extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Similar documents
Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Experiment GC : Analysis of BTEX by GC-FID

Extended Application Note

Rapid and Reliable Detection of Dissolved Gases in Water

no peaks missing peaks too small peaks increasing retention decreasing retention

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

GCMSD-Headspace Analysis SOP

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Chemical Analysis for Methyltestosterone

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Retention Time Locking: Concepts and Applications. Application

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Automated Determination of Dissolved Gases in Water Anne Jurek

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Determination of Oxygenates in Gasoline by Gas Chromatography and Oxygen Selective Flame Ionization Detector Method

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

CORESTA RECOMMENDED METHOD Nº 67

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

APPLICATION NOTE. GC Integrated Permeation Device

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

ISO INTERNATIONAL STANDARD

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

RayBio Human vwf ELISA Kit

How to Achieve Optimal Weighing Performance

Proficiency Testing Corrective Action Checklist: PT PROVIDER: PT EVENT: TEST:

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates

AKTA pure 25 New Owner s Intro

LEVANT WORMSEED FOR HOMOEOPATHIC PREPARATIONS CINA FOR HOMOEOPATHIC PREPARATIONS

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

FROG-4000 TM Chemical Analysis System

Video Notes. LC Troubleshooting Series Retention Time Shifts

Dilution Sequence Protocol

Improved Performance in Capillary Electrophoresis using Internal Standards

MV5: Automated concentration system

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

Multiple Gas#5 GC configuration Jan 2016

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

May 1, By John Heim,Doug Staples

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Setting a New Standard in Flash Chromatography Performance

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

Dilution Sequence Protocol

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

Dilution Sequence Protocol

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

<788> PARTICULATE MATTER IN INJECTIONS

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

VOLUMETRIC TECHNIQUES

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

General Accreditation Guidance. User checks and maintenance of laboratory balances

CHM Introductory Laboratory Experiment (r17sd) 1/13

ISO INTERNATIONAL STANDARD

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

STANDARD OPERATING PROCEDURES

CHM250 Calibration and Measurement Lab. Balance Calibration

CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 40 Ultra-Sensitive Kit

HUMAN IL6 KITS PROTOCOL

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Agilent Dimension Software for ELSD User Manual

M/s. Thermo fisher Scientific

Laboratory Activity Measurement and Density. Average deviation = Sum of absolute values of all deviations Number of trials

Gradients in ion chromatography. Optimal separation with shorter analysis time

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

BASIC LABORATORY TECHNIQUES (Revised )

Setting up and running a column

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit

LCMSMS Solutions For Steroid Analysis

Transcription:

Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s): Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution/Identification C: Limit of EG and DEG Expert Committee(s): Excipient Monographs 1 No. of Commenters: 10 Comment Summary #1: The commenter tested the proposed EG/DEG limit test for Sorbitol Solution and did not see peaks in their Standard solution for EG and DEG (EG and DEG, 0.08 mg/ml each). Response: Through communications with USP, the commenter identified and resolved what turned out to be an instrument problem (column related). Comment Summary #2: The commenter questioned why a 96:4 Acetone:DI Water diluent was being used. Sorbitol is very soluble in DI Water and the commenter could not get the sorbitol to go into solution with the acetone diluent. Response: The Expert Committee recommended adding a clarification note to state, [NOTE Acetone is used to precipitate sorbitol] in the Sample solution preparation section of the proposed method. This statement was incorporated because the Expert Committee determined that during a USP Web Meeting held on 16-17 March 2009, discussions on previous USP laboratory experiments conducted during the method development stage indicated that sorbitol solution samples prepared in water were unsuitable for the GC method, and as a result, requires a water- and sorbitol-free sample preparation. In order to prepare water- and sorbitol-free test solutions that would be suitable for the GC analysis, precipitation of sorbitol with water miscible organic solvents was explored. Sorbitol solution samples prepared by precipitation with Acetone:DI Water mixture (96:4, v/v) yielding adequate EG and DEG recoveries gave the desired results on the GC columns. It produced: No splitting or fronting of EG or DEG peaks was observed. Late-eluting sorbitol related peaks were greatly minimized No carry-over was observed. Comment Summary #3: The commenter requested USP s detailed sample preparation for Sorbitol Solution and questioned why there is a need to dilute and vortex as described in the sample procedure. Response: The Expert Committee determined that both EG and DEG are highly soluble in water and acetone, while sorbitol is only soluble in water. The detailed sample procedure involving dilution with diluent (Acetone:DI water 96:4, v/v) in three portions and vortexing after each addition was designed to achieve close to 100% recoveries of both analytes from the water-based Sorbitol Solution by keeping a balance between two processes occurring simultaneously, resulting in the precipitation of sorbitol and Page 1

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Comment Summary #4: The commenter suggested that the System suitability requirement of NLT 30 between EG and DEG in the Sorbitol method seems excessive. Response: The Expert Committee determined that the Propylene Glycol peak is quite often observed in the Sample solution after Ethylene Glycol peak and before Diethylene Glycol peak, therefore, the Resolution between Ethylene Glycol and Diethylene Glycol must ensure separation and agreed that this resolution is easily achieved under the stated conditions. Comment Summary #5: The commenter questioned the resolutions the USP laboratory is routinely getting in sorbitol since the retention times are 1.2 and 3.1 Response: Based on the laboratory data, the Expert Committee determined that the Resolution is about 55. Also, the Expert Committee determined that this is the combined Resolution of Diluent peak (14) and DEG peak (41). Comment Summary #6: The commenter indicated the requirement for holding a 25-mL volumetric flask on a vortexer for 12 minutes seemed excessive. Response: Based on the laboratory data, the Expert Committee determined that due to volume expansion during an initial trial in the lab, sonication may not be a viable option as it is difficult to achieve accurate dilution volume by addition of diluents in three aliquots. Also, poor recoveries of EG and DEG were observed. The Expert Committee determined that another option could be by shaking and mixing well by available alternative or equivalent methods. Comment Summary #7: The commenter suggested specifying the phase for the sample extract to avoid confusion because two phases are present and the analyst only should filter the top phase. Response: Comment incorporated. The Expert Committee recommended replacing the phrase Filter a portion of the test solution so obtained with, Pass a portion of the supernatant layer obtained. Comment Summary #8: The commenter observed the appearance of coating (sorbitol?) on the walls of a glass storage vial after filtration and questioned: Does this indicate timing is important? Is adequate precipitation of sorbitol time and temperature dependent? Requested that perhaps more detail on timing of extract preparation is warranted. Response: Based on the laboratory data, the Expert Committee determined that the sample preparation including extraction has been optimized to achieve spike recovery greater than 90%. Most importantly, any timing for extract preparation or storage was not identified in the method development and validation studies because sorbitol was not the analyte of interest. The USP laboratory did observe some precipitation of sorbitol in vials stored in a refrigerator. Page 2

Comment Summary #9: The commenter indicated that working with such small sample volumes (i.e., a 1 μl injection) a 10:1 split magnifies any mechanical difficulty with the equipment. A lesser split (perhaps 5:1) would allow greater repeatability, while retaining the sensitivity the assay is intended to achieve. Response: Comment not incorporated. The Expert Committee determined that a 10:1 split is commonly used in GC separations. During the course of USP s laboratory method development and validation experiments, no issues were encountered with regard to repeatability or sensitivity warranting the use of a lesser split. Comment Summary #10: The commenter indicated that the column has a maximum programmable temperature of 280, while the method calls for a final oven temperature of 300. The temperature caused no problem in this initial study, it may however, reduce the life of the column with continuous use. Response: The Expert Committee determined that according to the manufacturer s specification, the column is programmable up to 300 for a gradient method. Addition of a temperature hold for 5 minutes at 300 did not lead to any deterioration in the column performance during the course of extensive method development and validation studies. Comment Summary #11: The commenter indicated that the USP procedure describes the Standard solution as containing 0.08 mg/ml EG and 0.08 mg/ml of DEG in diluent but it does not include a protocol for preparing the Standard solution. The commenter asked that USP provide instructions for the GC Standard Solution preparation Response: The Expert Committee determined that according to the redesigned USP monograph format, if there is no special preparation involved in Standard solution and in Sample solution, the detailed procedures are not specified. USP General Notices states When a specified concentration is called for in a procedure, a solution of other normality or molarity may be used provided that allowance is made for the difference in concentration and that the change does not increase the error of measurement. Unless otherwise indicated, analyte concentrations shall be prepared to within ten percent (10%) of the indicated value. In the special case in which a procedure is adapted to the working range of an instrument, solution concentrations may differ from the indicated value by more than ten percent (10%), with appropriate changes in associated calculations. Any changes shall fall within the validated range of the instrument. Comment Summary #12: The commenter indicated they had questions on the stability of the GC Standard solution and whether or not the Standard solution needs to be prepared daily or weekly and whether, if prepared weekly, it should be refrigerated. Response: Based on the laboratory data, the Expert Committee determined that during the method validation standard solutions were prepared fresh and used immediately. Comment Summary #13: The commenter indicated they had questions on the stability of the prepared samples and whether or not they can be stored overnight or should be analyzed within a certain time period. Response: Based on the laboratory data, the Expert Committee determined that during the method validation sample solutions were prepared fresh and used immediately. Page 3

Comment Summary #14: The commenter suggested a GC method similar to the GC method proposed in the USP drafts. The commenter stated that two major differences were observed between the USP proposed method and their method: 1. The USP-proposed GC method does not include a derivatization step. As polyols are not volatile compounds, the derivatization step helps to obtain volatile compounds for the GC analysis. This also helps avoid clogging of the column. 2. The method developed by the commenter applies a derivatization step. Also, the GC method proposed by USP uses external standards. The GC method submitted by the commenter uses internal standard. Usually, external standards require work with accurate volumes and weights. This leads frequently to human and systematic errors. The use of internal standards decrease the risk of errors during the preparation of samples and solutions and the risk of deviation linked to the sampling. Response: Comment not incorporated. The Expert Committee determined that the method submitted by the commenter is only validated for Ethylene Glycol. There is no data supporting its use for Diethylene Glycol. Comment Summary #15: The commenter indicated that in the draft monographs, the relative retention times are listed for Ethylene Glycol and Diethylene Glycol as 1.0 and 2.6 respectively, but one laboratory obtained slightly different relative retention times of 1.0 and 1.8 respectively. The commenter noticed that General Chapter <621> Chromatography states: Relative retention times may be provided in monographs for information purposes only, to aid in peak identification. There are no acceptance criteria applied to relative retention times. The commenter mentioned that although the above statement indicates that relative retention times are informational, they believe that relative retention times may be interpreted as a requirement in this particular case since there is such high visibility for this particular identification test. Therefore, the commenter requested either that relative retention times be adjusted according to data provided to USP, or that relative retention times be removed from the monograph since they are only informational. Response: Comment incorporated. The Expert Committee recommended to revise the Note to read as [NOTE Diethylene Glycol elutes after Ethylene Glycol in the chromatogram.] Comment Summary #16: The commenter suggested the following changes to paragraph 1. Bold text reflects commenter recommendations. First sentence should read: Because of the serious hazards associated with the use of Diethylene Glycol and Ethylene Glycol-contaminated materials With regards to last sentence: the testing of USP Sorbitol Solution should demonstrate the absence at NMT 0.10% o There should be some consideration of daily intake Response: The Expert Committee determined that the previously posted BRIEFING will not appear as part of the Revision Bulletin. The spec limits for Diethylene Glycol and Ethylene Glycol were recommended by the Food and Drug Administration (FDA), please refer to FDA letter received by USP on 12-Jan-2009. Page 4

http://www.usp.org/pdf/en/hottopics/2009-01-12- FDALetterPropyleneGlycolSorbitalSolution.pdf Comment Summary #17: The commenter suggested the following changes to paragraph 2.).. Second sentence should read: The Gas Chromatography procedure for the limit of Diethylene Glycol and Ethylene Glycol is based on analysis performed with the J & W Scientific DB-1701 brand of G46 column Response: The Expert Committee determined that the previously posted BRIEFING will not appear as part of the Revision Bulletin. Comment Summary #18: The commenter provided the following comments and suggested the following two changes to paragraph 3.Bold text reflects commenter recommendations. 1. Standard solution: The concentration of Standard solutions may be too low for detection with older equipment 2. Sample Solution should read: a) Transfer 2.0 g of Sorbitol Solution to a 25ml volumetric flask. Add 1.0ml of diluent to the flask and vortex the flask for 3 min. Add the remaining Diluent to the flask to volume in 3 equal portions and bring up solution to a final volume of 25ml. Vortex the flask for about 3 minutes after each addition of Diluent. Filter a portion of the extract so obtained through a 0.45-um nylon filter. b) Commenter questioned if a volume should be assigned to the portion and asks if 10mL would be appropriate. Response: The Expert Committee determined for comment 1, that validation studies and public reports indicated that the concentration of Standard solution in current proposed methods for Sorbitol Solutions is appropriate for detection with any basic GC instrument via direct injection. However, USP General Notices also states that When a specified concentration is called for in a procedure, a solution of other normality or molarity may be used, provided that allowance is made for the difference in concentration and that the change does not increase the error of measurement. USP hosted an open microphone Web Meeting on 10 August 2009 to collect public comments on USP proposed methods. After that meeting, USP posted all the presentation slides made during the web meeting on USP website under Hot Topics for Propylene Glycol and Sorbitol Solutions. Laboratory validation data and important chromatograms are all included in these presentation slides. For comment 2a), the Expert Committee recommended the following additional changes to the Sample solution for clarification purposes by using the phrase Pass a portion of the supernatant layer obtained to replace Filter a portion of the test solution so obtained, and by adding about as suggested by the commenter. The committee determined that the suggestion of using and bring up solution to a final volume of 25 ml is not necessary because in the USP monograph format, a 25- ml volumetric flask is used in the experiment, Add the remaining Diluent to the flask to volume should be equivalent to Add the remaining Diluent to the flask to 25 ml Page 5

For comment 2b), the Expert Committee determined that it is unnecessary to provide a defined volume for the portion as long as there is an adequate quantity of filtrate to be used in the GC analysis. Comment Summary #19: The commenter suggested the following changes on Chromatographic System. Bold text reflects commenter recommendations. Injection Type: Split injection; Split Ratio is about 10:1 Response: Comment incorporated. The Expert Committee changed the current text: Injection mode: Split 10:1 to Injection type: Split injection; Split ratio is about 10:1 Comment Summary #20: The commenter suggested the following changes on System suitability. Bold text reflects commenter recommendations. Note above chart at top of the page should read: o NOTE-See the relative retention times in table below o Title for second column in chart should read Relative Retention Time (Approximately) Response: Comment incorporated. The Expert committee changed the Note to read as [NOTE Diethylene glycol elutes after ethylene glycol in the chromatogram.] Comment Summary #21: The commenter indicated they had questions on the Resolution under Suitability Requirements and whether or not the Resolution of NLT 30 is too large or should it be 3.0 or smaller? Response: Comment not incorporated. Please refer to response to Comment Summary #4. Comment Summary #22: The commenter suggested the following changes on Acceptance criteria. Bold text reflects commenter recommendations. 1. The areas of Ethylene Glycol or Diethylene Glycol peaks in the Sample solution is NMT the peak areas of Ethylene Glycol or Diethylene Glycol in the Standard solution, respectively, corresponding to NMT 0.10% ethylene glycol and NMT 0.10% diethylene glycol in Sorbitol Solution. 2. The commenter also commented that Standard solution is low, extract concentration is very low, possibly needs to be higher. The commenter suggested that it may need an internal standard to demonstrate the formula for calculations. Response: For comment 1, the Expert Committee recommended the creation of separate acceptance criteria for Ethylene Glycol and Diethylene Glycol in order to clarify the Acceptance criteria as follows: Acceptance criteria: Diethylene Glycol: The peak area of Diethylene Glycol in the Sample Solution is not more than the peak area of Diethylene Glycol in the Standard solution, corresponding to not more than 0.10% of Diethylene Glycol in Sorbitol Solution. Ethylene Glycol: The peak area of Ethylene Glycol in the Sample Solution is not more than the peak area of Ethylene Glycol in the Standard solution, corresponding to not more than 0.10% of Ethylene Glycol in Sorbitol Solution. For comment 2, the Expert Committee will consider revising Standard solution in the future if the necessary supporting data is submitted by the commenter. The Expert Page 6

Committee determined that extensive method development and validation studies as well as public reports demonstrated that Standard solution and extract concentration in current proposed methods for Sorbitol Solution are appropriate for detection with any basic GC instrument via direct injection. However, USP General Notices also states that When a specified concentration is called for in a procedure, a solution of other normality or molarity may be used, provided that allowance is made for the difference in concentration and that the change does not increase the error of measurement. Extensive method development and validation studies also clearly exhibited that the method works well using external standards thus an internal standard is not necessary for this limit test. The Expert Committee suggests that the commenter refer to 10-Aug meeting presentation slides showing data for Specificity, Accuracy, Precision, LOD of the methods etc posted on USP website. http://www.usp.org/pdf/en/hottopics/ppgwebmeetingaug2009.pdf Expert Committee-initiated Change #1: The Expert Committee emphasized the use of USP Reference Standards to revise the standard preparation In the Standard solution section, to read as below. Standard solution: 0.08 mg/ml of USP Diethylene Glycol RS and 0.08 mg/ml of USP Ethylene Glycol RS in Diluent. Expert Committee-initiated Change #2: The Expert Committee checked an on-line catalog and found that that the liner used for the limit test for EG and DEG in Sorbitol containing solutions did not have "low pressure" as a part of its description. The Expert Committee removed "low pressure" from the description of the liner in the proposed monographs. POSITIVE FEEDBACK Monograph/Section(s): Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution/Identification C: Limit of EG and DEG Expert Committee(s): Excipient Monographs 1 No. of Commenters: 4 Comment Summary #1: Pleased to find that the method works. The precipitation sample preparation approach was a clever addition to the GC measurement system. Nice job! Comment Summary #2: Procedure seems acceptable as EG and DEG limit test (NMT 0.1%). Comment Summary #3: No modifications to the procedure seem necessary. It is a very simple and straightforward procedure. Comment Summary #4: EG and DEG are easily separated by gas chromatography. This is likely due to the significant difference in their boiling points (EG at 197, and DEG at 245 ) Page 7

Comment Summary #5: The GC parameters and temperature profile are very simple and the run time is short. The time between injections is approximately 20 minutes. Any basic GC instrument with direct injection should work fine for this type of analysis. Comment Summary #6: Sample preparation is quick and easy. A polyol syrup (2.0 g) is vortexed with increasing amounts of Diluent (96:4 acetone/water) in a 25 ml volumetric flask. The sample is syringe-filtered into a GC vial prior to injection. It takes about 10 minutes to prepare a GC sample for analysis. Comment Summary #7: During the open microphone web meeting, USP stated it will be adding clarification to the Sorbitol Solution NF monograph to indicate that the Sorbitol is supposed to precipitate out after addition of the organic dilution (acetone). Our laboratory also noted this during their testing. We fully endorse USP adding this clarification to the monograph. Comment Summary #8: One large polyol manufacturer reported to USP that based on comments they received from several companies with which they interacted, they confirmed with USP that the method proposed by USP seems to be satisfactory and easy to use. Page 8