low levels should be used for more stable operation. 5

Let the controller operate for at least 30min. If the level of CO2 does not reach 5% in several minutes, increase input pressure or/and DC/AC. Repeat if necessary. 3 The input pressure can be reduced if you have indications that the flow of CO2 during operation is too high. IMPORTANT: small incubators require only very little flow of CO2 gas. 4aIf you notice unstable operation, or the indications of wrong CO2 level inside the incubator, perform fine tuning procedure again. 4bNOTE: self-calibration assumes that the controller was not functional and the incubator was open for at least 30min before turning it ON and that the surrounding environment is clean air. If you have problems with your experiments and suspect, that the controller does not provide correct CO2 level, turn ON the controller after the incubator being without the lid for 30min in well ventilated room without any CO2 source attached. USING THE FRONT DISPLAY CONTROLS Rotate the knob clock-wise to open the INPUT port, and rotate other direction to close - the display will show INPUT CLOSED Press the knob to get to CO2 settings and rotate to adjust: SET CO2% 5.0 Press again to go to SET FLOW level setting and rotate to adjust: SET FLOW 200 Press again to adjust input pressure (in PSI) input PSI 1.0 Rotate the pressure regulator on the back of the controller to increase pressure if required. Note: the pressure will not decrease until the controller starts supplying the gas to the output to relieve the internal pressure, so the lid may need to be open to decrease the input pressure. 6

Front Panel Controls 0.15 0.838 x 1.336 CO2% 0.41 SET 5.1 5.0 0.40x 0.32 0.15x0.35 output % 48 0.15.753 x 0.513 0.15x0.35 Front Panel Controls Knob Sets CO2 level, FLOW rate, input PSI pressure, DC level and AC level, % 7

Inputs, Outputs and Back Panel controls SENSOR I O INPUT PRESSURE OUTPUT Inputs & Outputs INPUT port OUTPUT port DB-9 port PRESSURE regulator Connects to a source of CO2. Maximum input pressure is 150 PSI. Connects to the incubator to supply CO2. Connects to the incubator to monitor CO2 level. Adjust input pressure inside the controller. Back Panel Controls Input Pressure regulator Turn CLOCK-wise to increase inside pressure and turn ANTI-clockwise to reduce available pressure.. 8

Using Heated Humidifier CO2-500ML A heated humidifier can be used to pre-heat and saturate the gas mixture with water, before the mixture enters the incubator. The humidifier consists of a heated base and a reservoir, which needs to be filled with distilled water. Fill the reservoir just enough to observe bubbles of gas coming out of input tubing, which has a female luer connector. The input tubing should be connected to BLACK output tubing coming out from a CO2 controller., with the checkvalve placed between After connecting tubing, place the reservoir on the base. Plug provided DC power adapter into the base and a wall power outlet. Turn the humidifier ON - an LED indicator will be ON. Let the base to warm up to facilitate water evaporation. After gas mixture enters the reservoir, it will be heated and mixed with water vapors. NOTE: You can use the reservoir as an indicator of gas mixture flow rate. Usually, enough gas flow is provided to the incubator, as long as you can observe stream of bubbles coming up from the inflow tubing. Using provided tubing, or any other tubing, connect the outflow MALE luer port to the incubator. Turn the CO2 controller ON to provide gas flow inside the incubator. Warranty This product is warranted to be free from defects in material and workmanship for the duration of one year. Normal wear, or damage resulting from abuse, accident, alteration, misuse, service by an unauthorized party or shipping damage, are excluded from this warranty and are not covered. Bioscience Tools will repair or replace the defective product covered by this warranty free of charge if it is returned, postage prepaid, to Bioscience Tools, ph: 1-877- 853-9755. 9

10