Operation of the Perkin Elmer TGA-GC/MS

Similar documents
5890II GC Standard Operating Procedure 9/2/2005

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Procedures for operation of the TA Instruments DSC

System Overview. TCD Detector temperature setpoint Regulator

Thermo K-Alpha XPS Standard Operating Procedure

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

GCMSD-Headspace Analysis SOP

JABATAN KIMIA Kategori Arahan Kerja

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

Addressing the World Shortage of Helium

Vapor Sorption Analyzer

A New GC/MS System Designed for Helium Carrier Gas Conservation

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

Quantitative Analysis of Hydrocarbons by Gas Chromatography

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Mass Spec will not Autotune

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

FOURIER TRANSFORM INFRARED SPECTROSCOPY

SOP: Buck Scientific BLC-20P HPLC Operation

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

FLUORESCENCE DETERMINATION OF OXYGEN

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510)

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Hiden QIC Series Gas Analysers Real-time gas analysis systems

1)! DO NOT PROCEED BEYOND THIS MARK

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

Quick Start Guide for SA-9600 Series Surface Area Analyzer

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

The MRL Furnaces USED FOR THIS MANUAL COVERS

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

CP-3800 GC Operator s Manual

SHIMADZU LC-10/20 PUMP

Calibration of a GC955

Best Practice for Identifying Leaks in GC and GC/MS Systems

Updated: Variable Temperature

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Boyle s Law: Pressure-Volume Relationship in Gases

FROG-4000 TM Chemical Analysis System

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

AKTA pure 25 New Owner s Intro

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

P3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Ammonia Treatment Instructions Page 1 of 5

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Xactix XeF2 OPERATION MANUAL

Adsorption Unit Operating Procedures: CO2 Adsorption on to Activated Carbon

1. Study the performance of a binary distillation column operated in batch mode.

Agilent Inlet EPC Leak Test

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

FROG-4000 TM Chemical Analysis System

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Customer Responsibilities

Prep LC Training Manual

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Figure 1: Plug in ion inlet of mass spec

Agilent 7697A Headspace Sampler

FROG-5000TM Chemical Analysis System User s Manual

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Generating Calibration Gas Standards

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Armfield Distillation Column Operation Guidelines

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

Quadruple mass spectrometers (transducer type)

Operator Quick Guide ORBISPHERE 3654

G200 Analyzer Range. Operating Manual

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

3 Flow and Pressure Control

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Prep LC Training Manual S.V.

PerkinElmer Aanalyst100

User Manual. Quantos Automated Dosing Powder Module

Gas Diluter Portable Gas Dilution System User s Manual

Inventory User Guide

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

Agilent 490 Micro GC Natural Gas Analyzer

G200 RANGE G200 AND G210 OPERATING MANUAL

Using the UltraRAE. Firmware 2.35

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

Determination of Zn using Atomic Absorption with Multiple Standard Additions

Using the Akta Prime plus October 22, 2012

Residual Gas Analysis Systems for Industry

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Multiple Gas#5 GC configuration Jan 2016

Arizona State University NanoFab XACTIX ETCHER. Rev A

Basic ICP Operating Procedures

Transcription:

Operation of the Perkin Elmer TGA-GC/MS Summary of the TGA-GC/MS: The TGA-GC/MS allows a user to decompose a sample by heating, measure its loss of mass, and simultaneously analyze the chemical composition of the sample. The instrument consists of three primary components: a thermogravimetric analyzer (TGA), a gas chromatograph (GC), and a mass spectrometer (MS). It can run in three operating modes: TGA, TGA-MS, and TGA- GC/MS. The TGA can operate at temperatures ranging from room temperature to 1100 C under flowing air, nitrogen, or helium. The mass spectrometer (MS) is used to measure the composition of the evolved the gases, while the gas chromatograph (GC) can be used to slow down the flow of each component through the column and better distinguish a sample s chemical composition. Figure 1. Perkin Elmer TGA-GC/MS Assembly. Modes of Operation: I. TGA II. TGA-MS III. TGA-GC/MS *Note: For TGA-GC/MS operation you will need to run a TGA-MS experiment, find the decomposition point of your sample, configure the GC to trigger at the decomposition point of your sample, and then run a TGA-GC/MS experiment. 1

Safety Concerns and Precautions Inspect the instrument to ensure that it is configured properly for your experiment. Check the temperature of the TGA before lowering the furnace. Warning When using the MS, helium must be used as the sample gas and purge gas. Warning Ensure that your sample pan is stable at the experimental temperatures that you will utilize. Aluminum pans cannot be heated above ~550 C, alumina pans cannot be heated above 700 C. Warning The flow rate of the balance purge must be higher than the sample purge, otherwise corrosive gases may enter the balance and cause damage. Typical settings are: Sample Purge: He = 30 ml/min; Balance Purge: He = 50 ml/min. Preliminary Instrument Inspection: Make sure the instrument is configured properly before starting your experiment. Contact the IAC staff if you have any questions. Helium gas must be used for experiments utilizing the MS (air cannot be used). Check the purge gas valve located behind the TGA. For GC/MS use, the valve must be opened for helium flow (pointed towards the copper tubing). For TGA use, the valve should be opened for nitrogen flow (pointed towards the plastic tubing). Figure 2. Purge gas valve (a) when set to helium for TGA-GC/MS or TGA-MS operation; and (b) the purge gas valve when set to nitrogen for TGA operation. I. TGA Operation: 1. Log onto the Instrument Log onto the PC in the IAC. 2. Turn on the TGA Flip the Toggle Switch on the back of the instrument. 3. Open Pyris Click the Pyris Software icon on the PC. 4. Load the Autosampler Place your empty pans in the TGA s autosampler. 2

5. Tare your Empty Pans Press the autosampler button and tare each empty pan by using the following buttons: Load Sample, Raise Furnace, Zero Weight, Lower Furnace, and Unload Sample. 6. Load Each Pan Weigh ~1-10 mg of sample and load each sample pan. 7. Check Purge Gas Settings Go to Tools, Preferences, Purge Gas Tab; Sample Purge A and Balance Purge should be set to nitrogen or air (only for TGA analysis). 8. Create a TGA Method Press the Method Editor button. Press File then Open Method. A series of preset methods have been created in Pyris, select General TGA. Select Save Method as to create a unique method for your samples. Enter sample information, weight, file name, etc. 9. Enter Initial State Information Select the Initial State tab and enter the proper settings for your sample. Nitrogen or air can be used only when the GC/MS is not in use. 10. Enter the Program Info Set the instrument to ramp and dwell at a desired rate. 11. Start the TGA Run In Pyris hit the Start/Stop button: 12. Monitor the Experiment Ensure that the experiment heats to the desired temperature. 13. Remove Samples from Furnace After the run has completed, allow the furnace to cool. Use the autosampler to remove your sample pans from the furnace. 14. Turn off the Instrument Set the temperature of the furnace to 30 C. Flip the Toggle Switch on the back on the TGA. Close Pyris and log off of the instrument. II. TGA-MS Operation: 1. Log onto the Instrument Log onto the PC in the IAC. 2. Open the Helium Tank Open the main valve on the cylinder and the smaller valve connected to the copper tubing. Leave the regulator in its standard position at 90 psi. 3. Turn on the MS Flip the Toggle Switch on the side of the instrument. 4. Turn on the GC Flip the Toggle Switch on the side of the instrument. Wait a few minutes for the green light on the front of the MS to turn on. 5. Turn on the TGA Flip the Toggle Switch on the back of the instrument. 6. Turn on the Controller Box Flip the Toggle Switch on the back of the TL8500. The controller will begin heating the transfer line. Depending on the sample being analyzed, the final temperatures should be set between ~270-330 C. 7. Turn on the Pump Press the green Pump Button on the TL8500 controller box after it reaches 50 C to initiate the pump. 8. Open Pyris Software Pyris is used to control the TGA. 9. Open Turbomass Software Turbomass controls the GC/MS. 3

Configuring the GC/MS in Turbomass: 10. Pump out the GC/MS Open the Tune Page in Turbomass by clicking on the glasses icon. Click Options then Pump/Vacuum System On. 11. Allow the System to Equilibrate Wait 15 minutes for the instrument to equilibrate. The MS should reach 200 C and the pressure on the tune page should be in the green. 12. Check the Tuning of the MS Check to make sure the instrument is properly tuned. a) Click the Press to Operate button in the bottom right corner of the tune page. b) Click the Gas option on the top tool bar. c) Press Reference Gas On allow a few minutes for the instrument to equilibrate. d) Check the ions: The nitrogen ion (ion 28) should be less than the reference gas ion (ion 69) and there should be a small quantity of ion 502; otherwise there may be a leak in the system. 13. Turn off the Reference Gas Turn off the reference gas by clicking the reference gas icon. 14. Turn off the Filament Press the Press for Standby button on the tune page. 15. Create an Experiment in Turbomass Click Add Sample and enter sample details. Enter an appropriate file name, sample ID, conditions, notes, etc. 16. Select a MS and GC Method General experimental methods have been created for sample analysis. However, the instrument can also be tailored to meet the needs of your sample. For a general TGA-MS experiment select: MS Method TG-MS 15min 31-300 amu GC Method TG-MS 17. Start the Run in Turbomass Highlight the row that you would like to run and press the Start Run button. A green dot should appear next to the row. At this point the GC/MS is waiting to be triggered by the TGA. Configuring the TGA in Pyris: 18. Load the Autosampler Place your empty pans in the TGA s autosampler. 19. Tare the Empty Pans Press the Autosampler button and tare each empty pan by using the following buttons: Load Sample, Raise Furnace, Zero Weight, Lower Furnace, and Unload Sample. 20. Load Each Pan Weigh ~1-10 mg of sample and load it into each pan. 21. Check Purge Gas Settings Helium must be used for TGA-MS and TGA-GC/MS analysis. Go to Tools, Preferences, Purge Gas Tab. Sample Purge A and Balance Purge should be set to helium. 22. Create a TGA-MS Method Press the Method Editor button. Press File then Open Method. A series of preset methods have been created in Pyris, select General Helium TG-MS. Select Save Method as to create a unique method for your samples. Enter proper sample information, weight etc. 4

23. Enter Initial State Information Select the Initial State info tab and enter the proper settings for your sample. Helium must be used when utilizing the MS. 24. Enter the Program Info Below is an example program used for TGA-MS. Make sure the end condition is: Go to 50 C, this will prevent the furnace from dropping immediately at the end of the run and exposing the MS to air. Check to make sure the total TGA method is longer than the MS and GC methods. Ensure that Helium Gas is selected during MS in the bottom right corner of Pyris. Figure 3. General TG-MS method in Pyris. 25. Start the TGA-MS Run In Pyris hit the Start/Stop button: 26. Monitor the Experiment Ensure that the TGA triggers the GC/MS. 27. Remove Samples from Furnace After the run has completed, use the autosampler to remove your sample pans from the furnace. 5

Figure 4. Turbomass and Pyris during the initial stages of TG-MS operation. III. TGA-GC/MS Operation: Note: For TGA-GC/MS operation you will utilize the same heating method utilized for TGA-MS operation; however, you will set the Pyris TGA-GC/MS method to trigger the GC/MS at the time in which a peak was detected in the TGA-MS chromatogram. 1. Starting from Step #15 in the above TGA-MS Procedure: 2. Create an Experiment in Turbomass Click Add Sample and enter sample details. 3. Select a MS and GC Method General experimental methods have been created for sample analysis. For a general TGA-GC/MS experiment select: MS Method TG-MS 15min 31-300 amu GC Method TG-GC-MS 15 min 4. Start the Run in Turbomass Highlight the row that you would like to run and press the Start Run button. A green dot should appear next to the row. At this point the GC/MS is waiting to be triggered by the TGA. 5. Load the Autosampler Place your empty pans in the TGA s autosampler. 6. Tare the Empty Pans Press the autosampler button and tare each empty pan by using the following buttons: Load Sample, Raise Furnace, Zero Weight, Lower Furnace, and Unload Sample. 7. Load Each Pan Weigh ~1-10 mg of sample and load it into each pan. 8. Check Purge Gas Settings Helium must be used for TGA-MS and TGA-GC/MS analysis. Go to Tools, Preferences, Purge Gas Tab. Sample Purge A and Balance Purge should be set to helium. 6

9. Create a TGA-GC/MS Method Press the Method Editor button. Press File then Open Method. A series of preset methods have been created in Pyris, select General Helium TG-GC-MS. Select Save Method as to create a unique method for your samples. Enter sample information, weight etc. 10. Enter Initial State Information Select the Initial State info tab and enter the proper settings for your sample. Helium must be used when utilizing the MS. 11. Enter the Program Info Below is an example program used for TGA-GC/MS. Make sure the End condition is: Go to 50 C, this will prevent the furnace from dropping and exposing the MS to air. Check to make sure the total TGA method is longer than the MS and GC methods. Ensure that Helium Gas is selected in the bottom right corner. Figure 5. General TGA-GC/MS method in Pyris. 12. Program GC/MS Trigger Time In the Program tab on the method editor page, edit External Channel 1 to turn on and turn off at a desired point in time during your experiment. This trigger should occur at the point in time where a broad peak was observed during the TGA-MS experiment. 13. Start the TGA-GC/MS Run In Pyris hit the Start/Stop button: 14. Monitor the Experiment Ensure that the GC/MS is triggered at the proper time. 15. Remove Samples from Furnace After the run has completed, use the autosampler to remove your sample pans from the furnace. 16. Turn off the Instrument Set the temperature of the furnace to 30 C. Flip the Toggle Switch on the back on the TGA. Close Pyris and log off of the instrument. 7

Figure 6. Turbomass and Pyris during TGA-GC/MS operation. GC/MS Shutdown Procedure: *Note: The GC/MS should be turned off and the helium cylinder should be closed if you are the last user of the week. 1. Open Turbomass 2. Open the Tune Page Click the Glasses Icon to open the Tune Page. 3. Change Temperature Set Points Change both the Inlet GC Line Temperature and the Source Temperature (in the source parameters area) from 200 C to 50 C. *Note: Allow the temperatures to drop below 100 C before continuing. 4. Turn Off the Vent/Vacuum System Once the temperature has dropped below 100 C, go to Options at the top of the menu bar and select Vent/Vacuum System Off. 5. Turn off the Turbo Pump Open the Diagnostics tab and change Turbo Speed from 100% to 0%. Allow several minutes for the turbo pump to power down to zero. 6. Exit Turbomass Close the Tune Page and Turbomass windows. Allow the software to fully close. 7. Turn Power Off on the GC & MS Flip the Toggle Switches on the sides of both the GC and MS. 8. Close the Helium Tank Close the main valve on the helium tank. *Note: Check the helium cylinder to make sure there is adequate helium for future operation. Contact IAC staff if the pressure on the helium cylinder is low. 8

Appendix Pyris Software Figure 7. Pyris software window layout. Instrument View Method Editor Pyris Player Data Analysis Start/Stop Go to Temp Go to Load Go to Hold Autosampler Control Sample Weight Zero Weight Raise Furnace Lower Furnace Reset Monitor Antistatic Fan Clean Furnace 9

Figure 8. Method Editor window tabs. 10

Turbomass Software Figure 9. Turbomass software Figure 10. Turbomass tune page 11

Method Editor View Spectrum View Chromatogram View Map View Quantify Results Library Search Combine Functions Strip Mw Calculator Environmental Reports Add Sample Sample List Wizard Inlet Editor Tune Page MS Method Editor Scope Parameters Reinitialize Instrument Noise Level Standard Ultra Tune Custom Ultra Tune Reference Gas CI Gas Insert Sample Customize Field Display 12

Emergency Information: Medical Emergencies: Contact 911 and Public Safety (609) 258-1000 Room / facility emergencies: Contact Public Safety (609) 258-1000 Issues related to the instrument: 1. Contact IAC Staff. 2. If unsure, leave system as is. 3. Try to power down the system. Audible/Siren Emergency Alerts: Follow previous steps 2 & 3 and leave the building. Emergency Contact Information: Nan Yao: Office (609) 258-6394; Cell (908) 922-2236 Email: nyao@princeton.edu John Schreiber: Office (609) 258-0034; Cell (215) 431-4670 Email: js51@princeton.edu Paul Shao: Office (609) 258-3851; Cell (847) 721-086 Email: pshao@princeton.edu Daniel Gregory: Office (609) 258-4384; Cell (302) 542-3182 Email: daniel.gregory@princeton.edu Yao-Wen Yeh: Office (609) 258-7956; Cell (848) 248-8058 Email: yyeh@princeton.edu 13