PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS

Similar documents
STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

QUICK START GUIDE TO DRINKING WATER MONITORING

Oxygen Measurement Systems High Reliability and Wide Application Coverage

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

Columbus Instruments

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

DISSOLVED OXYGEN SENSOR BT34i

Dissolved Oxygen Sensor

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

CSA Sample Draw Aspirator Adapter Operator s Manual

Vapor Sorption Analyzer

Aerobic Respiration. Evaluation copy

Making Measurements Leaks

Technical Data Sheet MF010-O-LC

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

& Dissolved Oxygen Sensor. Instruction Manual

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

TQC Bresle Chloride test 2008

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

4 CONTROL-CHART EXAMPLES

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Application Note AN-107

Instrument Operating Manual

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

OxyProbe II Instruction Manual

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

Installation, Operation, and Maintenance Manual

Applications Note: Use of "pentane equivalent" calibration gas mixtures

RK-IR Sample Draw Aspirator Adapter Operator s Manual

STERILIZABLE DISSOLVED OXYGEN SENSORS

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. C

Automatic Permeability Testing: The Challenges and Solutions Author: Alyce Hartvigsen, PBI-Dansensor A/S, Ringsted, Denmark

White Paper. Overview of MAP Headspace Gas Measurement: Sample Volume Considerations:

Navigator 600 Silica analyzers

Instruction Sheet. Calibration of Multiple 1720 and FT660 Turbidimeters

12mm Dissolved Oxygen Sensor Care and Use Instructions

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Kube Engineering Quinn E. Eddins, B.S.E.E Phone

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

HOW DOES A CHECK VALVE WORK? A GUIDE TO THE 8 STAGES OF CHECK VALVE OPERATION PLASTIC ONE-WAY IN-LINE SPRING-LOADED

LogIT Dissolved Oxygen Probe

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

How To Use Getters and Getter Pumps

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

Oxygen & Binary Gas Measurement and Analysis

APPLICATION NOTE. GC Integrated Permeation Device

Operator Quick Guide ORBISPHERE 3654

Compatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Oxygen Sensor with Electrode

Relative Humidity Calibration Kit

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

Verification Of Calibration for Direct-Reading Portable Gas Monitors

Four common PRESSURE CALIBRATION PAIN POINTS

Introduction to Ethylene Oxide (EO/EtO) Sterilization

Fisher DVI Desuperheater Venturi Inline

2B Technologies, Inc.

Degassing ERC KK Saitama, Japan

Experiment 18 Properties of Gases

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. K

Quick Start Guide LIQ-QSG-499AOZ, Rev M July Rosemount 499AOZ. Dissolved Ozone Sensor

Vinmetrica Dissolved Oxygen System User Manual

NT High-temperature flowmeter model Simultaneous flow and pressure outputs for high-temperature applications

ACV-10 Automatic Control Valve

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

LIQUID METER PROVING TECHNIQUES CT 4095

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

O3 3E 1 F Gas Sensor Module

Best Practice for Calibrating LTH Conductivity Instruments

Vortex flowmeters. Product family introduction Principle of operation Product review Applications Key product features

Operation & Maintenance Manual Place this manual with valve or person responsible for maintenance of the valve

Installation, Operation, and Maintenance Manual

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

Free Chlorine Sensor For additional information, please visit our website at

AMT-Ex Dewpoint Transmitter

Pipettor. User Manual

1020 Industrial Drive, Orlinda, TN fax

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

Calcimeter Instruction Manual

Quick Installation Guide Doc nr.: C Date:

PRODUCT SHEET. Order probe only as RXPROBE02

User guide DIAPHRAGM SEALS. Industrial instrumentation for Pressure and Temperature

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

OPERATING MANUAL CHEMTRAC 840M MANUAL RETRACTABLE PROCESS HOLDER

Protective Atmospheres, Measurement Technologies and Troubleshooting Tools

Installation and Use Manual

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

HI 2314 HI 2315 HI 23151

CORESTA RECOMMENDED METHOD N 6

Dissolved Oxygen Cartridge Electrode Systems

Mass Spec will not Autotune

NEULOG OXYGEN LOGGER SENSOR GUIDE

FX134, FX1234, FX3030 Problem and Solutions

GILMONT ACCUCAL FLOWMETERS

PIPETMAN F ENGLISH. Other languages available on our website:

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION

Transcription:

APPLICATION NOTE PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS Improved sensor performance yields longer intervals between service, consistent on-line monitoring, and increased accuracy in readings Application description: Techniques for sample verification and optimization When measuring trace levels of oxygen in liquid samples, it is important that your oxygen analyzer system is well maintained and operates properly. Erratic, inconsistent, or out-of-spec O 2 concentration readings could indicate sampling problems or problems with the analyzer. Thus, it is essential to confirm that your analyzer is operating correctly before you look into problems with the sample or the sample processing. This is accomplished by the following steps: 1. Calibrate your sensor in air 2. Verify your sensor s zero reading 3. Check your sample lines for leaks Having completed all three steps as described in the following paragraphs, you will quickly ensure the highest quality oxygen measurements possible. 1. Calibrate sensor to the instrument Calibration of the oxygen sensor is the first step to verify that your oxygen analyzer system is giving a linear response to the oxygen concentration in your sample. To calibrate your oxygen sensor, first allow the sensor and membrane to equilibrate to the temperature of the room, so the measurement is being properly compensated by the thermistor in the sensor. Next, follow the calibration instructions in the operator s manual for your oxygen sensor and instrument. If the sensor does not calibrate, it s time for a sensor service, as described in the manual. 2. Verify sensor residual Checking the residual signal of your sensor is the second step in verifying proper oxygen readings. Orbisphere oxygen sensors are unique in that they should generate a negligible current when in complete anaerobic conditions (that is, no oxygen). Nevertheless, a damaged or improperly maintained sensor may generate a positive or negative current (and subsequently a positive or negative oxygen display) in anaerobic conditions. It is good practice, therefore, to occasionally check the "zero" point of a sensor. You can do this by placing the sensor in water that has been deaerated by supersaturating with sodium sulfite crystals. Or you may choose to expose the sensor to nitrogen or argon gas in a flow chamber. Note: Do not place the membrane surface of the sensor on any undissolved sulfite crystals; this prevents

scratching and damaging of the membrane. Refer to the following Residual Performance Check table for sensor response for this procedure. Note: This verification procedure is carried out on all Orbisphere sensors before leaving our manufacturing facility. Your sensor should achieve the minimum performance times when placed in a deaerated sample; response reflects integrity of sensor properly maintained sensors are typically faster than the times listed. Residual performance check Concentration span Expected response time (for specific membranes) Starting from Down to 2953A 2958A 2952A Saturation 100 ppb 2 minutes 5 minutes 10 minutes 100 ppb 10 ppb 15 minutes 30 minutes 1 hour 10 ppb 2 ppb 10 minutes 1 hour 2 hour 3. Check sample lines for leaks The final step in verifying your oxygen analyzer performance is checking for air leaks in the sample lines, fittings, and valves used to bring the sample to the oxygen sensor. All Orbisphere dissolved oxygen sensors have a flow insensitive range when used in an Orbisphere flow chamber. Thus, you can leak-check your sample lines for air leaks, by varying the flow rate via this flow chamber. First, set the flow rate to 100 ml/min and achieve a stable reading, then slowly increase the flow rate to 200 ml/min. This flow rate is applicable for leak testing with any membrane, regardless of the recommended flow rate of that membrane. If you have airtight sample lines, the displayed oxygen value should either increase (very slightly) or remain constant. If you see a significant decrease in the displayed value, it is probably due to an air leak in your sampling system. Decreasing the flow should cause the opposite effect. The explanation of this phenomenon is that a small leak (such as a pin hole in the sample line around a fitting or valve that is too small to leak water) in a flowing line causes a venturi effect that sucks in air. A two-fold increase in the flow rate will not increase the venturi effect by double and therefore the concentration of DO 2 will decrease. Indeed, it is normally observed that the leak rate remains constant and the DO 2 decreases in inverse proportion to the flow. Please note that even if water is leaking out of a sample line, air is probably getting into your sample. A leak can be located by resetting the flow rate to 100 ml/min, then systematically covering fittings, valves, or pump seals with water from a lab squirt bottle or nitrogen gas while watching the oxygen display. When the leak point is momentarily blocked, the oxygen display will decrease noticeably. Tighten or replace the leaking fitting or valve (following the manufacturer s recommendations).

Note: Plastic tubing (even those made with low permeability materials) can also contribute to the oxygen displayed, causing the same effect. Therefore, replace all sample plastic tubing with stainless steel tubing if possible. Refer to the table on the next page for typical oxygen leakage effects using various types of plastic tubing. Additional practical tips Several additional "good practice" measures can be taken to assure the best performance and response from your oxygen sensor and analyzer. 1. Cool the sensor (if possible) Your Orbisphere oxygen sensor can be safely exposed to sample temperatures up to 100 C, but can only yield accurate temperature compensated results up to 50 C to 60 C, depending on the membrane choice. However, the sensor maintenance interval becomes significantly longer as the sample temperature decreases down to 25 C and below. The addition of another heat exchanger (or a sample chiller in warm climates) can significantly increase the service interval, and quickly pay for itself. 2. Use barrier films if measuring in harsh conditions When measuring in harsh conditions, such as in higher than ideal temperatures, or in high levels of ammonia, ethanol-amines, hydrocarbons, or hydrosulfites, you will lengthen your maintenance interval if you use a barrier film (membrane mask). The barrier film limits the effective area of the membrane to only that covering the central electrode. The overall reaction rate of the sensor is thereby reduced, thus increasing the service interval. Note: Be sure to use the correct model of membrane holding ring for your membrane and barrier film to assure a proper fit. 3. Turn off the instrument when not measuring When an Orbisphere oxygen sensor is operated in relatively high DO 2 conditions for long periods, it will dramatically shorten the maintenance interval. This is easily avoided. When your generating unit shuts down, and measurements are not required, simply switch off your analyzer. As long as the instrument is off, the sensor will not react electrochemically, thus lengthening your maintenance interval. If the unit will be down for months, it is advisable to remove the sensor, remove its membrane, rinse, and store dry until needed.

Common plastic materials used for sample tubing, showing leakage of oxygen into water flowing through tubing due to the permeability of the material and listed from least permeable to most: Leakage of oxygen through plastic tubing walls Material O 2 pickup in water, ppb/meter O 2 reading at 10 minutes, ppb Polyvinylidene chloride (Saran 1 ) 0.02 2.7 Nylone 0.03 3.8 Polychloro trifluoroethylene (Kel-F 2 ) 0.05 4.6 Polyvinyl fluoride (Tedlar 3 ) 0.05 4.7 Polyvinylidene fluoride (Kynar 4 ) 0.1 6.7 Polyethylene Terephthalate (Mylar 3 ) 0.12 6.9 Polyvinyl chloride (non-plasticized) 0.14 7.6 Polyacetal (Delrin 3 ) 0.2 7.8 Ethylene/Monochlorotrifluoroethylene copolymer (Halar 5 ) 0.43 13.6 Ethylene/Tetrafluoroethylene copolymer (Tefzel 3 ) 1.70 27 High density polyethylene (opaque) 2.04 30 Polypropylene 3.2 38 High density polyethylene (clear) 3.9 42 Polycarbonate (Lexan 6 ) 5.1 46 Polystyrene 5.3 48 Low density polyethylene 8.5 60 Fluorinated ethylene/propylene (FEP) 13 77 Tetrafluoroethylene (PTFE) 19 94 Natural rubber (Latex) 60 150 Silicone rubber (Silastic) 1700 1700 4. Clean the sensor thoroughly As Orbisphere sensors have improved over the years, cleaning techniques have also evolved. In addition to the standard chemical cleaning method described in the Sensor Service instructions supplied with the Orbisphere analyzer, Orbisphere strongly recommends the use of a model 32301 Sensor Cleaning and Regeneration Center, which provides a thorough electrochemical sensor cleaning. The 32301 Cleaning Center also can be used to perform sensor diagnostics when an electrical fault is suspected.

5. Use Orbisphere support When you purchase an Orbisphere analyzer, you also purchase Hach ServicePlus support and service support. Please do not hesitate to call your local Hach sales office or national Hach sales and service office if you have any problems, questions, or comments. 1 Saran is a registered trademark of Dow Chemical. 2 Kel-F is a registered trademark of 3M. 3 Delrin, Mylar, Tedlar, and Tefzel are registered trademarks of DuPont. 4 Kynar is a registered trademark of The Pennwalt Corporation. 5 Halar is a registered trademark of Ausimont U.S.A., Inc. 6 Lexan is a registered trademark of General Electric. Operating check list The following checklist is provided as a quick reference to the tips and recommended practices in this Application Note. Practical tips Calibrate your sensor Verify your sensor s residual (zero) reading Check your sample lines for leaks Other recommended practices Cool the sensor Use barrier film in harsh conditions Turn off Instrument when not measuring Clean sensor thoroughly Use Orbisphere (Hach ServicePlus) support FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: Tel: 800-227-4224 E-Mail: techhelp@hach.com To locate the HACH office or distributor serving you, visit: www.hach.com LIT2847 Hach Company, 2015. All rights reserved. In the interest of improving and updating its equipment, Hach Company reserves the right to alter specifications to equipment at any time.