Quick Start Guide LIQ-QSG-499ACL-03, Rev I July Rosemount 499ACL-03. Monochloramine Sensor

Similar documents
Quick Start Guide LIQ-QSG-499AOZ, Rev M July Rosemount 499AOZ. Dissolved Ozone Sensor

Free Chlorine Sensor For additional information, please visit our website at

Dissolved Oxygen Sensor

Rosemount 140/141/142

Rosemount 400 and 400VP

Specify the Right Remote Diaphragm Seal Solution for Vacuum Applications

61 Series Pilots for Pilot-Operated Pressure Reducing Regulators

Fisher Lever Lock Handlever Actuators for POSI SEAL A81 Rotary Valves

1805 Series Relief Valves

1305 Series Pressure Reducing Regulators

64 Series Pressure Reducing Regulators

1805 Series Relief Valves

Type 1367 High-Pressure Instrument Supply System with Overpressure Protection

299H Series. Introduction. P.E.D. Categories. Specifications. Installation. Warning. Installation Guide English September 2012

Oxygen Measurement Systems High Reliability and Wide Application Coverage

VALVCHEQ BACKFLOW PREVENTERS FIGURE RP03

Fisher Lever-Lock Handlever Actuators for POSI-SEAL A81 Rotary Valves

VALVCHEQ BACKFLOW PREVENTERS FIGURE RP03

Y692VB Series Vacuum Breaker

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

T208VR Series Tank Blanketing Vacuum Regulator

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE

Model PDT Dewpoint Transmitter

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

Types ACE95 and ACE95Sr Tank Blanketing Valves

Fisher DVI Desuperheater Venturi Inline

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER TRANSPORTABLE CRUDE OIL CONTAINER

Types 1808 and 1808A Pilot-Operated Relief Valves or Backpressure Regulators

1301 Series High-Pressure Regulators

Type ACE95jr Tank Blanketing Valve

Model B20/63. Combined Chlorine Recorder. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

Non-Membrane Disinfection Sensors Instruction Manual

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

Type EZH Relief or Backpressure Regulator

Types ACE95 and ACE95Sr Tank Blanketing Valves

Types ACE95 and ACE95Sr Tank Blanketing Valves

Model B20/63. Combined Chlorine Recorder. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

Pressure Dump Valve Service Kit for Series 2300 Units

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

10 DRIP NOZZLE WATERING DRIP KIT Model / Model 46095

EL-O-Matic E and P Series Pneumatic Actuator SIL Safety Manual

QUICK START GUIDE TO DRINKING WATER MONITORING

ERNG-X TESCOM Product Manual

ThermoVault Max. Extreme Temperature Thermal Barrier. Product User Guide

ATS430 turbidity sensor Retractable insertion assembly

Constant Pressure Crude Oil Container Model CPCCP

DTG - LCD Digital Temperature Gauge USER MANUAL

GILMONT ACCUCAL FLOWMETERS

LRS(H)4 Pressure-Reducing Regulator User Manual

Type R622 Pressure Reducing Regulator

AP100 series ph/redox (ORP) cartridge sensors with water-wash option

Rosemount Integral Orifice Flow Meter Series. Reference Manual , Rev KA December 2017

Petro-Chemical & Analyzer Sampling Systems

600 o F High Temperature Aging Cell Instruction Manual

TECHNICAL DATA CAUTION

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Type 310A-32A Pressure Reducing Regulator and Type 310A-32A-32A Working Monitor Regulator

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

TECHNICAL DATA. Page 1 of 12

Model B20/62. Free Chlorine Recorder. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

TECHNICAL DATA. Pressure Regulation 531a. April 24, 2009

Air Consumption Recommended Usage Siphon Feed

Instruction Manual Updated 7/26/2011 Ver. 2.2

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER

Stainless Steel 142 mm Filter Holder YY

9408 Series Dissolved Oxygen Systems. Dissolved oxygen systems designed for reliable operation with minimal operating costs

PENBERTHY SUBMERSIBLE AUTOMATIC SUMP DRAINER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

2 Sentry MVS Installation, Operation & Maintenance

Model Series 62 Constant Differential Relay

Aquor House Hydrant V2

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

RDO Optical Dissolved Oxygen Sensor

Instruction Manual. PE-25/32/37 Tapping Machines. POLYTAPP Valves

ph measuring system for hygienic process applications

Model B Wire Wet Oxidant Gas Transmitter. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

Rosemount 1195 Integral Orifice Primary Element

V DFX Series. Direct-Operated Tank Blanketing Regulator Manual D103808X012

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

9408 series Dissolved oxygen systems

Walchem Disinfection Sensors WCDB, WFCB, WOZB and WPAB Instruction Manual

STERILIZABLE DISSOLVED OXYGEN SENSORS

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

630 Series Regulators and Relief Valves

RC 195 Receiver-Controller

Installation Manual , Rev BB December Micro Motion H-Series Hygienic Coriolis Flow and Density Sensors

EcoPRO-Series 4-20mA SUBMERSIBLE PRESSURE TRANSMITTER INSTRUCTIONS

Autocalibration Systems for In Situ Oxygen Analyzers

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Differential Pressure Transmiter

MEGR-1627 Instruction Manual

Fisher DSA Steam-Atomized Desuperheater

MODEL MN-PWV PRESSURIZED WATER VESSEL ASSEMBLY OPERATING MANUAL

TracPhone V7-IP Gyro Replacement Instructions

Types S100K and S102K Pressure Regulators

Model 1159 Remote Diaphragm Seals Manual

Installation, Operation, and Maintenance Manual

English. Introduction. Safety Instructions. All Products. Inspection and Maintenance Schedules. Parts Ordering. Specifications WARNING WARNING

Transcription:

Quick Start Guide LIQ-QSG-499ACL-03, Rev I July 2017 Rosemount 499ACL-03 Monochloramine Sensor

Safety information CAUTION! SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor material may not be compatible with process composition and operating conditions. Application compatibility is entirely your responsibility. CAUTION! EQUIPMENT DAMAGE Do not exceed pressure and temperature specifications. Pressure: 65 psig (549 kpa abs) max. Temperature: 32 to 122 F (0 to 50 C)

Contents Contents Chapter 1 Plan...1 1.1 Unpacking and inspection... 1 1.2 Product description...1 1.3 Specifications...1 Chapter 2 Install...3 Chapter 3 Wire...5 Chapter 4 Calibrate... 11 4.1 Zero point calibration...11 4.2 Full scale calibration... 11 Chapter 5 Maintenance... 13 5.1 Cleaning the membrane...13 5.2 Replacing the electrolyte solution and membrane... 13 Chapter 6 Accessories... 15 Quick Start Guide i

Contents ii Rosemount 499ACL-03

Plan 1 Plan 1.1 Unpacking and inspection 1. Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. 2. Save the box. 3. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount immediately. 1.2 Product description Figure 1-1: Rosemount 499ACL-03 Sensor Parts A. Membrane retainer B. Membrane assembly C. O-ring D. Cathode E. Electrolyte fill plug (wrap with pipe tape) F. Pressure equalizing port G. Sensor cable (integral cable shown) 1.3 Specifications Table 1-1: Sensor specifications Physical characteristics Pressure Temperature (operating) Process connection Specifications Sensor must drain to open atmosphere. No back pressure. 0 to 50 C (32 to 122 F) 1 in. MNPT Quick Start Guide 1

Plan Table 1-1: Sensor specifications (continued) Physical characteristics Specifications Wetted parts Noryl (1), Viton (2), silicone, wood, and Zitex (PTFE) (3) Cathode Gold mesh Process connections Sensor must be used in flow cell PN 24091-01 (1) Noryl is a registered trademark of General Electric. (2) Viton is a registered trademark of DuPont Performance Elastomers. (3) Ziton is a registered trademark of Performance Plastics Corp. Table 1-2: Other specifications Type PN Wetted materials Process connection Maximum temperature Maximum pressure Low flow cell with bubble sweeping nozzle (1) 24091-01 Polycarbonate/polyester, 316 stainless steel, and silicone Compression fitting for 1/4 in. O.D. tubing 158 F (70 C) 90 psig (722 kpa abs) (1) Flow cell must drain to open atmosphere. Do not install the sensor in a pressurized line. Temperature and pressure specifications for the low flow cell exceed the temperature and pressure specifications for the sensor. 2 Rosemount 499ACL-03

Install 2 Install Install sensor in the low flow cell (PN 24091-01) only. Keep the flow as constant as possible between 1 and 4 gph (3.8 to 15 L/hr). The flow cell must drain to open atmosphere. Figure 2-1: Sensor Orientation Quick Start Guide 3

Install Figure 2-2: Low Flow Cell (PN 24091-00) 4 Rosemount 499ACL-03

Wire 3 Wire NOTICE For additional wiring information on this product, including sensor combinations not shown here, please refer to the Liquid Transmitter Wiring Diagrams. Figure 3-1: Rosemount 499ACL-03-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Quick Start Guide 5

Wire Figure 3-2: Rosemount 499ACL-03-54-60/499ACL-03-54-VP Sensor Wiring to Rosemount 1056 and 56 Transmitters 6 Rosemount 499ACL-03

Wire Figure 3-3: Rosemount 499ACL-03-54 Sensor Wiring to Rosemount 1066 Transmitter Quick Start Guide 7

Wire Figure 3-4: Rosemount 499ACL-03-54-60/499ACL-03-54-VP Sensor Wiring to Rosemount 1066 Transmitter Figure 3-5: Rosemount 499ACL-03-54-VP Sensor Pin-out Diagram (Top View of Connector End of Sensor) 8 Rosemount 499ACL-03

Wire When making a connection through a junction box (PN 23550-00), wire point-to-point. NOTICE Use a wire nut and pigtail (included) when connecting several wires to the same terminal. Quick Start Guide 9

Wire 10 Rosemount 499ACL-03

Calibrate 4 Calibrate 4.1 Zero point calibration Even in the absence of monochloramine, the Rosemount 499ACL-03 sensor generates a small signal called the zero current. Failing to correct for the zero current can introduce a bias, particularly if the monochloramine concentration is small (<0.4 ppm). Zero the sensor when it is first placed in service and every time the fill solution is changed. To zero the sensor: Procedure 1. Pour a cup of deionized or bottled water. 2. Place the sensor in the water. 3. Wait until the sensor current has reached a stable low value (at least two hours). 4. Follow the transmitter prompts for zeroing the sensor. Note Refer to the manual for the transmitter you are using (56, 1056, or 1066). The zero current should be between -10 and +15 na. For more information, refer to the Rosemount TCL Manual. 4.2 Full scale calibration Because stable dilute monochloramine standards are not available, the sensor must be calibrated against the results of a laboratory test run on a grab sample of the process liquid. 1. Place the sensor in the flow cell. 2. Start the sample and reagent flow. 3. Adjust the sample flow to between 1 and 4 gph. 4. Adjust the concentration so that it is near the upper end of the operating range. 5. Wait for the readings to stabilize. 6. Follow the transmitter prompts to complete the calibration. Note Refer to the manual for the transmitter you are using (56, 1056, or 1066). Be sure taking the sample does not alter flow to the sensor. 7. After calibration, go to the Diagnostics menu and check the sensitivity. Quick Start Guide 11

Calibrate The sensitivity should be between 250 and 450 na/ppm. For more information, refer to the transmitter manual. 12 Rosemount 499ACL-03

Maintenance 5 Maintenance WARNING! PRESSURIZED SPRAY INJURY Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level! 5.1 Cleaning the membrane Keep the membrane clean and free from dirt and algae. Periodically inspect the membrane. If it appears fouled, clean the membrane with water sprayed from a wash bottle. CAUTION! EQUIPMENT DAMAGE Do not wipe the membrane with a tissue. Do not touch the membrane. Doing so may damage the cathode, making the sensor unusable. 5.2 Replacing the electrolyte solution and membrane WARNING! CORROSIVE SUBSTANCE Fill solution is corrosive. Avoid contact with skin and eyes. Consult MSDS for safety information. Procedure 1. Unscrew the membrane retainer. 2. Remove the membrane assembly and O-ring. See Figure 1-1. 3. Hold the sensor over a container with the cathode pointing down. 4. Remove the fill plug. 5. Allow the electrolyte solution to drain out. 6. Remove the old pipe tape from the plug. 7. Wrap the plug with one or two turns of pipe tape. 8. Prepare a new membrane. Quick Start Guide 13

Maintenance a. Hold the membrane assembly with the cup formed by the membrane and membrane holder pointing up. b. Fill the cup with electrolyte solution. 9. Hold the sensor at about a 45 angle with the cathode end pointing up. 10. Add electrolyte solution through the fill hole until the liquid overflows. 11. Tap the sensor near the threads to release trapped air bubbles. 12. Add more electrolyte solution if necessary. 13. Place the fill plug in the electrolyte port and begin screwing it in. 14. After several threads have engaged, rotate the sensor so that the cathode is pointing up and continue tightening the fill plug. Do not overtighten. 15. Place a new O-ring in the groove around the cathode post. 16. Cover the cathode with electrolyte solution; then place the membrane assembly over the cathode. 17. Screw the membrane retainer in place. 18. Hold the sensor with the cathode end pointing down. 19. Give the sensor several sharp shakes to dislodge air bubbles trapped behind the cathode. The sensor may require several hours operating at the polarizing voltage to equilibrate after the electolyte solution has been replaced. 14 Rosemount 499ACL-03

Accessories 6 Accessories Part # Description 23750-00 Electrolyte fill plug with wooden osmotic pressure relief port 9550094 O-ring, Viton 2-014 33521-00 Membrane retainer 23501-09 Monochloramine membrane assembly: includes one membrane assembly and one O-ring 23502-09 Monochloramine membrane kit: includes three membrane assemblies and three O-rings 9210372 Monochloramine sensor fill solution, 4 oz (120 ml) Quick Start Guide 15

Accessories 16 Rosemount 499ACL-03

Accessories Quick Start Guide 17

Accessories 18 Rosemount 499ACL-03

Accessories Quick Start Guide 19

LIQ-QSG-499ACL-03 Rev I 2017 www.emerson.com/rosemountliquidanalysis Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317 Toll Free +1 800 999 9307 F +1 952 949 7001 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis EUROPE Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland T + 41 (0) 41 768 6111 F + 41 (0) 41 768 6300 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis MIDDLE EAST AND AFRICA Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033 T +971 4 811 8100 F +971 4 886 5465 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis ASIA-PACIFIC Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore T +65 777 8211 F +65 777 0947 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis 2017 Emerson Automation Solutions. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and, while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, expressed or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.