VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control

Similar documents
CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

FUNDAMENTALS OF GAS CHROMATOGRAPHY, GAS QUALITY & TROUBLESHOOTING Larry Ewing Chandler Engineering Company LLC

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Chapter 13 Gases, Vapors, Liquids, and Solids

Technical Bulletin, Communicating with Gas Chromatographs

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

GAS CHROMATOGRAPHY MAINTENANCE USING UNCERTAINTY BASED CBM

TECHNIQUES OF COMPOSITE SAMPLING. Kris Kimmel. YZ Systems, Inc Pollok Drive Conroe, Texas 77303

ISO INTERNATIONAL STANDARD

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Improved Reliability in Natural Gas Energy Measurement with the Revised ISO Standard. Gerard Nieuwenkamp - VSL

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

GAS CHROMATOGRAPHY AND SAMPLE LET DOWN SYSTEM CONDITIONAL BASED MONITORING AND LIVE UNCERTAINTY CALCULATION

Instrument Performance Evaluation, comparing ISO and NORSOK I-106 Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk Analyse North Sea Flow

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

1 PIPESYS Application

Research and Development. Money in the Pipeline

Multiple Gas#5 GC configuration Jan 2016

Application Worksheet

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Figure Vapor-liquid equilibrium for a binary mixture. The dashed lines show the equilibrium compositions.

INTEROFFICE MEMORANDUM MODIFICATIONS TO DEW POINT CONTROL PROCESS

Agilent 490 Micro GC Natural Gas Analyzer

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

Protective Atmospheres, Measurement Technologies and Troubleshooting Tools

Modulating Valves for Atmospheric, Infrared, and Direct Fired Burners

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

The Application of Temperature and/or Pressure Correction Factors in Gas Measurement

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

* William A. Sala, Jr. John A. Roscher. Columbia Gas Transmission, LLC * Sorana Linder 700 Louisiana Street, Suite 700 Director, Regulated Services

Calibration and Switching Module

PURE SUBSTANCE. Nitrogen and gaseous air are pure substances.

Micro Motion 3098 Gas Specific Gravity Meter

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter

Fundamentals of NGL Sampling Systems. By Dominic Giametta & Jim Klentzman

Best Practice Guide, Servomex 2700

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

Instruction Manual No. 742, 8/98

Example: Calculate the density of methane at 50 psig and 32 ⁰F. H.W. In previous example calculate the density of methane in gm/m 3.

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

ENSURING AN ACCURATE RESULT IN AN ANALYTICAL INSTRUMENTATION SYSTEM PART 1: UNDERSTANDING AND MEASURING TIME DELAY

Model Series 62 Constant Differential Relay

HumiSys HF High Flow RH Generator

Instruction Manual 742 5/1/2009. Eclipse Ratio Regulators ES Series Version 1

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Optimization of Separator Train in Oil Industry

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Name Chemistry Pre-AP

Bernhard Thomas September EnCal 3000 prochain Energy in Gas with ROI

North Sea Flow Measurement Workshop th October 2017

Chapter 22 Testing Protocols

Applied Technology and Best Practices in CEE. Conference

COLLECTING AND HANDLING OF NATURAL GAS SAMPLES FOR CUSTODY TRANSFER. Christopher L. Grant, Dr. Darin L. George, and Jacob L.

OVERVIEW. Flow Coefficient C v. Operating Conditions. Specific Gravity

4.3 Objectionable Properties. The gas received and delivered hereunder:

Applications Note: Use of "pentane equivalent" calibration gas mixtures

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

SG TF. Q = 963 Cv (P1 - P2)(P1 + P2) TECHNICAL INFORMATION. Flow Q (GPM)

North American Orifice Metering System

The Discussion of this exercise covers the following points:

Oxygen Generating Systems Ozone Medical Welding Glassblowing Aquaculture Pulp & Paper Wastewater

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Gas Leak Measurement of TRT Barrel Module

MMS Mercury Monitoring System for Natural Gas

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

H o w t o U s e a R e g u l a t o r t o R e d u c e T i m e D e l a y i n a n A n a l y t i c a l S y s t e m

LESSONS LEARNED FROM THE API MPMS, CHAPTER 14.1 GAS SAMPLING RESEARCH PROJECT

This test shall be carried out on all vehicles equipped with open type traction batteries.

Gas Billing Factors Guideline

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

A NEW PROCESS FOR IMPROVED LIQUEFACTION EFFICIENCY

EnCal 3000 Gas Chromatograph Commissioning and Maintenance Manual Biogas and THT-Devices

Laser Spectrometers for Online Moisture Measurement in Natural Gas. Ken Soleyn GE M&C

Flow meter. bellow vaporizer. APL valve. Scavenging system

Gas Measurement Fundamentals Certification. Curriculum

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

Vehicle-mounted meters, pump supplied

Generating Calibration Gas Standards

FUNDAMENTALS OF ENERGY DETERMINATION

Vehicle- or rack-mounted liquefied gas meters, pump supplied

5890II GC Standard Operating Procedure 9/2/2005

Oil Calculator Pro User Manual 2014

Test Report # Rev 0. Adiabatic Compression With Constant Bleed Valve

Installation, Operation, and Maintenance Manual

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Inerting System Design for Medium Speed Vertical Spindle Coal Pulverizers TABLE OF CONTENTS

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Transcription:

VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control Houston, Texas ABSTRACT The on-line gas chromatograph (GC) has been widely used for natural gas quality analysis and energy measurement at custody transfer locations since the early 80 s. The energy measurement and relative density measurement provided by the GC can have a large effect on unaccounted for Gas because the gravity effects volume calculations and the heating value effects energy rates. The gas chromatograph measures component concentrations first and then calculates physical properties of the gas such as heating value and relative density. Units of energy measurement can be British Thermal Units () Mega Joules (MJ) or Kilo Calories (KJ). This paper will refer to for energy units, ISO (KJ) or SI (MJ) metric units can be used interchangeably. INTRODUCTION The Gas Chromatograph (GC) can be a fairly complex instrument that requires a certain degree of specialization in order to maintain optimal measurement. Optimizing GC performance depends upon a thorough understanding of the chromatogram and its role in setting up a GC. The following paper will discuss the various operational components of a GC that need to be set up, in order of importance, to verify optimal measurement in a typical natural gas application. PRINCIPLE OF OPERATION The GC oven contains 3 main components mounted together in an electrically heated heat-sink-oven: GC Valves Micro-packed Columns TCD Detectors The GC valves control sample injection and column switching. Components are separated into groups and diverted into specific columns for further separation. This creates a critical split between certain key components. Splits between Hexanes + (C 6+ ) and Normal Pentane (NC 5 ) are controlled by V-2 off time. Splits between Ethane (C 2 ) and Propane (C 3 ) are controlled by V-3 off time. The component separations are as follows in Table 1. The valve timing, inhibit on and off times, and other timed events must be properly configured to ensure that component groups are diverted to the proper columns and all relevant peaks are properly integrated. Chromatograms must be used in conjunction with raw data and analysis data to ensure that all timed events are optimized. Column Configuration Column 1 (controlled by valve 2 off-time) Column 2 (controlled by valve 3 off-time) Column 3 (controlled by valve 3 on-time) Components Separated C 6+ C 3, IC 4, NC 4, NeoC 5, IC 5, NC 5 N, C 1,CO 2, C 2 TABLE 1. SAMPLE CONDITIONING AND SAMPLE TRANSPORT The sample conditioning system is a very important component in any analytical system. This is a frequently overlooked component that must deliver a representative sample from the pipeline to the GC without changing the composition. This is not a difficult task when low gas is measured in warm climates. When high gas is measured the potential exists to lose the heavier components as they drop into the liquid phase due to sample cooling. This cooling effect, commonly known as the Joules-Thompson (JT) effect, will bias energy readings low because the high contributors drop out of the vapor-phase sample before the GC even has a chance to measure it. The JT effect occurs at the point where the greatest pressure drop occurs. Usually this is the point where line pressure is reduced to 20-50 PSIG for sample transport to the GC. In addition, ambient temperatures contribute to cooling in the sample transport lines if left unheated in winter conditions. If the JT effect causes a loss of some of these high contributors energy measurement will be biased low as a result. Site inspections should take into account the need for heating on sample transport lines in cooler climates. Potential exists for errors biased in the negative direction when sample pressure is reduced by more than 700 PSIG on 1100 Gas or higher. The Sample Transport System must include sample probes to ensure that a representative sample is collected from within the flowing stream rather than from the walls of the pipeline. Sample probes provide a dual function: 1-They minimize liquid-contaminants from entering the sample transport line from the pipeline walls. 2-They extract a representative sample from the middle of the flowing stream for the sample transport system. Integral sample probe/regulators can be useful for high samples because they reduce the pressure at the 2002 PROCEEDINGS PAGE 107

sample tip where higher gas temperatures can help minimize the JT effect. By visually inspecting the sample transport system (including sample probes) and noting maximum pressure drops relative to the energy content of the gas an engineer/technician can get a feel for the potential for Joules-Thompson effect. The Stream Switching System can be double block and bleed, solenoid purge or single block with no bleed. In in custody transfer applications the latter is more common (see Figure 1). In this case each stream should be pressure balanced to minimize changes in sample loop volume due to pressure changes on differing steams. By halting GC operation and manually actuating each stream the engineer/technician can verify proper setup (pressure balancing) on the sample system. For example: Stream 1 pressure should be set at the pressure regulator (usually mounted at the sample probe) to indicate the same sample flow at F1 as streams 2,3 and the calibration stream (Cal Std). All stream flow rates can be adjusted in this manner with pressure using the common sample roto-meter to balance flow. Sample flow rate is not critical, so any point roughly mid scale (0-100cc/min) is adequate. When streams are pressure balanced the amount of normalization required during the analysis will be minimized. Sample filters should also be checked and replaced at least once per year. INSTRUMENT PERFORMANCE Linearity Linearity can be verified with the use of three known standards. The on-site calibration standard may suffice as one of the three standards, preferably the middle one. Each standard should have component concentrations that yield a value over the full operational range for a given installation. A meter station that may have swings from 1020 on stream 1 to 1250 on stream 3 should have a component mix roughly as follows: Component Std Heating Std Heating Std Heating #1 Value #2 Value #3 Value Hexanes + 0.01 0.53 0.03 1.59 0.2 10.58 Propane 1 25.16 3 75.48 7 176.13 Iso-Butane 0.5 16.26 0.1 3.25 0.4 13.01 N-Butane 0.5 16.31 0.1 3.26 0.4 13.05 Neo-Pentane 0.08 3.19 0 0.00 0 0.00 Iso-Pentane 0.08 3.20 0.3 12.00 0.8 32.01 N-Pentane 0.1 4.01 0.3 12.03 0.8 32.07 Nitrogen 3.31 0.00 2 0.00 1 0.00 Methane 89.5 904.15 87.67 885.47 77.4 781.74 Carbon Dioxide 2.9 0.00 1.5 0.00 1 0.00 Ethane 2 35.39 5 88.48 11 194.66 Total 100 1008.20 100 1081.56 100 1253.24 TABLE 2. REPEATABILITY Several consecutive runs with on-site calibration gas will yield results that will determine the analyzer repeatability. As a general rule the first run should be disregarded. The next 5 or 10 consecutive runs should yield results that are within at least ±.5, ±.25 is not uncommon. The same repeatability test over a 24-hour period in a temperature-controlled environment should yield no worse than ±.5. If the installation is in an uncontrolled environment a test over 24 hours should yield results no worse than ±1. Once repeatability has been established careful review of historical calibration reports can reveal a lot about the instrument performance. GPA report 2286-95 serves as a useful guideline for mole % repeatability standards. Accuracy FIGURE 1. Repeatability is necessary in order to obtain accuracy. Once repeatability has been determined instrument calibration Component Mol % Repeatability Reproducibility Range Nitrogen 1.0-7.7 2 7 Carbon Dioxide 0.14-7.9 3 12 Methane 71.6-86.4 2.7 Ethane 4.9-9.3 1 2 Propane 2.3-4.3 1 2 Iso-Butane.26-1.0 2 4 N-Butane.6-1.9 2 4 Iso-Pentane.12-.45 3 6 N-Pentane.14-.42 3 6 Hexanes+.1-.35 10 30 Note: 1-Heating Values are based on GPA 98 Engineering Data Book Volume II Section 23. 2- based on 60F and 14.696 base conditions. TABLE 3. Note: GPA-2186-95 Section 9 Precision PAGE 108 2002 PROCEEDINGS

will improve accuracy. The calibration updates retention times and response factors for each component. New response factors can deviate from the old response factors by up to 10%. Response factor (RF) deviations in excess of roughly 5% indicate that performance may be deteriorating slightly during on-line operation. There is a very important relationship between components that are on either side of a column split. If Response Factor (RF) deviations in excess of 2% are over time over time it is an indication that something may not be right. The critical component splits on a 4 or 6 minute Danalyzer are C 2 /C 3 and IC 5 /C 6+. A problem can be revealed by careful attention to the RF deviations for each pair. For instance, high RF deviations for NC 5 and C 6+ may indicate a problem with V2 off time (see Table 1). If there is a problem of this nature instrument linearity will be adversely effected. A single point repeatability check will not indicate a problem. In addition, calibration will do nothing to correct the problem. As time progresses these types of set up problems will show up as RF Deviation alarms. These are factory set to default values of 10% RF deviation for any component. PHYSICAL PROPERTY CALCULATIONS HEATING VALUES FOR IDEAL GASES USING GPA 2145 1996 VS. 2000 FOR IDEAL GASES Once the preceding GC components have been optimized there are some minor configuration parameters that can be checked to ensure that contractual obligations are in compliance. Every few years the GPA publishes new values for physical properties of select hydrocarbons. At the bottom of the table is a statement of accuracy that reads: Numbers in this table do not have accuracy s greater than 1 part in 1000. In some cases, extra digits have been added to calculated values to achieve internal consistency or to permit recalculation of experimental values. The largest % change made to any value in the 2000 table is for ethane and it is changed by 1 part in 17,696. This is a change that is more than 17 times less than the stated accuracy. The stated accuracy has not changed with the 2000 data. The net effect on component concentrations is minimal, is well below the repeatability of all TCD based field GCs. In addition, the changes have an even smaller effect on relative density calculations. These updated values have very little effect on the accuracy of the heating value calculation in natural gas when measured on a GC. The 2000 GPA data included very minor constituent changes in natural gas, the overall change is negligible. The difference in physical property measurement can be seen in table 4 using two identical natural gas streams with the 96 data and the 2000 data. The GPA 2000 data is currently under review. Typical GPA 2145-96 GPA 2145-2000 Natural Gas Rich Gas 1300 +0.0138 Typical US 1078 +0.0032 Gulf Coast TABLE 4. Most GCs operate with a stated repeatability of no better than 0.1 at ideal conditions. 1 is more common. There is questionable value in updating ideal gas heating values that are 10-100 times less than the repeatability of the instrument that is using the numbers. PRESSURE BASE FOR CALCULATIONS Pressure base, on the other hand, can have a large effect on the calculation. Care must be taken to ensure that the pressure base for calculations is the same as the pressure base used in ancillary computers that rely on the GC data for volume and flowing energy calculations. Typical Composition Pressure Pressure Pressure Natural Base @ Base @ Base @ Gas 14.73 14.696 15.023 Nitrogen 2.46 Methane 89.709 CO 2 1.001 Ethane 5.001 Propane.999 Iso-Butane.300 N-Butane.300 Iso-Pentane.100 N-Pentane.100 C 6+.03 Relative.6222.6222.6223 Density Gross Dry 1053.70 1051.26 1074.73 Gross Sat 1035.36 1032.97 1056.03 Net Dry 951.21 949.02 970.2 Net Sat 934.66 932.5 953.32 Note: Base temperature @ 60 F SUMMARY TABLE 5. To verify GC operation, historical calibration reports and configuration reports are necessary to ensure that all configuration parameters and calibration parameters are correct and within specifications. Three certified standards should be used to check repeatability, linearity and accuracy throughout the measured heating value range for all streams. Sample conditioning and sample transport systems should be carefully considered for possible JT effect on rich gas samples. 2002 PROCEEDINGS PAGE 109

APPENDIX A General description a typical audit procedure using two certified standards: A second cylinder of certified calibration gas can be used. Temporary tubing to the chromatograph should be installed along with a bleed vent to allow for purging. Gas will be introduced to the unit and the resulting As-found analysis reports to be saved. The unit will be re-calibrated and the second standard can be introduced to record As-left measurements. Specifications for the second standard as follows: 1) Hydrocarbon dew point: 20 degrees F. or lower. 2) Cylinder contents to include all components that are measured by the on-line gas chromatograph. 3) The component concentrations will be higher than those in the station calibration gas, except methane which must be the balancing component. Handling the cylinder 4) The cylinder will be heated for hours if/when exposed to temperature below the hydrocarbon dew point degrees F. 5) The cylinder must not be used below 25 p.s.i.g. 6) The cylinder regulator must be equipped with the inlet CGA fitting that corresponds to the gender and the left hand thread direction. Suggested guidelines for sample system tubing and fittings: 1) Maximum 1 /8 stainless steel tubing must be used. 2) Tubing must be clean (free of any/all contaminants) and dry. 3) Temporarily install a manual three-way valve to select the calibration gas or the referee without allowing air to enter the system. Suggested purging procedure: 1) Install the regulator and open the cylinder block valve and purge the air from the system. 2) While the gas is slightly purging, tighten the CGA fitting into the regulator block valve. 3) Connect the delivery tubing from the cylinder regulator to the three-way manual valve. 4) Attach the second standard as seen in Figure 2. 5) Use the purge valves to eliminate air. 6) Adjust the pressure regulator to deliver the same pressure as that used for the existing calibration cylinder (see pressure balance procedure on page 2). 7) Set the temporary three-way valve to allow the existing calibration gas to flow to the sample flow-meter. Testing the chromatograph: 1) Idle the chromatograph. 2) Using the keypad or the personal computer, issue the proper command to run the calibration gas as an unknown sample. Do not calibrate the GC yet. Verifying repeatability to obtain the As-found results on the station calibration gas: 3) Observe the results and pay particular attention to the measured results for nitrogen. As soon as the nitrogen measurements are repeated for at least three consecutive runs, then reports are to be saved as As-found documentation. 4) Idle the GC. Preparing to run the second standard: 5) Using the three-way selector valve, select the second cylinder. Make sure that the flow rate is the same for the referee cylinder as for the existing calibration cylinder. 6) Using the keypad or the personal computer, issue the proper command to run the referee gas as an unknown. Do not calibrate the GC yet. Verifying repeatability to obtain the As-found results on the referee gas: 7) Observe the results and once again verify that the referee gas is repeating the nitrogen measurement. Once repeatability is established, save the reports as As-found documentation. 8) Switch the selector valve back to the existing calibration gas. 9) Idle the chromatograph. 10) Run the unit to establish repeatability checking the nitrogen measurement as seen above. Calibrating the GC: 11) Calibrate the unit now. Obtaining as-left results: 12) Check the repeatability of the GC on the existing station calibration gas and save the reports as As-left documentation. 13) Idle the chromatograph. 14) Switch the selector to the referee cylinder. 15) Issue the command to run the referee gas as an unknown and verify repeatability as above. 16) Save the reports of the referee as As-left documentation. 17) Issue the proper command to return the GC to automatic service. 18) Remove the temporary tubing and reinstall the delivery tubing from the existing station calibration cylinder. 19) Purge the calibration gas delivery line of air. APPENDIX B Base conditions The conditions of temperature and pressure to which measured volumes are to be corrected. In the case of energy, the base conditions affect the volume of gas in a cubic foot that contained the measured volume for a determination. (British Thermal Unit) a measure of the quantity of heat contained within 1 cubic foot of volume. The classic definition is that one will raise the temperature of water one degree Fahrenheit at or near 60 F (15 C). The AGA has defined a IT as equal to PAGE 110 2002 PROCEEDINGS

Purge Cal To GC net The term net quantity subtracts the heat required to vaporize the water and represents actual useful heat energy. Gross is commonly used in the United States for custody transfer; net in Europe. During combustion water is formed. If that water remains in the ideal gas state the heating value is termed net. Once net is determined gross can then be calculated on a dry and wet basis. Other terms sometimes used for net and gross are: inferior(net) and superior (gross) or lower and higher heat value. Existing Cal REF Existing On Line GC Sample System Referee Cylinder - Details. Related cylinder, valves and tubing are shown bold. Tubing size 1/8" typical. 1055.056 Joule. Either definition will yield similar heat values, but it is important to know which basis is being used to state values. Normal pipeline gas has about 1023 s heat value per 1000 SCF. A parallel metric unit is gram-calorie, the amount of heat needed to raise 1 gram of water 1 C. Another, and more commonly used, is Megajoules with gas heat typically stated in MJ/1000 m 3 (volume) or MJ/Kg (mass). actual The calculation of actual is based on the field determination of water content of natural gas and yields a value that is between wet and dry values. It is converted from dry as follows: Actual = dry @pb * (1- H20/1,000,000* (21.0649 FT3/LB *14.696 pb) ). dry Heating value can be expressed as dry, common pipeline practice requires 7 pounds or less of water vapor per MMSCF to be considered dry. saturated The term saturated, also commonly referred to as wet assumes the gas has all the water vapor it can hold (at 60 F and the indicated pressure base) before the water starts to condense and form droplets. Wet and dry refer to gas before combustion. Gross The amount of energy transfered as heat per mass or heat per mole from the complete, ideal combustion of the gas with oxygen (from air) at base temperature in which all water (from combustion) condenses to liquid. If the heating value has a volumetric basis rather than mass a base pressure must also be specified. Combustion creates water as part of the process. It is this difference between the vapor phase water content after combustion that gives rise to the terms net and gross. A gross quantity represents total heat energy in a quantity of gas a very small portion of which is not actually contributing to useful heat. A Daniel chromatograph measures on a gross basis as combustion does not occur. Once gross is determined, net can then be calculated on a dry and wet basis. Saturated gas for the purpose of maximum lawful price calculations under the NGPA, the gas tested contains the maximum amount of water vapor possible under conditions of 14.73 psia and 60 F. The total quantity of heat is determined by multiplying a volume of gas at a specified pressure and temperature by the heating value per unit volume saturated with water at the same pressure and temperature base. No further corrections for the presence of water shall be made. The content of one cubic foot of natural gas at defined conditions is the number of s produced by the complete combustion of such cubic foot of gas at constant pressure with air of the same temperature and pressure as the gas, when the products of combustion are cooled to the initial temperature of the gas and air and when water formed by such combustion is condensed to a liquid state at the initial temperature of the gas. critical temperature and pressure temperature above which a fluid cannot exist as a liquid; vapor pressure of a fluid at the critical temperature. These values are used to calculate compressibility. Dekatherm 1 dekatherm = 1 MM (1 million s) or 1000 SCF x 1000. density, relative (gas) The ratio of a specific weight of a gas to the specific weight of air at the same base conditions. density, relative (liquid) The relative density of a liquid is the ratio of the fluid at a temperature to the density of pure water at a specified base temperature. (Also known as specific gravity) dew point temperature at which water vapor will begin to condense. A typical pipeline spec for maximum water vapor content is 7 lbs/mmscf. At 1000 psia, the dew point is 32 F. At 100 psia, the dew point drops to 15 F (and water dron uts wold form with possible line freezing). mole % the amount of a substance expressed as the percentage of its molecular volume compared to the total of molecular volume for all substances in a sample. The relative percentage of a component in a mixture of components in the sample stream is determined by comparing the peak area to the sum of peak areas for all components. Mole % = peak area / response factor. 2002 PROCEEDINGS PAGE 111

response factor the factor that defines how much detector output will be generated for a given amount of a sample component. (RF equals peak area for a known component in the cal gas divided by the known amount of the component.) retention time the time measured between the start of analysis to the top of a peak. specific gravity (SG) (see relative density) The ratio of the density of a substance to the density of a standard (normally water for liquids and air for gasses). SG is the reciprocal of relative density. Density is mass per unit volume of a fluid, but typically expressed as weight per unit volume lb/gal, lb/bbl, grams/cubic cm, etc.). water vapor defined as pounds of water per MMSCF in the U.S. As pressure increases, water vapor is compressed into its liquid state. For example, at 60 F, a natural gas stream holding 17.9 lbs/mmscf of water at 1000 psia will hold 128 lbs/mmscf at 100 psia. Lowering the temperature from 60 to 40 F at 100 psia will reduce water content to 61 lbs/mmscf. CONVERSION FACTORS = 252.2 calories = 0.252 kilocalories = 1055 Joule 1000 /SCF = 37.27 megaj/m 3 Gallon (Imperial) = 1.2 Gallon (US) ft 3 /sec = 0.000589 liters/min F = 9/5( C) + 32 C = 5/9( F -32) For the NGPA Pricing Method... REFERENCES GPA Engineering Data Book Volume 11 section 23 GPA Report 2145 & 2286-95 ACKNOWLEDGMENTS Charlie Cook, Daniel Measurement & Control, Appendix A Wobbe Index An index, used to determine the size of a gas burner tip, calculated by dividing total saturated by the square root of specific gravity. Murray Fraser PAGE 112 2002 PROCEEDINGS