Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Similar documents
Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Recording established using the VO2 and RER graph template file (h19.gtl)

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

BSL PRO Lesson H19: V O 2, V CO 2, RER, O 2 deficit and EPOC Measurement. Objectives: Equipment:

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7.

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts

Air Flow & Gas Analysis Accessories

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

Resting, Active, and Exercising Metabolic Rates

Review of gas frac9ons in atmospheric air 2/2/16. Metabolic Calculations of Indirect Calorimetry. Expired gas analysis.

Content Display. - Introduction. Laboratory Unit : Lab 2 - Measurement of Oxygen Consumption. KINE xxxx Exercise Physiology

Overview. Front Panel: Keypad and Display

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities

PRODUCT SHEET. See also: Student Accessory Pack BSL-ACCPACK and BSLACCPACK-11B AFT7 AFT7L AFT12 AFT6A AFT27 AFT16

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

PRODUCT SHEET. Order probe only as RXPROBE02

Gas exchange measurement in module M- COVX

GE Healthcare. CARESCAPE R860 Nutritional Assessment Tools Appliguide

Figure 1. A schematic diagram of the human respiratory system.

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Lesson 12 New Procedure PULMONARY FUNCTION I

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Application Note AN-107

CARDIOVIT AT-104 ergospirometry

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

BREATH-BY-BREATH METHOD

RM-80 respiration monitor

Lung Volumes and Capacities

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

MIL-STD-883G METHOD

VT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features

VT650. Gas Flow Analyzer. Technical data. Key benefits and features:

FOURIER TRANSFORM INFRARED SPECTROSCOPY

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

TR Electronic Pressure Regulator. User s Manual

Respiratory Response to Physiologic Challenges. Evaluation copy

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

Biology Unit 2, Structure of Life, Lab Activity 2-3

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

Overview. 2 Module 13: Advanced Data Processing

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Selecting the right pressure sensor for your application

Columbus Instruments

Pneumatic high-pressure controller Model CPC7000

Protective Atmospheres, Measurement Technologies and Troubleshooting Tools

LUMAZOTE T ECHNICAL BRO CHU RE

Biology 13A Lab #12: The Respiratory System

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

Blood Parameter Monitoring System 550

HESTA Issues affecting the accuracy of Metabolic Carts. Danny Rutar

Technical Data Sheet MF010-O-LC

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

TSE Blood Pressure Monitor Invasive series

Test Plans & Test Results

IMPC phenotyping SOPs in JMC

Improve Process Reliability

Cover Page for Lab Report Group Portion. Pump Performance

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

VT900. Gas Flow Analyzer. Technical data. Key benefits and features:

Process Control Loops

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Chapter 13 Gases. H. Cannon, C. Clapper and T. Guillot Klein High School. Pressure/Temperature Conversions

HumiSys HF High Flow RH Generator

Pneumatic high-pressure controller Model CPC7000

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

Quality assurance of Modular Metabolic Systems. Danny Rutar, Redback Biotek, Ireland/Australia Jeffrey Swen(Intern), University of Ulster, Belfast.

Phosgene Sensor. Sensoric COCl2 3E 1

RESPIRATORY REGULATION DURING EXERCISE

12. Laboratory testing

Applications Note: Use of "pentane equivalent" calibration gas mixtures

ALVEOLAR - BLOOD GAS EXCHANGE 1

Mouse and Rat Tail Cuff Blood Pressure Systems from 1-24 Animals. Mouse rat tail cuff blood pressure system components

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

Datasheet: K-30 ASCII Sensor

Determination of the Frequency Response of an End Tidal CO2 Analyser

This test shall be carried out on all vehicles equipped with open type traction batteries.

Let s talk about Capnography

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

Independent Health Facilities

AC : ACCURATE CRYOCHAMBER FOR A SMALL LABORATORY WITH SMALL BUDGET

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Transcription:

Application Note 268 Metabolic Analysis Systems and Propane-based Verification 8.10.11 This application note examines a conventional equipment setup for metabolic measurements. The setup is exposed to a propane-based, combustion verification source that emulates subject s steady-state consumption of oxygen and production of carbon dioxide. A propane-based, combustion source is a very reliable and accurate metabolic gas generator that can be used to verify the performance of a metabolic measurement system. Metabolic Analysis System Setup For this application note, the three input signals have been assigned to MP150 data collection channels: CO 2 e Carbon dioxide concentration via CO2100C module, set to a gain of 1% CO 2 per volt O 2 e Oxygen concentration via O2100C module, set to a gain of 10% O 2 per volt Fi Inspired flow from TSD107B transducer via DA100C amplifier, set to a gain of 1,000 Figure 1 illustrates the typical connections when performing metabolic analysis. Figure 1 Typical connections when performing metabolic analysis The subject breathes through the non-rebreathing T-valve facemask/mouthpiece (AFT25) which attaches directly to the TSD107B air flow transducer on the inspiration side and the AFT15A mixing chamber on the output side. When the subject inspires, air is drawn through the TSD107B, as shown by the arrows, and then passes through the AFT25 non-rebreathing T-valve into the subject s lungs. The TSD107B is placed on the inspiration side to eliminate any effects associated with expired air humidity and associated condensation. When the subject expires, air is directed out to the 5 liter mixing chamber (AFT15A). The CO2100C and O2100C gas concentration sampling modules connect to the mixing chamber via separate AFT20 gas sampling interface kits. The gas sampling interface kits act to filter and dehumidify (via Nafion tubing) the expired gases residing within the mixing chamber. Page 1 of 7

The non-rebreathing "T" valve directs only expired air to the 5 liter mixing chamber. Because only expired air is directed to the chamber, the mixing chamber acts to average respiratory outflows. This averaging effect causes the CO 2 and O 2 concentrations to vary in accordance to the mean values resident in several expired breaths. The size of the mixing chamber determines the extent of the averaging effect. For example, assuming the subject s expired breath volume is typically 0.5 liter (as during resting metabolic rate measurements), the AFT15A mixing chamber will average about 10 expired breaths. The TSD107B airflow transducer signal is amplified by a DA100C amplifier, which is attached to the MP150 system along with the CO2100C and O2100C modules. Accordingly, there are three channels of analog data that are acquired to properly calculate metabolic Oxygen Consumption (VO 2 ) and the Respiratory Exchange Ratio (RER). AcqKnowledge is setup to record on three analog channels and multiple calculation channels. Data is acquired to computer memory, or hard disk, with an acquisition time of at least several minutes. Once the setup is completed, with all the plumbing fixed and secure, gas concentration calibration can proceed. If airflow plumbing, gas sampling line plumbing or gas measurement module pump speed are changed, it s generally necessary to repeat the gas measurement module calibrations. Gas Concentration Calibration After the software setup, it s important to calibrate the system for the following three variables: 1) Carbon Dioxide concentration (calibrate in range of.04% to 4% CO 2 ) 2) Oxygen concentration (calibrate in range of 20.93% to 16% O 2 ) 3) Inspired Flow (calibrate within the range of 0 to 13 liters/sec) To calibrate the gas concentration sensors, the mixing chamber is flooded with the appropriate gas. First flood the chamber with fresh (ambient) air by attaching the calibration syringe to the subject mouthpiece and cycling fresh air into the mixing chamber. Monitor the gas concentration changes using AcqKnowledge. When the levels appear stabilized with ambient air, obtain the first calibration point (CO 2 at 0.04% and O2 at 20.93%). Then flood the chamber with the desired gas mixture (recommended - BIOPAC s GASCAL and GASREG: 4% CO 2, 16% O 2 and 80% N 2 ). When the levels appear stabilized, obtain the second calibration point. In this fashion, the CO200C module will be calibrated between 0.04% to 4% carbon dioxide and the O2100C module will be calibrated between 20.93% and 16% oxygen. The airflow transducer -pneumotach- (TSD107B) has a factory scale factor of 90uV per liter/second. This scale factor implies that when 1 liter/second of air is moving through the TSD107B, then it will output a signal of 90uV. The TSD107B is attached directly to the DA100C, set to a gain of 1000. This amplification results in a 90mV signal under the same airflow conditions. Calibration is estimated using the following procedure: 1) Record the first calibration value (value X) with no airflow through TSD107B 2) In the scaling dialog window, X volts will map to 0 liters/second 3) Calculate the inverse of the factory scale factor 1/0.090 = 11.111 (liters/sec per volt) 4) Add 11.111 to X to obtain value Y 5) In the scaling dialog window, Y volts will map to 1 liter/second The second point calibration is estimated by adding the inverse of the factory scale factor to the offset recorded by the first calibration point. The TSD107B can be precisely calibrated by using a calibration syringe to inject a specific volume of air through the airflow transducer. While recording the airflow signal in AcqKnowledge, note that the signal will rise and then fall in accordance to the syringe movement. The area pop-up measurement in AcqKnowledge is used to perform an area calculation on the airflow signal as shown in Figure 2. Page 2 of 7

Figure 2 Airflow signal area calculation Assuming a 2-liter calibration syringe (BIOPAC AFT26) is used, the measured area should be 2 liters. If the measured area is something different, then the scale factor of 11.111 liters/second (used in scaling dialog window) should be adjusted upwards or downwards in direct proportion until the area measurement returns 2 liters. Other factors to be recorded for the procedure are the following: Ta Pb PH 2 O Ambient temperature Ambient barometric pressure in mmhg Pressure of water vapor at ambient temperature in mmhg These values are important when performing the conversion of gas at Ambient Temperature and Pressure (ATP) to Standard Temperature and Pressure, Dry (STPD). VO 2 and RER Measurement Setup Analysis Because both the carbon dioxide and oxygen concentrations are known, it s possible to determine the concentration of expired nitrogen using the following formula: N 2 e = 100 [O 2 e + CO 2 e] Since the TSD107B is placed in the "Flow In" line, the flow of inspired air can be determined over a running 60- second interval, by integrating the flow signal using the expression (60 seconds equals 6000 samples at a 100Hz sampling rate): Convert gas at ATP to STPD: Vi = Integrate (Fi) over 60 second period Vis = Vi*(273/(273+Ta))*((Pb-PH 2 0)/760) Where Ta is ambient temperature (e.g. nominally 24 deg C) Pb is ambient barometric pressure (e.g. nominally 745 mmhg) PH 2 0 is ambient pressure of water vapor (e.g. nominally 22.4 mmhg) Calculate the expired volume from the inspired volume by using the Haldane transformation (Nitrogen factor): Ves = (Vis*79.03) / N 2 e Where the value 79.03 is the percent nitrogen in ambient air Page 3 of 7

The real-time oxygen consumption volume can be determined by using the expression: VO 2 = (1/100)*[(Vis*20.93) (Ves*O 2 e)] Where the value 20.93 is the percent oxygen in ambient air The real-time carbon dioxide production volume can be determined by using the expression: VCO 2 = (1/100)*[(Ves* CO 2 e) (Vis*.04)] Where the value.04 is the percent carbon dioxide in ambient air Finally, RER can be calculated by: Definitions CO 2 e O 2 e Fi N 2 e Vi Vis Ves VO 2 RER = VCO 2 / VO 2 Carbon Dioxide fractional concentration in expired air Oxygen fractional concentration in expired air Inspired air flow (ATP) Nitrogen fractional concentration in expired air Inspired air volume (ATP) Inspired air volume (STPD) Expired air volume (STPD) Volume of oxygen consumed (STPD) VCO 2 Volume of carbon dioxide produced (STPD) RER Note Respiratory Exchange Ratio All volume-related units are expressed in liters per minute, because the integration interval is over a 60 second period. Figure 3 illustrates typical values obtained when performing gas concentration and flow measurements from a subject. The CO2100 module was set to a gain of 1% CO 2 per volt, the O2100 module was set to a gain of 10% O 2 per volt and the DA100C amplifier (connected to TSD107B) was set to a gain of 1000. Figure 3 Typical gas concentration and flow measurements Page 4 of 7

The preceding graph illustrates the complete real-time Oxygen Consumption and Respiratory Exchange Ratio measurement taken from a subject riding a stationary bicycle. Both raw and calculated waveforms are displayed. Note that the VO 2, VCO 2, Weir Calculation and RER measurements (bottom four channels) vary smoothly with time. The graphical and continuous nature of this recording and calculation method provides significant information regarding the changes of these variables as a function of time. This measurement setup will provide a reading that, at any time, indicates the oxygen consumed, the carbon dioxide produced and the respective respiratory exchange ratio in the last 60 seconds. Accordingly, the peak VO 2 reading at 24 minutes - of 4.7 liters of oxygen consumed is indicative of the amount consumed by the subject in the previous 60 seconds. The RER at the same point in time is determined to be 0.91 and the Weir calculation indicates a peak value of 23.1 kcal/min. Gas Concentration Sensor Response Time A very important issue associated with metabolic systems calibration is the issue of gas concentration sensor response time. Because the RER is calculated as VCO 2 /VO 2, the response times of the carbon dioxide and oxygen sensors must match very closely. Figure 4 Gas concentration sensor response times As an example, consider the gas sensor response times indicated in Figure 4. In this rise-time and fall-time test, nitrogen is used to flood the chamber to bring both the CO 2 and O 2 concentrations to 0%. After 2-3 minutes, the chamber is flooded with a gas mixture containing 4% CO 2, 16% O 2 and 80% N 2 (BIOPAC GASCAL). The first two waveforms are representative of gas sensors that have very different response times. If these sensors are used in a metabolic system, they will only provide accurate RER measurements about 100 seconds after a quick change in gas concentrations. For these sensors, because the CO 2 sensor responds much slower than the O 2 sensor, the RER will be underestimated for quickly rising VCO 2 (and falling VO 2 ) and overestimated for quickly falling VCO 2 (and rising VO 2 ). However, with steady-state conditions, as with resting and basal metabolic rate measurements, RERs will be reported correctly even on gas systems that have very different response rates between the oxygen and carbon dioxide sensors. This is because the rate of change in gas concentration is very slow, so both sensors can track equivalently. The bottom two waveforms are representative of gas sensors that have very similar (and fast) response times. These sensors (CO2100C and O2100C) will accurately report RER under a very wide range changing gas concentrations. Page 5 of 7

In the case of the CO2100C and O2100C, both measurement units are very high speed (under 500 msec response time), so RERs are reported very accurately even under rapidly changing gas concentration conditions. Generally used, the outputs from the CO100C and O2100C are low pass filtered (1 Hz LPF) in AcqKnowledge to create smoothly varying waveforms. Propane-based Verification of Metabolic System Figure 5 illustrates the setup to perform a propane-based combustion verification of the previously referenced metabolic analysis system. Figure 5 Typical connections when performing propane-based combustion verification of metabolic analysis system In this situation, a propane-based combustion chamber replaces the subject. Also, a fan is added in-line to the AFT15A output. The fan draws air from the combustion chamber and then to the AFT15A where the resulting gas concentrations are sampled by the CO100C and O2100C. The combustion chamber has dimensions of 55cm x 81cm x 183cm, providing an internal volume of 815,265 cm 3 or approximately 815 liters. Inside the chamber is a burning propane torch. The propane torch is set to a moderate flame size and the resulting flame is well-oxygenated. The fan draws air through the AFT15A and then from the top of the combustion chamber. Fresh air enters through the bottom of the combustion chamber to continue to source oxygen to the propane torch. Inside the combustion chamber, the propane torch sits on a sensitive scale that measures weight with a resolution of 0.1grams. As the propane burns, it produces 11.92 kcals per gram of propane used. The scale determines the amount of propane consumed by the torch over some period of time. As the propane burns it has an effective RER of 0.60. This means that the volume of carbon dioxide produced by the torch is 60% of the volume of oxygen consumed by the torch. For industrial propane, the effective RER is in the region of 0.58 to 0.60. For instrument grade propane, the RER is 0.60. The Weir equation can also be evaluated with the combustion-based calibration test. The Weir equation is: kcal = (3.9*VO 2 ) + (1.1*VCO 2 ) - all dimensions on a per minute basis VO 2 and VCO 2 are the volume of oxygen consumed and the volume of carbon dioxide produced, respective, in units of liters/minute. Page 6 of 7

The graph in Figure 6 demonstrates proper metabolic system behavior, when subjected to propane-based verification. Once stabilized, after approximately 60 seconds, VO 2 is 0.297 liters/min, VCO 2 is 0.180 liters/min and the Weir calculation is 1.36 kcal/min and RER is approximately 0.6. Figure 6 Proper metabolic system behavior, when subjected to propane-based verification Artificial subject measurements are shown in Figure 7. In this case, the subject shown in Figure 1 is replaced with the propane-based combustion chamber. As illustrated in the graph, the RER is in the range of 0.58 +/- 0.01 throughout the duration of the test, even though the airflow is markedly changed over the course of the recording. VO 2 maximum is 1.645 liters/min at 90 seconds into the recording, and at that time the Weir calculation is 7.46 kcal/min and RER is 0.575. Figure 7 Artificial subject measurements Page 7 of 7