Gas Leak Measurement of TRT Barrel Module

Similar documents
This GC has two columns (I think new columns can be ordered from Restek)

Modulating Valves for Atmospheric, Infrared, and Direct Fired Burners

Introductory Lab: Vacuum Methods

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

ELEVATION CORRECTION FACTORS FOR RADON MONITORS*

Long term stability tests of INO RPC prototypes

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

LESER Deutschland Standard Functional Tightness Test (Cryo) Content

Chromat Calibration Updated October 27th, 2017

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

Ultimate Performance in Gas Chromatography

Multiple Gas#5 GC configuration Jan 2016

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Model Description

New improved robot for gas kinetics

6890 GC Site Preparation

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

This test shall be carried out on all vehicles equipped with open type traction batteries.

Manual for continuous distillation

Gas Diluter Portable Gas Dilution System User s Manual

EN1279 part 3 assessment of insulating glass: IGU with Duralite sealant

Hazards associated with the gas system and how to mitigate them

The University of Hong Kong Department of Physics Experimental Physics Laboratory

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

The Helium Leak Detector

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

A Reliable and Tracer Gas Independent Leak Detector for Food Packages

Marc McMullen. Physics Detector Support Group

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Straw CO 2 Leak Testing

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

Retention Time Locking: Concepts and Applications. Application

Self-operated Pressure Regulators for special applications

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

Quantitative Analysis of Hydrocarbons by Gas Chromatography

CHEM 3351 Physical Chemistry I, Fall 2017

Module 6. Tightness Testing

Digital Mass Flow Meters and Controllers for Gases

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Mass Flow Meter (MFM) for Gases

Your local gas generation partner. Precision series Modular gas generation solution for GC.

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

INDEPENDENT TESTING OF SPARKLIKE LASER NON-DESTRUCTIVE INSULATING GLASS GAS FILL ANALYZER

The Discussion of this exercise covers the following points:

Flow in a shock tube

The Transition Radiation Detector of the AMS-02 Experiment

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Cover Page for Lab Report Group Portion. Pump Performance

Leak Detection. Kirsten Zapfe, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, Special thanks to S. Holm, N. Mildner and A.

LEAKHUNTER Plus Model 8066

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

M/MR Modulating Valves TABLE OF CONTENTS DESCRIPTION SPECIFICATIONS. For Atmospheric, Infrared, and Direct Fired Burners

Comparative Study of Oxygen Permeation Through Polymers and Gas Barrier Films

SAMPLE RH = P 1. where. P 1 = the partial pressure of the water vapor at the dew point temperature of the mixture of dry air and water vapor

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

During today's webinar we're going to look at the following topics:

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Under pressure pushing down

Zebron Gas Management

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Fluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model

PENBERTHY SERIES LM, ELL, FL, GL, GH, U, L AND 2NC FOR PUMPING GASES

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs

INSTRUCTOR RESOURCES

CORESTA RECOMMENDED METHOD Nº 67

3 Flow and Pressure Control

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Svensk läkemedelsstandard

Chapter 6 Leak detection

Generating Calibration Gas Standards

Memorandum Background: Results and Discussion:

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Saturday Sept 16, 2006

ANALYSIS OF GASES PRODUCED BY THREE UNDERWATER CUTTING DEVICES

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Bubble Tube Installations

Transcription:

DUKHEP01-07-02 Gas Leak Measurement of TRT Barrel Module Shannon Watson, Seog Oh Department of Physics Duke University Durham, NC, 27708, USA 1. Introduction One of the requirements for the TRT module acceptance is the gas tightness. The present leak specification for the barrel module is 0.1 mbar/bar/min. The leak rate is calculated by pressurizing the module with argon and measuring the pressure drop as a function of time. In order to reach the required sensitivity, the measurement takes several hours. Although the temperature, pressure and humidity in the room vary enough to distort the measurement during this period, the leak rate can be calculated by measuring the absolute room pressure and the temperature and making the necessary corrections. However, it is not clear if the measurement using this technique reflects the true leak rate from the module when the module is filled with the ionization gas (Xe/CF 4 /CO 2-70/20/10 mixture). In order to understand the correlation, we have constructed a gas tight box with space to put a module inside. The box is connected to a GC (gas chromatograph) to determine the gas composition in the box. The basic principle is to fill the box and the purging volume with CO 2 and to fill the straws with Xe. The purging volume is the volume inside the shell other than straws. The pressure inside the straws is kept higher than the box pressure. Periodically, the gas inside the box is sampled in order to measure Xe content inside the box. By knowing the volume of the box and the module, and the Xe concentration in the box (from the GC), the Xe leak rate can be calculated. Another interesting measurement is the CO 2 permeation rate. It is well known that CO 2 permeates through kapton much better than other gases. Since straws are surrounded with CO 2, the permeation process could affect the ionization gas composition and change the operating characteristics of the detector. This rate can be easily measured using our setup. In this paper, we report on two measurements. One is the leak rate measured using two techniques (pressure drop method, and using the box with GC), and the other is the CO 2 permeation rate through straws.

2. Setup For this study, we have chosen module 2.06 because this module exhibits some leak. The measurement using the pressure drop shows that the leak rate for the module is 0.3 mbar/bar/min. The operating principle for the pressure drop measurement is fairly simple and won t be described here. For the setup using the box with GC, a picture of the setup is shown below. The dimension of the box (made of aluminum and sitting on the long table in the picture above) are 30 cm x 30 cm x 170 cm. The box was designed to accommodate the largest module (type III). For type II, about 1/3 of the space inside is filled with ballast material to reduce the overall volume. The GC (red color) is sitting on the top of the rack in the picture. The following is a schematic of the operation. in in the figure indicates the input gas direction.

3. Data taking and analysis a) CO 2 permeation rate One of the interesting quantities is the CO 2 permeation through straws. We have previously measured the permeation rate using the first prototype module (type I) and the result was 5x10 5 cc/cm of straw/min, or 0.0075 cc/min/barrel straw. With the new straws, it is expected that the permeation rate is not to be as large as the old straws used in the prototype. This permeation rate is measured by filling the box and the purging volume with CO 2 and the straws with Xe. The purging volume inside the shell is flushed with CO 2 by a small aquarium pump operated between the purging volume and the box. The pressure inside the straws with respect to the box is measured as a function of time using a differential pressure gauge. For this measurement, the pressure inside straws is maintained slightly above the box pressure by ~1 mbar. Figure below shows the measurement. The x-axis is the time in second and the y-axis is the pressure inside straws in mbar. (We assume that the argon permeation and the leak rate ( p ~1 mbar) are small compared to the CO 2 permeation rate.) The oscillation shown in the data is due to the characteristic of the differential pressure gauge. From the figure, the permeation rate of about ~0.04 cc/min

for the type II module or 0.05x10-5 cc/min/(cm of straw) is obtained, which is about a factor of ~100 less than the result from the old straws. The rate corresponds to about extra ~5 cc/min of CO 2 in the ionization gas for the entire barrel TRT. Pressure (mbar) Series1 Linear (Series1) 1.5 1.4 1.3 1.2 1.1 1 0.9 0.8 0.7 0.6 0.5 P.vs. time 0 1000 2000 3000 4000 5000 Time (second) b) Ar/Xe leak rate The next was to measure the Xe leak rate. Because of the cost, the argon leak rate is measured first to debug the system. As in the permeation measurement, the straws are filled with argon and the box and the purging volume are filled with CO 2 (see the schematic shown above). The straws are over pressurized (20 mbar) compared to the box. This is accomplished with a computer-controlled solenoid valve attached to the gas outlet. The pressure difference between the straws and the box is measured (using the differential pressure gauge) and the solenoid valve is opened or closed depending on the pressure difference. We note that 5-10 cc/min of argon (or Xe later) is flowed continuously through the module for this test.

Figure below (Figure GC1, the figure number is on the top of the figure) shows the GC analysis of the gas from the box at the beginning. The first peak is oxygen (t~1.2min) and the second peak is nitrogen (t~1.5 min). The peaks at the beginning are due to the gas injection. The vertical axis is what we call retention time. This measures the time in minutes that a particular gas takes to pass through a GC column. The integrated area below a peak is proportional to the gas concentration. The conversion of the area to the concentration requires a calibration and it is done by injecting a known amount of the gas into the column (typically 1cc). In this GC column (molecular sieve column), unfortunately, the argon peak overlaps with the oxygen peak.

The next figure (Figure GC2) shows the same but after one day of running. As shown in the figure, the nitrogen peak has increased very slightly while the argon/oxygen peak increased by a factor of ~2. The increase of the nitrogen peak tells us that the box is not completely gas tight. Although the argon and oxygen peak overlap, the leak rate can be calculated from the peaks in two plots since the ratio of the oxygen to nitrogen is known (we assume that air leakage to the box preserves the ratio). The argon leak rate from the module is calculated to be 0.04 cc/min, which correspond to 0.2 mb/bar/min. We again note that the pressure difference of 20 mbar was maintained between the straws and the box. This is somewhat lower than 0.3 mb/bar/min measured using the pressure drop method.

The last step was to measure the Xe leak rate. The setup is identical to the argon leak rate measurement except argon was replaced with Xe. Figure GC3 shows the first measurement (time=0). Similar to the earlier plots, there are two small peaks corresponding to oxygen/argon and nitrogen respectively. Thus is after flushing the box with CO 2 to purge air and argon left from the previous argon leak measurement. Calibration with Xe gas shows that the Xe peak should appear at a retention time of 8 minutes, thus Xe peak identification is not a problem Figure GC4 below show the GC measurement after two days. As shown in the figures, there is evident Xe peak. From the figures, the leak rate of 0.04 cc/min is obtained and it corresponds to 0.2 mb/bar/min leak rate. This is consistent with the argon leak rate.

The table below summarizes the leak test results. Module Leak (P method) Leak (with GC) Leak (with GC) Type Argon Xe 2.06 ~0.3 mb/bar/min ~0.2 mb/bar/min ~0.2 mb/bar/min The Xe leak rate is measured for two other pressure differences between the straws and the box in order to estimated the Xe permeation rate. The figure below shows the leak

rate as a function of the pressure difference. Extrapolating the curve to p =0, we obtain less than 0.01 cc/min for the Xe permeation rate for a type II module. Leak.vs. P Series1 Leak Rate (cc/min) 0.05 0.04 0.03 0.02 0.01 0 0 5 10 15 20 25 P (mbar) 4. Summary We have measured the CO 2 and Xe permeation rate and Ar and Xe leak rate using production module 2.06. The CO 2 permeation rate is 0.04 cc/min for the type II module (or ~10-4 cc/min for a barrel straw or ~5 x 10-7 cc/(cm of straw)/min). This corresponds to an extra ~5 cc/min of CO 2 in the ionization gas for the barrel TRT. We also measured the Xe permeation rate and it is less then 0.01cc/min for the type II module. The Ar/Xe leak rate was measured using two methods. One measures the pressure drop due to the leak in the straws and the other measures the leaked out Ar/Xe in the box using GC. The results from both methods are reasonably consistent with each other. Using our measurement, the leak rate specification 0.1 mb/bar/min corresponds to the leak rate of ~0.015 cc/min for the type II module, when the pressure difference is 20 mbar.